    Pneumatic Comparators (Metrology)

    5.8.

    Air gauging has rapidly increased during some past time due to the following importantcharacteristics:(a) Very high amplifications are possible. It can be used to measure diameters, length,squareness, parallelism, concentricity, taper, centre distance between holes and othergeometric conditions.(b) As no physical contact is made either with the setting gauge or the part beingmeasured, there is no loss of accuracy because of gauge wear. For this reason, air spindle andair snap gauges last very long. Also very soft parts which are easily scratched, can be gauged.(c) Internal dimensions can be readily measured not only with respect to tolerance

     boundaries but also geometric form. In other words, while measuring a bore it can revealcomplete story of size, taper, straightness, camber and bell mouth etc.(d) It is independent of operator skill.(e) High pressure air gauging can be done with cleansing of the parts which helps toeliminate errors due to dirt and foreign matter.(f) Gauging pressures can be kept sufficiently low to prevent part deflection.(In general, high pressure gauges are suitable for those parts in which tolerances arerelatively large and low pressure air gauges are preferable for highly precise work.)(g) Dimensional variations throughout the length of shaft or cylinder bore can beexplored for out of roundness, taperness, concertricity, regularity and similar conditions.

    (h) Not only it measures the actual size, but it can also be used to salvage oversized pieces for rework or to sort out for selective assembly, i.e., it is suitable both for variableinspection (measurement of size) and attribute inspection (GO and NO GO) gauging and limits.(i) The total life cost of the gauging heads in much less.(j’) It is accurate, flexible, reliable, universal and speedy device for inspecting parts inmass production.(k) It is best suited for checking multiple dimensions and conditions on a part simul-taneously in least possible time. It can be used for parts from 0.5 mm to 900 mm diameterhaving tolerance of 0.05 mm or less. It can be easily used for on line measurement of parts asthey are being machined and take corrective actions.5.8.1.

    Systems of Pneumatic Gauges

    . Based on the physical phenomena on whichthe operation of pneumatic gauges is based, these may be classified as :(i) Flow or velocity type, (ii) Back pressure type.Flow or velocity type pneumatic gauges operate by sensing and indicating the momen-tary rate of air flow. Flow could be sensed by a glass tube with tapered bore, mounted over a

    graduated scale. Inside the bore a float is lifted by the air flow.Velocity of air in velocity type pneumatic gauges can also be sensed by sensing thevelocity differential i.e., differential pressure across a venturi chamber. Such systems have

    http://what-when-how.com/

  • 8/19/2019 Pneumatic Comparators (Metrology)

    2/17

    3/12/2016 Pneumatic Comparators (Metrology)

    http://what-when-how.com/metrology/pneumatic-comparators-metrology/ 2/17

    Fig. 5.32. Free flow air gauge. quick response. These permit use of large clearance between nozzle and object surface,

    resulting in reduced wear of the gauging members. There is less air consumption. Magnifica-tion of the order of 500 to 5000 times is possible.5.8.1.1.

    Free Flow Air Gauges (Flow or velocity type).

    In this case the compressed airafter the filtering and pressure reducing unit flows through a tapered glass tube containing asmall metal float and then through a plastic tube to the gauge head having two diametricallyopposed orifices for air escapement into atmosphere (Refer Fig. 5.32). The position of the tube

    is dependent upon the amount of air flowing through the gauge head, which in turn isdependent upon the clearance between the bore to be measured and the gauge head. Fig. 5.34shows a curve between the air flow and the clearance between the part and the orifice in gaugehead.

    Fig. 5.33. Zero and magnification adjustment in flow type pneumatic comparator. 

    Fig. 5.34. Characteristic of air flow versus clearance of flow. The flow velocity type pneumatic comparator with zero adjustment and magnificationadjustment is shown in Fig. 5.33. Magnification can be changed by passing some of the airsupply, using a screw at the inlet to the tapered glass tube. The float can be zeroed by a bleedvalve installed at the top of the tube. Size is measured by the velocity of air in a tapered glasstube which is measured by the height of the float in tube.The straight portion of the curve is utilized for the measuring range. It provides highamplification (10 : 1) and thus within the linear range, it is possible to read accurately upto

    microns depending upon scale length, or classify the sizes quickly and accurately. Theamplification can be changed by quick change of tube, float and scale. Air gauge amplificationand range are based on the tooling and instrument standards of manufacturer. The amplifica-tion and instrument are selected by considering the total tolerance spread and choosing theinstrument that covers the range. About 50 to 100 mm of column is usually allowed for the

    http://lh5.ggpht.com/_X6JnoL0U4BY/S0zNnS_7i4I/AAAAAAAAEKk/0QtNeCUWxWM/s1600-h/tmpB40_thumb1.pnghttp://lh3.ggpht.com/_X6JnoL0U4BY/S0zNjHVmfKI/AAAAAAAAEKc/c29t2LHhjd8/s1600-h/tmpB39_thumb1.pnghttp://lh3.ggpht.com/_X6JnoL0U4BY/S0zNhxf_69I/AAAAAAAAEKU/MoeFMDLjRVE/s1600-h/tmpB38_thumb1.png

  • 8/19/2019 Pneumatic Comparators (Metrology)

    3/17

    3/12/2016 Pneumatic Comparators (Metrology)

    http://what-when-how.com/metrology/pneumatic-comparators-metrology/ 3/17

    actual tolerance spread.

