Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS...

103
Physics of Ultrasound Ultrasound Imaging and Artifacts รศ.นพ.เดโช จักราพานิชกุล สาขาหทัยวิทยา, ภาควิชาอายุรศาสตร์ คณะแพทยศาสตร์ศิริราชพยาบาล

Transcript of Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS...

Page 1: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Physics of UltrasoundUltrasound Imaging and Artifacts

รศ.นพ.เดโช จกัราพานิชกุลสาขาหทยัวทิยา, ภาควชิาอายรุศาสตร์ คณะแพทยศาสตรศ์ริริาชพยาบาล

Page 2: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Diagnosis‐TTE‐ TEE‐ ICE‐3D‐ 4D 

Evaluation ofCardiac Anatomy

HemodynamicAssessment

Multidimensional Echocardiography

Page 3: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Ultrasound Physics

Page 4: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Wave Motion VS Circular Motion

Page 5: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Ultrasound Waves

•Diagnostic medical ultrasound typically uses transducers with a frequency between 1-20 MHz

• Humans can hear sound waves with frequencies between 20 Hz and 20 KHz• Adult echocardiogram: 2-4 MHz• Pediatric echocardiogram: 4-8 MHz

Page 6: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Frequency

Frequency (f) time = number of cyclesFrequency (f) 1 period = 1 cycle

One hertz (Hz) = One cycle per second

Frequency (f) = 1period

Time

Period ()1 cycle

Page 7: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Wavelength

Wavelength (λ) times frequency (f) equals the propagation velocity (c):

c f

1.54 (mm) f (MHz)

Propagation velocity in the heart is 1540 m/s, the wavelength for any transducer frequency can be calculated as

Page 8: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers
Page 9: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Interaction with Tissue:Speed of Sound

• As sound travels through tissue it compresses and expands the tissue. • Where the tissue is compressed, the speed of sound is higher.

Compressed(high pressure)

ExpandedCompressed(high pressure)

Expanded

Distance

Page 10: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Nonlinear Propagation

• As the wave passes through tissue, the top of the waveform gets pulled forward to be non‐sinusoidal shape.• Propagation in which speed depends on pressure and the wave shape changes is called “nonlinear 

propagation”. • Harmonic frequency is generated from this nonlinear propagation.

Compressed(high pressure) ExpandedCompressed

(hight pressure)Expanded

Distance

Higher‐pressure portionsof the wave travel faster

Page 11: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Harmonic Imaging

Near FieldNo harmonics being generated

Signal has not traveled enough to be distorted

Near Mid FieldHarmonics Increasing

Harmonics beginning to be produced as signal travels through tissue

Mid FieldHarmonics Unchanging

Additional harmonics generated and attenuated in equal proportion

Far Mid FieldHarmonics Decreasing

Harmonics being attenuated faster than being produced

Far FieldFundamental Frequency Only

Increa

sing De

pthMaxim

um 

Harmonic Effect

Page 12: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Constructive Interference

• The waves are “in phase” with each other.• The waves reinforce each other, resulting in an “intensified” sound.

Distance

Compressed(high pressure)

ExpandedCompressed(high pressure)

Expanded

Page 13: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Destructive Interference

• The waves are “out of phase” with each other.• The waves cancel each other, resulting in a “diminished” sound.

Distance

Page 14: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Harmonic Signal• Fundamental ultrasound signals are generated from the transducer passing the fat

layer twice (transmit and receive)• Harmonic signals are generated from the tissue and transmitted to transducer

passing through the fat layer once (on receive)

Harmonic signal

Fundamental signal

Page 15: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Harmonic Imaging

Amplitude

(Velocity)f0 2f0

Fundamental frequency bandwidth

2nd Harmonic frequency bandwidth

Page 16: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Ultrasound-Tissue Interaction

Specular reflector

Moving RBC

Reflection

AttenuationRefraction

Scatter

Small structures (< 1 wavelength in lateral dimension) result in scattering

Page 17: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Biologic Effect of Ultrasound

