Original operating instruction Electronic Analyser

70
Operation Manual EPM-XP plus Software Release 1.3.0.2 25.09.2020 Page 1 of 70 Original operating instruction Electronic Analyser

Transcript of Original operating instruction Electronic Analyser

Page 1: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 1 of 70

Original operating instruction

Electronic Analyser

Page 2: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 2 of 70

Revision History

Version Brief Description of Changes Date of Issue

1 EPM-XPplus First version 05-August-2019

1.1.0.0 New functions: Trending, Save reference files, Optimization of Pcomp calculation

11-October-2019

1.2.2.0

New functions: Time indication of loading data set from hard disc, Indication of stored ASCII file on hard disc, Indication of reference curve, direct data transfer to IPE software, small bug fix solved. New firmware release: 1.36

16-December 2019

1.2.2.0 New firmware release: 1.38 14. January 2020

1.2.3.0 New functions: p/volume (log) diagram, new optimizing function, small bug fixed solved

06. March 2020

1.2.4.0

Correction of rounding error of older measurements Bug fix temperature indication of measuring cell solved Selection of *.csv format for export as “default value” or old format “EPM-XP”

25.March 2020

1.3.0.2

New functions:

• Updated engine database

• Support for engine installation templates

• Manual overwrite of calculated scavenge air pressure

• Derivative curves of all cylinders simultaneous visible

• Calculation of ignition point

• Adjustable filtering of pressure curves

• Loading of a measurement file as reference

• Report with additional statistic values and deviation highlighting

25. September 2020

Page 3: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 3 of 70

Imprint

IMES GmbH may be contacted via the following:

Mailing address:

IMES GmbH

Dr.-Herbert-Kittel-Straße 2

87600 Kaufbeuren – Germany

Telephone and e-Mail:

Sales: [email protected]

General Information: [email protected]

Phone: +49 8341 9661730

Fax: +49 8341 96617318

For further information about other IMES products, please visit our website: www.imes.de

Copyright

Copyright © 2020 IMES GmbH. All rights reserved. All data is for information only and not guaranteed for

legal purposes. Information has been carefully checked, however, no responsibility is assumed for inaccuracies.

This manual may not be copied, translated, changed, reproduced or distributed without the written consent of

IMES GmbH.

Page 4: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 4 of 70

Contents

Contents __________________________________________________________________ 4

List of figures ______________________________________________________________ 6

1 Important information __________________________________________________ 9

1.1 Use of the operator manual __________________________________________________ 9

1.2 Copyright ________________________________________________________________ 9

1.3 Warning Notices ___________________________________________________________ 9

1.4 Address of the manufacturer _______________________________________________ 10

1.5 Obligations of the operator _________________________________________________ 11

2 Safety _______________________________________________________________ 11

2.1 Intended use _____________________________________________________________ 11

2.2 Personnel requirements ____________________________________________________ 12

2.3 Safety notes ______________________________________________________________ 12

3 Transportation and packaging ___________________________________________ 12

3.1 Safety notes for transportation ______________________________________________ 12

3.2 Scope of delivery _________________________________________________________ 13

3.3 Packaging _______________________________________________________________ 13

4 Technical data ________________________________________________________ 13

4.1 Dimensions and weight of instrument case ____________________________________ 13

4.2 Other parameters _________________________________________________________ 13

5 Scope of supply _______________________________________________________ 14

6 Description __________________________________________________________ 14

6.1 Introduction _____________________________________________________________ 14

6.2 Measure functions ________________________________________________________ 15

6.3 Functional description _____________________________________________________ 15

7 Operation ____________________________________________________________ 15

7.1 Operator push buttons ____________________________________________________ 15

7.2 Operating functions _______________________________________________________ 16

7.3 Install rechargeable batteries _______________________________________________ 17 7.3.1 Install batteries to correct pole. ____________________________________________________ 18 7.3.2 Confirm installed battery type ____________________________________________________ 18

