P D F - X Change w Click to buy NOW! mo w.t . c racker-s o f t w a r...
Transcript of P D F - X Change w Click to buy NOW! mo w.t . c racker-s o f t w a r...
ChapterV Conducting polymer thin film
modified electrochemical sensor
for uric acid and glucose
0.0 0.2 0.4 0.6 0.8 1.0
-20
0
20
40
60Glucose response
at SPEEK/PDMA/CuNFs/GCE
I/
A
E / V
Copper nanoflower
Click t
o buy NOW!
PDF-XChange
ww
w.tracker-software
.comClic
k to buy N
OW!PDF-XChange
ww
w.tracker-software.c
om
Chapter V: Conducting polymer............................. Uric acid and Glucose
[132]
5.1. Polyelectrolyte for uric acid detection in the presence of ascorbic
acid
Simple and rapid method for uric acid (UA) determination in the presence of
ascorbic acid (AA) and other interferents has received considerable attention [1].UA is a
primary end product of purine metabolism [2]. For a healthy person, UA concentrations
in serum and urinary excretion were 0.24-0.52 mM and 1.49-4.46 mM, respectively [3].
AA is a vital vitamin of human diet, and used for prevention and treatment of some
diseases. Thus, concentration of UA and AA in biological fluids (human urine), always
changed and reflected by body metabolism [4]. Therefore, a simple sensor for their
simultaneous detection in biological fluid or human urine is important for medical
diagnosis.
For determining UA concentration, numerous methods such as in ion-exchange
column chromatography [5], enzyme analysis [6], HPCE [7], were developed. However,
cost and complexity of these methods limited their applicability. Electrochemical
methods are quick, simple and sensitive [8]. During direct determination of UA,
interference of AA is a main problem, because their almost same oxidation potential.
High concentration of AA (as in biological fluids) often interferes during UA detection.
Thus, detection method should be more selective and sensitive especially for biological
fluids. Chemically modified electrodes were reported for determination of UA in
presence of AA. For instance, GC electrode electrochemically modified with Luminol
[9], Tyron [10], conducting polymers [11], silica gel [12], methylene blue adsorbed on a
phosphorylated zirconia-silica composite [13], thiadiazole [14], MWCNT-Nafion [15],
mesoporous carbon/Nafion composite [16], poly(Evans Blue) [17] and copper [18], were
reported in the literature. But, these methods are difficult, cost intensive, less stable and
consumed more time. UA sensors were fabricated with positively or negative charged
films, self-assembled monolayer and LBL technique using poly(dimethylsiloxane) [19],
PEDOT/SDS [20], PDDA/Uricase enzyme [21], CPB [22,23], Nafion [24] and PDDA
[25]. But, these techniques are also difficult and less sensitive. Thus, there is an
expanding demand to develop simple, reliable, sensitive, and stable sensor with improved
characteristics for simultaneously sensing of AA and UA.
Click t
o buy NOW!
PDF-XChange
ww
w.tracker-software
.comClic
k to buy N
OW!PDF-XChange
ww
w.tracker-software.c
om
Chapter V: Conducting polymer............................. Uric acid and Glucose
[133]
Polyelectrolytes were polymers whose repeating units bear an electrolyte group.
These groups will dissociate in water, making the polymers charged and properties are
thus similar to electrolytes (salts) and polymers (high molecular weight compounds).
They sometimes called polysalts, their solutions are electrically conductive. They were
widely used for: molecular self-assembly techniques for thin film deposition of
electrically conducting polymers [26], conjugated polymers for light emitting devices
[27], nanoparticles [28], in second order nonlinear optical (NLO) devices [29] and
bimolecular sensor [21]. Cationic polyelectrolyte (PDDA) was explored for simultaneous
detection of UA and AA at GC electrode, because of its electro-conductive nature and
electrostatic interaction with charged molecules [30,31].
Thus, we are exploiting the polyelectrolyte environment for simultaneously
detection of UA and AA at GC electrode by CV and DPV. Both (UA and AA) exhibited
well separated oxidation peak potentials (360 mV). This method was also validated for
human urine samples with 3.0–8.0% measuring error. Developed approach offered an
effective tool for simultaneous determination of different biomolecules.
5.1.1. Materials
Ascorbic acid (AA) and poly (diallyldimethylammonium chloride) (PDDA) (low
molecular weight, 35 wt. % in water) were received from Sigma-Aldrich, while UA was
received from Chemika-Biochemika reagent, and used without further purification. All
other chemicals of AR grade were received from S.D. Fine Chemicals (India).
5.1.2. Electrochemical studies
All electrochemical experiments were performed using a potentiostat/galvanostat
Autolab®PGSTAT-302 N (Eco Chemie B.V., The Netherlands), in single-compartment
three-electrode cell. GC electrode (3.0 mm diameter) was polished with 0.5 µm alumina
slurry on a flat surface, rinsed thoroughly with deionized water, and sonicated for 2 min
immediately before its use and worked as working electrode . Platinum wire was used as
an auxiliary electrode. All potentials were recorded using saturated calomel reference
electrode (SCE). Aqueous solutions were prepared through deionized water (ρ > 18.2
MΩ, Millipore Milli-Q system) deoxygenated by purging with nitrogen gas prior to start
of each experiment.
Click t
o buy NOW!
PDF-XChange
ww
w.tracker-software
.comClic
k to buy N
OW!PDF-XChange
ww
w.tracker-software.c
om
Chapter V: Conducting polymer............................. Uric acid and Glucose
[134]
5.1.3. Analysis of urine
The utilization of the proposed method in natural sample analysis was also
investigated by direct analysis of human urine samples, collected from our lab workers.
UA was dissolved in urine to make a stock solution (50 ml) with 5 mM concentration.
Standard successive additions of 1 µl of 5 mM UA in urine were added to the buffer pH:
4.0 containing 0.028% of PDDA.
5.1.4. Results and discussion
5.1.4.1. Electrochemical oxidation of AA and UA
Cyclic voltagramms for AA oxidation at glassy carbon electrode in 50 ml of 0.2 M
phosphate buffer (pH: 4) without or with PDDA (cationic polyelectrolyte) were
investigated (Fig. 5.1.1(A)). The electrochemical oxidation of AA is dependent on the pH
of solution. In acid solution, only oxidation peak for AA was observed, whereas in basic
solution a second oxidation peak was also observed. At pH 4.0, the oxidation of AA
involved two electrons and protons (pKa of AA: 4.17) [32]. While at high pH (>4.17),
oxidation process involved loss of a proton and two electrons due to anionic nature of AA
[32,33]. Oxidation of AA at GC electrode is irreversible and requires high over-potential.
Also, no reproducible electrode response was obtained due to electrode fouling by
adsorption of AA oxidation product. AA oxidizes at 250 mV vs SCE, while it showed
-0.4 -0.2 0.0 0.2 0.4 0.6 0.8 1.0-4
-2
0
2
4
6
A
b
a
I /
A
E / V S SCE
-0.4 -0.2 0.0 0.2 0.4 0.6 0.8 1.0
-4
0
4
8
B
b
a
I /
A
E / V S SCE
Fig. 5.1.1. Cyclic voltagramms for: (A) 0.5 mM AA and (B) 0.5 mM UA in 0.2 M phosphate
buffer solution (pH: 4.0) at 50 mV/s scan rate with: (a) without PDDA; (b) with of PDDA
solution at GC electrode.
Click t
o buy NOW!
PDF-XChange
ww
w.tracker-software
.comClic
k to buy N
OW!PDF-XChange
ww
w.tracker-software.c
om
Chapter V: Conducting polymer............................. Uric acid and Glucose
[135]
negative shift of about 50 mV with slightly enhanced peak current in the presence of
cationic PDDA. This may be attributed to favorable electrostatic attraction between the
cationic PDDA and anionic AA, which further increased with AA concentration within
the film.
At pH: 7.34 of blood and normal tissues, UA exists as urate (anionic form; pKa:
5.75) [3]. Cyclic voltagrams for UA without or with 0.028% of PDDA (50 ml of 0.2 M
PBS, pH: 4) at GC electrodes are depicted in Fig. 5.1.1B(a,b). UA oxidation was
irreversible at GC electrode or metal
electrode, while at graphite electrode it was
quasi-reversible. Electrochemical oxidation
of UA involves two electrons and protons
process that lead to formation of unstable
diimine species, which then attacked by
water molecules in a step-wise fashion and
converted to imine-alcohol and uric acid- 4,5
diol. The uric acid-4,5 diol is unstable and
decomposes, in absence of PDDA, UA
oxidized at 500 mV vs SCE, while negative
shift of about 50 mV with small increase in peak current was observed in presence of
PDDA. At pH: 7.2, UA exists as anion, which is more hydrophobic than AA because of
its low solubility of UA in water. Thus, negative shift of oxidation potential and increase
in current response were observed due to favorable electrostatic interaction between
PDDA and anionic UA as depicted in Fig. 5.1.2 [30,31,34].