Page 18: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Medical Ultrasound Safety:Adverse Biological Effects

•Thermal bioeffects•Heating of soft tissue and bone

•Nonthermal bioeffects•Cavitation

Page 19: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Thermal Index (TI)

•The ratio of attenuated acoustic power at a specified point to the attenuated acoustic power required to raise the temperature at that point in a specific tissue model by 1C

• TIB: Bone thermal index• TIC: Cranial-bone thermal index• TIS: Soft tissue thermal index

Page 20: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Acoustic Output Limits

•Non-ophthalmic applications• Ispta.3 720 mW/cm2

•MI 1.9•TI 6.0

•Ophthalmic applications• Ispta.3 50 mW/cm2

•MI 0.23•TI 1.0

Page 21: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Mechanical Index (MI)

•Mechanical bioeffects that occur when a certain level of output is exceed

Mechanical index (MI) = Peak negative pressureFrequency√

Page 22: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Output Display•Thermal index (TI)

•Soft tissue (TIS)•Bone (TIB)•Cranial bone (TIC)

•Mechanical index (MI)•Range 0.0 to 1.9

Page 23: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Acoustic Impedance (Z)

•Measure of how ultrasound traverses the medium and depends on• Density of the medium (p)• Propagation velocity of ultrasound through the medium (v)

Z = pv

Page 24: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Acoustic Impedance

Fluid Soft tissue

Fibrous tissue Solid (calcium)

Page 25: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Applied Ultrasound

Page 26: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Ultrasound Transducer

•Device that converts a signal in one form of energy to another form of energy•Mechanical rotating transducer (single element)•Electronic phased array transducer (multiple elements)

• Linear array• Annular array• Matrix array• Circular array• Curved array

Page 27: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Piezoelectric Effect

Piezoelectricelement

Piezoelectricelement

Inverse piezoelectric effect converts electrical energy into mechanical energy.

• Electrical voltage causes a change in length or vibration of piezoelectric material to generate a sound wave. 

Direct piezoelectric effect, also called generator or sensor effect, converts mechanical energy into electrical energy.

• Mechanical stress generates an electric charge proportional to that stress.

Page 28: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Transducer With 20 PZT Elements

Piezoelectricelements

Page 29: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Transducer

Electronic phased array

Mechanical annular array

Page 30: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Ultrasound Beam

Main lobe

Side lobe

Main lobeSide lobes

Page 31: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Ultrasound Beam

(RadioGraphics 2009; 29:1179–1189)

Page 32: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Side Lobes

•Secondary and smaller acoustic beams falling outside at predictable angle located around the main lobe

•Created by a single crystal transducer

Page 33: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Ultrasound Images

Motion mode Brightness mode

Amplitude mode

Page 34: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Ultrasound Images

Motion mode Brightness mode

Amplitude mode

Page 35: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Creating a Two-dimensional Image

• Ultrasound beam is electronically steered through a sector arc of 80°

• Speed of imaging rate of 25 frames/s.

Page 36: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Creating a Two-dimensional Image

The time needed to acquire all the data for one image frame is directly related to the number of scan lines

There is tradeoff between scan line density and image frame rate

Page 37: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Depth VS Time elapsed

Timing is proportional to distance from the transducer

Page 38: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Resolution

•The ability to distinguish two objects that are close together• Temporal resolution

• The ability to accurately locate structures or events at a particular instant in time• Spatial resolution

• Axial resolution: the ability to distinguish two objects that are close together along the axis of the ultrasound beam

• Lateral resolution: in the direction perpendicular to the beam’s axis

Page 39: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Temporal Resolution

•Dependent on frame rate •Can be improved by

• Minimizing depth - the maximum distance from the transducer as this affects the PRF

• Narrowing the sector to the area of interest - narrowing the sector angle• Minimize line density (but at the expense of lateral resolution)