7.4 Charge battery ___________________________________________________________ 19

7.5 Measurements with EPM-XPplus _____________________________________________ 19

8 Installation of visualisation- and USB driver software ________________________ 21

9 Visualization software __________________________________________________ 24

9.1 The Application Window __________________________________________________ 24

9.2 Connecting to the EPM-XPplus device _________________________________________ 25

Page 5: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 5 of 70

9.3 Download measured data to PC _____________________________________________ 26

9.4 Open saved data __________________________________________________________ 27

9.5 Visualization of measured data on PC ________________________________________ 28 9.5.1 P/alpha diagram _______________________________________________________________ 30 9.5.2 P/volume diagram ______________________________________________________________ 32 9.5.3 P/volume (logarithmic) __________________________________________________________ 32 9.5.4 Engine report _________________________________________________________________ 33 9.5.5 Balance diagrams ______________________________________________________________ 34 9.5.6 Trending _____________________________________________________________________ 35 9.5.7 Save to ASCII format ___________________________________________________________ 37 9.5.8 Change Engine data ____________________________________________________________ 37 9.5.9 Engine Database Editor _________________________________________________________ 38 9.5.10 Installation Template Editor ______________________________________________________ 38 9.5.11 Solver configuration ____________________________________________________________ 39 9.5.12 Color setup ___________________________________________________________________ 39 9.5.13 Optimize TDC calculation _______________________________________________________ 40 9.5.14 Save as reference curve _________________________________________________________ 41 9.5.15 Open a reference file____________________________________________________________ 41 9.5.16 Delete measurements on EPM-XPplus _______________________________________________ 42

9.1 Online measurements on engines with EPM-XPplus _____________________________ 43 9.1.1 Preparation of EPM-XPplus for online measurements on 2-stroke engines __________________ 43 9.1.2 Preparation of EPM-XPplus for online measurements on 4-stroke engines __________________ 44 9.1.3 Preparation of EPM-XPplus for online measurements via Ethernet cable ___________________ 46 9.1.4 Visualization Software __________________________________________________________ 47 9.1.5 Screen page “Online” ___________________________________________________________ 48 9.1.6 Screen page “Offline” ___________________________________________________________ 53

9.2 Update firmware to EPM-XPplus from visualisation software _____________________ 56 9.2.1 Check actual firmware version ____________________________________________________ 56

10 Accuracy check _______________________________________________________ 57

10.1 Check on pressure calibrator _______________________________________________ 58

10.1 Example of static pressure indication ________________________________________ 60

11 Cleaning ____________________________________________________________ 61

11.1 Periodically cleaning ______________________________________________________ 61

11.2 Cleaning in case of hard deposit _____________________________________________ 61

11.3 Cleaning procedure of adaptor ______________________________________________ 62

11.4 Cleaning procedure of cylinder pressure sensor _______________________________ 62

12 Basic check for fault finding ____________________________________________ 63

13 Nomenclature ________________________________________________________ 63

14 Option ______________________________________________________________ 64

14.1 IMES IPE Performance evaluation software __________________________________ 64

14.2 Ethernet cable 20m _______________________________________________________ 65

14.3 USB Extender for connection to Ethernet cable ________________________________ 66

14.4 Thompson adaptor for 4-stroke continuous online measurement __________________ 67

14.5 Pressure pump coupling for Thompson adaptor _______________________________ 68

14.6 Thread cleaner for Thompson thread ________________________________________ 69

15 EU Declaration of Conformity ___________________________________________ 70

Page 6: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 6 of 70

List of figures Figure 1: EPM-XPplus incl.all components in instrument case ................................................14 Figure 2: Push buttons .........................................................................................................15 Figure 3: 4 pcs. re-chargeable battery in the transport case .................................................17 Figure 4: Back side of EPM-XP incl. protection cover ...........................................................17 Figure 5: Battery case ..........................................................................................................17 Figure 6: Battery holder ........................................................................................................18 Figure 7: 4 x NiMh AAA 930 mAh installed on battery holder................................................18 Figure 8: Selection of installed battery type ..........................................................................18 Figure 9: Indication during charging process ........................................................................19 Figure 10: Indication charging process completed ................................................................19 Figure 11: Indicator cock ......................................................................................................19 Figure 12: Adaptor installed on indicator cock ......................................................................19 Figure 13: Indication of 2-stroke mode .................................................................................19 Figure 14: Indication of 4-stroke mode .................................................................................19 Figure 15: Warming up sensor .............................................................................................20 Figure 16: Indication during measurement............................................................................20 Figure 17: Indication of measurement results on display for 3s ............................................20 Figure 18: Indication over temperature on measuring cell of sensor .....................................20 Figure 19: Dialog of installation ............................................................................................21 Figure 20: Select destination location ...................................................................................21 Figure 21: Start menu folder .................................................................................................22 Figure 22: Select additional tasks .........................................................................................22 Figure 23: Ready to install ....................................................................................................22 Figure 24: Ready to install ....................................................................................................23 Figure 25: Installation completed ..........................................................................................23 Figure 26: Start up window of EPM-XPplus visualization software ..........................................24 Figure 27: EPM- XPplus connected via USB cable to PC .......................................................25 Figure 28: Connect button ....................................................................................................25 Figure 29: Connect button with connection established ........................................................25 Figure 30: Selection of engine configuration .........................................................................26 Figure 31: Entering of specific engine data ...........................................................................26 Figure 32: Selection of directory on hard disc of PC for saving data .....................................27 Figure 33: Downloading and TDC calculation process of measured data .............................27 Figure 34: Button “Open” ......................................................................................................27 Figure 35: Stored measurement file......................................................................................27 Figure 36: File view ..............................................................................................................28 Figure 37: Cylinder selection panel ......................................................................................28 Figure 38: Cylinder details panel ..........................................................................................29 Figure 39: Display mode panel .............................................................................................29 Figure 40: Zoom in ...............................................................................................................29 Figure 41: Zoom out .............................................................................................................30 Figure 42: Functions of p/alpha diagram ..............................................................................30 Figure 43: Pressure curve adjustments ................................................................................31 Figure 44: p/volume diagram ................................................................................................32 Figure 45: p/volume (log) diagram ........................................................................................32 Figure 46: Engine report diagram .........................................................................................33 Figure 47: Engine report configuration dialog .......................................................................34 Figure 48: Balance diagram .................................................................................................34 Figure 49: Overview of bush puttons ....................................................................................35 Figure 50: Selection criteria for measurement files ...............................................................35 Figure 51: Selection of measurement files ............................................................................35