The CVs of UA at different scan rates in the presence of PDDA on GC electrodes
are presented in Fig. 5.1.3(A). Oxidation peak potential was shifted to positive direction
with and linear relationship between peak oxidation current and square root of scan rate
(10-400 mV/s) (Fig. 5.1.3(B)), suggested diffusion-controlled UA oxidation in the
presence of PDDA on GC electrode.
GCE
PDDA
UA
AA
GCE
PDDA
UA
AA
Fig. 5.1.2. Schematic drawing of poly
electrolyte electrostatic interaction between
UA and AA anion.
Click t
o buy NOW!
PDF-XChange
ww
w.tracker-software
.comClic
k to buy N
OW!PDF-XChange
ww
w.tracker-software.c
om
Chapter V: Conducting polymer............................. Uric acid and Glucose
[136]
The rate constant (k) and number of electrons released during UA oxidation (n) were
determined by the following equation [35]:
)1.1.5(78.301)/(log39.201pE
where k is rate constant, n the number of electrons transferred and ΔEp is the potential
peak difference. Electrochemical oxidation of UA was proceeds by a 2e-, 2H
+
mechanism, resulting unstable diimine [15-18, 22-25]. The proposed mechanism for the
electrode reaction of UA is highlighted as follows:
ΔEp values were experimentally determined and used for estimation of k values for the
UA electrochemical reaction (Eq. 5.1.2) and presented in Table 5.1.1. In these systems
strong adsorption of reactant and product are involved, the full peak width at half-height
(EFWHM) was determined by the following equation [35]:
Number of electrons transfer (n) in the electrochemical reaction are included in Table
5.1.1., which further confirmed UA oxidation in presence of PDDA, corresponds to (Eq.
5.1.2) reaction.
-0.2 0.0 0.2 0.4 0.6
-3
0
3
6
9A
a
g
I /
A
E / V ( s SCE)
8 12 16 20 241
2
3
4
5
B
I/A
Fig. 5.1.3. (A) CVs for 0.1 mM UA in 0.2 M phosphate buffer solution (pH: 4.0) in the presence
of 0.028% of PDDA at GC electrode at different scan rates: (a) 10, (b) 20, (c) 50, (d) 100, (f)
200, (e) 300, (g) 400 mV/s and (B) The relationship between the Ip and v1/2
for UA.
)3.1.5(6.90n
EFWHM
Click t
o buy NOW!
PDF-XChange
ww
w.tracker-software
.comClic
k to buy N
OW!PDF-XChange
ww
w.tracker-software.c
om
Chapter V: Conducting polymer............................. Uric acid and Glucose
[137]
Table 5.1.1. Variation of the parameters obtained from Fig. 5.1.3, as a function of the
potential scan rate.
5.1.4.2. Effect of PDDA concentration on AA and UA oxidation
Effect of PDDA concentration on UA and AA oxidation was explored by variations
of anodic peak potential with
current (Fig. 5.1.4). Initial change
of Epa and Ipa with addition of
PDDA, may be because of: (i)
adsorption of PDDA on electrode
surface; and/or (ii) change in
electrical activity and
concentration at electrode surface.
Adsorption of PDDA was
responsible for electrode over-
potential and electron transfer rate
[23,30,31]. After complete
coverage of the electrode surface,
PDDA formed polyelectrolyte bridging in the bulk without effecting electrode oxidation
process, which was confirmed by the unexpected change in Epa and Ipa in lower PDDA
percentage.
All experiments were carried out using PDDA solution (0.028%, w/v) (sufficient to
saturate the electrode surface with PDDA charged species). Oxidation of AA or UA at
υ/mV s-1
ΔEp/mV EFWHM/mV n k/s-1
10 67 120 0.7 0.14
20 64 12 0.7 0.30
50 66 119 0.7 0.75
100 75 119 0.7 1.30
200 121 111 0.8 1.58
300 140 103 0.9 1.93
400 163 87 1.0 1.98
0.00 0.05 0.10 0.15 0.20 0.255.1
5.4
5.7
6.0
6.3
peak potential
peak current
PDDA/%
I pa/
A
0.39
0.40
0.41
0.42
0.43
0.44
0.45
Ep
a/ V
Fig. 5.1.4. Variation in anodic peak current (Ipa) and
anodic peak potential (Epa) UA (0.5 mM) with PDDA
percentages at GC electrode, respectively.
Click t
o buy NOW!
PDF-XChange
ww
w.tracker-software
.comClic
k to buy N
OW!PDF-XChange
ww
w.tracker-software.c
om
Chapter V: Conducting polymer............................. Uric acid and Glucose
[138]
GC electrode is irreversible with high over-potential [36]. Thus electrode responses were
non-reproducible because of surface fouling. Due to presence of PDDA, electrode
response was stable and showed anti-fouling properties without any reduction in current
(for more than 30 cycles). In the absence of PDDA electrode surface fouling decrease
40% response after 30 cycles.
5.1.4.3. Effect of pH on AA and UA oxidation
The effect of solution pH on peak currents and potentials during AA and UA
oxidation (0.5 mM concentration each) was studied in 0.2 M phosphate buffer in presence
of PDDA (0.028%). At pH (2.0-7.0), both peak potentials shifted towards negative
direction. But, at pH (2.0-4.0), peak currents were increased, while reduced in 5.0-7.0 pH
range. This observation may be explained due to facilitated deprotonation step in
oxidation process at high pH [37]. Anodic potential for AA oxidation showed 55 mV/pH
slopes between pH (2.0-4.0) range (1e-/1H
+ oxidation) [38]. While at high pH, 30 mV/pH
slope value suggested involvement of 2e-/1H
+ [33]. Below pH: 4.17 (pKa), AA exists as
ascorbate anion (A-), thus A
- and AA may co-exist in 1:1 ratio [39]. Hence observed peak
current at pH 4.0 corresponded to the presence of the both the species in solution. 51
mV/pH slope for anodic potential in pH range: 2.0–7.0 confirmed 1e-/1H
+ UA oxidation
process [38]. UA oxidation involved two electrons to produce 4,5-dihydroxy-4,5-
dihydrouricacid (unstable) [34]. The oxidation peak current for UA was increased with
pH (2.0-4.0). Hence, pH: 4.0 in presence of PDDA is optimum condition for maximum
peak currents and peak separation (360 mV).
5.1.4.4. Simultaneous determination of AA and UA
Electrochemical sensing of biomolecules using GC electrode suffered from
interference of AA (mM concentration in biological fluids); because the fouling effect of
AA oxidation product is not negligible and the AA oxidation potential is quite close to
most of biomolecules. Herein, we tried simultaneous detection of UA and AA. In the
presence of PDDA, CV studies at GC electrode, revealed comparatively high separation
of UA and AA. The CVs for UA and AA (0.5 mM each) in 0.2 M phosphate buffer (pH:
4.0) with and without PDDA (0.028%) at GC electrodes are presented in Fig. 5.1.5. In the
absence of PDDA, both showed single peak at a potential different from their individual
oxidation potentials. At GC electrode, repetitive oxidation potential of AA was gradually
Click t
o buy NOW!
PDF-XChange
ww
w.tracker-software
.comClic
k to buy N
OW!PDF-XChange
ww
w.tracker-software.c
om
Chapter V: Conducting polymer............................. Uric acid and Glucose
[139]
-0.3 0.0 0.3 0.6 0.9
0.0
0.5
1.0
1.5
2.0
2.5
c
b
a
I /
A
E / V S SCE
Fig. 5.1.6. Differential pulse voltammograms for:
(a) AA (0.50 mM); (b) UA (0.50 mM); and (c)
mixture of AA and UA (0.50 mM, each) in the
presence of PDDA at GC electrode in 0.2 M
phosphate buffer solution (pH: 4.0).
-0.8 -0.4 0.0 0.4 0.8 1.2
-4
0
4
8
12
b
a
I /
A
E / V s SCE
Fig. 5.1.5. CVs for mixtures of AA and UA (0.5 mM,
each) in 0.2 M phosphate buffer solution (pH: 4) at 50
mV/s scan rate: (a) without PDDA; (b) with 0.028%
of PDDA, at GC electrode.
shifted towards anodic side, i.e.,
the non-reproducible electrode
response [36]. In absence of
PDDA, AA may affect responses
of UA at GC electrode [34]. In the
presence of cationic PDDA, clear
separation between AA and UA
peak potentials was observed. For
confirmation, the responses of UA
and AA were investigated by the
more sensitive method,
differential potential voltammetry
(DPV).
DPV curves for the oxidation of AA and UA (0.50 mM each, separately or mixture)
in the presence of PDDA in 0.20 M
phosphate buffer solution at GC
electrode are presented in Fig. 5.1.6.