Page 40: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Axial Resolution

• The ability to recognize two different objects at slightly different depths from the transducer along the axis of the ultrasound beam

where SPL = λ × no. of cycles

• It is improved by• Higher frequency (shorter wavelength) transducers but at the expense of penetration• Maximize line density (at the expense of frame rate, i.e. temporal resolution)

2

Page 41: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Lateral Resolution

Lateral resolution = FD

Page 42: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Ultrasound Imaging

Page 43: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Medical Image Orientation

Page 44: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Gain Control

•The degree of amplification of the returning ultrasound signal•Affects all parts of the image equally•Seen as a change in “brightness” of the images on the entire screen

Page 45: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Time Gain Compensation (TGC)

Depth

Signal Amplitude

Page 46: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Reject•Filters out low signal decrease noise in the image

No reject High reject

Page 47: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Dynamic Range (Compression)

•Determines the number of gray shades used to map the gray scale image on the display

• Higher more shades of gray (softer looking image)

• Lower fewer shades of gray (sharp looking image)

Page 48: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Wide Dynamic Range

Page 49: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Narrow Dynamic Range

Page 50: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

High Compression

Page 51: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Low Compression

Page 52: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Harmonic Imaging

Page 53: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Type of Doppler Study

•Pulse wave Doppler•Doppler information is generated from a small “gate” that is interrogated

•Continuous wave Doppler•One crystal constantly sends, another constantly receives•Allows evaluation of very high velocity

Page 54: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers
Page 55: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Color flow Doppler

• This is a variety of pulse wave Doppler. Multiple points in the ‘region of interest’ are analysed and colour coded rapidly.

Page 56: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Doppler Filter

•Doppler signal from tissue VS blood•Blood flow has low amplitude but high frequency•Myocardial motion has high amplitude but low frequency

•High pass filter (Wall filter)• Remove low frequency signals from the display• Display Doppler signal from blood flow only

•Low pass filter• Remove high frequency signals from the display• Display Doppler signal from tissue only

Page 57: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Doppler of Blood

Frequency

Amplitude

Tissue

Blood

(Velocity)

High‐Pass Filter

Page 58: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Doppler of Tissue

Frequency

Amplitude

Tissue

Blood

(Velocity)

Low‐Pass Filter

Page 59: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Doppler Filter

Frequency

Amplitude

Boost

Cut

Pass band Pass band

Transitional zone

Boost

Cut

High‐Pass Filter(Wall Filter)

Low‐Pass Filter

PassPass

Page 60: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

High Pass Filter (Wall Filter)

Frequency

Amplitude

Boost

Cut

Pass band

Transitional zone

Boost

Cut

High‐Pass Filter(Wall Filter)

Low‐Pass Filter

Pass

Page 61: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Low Pass Filter

Frequency

Amplitude

Boost

Cut

Pass band

Transitional zone

Low‐Pass Filter

Pass

Page 62: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Doppler Display

• The perceived returning frequency is lower than the transmitted frequency, it will be plotted below the zero baseline as a negative Doppler shift

• The spectrum displays echo amplitude by varying the brightness of the display

Page 63: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Spectral Display

Low amplitude

Frequency

Time

Page 64: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Spectral Display

Mid amplitude

Frequency

Time

Page 65: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Spectral Display

Frequency

Time

High amplitude

Page 66: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Sampling of Received Doppler Signal

Page 67: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Sampling Frequency

Page 68: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers
Page 69: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Aliasing

Page 70: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Aliasing of Doppler Signal

•Erroneous display of velocities that have exceeded the Nyquist limit• The velocity exceeds the rate at which the pulsed wave system can record it properly

•Spectral trace is cut-off at a given velocity with placement of the cut section in the opposite channel or reverse flow direction