Page 7: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 7 of 70

Figure 52: Selection criteria for values and cylinder ..............................................................36 Figure 53: Display of trending function .................................................................................36 Figure 54: File entry in menu bar ..........................................................................................37 Figure 55: CSV file selection ................................................................................................37 Figure 56: Engine Configuration Dialog ................................................................................37 Figure 57: Engine Database Editor .......................................................................................38 Figure 58: Installation Template Editor .................................................................................38 Figure 59: Engine Configuration Dialog ................................................................................39 Figure 60: “Solver Configuration” Dialog ...............................................................................39 Figure 61: “Color Setup” Dialog ............................................................................................39 Figure 62: Cylinder pressure curve incl. motored curve ........................................................40 Figure 63: Shifted Cylinder pressure curve incl. motored curve ............................................40 Figure 64: Button “Save REF” ..............................................................................................41 Figure 65: Save the reference on hard disc of PC ................................................................41 Figure 66: Button “Open REF” ..............................................................................................41 Figure 67: Selection of reference curve on hard disc of PC ..................................................41 Figure 68: Display of selected reference curve in comparison to a cylinder curve ................42 Figure 69: Close reference curve .........................................................................................42 Figure 70: Button “Del.Meas”................................................................................................42 Figure 71: Indicator cock ......................................................................................................43 Figure 72: Adaptor installed on indicator cock ......................................................................43 Figure 73: EPM-XPplus connected to notebook ....................................................................43 Figure 74: Disconnection of spring .......................................................................................44 Figure 75: Disconnection of sensor from adaptor .................................................................44 Figure 76: Connection of HTT-04 sensor on special cooling adaptor ....................................44 Figure 77: Indicator cock ......................................................................................................45 Figure 78: Adaptor installed on indicator cock ......................................................................45 Figure 79: EPM-XPplus connected via USB cable to notebook ..............................................45 Figure 80: Overview of online measurement with EPM-XPplus...............................................46 Figure 81: Overview of Ethernet connection to EPM-XPplus ..................................................46 Figure 82: Indication on display for successful connection to PC ..........................................47 Figure 83: Connect button ....................................................................................................47 Figure 84: Connect button with connection established ........................................................47 Figure 85: Indication of online pressure curve ......................................................................48 Figure 86: Example for over temperature on measuring cell of sensor .................................48 Figure 87: Button “Pressure curve save” ..............................................................................49 Figure 88: Register for saving the file ...................................................................................49 Figure 89: Indication of online pressure curve during saving process ...................................50 Figure 90: Button “Working cycle save” ................................................................................50 Figure 91: Indication of online Pmax data .............................................................................51 Figure 92: Record button ......................................................................................................51 Figure 93: Register for saving the file ...................................................................................51 Figure 94: Indication of Pmax data during recording process ...............................................52 Figure 95: Example for over temperature on measuring cell of sensor .................................52 Figure 96: Screen page “Offline” ..........................................................................................53 Figure 97: Button “100 Pressure curve open” .......................................................................53 Figure 98: Register of saved file ...........................................................................................53 Figure 99: Indication of pressure cycles ..............................................................................54 Figure 100: Button “Working cycle open” ..............................................................................54 Figure 101: Screen page “working cycle offline” ...................................................................54 Figure 102: Register of saved working cycle data file ...........................................................55 Figure 103: Selection of saved Pmax working cycle .............................................................55 Figure 104: Indication of selected Pmax data .......................................................................55 Figure 105: Button “Connect” ...............................................................................................56