    In the gauging head, the air escapement orifices are recessed below its cylindricalsurface so that the orifices never contact the part being gauged. Thus the surface wear will notaffect the accuracy till it is worn down to orifice level. Also the orientation of gauge or the wayoperator holds the gauge is of no consequence and same readings will be obtained for givendiameter. On the gauge, knobs are also provided for adjusting float position and calibration.Air gauge is set by placing masters for maximum and minimum tolerances on spindlealternatively and adjusting the float position for each master by turning the knurled knobs at

    the base of the instrument.Free-flow column type gauges are usually assembled together side by side and thusmultiple interrelated readings can be seen at a glance. This is the big advantage of air gaugingthat the multiple dimensions and conditions can be inspected with great ease, accuracy andspeed.Pneumatic circuits can be arranged to determine dimensional differences like taper(comprising the diameter of bore at different points along a part), bore centre distance and alsoto select parts to assemble to predetermined clearances or interference fits.5.8.1.2.

    Back Pressure Gauges.

    The basic principle and the theory of pneumaticgauging in the back pressure gauges is described below. (Refer Fig. 5.35)Air from a constant pressure sourceflows to the atmosphere through two orificesOc and Om in series.P is the pressure upstream of the first

    orifice andp is the pressure between the twoorifices, both measured with reference tothe atmospheric pressure as datum.

    Fig. 5.35. Theory of pneumatic gauging. The relationship between p and P will depend upon the relative sizes of the two orifices :

     p being equal to P when Om is blocked and tends to zero as Om is increased indefinitely. Let C

     be the geometrical area of Oe and M that of Om.Then if p and C are kept constant while M is varied, the relationship between thedimensionless quantitiesp/P and MIC is of the type shown in Fig. 5.36.(The general form of this curve is quite well predicted by an analysis employingBernouli’s equation for flow of a compressible fluid.)We are interested in linear portion of this curve.For design purposes we follow an empirical approach which is based on an experimentalstudy at N.P.L. (London) of the relationship between pressures and orifices areas.The characteristics of p/P and MIC are determined experimentally for pressure Pvarying from 2 to 75 pounds per sq. inch (0.13 to 5 kg/cm2) and inspection of any one of theseshows that within the range 0.6p/P to 0.8p/P, the curve approximates to a straight line, the

    equation for which mav be written as

    http://lh4.ggpht.com/_X6JnoL0U4BY/S0zNoTR2fLI/AAAAAAAAEKs/CjI2-y8QZWI/s1600-h/tmpB41_thumb1.png

  • 8/19/2019 Pneumatic Comparators (Metrology)

    4/17

    3/12/2016 Pneumatic Comparators (Metrology)

    http://what-when-how.com/metrology/pneumatic-comparators-metrology/ 4/17

    Fig. 5.36. Characteristic curve of pneumatic gauge. Examination of the family of curves shows that constant A, the intercept on the p/P axis isclosely constant over the range of pressures investigated and for practical purposes, the valueof A = 1.10 can be adopted for any value of P likely to be used.The slobe b of straight portion characteristics is however not independent of P, itsnumerical value decreases as P increases and the limiting values found in the investigationare as under :

     b = 0.6 when P = 0.13 kg/cm2 b = 0.4 when P = 5 kg/cm2.5.8.1.3.

    Area of escape orifice.

    Whenthe clearance between the surface and thenozzle face is zero, no air escapes from thenozzle and the area of the escape orifice is zero.When the clearance between the sur-face and the nozzle face is very large, the areaof the escape orifice is ^ d2, where d is the

    diameter of the nozzle.Between these extremes, especiallywhere the clearance is small and where airgauging can be employed, the area of theescape orifice is n dl, that is, the area of thecurved surface of the cylinder shown inFig. 5.38.5.8.1.4.

    Range of linear measurement.

    The condition 0.6 < p/P < 0.8 defines asection of the characteristics which ex-

     perimental investigation has shown to belinear to within 1%.

    http://lh4.ggpht.com/_X6JnoL0U4BY/S0zNrCdj-xI/AAAAAAAAEK8/h15P6bltV2Q/s1600-h/tmpB43_thumb1.pnghttp://lh6.ggpht.com/_X6JnoL0U4BY/S0zNpq_DWFI/AAAAAAAAEK0/pzQUl-KiKqA/s1600-h/tmpB42_thumb1.png

  • 8/19/2019 Pneumatic Comparators (Metrology)

    5/17

    3/12/2016 Pneumatic Comparators (Metrology)

    http://what-when-how.com/metrology/pneumatic-comparators-metrology/ 5/17

    Fig. 5.37. Experimentally determined characteristics for different operating pressures. 