Without PDDA, individual response
data for AA and UA were not
included. It was reported that in this
condition, indistinguishable oxidation
of AA and UA occurred at same peak
potential (600 mV) [22]. While in
presence of PDDA, two well defined
oxidation peaks at 100 mV and 460
mV were observed with about 360
mV separations, which were large
enough for their simultaneously determination. Observed peak separation was compared
with reported values [22,36,40-43] on different modified electrodes in Table 5.1.2.
Click t
o buy NOW!
PDF-XChange
ww
w.tracker-software
.comClic
k to buy N
OW!PDF-XChange
ww
w.tracker-software.c
om
Chapter V: Conducting polymer............................. Uric acid and Glucose
[140]
Table 5.1.2. Comparison of some characteristics of the different modified electrodes for
the determination of UA in the presence of AA.
Remarkable study was performed by recording DPVs curves at different UA
concentration (1.0 - 60.0 M) with 0.2 mM fixed AA concentration (Fig. 5.1.7(A)). The
peak current increased linearly with UA concentration with 0.994 correlation coefficient,
0.05 µA µM-1
sensitivity and 0.047 µA intercept. The precision (n = 6) assessed as
relative standard deviation (RSD) was 2% standard deviation (Fig. 5.1.7(B)). The peak
potential of AA was less in compare with UA, while peak currents of UA with or without
Electrode Sensitivity
(µA/ µM)
Determination
Range (µM)
Peak
Separation (V)
References
PEDOT/GC 0.48 1-20 0.365 39
OMC/Fc 0.0224 60 - 390 0.308 40
CT/CPB/GC - 2 - 600 0.260 22
Acetate buffer/GC 18.57 5 - 10 0.310 35
BBNBH /TN/ CPE - - 0.250 41
f-OMC/IL gel - 0.1 – 100 0.300 42
Polyelectrolyte/GC 0.05 1 - 60 0.360 In this work
-0.3 0.0 0.3 0.6 0.92
4
6
8A
a
g
I/A
E / V s SCE0 10 20 30 40 50 60
0.0
0.5
1.0
1.5
2.0
2.5
3.0B
I/A
UA/ M
Fig. 5.1.7. (A) Differential pulse voltammograms of AA and UA in the presence of PDDA at GC
electrode in 0.2 M phosphate buffer solution (pH: 4.0). [AA] was kept constant at 0.2 mM; [UA]
was changed from (a) 1µM, (b) 10 µM, (c) 20 µM, (d) 30 µM, (e) 40 µM, (f) 50 µM and (g) 60
µM, (B) Relationship between the anodic peak current and the concentration of UA. Linear fit
equation; Ip/ A=0.05 ([UA]/ M) + 0.067 (Ip/ A), R2 = 0.994).
Click t
o buy NOW!
PDF-XChange
ww
w.tracker-software
.comClic
k to buy N
OW!PDF-XChange
ww
w.tracker-software.c
om
Chapter V: Conducting polymer............................. Uric acid and Glucose
[141]
AA were almost same. Thus in presence of PDDA, oxidation of AA did not affect the UA
oxidation at GC electrode. PDDA acted as antifouling agent and covered the GC
electrode surface after formation of complex with AA oxidation product. Thus, in the
presence of AA (0.2 mM) and PDDA, sensing of UA at very low concentration level (1
µM) at GC electrode is possible. In biological conditions, AA concentration is always
higher than UA, thus about 200 times concentration ratio of AA and UA was undertaken
to mimic the physiological and biological conditions. Observations revealed the
feasibility for simultaneous detection of AA and UA in the presence of PDDA, as
separation of peak potentials were rather appreciable.
5.1.4.5. Effect of interfering substances
Table 5.1.3. Effect of interferents on DPV current response for sensing of UA (1µM) at
GC electrode in PDDA aqueous media.
aThe response of peak current for 1µM UA was 1.68µA in the absence of interferents. The
positive and negative signs are the increase and decrease in the current response, respectively.
In biological samples, ascorbic acid, glucose, urea, citrate and oxalate, etc. coexist
with UA and interfere in its selective detection. Thus, prior to the application of sensor in
Interferents Concentration of
interferents/mM
Change in current
response for 1µM UA/µAa
Glucose 0.1
0.2
0.3
-0.056
-0.059
-0.066
Urea 0.1
0.2
0.3
-0.025
-0.040
-0.050
Citrate 0.1
0.2
0.3
0.037
0.010
0.011
Oxalate 0.1
0.2
0.3
0.027
0.031
0.012
Click t
o buy NOW!
PDF-XChange
ww
w.tracker-software
.comClic
k to buy N
OW!PDF-XChange
ww
w.tracker-software.c
om
Chapter V: Conducting polymer............................. Uric acid and Glucose
[142]
biological samples we often need to check the effect of interferents on the response
specific analyte. Under the optimized conditions, 1.0 µM UA was sensed in the presence
of different interferents (0.1–0.3 mM) with 0.028% PDDA in 0.2 M phosphate buffer
solution, to explore the effect of interferents. No shift in oxidation potential of UA in
presence of interferents was detected. The oxidation current responses for UA in the
absence and presence of the interferents of 0.1–0.3 mM were checked and presented in
Table 5.1.3. In all the cases, change in peak current response for 1 µM UA in the
presence of different interferents was within ±0.03 µA. Thus, the UA and AA both can be
detected simultaneously in the presence of different interferents (similar to biological
systems) in polyelectrolyte media at GC electrode.
5.1.4.6. Reproducibility and stability test
The reproducibility of the proposed method was investigated by comparing the
peak current of DPV response to UA and AA at ten times added PDDA to solution at GC
electrode independently. For relative standard deviation (RSD) were 1.2% and 1.7% at
UA and AA concentration of 0.5 mM, respectively, which showed a satisfactory value.
On the other hand, the oxidation currents of UA and AA at this condition almost did not
change with potential scanning between 0.0 and 0.8 V (0.2 M PBS, pH 4.0) during 200
cycles (decreased less than 2%). The stability of this method was assessed by the PDDA
solution mixed with 0.5 mM UA and AA, stored air at room temperature for 24 h. The
results showed that there was no apparent decrease in current response to AA and UA
after almost 25 days.
5.1.4.7. Determination of UA in human urine samples
The proposed method was applied for the determination of AA and UA in human
urine. Table 5.1.4 presents the results obtained from three parallel measurements. UA
concentrations were detected about 2.61, 2.65 and 2.68 µM, which were in the range
human urine samples [44, 45]. Further, 1 µM of UA was added to the samples and it was
further detected with 3.0–8.0% measuring error. These observations validated the
suitability of developed sensor for detecting UA concentration in biological samples.
Click t
o buy NOW!
PDF-XChange
ww
w.tracker-software
.comClic
k to buy N
OW!PDF-XChange
ww
w.tracker-software.c
om
Chapter V: Conducting polymer............................. Uric acid and Glucose
[143]
Table 5.1.4. Results from simultaneous detection of AA and UA in urine samples by
proposed methoda
Urine samples
Detected
(µM)
Added
(µM)
Found
(µM)
Measurement errors
(%)
Sample 1 UA
AA
2.61
ND
1.00
0.50
3.58
0.49
3.0%
1.0%
Sample 2 UA
AA
2.65
ND
1.00
0.50
3.57
0.49
8.0%
1.0%
Sample 3 UA
AA
2.73
ND
1.00
0.50
3.68
0.48
5.0%
2.0%
ND= not detected .aThe results are average of three determinations
5.1.5. Conclusions for UA detection in the presence of AA in polyelectrolyte
In this chapter, a simple and reliable approach for the simultaneous detection of UA
and AA in cationic polyelectrolyte media (PDDA) at GC electrode is described. AA and
UA interacted with PDDA and facilitated their accumulation at GC surface, resulting in
the enhancement of oxidation current. Better separation of oxidation peaks of UA and
AA were achieved. It was also observed that UA and AA both can be detected
simultaneously in the presence of different interferents (similar to biological systems) in
polyelectrolyte media GC electrode. This method was also validated for the
determination UA in human urine samples with 3.0 – 8.0% measuring error. Reported
methodology is advantageous compared with chemically modified electrode because of
its simplicity and low cost. Furthermore, reported method is suitable for simultaneous
detection of AA and UA with good sensitivity and selectivity in biological fluids.
5.2. Ion conducting polymer-metal nanoparticle composite thin film for
glucose detection
Clinical conditions of diabetes mellitus are well known and understood, yet remain
growing concern as the prevalence of the disease increases worldwide at an alarming rate.
A number of life-threatening and life-impeding conditions greatly affect the diabetic
community, resulting in a much greater risk of cardiac, nervous, renal, ocular, cerebral
and peripheral vascular diseases. It was estimated that 2.8% of the world population was
affected by diabetes in 2000 [46]. Treatment has become a far more sophisticated science,
Click t
o buy NOW!
PDF-XChange
ww
w.tracker-software
.comClic
k to buy N
OW!PDF-XChange
ww
w.tracker-software.c
om
Chapter V: Conducting polymer............................. Uric acid and Glucose
[144]
with self-testing becoming increasingly more compact and accurate, and continuous
glucose monitoring now obtained from practical commercial sensors. Therefore
continuous monitoring of glucose level in blood is essential which demands the
development of a reliable, fast and ‘interference-free’ method. This has been one of the
challenging tasks for both analytical chemists and biomedical engineers for a long time.