Page 8: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 8 of 70

Figure 106: Selection “Tools”................................................................................................56 Figure 107: Selection “Firmware” .........................................................................................56 Figure 108: Load-and update firmware .................................................................................56 Figure 109: Indication of main menu .....................................................................................56 Figure 110: Indication of device info .....................................................................................57 Figure 111: Indication of actual firmware ..............................................................................57 Figure 112: Calibration certificate of EPM-XPplus unit ............................................................57 Figure 113: Thompson Adaptor mounted on pressure plate .................................................58 Figure 114: Example for indication of static pressure @ 0 bar ..............................................58 Figure 115: Example for indication of static pressure @ 50 bar ............................................59 Figure 116: Indication of main menu .....................................................................................59 Figure 117: Indication of measuring mode ............................................................................59 Figure 118: Indication of sensor test .....................................................................................59 Figure 119: Indication @ 0 bar on display ............................................................................59 Figure 120: Indication @ 50 bar ...........................................................................................60 Figure 121: Indication on pressure pump .............................................................................60 Figure 122: Periodical cleaning of gas channel ....................................................................61 Figure 123: Disconnection of spring .....................................................................................61 Figure 124: Disconnection of sensor from adaptor ...............................................................61 Figure 125: Drill 2.5 mm .......................................................................................................62 Figure 126: Drill in gas channel of adaptor for cleaning ........................................................62 Figure 127: Special drill tool .................................................................................................62 Figure 128: Special drill tool in gas channel of sensor ..........................................................62 Figure 129: Example IPE software .......................................................................................64 Figure 130: Ethernet cable ...................................................................................................65 Figure 131: USB-Ethernet Extender .....................................................................................66 Figure 132: Thompson adaptor for 4-stroke online measurement .........................................67 Figure 133: Pressure pump coupling ....................................................................................68 Figure 134: Thread cleaner for Thompson thread .................................................................69

List of tables

Table 1: Dimensions and weight ...........................................................................................13 Table 2: Technical specification ............................................................................................13 Table 3: Functions of push buttons .......................................................................................17 Table 4: Static pressure comparison ....................................................................................60 Table 5: Basic check for fault finding ....................................................................................63 Table 6: Nomenclature .........................................................................................................63

Page 9: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 9 of 70

1 Important information Please study this manual carefully, before using the

equipment. This will ensure that you will receive maximum

benefits from using this engine performance tuning tool with

its versatile functionality and it will guarantee optimum

benefits over its lifetime.

1.1 Use of the operator manual

We strongly recommend you to read the complete manual

before using the equipment.

It was our intention to structure this manual in a clear

layout, to enable you to get easy and instant access to the

information you are looking for. Please keep this manual in

a safe place where it is always available for easy access.

1.2 Copyright

These operating instructions has to be treated

confidentially. They may only be used by the persons

authorized to this effect. They may only be given to third

parties with the written consent of IMES GmbH.

All documents are protected under copyright law.

Dissemination and reproduction of documents or parts of

documents as well as the utilization and communication of

their contents are prohibited, unless expressly permitted.

Infringements are punishable and liable to damages.

We reserve all rights to exercise industrial property rights.

1.3 Warning Notices

In the present operating instructions, the following warning notices are used:

WARNING

Risk of injury

Consequences of non-observance …

Prevention

A warning notice of this risk level signals a potentially dangerous situation. If the dangerous situation is not

prevented, this could lead to death or serious injury to persons. Follow the instructions in this warning notice to

prevent the possible risk of death or serious injury to persons.

Page 10: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 10 of 70

CAUTION

Property damage due to …

Consequences of non-observance …

Prevention

A warning notice of this risk level signals a potentially dangerous situation. If the dangerous situation is not

prevented, this could lead to damage to property. Follow the instructions in this warning notice to prevent

damage to property.

NOTE

Text of the note …

A note identifies additional information that facilitates the handling of the device.

1.4 Address of the manufacturer

IMES GmbH

Dr.-Herbert-Kittel-Str. 2

87600 Kaufbeuren

Germany

Tel.: + 49 8341 966173 0

Fax: +49 8341 96617318

e-mail: [email protected]

Web: www.imes.de

Page 11: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 11 of 70

1.5 Obligations of the operator

The EPM-XPplus may be a source of danger if used in an improper manner or not in proper condition.

