CONFIDENTIAL & PROPRIETARY DO NOT DUPLICATE OR DISTRIBUTE CONFIDENTIAL & PROPRIETARY DO NOT DUPLICATE OR DISTRIBUTE
R O M E O® 2 PAD
BIOMECANIC and CARACTERISTIC
P O S T E R I O R A X I A L D E V I C E
Gianluca Maestretti CO-CHIEF OF ORTHOEPEDICS DEPARTEMENT
CHIEF OF SPINAL UNIT
Hospital Cantonal Fribourg
CONFIDENTIAL & PROPRIETARY DO NOT DUPLICATE OR DISTRIBUTE
1.Titanium plate
2.Central PEEK cage
3.Titanium plate
4.Revision screw
The implant consists of:
C O N C E P T & D E S I G N
CENTRAL PEEK CAGE:
The PEEK cage in the central part of the implant is designed to help the fusion process, like the interbody PEEK cages for lumbar interbody fusion.
The graft window in the middle offers the possibility to insert allograft or autograft.
CONFIDENTIAL & PROPRIETARY DO NOT DUPLICATE OR DISTRIBUTE
C O N C E P T & D E S I G N
CONFIDENTIAL & PROPRIETARY DO NOT DUPLICATE OR DISTRIBUTE CONFIDENTIAL & PROPRIETARY DO NOT DUPLICATE OR DISTRIBUTE
Purpose of the study:
Evaluate the mechanical properties and performances of the ROMEO®2 PAD
interspinous device (Posterior Axial Device) by the mean of static and dynamic tests in a corpectomy model according to the principle of ASTM F1717-11
Additional Dynamic Compression-tension test performed to characterize the attachment of the device to the spinous process
CONFIDENTIAL & PROPRIETARY DO NOT DUPLICATE OR DISTRIBUTE CONFIDENTIAL & PROPRIETARY DO NOT DUPLICATE OR DISTRIBUTE
List of predicate devices:
SPIRE PLATE (Medtronic) K032037
ASPEN (Lanx) K071877
AFIX (Nuvasive) K073278
AXLE PEEK (X-Spine) K112592
CONFIDENTIAL & PROPRIETARY DO NOT DUPLICATE OR DISTRIBUTE CONFIDENTIAL & PROPRIETARY DO NOT DUPLICATE OR DISTRIBUTE
Static compressive tests on subassemblies (H08 & H18)
115 kg
CONFIDENTIAL & PROPRIETARY DO NOT DUPLICATE OR DISTRIBUTE CONFIDENTIAL & PROPRIETARY DO NOT DUPLICATE OR DISTRIBUTE
Static compression & torsion tests / Dynamic compression test
3D view Front view Side view
40
Device assembly on testing blocks
CONFIDENTIAL & PROPRIETARY DO NOT DUPLICATE OR DISTRIBUTE
Static compression bending test
Method: Based upon ASTM F1717-11 with modified test blocks that include mock spinous process Constant displacement speed set up at 5 mm/min Vertical force applied to the assembly 40mm behind the anchors of the ROMEO®2 PAD
CONFIDENTIAL & PROPRIETARY DO NOT DUPLICATE OR DISTRIBUTE
Static compression bending test
Results: ROMEO®2 PAD yield load 60% greater than predicate(s) ROMEO®2 PAD compression stiffness 63% superior to predicate
60
40
20
0
CONFIDENTIAL & PROPRIETARY DO NOT DUPLICATE OR DISTRIBUTE
Static torsion test
Method: Based upon ASTM F1717-11 with modified test blocks that include mock spinous process Speed of rotation equal to 60°/min
CONFIDENTIAL & PROPRIETARY DO NOT DUPLICATE OR DISTRIBUTE
Static torsion test
Results: ROMEO®2 PAD yield torque at least equal to predicate
CONFIDENTIAL & PROPRIETARY DO NOT DUPLICATE OR DISTRIBUTE
Dynamic compression bending test
Method: Based upon ASTM F1717-11 with modified test blocks that include mock spinous process Duration: 5 million cycles Frequency equal to 4Hz Test performed in a sodium chloride solution at 37°C ± 2°C
CONFIDENTIAL & PROPRIETARY DO NOT DUPLICATE OR DISTRIBUTE
Dynamic compression bending test
Results: ROMEO®2 PAD run-out 67% superior to predicate
CONFIDENTIAL & PROPRIETARY DO NOT DUPLICATE OR DISTRIBUTE
Dynamic compression-tension test
Device assembly on testing blocks
CONFIDENTIAL & PROPRIETARY DO NOT DUPLICATE OR DISTRIBUTE
Method Dynamic compression-tension test
Method: ROMEO®2 PAD tested on specific blocks made of rigid polyurethane foam with grade 40 according to the standard ASTM F1839 (density = 640.7 kg/m3) simulating cortical bone Duration: 5 million cycles Frequency equal to 4Hz Device submitted to a sinusoidal load between a maximal and a minimal value
CONFIDENTIAL & PROPRIETARY DO NOT DUPLICATE OR DISTRIBUTE
Dynamic tension-compression test
Results: ROMEO®2 PAD run-out 33% superior to predicate
200
150
100
50
0
CONFIDENTIAL & PROPRIETARY DO NOT DUPLICATE OR DISTRIBUTE CONFIDENTIAL & PROPRIETARY DO NOT DUPLICATE OR DISTRIBUTE
THANK‘ YOU
P O S T E R I O R A X I A L D E V I C E