Many researchers were focusing on the development of sensors based on the
electrocatalytic oxidation of glucose [47-50].
Enzymatic glucose sensors were good selective and high sensitive, although
enzymatic sensors offer reliable results [51], the most common and serious problem is the
instability of the enzyme in the sensors. Activity of glucose oxidase (GOx) can be
seriously affected by temperature, pH, oxygen, humidity, toxic chemicals and so on.
Furthermore, a complicated procedure, including functionalization, cross-linking,
adsorption, or entrapment, is required for the immobilization of the enzyme on the
electrode surfaces, and the activity of the GOx may decrease [52]. To overcome or
alleviate the drawbacks of enzymatic glucose sensors, more methods were developed to
determine glucose without using enzymes. Hence, attempts were made to develop
enzyme free sensors. Various metal electrodes such as Pt, Au, Cu, Bi, Hg, and Ag have
been proved to be highly electro-active in the glucose oxidation requiring a high over
potential. But, most of these electrodes were found to suffer the problems of low
sensitivity and poor selectivity caused by surface poisoning from the adsorbed
intermediates and the interference from chloride [53]. More recently, some of these
drawbacks were overcome by taking advantages of special physical and chemical
properties of different nanomaterials, such as nanoparticles [54-56], nanowires [57], and
nanoporous films [58,59]. When utilizing nanomaterials, it becomes possible to minimize
the influence of interferences such as AA and UA [60]. Nonetheless, low sensitivity and
costliness still limits the practical application of many nanomaterials based sensors.
For the electrochemical sensor and biosensor applications, the electrode modifying
material is expected to possess several characteristics such as good electron transduction
capability, physical or chemical environment for the bioactivity, easy accessibility
towards the analyte and large surface area. Literature reveals that all these important
characteristics cannot be inbuilt in a single material. Hence, there is always a demand for
Click t
o buy NOW!
PDF-XChange
ww
w.tracker-software
.comClic
k to buy N
OW!PDF-XChange
ww
w.tracker-software.c
om
Chapter V: Conducting polymer............................. Uric acid and Glucose
[145]
the development of composite materials, comprising two or more components, to achieve
adequate sensitivity and stability for the biosensors [51,61,62].
Polyaniline and its derivatives are superior polymeric materials with high
conductivity, good redox reversibility, strong attachment to the electrode surface, and
stability [62, 63]. It has been used in many practical applications, such as electrocatalysis
[64], chemo sensors [65], and biosensors [66]. Sulfonated Poly (ether ether ketone)
(SPEEK), sulfonated ionomer, has been mixed with conducting polymer to develop a
catalyst, gas separation, storage material and electrodialysis because of its enhanced
redox activity, conductivity, and mechanical behavior [67-69]. The incorporation of
SPEEK into polymer film minimizes the brittleness of the pure polymer, improves the
solubility and enhances the long-term stability of sensor.
Copper and copper oxide based materials were of great interest for the electro-
oxidation of glucose for a long time [70-73]. Recently, there is a great amount of interest
on copper and copper oxide nanomaterials for the enzyme free detection of glucose.
Cherevko et al. reported porous CuO electrode for glucose oxidation [74]. Xu et al.
developed copper nanoparticles functionalized with dimethylglyoxime modified glassy
carbon for the amperometric detection of glucose in alkaline medium [75]. Kang et al.
reported Cu nanoclusters deposition on layer of CNT prepared on glassy carbon
electrode. The obtained sensor showed very high sensitivity and good stability due to the
increase in electroactive surface area, the synergistic electrocatalytic activity combining
Cu nanoclusters with CNTs, and the three-dimensional porous structure of the Cu
nanoclusters [76]. But, no one attempt for copper nanoparticles incorporated with soluble
conducting polymer for glucose oxidation in alkaline medium, due to polymers as
electro-inactive in basic media and insolubility in solvent. Recently, over-oxidized
polypyrrole modified electrodes fabricated for glucose detection in alkaline medium with
good sensitivity and stability [77,78].
Every year authors are being reported on enzyme free glucose sensor, though they
lack the required sensitivity and/or selectivity or the method of fabrication may be
cumbersome. In order to further improve the sensitivity and selectivity, we propose an
enzyme free glucose sensor based on the SPEEK/PDMA/CuNFs/GCE. Nano-flower Cu
particles were growth by novel one step method. We used the anion-cation pair [SPEEK--
Click t
o buy NOW!
PDF-XChange
ww
w.tracker-software
.comClic
k to buy N
OW!PDF-XChange
ww
w.tracker-software.c
om
Chapter V: Conducting polymer............................. Uric acid and Glucose
[146]
Gluconolactone
Glucose
CuNFs coated PDMA /SPEEK
DMA and SPEEK mixture
DMASPEEK
2e-, 2H+
PDMA growth on SPEEK
APS
CuC
l2
Gluconolactone
Glucose
CuNFs coated PDMA /SPEEK
DMA and SPEEK mixture
DMASPEEK
2e-, 2H+
PDMA growth on SPEEK
APS
CuC
l2
Fig. 5.2.1. Fabrication of SPEEK/PDMA/CuNFs/GCE sensor.
DMA+]
as reducing agent for [CuCl2] reduction, and concomitantly deposits on/in the
PDMA and SPEEK as a nanosized Cu0 metal particles. Modified composite electrode
was then fabricated and electrochemically characterized to explore their catalytic activity
for glucose oxidation in alkaline medium. Our results demonstrate advantages such as
better sensitivity to glucose, fast current response, stability, and insensitivity to
interferences.
5.2.1. Materials and methods
D(+)-glucose, ascorbic acid, uric acid and poly (ether ether ketone) (PEEK) were
purchased from Fluka Chemicals. N,N’-dimethyl aniline (DMA), Copper chloride
(CuCl2), ammonium persulphate (APS), N,N’- dimethylacetamide, potassium hydroxide
(KOH) and hydrochloric acid (HCl) were of analytical grades and used as received.
Glucose solutions were prepared in 0.1 M KOH solution fresh at the time of experiment.
Double distilled water was used throughout the experiment.
5.2.2. Fabrication of SPEEK/PDMA/CuNFs/GCE sensor
The fabrication of sensor was presented in Fig. 5.2.1. Sulfonation of PEEK was
obtained by slow addition of dried powdered PEEK in conc. sulfuric acid by 10% (w/v)
Click t
o buy NOW!
PDF-XChange
ww
w.tracker-software
.comClic
k to buy N
OW!PDF-XChange
ww
w.tracker-software.c
om
Chapter V: Conducting polymer............................. Uric acid and Glucose
[147]
material to acid ratio at room temperature under constant stirring for 72 h. The moisture
contamination was rigorously excluded to ensure reproducibility of sulfonation level,
which was found to be 61% by 1H-NMR spectra and ion-exchange capacity studies. It
was believed that the presence of moisture contamination renders the formation of
pyrosulfonate intermediate to inter and intra-molecular sulfone crosslink. After the
reaction, polymer was precipitated with at least five-fold volume of deionized water and
shredded to obtain fine powder. This was then filtered and washed with water until the
last trace of acidity was removed [69]. Mechanism of nanocomposites formation was
reported earlier [62]. SPEEK (8.0 mg) was dissolved in 10 ml of N,N’-dimethylacetamide
and it was mixed with clear solution 4 mmol of DMA monomer under constant stirring in
nitrogen atmosphere for 2 h. 20 ml (1mM) of CuCl2 (in 1.0 M HCl) and 20 ml (0.2 M) of
ammonium persulphate (in 1.0 M HCl) was added drop wise for ternary mixture and
binary mixture, respectively. The color of the mixture turned into green indicating the
occurrence of polymerization of dimethylaniline and the colloidal solution was
evaporated into paste. The green paste was grounded well with a mortar and this
suspension was drop coated onto the surface of GCE (Glassy carbon electrode (GCE) was
polished by 1.0 and 0.06 μm alumina powder slurry on the emery sheet and chamois
leather, respectively. Then it was rinsed with Milli-Q water for 5-10 min and ultra
sonicate in nitric acid (1:1), acetone and redistilled water (10 min each), respectively) and
dried to obtain SPEEK/PDMA/CuNFs/GCE. For a comparative study,
SPEEK/PDMA/GCE and PDMA/CuNFs/GCE electrodes were also fabricated.
5.2.3. Electrochemical studies and determination of glucose
A conventional three electrode cell assembly was used for the electrochemical
measurements. SPEEK/PDMA/CuNFs/GCE and SPEEK/PDMA/GCE sensor were used
as the working electrode. Saturated calomel electrode (SCE) and platinum wire were used
as reference and counter electrodes, respectively. The electrocatalytic activity of
SPEEK/PDMA/CuNFs/ GCE was evaluated using cyclic voltammetry in 0.1 M KOH. A
small magnetic bar provided the convective transport. Amperometric measurements were
performed with SPEEK/PDMA/CuNFs/ GCE sensor at fixed potential (0.5 V s SCE) in
0.1 M KOH solution under continuous stirring for various concentrations of glucose in an
Click t
o buy NOW!