The operator is obliged to operate the EPM-XPplus only in proper condition. Hazard areas arising between IMES

devices and customer equipment are to be secured by the operator.

The operator must designate and instruct responsible persons:

• Only employ trained or instructed personnel.

• Clearly define the responsibilities of the personnel for operation, maintenance, and repair.

• Regularly check if the personnel is working with due safety and risk awareness whilst observing the operating

instructions.

• Before starting work, the personnel assigned to work with the EPM-XPplus must have read and understood the

operating instructions, in particular the "Safety" chapter, as well as the applicable regulations.

• Keep the operating instructions and applicable regulations accessible to the operating and servicing personnel.

• Determine the responsibility of the user and allow him to reject all third party instructions that are in breach of

safety regulations.

• For safe operation of this EPM-XPplus it is required to wear safety glasses and gloves. Further protective

equipment has to be worn according to the applicable regulations of the operator.

2 Safety

2.1 Intended use

WARNING

Risk of injury due to non-intended use. EPM-XPplus may be a source of danger if operated not in compliance with the intended use and/or being used otherwise.

• Use the EPM-XPplus for its intended purpose only.

• Observe the procedure described in the operating instructions.

The EPM-XPplus may only be used for measuring combustion pressure on engines. Any other use of the

EPM-XPplus is impermissible. The manufacturer/supplier accepts no liability for damage resulting from non-

intended use. The risk exclusively rests with the user/operator.

Page 12: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 12 of 70

2.2 Personnel requirements

WARNING

Work on/with the EPM-XPplus may only be carried out by persons who, because of their education and qualification, are qualified to do so. Furthermore, they must have been commissioned to do so by the operator.

Personnel to be trained, instructed or briefed or operating personnel receiving general training may only be

allowed to work with the device under the permanent supervision of an experienced operating person.

Maintenance and repair work may only be carried out by the manufacturer.

2.3 Safety notes

Observe the following notes for preventing accidents while using the EPM-XPplus:

• Operate the EPM-XPplus only in proper condition.

• Adjustment and/or maintenance work is to be carried out by authorized professionals.

• Before using the EPM-XPplus, check it for defects and deficiencies visible on the outside. Immediately notify

any changes (including changes in operating behavior) to the competent staff/person. If necessary, immediately

put the device out of service and lock/tag it out!

• Keep the operating instructions near the EPM-XPplus so as to be readily available at any moment.

• Non-observance of the safety notes contained in these operating instructions may result in property damage,

injuries or even death of persons.

3 Transportation and packaging

3.1 Safety notes for transportation

Page 13: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 13 of 70

3.2 Scope of delivery

For the exact scope of supply please refer to the order documents and compare to the delivery note.

Check for completeness

Check the entire delivery for completeness based on the attached delivery note. In this context, we are referring

to our conditions of sale and delivery.

Report any damage

Immediately report to the carrier, insurance company and the delivering factory any damage to the device due to

improper packaging or transportation identified after delivery. Take action to mitigate the damage that has

occurred and prevent further damage.

3.3 Packaging

Unless an agreement was made with IMES GmbH regarding the return of the packaging material, the packaging

material remains with the customer.

The most commonly used packaging materials cardboard.

4 Technical data

4.1 Dimensions and weight of instrument case

Length: 45 cm

Width: 35 cm

Heigh: 11 cm

Weight of

instrument case

2,7 kg

Table 1: Dimensions and weight

4.2 Other parameters

Measuring range 0...300 bar

Accuracy (EPM-XPplus unit incl. HTT® sensor) +/- 0,5 % full scale

Sampling resolution 0,05°CA on two-stroke ; 0,1°CA on four stroke

Resonance frequency HTT®-sensor >100 kHz

Max. temperature at measuring cell 300 °C (short time 1 minute 350 °C)

Storing capacity 20 measurements / engine

Online measurement Via USB- / Ethernet cable

Interface USB—HID

Battery 4 x NiMh AAA 930 mAh

Weight incl. sensor and adaptor 1100 g

Dimension incl. protection cover 210 x 100 x 80 mm

Table 2: Technical specification

Page 14: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 14 of 70

5 Scope of supply

Figure 1: EPM-XPplus incl.all components in instrument case

6 Description

6.1 Introduction

The electronic indicator EPM-XPplus is a handy-, battery

powered-, portable device to measure and evaluate cylinder

pressure on 2- and 4-stroke diesel engines at speed up to 1800

RPM.

