Circulation Journal Vol.81, May 2017 Circ J 2017; 81: 668 – 674 doi: 10.1253/circj.CJ-16-1161 have not been investigated to determine if, and how, they affect Cryo-Abl outcomes. We hypothesized that the angle of the IPVs from the LA (PV angle) and their take-off position (PV position) are related to the difficulty of cryoballoon wedging into the PVs. Methods Patients We retrospectively investigated 30 consecutive patients with PAF who underwent enhanced computed tomogra- phy (CT) to evaluate the anatomy of their LA and PVs before Cryo-Abl was performed between October 2015 and July 2016. Patients with a common PV canal were not included because a large diameter of the ostium of the PV can be an independent factor related to difficulty with Cryo-Abl. Cryo-Abl Procedure All Cryo-Abl procedures were performed under a local C ryoballoon ablation (Cryo-Abl) is a useful treat- ment for curing paroxysmal atrial fibrillation (PAF) by creating complete circular lesions around each pulmonary vein (PV). 13 In many cases, a “one shot” application successfully isolates the PV from the left atrium (LA). However, incomplete circumferential balloon con- tact to the PVs may result in residual connecting gaps, especially for the inferior PVs. When the first attempt fails, multiple freezing procedures and/or focal (“touch-up”) ablation are required to accomplish isolation of the PVs. Therefore, careful selection of patients for Cryo-Abl before the procedure may shorten the procedural time and con- tribute to reducing the cost by avoiding use of an addi- tional ablation catheter. The inferior PVs (right and left: RIPV and LIPV, respec- tively) are difficult targets for creating complete isolation with current cryoballoon technology. 3 Previous reports have proposed that several anatomical characteristics of the IPVs (oval shape of the ostium and early branching of the RIPV) inhibit the efficacy of Cryo-Abl. 46 However, the exact orientation (angulation) and position of the IPVs Received November 17, 2016; revised manuscript received January 5, 2017; accepted January 16, 2017; released online February 17, 2017 Time for primary review: 22 days Division of Cardiology, Department of Medicine, Faculty of Medicine, Kindai University, Osaka-sayama, Japan Mailing address: Takashi Kurita, MD, PhD, Division of Cardiology, Department of Medicine, Faculty of Medicine, Kindai University, 377-2 Onohigashi, Osaka-sayama 589-8511, Japan. E-mail: [email protected] ISSN-1346-9843 All rights are reserved to the Japanese Circulation Society. For permissions, please e-mail: [email protected] Particular Morphology of Inferior Pulmonary Veins and Difficulty of Cryoballoon Ablation in Patients With Paroxysmal Atrial Fibrillation Ryobun Yasuoka, MD, PhD; Takashi Kurita, MD, PhD; Yasuhito Kotake, MD, PhD; Naotaka Hashiguchi, MD, PhD; Koichiro Motoki, MD, PhD; Kazuhiro Kobuke, MD, PhD; Yoshitaka Iwanaga, MD, PhD; Shunichi Miyazaki, MD, PhD Background: The CRYO-Japan PMS study indicated that cryoballoon ablation (Cryo-Abl) has a lower acute success rate of pul- monary vein isolation (PVI) for the right and left inferior PVs (RIPV and LIPV, respectively) than for the superior PVs. This study aimed to determine if the orientation and position of the inferior PVs are related to the difficulty of acute success of PVI. Methods and Results: We investigated 30 consecutive patients who underwent Cryo-Abl. A “difficult PV” was defined as the requirement for >2 cooling applications and/or touch-up ablation to achieve PVI. We measured the ventral angle between the verti- cal line and the direction of each PV trunk (PV angle) on the transverse plane of enhanced CT images. PV position was defined as the difference in the levels between the bottom of the RIPVs and the non-coronary cusp of the aorta. PV angle <105° and PV position <1.250 mm were independent factors of difficult RIPV isolation (PV angle: odds ratio (OR)=23.80, confidence interval (CI) −3.15528 to −0.53622, P=0.002; PV position: OR=12.14, CI −2.77301 to −0.23160, P=0.014). PV position <16.875 mm was also related to the difficulty of LIPV isolation (OR=5.78, CI −1.77095 to −0.09474, P=0.027). Conclusions: RIPV with ventral orientation may require difficult maneuvers to advance an ablation system towards it. Low take-off of the inferior PVs may cause non-coaxial configuration of balloon catheters towards the direction of these veins. Key Words: Anatomy; Atrial fibrillation; Cryoballoon ablation; Pulmonary veins ORIGINAL ARTICLE Arrhythmia/Electrophysiology

Transcript of Particular Morphology of Inferior Pulmonary Veins and ...

