Peter Burgess, Lynsey Keightley, Clare Lee, Max … · Peter Burgess, Lynsey Keightley, Clare Lee,...

77
Good Practice Guide No. 29 The Examination, Testing and Calibration of Installed Radiation Protection Instruments Peter Burgess, Lynsey Keightley, Clare Lee, Max Pottinger, Mike Renouf, David Williams. Consultation Draft Issue 2

Transcript of Peter Burgess, Lynsey Keightley, Clare Lee, Max … · Peter Burgess, Lynsey Keightley, Clare Lee,...

Good Practice Guide No. 29

The Examination, Testing and Calibration of Installed

Radiation Protection Instruments

Peter Burgess, Lynsey Keightley, Clare Lee,

Max Pottinger, Mike Renouf, David Williams.

Consultation Draft

Issue 2

CONSULTATION DRAFT

The Examination, Testing and Calibration

of Installed Radiation Protection

Instruments

This document has been prepared by a working group of the

Ionising Radiation Metrology Forum. It is intended to form the

second edition of the established NPL Guide GPG29. The

working group invite you to comment on the technical content of

this document before publication. If you wish to submit

comments, please send them to [email protected] by 17

February 2012.

Lynsey Keightley

National Physical Laboratory

Tel 020 8943 6435

ii

Measurement Good Practice Guide No. 29

The Examination, Testing and Calibration of Installed

Radiation Protection Instruments

Peter Burgess

Nuvia

Lynsey Keightley

National Physical Laboratory

Clare Lee

National Physical Laboratory

Max Pottinger

James Fisher Nuclear

Mike Renouf

Sellafield Ltd

David Williams

Magnox Ltd

iii

Queen‟s Copyright Printer and Controller of HMSO, 2011

ISSN XXXX-XXXX

National Physical Laboratory

Hampton Road, Teddington, Middlesex, TW11 0LW

Extracts from this report may be reproduced provided the source is acknowledged and the

extract is not taken out of context.

Approved on behalf of the Managing Director, NPL

by XXXX

iv

This Good Practice Guide has been written by a working party of the Ionising Radiation

Metrology Forum. Membership of the working party was as follows:

Working Group

Peter Burgess Nuvia

Lynsey Keightley National Physical Laboratory

Clare Lee National Physical Laboratory

Max Pottinger James Fisher Nuclear

Mike Renouf Sellafield Ltd

David Williams Magnox Ltd

v

Foreword

This Good Practice Guide has been written by the UK Ionising Radiation Metrology Forum1

in collaboration with the radiation user community. It describes recommended procedures

for the examination, testing and calibration of installed radiation protection instruments. Test

procedures recommended in this document are not legally binding: they are general methods

based on current accepted good practice.

The current statutory requirement for installed radiation protection instrument tests is stated

in the Ionising Radiations Regulations 1999, Regulation 19. All Employers who work with

ionising radiation must ensure that levels are adequately monitored and instruments are

suitable for this purpose.

Although the testing regimes presented here are for general application, Qualified Persons

responsible for the calibration of radiation protection instruments may modify them, with the

agreement of the Radiation Protection Adviser, as necessary to suit their particular purpose,

provided that the Employer is satisfied that the overall quality of the testing is not

compromised.

1 The Ionising Radiation Metrology Forum consists of representatives of UK establishments and organisations

actively involved in radiation measurement for protection purposes. It is the aim of the forum to facilitate the

exchange of information regarding UK calibration facilities and their efficient use by those required to comply

with these regulations.

vi

CONTENTS

Foreword

2.1 Type Tests .................................................................................................... 8

2.2 Tests Before First Use .................................................................................. 8

2.3 Periodic Tests ............................................................................................... 9

2.4 Routine Tests.............................................................................................. 10

2.5 Retest After Repair..................................................................................... 10

2.6 Analysis of Test Results ............................................................................. 11

3.1 Gamma Dose Rate Monitors ...................................................................... 14

3.2 Personnel Contamination Monitors ........................................................... 15

3.2.1 Hand and Foot Monitors ................................................................ 15

3.2.2 Frisking Monitors ........................................................................... 15

3.2.3 Personnel Exit Monitors ................................................................. 15

3.3 Portal Monitors .......................................................................................... 16

3.4 Small Article Monitors............................................................................... 16

4.1 Functional Check ....................................................................................... 25

4.2 Background Indication ............................................................................... 25

4.3 Alarm Test.................................................................................................. 25

4.3.1 Operational High Level Alarm ....................................................... 25

4.3.2 Detector Fail Alarm ........................................................................ 26

4.4 Response to High Dose Rates .................................................................... 26

4.5 Linearity of Response ................................................................................ 27

4.6 Energy Dependence of Gamma Monitors .................................................. 28

4.7 Directional Dependence ............................................................................. 29

5.1 Functional Check ....................................................................................... 31

5.2 Energy Threshold Check ............................................................................ 31

5.3 Background Indication ............................................................................... 32

5.4 Response to Contamination ....................................................................... 32

5.5 Count Rate Alarm Test .............................................................................. 33

5.6 Response to a High Activity Source .......................................................... 33

5.7 Uniformity of Response ............................................................................. 33

6.1 Functional Check ....................................................................................... 35

vii

6.2 Energy Threshold Check ............................................................................ 35

6.3 Background Indication ............................................................................... 35

6.4 Count Rate Alarm Test .............................................................................. 36

6.5 Response to Contamination ....................................................................... 36

6.6 Response to High Activity Source ............................................................. 36

6.7 Uniformity of Response ............................................................................. 36

7.1 Functional Check ....................................................................................... 39

7.2 Energy Threshold Check ............................................................................ 39

7.3 Background Indication ............................................................................... 39

7.4 Count Rate Alarm Test .............................................................................. 40

7.5 Response to Contamination ....................................................................... 40

7.6 Linearity of Response ................................................................................ 40

7.7 Response to High Activity Source ............................................................. 41

7.8 Spatial Response ........................................................................................ 41

8.1 Source Considerations and Jigs.................................................................. 43

8.2 Workplace Contamination Monitors .......................................................... 44

8.3 Workplace SAMS ...................................... Error! Bookmark not defined.

8.4 Workplace Portals ...................................... Error! Bookmark not defined.

9.1 Calibration Laboratory ............................... Error! Bookmark not defined.

9.2 Workplace Testing ..................................... Error! Bookmark not defined.

9.3 Test Label ................................................... Error! Bookmark not defined.

viii

TABLES

Table 1: Summary of Tests Before First Use and Periodic Tests ........................................... 12

Table 2: Tests Required for Gamma Dose Rate Monitors ...................................................... 18

Table 3: Tests Required fro Personnel Contamination Monitors ........................................... 20

Table 4: Tests Required for Portal Monitors .......................................................................... 21

Table 5: Tests Required for Small Articles Monitors ............................................................. 22

Table 6: List of Suitable Check Radionuclides....................................................................... 65

ILLUSTRATIONS

Figure 1: Testing Regimes for Installed Instruments ................................................................ 7

Figure 2: Hand Contamination Monitor Calibration Source .................................................. 44

Figure 3: Frisking Contamination Monitor Calibration Jig .................................................... 45

Figure 4: Exit Contamination Monitor Calibration Jig (Ladder) ............................................ 46

Figure 5: Typical Steel Walled Energy Compensated GM Detector ...................................... 62

Figure 6: Pancake GM Detector ............................................................................................. 63

1

Introduction

IN THIS CHAPTER

11

Introduction

Measurement Good Practice Guide No. 29

2

The examination and testing of radiation protection instruments is a legal requirement

for those carrying out work with ionising radiations1, 2

. Sufficient equipment must be

available to comply with the regulations and the instruments must be examined, tested

and calibrated at appropriate intervals to ensure that they remain fit for use. Periodic

examination and testing of installed equipment would normally take place in the

workplace. This minimises risk of damage caused by removal, transport and re-

installation of equipment: it also permits testing of auxiliary indicators, such as remote

warning lights.

This Good Practice Guide provides recommended procedures for the general

examination, testing and calibration of installed radiation protection instruments. The

primary purpose of such equipment is protection of personnel and includes personnel

exit monitors and frisking equipment, portal monitors, Small Articles Monitors

(SAMS) and area gamma monitors. The scope of this document does not extend to

equipment used for the monitoring of airborne radioactive particulates, this guidance is

provided in GPG823, nor does it extend to installed environmental protection

instruments. Additionally, this document does not provide guidance associated with

the maintenance and examination of engineering controls. This guidance follows a

similar format to GPG144, which provides advice for portable radiation protection

instruments. Recommendations made in documents published by national and

international organisations, including the United Kingdom Accreditation Service

(UKAS), the International Organisation for Standardisation (ISO), the International

Electrotechnical Commission (IEC) and International Atomic Energy Agency (IAEA)5

have been consulted during the preparation of this Guide.

The objective of testing is to demonstrate that the instrument is suitable and fit for use.

The testing regimes contained herein have no legal standing and Employers may

implement their own schemes, provided they ensure compliance with the relevant

regulations.

The procedures detailed in this guidance provide the minimum level of testing that is

recommended for instruments used in normal operating conditions. There may be

special cases where testing requirements will go beyond these recommendations,

where instruments are used in conditions outside those envisaged in the standards

above. In such circumstances, the Employer may need to design appropriate test

procedures.

Due to the varied nature of the instruments covered in the Guide and their

applications, it is not always possible to specify complete calibration geometries and

suitable radionuclides. In these instances it is the responsibility of the Qualified

Person, in conjunction with the Radiation Protection Advisor, to define suitable test

Measurement Good Practice Guide No. 29

3

protocols to be employed for each instrument type to suit the application for which it

is used and the environment in which it operates.

A glossary of terms is contained in Section 11.

The types of instrument that are covered by this guidance are described in detail in

Section 3.

4

Testing Regime

IN THIS CHAPTER

22

Type Tests

Tests Before First Use

Periodic Tests

Routine Tests

Retest After Repair

Analysis of Test Results

Measurement Good Practice Guide No. 29

5

For the purposes of this guidance, a test is defined as a procedure to evaluate an

instrument‟s performance in order to establish its suitability, or its continued fitness,

for a particular type(s) of measurement in operational radiation protection. A test will

involve an element of calibration, which may be defined as the measurement of the

response of the instrument to known radiation fields. It is important to recognise that

the terms test and calibration are not synonymous: this is because a test will also

involve a degree of examination, which may include, for example, an inspection of

the mechanical and electrical state of the instrument.