    Fig. 5.38. Area of the escape orifice. 

    Fig. 5.39. Range of linear measurement. 

    5.8.1.5. Pneumatic Sensitivity, i.e. how p varies by variation of M. 

    5.8.1.6.

    Changing sensitivity (Magnification).

    A plot of pressure against escapeorifice area for a number of different sizes of control orifice will show that sensitivity increasesas the diameter of the control orifice decreases, i.e., for small control orifices the change in

     pressure is greater for a given change in escape orifice area. Utilising this property, it is possible to set precisely the sensitivity (magnification) by incorporating a variable control

    http://lh5.ggpht.com/_X6JnoL0U4BY/S0zN2OHIcnI/AAAAAAAAELk/tp5ywHI-BZg/s1600-h/tmpB48_thumb1.pnghttp://lh5.ggpht.com/_X6JnoL0U4BY/S0zN0TrXQdI/AAAAAAAAELc/6MVHTPQG8tI/s1600-h/tmpB47_thumb1.pnghttp://lh5.ggpht.com/_X6JnoL0U4BY/S0zNwAwg4QI/AAAAAAAAELU/p2rLizJAKxo/s1600-h/tmpB46_thumb1.pnghttp://lh5.ggpht.com/_X6JnoL0U4BY/S0zNulpCYOI/AAAAAAAAELM/O1nMRc_7swY/s1600-h/tmpB45_thumb1.pnghttp://lh6.ggpht.com/_X6JnoL0U4BY/S0zNs8c-BfI/AAAAAAAAELE/aQV1IuruNi4/s1600-h/tmpB44_thumb1.png

  • 8/19/2019 Pneumatic Comparators (Metrology)

    6/17

    3/12/2016 Pneumatic Comparators (Metrology)

    http://what-when-how.com/metrology/pneumatic-comparators-metrology/ 6/17

    orifice.5.8.1.7.

    Overall Magnification.

    In a practical pneumatic measuring apparatus thearea M will be associated with the measuring head and change in M will be the result of achange in the dimension which is

     being measured e.g. a change in theseparation L between nozzle andsurface (Fig. 5.40).The overall magnification ofthe apparatus, i.e. the ratio of thelinear movement of the pointer orindex of the pressure measuring in-strument to the change in the

    Fig. 5.40. Variation of M w.r.t. L. 

    This condition requires the measuring head to be correctly designed. The final escape-ment of the air from the nozzle to the atmosphere is taken as being through an area of thecurved surface of the cylinder of length L and diameter D, where L is the separation betweenthe nozzle surface and the surface to be gauged and D is the internal diameter of the nozzle.

    5.8.1.8.

    Response Speed.

    For a back pressure system the speed of response is not as

    fast as for free-flow type, because some time is required for the pressure to build up. The speedof response becomes of concern when the gauginghead is separated from indicating instrument by long distance.A pneumatic measuring system will not correctly measure displacements of frequencygreater than about 2 cycles/second, because of its slow speed of response. The response is

    http://lh4.ggpht.com/_X6JnoL0U4BY/S0zOESlqjqI/AAAAAAAAEMY/F9uIOgNVIzo/s1600-h/tmpB53_thumb1.pnghttp://lh4.ggpht.com/_X6JnoL0U4BY/S0zOC5vgjII/AAAAAAAAEMQ/Eb4eslmuw6Y/s1600-h/tmpB52_thumb1.pnghttp://lh6.ggpht.com/_X6JnoL0U4BY/S0zN_G1xZ_I/AAAAAAAAEME/CgZ1viGW-H8/s1600-h/tmpB51_thumb1.pnghttp://lh4.ggpht.com/_X6JnoL0U4BY/S0zN41w4hNI/AAAAAAAAEL0/sDKe8Ug3fKw/s1600-h/tmpB50_thumb1.pnghttp://lh5.ggpht.com/_X6JnoL0U4BY/S0zN3maG_wI/AAAAAAAAELs/IMeelHBle4Q/s1600-h/tmpB49_thumb1.png

  • 8/19/2019 Pneumatic Comparators (Metrology)

    7/17

    3/12/2016 Pneumatic Comparators (Metrology)

    http://what-when-how.com/metrology/pneumatic-comparators-metrology/ 7/17

    considerably slower compared to the electrical system because of the following reasons.Between the control orifice and the measuring head, there exists a closed volume associatedwith the measuring instrument used to measure the pressure p. A dimensional change, e.g., adisplacement of the surface alters the flow so changing the pressure p. The time needed toestablish the new value of p depends on the total volume and on the rate of air flow into andout of it. The latter in turn depends upon the operating pressure P and the size of the controlorifice and orifice in the measuring head. These orifices are related in size and determine the

     pneumatic sensitivity, the smaller orifices (corresponding to the higher magnification), having