Circulation Journal Vol.81, May 2017

668 YASUOKA R et al.Circ J 2017; 81: 668 – 674doi: 10.1253/circj.CJ-16-1161

have not been investigated to determine if, and how, they affect Cryo-Abl outcomes.

We hypothesized that the angle of the IPVs from the LA (PV angle) and their take-off position (PV position) are related to the difficulty of cryoballoon wedging into the PVs.

MethodsPatientsWe retrospectively investigated 30 consecutive patients with PAF who underwent enhanced computed tomogra-phy (CT) to evaluate the anatomy of their LA and PVs before Cryo-Abl was performed between October 2015 and July 2016. Patients with a common PV canal were not included because a large diameter of the ostium of the PV can be an independent factor related to difficulty with Cryo-Abl.

Cryo-Abl ProcedureAll Cryo-Abl procedures were performed under a local

C ryoballoon ablation (Cryo-Abl) is a useful treat-ment for curing paroxysmal atrial fibrillation (PAF) by creating complete circular lesions around

each pulmonary vein (PV).1–3 In many cases, a “one shot” application successfully isolates the PV from the left atrium (LA). However, incomplete circumferential balloon con-tact to the PVs may result in residual connecting gaps, especially for the inferior PVs. When the first attempt fails, multiple freezing procedures and/or focal (“touch-up”) ablation are required to accomplish isolation of the PVs. Therefore, careful selection of patients for Cryo-Abl before the procedure may shorten the procedural time and con-tribute to reducing the cost by avoiding use of an addi-tional ablation catheter.

The inferior PVs (right and left: RIPV and LIPV, respec-tively) are difficult targets for creating complete isolation with current cryoballoon technology.3 Previous reports have proposed that several anatomical characteristics of the IPVs (oval shape of the ostium and early branching of the RIPV) inhibit the efficacy of Cryo-Abl.4–6 However, the exact orientation (angulation) and position of the IPVs

Received November 17, 2016; revised manuscript received January 5, 2017; accepted January 16, 2017; released online February 17, 2017 Time for primary review: 22 days

Division of Cardiology, Department of Medicine, Faculty of Medicine, Kindai University, Osaka-sayama, JapanMailing address: Takashi Kurita, MD, PhD, Division of Cardiology, Department of Medicine, Faculty of Medicine, Kindai

University, 377-2 Onohigashi, Osaka-sayama 589-8511, Japan. E-mail: [email protected] All rights are reserved to the Japanese Circulation Society. For permissions, please e-mail: [email protected]

Particular Morphology of Inferior Pulmonary Veins and Difficulty of Cryoballoon Ablation in Patients With

Paroxysmal Atrial Fibrillation

Ryobun Yasuoka, MD, PhD; Takashi Kurita, MD, PhD; Yasuhito Kotake, MD, PhD; Naotaka Hashiguchi, MD, PhD; Koichiro Motoki, MD, PhD; Kazuhiro Kobuke, MD, PhD;

Yoshitaka Iwanaga, MD, PhD; Shunichi Miyazaki, MD, PhD

Background: The CRYO-Japan PMS study indicated that cryoballoon ablation (Cryo-Abl) has a lower acute success rate of pul-monary vein isolation (PVI) for the right and left inferior PVs (RIPV and LIPV, respectively) than for the superior PVs. This study aimed to determine if the orientation and position of the inferior PVs are related to the difficulty of acute success of PVI.

Methods and Results: We investigated 30 consecutive patients who underwent Cryo-Abl. A “difficult PV” was defined as the requirement for >2 cooling applications and/or touch-up ablation to achieve PVI. We measured the ventral angle between the verti-cal line and the direction of each PV trunk (PV angle) on the transverse plane of enhanced CT images. PV position was defined as the difference in the levels between the bottom of the RIPVs and the non-coronary cusp of the aorta. PV angle <105° and PV position <1.250 mm were independent factors of difficult RIPV isolation (PV angle: odds ratio (OR)=23.80, confidence interval (CI) −3.15528 to −0.53622, P=0.002; PV position: OR=12.14, CI −2.77301 to −0.23160, P=0.014). PV position <16.875 mm was also related to the difficulty of LIPV isolation (OR=5.78, CI −1.77095 to −0.09474, P=0.027).

Conclusions: RIPV with ventral orientation may require difficult maneuvers to advance an ablation system towards it. Low take-off of the inferior PVs may cause non-coaxial configuration of balloon catheters towards the direction of these veins.