Type Tests are laboratory tests that establish and confirm an instrument‟s

specification. The Type Tests are normally carried out by, or on behalf of, the

instrument manufacturer.

Manufacturer‟s Production Tests confirm a tolerance in production, confirming that

each instrument conforms to type and so meets specification. These tests may be

considered as factory acceptance tests (FAT) by an employer but they must be

undertaken by or under the direct supervision of a Qualified Person on behalf of the

Employer if they are intended to satisfy Test Before First Use requirements.

The tests required for compliance with current regulations are the Tests Before First

Use (TBFU) and Periodic Tests. The findings of these tests must be compared with

any previous test information and the appropriate Type Test to confirm that the

instrument is meeting its specification and is suitable for its intended use. TBFU and

Periodic Tests are carried out by, or on behalf of, the Employer. Table 1 lists

recommended TBFU and Periodic Tests.

Commissioning of an instrument may well include TBFU and/or establishing the

Baseline for subsequent Periodic Tests. Commissioning may also include the testing

of interfaces between instruments and their remote displays or safety or warning

devices.

Routine Tests and function checks are also recommended. A Routine Test includes

testing an instrument‟s response to traceable sources, e.g. alarm level testing. A

function check is a simple test carried out to ensure that the instrument appears to be

working correctly - it does not necessarily require the use of a radioactive source. For

gamma monitors it may include observation of response to any local radiation field

and/or any detector priming source. For contamination monitors a function check is

usually carried out together with routine maintenance such as detector cleaning,

checking foils, etc. Further details on recommended function checks can be found in

Sections 4 to 7.

Measurement Good Practice Guide No. 29

6

A Test after repair is required to ensure that instrument performance has been re-

established after repair. Depending on the nature of the repair, the scope of Test after

repair may be anything from function checks through to a new TBFU to provide a new

Periodic Test baseline.

All tests should be traceable and repeatable. A full record of test results, including

details of any significant adjustments made to the instrument, should be kept for a

minimum period of 2 years.

A summary of these tests and checks is shown in Figure 1 identifying the different

ways TBFU can be achieved in practice. The diagram is intended to show a variety of

possible routines but is not considered to be exhaustive.

Measurement Good Practice Guide No. 29

7

Manu

factu

rer's r

espo

nsib

ility

/accounta

bili

ty

Em

plo

yer's r

espo

nsib

ility

/accounta

bili

ty

Test

Befo

re F

irst

Use

afte

r

transport

and insta

llation

Test

Befo

re F

irst

Use

at a

test ho

use

Tra

nsport

Insta

ll

Com

mis

sio

n

Peri

od

ic T

est

Base

line

Ro

uti

ne T

est

Test

aft

er

rep

air

Typ

e T

est

(in lab

ora

tory

)

Facto

ry A

ccepta

nce T

est

Test

Befo

re F

irst

Use

at

manu

factu

rer's p

rem

ises

Test

Befo

re F

irst

Use

in

em

plo

yer's w

ork

shop a

fter

transport

Peri

od

ic T

est

Base

line

Insta

ll

Com

mis

sio

n

Ro

uti

ne T

est

(to v

alid

ate

baselin

e)

Test

aft

er

rep

air

Measurement Good Practice Guide No. 29

8

It is the responsibility of the Employer to ensure that an instrument is suitable for the

intended use before purchase. Decisions about instrument selection should be made

taking into account advice from an RPA, information from the manufacturer and other

authoritative data that might be available.

The body of information regarding the characteristics and expected performance of

instruments is called Type Test data and is usually based on recommendations from

international organisations such as IEC, ISO, etc. A number of IEC documents exist

which detail the tests that are appropriate for the Type Testing of particular types of

instrument. Typical documents for testing installed instrumentation are BS IEC

605326 for X- and gamma-ray dose rate monitors and BS EN 61098

7 for installed

contamination monitors and portal monitors. Although no standard fully addresses

contamination frisking monitors, parts of BS EN 603258, the standard for portable

contamination monitors, may also be applied to this type of equipment. Currently there

is no standard covering SAMs.

Type Tests are very comprehensive and may require specialised facilities: the tests

should be performed by someone with appropriate expertise and insight into the use of

instruments, in a laboratory with secondary standard or similar status, using

International Commission on Radiation Units and Measurements9 specified

measurement quantities, ISO876910

specified calibration sources and ISO403711

specified radiation beams.

For most new instruments, the manufacturers or suppliers provide Type Test data that

will enable the Employer to decide the necessary scope of TBFU. In the absence of

Type Test data, other sources of information, for example, published peer reviewed

evaluations, may be useful. When Type Test data are not available, or are insufficient

in the judgement of the QP and RPA, the Employer should perform their own Type

Test to establish their own baseline data at the TBFU stage.

Assuming that the instrument is delivered in good condition and set up according to its

specifications, the TBFU should demonstrate that the instrument conforms to type and

confirm its suitability for the intended use. The practicalities of TBFU are dependent

on whether an instrument is installed before or after TBFU. A gamma monitor, frisk

probe and ratemeter or a hand and foot monitor can be tested before installation but

this may be impractical for a large installed personnel exit monitor.

Measurement Good Practice Guide No. 29

9

The TBFU should be undertaken by a Test House on behalf of the Employer. Tests

must be undertaken by or under the immediate supervision of a Qualified Person on

behalf of the Employer.

Figure 1 illustrates the stages at which the TBFU, Periodic Test and any Routine Test

should be performed, and identifies which of those test results form the baseline data

for all subsequent tests. If the results of tests performed after installation, differ

significantly to baseline data obtained before installation, this should be investigated.

Table 1 summarises the tests required for the TBFU. Recommended procedures for

each of the tests are provided in Sections 4 to 7. Some of these tests may need to be

repeated periodically as the performance of an instrument can vary with age, key

components may deteriorate or fail, and damage may occur during use; these are some

of the reasons for the subsequent Periodic Tests.

It is the responsibility of the Employer to define the frequency of Periodic Tests based

upon considerations of the age of the equipment, the environment in which it is used,

the frequency of use, etc. It is the recommendation of this guidance that examination,

testing and calibration should be performed at least annually. However, the

requirements of any regulations published in the future must be adhered to.

The purpose of Periodic Testing is to check that the performance of an instrument has

not significantly deteriorated, that it remains fit for the intended purpose and to

confirm the performance findings of the TBFU. Although it is more than just a simple

check, highly specialised facilities are not necessarily required for Periodic Testing;

the facilities should be suitable to allow measurements to be made to a known

accuracy.

Table 1 summarises the tests required for the Periodic Test. Recommended procedures

for each of the tests are provided in Sections 4 to 7.

As the lifetime of the instrument progresses, the instrument may have suffered from

wear and tear or misuse, therefore attention should be paid to the performance and

condition of its electrical and mechanical systems. Cables, connectors and detector

windows for example should be examined and any necessary repairs carried out before

the radiation response of the instrument is tested. Section 2.5 provides advice on the

scope of testing after repair.

Measurement Good Practice Guide No. 29

10

The critical role that many installed monitors play in maintaining safe working

conditions is such that a subset of the Periodic Tests should be conducted on a more

frequent basis e.g. weekly or monthly depending on the instrument type and use.

Table 1 indicates which tests should be included in Routine Tests for each instrument

type.

The Employer should be satisfied, on the basis of a risk assessment, that the frequency

of Routine Tests and the recommended subset of tests are sufficient for his own work

situation. Typical criteria to consider in the risk assessment would include the time

between breakdowns, the probability and consequences of a failure, and the occupancy

of the area which the monitor serves.

The effect of any repairs or adjustments to an instrument should be considered and

tests repeated if necessary. The scope of tests after repair can be subdivided into 3

classes:

Simple mechanical repairs such as the tightening of screws, replacement of

feet, etc. Only a function check is required.

Repairs which could influence the radiation performance but in a manner that

is easy to check, such as a detector foil replacement. A routine or periodic test

may be required.

Repairs which could have a major influence on the radiation performance, such

as the replacement of a high dose rate Geiger-Műller (GM) detector, will

generally require an initial dose rate measurement to establish the new

response factor, followed by adjustment and a repeat of the TBFU. The

baseline for future periodic tests is re-established by this TBFU.

If a repair is likely to change the radiation response of the instrument, the magnitude

of the adjustment should be recorded and reported to the Employer. This may be

achieved by recording before and after repair readings. The purpose of taking a

reading before the repair is that the Employer may have recorded measurements with

the instrument prior to the repair and calibration; if the change in response was

significant, the Employer may decide to either repeat this measurement or normalise

previous recorded measurements.

Measurement Good Practice Guide No. 29

11

In order to confirm that the instrument still conforms to type and remains fit for

purpose, the results of the TBFU should be compared with the Type Test data; these

TBFU results then form the baseline for all subsequent tests.

A full record of test results must be kept in accordance with the regulations. It is good

practice to maintain details of any adjustments made to the instrument. Current test

results should be compared with previous results and any significant changes noted

and investigated, even if all the results fall within specification. For example, the

performance of an instrument should be regarded as suspicious if a previously

consistent response is now significantly different, even if it is still within acceptable

limits.

Whenever an instrument is adjusted during the course of testing, a statement indicating

the nature and magnitude of the adjustment should be made on the test report.

An instrument may fail the TBFU or Periodic Tests if the results of any component of

the appropriate tests are not within the acceptable limits defined in Tables 2 to 5, or if

the instrument‟s performance is deemed unsatisfactory by the QP. In this way the

TBFU record is maintained as the baseline for subsequent tests.

Measurement Good Practice Guide No. 29

12

INSTRUMENT

TESTS BEFORE

FIRST USE

PERIODIC TESTS

Gamma Dose Rate Monitors

Functional Check

Background Indication

Alarm Test

Response to High Dose

Rates

Linearity

Energy Dependence

Directional Dependence

Functional Check

Background Indication

Alarm Test

Response to High Dose Rates

Personnel Contamination

Monitors

Functional Check

Energy Threshold Check

Background Indication

Response to Contamination

Count Rate Alarm Test

Response to High Activity

Source

Uniformity of Response

Functional Check

Energy Threshold Check

Background Indication

Response to Contamination

Count Rate Alarm Test

Portal Monitors

Functional Check

Energy Threshold Check

Background Indication

Count Rate Alarm Test

Response to Contamination

Response to High Activity

Uniformity of Response

Functional Check

Energy Threshold Check

Background Indication

Count Rate Alarm Test

Response to Contamination

Small Article Monitors

Functional Check

Energy Threshold Check

Background Indication

Count Rate Alarm Test

Response to Contamination

Linearity

Response to High Activity

Spatial Response

Functional Check

Energy Threshold Check

Background Indication

Count Rate Alarm Test

Response to Contamination

13

Instruments

IN THIS CHAPTER

33

Gamma Dose Rate Monitors

Personnel Contamination Monitors

Portal Monitors

Small Article Monitors

Measurement Good Practice Guide No. 29

14

Installed radiation protection instruments will normally include an audible and/or

visual alarm indication as a minimum. Some instruments may also have a readout

meter in the form of an analogue and/or digital reading. Instruments without alarm

functions, referred to as meters, are included within the following descriptions.