    a more restrictive effect on the air flow and so slowing the response.It has been examined theoretically and experimentally that response is slowed by usinga large operating pressure P and a large volume, but by these high sensitivity (magnification)is obtained. In any practical pneumatic measuring system, the overall response will beinfluenced by dynamic characteristics of the pressure measuring device. Thus it follows thatthe use of a low operating pressure will not improve the overall response if a low pressuremeasuring device of slow response is used to measure the pressure changes. High sensitivitywill inevitably be associated with slow response and the only factor left to the designer is thevolume, which should be made as small as possible for quick response.Since gauge is always located at some distance from the control unit, the effect of

    variations in the gauging position does not reach the control unit instantaneously, though thesize variations of the object will promptly affect the air flow at nozzles. The time gap betweenthe sensing and indication is known as response time which depends upon : Length of air line

     between the nozzles and indicator, (ii) type of gauge system, and (Hi) the design of control unit.Response in case of flow type pneumatic gauges is relatively quick. Response in case of back

     pressure gauges is slow, the compressibility of air also contributing to the delayed transmissionof the variations sensed at the nozzles.Response time of back pressure type pneumatic gauges can be improved by utilisingfollowing devices :(i) Using filled system pressure gauge, thereby reducing volume of air.

    (ii) Restricting the unimpeded escape of air through the orifices when the gauging headis not in operation, by using a spring-charged cover sleeve around the gauge head.(iii) Counteracting the unrestricted air escape by an auxiliary air supply relay whoseoperation automatically discontinues as soon as a specific back pressure develops during theactual gauging process.(iv) Using a high speed relay to compensate for additions to the volume of the instrumentsystem.5.8.1.9.

    Zero Setting.

    It is accomplished by means of a bleed valve and consists inadjusting the indicating element of the gauge to that marking on the scale which was selectedto signal coincident with the nominal limit size represented by a setting master.5.8.1.10.

    Datum Control.

    If means be provided to change the pressure in the cavity(between control orifice and measuring orifice) using a variable bleed to atmosphere, a datumor zero can be provided which varies the pressure ‘p’ when the escape orifice area remainsfixed. This addition to a circuit provides means of accommodating small differences whichinevitably occur in the manufacture of gauge heads. Limited use of a datum control in the formof a bleed to atmosphere has an insignificant effect on linearity.However, this system depends highly upon the pressure regulator to maintain the

  • 8/19/2019 Pneumatic Comparators (Metrology)

    8/17

    3/12/2016 Pneumatic Comparators (Metrology)

    http://what-when-how.com/metrology/pneumatic-comparators-metrology/ 8/17

    supply pressure within very close pressure limits. Thus the pressure regulator is a criticalcomponent in this circuit. This problem is overcome in differential back pressure circuit in

    Fig. 5.41. Practical back pressure circuit. 

    which accuracy is maintained regardless of some variation in the regulator performance whichcontrols supply pressure. Fig. 5.42 shows a practical flow-responsive system which is incommon use. This system employs variable orifices for sensitivity and datum control.

    Fig. 5.42. Practical flow responsive system. 

    Fig. 5.43. Plug for pneumatic comparator. 5.8.1.11. Am

    plification Adjustment.

    This permits different range of gauge indica-tions on same scale length and is carried out with a precision valve of the control unit. Bothzero setting and amplification adjustment should be checked from time to time depending uponexperience.5.8.1.12.

    Jet Recession.

    It has already been seen that when the surface being measuredis very close to the nozzle face, equal increments of change in clearance do not produce equal

    increments of pressure change. The system is not linear under these conditions.Because of this the faces of the jets on air plug gauges are ground below the bodydiameter of the plug as shown Fig. 5.43. This grinding back is called jet recession and it is themeans by which the non-linear portion at the high pressure (low flow), end of the pres-sure/clearance curves is avoided.

    http://lh5.ggpht.com/_X6JnoL0U4BY/S0zOUQUXg3I/AAAAAAAAEMw/AbZ6G-wk9RM/s1600-h/tmpB56_thumb1.pnghttp://lh5.ggpht.com/_X6JnoL0U4BY/S0zOK73VgfI/AAAAAAAAEMo/qZqKBSaiEWY/s1600-h/tmpB55_thumb1.pnghttp://lh3.ggpht.com/_X6JnoL0U4BY/S0zOF0jn75I/AAAAAAAAEMg/xpaSEwySOJo/s1600-h/tmpB54_thumb1.png

  • 8/19/2019 Pneumatic Comparators (Metrology)

    9/17

    3/12/2016 Pneumatic Comparators (Metrology)

    http://what-when-how.com/metrology/pneumatic-comparators-metrology/ 9/17

    5.8.1.13.

    Measurement of Bore. [Refer Fig. 5.44 (a) and (6)]

    In the above discussion we have as-sumed a jet separation of the nozzle from theflat surface. This is also applicable to measure-ment of bore, where L is modified as :L = Lx + L2, Lx – Radial gap for one jetand L2 = Radial gap for other jet.Theoretically Li should be equal to L2,

     but even if L\ * L2 the back pressure would beclosely equal to that corresponding to the con-dition L1 = L2 = L/2. This is a big advantage asoperator need not be very meticulous aboutorientation of the measuring head in the bore,and thus the readings from operator tooperator will be uniform as they do not depend

    upon a high degree of operator skill or sense offeel.5.8.1.14.