PDF-XChange
ww
w.tracker-software
.comClic
k to buy N
OW!PDF-XChange
ww
w.tracker-software.c
om
Chapter V: Conducting polymer............................. Uric acid and Glucose
[148]
Fig. 5.2.2. The FTIR spectra of: (A) SPEEK, (B) DMA/SPEEK, and (C) PDMA/SPEEK
electrochemical cell. After stabilization of the baseline, glucose solution was injected and
the plot of current versus time was recorded.
For determination of glucose in blood serum (BS), serum sample was collected
from normo-glycemic person. 1.0 ml of the serum was diluted to 5.0 ml in phosphate
buffer solution (pH: 7.0) (now referred as diluted blood serum (DBS)). Different samples
of blood serum (BS), were prepared as described; (b) 0.5 ml DBS + 0.10 mM glucose, (c)
1.5 ml of DBS + 0.58 mM glucose, and (d) 1.5 ml DBS + 0.30 mM glucose, were added
to 10 ml 0.1 M KOH, respectively and the amperometric response was recorded at 0.5 V.
5.2.4. Results and discussion for detection of glucose
5.2.4.1. Structural characterization
FTIR spectra for SPEEK, DMA/SPEEK and PDMA/SPEEK films were presented
in Fig. 5.2.2. FT-IR spectra of the SPEEK was found that the aromatic C-C stretching
vibration was observed at 1599-1412 cm-1
with sharp to medium intensity, while the
absorption at 3092-3006 cm-1
with medium intensity was attributed to aromatic C-H
stretching vibration in SPEEK. In plane C-H deformation bands were observed at 1290-
1000 cm-1
, and C-H out of plane bending at 955-845 cm-1
. Peaks at 2192-1788 cm-1
were
characteristic of 1,2-disubstituted and 1,2,4-trisubstituted aromatic moiety. The
Click t
o buy NOW!
PDF-XChange
ww
w.tracker-software
.comClic
k to buy N
OW!PDF-XChange
ww
w.tracker-software.c
om
Chapter V: Conducting polymer............................. Uric acid and Glucose
[149]
adsorption band at 1729 cm-1
was attributed to the carbonyl group stretching vibration,
and at 1024 cm-1
to O=S=O vibrations of sulfonic acid group. Few peaks with medium to
sharp intensity for –OH group of –SO3H were observed at 3644-3392 cm-1
. The
absorption bands associated with the aromatic ring in plane skeletal deformation
vibrations at 1601, 1510 and 1467 cm-1
.The band at 750 and 696 cm-1
, indicate that there
is head to tail coupling in the polymers. Absorption bands around at 1321 and 1243 cm-1
are due to C-N stretching vibrations of tertiary amines of DMA/SPEEK , but in the
PDMA/ SPEEK composite was observed in the higher frequency region shifted to 1359
and 1269 cm-1
[79]. Observed band at 2850-3000 cm-1
indicated CH3 groups of the
composites. SPEEK exhibited characteristic absorption bands at 3450 cm−1
(assigned to
OH vibration from sulfonic group interacting with molecular water), 1633 cm−1
(a
carbonyl absorption band), and 1050-1080 cm−1
(sulfur–oxygen (O=S=O) symmetric
vibration) [80].These results confirmed growth of polymerization of dimethylaniline in
SPEEK matrix caused by the ionic interactions of cationic form of DMA and SO3
(SPEEK).
Fig. 5.2.3(A) shows the XRD of SPEEK/PDMA/CuNFs .The diffraction pattern
clearly shows three major peaks at 43.50, 50.00, and 70.48 in the range of 10 to 80 ,
which can be assigned to the diffraction from the [111], [200],and [220] planes,
respectively, of the face-centered cubic lattice of Cu0 , which match well the standard
XRD data (JCPDS 004-0836) and no additional peaks of Cu or of other impurities or
oxidation of Cu2O/CuO were seen [81]. The XRD results clearly indicate that CuNFs
were present on the surfaces of PDMA/SPEEK. SEM was applied to observe the surface
morphology of the nanocomposites (Fig. 5.2.3(B-D)). SEM image of PDMA/SPEEK
shows the existence of globular porous structure (sizes in the range of 200 nm to 1 m);
while for SPEEK/PDMA/CuNFs composites showed nanostructure with high density,
shoe flower (hibiscus flower) shaped Cu nanocrystals were produced and randomly
distributed on polymer composite.
Click t
o buy NOW!
PDF-XChange
ww
w.tracker-software
.comClic
k to buy N
OW!PDF-XChange
ww
w.tracker-software.c
om
Chapter V: Conducting polymer............................. Uric acid and Glucose
[150]
a b
d
(111)
20 30 40 50 60 70 802(
Rel
ativ
e in
tens
ity
(200)(220)
A B
2 m
C
2 m
D
200nm
a b
d
(111)
20 30 40 50 60 70 802(
Rel
ativ
e in
tens
ity
(200)(220)
A B
2 m2 m
C
2 m2 m
D
200nm200nm
Fig. 5.2.3 (A) XRD of SPEEK/PDMA/CuNFs and SEM images of (B)
SPEEK/PDMA, (C) and (D) SPEEK/PDMA/CuNFs composite.
5.2.4.2. Electrochemical performance towards glucose
Cyclic voltammograms (CVs) of the modified electrodes were conducted in 50ml
0.1 M KOH in the absence and presence of the glucose, respectively. As shown in Fig.
5.2.4(A), no redox peaks can be observed at bare GCE; while at the SPEEK/PDMA/GCE
exhibits a pair of one electron quasi-reversible redox wave was observed in the absence
and presence of the glucose, respectively (Fig. 5.2.4(B)). This observation is quite
interesting. Because of PDMA provides adequate redox behavior to the electrode surface
and facilitates electron transfer at the electrode. Thus, an enhanced electrocatalytic
0.0 0.2 0.4 0.6 0.8
2
4
6
8 A
b
a
I/A
E / V S SCE0.0 0.2 0.4 0.6 0.8
-1
0
1
2
3
4
b
a
B
I/A
E / V( s SCE)0.0 0.2 0.4 0.6
-10
0
10
20
30
40
C
a
b
I/A
E/V S SCE Fig. 5.2.4. Cyclic voltammograms (CVs) of (A) bare GCE, (B) SPEEK/PDMA/GCE and (C)
SPEEK/PDMA/CuNFs/GCE in 50ml of 0.1 M KOH solution: (a) absence of glucose and (b)
presence of 1.0 mM glucose, respectively.
Click t
o buy NOW!
PDF-XChange
ww
w.tracker-software
.comClic
k to buy N
OW!PDF-XChange
ww
w.tracker-software.c
om
Chapter V: Conducting polymer............................. Uric acid and Glucose
[151]
0 25 50 75 1000
5
10
15
20
25
y = 0.225x + 0.75
R² = 0.996
I/A
/mV/S
0.0 0.2 0.4 0.6
-20
0
20
40
60
I/
A
E / V S SCE
Fig. 5.2.6. CVs of SPEEK/PDMA/CuNFs/GCE in
0.1M KOH (50 ml) and 0.5mMglucose at various
scan rates: 10-100 mV/s. Inset: Peak current vs. scan
rates.
current was witnessed at SPEEK/PDMA/GCE. But in above both cases of electrodes no
changes in current during the addition of glucose were seen.
In the case of SPEEK/PDMA/CuNFs/GCE, a significantly high peak current with
irreversible process than GCE and SPEEK/PDMA/GCE was observed. Thus, when
CuNFs were grafted onto
PDMA/SPEEK the electroactivity
of the electrode was significantly
improved for glucose detection
(Fig. 5.2.4(C)). PDMA and CuNFs,
synergistically contributes to the
electrocatalytic process. The above
results demonstrate that
modification of GCE by
SPEEK/PDMA/CuNFs provides
adequate electron transfer path and
enhance catalytic current for
glucose oxidation [82]. The performance of SPEEK/PDMA/CuNFs /GCE for glucose
oxidation was assessed in acidic, alkaline and neutral medium (Fig. 5.2.5). No current
response was detected in acidic or neutral medium. In alkaline medium (0.1 M KOH)
current response increased with
decrease in background noise due
relatively easy oxidation of glucose
[76]. Earlier reports also supports
our observations for enhanced
electro-catalytic activity transition
metals towards oxidation of
carbohydrates, in alkaline medium
[75,76,83] Thus, SPEEK/PDMA/
CuNFs/GCE sensor was suitable for
glucose oxidation in alkaline
medium(0.1 M KOH).