Key Words: Anatomy; Atrial fibrillation; Cryoballoon ablation; Pulmonary veins

ORIGINAL ARTICLEArrhythmia/Electrophysiology

Circulation Journal Vol.81, May 2017

669IPVs and Cryoballoon Ablation

avoid injury to the nerve. If PV isolation was still not achieved with several freezing procedures, point lesions in the connecting gap (touch-up ablation) were added using an irrigated ablation catheter (Flexability; SJM) until achievement of PV isolation.

Five operators (operators A–E) participated in the pres-ent study. Every procedure was performed by a team of 2 or 3 operators. All operators’ agreement was required for switching to a touch-up radiofrequency ablation when the first cooling application failed.

Definition of a “Difficult PV”Requirement for multiple freezing procedures and/or the touch-up ablation procedure to create complete PV isola-tion was defined as a difficult PV. Other PVs were classified as easy. Early termination of freezing because of a decline in esophageal temperature and local excessive cooling were not included as difficult PVs.

Enhanced CT ImagingA standard method for imaging and enhancement of CT was used in the present study. An ECG-gated 64-slice multidetector CT system (GE Health Care) was used to perform CT with the test bolus tracking method.8 Administration of propranolol or landiolol intravenously was performed prior to CT scanning to reduce heart rate. To estimate scan timing, a test bolus of contrast agent (10 mL of Iopamiron370®) was administered. The main bolus of contrast agent (0.8 mL/kg of Iopamiron370®) was then injected. The scan was started at the time of maximum enhancement of the ascending aorta. The scan parameters were as follows: 120 kV, 500–700 mA, 0.16 helical pitches, 350 ms gantry rotation time, and 0.625×0.625-mm slice thickness by interval.

We constructed a CT image at 75% of the R-R interval to reduce the influence of geometrical movement. We measured the ostial dimensions, PV angle, and PV position on CT images using a general medical recording system

and general anesthetic (lidocaine and dexmedetomidine). A 20-polar catheter was inserted into the coronary sinus and a 10-polar catheter was inserted into the right ventri-cle. After transseptal puncture, a deflectable circumferen-tial catheter (Lasso catheter; Biosense Webster or Optima; St. Jude Medical [SJM]) was delivered to the PV or LA through an 8.5F sheath. A 28-mm cryoballoon (Arctic Front Advance; Medtronic, CA, USA) was also inserted in the LA guided by an 11F FlexCath sheath (CryoCath; Medtronic). PV/LA angiography was performed to identify the anatomy of the PV using standard angulations (right anterior oblique view 0° and left anterior oblique view 55°). Pre-ablation voltage mapping was performed by a circumferential catheter on a 3D mapping system (Ensite System Velocity; SJM).

The cryoballoon was inflated within the LA and advanced into the ostium of each PV guided by a particu-lar ring wire (Achieve; Medtronic). Retrograde PV angiog-raphy was performed to verify complete PV occlusion by the cryoballoon. We performed Cryo-Abl in sequential order from the left superior PV (LSPV), LIPV, right supe-rior PV (RSPV), and RIPV. In all cases, freezing was continued for 180 or 200 s, except in cases of achieving local excessive cooling (<−60°C), delayed cooling (>−40°C until 60 s), or a decline in temperature of the esophagus (<25°C).

We defined successful isolation of the PVs as follows: (1) disappearance of PV potentials that were recorded by a guiding electrode (Achieve) during freezing, (2) absence of PV potentials confirmed by a circumferential catheter after freezing, and (3) no conduction from the PV to LA during high-output pacing from a circumferential catheter placed in each PV. In cases of failure of PV isolation, freezing was repeated by changing the position of the guiding catheter (Achieve) to another PV branch, and/or using the pull-down technique.7 During ablation of the right-sided PVs, stimulation of the phrenic nerve by high-output pacing from a catheter in the superior vena cava was continued to

Figure 1. Orientation of the inferior pul-monary veins (definitions of PV angle and PV position). The PV angle was defined as the ventral angle between the vertical line and the direction of each PV trunk (A,D). Case 1 shows ventral orien-tation of the RIPV in which the angle is 95° (A). Case 2 shows dorsal orientation of the RIPV in which the angle is 125° (D). In case 1, the level of the RIPV (B) was 21 slices lower than the non-coro-nary cusp (C), and the position of the RIPV was calculated as −13 mm. In case 2, the level of the RIPV (E) was 8 slices higher than the non-coronary cusp (F), and the RIPV position was calculated as 5 mm. LIPV, left inferior PV; RIPV, right inferior PV.

Circulation Journal Vol.81, May 2017

670 YASUOKA R et al.

depend on the anatomy of the PVs or LA. The longitudinal line of the PV trunk was drawn between the 2 distinct middle points of proximal and distal traverse (intersection) lines on a PV trunk. The proximal traverse line was placed on the PV ostium as previously defined, and the distal traverse line was placed approximately 1 cm distal to the ostium. If a branch bifurcated within 1 cm distal to the ostium (defined as early branching), the distal traverse line was placed just proximal to the bifurcation.