    Measuring Heads Measuring heads fall under two categories,

    viz. direct head [Fig. 5.45 (a) and(6)1, and indirect or contact head [Fig. 5.45 (c) and (d)]. Tapered nose type direct head [Fig.5.45 (a)] is quite popular as it permits easy

    access in constricted measuring conditions.The ratio of land diameter (overall diameter attapered small end) compared to jet diameter istwice in size. Bigger ratio would affect the escapement of air and the characteristics of thesystem.Head at Fig. 5.45 (b) provides good protection to the nozzle due to incorporation of guardring and escapement holes. In the case of indirect measuring heads, the jet is protected fromaccidental damage. The size of air escapement is controlled by a needle valve or flat plate [Fig.5.45 (c) and (d) respectively] which move due to movement of measuring plunger. By changingthe taper of needle valve, the range of measurement can be changed. Parabolic needle provides

    linear response. Fig. 5.45 (e) shows the plug gauge used for measurement of diameter, lobing,taper etc. The measuring side is made somewhat smaller than the bore so that it enters muchmore conveniently.5.8.2.

    Solex Pneumatic Gauge. (Fig. 5.46).

    This instrument is produced commercially by Solex Air Gauges Ltd. This is generally designed for internal measurement, but withsuitable measuring head it can be used for external gauging also.

    It is obvious from the equation for sensitivity that in order that sensitivity (magnifica-tion) remains constant, the supply pressure P must be constant.Thus it is necessary to have a simple pressure regulator which may control the pressureof air from the normal supply line, and if necessary the pressure might be reduced also. Thearrangement used in Solex gauge is to pass the high pressure air after filtering, through a flow

  • 8/19/2019 Pneumatic Comparators (Metrology)

    10/17

    3/12/2016 Pneumatic Comparators (Metrology)

    http://what-when-how.com/metrology/pneumatic-comparators-metrology/ 10/17

    valve. There is a tank in which water is filled upto a certain level and a dip tube is immersedinto it upto a depth corresponding to air pressure required. In Fig. 5.46, it is represented byH. Since air is sent at higher pressure than required one, some air will leak out from the dig

    tube and bubble out of water and the airmoving towards control orifice will be atdesired constant pressure H.

     Now-a-days diaphragm type pressure regulators are readily avail-able in the market and they are betterfor regulating the pressure than theabove device. The air at reduced pres-sure then passes through the controlorifice and escapes from the measuring

     jets. The back pressure in the circuit isindicated by the head of water displaced

    in the manometer tube. The tube isgraduated linearly to show changes in

     pressure resulting from changes in in-ternal diameter of the work measured.This instrument is capable of measur-ing to the accuracy of microns.It is very obvious from Fig. 5.46that the diameter being measured atany instant is corresponding to the por-tion against two jets. Now to find the

    concentricity (roundness of any job atany section), the workpiece may berevolved around measuring gauge. If nochange in reading is there, then it is

     perfectly round hole. Similarly thediameter can be noted down at several

     places along the length of bore and thustapering of hole is determined. Thismethod is, therefore, best suited formeasuring roundness and taperness ofcylinder bores and gun barrel bores.By having suitable measuringhead this can be used for external gaug-ing, and head in this case will be as

    http://lh4.ggpht.com/_X6JnoL0U4BY/S0zOZrEt6wI/AAAAAAAAEM4/wLoyCAnMB4U/s1600-h/tmpB57_thumb1.png

  • 8/19/2019 Pneumatic Comparators (Metrology)

    11/17

    3/12/2016 Pneumatic Comparators (Metrology)

    http://what-when-how.com/metrology/pneumatic-comparators-metrology/ 11/17

    (a), (6) Direct heads (c) Forward head (d) Reverse head (e) Plug gauge. Fig. 5.45. Measuring heads. 

    Fig. 5.46. Solex Air Gauge. shown in Fig. 5.47. This can best reveal any lobing effectalso. It is also possible to have arrangement to measurethe length of slip gauge by having the flat table and one

     jet at the top.5.8.2.1.

    Overall Magnification and Range.

    From equation dpIdM = 0.40 P/Mc, the pneumatic sen-sitivity of a pneumatic measuring apparatus can beincreased by increasing the operating pressure P, but

    controlled by the length of scale of pressure measuringinstrument corresponding to pressure changes from 0 toP. If this scale length is to remain of convenient mag-nitude, increasing the operating pressure is not a

    suitable method for improving the overall magnification. The only effective method forobtaining the higher magnification is, therefore, to reduce average separation between nozzleand surface, which at the same time, of course, reduces the range of linear measurement.5.8.2.2.

    Limitations of empirical approach.