A gamma dose rate monitor is designed to display an audible or visual alarm (or both)

when the local dose rate exceeds the preset alarm threshold on the instrument. It may

also display the local dose rate on a meter. Most modern gamma monitors have a

detector fail alarm facility which is often held off with a radioactive priming source.

This facility allows the instrument to perform self-checks continuously: the instrument

electronics will inform the user if there is a detector fault such that the expected pulse

rate from the detector is not received by the associated electronics. Instruments

without either a priming source or detector fail alarm will have no built in detector

checks.

For the tests defined in this guidance, installed gamma dose rate monitors may be

divided into three classes:

Class A Monitors with a priming source and with a detector fail alarm.

Class A gamma monitors have a priming source fitted to the detector to produce a

definite count rate or current at background levels. These are generally instruments

with low sensitivities where the natural background count rate is too low to be able to

give confidence that the detector is functioning.

Class B Monitors with a detector fail alarm only.

Class B gamma monitors are similar to Class A monitors, but are generally more

sensitive and do not require a priming source to trigger detector fail alarm.

Class C Monitors with no priming source and with no detector fail alarm.

Class C gamma monitors do not have detector fail alarm functionality. Installation of

these instruments for radiation protection purposes is not generally recommended,

however, they may continue to serve a useful function in low risk situations.

These classes are relevant to the background indication and influence the frequency of

the alarm test and functional check.

The tests required to establish the linearity, energy dependence, directional

dependence and other relevant characteristics of installed gamma dose rate monitors

Measurement Good Practice Guide No. 29

15

are detailed in Section 4. Table 2 is a quick reference guide for these monitors and

provides a brief description of each of the tests.

The three types of personnel contamination monitor are described below. Modern

installed contamination instruments monitor the background count rate while they are

not monitoring personnel. The monitor therefore compensates for any changes in

background. Tests are required to ensure that the alarm threshold will operate at an

appropriate contamination level.

It is important that the alarm threshold on these instruments is set to activate at the

appropriate surface contamination level. Reference should be made to „The Selection

of Alarm Levels for Personnel Exit Monitors‟ published by Industry Radiological

Protection Co-ordination Group12

.

Typically these instruments are supplied to monitor hands only, feet only or both

simultaneously. The detectors are static and the hands and/or feet are monitored when

in contact with the detector. The monitoring time is depending on how the instrument

has been set up. An outline of the tests required for hand and foot monitors is given in

Table 3 and more information is provided in Section 5. For those monitors fitted with

frisking probes, the probes should be tested as Section 3.2.2 below.

These instruments consist of a probe connected to a monitoring assembly. The users

monitor themselves by moving the probe slowly over their body. Other types of

contamination monitor may have position sensors, e.g. for hands, feet, closeness of the

body to arrays of detectors, but frisking monitors rely entirely on the self-monitoring

method adopted by the user. An outline of the tests required for frisking monitors is

given in Table 3 and more information is provided in Section 5.

In some situations a portable contamination monitor may be fixed to a location and

used as a frisking monitor. In this situation, the monitor should be tested in

accordance with the recommendations of GPG144.

These instruments typically consist of an array of detectors to monitor for

contamination on the body, hands and feet. The user is normally required to position

themselves against this array for a short, period of time.

Measurement Good Practice Guide No. 29

16

A brief description of the tests required for exit monitors is provided in Table 3 while

detailed information can be found in Section 5. Note that some detector housings may

contain multiple separate detectors each with more than one channel; each channel

should be tested.

3.2.3.1

These instruments are designed to monitor alpha and/or beta emitting contamination

on the body. Typically this is achieved by an array of detectors mounted close

together. These also incorporate the hand and foot detectors as described in Section

3.2.1.

3.2.3.2

Alpha and/or beta contamination monitoring instruments may be supplemented with

additional gamma scintillation detectors. These detectors are designed to be sensitive

to higher energy gamma emissions, typically of energy in excess of 60 keV. They are

designed to monitor contamination where alpha and beta emissions may be shielded

by clothing or the user‟s body, or where only photons are emitted. Each detector may

be configured with an alarm. An outline of the tests required for these monitors is

provided in Section 6.

These instruments are used to monitor personnel for beta and/or gamma emitting

contamination. They are designed for both „walkthrough‟ mode and well as „standing

stationary‟ mode. The guidance provided in this document is intended to satisfy the

IRRs. However, similar systems may be employed for other purposes and the

principles of this guidance may be extended to cover these applications.

An outline of the tests required for Personnel Portal Monitors is given in Table 4 and

more information is provided in Section 6. Each detector should be tested.

Small Articles Monitors (SAMs) are designed to monitor photon emitting

contamination on, or in, articles such as tools, hand held instruments and personal

artefacts. Typically they have 2, 4 or 6 gamma scintillation detectors which are

mounted around a cuboid chamber. Alarms are normally set based upon the summed

response of the detectors. Doors are often employed to control the release of the

articles from the controlled area.

SAMs may also be incorporated within an exit monitor and may utilise alpha and/or

beta detectors as well as gamma detectors.

Measurement Good Practice Guide No. 29

17

SAMs may be designed with an internal or external weighscale, for the purpose of

measuring the specific activity of an article. The specific activity mode of operation is

outside the scope of this guidance document, however, it may be convenient to test the

functionality of the weighscale at the same time as the radiation testing.

An outline of the tests required for SAMs is given in Table 5 and more information is

provided in Section 7.

Measurement Good Practice Guide No. 29

18

TEST REQUIRED

COMMENTS

PASS / FAIL CRITERIA

TESTS

BEFORE

FIRST USE

PERIODIC

TESTS

ROUTINE

TESTS

DETAILED

REFERENCE

FUNCTIONAL CHECK

Check indicator lights are functioning; visual check of

physical condition. Check alarm using the check function,

if available. Check display operates correctly.

Good physical condition, all indicators

functional

Yes

Yes

Yes

Section 4.1

BACKGROUND INDICATION

Check and note the background indication.

Class A monitors will display an

elevated background.

No significant deviation in indication from

previous test, unless there has been a change in

the ambient conditions.

Yes

Yes

Yes

Section 4.2

ALARM TEST

OPERATIONAL HIGH LEVEL

Expose the instrument to a dose rate no more than 50 %

above the preset alarm threshold.

Test to be performed where

practicable and operator doses are

ALARP

Satisfactory initiation of alarm indication

Yes

Yes

No*

Section 4.3.1

DETECTOR FAIL

Confirm the detector fail alarm operation

Low dose rate test only applies to

Class A and Class B monitors.

Satisfactory operation in accordance with

manufacturer‟s specification

Yes

Yes

No

Section 4.3.2

RESPONSE TO HIGH DOSE RATES

Expose the instrument the dose rate equivalent to at least

the maximum foreseeable dose rate. A minimum dose rate

of 10 mSv h-1 should be used.

Test to be performed where

practicable and operator doses are

ALARP

Satisfactory indication of dose rate or overload

condition for each detector used

Yes

Yes

No

Section 4.4

LINEARITY

Mount the instrument in the calibration orientation, with its

reference point at the point of test in the radiation field of a 137Cs source. Measure the instrument‟s response to the

field for each range or decade of the instrument, up to the

maximum dose rate it could reasonably encounter in the

workplace, even in accident conditions.

60Co may be used if 137Cs is not

available.

Where more than one display

exists, check all displays.

Agreement to within 30% of type test data.

Yes

No

No

Section 4.5

ENERGY DEPENDENCE

Mount the instrument in the calibration orientation, with its

reference point at the point of test in the radiation field of

an appropriate low energy photon source. The dose rate

from the source should be adjusted until the instrument

reading is close to one of those obtained for 137Cs in the

linearity test. Determine the instrument‟s response to the

low energy source.

Filtered X-radiation from the ISO

low or narrow series may also be

used, particularly for high dose rate

detectors.

The ratio of the low energy response to the

response from the linearity test should agree

within 30% of that in type test data.

Yes

No

No

Section 4.6

Measurement Good Practice Guide No. 29

19

DIRECTIONAL DEPENDENCE

Using the same dose rate and photon energy as for the energy dependence test, determine the instrument‟s response at ± , where is an angle between 45° and 90°.

The angle will typically be 90°, however it may be more appropriate to use a smaller angle if the instrument does not have a useful response at 90°.

Agreement to within 30% of type test data.

Yes

No

No

Section 4.7

* Category C monitors will require this check as part of the Routine Test

Measurement Good Practice Guide No. 29

20

TEST REQUIRED

COMMENTS

PASS / FAIL CRITERIA

TESTS

BEFORE

FIRST USE

PERIODIC

TEST

ROUTINE

TEST

DETAILED

REFERENCE

FUNCTIONAL CHECK

Visual check of physical condition. Clean foot monitors if

appropriate. Check gas flow and supplies.

Good physical condition, all indicators

functional

Yes

Yes

Yes

Section 5.1

ENERGY THRESHOLD CHECK

For alpha/beta detectors, evaluate the proportion of counts

from an alpha emitting source in the beta channel and

counts from a beta emitting source in the alpha channel The beta source used should have a high energy beta

emission e.g. 90Sr + 90Y and the alpha source should be 241Am.

The beta-in-alpha ratio should not exceed

0.01. The alpha-in-beta ratio should not differ

by more than 10% from the previous Periodic

Test

Yes

Yes

Yes

Section 5.2

BACKGROUND INDICATION

Check whether the background indication differs from the

normal level.

No significant deviation in indication from

previous test, unless there has been a change

in the ambient conditions.

Yes

Yes

Yes

Section 5.3

RESPONSE TO CONTAMINATION

Check the efficiency of the detector with the appropriate

nuclides given in Appendix C.

Results should agree within 30% of

baseline and type test data.

Yes

Yes

No

Section 5.4

COUNT RATE ALARM TEST

Check with a traceable source of the appropriate emission

type to demonstrate that the alarm is working correctly at

the each alarm threshold.