0.0 0.2 0.4 0.6
0
4
8
12
a
bcI/
A
E / V( s SCE)
Fig. 5.2.5. CVs of SPEEK/PDMA/CuNFs/GCE for
1mM glucose oxidation in different media a) 0.1 M
HCl, b) 0.1 M PBS (pH: 7) and (c) 0.1 M KOH .
Click t
o buy NOW!
PDF-XChange
ww
w.tracker-software
.comClic
k to buy N
OW!PDF-XChange
ww
w.tracker-software.c
om
Chapter V: Conducting polymer............................. Uric acid and Glucose
[152]
0 2000 4000 6000 80000
6000
12000
18000
24000
c
b
a
Z''/
Ohm
Z'/OhmFig. 5.2.7. Nyquist plots of EIS in 0.1 M KOH at
(a) bare GCE, (b) PDMA/SPEEK/GCE and (c)
SPEEK/PDMA/ CuNFs/GCE.
0 5 10 15 200
5
10
15
20
25
30
I/A
Cu2+
(mM)
Fig. 5.2.8. Variation of copper concentration with
peak current of glucose oxidation (0.1 M KOH,
1.0 mM glucose) at SPEEK/PDMA/CuNFs/GCE.
CVs were recorded in a solution of 0.5 mM glucose at SPEEK/PDMA/CuNFs/GCE
for different scan rates (Fig. 5.2.6). It can be seen that the potentials and peak currents of
the glucose irreversible process (anodic peak) are dependent on the scan rates. Larger
peak-to-peak separations were observed with increase in scan rates. The anodic peak
current values increased linearly with the scan rates (correlation coefficient values of
0.996 for anodic peak) in the scan rate range of 10-100 mV/s (Inset Fig. 5.2.6). The
linearity between peak current and scan rate suggests that the electrochemical kinetics is
controlled by the adsorption of glucose.
An impedance spectrum (Nyquist plot) consists of a semicircle portion observed
at higher frequency range
corresponding to the electron-
transfer-limited process and a linear
segment at lower frequencies
representing the diffusion limited
process. The diameter of the
semicircle in the Nyquist plot equals
the electron-transfer resistance (Rct),
which is related to the electron-
transfer kinetics of the redox probe at
the electrode surface. EIS of GCE,
SPEEK/PDMA/GCE, and SPEEK/
PDMA/CuNFs/GCE in 0.1 M KOH
was presented in Fig. 5.2.7. It shows
a decrease in diameter of the
semicircular portion for
SPEEK/PDMA/CuNFs/GCE and
hence the electron transfer resistance
decreases in this electrode. From the
Fig. 5.2.7, it is obvious that the bare
GCE exhibits maximum electron
transfer resistance and it decreased
Click t
o buy NOW!
PDF-XChange
ww
w.tracker-software
.comClic
k to buy N
OW!PDF-XChange
ww
w.tracker-software.c
om
Chapter V: Conducting polymer............................. Uric acid and Glucose
[153]
for ternary composite [84]. Variation of peak current of glucose oxidation (0.1 M KOH,
1.0 mM glucose) at SPEEK/PDMA/CuNFs/GCE with copper concentration in presence
of dimethylaniline (fixed concentration) was also investigated (Fig. 5.2.8). Further, with
increase in copper chloride concentration, peak current increased and beyond 10 mM it
limited, because agglomeration of copper particles in the polymer matrix and thus
reduction of its surface area.
As said by the related studies on the electrochemical oxidation of glucose in
alkaline media at metals modified electrodes, clear mechanism has not been proposed yet
[74,84-86].According to them, the oxidation was triggered by the deprotonation of
glucose and isomerization to its enediol form followed by adsorption onto the electrode
surface and oxidation by different oxidation states of metals. During the glucose
oxidation occurs at 0.5 V, the oxidation might not simply produce gluconic or glucuronic
acid but might instead entail C-C bond cleavage generating lower-molecular-weight
products such as formats and carbonates, which involves a 12 electron transfer. Thus, the
current response for a definite concentration of glucose has to be very high on CuNFs
modified enzyme free electrodes as compared to other enzyme free electrodes.
5.2.4.3. Amperometric sensing of glucose
To find the optimal conditions for the amperometric sensing of glucose, the effect
of applied potential on the response current of the sensor was investigated. The
amperometric curves of modified electrode in 0.1 M KOH solution with successive
addition of 0.5 mM glucose at applied potentials of 0.3 - 0.6 V was shown in Fig. 5.2.9.
In the range of 0.3 - 0.6 V the oxidation current of glucose increased with increasing
potential. With increasing potential further the side reaction takes place on the electrode,
which results in the increase of the background current and leads to the inhibition of
glucose oxidation. Thus, a constant potential of 0.5 V was chosen for further
amperometric investigations.
Click t
o buy NOW!
PDF-XChange
ww
w.tracker-software
.comClic
k to buy N
OW!PDF-XChange
ww
w.tracker-software.c
om
Chapter V: Conducting polymer............................. Uric acid and Glucose
[154]
100 200 300 400
0
5
10
15
20
25
30
I/A
Time/sec
0.3 V
100 200 300 400
0
10
20
30
40
Time/sec
0.4 V
100 200 300 4000
15
30
45
60
I/
A
Time/sec
0.5 V
100 200 300 400
0
10
20
30
40
Time/sec
0.6 V
Fig. 5.2.9. Amperometric response of SPEEK/PDMA/CuNFs/GCE with successive addition of
0.5 mM glucose in 0.1 M KOH (50 ml) at different applied potentials.
Amperometric measurements were performed to analyze the performance of
SPEEK/PDMA/ CuNFs/GCE sensor towards increasing glucose concentrations in 0.1 M
KOH (50 ml). Fig. 5.2.10 shows well-defined, stable, and fast amperometric responses
were observed at an applied potential of 0.5 V with successive step additions of glucose
in stirred condition (300 rpm). After successive glucose addition in step-wise, current
response increased at regular interval. It is clear that the response current of the modified
electrode increases to steady-state values upon the addition of glucose. This sensor
showed a very fast response, as the response reaches 95% of the steady-state value within
6s.
Inset of Fig. 5.2.10 shows the calibration curves for nanoflowers electrode.
Amperometric sensitivity and detection limit of the nanoflowers sensor was 19.3 µA/mM
(R2 = 0.994) and 0.1 µM with the linear range of 0.5-5 mM, respectively. CuNFs sensor
sensitivity is superior than that of the literatures for mesoporous platinum (10 µA/mM),
Cu-nanocluster/ CNT-Nafion composite (17.76 µA/mM),over oxidized polypyrrole
nanofiber/cobalt(II) phthalocyanine tetrasulfonate (5.695 µA/mM), oil/MWCNT/Cu-
nanoparticle composite (59.2 nA/mM), NiO/MWCNT (1.79 µA/mM), platinum
Click t
o buy NOW!
PDF-XChange
ww
w.tracker-software
.comClic
k to buy N
OW!PDF-XChange
ww
w.tracker-software.c
om
Chapter V: Conducting polymer............................. Uric acid and Glucose
[155]
nanotubes (1 µA/mM), NiO (50
nA/mM), MWCNTs (5 µA/mM),
porous Au electrode (11.8 µA/mM),
Pt-Pb alloy nanoparticles/MWCNTs
(17.8 µA/mM) and FeOOH nanowires
(12.1 µA/mM) [60,76,78, 87-94]. The
above elctrocatalytic studies of
SPEEK/ PDMA/CuNFs/GCE sensor
reveal the properties of high
sensitivity, low detection limit and
fast response time. This is attributed
to the synergetic effects by combining
Cu-nanoflowers with PDMA in the
presence of SPEEK, which greatly
increase the electrocatalytic active areas and promote electron transfer in the oxidation of
glucose.
The electrocatalytic behavior of as-prepared PDMA/CuNFs/GCE was also
investigated compared with SPEEK/PDMA/CuNFs/GCE. As shown in Fig. 5.2.11(A), an
obvious increase of anodic current of PDMA/CuNFs at 0.0 - 0.6 V can be observed when
0.5 mM of glucose was added, but the catalytic current is significantly lower than that of
0 1 2 3 4 50
25
50
75
100R
2= 0.994
Cu
rre
nt
(A
)
Glucose(mM)
0 100 200 300 400 500 6000
20
40
60
80
100b
a
0.5 mM
3.0 mM
I/A
Time (sec) Fig. 5.2.10. Amperometric responses of (a)
SPEEK/PDMA/GCE and (b) SPEEK/PDMA/
CuNFs/GCE to successive addition of 0.5 mM
glucose in 50ml of 0.1 M KOH solution at an
applied potential of 0.5 V ( s. SCE); Inset:
calibration plot of the glucose concentration vs.
current response.
0.1 0.2 0.3 0.4 0.5 0.6-20
0
20
40
A
b
a
I/A
E/V S SCE 100 200 300 400
0
20
40
60
a
b
B
I/A
Time (sec)
Fig. 5.2.11. A) CVs and (B) Amperometric curves of different electrodes for 1 mM glucose
oxidation in 0.1 M KOH, respectively; (a) PDMA/CuNFs/GCE, and (b) SPEEK/PDMA/CuNFs/
GCE.