Page 71: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Nyquist Frequency

Measurements of frequency shifts (and, thus, velocity) will be appropriately displayed only if the pulse repetition frequency (PRF) is at least twice the maximum velocity (or Doppler shift frequency) 

encountered in the sample volume

Page 72: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

The Nyquist Limit

• A sampled waveform thus needs at least two sample points per cycle. • Thus the wave's frequency must not be above half the sampling frequency. • This limit is called the Nyquist limit of a given sampling frequency

Page 73: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Basic Assumption Used in Imaging Systems

1. Sound travels in a straight line2. Sound travels directly to a reflector and back3. Sound travels in soft tissue at exactly 1540 m/s4. Reflections arise only from structure positioned in the beam’s main axis5. The imaging plane is very thin6. The strength of a reflection is related to the characteristics of the tissue

creating the reflection

(Understanding Ultrasound Physics. Sidney K Edelman, 4th ed 2012)

Page 74: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Artifacts

•Caused by violation of the assumptions•Direction

• Ultrasound path between transducer and reflector is not in straight line• Transmit path• Receive path

•Reflection•Reflection from side lobe ultrasound beam•Speed is not correct•Attenuation

Page 75: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Artifacts due to Direction of Ultrasound

Page 76: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Refraction

Page 77: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Mirror Image

Page 78: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Reverberation• A is the true anatomical structure strongly reflective of ultrasound• A1 is the artefact generated by the returning ultrasound beam being re-reflected from the

transducer and again reflected from the true anatomical structure at A.

As this doubly reflected beam has travelled twice as far it is seen at twice the distance from the transducer. It is weaker but otherwise identical to the real structure at A

A A1

Page 79: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Reverberation

Page 80: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Reverberation

Page 81: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Reverberation

Page 82: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Reverberation and Shadowing

Page 83: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Comet Tail (Ring Down Artifact)

•Reverberation with the spaces squeezed

•The reflecting surfaces are located in a medium with a very high propagation speed, such as mechanical heart valve

•Appearance of “solid hyperechoic line” directed downward

Page 84: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Comet Tail (Ring Down Artifact)

Page 85: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Comet Tail (Ring Down Artifact)

Page 86: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Ultrasound Beam

(RadioGraphics 2009; 29:1179–1189)

Main lobe

Side lobes

http://folk.ntnu.no/stoylen/strainrate/Ultrasound/#Depth_res

Page 87: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Side Lobe Artifacts

•Created by the emitted side lobes reflecting to a strong reflector, laterally positioned targets

•Erroneously displayed and interpreted by the machine as if they originated from the central ultrasound beam

•The depth of the artifact depends on time taken to and from transducer•The intensity of the artefact decreases with distance. •Common sources of the artifact are

• The atrioventricular groove• The fibrous skeleton of the heart

Page 88: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Side Lobe Artifact

Side lobeSide lobe Main

Page 89: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Side Lobe Artifact

Page 90: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Side Lobe Artifact

Page 91: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Side Lobe Artifact

Page 92: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Side Lobe Artifact

Page 93: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers
Page 94: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Artifacts due to Speed of Ultrasound

Page 95: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Range Ambiguity Artifact

•“All reflections are received by the transducer before the next pulse is transmitted.”

• If a reflection is created by a very deep structure and arrives at the transducer after the next pulse has been transmitted

•This very deep reflection will be interpreted as being created by the second pulse and is placed at a very shallow depth

Page 96: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Range Ambiguity Artifact

Page 97: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Artifacts due to Attenuation of Ultrasound

Page 98: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Acoustic Shadowing

Page 99: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Edge Shadow

Page 100: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Spit Image•If the near shadow is in the center of the probe, the ultrasound beam is splitted in two, resulting in two apparent apertures

Page 101: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Spilt imageCaused by a near shadow in the middle of the probe footprint

Page 102: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Enhancement

Page 103: Physics of Ultrasound UltrasoundImaging and Artifacts · Ultrasound Physics. Wave Motion VS Circular Motion. Ultrasound Waves •Diagnostic medical ultrasound typically uses transducers

Enhancement