Satisfactory initiation of alarm indication

Yes

Yes

Yes

Section 5.5

RESPONSE TO HIGH ACTIVITY SOURCE

Expose the detector to a source of activity at least 10 times

greater than that used for the alarm test .

Satisfactory initiation of alarm indication

Yes

No

No

Section 5.6

UNIFORMITY OF RESPONSE

Use a 50 cm2 source to determine the instrument response

at various positions over the detector window. Calculate

the mean response over the whole window. Very specific

size

Only instruments with detector

areas in excess of 150 cm2 need be

tested. Frisking probes should be

tested in accordance with GPG14

No individual area should have a response,

outside a factor of 2 of the mean response.

Yes

No

No

Section 5.7

Measurement Good Practice Guide No. 29

21

TEST REQUIRED

COMMENTS

PASS / FAIL CRITERIA

TESTS

BEFORE

FIRST USE

PERIODIC

TEST

ROUTINE

TEST

DETAILED

REFERENCE

FUNCTIONAL CHECK

Check indicator lights are functioning; visual check of

physical condition. Check alarm using the check function,

if available. Check display operates correctly.

Good physical condition, all indicators

functional

Yes

Yes

Yes

Section 6.1

ENERGY THRESHOLD CHECK

Using a source and method recommended by the

manufacturer, confirm the high voltage is set appropriately

High voltage set to optimise performance

Yes

Yes

Yes

Section 6.2

BACKGROUND INDICATION

Check whether the background indication differs from the

normal level.

No significant deviation in indication from

previous test, unless there has been a

change in the ambient conditions.

Yes

Yes

Yes

Section 6.3

COUNT RATE ALARM TEST

Check with a calibrated source of the appropriate

radionuclide to demonstrate that the alarm is working

correctly at the alarm level. Each alarm utilised should be

activated individually

Satisfactory initiation of alarm indication

Yes

Yes

Yes

Section 6.4

RESPONSE TO CONTAMINATION

Check the efficiency of the detector with the appropriate

nuclides given in Appendix C. Each detector should be

tested.

No specific method provided,

geometry should be reproducible

and representative

Results should agree within 30% of baseline data, and type test data.

Yes

Yes

Yes

Section 6.5

RESPONSE TO HIGH ACTIVITY

Confirm the instrument response to activity far in excess of

alarm levels

Instrument operation as per type test

Yes

No

No

Section 6.6

UNIFORMITY OF RESPONSE

Perform a series of measurements in the vertical and

horizontal plane.

Specific methods detailed in the

detailed reference

Results should agree within 30% of type test data.

Yes

No

No

Section 6.7

Measurement Good Practice Guide No. 29

22

TEST REQUIRED

COMMENTS

PASS / FAIL CRITERIA

TESTS

BEFORE

FIRST USE

PERIODIC

TEST

ROUTINE

TEST

DETAILED

REFERENCE

FUNCTIONAL CHECK

Check indicator lights are functioning; visual check of

physical condition. Check alarm using the check function,

if available. Check display operates correctly.

Good physical condition, all indicators

functional

Yes

Yes

Yes

Section 7.1

ENERGY THRESHOLD CHECK

Using a source and method recommended by the

manufacturer, confirm the high voltage is set appropriately

High voltage set to optimise performance

Yes

Yes

Yes

Section 7.2

BACKGROUND INDICATION

Check whether the background indication differs from the

normal level.

No significant deviation in indication from

previous test, unless there has been a

change in the ambient conditions.

Yes

Yes

Yes

Section 7.3

COUNT RATE ALARM TEST

Check with a calibrated source of the appropriate

radionuclide to demonstrate that the alarm is working

correctly at the alarm level. Each alarm utilised should be

activated individually

Satisfactory initiation of alarm indication

Yes

Yes

Yes

Section 7.4

RESPONSE TO CONTAMINATION

Check the efficiency of the detector with the appropriate

nuclides given in Appendix C. Each detector should be

tested.

No specific method provided,

geometry should be reproducible

and representative. Where sum

zones are used in measurement,

this operation should also be

tested.

Results should agree within 20% of baseline data and type test data.

Yes

Yes

Yes

Section 6.5

LINEARITY OF RESPONSE

Measurements should be performed over the measurement

range anticipated in operation

Results should agree within 20% of

baseline data and type test data.

Yes

No

No

Section 7.6

RESPONSE TO HIGH ACTIVITY

Confirm the instrument response to activity far in excess of

alarm levels

Instrument operation as per type test

Yes

No

No

Section 7.7

SPATIAL RESPONSE

Confirm the detector response in centre of the chamber

relative to the summed detector response

Results should agree within 30% of

baseline data and type test data.

Yes No No

Section 7.8

Measurement Good Practice Guide No. 29

23

Measurement Good Practice Guide No. 29

24

Specific Tests for

Gamma Dose

Rate Monitors

IN THIS CHAPTER

44

Functional Check

Background Indication

Alarm Test

Response to High Dose Rates

Linearity of Response

Energy Dependence of Gamma Monitors

Directional Dependence

Measurement Good Practice Guide No. 29

25

Table 1 lists the tests, which are applicable to the TBFU and Periodic Tests for gamma

monitors. Table 2 provides a brief summary of the tests for gamma monitors and

analysis of test results: the tables are not comprehensive and should not be used

without reference to the detailed information in the sections of text. The tests may be

performed in an order that is convenient to the Test House. If the response to high

dose rates is not tested first, it is important to check that this test, when conducted

subsequently, has not adversely affected other aspects of the instrument performance.

Full procedures for the performance of all of the tests are provided in the remainder of

this Section.

Perform a visual check of the physical condition of the instrument. Check that

indicator lights are functioning. Note the threshold at which the alarm is activated.

Check the alarm using any check function available; this should ensure that both

audible and visual indicators work correctly.

If the instrument has a digital display and has a display check function, check that all

segments of the display work correctly.

If Class C instruments are in service, due to the absence of low level alarm, functional

checks should be performed with increased frequency (e.g. daily) to maintain

confidence that they still provide the required level of protection.

Obtain a background measurement and compare with baseline data; any significant

deviation in indication should be investigated unless there has been a change in

ambient conditions.

Where the function exists, expose the instrument to a dose rate no more than 50%

above the defined alarm threshold and confirm that the alarm is activated.

Some instruments use two detectors, one of which is relatively sensitive and operates

at low dose rates and a much less sensitive one which takes over automatically and

provides the indication at higher dose rates. Normally, the operational alarm will be

triggered by the sensitive detector and a simple alarm function test will often not check

the high dose rate detector. It is important to confirm that the alarm operates correctly

up to the maximum radiation level that could be encountered.

Measurement Good Practice Guide No. 29

26

If the monitor is routinely exposed to a reproducible radiation field e.g. radiation cell

interlock monitor, proof that the monitor responds in an expected manner may satisfy

this check.

Many instrument designs are equipped with an alarm latch, which holds the alarm on,

even if the dose rate has dropped below the alarm level. Where the function exists, the

operation of the alarm latch should be confirmed.

The response time for the alarm to be initiated should be recorded and be within the

response time specified by the RPA.

Class C monitors must undergo an alarm test on a routine basis. Where the instrument

is operating in an area with an enhanced background dose rate, such that instrument

reading is always within the first decade of the measurement, the Employer may

justify the relaxation of the routine alarm test. However the Employer should note, on

a routine basis, that the monitor continues to detect the enhanced background dose

rate.

Where practicable, activate the Detector Fail Alarm to ensure it operates in accordance

with manufacture‟s specifications. Where the Detector Fail Alarm does not operate as

anticipated this should be investigated. The instrument manufacturer should be

consulted as to an appropriate method to conduct this test.

Failure of equipment or operational procedure could lead to dose rates far beyond

those routinely encountered. This possibility should have been recognised in a risk

assessment and a suitable instrument selected that has been type tested up to a

sufficiently high dose rate. Where practicable, the instrument should be tested to at

least the maximum dose rate it could encounter. Where this is not practicable, the

instrument should be tested to as high a dose rate as practicable and then an analysis

made of the instrument function to confirm that there is no reasonable possibility of its

failure to danger at dose rates above those tested. This requires a detailed

understanding of how the instrument operates; in particular the detector, polarising

supply, input amplifier and subsequent electronics.

It is important to ensure that the high dose rate test has not damaged the instrument,

for example, leading to a high background count rate after the test. The test should

therefore take place early in the testing routine, before the alarm test. It is also good

practice to check the instrument indication at an elevated dose rate level before and

Measurement Good Practice Guide No. 29

27

after the high dose rate test. Any significant change should cast doubt on the

appropriateness of the instrument for that application. It should be noted that Geiger

Muller detectors have a life of only about 1011

counts, so high dose rate exposure test

times should be kept short. This does not mean that such instruments are unsuitable for

potential very high dose rates, provided such levels are infrequent and of short

duration.

Where more than one detector is used to provide operation over an extended dose rate

range, the performance of the low dose rate detector should be confirmed during the

high dose rate test. Typically this is displayed as OFLOW or ------. It is not unknown

for the low dose rate detector to fail at high dose rates therefore impacting upon

operation of the instrument as a whole.

The instrument should be tested with the alarm latch off, to ensure that the alarm does

not stop at very high dose rates. It should then be turned on and checked by exceeding

the alarm level, observing a correct response, then reducing the dose rate below the

alarm level and confirming that the alarm indications continue.

Where in-situ periodic testing is chosen, instruments should normally be tested up to at

least 10 mSv h-1

. This can be achieved safely in the workplace using a collimated

source mounted close to the detector using a jig, which provides reproducible

geometry.

Where there is a possibility of dose rates much greater than 10 mSv h-1

then, where

practicable, the instrument should be exposed during the periodic test to a dose rate

which will confirm that the instrument functions correctly at these dose rates.

Additional guidance for deriving reference dose rates and performing tests is provided

in Appendix A.

The instrument under test should be mounted in the calibration orientation, with its

reference point (marked calibration point on the detector or detector housing), or in

the absence of a marked calibration point, the geometric centre of the detector, at the

point of test in the radiation field from 137

Cs gamma radiation; 60

Co may be used as an

alternative source. A combination of both gamma radiations may be used, where the

appropriate range of dose rates from 137

Cs is not available. The instrument‟s response

to the field should be measured for at least one dose rate in each range or decade of

the instrument, up to the maximum dose rate which it could reasonably encounter in

the workplace, even under accident conditions. If the response of an instrument is

Measurement Good Practice Guide No. 29

28

found to be unsatisfactory, it may be possible in some cases for the instrument to be

adjusted to give an acceptable response over its range of use. Any adjustment made

should be reported on the test report by displaying a before and after adjustment

reading (or response).