    From the view-point of air flow, theeffective area of an orifice is not usually identical with its geometrical area. If two orifices aremade by producing holes of identical geometrical area in two thin discs, their effective areas

    may be appreciably different as a result of edge effects on the air flow arising at the peripheriesof the orifices. Again the relationship between effective area and geometrical area is unlikelyto be the same for air flow through an orifice and the jet of air from a nozzle. In the experimentaldetermination of the p/P, MIC characteristics the value of M and C used were the geometricalareas of orifices Om and Oc. Therefore due to effective area being different from the geometrical

    http://lh4.ggpht.com/_X6JnoL0U4BY/S0zOdhJRlnI/AAAAAAAAENQ/Gh_FbvwjKhQ/s1600-h/tmpB60_thumb1.pnghttp://lh6.ggpht.com/_X6JnoL0U4BY/S0zOcaYQN5I/AAAAAAAAENI/fzovlTFTkrc/s1600-h/tmpB59_thumb1.pnghttp://lh5.ggpht.com/_X6JnoL0U4BY/S0zOa7apzSI/AAAAAAAAENA/N-tBMv0C6EY/s1600-h/tmpB58_thumb1.png

  • 8/19/2019 Pneumatic Comparators (Metrology)

    12/17

    3/12/2016 Pneumatic Comparators (Metrology)

    http://what-when-how.com/metrology/pneumatic-comparators-metrology/ 12/17

    area, the empirical equation obtained by analysing these characteristics would not be expectedto provide a completely accurate numerical forecast of performance.

     Nevertheless, experience has shown that they do give a first approximation sufficientlyreliable to permit the required performance to be obtained by a single-step corrective adjust-ment of the control orifice.5.8.3.

    Differential Comparators.

    A later development brought out the balancedcircuit type of air-gauge. In this equipment a differential pressure indicating mechanism,connected across the two air-paths and a built-in gauge zeroing valve is provided. Such a

     balanced circuit is shown schematically in Fig. 5.48. An air gauge based on this balanced circuitis called ‘Differential Comparator’.

    Three equally spaced measuring orifices (jets) reveal any lobbing effect also. Fig. 5.47 

    Fig. 5.48. Differential Circuit. Compressed air from a suitable source, after passing through air-drier and filter isregulated for constant pressure by a pressure regulator. Now the air flows into two channels,each of which has a control orifice Ocl and Oc2 (control orifices are also called master jets).From the control orifice Ocl, air flows to the measuring head where it meets further restriction

    of the workpiece or the master settings. The restriction of the workpiece builds up back pressure as explained earlier. At the same time, other half of the air is flowing through theother control orifice Oc2 to the reference jet Om. By closing or opening the valve of reference jetOm, the pressure in the space between Oc2 and Om is regulated (adjusted) to match the back

     pressure from the measuring jets, which is sensed by the pressure indicating device fittedacross the two channels as shown. At this adjustment of the reference jet, the pressure indicatorwould indicate equal pressure in the two channels and hence read zero on the scale. This zerosetting (adjusting of reference jet Om) is done with master workpiece whose dimension is exactnominal size.

     Now the variation of the dimension at the measuring head would cause change of back-pressure in channel A. This pressure would be different from the mean pressure which

    has been already set in the channel B (by reference jet). Now the difference of pressure of thetwo channels would be indicated by the pressure indicating device which can be directlycalibrated in terms of variation of dimension from the mean dimensions. Hence the instrument

     based on the measurement of differential pressure is called Differential Comparator.If the dimension causes a decrease in gap L as compared to La, this in turn decreases

    http://lh5.ggpht.com/_X6JnoL0U4BY/S0zOgAAU19I/AAAAAAAAENg/y3Xz5aymnn0/s1600-h/tmpB62_thumb1.pnghttp://lh3.ggpht.com/_X6JnoL0U4BY/S0zOe6ZrXqI/AAAAAAAAENY/8v_lGAnlUyI/s1600-h/tmpB61_thumb1.png

  • 8/19/2019 Pneumatic Comparators (Metrology)

    13/17

    3/12/2016 Pneumatic Comparators (Metrology)

    http://what-when-how.com/metrology/pneumatic-comparators-metrology/ 13/17

    M and hence increasing back pressure in channel A and vice versa. In these cases the pressureindicator would show readings on both sides of zero corresponding to ± deviation from mastersetting.5.8.3.1.

    Advantages of Differential Circuit over Single Channel Circuit,

    (i) Effectof change of operating pressure P. The operating pressure may vary slightly from the designedvalue. It can be shown that error due to change of pressure would be 0.6 to 0.8 times the changein pressure in single channel in case of differential circuit the error would be 0.1 times thechange in pressure.(ii) Zero setting of master gauge is an extra advantage.(iii) Rectification for control orifice. In a single channel system, the practical limitationsmay not give the perfectly correct and accurate dimension of the control orifice as designed.Sometimes it may go out of the useful range of the design and it may have to be discarded.Therefore, in order to avoid the error of manufacture and also to take into account thefact that the geometrical area is different from the effective area, we need a needle valve so

    that area may be adjusted accordingly.But in the differential circuit which incorporates a needle zero adjusting valve, theoff-set of the actual size of the control orifice from the designed value can be nullified byadjusting this valve.5.8.4.