Click t
o buy NOW!
PDF-XChange
ww
w.tracker-software
.comClic
k to buy N
OW!PDF-XChange
ww
w.tracker-software.c
om
Chapter V: Conducting polymer............................. Uric acid and Glucose
[156]
SPEEK/PDMA/CuNFs/GCE. The amperometric response of PDMA/CuNFs modified
electrode to successive addition of glucose is also studied (Fig. 5.2.11(B)). Compared
with SPEEK/PDMA/CuNFs/GCE, the response current of PDMA/CuNFs modified
electrode is unstable, vulnerable to external disturbances, and poorly reproducible.
5.2.4.4. Selectivity and Stability
Enzyme free sensors based on metals or alloys could easily lose their electroactivity
in the presence of Cl-, due to the formation of CuCl and/or CuCl complexes [74].
However, we did not observe any loss in electroactivity with our modified electrode
while sensing glucose in 0.1 M KOH with 0.1 M KCl; the current response remained
almost unchanged, confirming excellent stability of our sensor against chloride
poisoning.
Enzyme free sensor usually suffers from the inferences such as AA and UA,
because of these easily oxidized, have
higher electron transfer rates than
glucose. Normal physiological level of
glucose is 3-8 mM and that of AA and
UA is about 0.1 mM [61]. Therefore,
the selectivity of the modified
electrode was investigated against AA
and UA. Fig. 5.2.12 present
amperometric curve for modified
electrode in 0.1M KOH with
successive additions of AA and UA
(0.1 mM), and 1 mM glucose. As
shown in Fig. 5.2.12, the
SPEEK/PDMA/CuNFs/GCE successfully detected the glucose in the presence of UA and
AA, demonstrating almost negligible interferences from UA and AA. Similarly, there
was no obvious interference in the measure of the SPEEK/PDMA/CuNFs/GCE.
150 200 250 300 350 400
10
15
20
25
30
351 mM glucose + 0.1 mM UA
1 mM glucose + 0.1 mM AA
1mM glucose
I/
A
Time/secFig. 5.2.12. Amperometric responses of
SPEEK/PDMA/CuNFs/GCE to glucose (1.0 mM),
ascorbic acid (0.1 mM), and uric acid (0.1 mM) in
50 ml of 0.1 M KOH.
Click t
o buy NOW!
PDF-XChange
ww
w.tracker-software
.comClic
k to buy N
OW!PDF-XChange
ww
w.tracker-software.c
om
Chapter V: Conducting polymer............................. Uric acid and Glucose
[157]
Reproducibility and stability of the
modified electrode was investigated in
continuous operation mode. The electrode
was kept at 25 ºC in distilled water, and
tested with 0.5 mM glucose in 0.1 M
KOH for different days (Fig. 5.2.13).
Loss in current signal after first 5 days
was negligible (3%), but after 20 days,
loss in current signal was about 87.5 %.
This study has demonstrated long-term
stability and reproducibility of developed sensor, because of existing of SPEEK, has high
adhesion of CuNFs on PDMA and tightly immobilized on GCE surface.
Suitability and reliability of reported biosensor was tested in BS samples
(Preparation of BS samples described in section 5.2.3). Amperometric responses of
SPEEK/PDMA/CuNFs/ GCE to glucose solution and blood serum samples are presented
in Fig. 5.2.14 (a: 0.5 mM glucose; (b-d) blood serum in 0.1 M KOH (10 ml),
respectively). Two additions of known concentration of glucose were made and
correlated with the response obtained
in BS. At point ‘a’ addition of 0.1 ml,
0.5 mM (9.0 mg/dl) glucose solution
increases the effective concentration
with current by 9.04 µA. These
values when compared with the
values for glucose, yield
concentrations 3.51 mM (63.18
mg/dl), 4.09 mM (73.62 mg/dl) and
3.81 mM (68.96 mg/dl), their
corresponding current values were
65.1, 75.5, and 68.8 µA, respectively.
Results thus obtained were in
agreement with those from the photometric method, variation being less than 2.3%. Thus,
0 100 200 3000
50
100
150
200
250d
c
b
a
I/A
Time/sec
Fig. 5.2.14. Amperometric response of the modified
electrode for glucose solution and serum samples
tested in 0.1MKOH (10 ml) at 0.5 V applied
potential: a) 0.5 mM glucose solutions, b-d) blood
serum containing glucose concentrations.
0 5 10 15 20 25 30
88
92
96
100
I/I 0
X1
00
t (days)Fig. 5.2.13. Long-term stability of the
SPEEK/PDMA/CuNFs/GCE sensor.
Click t
o buy NOW!
PDF-XChange
ww
w.tracker-software
.comClic
k to buy N
OW!PDF-XChange
ww
w.tracker-software.c
om
Chapter V: Conducting polymer............................. Uric acid and Glucose
[158]
the SPEEK/PDMA/CuNFs/GCE biosensor can be used for the glucose detection in
human blood serum.
5.2.5. Conclusions for ion conducting polymer-metal nanoparticle composite thin
film for glucose detection
In the course of the current study we found that the stabilized new nanocomposites
can be formulated for enzyme free glucose sensor using a poly (dimethylaniline), SPEEK
and copper nanoparticles. The SPEEK/PDMA/CuNFs/GCE fabricated sensor
electrochemically boasts a number of advantages such as high sensitivity, good stability,
reproducibility, and quick response. Excellent selectivity was rendered by the
nanocomposites film and signals from oxidation of common interfering species such as
UA and AA can be effectively suppressed. The remarkable performance was attributed to
synergistic electrocatalytic activity combining a CuNFs and PDMA in the presence of
SPEEK, and the large electro-active surface area of CuNFs. Also, the developed sensor is
a potential candidate for routine glucose analysis in real blood serum samples. This
methodology would be extended for the development of other biosensors.
References
[1] Mazzali M, Kim YG, Hughes J, Lan HY, Kivlighn S, Johnson RJ, Am J
Hypertens 2000; 13:36A.
[2] Simoni RE, Ferreira LNL, Scalco FB, Oliveira CPH, Aquino FR, Oliveira MLC, J
Inherit Metab Dis 2007; 30:295
[3] Richards J, Weinman EJ, J Nephrol 1996; 9:160.
[4] Harper HA, Review of Physiological Chemistry, 16th ed. Lange Medical
Publications, California 1977; 406.
[5] Dai X, Fang X, Zhang C, Xu R, Xu B, J Chromatagr B 2007; 857:287.
[6] Jiang Y, Wang AY, Kan JQ, Sens Act B 2007; 124:529.
[7] Causse E, Pradelles A, Dirat B, Negre-Salvayre A, Salvayre R, Couderc F,
Electrophoresis 2007; 28:381.
[8] Qiao JX, Luo HQ, Li NB, Colloids Surf B 2008; 62:31.
[9] Kumar SA, Cheng HW, Chen SM, Electroanalysis 2009; 21: 228.
[10] Ensafi AA, Taei M, Khayamian T, Int J Electrochem Sci 2010; 5:116.
[11] Atta NF, ElKady MF, Galal A, Anal Biochem 2010; 1:78.
Click t
o buy NOW!
PDF-XChange
ww
w.tracker-software
.comClic
k to buy N
OW!PDF-XChange
ww
w.tracker-software.c
om
Chapter V: Conducting polymer............................. Uric acid and Glucose
[159]
[12] Nasria Z, Shams E, Electrochim Acta 2009; 54:7416.
[13] Arguello J, Leidens VL, Magosso HA, Ramos RR, Gushikem Y, Electrochim
Acta 2008; 54:560
[14] Kalimuthu P, John SA, Tang H, Hu G, Jiang S, Liu X, Bioelectrochemistry 2009;
77:13
[15] Li G, Yang S, Qu L, Yang R, Li J, J Solid State Electrochem 2010; DOI
10.1007/s10008-010-1088-7.
[16] Zheng D, Ye J, Zhou L, Zhang Y, Yu C, J. Electroanal Chem 2009; 625:82.
[17] Lin L, Chen J, Yao H, Chen Y, Zheng Y, Lin X, Bioelectrochemistry 2008; 73:11.
[18] Selvaraju T, Ramaraj R, Electrochim Acta 2007; 52:2998.
[19] Zhang QL, Xu JJ, Lian HZ, Li XY, Chen HY, Anal Bioanal Chem 2007;
387:2699.
[20] Atta NF, Galal A, Ahmed RA, Bioelectrochemistry 2010; 80:132.
[21] Moraes ML, Filho UPR, Oliveira Jr. ON, Ferreira M, J Solid State Electrochem
2007; 11:1489.
[22] Cao X, Xu Y, Luo L, Ding Y, Zhang Y, J Solid State Electrohem 2010; 14:829.