Where appropriate, the test report should give the instrument response or

calibration/multiplication factors which enable the user to convert the instrument

indication to dose rate, or quote that the instrument‟s response is acceptable within a

specified range of dose rates, or that it has been adjusted to be acceptable within the

range. The instrument responses in the known calibration fields should be within

±30% of the baseline data and ideally within ±30% of the true dose rate. Any untested

ranges or decades should be clearly indicated on the test report.

The energy dependence of instruments used to measure dose rates in the workplace is

governed by the type of detector and, in some cases, on the setting up of the electronic

system of the instrument. The following test is designed to confirm that the response

of the instrument does not vary with energy in a manner which is significantly

different to that quoted in the Type Test data. The test utilises the information

obtained in the linearity test described above, and combines it with a test procedure

using an 241

Am photon radiation source. This test should identify any major faults in

the detector.

Information at one energy (corresponding to 137

Cs or 60

Co) should have been obtained

in the linearity test described in Section 4.5. For many instruments, a test at a much

lower energy is required to confirm that the energy dependence corresponds, within

acceptable limits, to that quoted in Type Test data. This is because incorrect assembly

or the use of wrong materials during repair may have a negligible effect on instrument

response at high energies, while having a more significant effect at low energies.

The instrument should be mounted in the calibration orientation with its reference

point (marked calibration point of the detector or detector housing), or in the absence

of a marked calibration point, the geometric centre of the detector, at the point of test

in the radiation beam. The recommended radiation energy is 60 keV (241

Am gamma

radiation), although an appropriate X radiation quality from the ISO low or narrow

series of reference filtered X radiation11

may be used. The dose rate from the 241

Am or

X radiation should be adjusted until the instrument indication is close to one of those

from 137

Cs or 60

Co used in the linearity measurement so as to eliminate any effects of

non-linearity. The true dose rate, at the point of test in the 241

Am or X radiation field,

should then be determined and the instrument response or calibration/multiplication

factor derived.

Measurement Good Practice Guide No. 29

29

The ratio of the low energy response to that for 137

Cs or 60

Co gamma radiation should

be calculated and compared with the same ratio derived as the type test data. The ratios

should agree to within 30%. Caution should be exercised when comparing 241

Am to 137

Cs response ratio with a 60 keV X radiation to 137

Cs response ratio due to the fact

that the X-radiation is emitted with a range of energies while the emission from 241

Am

is effectively mono-energetic; as a consequence the instrument‟s response is unlikely

to be identical to the two radiations.

Where an instrument uses more than one detector, this test should be performed on

each detector.

The majority of instruments are intended to respond isotropically to radiation. This

characteristic is normally investigated during Type Testing. However, it is possible

during instrument manufacture to produce gross defects in directional dependence by,

for example, missing out components in the energy compensation filter of a Geiger-

Müller detector or in the internal energy-correction components of an ionisation

chamber. These errors may not be detected in the energy dependence test.

The directional dependence test can normally be performed by rotating an instrument‟s

detector housing in the horizontal plane about its calibration reference point and

measuring its response in each orientation. The response at 0° should be compared

with the response at ± ° where is an angle between 45° and 90°. Although typically

will be 90°, the Test House may prefer not to test it at an angle where a known blind

spot occurs. Note that in the workplace, since the monitor is likely to be mounted on a

wall, irradiation from ±90° is unlikely. Depending on the proposed use of the

instrument, it may also be necessary to carry out a similar test in the vertical plane,

especially for a cylindrical detector (or detector housing) which the Employer may

choose to mount in the +90° (end on) orientation. For photon dose rate monitors, the

same radiation quality, normally 241

Am gamma radiation, should be used as in the

energy dependence confirmation. A relatively low energy is preferable to 137

Cs or 60

Co gamma radiation because the test is much more sensitive at the lower energy. It is

essential that all detectors are tested. Appendix B discusses angles which are suitable

for testing different detector designs.

The ratio of the 60 keV response at the selected angle to the 60 keV response at 0°

should not differ from the type test data by more than 30%. If the difference exceeds

30%, the Employer may decide, after consultation with the Qualified Person and

taking account of the fixed geometry of the instrument, that the instrument is suitable

for use. In this situation, an appropriate remark should be made on the test report.

Measurement Good Practice Guide No. 29

30

Specific Tests for

Personnel

Contamination Monitors

IN THIS CHAPTER

55

Functional Check

Energy Threshold Check

Background Indication

Response to Contamination

Count Rate Alarm Test

Response to a High Activity Source

Uniformity of Response

Measurement Good Practice Guide No. 29

31

Table 1 lists the tests, which are applicable to the TBFU and Periodic Tests for

contamination monitors. Table 3 provides a brief summary of the tests for

contamination monitors and analysis of test results: the tables are not comprehensive

and should not be used without reference to the detailed information in the sections of

text.

Full procedures for the performance of all of the tests are provided in the remainder of

this Section. It is recommended that the first three tests in this Section are performed

in the order specified.

Perform a visual inspection including the case and display and, if necessary,

connecting cables. Check background response and clean foot monitors if required.

Check gas flow and supplies if appropriate; the flow rate out of the instrument should

not deviate significantly from the flow rate into the instrument. For scintillation

detectors, check for light sensitivity.

If the instrument has a digital display and has a display check function, check that all

segments of the display work correctly.

For gas flow systems, the detector response should be checked close to the outflow - if

there are leaks in the detector, the gas pressure will be least near the outflow.

The Energy Threshold is normally set by the high voltage.

For alpha/beta detectors, evaluate the proportion of counts from an alpha emitting

source in the beta channel and counts from a beta emitting source in the alpha channel

and compare with data from the previous plateau check. The beta source used should

have a high energy beta emission e.g. 90

Sr +

90Y

and the alpha source should be

241Am.

The beta-in-alpha ratio should not exceed 0.01. The alpha-in-beta ratio should not

differ by more than 10% from the previous Periodic Test; otherwise re-determine the

optimum operating voltage recommended by the manufacturer. Note that the alpha-

in-beta ratio may increase with the activity of the test source.

For beta only detectors, check the 14

C response and compare to previous records. For

instruments intended to measure low energy photon contamination, use 55

Fe instead of 14

C. If the response differs by more than 10% from that derived from the last plateau

check, re-determine the optimum operating voltage.

Measurement Good Practice Guide No. 29

32

The plateau check is a suitable component test to undertake after a simple repair e.g.

foil change.

Check the background count rate on each detector. The manufacturer‟s data should

define the expected background response rates. Any significant deviation from the

established baseline data should be investigated. If background count rate is elevated

it could be due to contamination on a detector window. An elevated background may

be acceptable if agreed with the RPA.

When the user places their body close to any detector, the instrument background may

change. Therefore the Background Indication Test should be performed in such a way

as to minimise the effect of a person present.

Obtain the response to contamination using at least one radionuclide of the appropriate

radiation emission type that the instrument is designed to measure. Appendix C lists

suitable radionuclides for each type of radiation emission. Where the instrument is

designed to measure alpha and beta contamination, the instrument should be tested to

both an appropriate alpha and beta emitting radionuclide. The response to photon

contamination test should be undertaken where there is a potential for contamination

from photon emitting radionuclides. The results should agree to within 30% of both

baseline data and of type test data.

This guidance does not specify a suitable distance between the test sources and the

detector, however it is important that the test geometry is reproducible and

representative.

If an instrument has many detectors e.g. an Exit Monitor, the time taken to test each

detector would be reduced significantly by using sources of greater activity than those

used to conduct the alarm test e.g. > 5 kBq. The high activity source will allow the use

of shorter counting times to provide statistically significant results. This is especially

true where the instrument detectors show a high count rate due to background gamma

radiation. Although the response test may be faster using a higher activity source, it

should be noted that some detectors might have a significantly lower response to high

activity sources, than to the lower activity alarm test sources. Therefore, if this

method is adopted, baseline data for the high activity source must be available and

used for the comparison.

Measurement Good Practice Guide No. 29

33

Test with a traceable source(s) to demonstrate that each alarm in use is working

correctly at the alarm threshold.

Due to the impracticalities of manufacturing sources of exact emission rate, it is

acceptable to use test sources which will give rise to count rates up to three times

above the alarm threshold.

Every detector should be tested at least once every three months using a systematic

procedure. Exit (Personnel) Monitors that have diagnostic software may allow testing

of more than a single detector at the same time.

It is necessary to have confidence that a monitor will continue to operate correctly at

count rates well in excess of those at which the count rate alarm is activated. Therefore

a high activity test should be performed using a source of activity at least ten times

that used for the alarm test (Section 5.5). A suitable source activity is 10 kBq and the

alarm should operate when exposed to the source. Ideally the source should be the

same radionuclide, construction and dimensions as the source used for the alarm test.

Where this is not practicable, the response from the high activity source should be

compared with the response from a source of the same radionuclide and construction,

but with an activity similar to that used for the alarm test. If the response varies

significantly between the two sources then this information should be considered when

selecting sources for the Response to Contamination test (Section 5.4).

This test is only required for instruments with scintillation detectors. Instruments with

a detector area in excess of 150 cm2 should be checked to ensure their response to

appropriate radiations is reasonably uniform over the whole area of the detector.

Frisking probes should be tested as in GPG 14. This test is designed to identify areas

of the detector which have an inadequate detection efficiency as a result of incorrect

assembly or defective materials.

This test will normally not require more than an eight-position check using a source of

area less than 50 cm2. The energy of the emissions from the source should be less than

or equal to the lowest energy radionuclide the instrument is intended to detect. To

determine the uniformity of a detector, divide the detector area into a number of

segments and measure the instrument response in each segment; no individual

segment should have a response outside a factor of 2 of the mean response.

Measurement Good Practice Guide No. 29

34

Specific Tests for

Portal Monitors

IN THIS CHAPTER

66

Functional Check

Energy Threshold Check

Background Indication

Count Rate Alarm Test

Response to Contamination

Response to High Activity Source

Uniformity of Response

Measurement Good Practice Guide No. 29

35

Table 1 lists the tests which are applicable to the TBFU and Periodic Tests for gamma

portal monitors. These tests are also applicable to the gamma detectors within an Exit

Monitor (Section 3.2.3.2). Table 4 provides a brief summary of the tests for gamma

portal monitors and analysis of test results: the tables are not comprehensive and

should not be used without reference to the detailed information in the sections of text.

The tests may be performed in an order that is convenient to the Test House.