Figure 1 shows a representative case (patient 1) of ventral orientation of the RIPV in which the angle was measured as 95° (Figure 1A). Figure 1 also shows dorsal orientation of the RIPV in which the angle was measured as 125° (patient 2, Figure 1D). The ostial diameters of the 2 PVs were also measured at the same level on the image.

The craniocaudal position of IPVs was also investigated on transverse CT images. We selected a non-coronary cusp of the aorta as a reference for the position of the PV, and defined the take-off level of the IPVs by measuring the dif-ference in vertical level between the bottom of the IPVs (Figure 1B,E) and the bottom of the non-coronary cusp (Figure 1C,F) by manually moving the transverse images. In case 1, the level of the RIPV (Figure 1B) was 21 slices caudal to the non-coronary cusp (Figure 1C). The RIPV position could then be calculated as −13 mm. In case 2, the level of the RIPV (Figure 1E) was 8 slices cranial to the non-coronary cusp (RIPV position was calculated as +5 mm, Figure 1F).

Statistical AnalysisAll data were repeatedly measured and analyzed without knowledge of the patients’ clinical backgrounds. Univariate and multivariate logistic regression analyses were per-formed to evaluate the relationships between the anatomi-cal characteristics of the IPVS (PV angle, PV position, early branching, and PV ostial dimensions) and the results of Cryo-Abl. Analysis using a receiver-operating charac-teristic (ROC) curve was performed to identify an appro-priate cutoff value of the CT parameters with which to discriminate patients with a difficult PV for Cryo-Abl. Statistical analysis was performed with JMP software (SAS Institute Inc., Cary, NC, USA). To evaluate the intra- and interobserver reliability, the PV angle was mea-sured twice by 2 independent cardiologists who were blinded to the results of Cryo-Abl. In 15 randomly selected cases, intraclass correlation coefficients (ICC) were calcu-lated using IBM SPSS Statistics 21.0. Continuous variables are presented as mean ± standard deviation. Categorical variables are presented as number and percentage. P<0.05 was considered statistically significant.

ResultsClinical and Echocardiographic VariablesClinical and echocardiographic variables of the 30 patients are shown in Table 1. No patients had concomitant heart disease, such as valvular heart disease, congenital heart disease, or cardiomyopathy.

Results of Cryo-AblThe overall number of applications of Cryo-Abl was 2.0±1.2 in the LIPV and 1.6±1.2 in the RIPV. The overall duration of applications of Cryo-Abl was 229±148 s in the RIPV and 265±167 s in the LIPV (Table 2). A difficult PV was observed in 12/30 (40.0%) patients for the RIPV and

(Synapse; Fujifilm, Tokyo, Japan).

Orientation of IPVs (Definitions of PV Angle and Position)The PV angle was measured on enhanced CT images in the 2 IPVs. The ostium of the IPVs was defined as the inter-section between the PV trunk and the posterior LA wall on a traverse plane. On the transverse plane at the level where the ostial diameter of the RIPV or LIPV was maximal, we measured the ventral angle between the vertical line and the direction (longitudinal line) of each PV trunk (Figure 1). The vertical line was perpendicular to both the transverse and sagittal plane on CT images, so it did not

Table 1. Baseline Characteristics of Patients Undergoing Cryo-Abl

Total

Age (years) 59±14

Sex (M/F) 23/7

Height (cm) 166±8    Weight (kg) 66±11

CHA2DS2-VASc 1.5±1.2

CHF 1 (3.3%)

Hypertension 13 (43.3%)

Age ≥75 years 3 (10.0%)

Diabetes 3 (10.0%)

Stroke 2 (6.7%)

Vascular disease 3 (10.0%)

Echocardiography

LA dimensions 40±5   LVEF 67±7  

Data are mean ± standard deviation (SD) or n (%). CHF, congestive heart failure; Cryo-Abl, cryoballoon ablation; LA, left atrial; LVEF, left ventricular ejection fraction.

Table 2. Results of Cryo-Abl in Patients With Paroxysmal Atrial Fibrillation

nFreezing

Touch-upTime (min) Frequency

Total

LIPV 30 265±167 2.0±1.2 2 (6.6%)   RIPV 30 229±148 1.6±1.2 7 (23.3%)

Easy PV

LIPV 19 180±37   1.4±0.6 0

RIPV 18 174±41   1.1±0.3 0

Difficult PV

LIPV 11 412±199 3.0±1.2 2 (20.0%)

RIPV 12 339±187 2.4±1.5 7 (58.3%)

Cryo-Abl, cryoballoon ablation; LIPV, left inferior pulmonary vein; RIPV, right inferior pulmonary vein.