The EPM-XPplus collect 10 consecutive pressure

measurements (cycles) and calculates peak pressure and

engine speed. The measured data is displayed in numerical

format on the LCD screen and can be stored in memory.

Up to 20 measurement data sets can be stored in the

EPM-XPplus. The stored data can be downloaded via USB-

interface to a personal computer for evaluation.

The scope of supply includes software for Windows which

allows additional evaluation and visualisation of the

collected data.

Additional continuous online measurement on one cylinder

to analyse actual engine condition by direct data transfer

via USB/Ethernet (max. length 50m) to PC

The scope of supply includes the following

components:

- EPM-XPplus unit incl. protection cover incl.

cylinder pressure sensor mounted on Thompson

adaptor

- Tool for to open- and close the Thompson adapter

- 4 pcs. 4 x NiMh AAA 930 mAh

- 1 x USB-cable (1m)

- 1 x USB stick (documentation English/German

and visualization software)

- 1 x Quickstep manual in English language (printed)

- 1 x Certificate of calibration (printed)

Page 15: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 15 of 70

6.2 Measure functions

Offline function

Pmax: Average of 10 cycles

Pressure curve: Average of 10 cycles

RPM: engine speed

Online function

Continuous online measurement on one cylinder

6.3 Functional description

The cylinder pressure will be measured with the

EPM-XPplus unit incl. high temperature pressure sensor and

Thompson adaptor at indicator cock on diesel engines.

Memory: 20 data sets

Application 2-stroke: 40 – 950 RPM

Application 4-stroke: 200 – 1800 RPM

NOTE

The EPM-XPplus turns off automatically after 10 minutes without pressing any push button

7 Operation

7.1 Operator push buttons

Figure 2: Push buttons

Arrow up ““

Arrow down ““

On/Off

Start measurement

Read measurement

2- or 4-stroke

Set-up menu

Page 16: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 16 of 70

7.2 Operating functions

Function Key board

push button

Display Notes

Switch On/Off

Indication after switch

“On/Off” button. The user

will be informed about

engine selection.

Basic settings will be dis-

played after switching on:

Date and time

Cyl:1 = cylinder no.1

2- or 4-stroke

Pmax in bar ; RPM: speed

After connection on

cylinder and opening on

indicator cock the Pmax

value in bar and engine

speed will be displayed.

Engine selection

Push key “2/4“ to select 2-

or 4-stroke engine

”Indication 2-stroke”

”Indication 4-stroke”

Setup

Push 3s key “2/4” to select

settings: Clear memory,

Battery type, Device info,

Measuring mode,

Selection of cylinder

Push arrow key ““ and

change the displayed value

to the number of the

selected cylinder.

”Indication cylinder no.1”

”Indication cylinder no.2”

Start measurement

During measurement, the

following information

“Recording cylinder” will

be indicated.

Display indication (1-2 s)

of measured cylinder

values. Then EPM-XPPLUS

unit switches

automatically to the next

cylinder.

Page 17: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 17 of 70

Table 3: Functions of push buttons

7.3 Install rechargeable batteries

IMES recommends to use rechargeable battery type Panasonic Eneloop AAA pro 930 mAh.

Figure 3: 4 pcs. re-chargeable battery in the transport case

Please disconnect EPM-XPplus from protection cover (Figure 4) and open battery case (Figure 5) and put re-

chargeable batteries into the battery holder (Figure 6).

Figure 4: Back side of EPM-XP incl. protection cover Figure 5: Battery case

Function Key board

push button

Display Notes

Show measurements

The measured value for

Pmax and - speed of the

selected cylinder will be

displayed.

Page 18: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 18 of 70

Figure 6: Battery holder

7.3.1 Install batteries to correct pole.

Figure 7: 4 x NiMh AAA 930 mAh installed on battery holder

Note: It is also possible to use Alkaline batteries type AAA

7.3.2 Confirm installed battery type

Start EPM-XPplus by pressing “On/Off”. The setup menu (Figure 8) asked for confirmation of installed battery

type. The standard battery type is NiMh rechargeable.

Figure 8: Selection of installed battery type

Select item by arrow key and confirm selection by pressing “2/4” button

Please check correct pole of each battery

Page 19: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 19 of 70

7.4 Charge battery

The EPM-XPplus has 4 pcs. re-chargeable battery type AAA

The battery charging (Figure 9) will start after the EPM-XPplus unit is connected via USB cable to PC and battery

voltage is lower than 5,3 V.