Full procedures for the performance of all of the tests are provided in the remainder of

this Section. It is recommended that the tests in this Section are performed in the order

specified.

Perform a visual inspection of the unit. Where diagnostic facilities exist, check that

the status indication lights are activated correctly, and the position sensors are

correctly identified by the monitor.

The energy thresholds are normally set by the high voltage. The method used for

assessing the optimum operating voltage varies significantly between manufacturers.

This guidance will not provide a specific energy threshold check criterion for each

monitor.

Using the gamma emitting source and procedure recommended by the manufacturer,

confirm the high voltage is set appropriately to optimise performance.

In the absence of a procedure recommended by the manufacturer, a source of photon

energy at or below the minimum energy of interest should be used to check no

significant variation from baseline data. Where the results of this test do not meet the

acceptance criteria, a high voltage scan should be undertaken.

The high voltage scan is essential if a detector has been repaired or replaced.

These monitors will have a background compensation feature. However, it is also

possible for a high background gamma dose rate on a single detector to reduce the

effectiveness of the background compensation of the whole instrument. Similarly, a

contaminated detector panel may also reduce the effectiveness of the compensation.

Check the background count rate on each detector. Any deviation greater than 20%

from the baseline data should be investigated.

Measurement Good Practice Guide No. 29

36

Test with traceable source(s) to demonstrate that each alarm set is working correctly

at the alarm threshold.

Each detector alarm utilised during operation should be activated individually.

Due to the impracticalities of manufacturing sources of exact emission rate, it is

acceptable to use test sources or geometries which will give rise to count rates up to

three times above the alarm threshold

The high voltage scan is essential if a detector has been repaired or replaced.

Obtain the response to contamination using at least one radionuclide that the

instrument is designed to measure. Appendix C lists suitable radionuclides. The results

should agree to within 30% of both baseline data and of type test data.

This guidance does not specify a suitable distance between the test sources and the

detector, however it is important that the test geometry is reproducible and

representative.

Each detector should be tested, and in situations where the instrument utilises alarms

based sum zone efficiencies, then these should also be tested.

The High Activity response test ensures that the instrument response does not deviate

significantly from the type test response at activities far in excess of the alarm levels

and is a suitable test of any dead time correction. A suitable test source should have

an activity in excess of 1 MBq of 137

Cs, or the maximum activity anticipated by the

employer if this is a lower value. Ideally the source should be the same radionuclide,

construction and dimensions, and positioned in the same geometry, as the source used

for the response to contamination test. Each detector should be tested. The test results

should agree to within ±30 % of Type Test data.

This test characterises the detection system, defining the range of sensitivity across the

measurement area. Measurements should be performed using a source of the reference

radionuclide (see Appendix C). The number of measurements to be performed should

be determined from the Type Test and the instrument design. This test is designed to

identify significant deviations in detection efficiency, compared to that shown in the

Measurement Good Practice Guide No. 29

37

Type Test.

Using the vertical scan information from the Type Test, test the response at the heights

above the foot plinth corresponding to the maxima and minima [from the plot] of each

vertical array. The horizontal distance from the array should correspond to that used

for the Type Test; this is typically 5 cm. For example an instrument with 2 vertical

arrays with 3 detectors each will require 14 test positions. Test results should agree to

within ±30% of type test data.

Measurement Good Practice Guide No. 29

38

Specific Tests for Small

Articles Monitors

IN THIS CHAPTER

77

Functional Check

Energy Threshold Check

Background Indication

Count Rate Alarm Test

Response to Contamination

Linearity of Response

Response to High Activity Source

Spatial Response

Measurement Good Practice Guide No. 29

39

Table 1 lists the tests, which are applicable to the TBFU and Periodic Tests for small

articles monitors. Table 5 provides a brief summary of the tests for small articles

monitors and analysis of test results: the tables are not comprehensive and should not

be used without reference to the detailed information in the sections of text. The tests

may be performed in an order that is convenient to the Test House.

Full procedures for the performance of all of the tests are provided in the remainder of

this Section. It is recommended that the tests in this Section are performed in the order

specified

Perform a visual inspection including the door hinges and internal cleanliness.

Where diagnostic facilities exist, check that the status indication lights are activated

correctly, and the door, lock and start button status is correctly identified by the

monitor.

The method used for assessing the optimum operating voltage varies significantly

between manufacturers. This guidance will not provide a specific energy threshold

check criterion for each monitor.

Using the gamma emitting source and procedure recommended by the manufacturer,

confirm the high voltage is set appropriately to optimise performance.

In the absence of a procedure recommended by the manufacturer, a source of photon

energy at or below the minimum energy of interest should be used to check no

significant variation from baseline data. Where the results of this test do not meet the

acceptance criteria, a high voltage scan should be undertaken.

The high voltage scan is essential, if a detector has been repaired or replaced.

These monitors will have a background compensation feature. However, it is also

possible for a high background gamma dose rate on a single detector to reduce the

effectiveness of the background compensation of the whole instrument. Similarly, a

contaminated internal liner may also reduce the effectiveness of the compensation.

Check the background count rate on each detector. Any deviation greater than 20%

from the baseline data should be investigated.

Measurement Good Practice Guide No. 29

40

Test with traceable source(s) to demonstrate that each alarm set is working correctly

at the alarm threshold.

Each alarm utilised during operation should be activated individually.

Due to the impracticalities of manufacturing sources of exact emission rate, it is

acceptable to use test sources or geometries which will give rise to count rates up to

three times above the alarm threshold.

Obtain the response to contamination using at least one radionuclide that the

instrument is designed to measure. Appendix C lists suitable radionuclides for gamma

detectors. The results should agree to within 20% of both baseline data and of type

test data.

This guidance does not specify the source position in the chamber, however it is

important that the test geometry is reproducible and representative.

In situation where the instrument only utilises alarms based on sum zone efficiencies,

then only the sum zone needs to be tested.

The source(s) used for this test may be calibrated using a SAM as a transfer standard.

This test is important to ensure the instrument is providing a consistent response over

the anticipated range of measurement. Sources should be positioned in a repeatable

geometry, where practicable matching that used during type testing. Measurements

should be performed over the measurement range the instrument may reasonably

encounter in the workplace. A minimum of two measurements should be made during

this test; it is highly recommended the lowest measurement is performed with a source

of activity less than that required to initiate an activity alarm while a second

measurement should be undertaken with sufficient activity to produce an indication at

least ten times the alarm level. To confirm linearity of the instrument, it is desirable to

have an additional test point between the lowest and highest activities. Test results

should agree to within ±20 % of type test and baseline data. Depending upon the

range of source activity used, it may be possible to combine the linearity and high

activity response tests into a single test protocol.

Measurement Good Practice Guide No. 29

41

The High Activity response test ensures that the instrument response does not deviate

significantly from the type test results at activities far in excess of the alarm levels and

is a suitable test of any dead time correction. A 300 kBq source of 137

Cs is suitable for

this test.

A spatial response test, as provided in Type Test data, is not necessary for TBFU,

Periodic and Routine Tests. Measurements should be performed using a source of the

reference radionuclide (see Appendix C). Place the test source in the centre of the

measurement chamber and calculate the response of each detector. The individual

detector response, normalised to the sum of all detector responses, should agree to

within 30% of Type Test data.

Measurement Good Practice Guide No. 29

42

Facilities and

Traceability

IN THIS CHAPTER

88

Source Considerations and Jigs

Workplace Contamination Monitors

Workplace SAMS

Workplace Portals

Measurement Good Practice Guide No. 29

43

The majority of the tests described in this document would be performed in the

workplace rather than the calibration laboratory. The facilities and traceability

required in order to undertake tests on an installed instrument in a calibration

laboratory are broadly similar to those required for a portable instrument. This

document will not discuss the facilities and traceability that are required for a

calibration laboratory, since they are described in detail in GPG144. Therefore the

focus of this section is on sources used for testing and mechanisms for achieving

reproducibility.

As the majority of testing of installed instruments is done in the workplace, the

standard of facilities may not be equivalent to those found in a calibration laboratory.

However, a QMS and thorough procedures for the documentation of these tests are as

important as the performance of the tests themselves.

In general, the Periodic Testing of gamma monitors will take place in the workplace.

The requirements are:

A suitable radiation source, capable of generating the dose rates of interest;

A means of positioning the source at a reference position, such as a jig;

The ability to store and deploy the source safely and with minimal personnel

exposure in the workplace.

A means of effectively producing traceability for each source and position is normally

achieved immediately after the TBFU of the first of any instrument type to be

purchased. See Appendix A.

The Test House should undertake a Periodic Test as soon as practical after installation,

and compare the results to those from the Test Before First Use which had taken place

in the calibration laboratory.

The Test House should repeat this process for each type of instrument installed in the

workplace. A procedure that uses the same calibration jig on different types of

instrument can give significantly different results due to different detector dimensions.

See Appendix A.

The process selected may be used for all subsequent Periodic Tests. However, the jig

and associated sources should undergo a recalibration at least every four years. As part

of the recalibration, an example of each type of instrument that is used should be sent

to a calibration laboratory in order to re-establish traceability via a transfer standard

and the subsequent intercomparison described above should be repeated.

Measurement Good Practice Guide No. 29

44

Elaborate calibration facilities are not normally required for contamination monitors.

However, sources incorporated into calibration jigs are generally used for the Periodic

Testing of installed instruments. Calibration jigs will typically allow the Test House

to establish a reproducible geometry quickly, as well as minimising the contact

between the person undertaking the test and the source. Figures 2 to 4 show some

examples of contamination monitor calibration jigs such as a “standard hand” for a

hand monitor (Figure 2); calibration jig for a frisking monitor (Figure 3), and a ladder

source for an Exit monitor (Figure 4). Source jigs, which expose two or more

detectors simultaneously e.g. a double sided standard hand, should only be used where

the instrument shows an indication from each detector independently. Otherwise

source jigs that expose only a single detector at any one time should be used.

Measurement Good Practice Guide No. 29

45

Measurement Good Practice Guide No. 29

46

Photograph of Isotrak

TM source reproduced with permission of Eckert and Ziegler

Contamination monitor calibration sources will typically be tertiary standards, which

have been compared, not with the primary standard, but with an appropriate secondary

standard. Tertiary standards should be calibrated against a secondary standard at least

every four years and be the subject of at least an intercomparison check every two

years. Causes of deviation from the expected output have included an apparent loss of

activity to the atmosphere, which has occurred with 14

C sources, and examples where

the source was contaminated by another radionuclide of longer half life, where the

Measurement Good Practice Guide No. 29

47

effects were initially insignificant, but which become more important as the primary

nuclide decays.