Table 3. Geographic Parameters of Inferior Pulmonary Veins

LIPV RIPV

Dimension (mm) 12±2 14±3

Angle (°) 110±12 109±16

Position (mm) 17±10 2±10

Early branching 0 (0.0%) 17 (56.7%)

Abbreviations as in Table 2.

Circulation Journal Vol.81, May 2017

671IPVs and Cryoballoon Ablation

ostial dimension of the LIPV was 12±2 mm (range: 8–15; median: 11 mm) and that of the RIPV was 14±3 mm (range: 10–21; median: 14 mm). The PV angle of the LIPV was 110°±12° (range: 77°–130°; median: 111°) and that of the RIPV was 109°±16° (range: 74°–136°; median: 72°). The position of the LIPV was 17±10 mm (range: −6 to 36; median: 17 mm) and that of the RIPV was 2±10 mm (range: −14 to 24; median: 2 mm) (Table 3).

Early branching was observed in 17 of 30 (56.7%) RIPV, but none of the patients had early branching of the LIPV (Table 3).

Intra- and Interobserver Reliability for the RIPV AngleFor the RIPV angle, intraobserver reliability (ICC: 1,1) and interobserver reliability (ICC: 2,1) showed almost per-fect agreement (r=0.927 and r=0.912, respectively).

Relationships Between CT Parameters and Difficulty of Cryo-Abl for the RIPVFigure 2A–C shows the relationships between CT param-eters and difficulty of Cryo-Abl for the RIPV. The dimen-sion of the RIPV was not significantly different between a difficult and an easy PV (13±2 vs. 15±3 mm, P=0.08). The PV angle of difficult RIPVs was significantly narrower than that of easy RIPVs (98°±14° vs. 116°±14°, P=0.002).

in 11/30 (33.3%) patients for the LIPV (Table 2). Touch-up ablation was required in 7/30 (23.3%) patients for the RIPV and in 2/30 (6.7%) patients for the LIPV. Conduction gaps after Cryo-Abl in the RIPV were observed at the anterior in 2, carina in 3, bottom in 3, and posterior in 3 patients. Conduction gaps in the LIPV were observed at both the carina and bottom in 1 patient, and posterior in 1 patient. Multiple freezing and touch-up ablation were not necessary for either of the SPVs. More frequent cryo appli-cations and a longer duration of freezing applications of Cryo-Abl were required with a difficult RIPV than an easy RIPV (2.4±1.5 vs. 1.1±0.3, P=0.002; 339±187 vs. 174±41s, P=0.002, respectively). More frequent cryo applications and a longer freezing time of applications of Cryo-Abl were required with a difficult LIPV than with an easy LIPV (3.0±1.2 vs. 1.4±0.6, P<0.001; 412±199 vs. 180±37s, respec-tively, P<0.001). Cryo-Abl procedures were mainly per-formed by operator A in 4 (13.3%), B in 4 (13.3%), C in 5 (16.7%), D in 10 (33.3%), and E in 6 (20%) cases, respec-tively.

Geographic Parameters of IPVs on CT ImagesPatients with a common ostium of the superior and infe-rior PVs were not included in this study. The middle branch of the right PV was observed in 2 patients. The

Figure 2. Relationships between CT parameters and difficulty of Cryo-Abl for the right inferior pulmonary vein (RIPV). The dimen-sions of the RIPV were not associated with difficulty isolation (A). The RIPV angle was significantly narrower in a difficult case than in an easy case (B). The position of the RIPV was also related to difficulty with isolation (C).

Figure 3. ROC analysis of CT parameters and difficulty of Cryo-Abl. RIPV angle cutoff of 105° discriminated a difficult PV with a sensitivity of 75% and specificity of 89% (A). RIPV position cutoff of 1.250 mm discriminated a difficult RIPV with a sensitivity of 83.3% and specificity of 72.2% (B). LIPV position cutoff of 16.875 mm discriminated a difficult LIPV with a sensitivity of 81.8% and specific-ity of 68.4% (C). LIPV, left inferior pulmonary vein; RIPV, right inferior pulmonary vein; ROC, receiver-operating characteristic.