    Non contact and Contact Type Pneumatic Gauging Elements.

     Non contacttooling is best suited for automatic gauging applications because of the advantages of no

    contact, clearing of oil or foreign particles from gauging area, etc.In the case of non-contact air gauge tooling, only the air coming out of the air escapementorifice touches the part to be measured, the air flow rate depending on the cross-sectional areaof the jet and the clearance between the jet and the part to be measured. It may have a single

     jet, two diametrically opposite jets or more evenly spaced jets. Single open-jet tooling can beused for checking outside diameter, height, depth, straightness, squareness, etc. and Fig. 5.49shows a few of such applications. Dual jet techniques can be applied for determining truediameter, out or round, bell mouth, thick-ness, etc. The various gauges may bedesigned either for presenting the gauges

    to part or vice versa.Many modern mechanical as-semblies demand that holes should beclosely controlled for straightness as wellas diameter. An air plug gauge forgauging hole straightness is shown inFig. 5.50.In the contact tooling, a mechani-cal member is incorporated between theair escapement orifice and the part to bemeasured. The air flow from the jet chan-ges due to displacement of this mechani-cal member when it contacts the part.The mechanical member could be a ball,lever, plunger or blade. A big advantage

  • 8/19/2019 Pneumatic Comparators (Metrology)

    14/17

    3/12/2016 Pneumatic Comparators (Metrology)

    http://what-when-how.com/metrology/pneumatic-comparators-metrology/ 14/17

    of contact type air gauge is that a much bigger measuring range (upto 2.5 mm) is possible i.e. it is suitable for wide rangeof gauging. Another advantage is that iteliminates surface roughness fromsize. It may be mentioned that open

     jet type method would be subject toerror for rough surfaces because it

    measures a combination of size andsurface finish ; further its range ofmeasurement is limited. Ball jetspindle gives a point reading ratherthan the average over a small areaand is best suited for gauging insidediameter of soft or porous parts andfor rough bores. Leaf jet spindle can

     be used for checking laminated bores, blind holes in which keyways etc. do not permit the useof open jet spindles at extreme bottom of blind holes etc. Blade jet spindles are used for

    inspecting gun bores in which oil grooves, or slots do not permit the use of ball jet or leaf jetspindles.Fig. 5.51 shows a small plunger type air gauging cartridge which is highly efficientsize-sensing element for wide range of gauging, tooling, fixturing, and machine controlapplications. It essentially consists of a spring loaded plunger. The spring tries to keep the

    Fig. 5.49. Applications of single open jet tooling. 

    Fig. 5.50. Straightness measuring gauge. 

    Fig. 5.51. Air Cartridge. 

     plunger outwards and when the part to be measured comes in contact with it or it comes incontact with the part it moves in and at the end restricts the orifice, thereby increasing the

     back pressure. The maximum and minimum limits of the plunger movement can be set withthe help of masters. Such cartridges can be secured in gauging position on various types offixtures and used for measurements like height, depth, flatness, concentricity, squareness,

    http://lh4.ggpht.com/_X6JnoL0U4BY/S0zOkqe9AJI/AAAAAAAAEN4/8J90FeIhCLQ/s1600-h/tmpB65_thumb1.pnghttp://lh5.ggpht.com/_X6JnoL0U4BY/S0zOi_HJXmI/AAAAAAAAENw/BQnzmkCOhDc/s1600-h/tmpB64_thumb1.pnghttp://lh5.ggpht.com/_X6JnoL0U4BY/S0zOhnouRcI/AAAAAAAAENo/AUAH_kxAF6c/s1600-h/tmpB63_thumb1.png

  • 8/19/2019 Pneumatic Comparators (Metrology)

    15/17

    3/12/2016 Pneumatic Comparators (Metrology)

    http://what-when-how.com/metrology/pneumatic-comparators-metrology/ 15/17

    inside/outside diameter, etc.Fig. 5.52 shows the same principle used in a test indicator which is very efficient forseveral applications. It has a tiny stylus capable of entering into small holes, keyways, slots

    Fig. 5.52. Pneumatic test indicator. 

    etc. and its movement causes the tapered end to act as a precision valve stem to regulate theamount of air flowing through an orifice. It is free of hysteresis or lag or drag in indicationswhen the stylus is moved in any direction across the workpiece.The contact type adjustable spindle kits and adjustable air snap gauges available inmarket are found to be very suitable for handling new designs, altered dimensions and variousother varied applications.5.8.5.

    Indirect Pneumatic Gauging Devices.

    The open jet has the disadvantage ofsmall measuring range. It can be overcome by using indirect pneumatic gauging devices byusing a gauging cartridge. Such cartridge employs a contacting stylus, the inner end of whichis tapered and forms the restriction in an escape orifice. The position of the stylus andconsequently the position of the taper in the orifice causes changes in the area of the orifice.Changes in the rate of taper change the measuring range of the cartridge. Measuring rangesupto 3 mm can be obtained with this type of cartridge.5.8.6.