The testing of workplace portal monitors generally involves the placing of sources of

specified radionuclide and activity in a defined position or positions. As such, it is the

effective activity of the sources which is important. There is a variety of mechanisms

for the measurement of effective activity. One mechanism particularly suitable for

lower activity sources involves the comparison with a suitable traceable source of the

same nuclide using an appropriate detector, such as an HPGe detector or a sodium

iodide scintillator, identifying that only the expected photopeak or peaks are present

and then comparing the photopeak count rates.

For higher activity sources, an alternative option is to measure the dose rate produced

by the source at a defined distance in low scatter conditions using a detector which has

a known response at the energy of interest. The measured dose rate can then be used to

calculate the effective activity using the specific gamma ray constant13

.

Low scatter conditions can be assured by mounting the source and detector on low

mass supports and ensuring that the source to detector centre distance is much less

than the distance from either element to floors or walls.

It is essential that the source to detector distance is at least 3 detector maximum

dimensions (depth or diameter). This allows the detector to be treated as a point

detector at its geometric centre with only minimal error.

The equipment can be quite simple. An ionisation chamber and electrometer can be

used, where the source activity is sufficient to generate a current with an acceptable

uncertainty when corrected for temperature, pressure and leakage. In many cases, an

energy compensated GM detector connected to a scaler timer is effective. These have

the merit of not requiring temperature and pressure correction and of generally being

more sensitive per unit volume. This allows the source to be mounted closer to the

detector. The statistical treatment of the results is also easier in the sense that it is

defined by the background count rate, the source count rate and the relevant measuring

times. No matter which method is used, the equipment employed must have a suitable

calibration traceable to national standards.

48

Certification of Tests

IN THIS CHAPTER

99

Calibration Laboratory

Workplace Testing

Test Label

Measurement Good Practice Guide No. 29

49

The results of tests performed under the current regulations should be communicated to

the Employer in a formal manner. The precise format of a test certificate or report is not

specified in the regulations and may depend on where the calibration or test takes place

and who it is performed by. In many instances, installed radiation protection equipment

is tested and calibrated by the Employer in the workplace and it may not be appropriate

to produce a formal certificate or report so long as an adequate record is made. Any

requirements relating to the record or production of a certificate or report should be

specified by the Employer‟s Radiation Protection Adviser. It is recommended where the

testing is performed by a Test House which is not the Employer, a formal certificate is

produced. If an instrument fails to meet the pass/fail criteria of any component of a test,

the calibration or test laboratory should prominently label the instrument as failed and

make some indication of the nature of the failure on the test document.

The information provided for an installed instrument that is tested in a Calibration

Laboratory or in the workplace by an external Test House is the same as that required for

a portable instrument. The following basic information should be provided by the Test

House:

a) the name and address of the customer or user;

b) a description of the instrument (including type, serial number and unique

identifier);

The intended use of the instrument. Where it is not possible to determine this then the

range over which it has been tested should be specified. For example, for a personnel

contamination, a suitable phrase may be “Monitoring of personnel for alpha and beta

surface contamination for radionuclides with a maximum energy in excess of 150 keV”.

c) the type of test, i.e. TBFU, Periodic Test or Retest After Repair;

d) any limitations of the calibration performed including identification of

the output modes or ranges not tested;

e) a basic description of the test, any specific instrument settings used

which may be readily modified by the user, any significant deviations

from the manufacturers recommended settings and any adjustments or

repairs performed;

f) the results of the tests including a statement of the uncertainty with the

Measurement Good Practice Guide No. 29

50

confidence level at which the uncertainty is quoted for numerical results;

g) a record of the background dose rate or count rate and any relevant

environmental conditions during the tests;

h) the value of the dose rate used for the high dose rate test or activity used

for the high activity source test;

i) the value of any conversion coefficient or P-factor applied to the results;

j) a statement that the test was carried out for the purpose of the

regulations and was successful or the test criteria were met;

k) the name and signature of the QP responsible for the test;

l) the name, address and contact details of the laboratory at which the test

was performed;

m) the date of the test;

n) the test certificate reference number.

In addition to the information listed above, further details should be provided when

surface contamination monitors have been calibrated. The sizes of the calibration sources

used for the determination of alpha, beta or photon response and the orientation in which

they were used with respect to the source should be indicated. Where a contiguous

portions measurement has been made, the value of any correction factor used should be

clearly stated.

The results of Periodic or Routine tests for an installed instrument do not typically require

the issue of a certificate for each instrument. However the Test House should indicate all

the information listed above through a combination of documented test procedures that

describe the tests to be undertaken and a test record which provides the information that

is unique to the instrument. In general the record should specify as a minimum the

following:

a) Location of the instrument;

b) Type and serial number of the instrument;

Measurement Good Practice Guide No. 29

51

c) Type of test e.g. Routine or Periodic;

d) a basic description of the test, any specific instrument settings used which may

be readily modified by the user, any significant deviations from the

manufacturers recommended settings and any adjustments or repairs

performed;

e) Results of the testing, including background indications;

f) Name and signature of QP responsible for the test.

As test reports are usually filed away for quality assurance purposes and tend not to

accompany instruments in the workplace, it is recommended that instruments which are

satisfactory are labelled after Test before First Use and Periodic testing: with the

following information

a) unique identifier of the instrument;

b) the date of calibration or test;

c) test reference (where the test is performed in a Calibration Laboratory or by an

external Test House).

If an instrument fails to meet the pass/fail criteria of any component of a test, the Test

House should prominently label the instrument as failed and make some indication of the

nature of the failure on the certificate, test report or other record. Where testing is

performed in the workplace it is especially important that the Test House indicates to

both the Employer and the user that the instrument is not fit for use.

Measurement Good Practice Guide No. 29

52

Quantities and Units

IN THIS CHAPTER

1100

Quantities and Units

53

Throughout this document the term „dose‟ has been used as a general term to refer to

the ICRU operational quantity ambient dose equivalent. The definition of this

quantity is described in detail in GPG144 and ICRU 47

9.

For gamma monitors designed to indicate dose rate, the instrument response is the

instrument reading divided by the true ambient dose equivalent rate.

For contamination monitors, the traceable quantity is the surface emission rate of the

reference source used to calibrate the instrument response for a particular

radionuclide. Instrument test reports often quote the instrument response in terms of

the number of source emissions but may also quote the response in terms of the

activity of contamination. To convert the instrument response from emissions (cps

per emission per cm2) to activity (cps per Bq per cm

2), a P-factor must be applied (see

Section 4.9.4 of GPG144).

For small articles monitors, gamma portal monitors, and the gamma detectors within

an exit monitor, the traceable quantity is the activity of the reference source or the

photon emission rate of the specific energies emitted from the source. In practice the

response is quoted in terms of the quantity 4π efficiency.

Measurement Good Practice Guide No. 29

54

Definitions

IN THIS CHAPTER

1111

Definitions

55

Ambient

(as applied to dose rate,

etc)

The normal background condition in the absence of any

artificial source of ionising radiation.

Backscatter The deflection of radiation or nuclear particles by scattering

processes through angles greater than 90° with respect to

the original direction of travel.

Baseline Data Test results (usually from a Periodic Test performed around

the time of installation of an instrument) against which all

future results of Periodic Tests and some Tests After Repair

will be compared. After significant repairs it may be

appropriate to redefine the baseline data. To be defined from

2.2 and 2.6.

Detection Efficiency The ratio between the observed count rate and the surface

emission rate of the source, expressed as a percentage.

Detector Fail Alarm

Low Level Alarm

An alarm that is activated when the pulse rate or current from

the detector falls below a certain level. Typically this alarm

will be activated by a failure of the detector or by damage to

the cabling between the detector and the processing unit of the

instrument. Historically, the term Low Level Alarm has been

used by some manufacturers to describe this function.

These operate at decreasing dose rate and generally act as an

additional fault alarm but could be used for process control.

Dual Probe

A radiation detector probe which can be used to make

independent measurements of two different types of

radiation e.g. alpha and beta.

Employer The person or establishment who is legally responsible for the

maintenance of the equipment.

Energy Compensated

(as applied to detectors)

The process of adjusting the response characteristic of a

detector system such that the measured response is

relatively independent of the energy of the radiation over an

identified range.

Fold-back

The behaviour of a detector system whereby the indication

reduces as the radiation intensity increases. In extreme

cases, the indication may drop to zero at very high exposure

rates.

Function-check Source An uncalibrated source which causes the detector to

indicate a response above background. Often used to test

Measurement Good Practice Guide No. 29

56

such items as alarm level settings, etc.

Priming Source A radioactive source, which is incorporated into the

instrument and is in close proximity to the detector. The

instrument measuring assembly will receive a constant signal

from the detector, which is above a typical background signal.

When the pulse rate from the detector drops below a

predefined level, this will initiate a low level fail alarm.

Qualified Person (QP)

A person who possesses the necessary expertise in

instrumentation, theory and practice appropriate to the

instrumentation to be tested. This person may have

responsibility for developing test protocols, taking account of

the intended use of the instrumentation, as stated by the

Employer.

Self-absorption The absorption of the ionising radiation emitted from a

radioactive source by the source material itself.

Specific Activity The activity of a given radionuclide per unit mass.

Sum Zone A pseudo detector that utilises the sum of the counts from 2

or more physical detectors. Alarm thresholds may be set for a

sum zone, in a similar way as for a single detector. Many

small articles monitors use a sum zone which is the sum of

counts from all detectors.

Test House This indicates the organisation undertaking the testing of the

instrument. This may include: an independent calibration

laboratory; a contract company who undertake the instrument

testing in the workplace; or a department under direct

management control of the Employer who undertakes testing

in the laboratory or in the workplace. Ideally the Test House

should may UKAS accreditation for all the testing that it

undertakes.

Time Constant

(of an instrument

response)

A measure of the time before the indicated response of an

instrument reaches a given percentage of true response

(often taken as 90%).

Traceability

The property of a result of a measurement whereby it can be

related to appropriate standards, generally international or

national standards, through an unbroken chain of

comparisons.

57

Appendix A:

Workplace Testing of a

Gamma Monitor

Measurement Good Practice Guide No. 29

58

On plant radiation tests on a gamma monitor benefit from the use of a test jig. This is

especially true for the high dose rate test. A well designed test jig will have the dual

benefit of providing a reproducible geometry for the tests and minimising the doses to

the Test House operator. Where the Employer owns more than one gamma monitor of

the same type, he should consider the design or purchase a test jig.