Circulation Journal Vol.81, May 2017

672 YASUOKA R et al.

take-off position (cutoff value <1.250 mm) and ventral PV angle (cutoff value <105°) were significant parameters for indicating difficulty of RIPV isolation (low take-off RIPV position: odds ratio (OR)=7.80, 95% confidence interval (CI) −1.93640 to −0.23963, P=0.010; ventral PV angle: OR=16.00, 95% CI −2.47058 to −0.51350, P=0.001). Early branching was not associated with difficulty of RIPV isola-tion (RIPV: OR=2.00, 95% CI −1.14276 to 0.39676, P=0.364, Table 4). When PV angle and position were ana-lyzed as categorical variables, multivariate logistic analysis showed low take-off position (cutoff value <1.250 mm) and ventral PV angle (cutoff value <105°) were significant parameters for identifying difficulty of RIPV isolation (low take-off RIPV: OR=12.14, 95% CI −2.77301 to 0.23160,

The position of difficult RIPVs was significantly more cau-dal (lower) than that of easy RIPVs (−3±6 vs. 5±10 mm, P=0.034). ROC analysis (area under the curve [AUC], 0.83) showed that a PV angle cutoff of 105° discriminated a difficult PV with a sensitivity of 75% and specificity of 89% (Figure 3A). ROC analysis (AUC, 0.76) showed that a RIPV position cutoff of 1.250 mm discriminated a diffi-cult RIPV with a sensitivity of 83.3% and specificity of 72.2% (Figure 3B). In patients with both of these factors predicting difficulty of Cryo-Abl (PV angle <105° and a low take-off position <1.250 mm), multiple applications of the freezing procedure were required in 5/6 (83%) patients and the touch-up procedure was required in 2/6 (33%) patients. Univariate logistic analysis showed that the low

Table 4. Logistic Analysis of Ccomputed Tomography Parameters and Difficulty of Cryo-Abl

OR 95% CI P value

Univariate

LIPV

Continuous variable

PV position 1.10 0.01004 to 0.19485 0.027

Categorical variable

Low take-off (<16.875 mm) 5.78 −1.77095 to −0.09474 0.027

RIPV

Continuous variable

PV angle 1.09 0.02967 to 0.16959 0.002

PV position 1.11 0.01174 to 0.22220 0.025

Categorical variable

Early branching (+) 2.00 −1.14276 to 0.39676   0.364

Ventral orientation (<105°) 16.00 −2.47058 to −0.51350 0.001

Low take-off (<1.250 mm) 7.80 −1.93640 to −0.23963 0.010

Multivariate

LIPV

Low take-off (<16.875 mm) – – –

RIPV

Continuous variables

PV angle 1.08 0.01535 to 0.15762 0.012

PV position 1.07 −0.04187 to 0.20443   0.222

Categorical variables

Ventral orientation (<105°) 23.80 −3.15528 to −0.53622 0.002

Low take-off (<1.250 mm) 12.14 −2.77301 to −0.23160 0.014

Abbreviations as in Table 2.

Figure 4. Relationships between CT parameters and difficulty of Cryo-Abl for the LIPV. The dimensions of the LIPV and PV angle in a difficult case were not significantly different from those in an easy case (A,B). The take-off position in difficult LIPVs was sig-nificantly lower than in easy LIPVs (C). LIPV, left inferior pulmonary vein.

Circulation Journal Vol.81, May 2017

673IPVs and Cryoballoon Ablation

uating the orientation of the PV.The CRYO-Japan PMS study demonstrated that IPVs

are harder targets for achieving PV isolation using the Cryo-Abl technique compared with SPVs.3 In the present study, we found that narrow angulation (ventral orienta-tion) of the RIPV and low take-off IPVs were anatomical characteristics that resulted in incomplete PV isolation using Cryo-Abl.

Ventral orientation of the RIPV (parallel to the atrial septal wall) causes difficulty in advancing the cryoballoon catheter towards the ostium of the RIPV by creating acute angulation of the relatively stiff catheter. However, in an RIPV with a wide angle (i.e., dorsal orientation), the cryo-balloon catheter can be easily advanced to the RIPV because a transseptal catheter tends to advance vertically to the atrial septum. Another possible factor that affects manipulation during Cryo-Abl is the distance between the atrial septum and each PV. A previous study, which evalu-ated the distance between the fossa ovalis and each PV using MRI, showed that the mean distance was shortest in the RIPV (20.2 mm) among the 4 PVs.9 The longitudinal diameter of a fully expanded cryoballoon is approximately 24 mm, which may be inadequately large, especially for patients with ventral orientation of the RIPV because the PV ostium deviates closely to the atrial septum. Although the RSPV shows a ventral orientation in almost all patients, isolation of this PV using a cryoballoon is much easier than for the RIPV. This may be related to the ade-quate distance between the atrial septal puncture site and the ostium of the RSPV.