[23] Shankar SS, Swamy BEK, Chandra U, Manjunatha JG, Sherigara BS, Int J
Electrochem Sci 2009; 4:592.
[24] Zen JM, Hsu CT, Talanta 1998; 46:1363.
[25] Manjunatha H, Nagaraju DH, Suresh GS, Venkatesh TV, Electroanalysis 2009;
21:2198.
[26] Sayre CN, Collard DM, J Mater Chem 1997; 7:909.
[27] Cao W, Dong H, Huang F, Shen H, Cao Y, Front Optoelectron 2008; 1:299.
[28] Prakash S, Rao CRK, Vijayan M, Electrochim Acta 2009; 54:5919.
[29] Wang X, Balasubramanian S, Li L, Jiang X, Sandman DJ, Rubner MF, Kumar J,
Tripathy SK, Macromol Rapid Commun 1997; 18:451.
[30] Zhang Y, Cai Y, Su S, Anal Biochem 2006; 350:285.
[31] Podgornik R, Licer M, Curr Opin Coll Interface Sci 2006; 11:273.
[32] Reudo M, Aldaz A, Burgos FS, Electrochim Acta1978; 23:419.
[33] Raj CR, Tokuda K, Oshaka T, Bioelectrochemistry 2001; 53:183.
[34] Roy PR, Okajima T, Oshaka T, J Electroanal Chem 2004; 561:75.
Click t
o buy NOW!
PDF-XChange
ww
w.tracker-software
.comClic
k to buy N
OW!PDF-XChange
ww
w.tracker-software.c
om
Chapter V: Conducting polymer............................. Uric acid and Glucose
[160]
[35] Tender L, Carter MT, Murray RW, Anal Chem 1994; 66:3173.
[36] John SA, J Electroanal Chem 2005; 579:249.
[37] Wang Z, Liang Q, Wang Y, Luo G, J Electroanal Chem 2003; 540:129.
[38] Yu AM, Zhang HL, Chen HY, Analyst 1997; 122:839.
[39] Giz MJ, Duong B, Tao NJ, J Electroanal Chem 1999; 465:72.
[40] Kumar SS, Mathiyarasu J, Phani KLN, Jain YK, Yegnaraman V, Electroanalysis
2005; 17:2281.
[41] Ndamanisha JC, Guo L, Biosens Bioelectron 2008; 23:1680.
[42] Beitoollahi H, Benvidi A, Naeimi H, Chinese Chem Lett 2010; 21:1471.
[43] Dong J, Hu Y, Zhu S, Xu J, Anal Bioanal Chem 2010; 396:1755.
[44] Kratz A, Lewandrowski KB. Normal reference laboratory values. N Engl J Med
1998; 339:1036.
[45] Iverson C, Flanagin A, Fontanarosa, American Medical Association Manual of
Style: A Guide for Authors and Editors. 9th
ed. Baltimore, Md: Williams &
Wilkins; 1998.
[46] Wild S, Roglic G, Green A, Sicree R, King H, Diabetes Care 2004; 27:1047.
[47] Clark JLC, Lyons C, Annals of the New York Academy of Sciences 1962;
102:29.
[48] Reach G, Wilson GS, Anal Chem 1992; 64:381A.
[49] Turner APF, Chen B, Sergey AP, Clinical Chemistry 1999; 45:1596.
[50] Wang J, Chem Rev 2008; 108:814.
[51] Gopalan AI, Lee KP, Ragupathy D, Lee SH, Lee JW, Biomaterials 2009; 30:5999.
[52] Park SJ, Boo HK, Chung TD, Anal Chim Acta 2006; 556:46.
[53] Vassilyev YB, Khazova OA, Nikolaeva NN, J Electroanal Chem 1985; 196:105.
[54] Cui HF, Ye JS, Zhang WD, Li CM, Luong JHT, Sheu FS. Anal Chim Acta 2007;
594:175.
[55] Jena BK, Raj CR, Chem Eur J 2006; 12:2702.
[56] Kang X, Mai Z, Zou X, Cai P, Mo J, Talanta 2008; 74:879.
[57] Bai Y, Sun Y, Sun C. Biosens Bioelect 2008; 24:579.
[58] Bai Y, Yang W, Sun Y, Sun C. Sens Act B 2008; 134:471.
[59] Deng Y, Huang W, Chen X, Li Z, Electrochem Commun 2008; 10:810.
Click t
o buy NOW!
PDF-XChange
ww
w.tracker-software
.comClic
k to buy N
OW!PDF-XChange
ww
w.tracker-software.c
om
Chapter V: Conducting polymer............................. Uric acid and Glucose
[161]
[60] Watanabe T, Einaga Y, Biosens Bioelectron 2009; 24:2684.
[61] Ragupathy D, Gopalan AI, Lee KP, Electrochem Commun 2009; 11:397.
[62] Prakash S, Rao CRK, Vijayan M, Electrochim Acta 2009; 54:5919.
[63] Murray RW Published by Wiley,NY, 1992.
[64] Suganandam K, Santhosh P, Sankarasubramanian M, Gopalan A, Vasudevan T,
Lee KP, Sens Act B 2005; 105:223.
[65] Ojani R, Raoof JB, Zarei E, J Electroanal Chem 2010; 638:241.
[66] Ojani R, Raoof JB, Zarei E, Electroanalysis 2009; 21:1189.
[67] Li X, Wang Z, Lu H, Zhao C, Na H, Zhao C, J Membr Sci 2005; 254:147.
[68] Roeder J, Zucolotto V, Shishatskiy S, Bertolino JR, Nunes SP, Pires AN, J
Membr Sci 2006; 279:70.
[69] Chakrabarty T, Kumar M, Rajesh KP, Shahi VK, Natarajan TS, Sep Pur Technol
2010; 75:174.
[70] Miller B, J Electrochem Soc 1969; 116:1675.
[71] Kano K, Torimura M, Esaka Y, Goto M, J Electroanal Chem 1994; 372:137.
[72] Luo MZ, Baldwin RP, J Electroanal Chem 1995; 387:87.
[73] Torto N, Ruzgas T, Gorton L, J Electroanal Chem 1999; 464:252.
[74] Cherevko S, Chung CH, Talanta 2010; 80:1.
[75] Xu Q, Zhao Y, Xu JZ, Zhu J, Sens Actuators B Chem 2006; 114:379.
[76] Kang X, Mai Z, Zou X, Cai P, Mo J, Anal Biochem 2007; 363:143.
[77] Shi J, Ci P, Wang F, Peng H, Yang P, Wang L, Ge S, Wang Q, Chu PK, Biosens
Bioelectron 2011; 26:2579.
[78] Ozcan L, Sahin Y, Turk H, Biosens Bioelectron 2008; 24:512.
[79] Kravets L, Gilman AB, Drachev AI, High Energy Chem 2005; 39:114
[80] Xing P, Robertson GP, Guiver MD, Mikhailenko SD, Wang K, Kaliaguine S, J
Membr Sci 2004; 229:95
[81] Yang J, Zhang WD, Gunasekaran S, Biosens Bioelectron 2010; 26:279.
[82] Yan W, Feng X, Chen X, Hou W, Zhu JJ, Biosens Bioelectron 2008; 23:925.
[83] Wang J, Analytical Electrochemistry Published by Wiley-VCH, NY, 2001.
[84] Babu TGS, Ramachandran T, Nair B, Microchim Acta 2010; 169:49.
Click t
o buy NOW!
PDF-XChange
ww
w.tracker-software
.comClic
k to buy N
OW!PDF-XChange
ww
w.tracker-software.c
om
Chapter V: Conducting polymer............................. Uric acid and Glucose
[162]
[85] Wu HX, Cao WM, Li Y, Liu G, Wen Y, Yang HF, Yang SP, Electrochim Acta
2010; 55:3734.
[86] Wang X, Zhang Y, Banks CE, Chen Q, Ji X, Colloids Surf B 2010; 78:363.
[87] Wang J, Chen G, Wang M, Chatrathi MP, Analyst 2004; 129:512.
[88] Shamsipur M, Naja M, Hosseini MRM, Bioelectrochemistry 2010; 77:120.
[89] Yuan J, Wang K, Xia X, Adv Func Mater 2005; 15:803.
[90] Mu Y, Jia D, He Y, Miao Y, Wu HL, Biosens Bioelectron 2011; 26:2948.
[91] Ye JS, Wen Y, Zhang WD, Gan LM, Xu GQ, Sheu FS, Electrochem Comm 2004;
6:66.
[92] Li Y, Song YY, Yang C, Xia XH, Electrochem Commun 2007; 9:981.
[93] Cui HF, Ye JS, Zhang WD, Li CM, Luong JHT, Sheu FS, Anal Chim Acta 2007;
594:175.
[94] Xia C, Ning W, Electrochem Commun 2010; 12:1581.
Click t
o buy NOW!
PDF-XChange
ww
w.tracker-software
.comClic
k to buy N
OW!PDF-XChange
ww
w.tracker-software.c
om