    Air gauging with electronic sensors.

    Air gauging system operate on eitherlow or high air pressure. While low pressure systems have greater sensitivity, quicker responsetime and minimal distortion when measuring flexible components, the high pressure systemsare self cleaning type and have a large measuring range. Basically air gauging comprises air

     jet gauges—such as ring or plug, and air operated liquid columns for multi-dimensionalmeasurement.

     Now-a-days electronic flowmeters are used in place of air operated liquid columns. Theyhave the advantage of measuring flow of air with the added benefit of electronic display. Suchinstruments can easily have 2 or 3-range selection to give an extra magnification factor.

    Tolerance limit lights can be incorporated to indicate whether parts are inside or outsidemanufacturing tolerances. Response time increases many folds.The versatility of air gauging is enhanced by the wide range of measuring tools like 2and 3 jet non-contact air plug gauges, setting rings and air jet ring gauges.5.8.7.

    Multi-gauging Systems.

    Multi-gauging systems are used to measure a numberof dimensions simultaneously. Parts to be gauged are compared with a setting master which

    simulates the component. The features gauged could be external/internal diameters, lengths,straightness, squareness, ovality, run-out of faces, etc. The measuring head gauging fixture isspecifically designed to suit the component to be measured and may be completely special orit may be built using a series of modular elements. It contains the means for sensing thedimensional difference between the components and the master which may take the form of

    http://lh6.ggpht.com/_X6JnoL0U4BY/S0zOmWEMk_I/AAAAAAAAEOA/2oGaYUCsBlk/s1600-h/tmpB66_thumb1.png

  • 8/19/2019 Pneumatic Comparators (Metrology)

    16/17

    3/12/2016 Pneumatic Comparators (Metrology)

    http://what-when-how.com/metrology/pneumatic-comparators-metrology/ 16/17

    Custom Search

    mechanical or electronic probes or air jets connected to the means for amplifying the difference.The amplifying and display of readings may take the form of dial gauges or some formof electronic or air/electronic system.Display may be analogue, digital or graphic and may be augmented by out of toleranceindication.Using electronic differential methods, the relationship between different features can

     be related to a common datum.The choice of system depends on number of factors like initial cost, dimensional

    tolerances of the features to be measured, complexity of the component, complexity of thefeatures to be measured, number of features to be measured, speed of measurement required,skill or otherwise of the user.The reasons affecting the choice of displays are given below :(i) Pointer displays—best where a rapid check of run-out or concentricity is required.(ii) Columns—natural choice where a considerable number of dimensions are involved.These are fastest and most convenient form of displaying the readings for everydimension.(iii) Digital—provides high accuracy over a longer measuring range and best for situa-tions where it is required to work in drawing dimensions. Can be viewed without strain over

    longer distances.(iv) Graphic on VDU—the most sophisticated display essential when statistical processcontrol systems are employed.Systems capable of dealing with very large number of inter-related dimensions can bedeveloped.Automatic inspection machines incorporate both automated loading hopper, magazine,

     pick and place robot and—automated segregation of inspected components. Automatic inspec-tion is essential where the complexity of the component is such that manual methods can notachieve the desired levels of accuracy.The results of inspection could be fed to electronic computer based system which may

    also control the machine operation. The use of such computer based processing also allows theresults obtained to provide a wide range of control facilities including feedback for control ofthe manufacturing process.Coordinate measuring machines accommodate multidimensional inspection by using asingle point contact to take successive measurements over the component profile. The contactmovement and processing of the dimensional information is under computer control, whichcan also provide similar facilities to those offered by multi-point gauging.

     Next post: Fluid Displacement Comparators [Fig. 5.53] (Metrology)

    Previous post: Electrical and Electronic Comparators (Metrology)

    Related Links

    MetrologyWhat is MetrologyPhysical Measurement (Metrology)

     Need of Inspection (Metrology)Measuring Instruments (Metrology)Selection of Instruments (Metrology)

    :: Search WWH ::

      Search

    http://what-when-how.com/metrology/selection-of-instruments-metrology/http://what-when-how.com/metrology/measuring-instruments-metrology/http://what-when-how.com/metrology/need-of-inspection-metrology/http://what-when-how.com/metrology/physical-measurement-metrology/http://what-when-how.com/metrology/what-is-metrology/http://what-when-how.com/category/metrology/http://what-when-how.com/metrology/electrical-and-electronic-comparators-metrology/http://what-when-how.com/metrology/fluid-displacement-comparators-fig-5-53-metrology/

  • 8/19/2019 Pneumatic Comparators (Metrology)

    17/17

    3/12/2016 Pneumatic Comparators (Metrology)

    Help Unprivileged Children  ¶ Careers  ¶ Privacy Statement  ¶ Copyright Information

    http://what-when-how.com/copyright-information/http://what-when-how.com/privacy-statement/http://what-when-how.com/careers/http://what-when-how.com/help-unprivileged-children/