PV Position and Difficulty of Cryo-AblIn the present study, we showed that a low take-off (caudal position) of the IPVs was associated with difficulty of Cryo-Abl. The bottom of the aortic non-coronary cusp was defined as the reference for PV position. This defini-tion was chosen because the cusp is one of the traditional markers for identifying the optimal site (fossa ovalis) for the transseptal approach.10 When the IPVs are located more caudally, acute downward bending of the catheter shaft from the fossa ovalis is required. In this situation, we could not avoid producing non-coaxial configuration of the catheter shaft along the IPVs, which may result in incomplete contact, particularly to the bottom of the PV.7

Clinical ImplicationsIn our study, touch-up RF ablation was necessary in 7/30 (23.3%) RIPVs and 2/30 (6.7%) LIPVs. Our novel analysis for predicting a difficult PV for Cryo-Abl based on CT images will be useful for selecting appropriate patients for this strategy, and avoiding having to use multiple ablation catheters.

Study LimitationsFirst, the design of our study was retrospective in a single institution with a small number of patients. Particularly, the number of patients is the ultimate limitation. Our results must be interpreted as limited data from an initial 30 patients. Our data should be reconfirmed by a multi-center study including a large number of patients. Second, we did not evaluate the long-term results of Cryo-Abl because official use of this novel technique in Japan was only approved in February 2014. In the future, we should address the relationship of PV anatomy and results of long-term follow-up by showing the exact site with re-

P=0.014; ventral PV angle: OR=23.80, 95% CI −3.15528 to −0.53622, P=0.002, Table 4). In multivariate logistic analysis of these parameters as continuous variables, only PV angle was a significant parameter for identifying diffi-culty of RIPV isolation (RIPV position: OR=1.07, 95% CI −0.04187 to 0.20443, P=0.222; RIPV angle: OR=1.08, 95% CI 0.01535 to 0.15762, P=0.012). We also analyzed the echocardiographic data, but left atrial dimensions (LAD) and left ventricular ejection fraction (LVEF) were not significantly different between difficult and easy RIPVs (LAD: 38±4 vs. 40±6 mm, P=0.333; LVEF: 68%±6% vs. 66%±8%, P=0.505).

Relationships Between CT Parameters and Difficulty of Cryo-Abl for the LIPVFigure 4A–C shows the relationships between CT param-eters and difficulty of Cryo-Abl for the LIPV. The dimen-sion and PV angle of difficult LIPVs were not significantly different compared with easy LIPVs (12±2 vs. 12±2 mm, P=0.804; 115°±7° vs. 108°±13°, respectively, P=0.106). The take-off position of difficult LIPVs was significantly lower than for easy LIPVs (11±8 vs. 20±10 mm, P=0.032). ROC analysis (AUC, 0.739) showed that an LIPV position cutoff of 16.875 mm discriminated a difficult LIPV with a sensitivity of 81.8% and specificity of 68.4% (Figure 3C). In the univariate analysis, a low take-off level of the LIPV (<16.875 mm) was also related to difficulty of LIPV isola-tion (OR=5.78, 95% CI −1.77095 to −0.09474, P=0.027, Table 4). In the univariate analysis, LIPV position was related to difficulty of LIPV isolation (OR=1.10, 95% CI 0.01004 to 0.19485, P=0.027, Table 4).

Similar to the RIPV, the LAD and LVEF were not sig-nificantly different between difficult and easy LIPVs (LAD: 41±6 vs. 39±4 mm, P=0.281; LVEF: 67%±6% vs. 67%±7%, P=0.941).

DiscussionThe major finding of this study was that the PV angle and take-off level of inferior PVs were significantly associated with difficulty in PV isolation using Cryo-Abl. In the case of an RIPV angle <105°, the requirement of multiple freez-ing procedures and/or touch-up catheter ablation (defined as a difficult PV) was approximately 23-fold higher than in cases of an angle ≤105°. In the case of an RIPV position <1.250 mm, the rate of a difficult PV was approximately 12-fold higher than in the case of an RIPV position ≥1.250 mm. In the case of an LIPV position <16.875 mm, the rate of a difficult PV was approximately 6-fold higher than in the case of an LIPV position ≥16.875 mm.

PV Angle and Difficulty of Cryo-AblPrevious reports have demonstrated that several anatomi-cal characteristics of the PVs (oval shape, early branching, and dull-edged shape) result in difficult maneuvers using cryo-ablation techniques.4–6 Only one study has proposed a relationship between the angle of the RIPV and acute success of RIPV isolation.6 In this study, the non-perpen-dicular angle between the axis of the PV and the ostial plane was a parameter that suggested difficulty of RIPV isolation. However, on some CT images, determination of the PV ostial plane is not easy because of obtuse (gradual) transition of the RIPV from the LA. In the present study, we drew a vertical line as an axial reference to the PV angle, which provided a distinct and easy method for eval-

Circulation Journal Vol.81, May 2017

674 YASUOKA R et al.

DisclosuresThe authors did not receive any grants concerning this study.