Figure 9: Indication during charging process

Figure 10: Indication charging process completed

The indicator can detect the charging status and continuous charging until the rechargeable batteries are fully

loaded (Figure 10). Then the charging process stops automatically. It is not recommended to interrupt the

charging process before it is automatically stopped.

7.5 Measurements with EPM-XPplus

• Start measurements on the engine

2- and 4- stroke application

Open indicator cock and close (Figure 11). Install sensor with adaptor on cylinder no.1 and open

indicator cock (Figure 12)

Figure 11: Indicator cock Figure 12: Adaptor installed on indicator cock

Press button “On/Off” at EPM-XPplus

Figure 13: Indication of 2-stroke mode Figure 14: Indication of 4-stroke mode

The EPM-XP starts always in the last selected engine mode (Figure 13 and Figure 14). Press button

“2/4” for to change adjusted engine mode.

Charging indicator

Charging completed indication

Indication of USB connection to PC

Page 20: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 20 of 70

Press “Measure” and the measurement starts after warming up sensor to minimum temperature of

50°C (Figure 16). After achieving minimum sensor temperature, the measurement starts incl. the

following indication (Figure 16)

Figure 15: Warming up sensor Figure 16: Indication during measurement

Note: It is possible to start without warming up sensor by pressing “Measure”

• Indication of measurement results

No. of cylinder Date and time engine type 2- or 4-stroke

Figure 17: Indication of measurement results on display for 3s

Engine speed Temperature of measuring element during measurement

After indication of measured cylinder values (Figure 17) the display switches automatically to next cylinder.

Repeat measuring procedure for all remaining cylinders.

WARNING

DANGER! The indicator valve ejects hot gas under high

pressure. Danger of sparks and burning. Hot gases and particles

may be ejected. Use safety gloves and safety glasses!

NOTE

The EPM-XPplus has an overheating protection of measuring cell

for temperature >300°C. In case of indication “Over temp.!”

(Figure 18) on display the indicator cock should be immediately

closed and Thompson adaptor disconnected for cooling down

sensor (10 min).

Figure 18: Indication over temperature on measuring cell of sensor

Page 21: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 21 of 70

8 Installation of visualisation- and USB driver software

For installation, the EPM_XPplus_1.3.0.2.exe file is necessary. The software installs USB driver and

visualization software in one step.

The installation will be started by double click on „ EPM_XPplus_1.3.0.2.exe“.

For installation of software the user must have administrator rights on PC.

Figure 19: Dialog of installation

Figure 20: Select destination location

The user must have administrator rights on the PC for soft- and USB driver installation.

Page 22: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 22 of 70

Figure 21: Start menu folder

Figure 22: Select additional tasks

Figure 23: Ready to install

Page 23: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 23 of 70

Figure 24: Ready to install

Figure 25: Installation completed

Page 24: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 24 of 70

9 Visualization software Start the visualization software by double click on the desktop icon or the EPM-XPplus entry in the start menu.

9.1 The Application Window

After startup the application shows the initial welcome page. The following menus and functions are available:

Figure 26: Start up window of EPM-XPplus visualization software

The menu bar at the top of the screen contains occasional needed functions.

The functional areas tab sheets give access to the following functions:

• The File tab is used for download and evaluation of measurement data from the EPM-XPplus device.

For further description see Section 9.2 ff.

• In the Online tab it is possible to watch and record live data from the EPM-XPplus device. This function

is described in Section 9.1.5.

• In the Offline tab one can view the data that was recorded in the online view. This function is described

in Section 9.1.6.

The button bar gives access to most often needed functions.

The status bar displays additional information and application messages.

Menu bar Functional areas tab sheets Button bar for fast access of main functions

Status bar

Data display area

Page 25: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 25 of 70

9.2 Connecting to the EPM-XPplus device

Please connect EPM-XPplus by USB cable to PC (Figure 27).

Figure 27: EPM- XPplus connected via USB cable to PC

In the Visualization software:

Press “Connect” button in the button bar

Figure 28: Connect button

When the connections is established, the connect button shows an activated symbol and the caption changes to

“Disconnect” (Figure 29).

Figure 29: Connect button with connection established

Page 26: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 26 of 70

9.3 Download measured data to PC

Press for to download measured data from EPM-XPplus to visualisation software.

The following dialog will appear.

Figure 30: Selection of engine configuration

Figure 31: Entering of specific engine data

Note: Engine cycle, Bore, Stroke and Conrod length are mandatory fields and must be filled with valid data.

Input to all other fields is optional.

Note: It is also possible to create engine installation templates to reduce the input of data in this step. This

function is described in chapter 9.5.10.