The components of a typical jig are:

1. A shielded source container

2. A suitable radiation source

3. A means of moving the source from the container to the exposure position

4. A collimator which defines the source position and limits the beam to that

which is useful

5. A means of mounting the collimator at a defined distance or distances and

orientation

6. A series of absorbers which can be used to change the dose rate without

having to house the source or move the collimator

A convenient means is to use an industrial radiography exposure unit.

The reference dose rate from such a source can be derived by determining the

indication with the source in position, in the jig, and then using conventional test

facilities to measure the true dose rate to which that corresponds. This effectively

makes the instrument used a tertiary standard. Note that the reference dose rate will

need to be determined for each instrument type (because of variations in detector size

and position within instruments). The reference dose rate should then be corrected for

radioactive decay using the appropriate half life. In most cases, 137

Cs is appropriate for

this test as it offers a reasonable dose rate per unit activity, a convenient half life and

an energy which is sufficient to penetrate instrument cases but can still be shielded

with reasonable collimator thickness.

In some situations, it may be impractical to manufacture a test jig. For example where

the accessibility of the monitor is poor, or where the monitor is of an unusual one-off

design. In such situations, the Test House may choose to use a hand held source,

normally incorporated into a rod. The Employer and the Test House should ensure

that the doses to the operator are as low as reasonably practicable (ALARP), especially

where the activity of the test sources is in excess of the equivalent of 1 MBq of 137

Cs.

For the high dose rate test, a dose rate indication of the order of 10 mSv h-1

is required.

This dose rate would be achieved by a projector style exposure unit, containing a

source of at least 10 MBq of 137

Cs, which is available from a number of

manufacturers. This unit has the additional benefits of clearly showing when the

source is exposed, and will include some radiation interlocks. A typical unit will be

portable with a mass of approximately 10 kg.

Measurement Good Practice Guide No. 29

59

Where a high dose rate is not required, a lightweight unit incorporating a less

penetrating radionuclide source may suffice. Such a unit would utilise the 60 keV

photon emissions from an 241

Am source. A source with an activity less than 1 GBq

should provide dose rates up to 1 mSv h-1

to the detector. The test radionuclide is

especially useful for uncompensated GM detectors where the response of the detector

to 60 keV photons is considerably greater than that of a compensated detector. Due to

the low penetration of 60 keV photons, the unit will require a minimum of shielding

and should therefore be hand held with a mass of the order of 0.5 kg. However the

Employer and the Test House should ensure that the gamma monitor detectors have a

reasonable response to low energy photons, before utilising such a device. Stainless

steel housed detectors may have a poor response to such a test jig.

The required source activity can be calculated from the detector dimensions and the

maximum dose rate required. Good practice is to place the source at 3 times the

largest dimensions of the detector. Under these conditions, the detector will follow the

inverse square law with sufficient accuracy. However, for large detectors, this is

impractical, as it would lead to an inconveniently high source activity. Therefore

testing often takes place with the source much closer, and sometimes touching the

detector. There are two problems with this:

It is difficult to predict the average dose rate in the detector.

The average dose rate is very sensitive to source position.

The first problem can be addressed before installation. An example of the instrument

type should be used to determine the apparent dose rate versus distance curve. This

can then be used to derive an expected indication from the test source.

The second problem requires that the source position is sensible, i.e. not in a position

where, for example, it is partially shielded which will generate large variations in

indication for small source movements. It also requires that the position is fixed with

respect to the instrument. This can be achieved in a variety of ways, such as gluing the

source holder to the instrument case, or fixing it to the wall alongside the instrument,

or the use of a jig specific to the detector type and geometry.

Instruments are often mounted on a wall, which will produce backscatter when

exposed to gamma radiation. This should be allowed for in the in situ testing process.

The effect of backscatter will be type specific and should not vary from example to

example of the same type. All that is required is to use the test sources with the

detector in free air and then to repeat the measurements with the detector on a suitable

wall. Thick concrete typically generates about 6% backscatter, thinner materials are

less effective.

Measurement Good Practice Guide No. 29

60

Alternative methods for conducting this test use work in progress as either the source

of radiation where there is sufficient accuracy in the activity, or in combination with a

transfer standard. Where a transfer instrument is used, it can be mounted alongside the

instrument under test. This reference instrument can be a very low sensitivity GM

detector attached to a relatively simple rate meter, which has been calibrated before

any testing campaign over the range of dose rates expected. Again, it may not be

possible to make measurements over the full possible dose rate range, but if the

function of the installed instrument can be confirmed up to the point where, for

example, any overload trigger operates, then that will give reasonable confidence that

the system is operating correctly.

Where the activity of the source used during the work in progress is known with

sufficient accuracy, one of the sources can be brought close enough to the detector in

question to perform the test. There are limitations with this, in that a close approach to

a large detector produces very non-uniform detector irradiation and makes it difficult

to calculate the reference value. If this approach is to be used, the dose versus distance

curve should be established before the detector is installed.

Finally, where an instrument is used within a cell where x-radiations are used, the kV

and current that just produces full scale indication on the monitor can be determined.

These values can then be used to generate a reference indication before and after each

full output test.

61

Appendix B:

Directional Dependence

Testing of a

Gamma Monitor

Measurement Good Practice Guide No. 29

62

Gamma monitors typically utilise GM detectors, ionisation chambers or, increasingly,

silicon diode detectors. These vary in terms of the quality of their polar response at

the low energies.

Conventional energy compensated steel walled GMs and most ionisation chamber

detectors have good and predictable responses at ±90° to the calibration (reference)

orientation for 241

Am gamma radiation. Therefore it is appropriate to test these

detectors at ±90 ° (Figure 4).

Figure 5: Typical Steel Walled Energy Compensated GM Detector

However, other detectors often do not respond effectively at low energies out to 900

because of aspects of their construction which makes it difficult.

These include:

energy compensated thin end window pancake GM detectors where the

detector tends to have thick side walls

cylindrical ionisation chambers with steel walls which tend to be long and

thin

detectors with a high atomic number gas filling and a high length to diameter

ratio where it is impossible to follow the normal energy compensation

techniques

most silicon diode detectors, where the sensitive volume is in the form of a

thin disc

For these detectors, the response at ±90° for 60 keV gamma radiation or 65 keV X

radiation is often close to zero.

GM +90

Anode Pin

Reference

Orientation (0°)

Main filter

elements

-90o

Measurement Good Practice Guide No. 29

63

The potential failure mechanism which can corrupt the response is a misalignment of

one or more components of the energy compensation filter. This can result in a

reduced response at one extreme angle and an enhanced response at the opposite

angle. In such cases, the directional dependence should be tested at the largest

convenient angle for which the detector is intended to produce a response within a

factor of two of the response in the calibration orientation. The type test information

should be consulted to determine a suitable angle. Frequently, ±60° to the calibration

direction is appropriate (Fig. 6).

An example is shown below in figure 6 for a typical end window energy compensated

GM detector.

Compensation

Filter 60

o

Test

Point

Reference

Orientation (0°)

GM

64

Appendix C:

Check Radionuclides

for Contamination

Monitors in Different

Workplaces

Measurement Good Practice Guide No. 29

65

Installed instrumentation may be found in a wide variety of industrial workplaces

including hospitals, universities and nuclear power stations. Although the same model

of instrument might be installed in each workplace, it may be required to measure very

different types of radioactive contamination. The radionuclides likely to be

encountered in each workplace should be considered. Test radionuclides should be

chosen with energies at, or below, the minimum energy of importance in the work

place. For alpha contamination 241

Am is the recommended radionuclide for such tests.

However, during TBFU and Type Tests, it is important that the instrument has a good

response to alpha radiation with energies significantly lower than the energy of alpha

radiation from an 241

Am test source. Table 4 gives examples of radionuclides, which

are suitable for Periodic and Routine Tests in different workplaces.

Workplace Personnel Exit Monitors SAMs and

Portals

Alpha Beta Photon Photon

Nuclear Power Station:

Magnox

241Am

36Cl,

14C,

90Sr +

90Y

N/A 137

Cs, 60

Co

Nuclear Power Station:

AGR, PWR

241Am

36Cl,

14C,

90Sr +

90Y

55Fe

137Cs,

60Co

Reprocessing Plant 241

Am 36

Cl, 14

C, 90

Sr + 90

Y

N/A 137

Cs, 60

Co, 57

Co

Hospital N/A

14C

129I

University 241

Am 36

Cl, 14

C, 90

Sr + 90

Y

55Fe,

129I

66

References

Measurement Good Practice Guide No. 29

67

1. The Ionising Radiations Regulations, 1999, HMSO, 1999.

2. Work with ionising radiation, approved code of practice, Ionising Radiation

Regulations 1999, HMSO, 1999.

3. National Physical Laboratory, Measurement Good Practice Guide No. 82, The

Examination and Testing of Equipment for Monitoring Airborne Radioactive

Particulate in the Workplace, NPL, 2006.

4. National Physical Laboratory, Measurement Good Practice Guide No.14, The

examination, testing and calibration of portable radiation protection

instruments, NPL, 1999.

5. International Atomic Energy Agency, Calibration of Radiation Protection

Monitoring Instruments, Safety Series No. 16, IAEA, 2000.

6. British Standard, Radiation protection instrumentation. Installed dose rate

meters, warning assemblies and monitors. X and gamma radiation of energy

between 50 keV and 7 MeV, BS IEC 60532, 2010.

7. British Standard, Radiation protection instrumentation. Installed personnel

surface contamination monitoring assemblies, BS EN 61098, 2007.

8. British Standard, Radiation protection instrumentation – Alpha, beta and

alpha/beta (beta energy > 60 keV) contamination meters and monitors,

BS EN 60325, 2004.

9. International Commission on Radiological Units and Measurements, Quantities

and Units in Radiation Protection Dosimetry, ICRU Report 51, 1993.

10. British Standard, Reference sources. Calibration of surface contamination

monitors. Alpha-, beta- and photon emitters, BS ISO 8769, 2010.

11. British Standard, X and gamma reference radiations for calibrating dosemeters

and doserate meters and for determining their response as a function of photon

energy. Radiation characteristics and production methods, BS ISO 4037-1,

1996.

12. Industry Radiological Protection Co-ordination Group, The Selection of Alarm

Levels for Personnel Exit Monitors,

Measurement Good Practice Guide No. 29

68

13. International Commission on Radiological Protection, Nuclear Decay Data for

Dosimetric Calculations, ICRP Publication 107, 2008.