References 1. Packer DL, Kowal RC, Wheelan KR, Irwin JM, Champagne J,

Guerra PG, et al. Cryoballoon ablation of pulmonary veins for paroxysmal atrial fibrillation. J Am Coll Cardiol 2013; 61: 1713 – 1723.

2. Kuck KH, Brugada J, Fürnkranz A, Metzner A, Ouyang F, Chun KR, et al; FIRE AND ICE Investigators. Cryoballoon or radiofrequency ablation for paroxysmal atrial fibrillation. N Engl J Med 2016; 374: 2235 – 2245.

3. Okumura K, Matsumoto K, Kobayashi Y, Nogami A, Hokanson RB, Kueffer F; CRYO-Japan PMS Study Investigators. Safety and efficacy of cryoballoon ablation for paroxysmal atrial fibril-lation in Japan. Circ J 2016; 80: 1744 – 1749.

4. Sorgente A, Chierchia GB, de Asmundis C, Sarkozy A, Namdar M, Capulzini L, et al. Pulmonary vein ostium shape and orienta-tion as possible predictors of occlusion in patients with drug-refractory paroxysmal atrial fibrillation undergoing cryoballoon ablation. Europace 2011; 13: 205 – 212.

5. Schmidt M, Dorwarth U, Straube F, Daccarett M, Rieber J, Wankerl M, et al. Cryoballoon in AF ablation: Impact of PV ovality on AF recurrence. Int J Cardiol 2013; 167: 114 – 120.

6. Knecht S, Kühne M, Altmann D, Ammann P, Schaer B, Osswald S, et al. Anatomical predictors for acute and mid-term success of cryoballoon ablation of atrial fibrillation using the 28 mm bal-loon. J Cardiovasc Electrophysiol 2013; 24: 132 – 138.

7. Chun KR, Schmidt B, Metzner A, Tilz R, Zerm T, Köster I, et al. The ‘single big cryoballoon’ technique for acute pulmonary vein isolation in patients with paroxysmal atrial fibrillation: A prospective observational single centre study. Eur Heart J 2009; 30: 699 – 709.

8. Yamaguchi T, Takahashi D. Development of test bolus tracking method and usefulness in coronary CT angiography. Nihon Hoshasen Gijutsu Gakkai Zasshi 2009; 65: 1032 – 1040 (in Japanese).

9. Schmidt B, Ernst S, Ouyang F, Chun KR, Broemel T, Bänsch D, et al. External and endoluminal analysis of left atrial anatomy and the pulmonary veins in three-dimensional reconstructions of magnetic resonance angiography: The full insight from inside. J Cardiovasc Electrophysiol 2006; 17: 957 – 964.

10. Croft CH, Lipscomb K. Modified technique of transseptal left heart catheterization. J Am Coll Cardiol 1985; 5: 904 – 910.

connection of the PV to the LA. Third, because we enrolled patients from the beginning of our experience with this procedure, our data should be interpreted with caution because there would have been a learning curve. The rela-tively high percentage of touch-up RF ablation might be caused to inexperience with the technique in the initial phase of our study. Because tips and techniques based on experience may improve achievement of PV isolation by Cryo-Abl, our results will change with further training and learning. However, our data may be useful especially for beginners who have not yet become highly skilled for Cryo-Abl to assist with predicting difficulty and to prepare back-up procedures (e.g., touch-up methods).

Fourth, the position of the IPVs should be defined using the septal punctured site as a reference, not a non-coronary cusp. However, precise identification of the puncture site on pre-procedural CT images was impossible in our study. Finally, definition of a difficult PV is the limitation of the present study because some operators may consider that PVs requiring 2cooling procedures were not difficult. In the CRYO-Japan PMS study, additional (touch-up) ablation was allowed if PVI was not achieved by >1 Cryo-Abl application. When the first cooling procedure is insuf-ficient, one operator may continue the balloon proce-dure further, but another may select touch-up ablation. Accordingly, if we defined a difficult PV as requirement for >3 applications, we may produce a selection bias that depends on the operator’s personal choice.

ConclusionsAnatomical characteristics (ventral orientation of the RIPV and low take-off of IPVs) may require difficult maneuvers to achieve complete PV isolation using Cryo-Abl. A tailored ablation strategy based on the anatomical features of the PVs may allow a reduction in procedural time and medical costs by avoiding additional use of touch-up ablation catheters.