Note: If report templates are defined (see Section 9.5.4), it is possible to create a customized engine reports. If

none is selected a default report will be created.

Enter ship name, location, engine id

Select engine model or input engine data manual

When ready, press “OK”

Select a report template if defined (optional)

Page 27: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 27 of 70

Figure 32: Selection of directory on hard disc of PC for saving data

After selection of directory the information of downloading process will appear.

Figure 33: Downloading and TDC calculation process of measured data

9.4 Open saved data

Press button “Open” (Figure 34) for opening a stored measurement file.

Figure 34: Button “Open”

Select measurement file and klick “Open” (Figure 35)

Figure 35: Stored measurement file

Page 28: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 28 of 70

9.5 Visualization of measured data on PC

After download of measurement data from the EPM-XPplus device or opening a stored measurement file the

application switches to the File tab and displays the data.

Figure 36: File view

Cylinder selection panel:

In the cylinder selection panel one can specify for which cylinder(s) the curves should be visible and data be

displayed.

Single measurement with reference

File file opened

Figure 37: Cylinder selection panel

Info bar with engine and measurement data Cylinder selection panel Cylinder details panel

Diagram area

Display mode selection

Diagram options panel

Indication of active

cylinder(s).

Click on the name to

set a cylinder as the

active one.

The curve of the

active cylinder is shown in the diagram

area and the data is

shown in the cylinder details panel

Enable or disable check box to set visibility of

all cylinders of a series

at once

Enable check box to set

permanent visibility of

the cylinder curve in the

diagram area

Page 29: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 29 of 70

Cylinder details panel:

The Cylinder details panel displays measured and calculated data for the active cylinder(s):

Figure 38: Cylinder details panel

Hint: Theses values are measured or calculated by the software, except the scavenge air pressure. For the

scavenge air pressure the software calculates an estimated value, but this value can be overridden by the user.

When the scavenge air pressure is overridden by the user this is indicated by an underlined value.

Display mode panel:

The display mode panel specifies the diagram style to be displayed:

Figure 39: Display mode panel

• p/alpha shows the cylinder pressure over crank angle position (see section 9.5.1)

• p/vol shows cylinder pressure over cylinder volume (see section 0)

• p/vol (log) shows cylinder pressure over cylinder volume with both axis logarithmic scaled (see section

9.5.3)

Diagram area:

The diagram area displays the pressure curves in an x/y-diagram.

To get into more details, the diagram area supports zooming. To zoom in click with the left mouse button to the

upper left corner of the region of interest, drag with the mouse to the lower right corner of the region and release

the mouse button.

Figure 40: Zoom in

Pmax: maximum (peak) pressure

Pmax pos: crank angle position of the peak pressure relative to TDC

Pcomp: maximum pressure of the (calculated) compression curve

IMEP: indicated mean effective pressure

iPower: indicated power

RPM: engine speed

Scav. airpess.: scavenge air pressure

Page 30: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 30 of 70

To revert to full screen, click with the left mouse button inside the diagram area, drag with the mouse towards

the left side and release the mouse button.

Figure 41: Zoom out

Diagram options panel:

The diagram options panel allows additional functions or display of data depending on the selected display

mode. For further descriptions see the following sections.

9.5.1 P/alpha diagram

The p/alpha diagram shows the pressure as a function of crank angle position. A number of additional tools and

display option are available:

Figure 42: Functions of p/alpha diagram

Adjustment of signal filter

level for noise reduction

Set visibility of

derivative curve(s)

Set visibility of the motored

curve for the active cylinder

Set visibility of indicators

for the ignition start point(s).

Set visibility of Pmax values

for 10 consecutive cycles

Set visibility of the

cursor line tool.

Page 31: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 31 of 70

In this display mode adjustments can be made to improve the quality of the calculated results:

Figure 43: Pressure curve adjustments

Arrow function to shift curves left

or right to adjust the TDC position.

The motored curve is calculated for the new position.

Click with the left mouse

button on the ignition start indicator and drag it with

the mouse to a new position

Page 32: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 32 of 70

9.5.2 P/volume diagram

The p/volume diagram shows the pressure as a function of the cylinder volume.

Figure 44: p/volume diagram

9.5.3 P/volume (logarithmic)

The p/vol (log) diagram shows the pressure as a function of the cylinder volume with both axis logarithmic

scaled.

Figure 45: p/volume (log) diagram

Page 33: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 33 of 70

9.5.4 Engine report

The engine report gives a comprehensive view of the measured engine data.

Figure 46: Engine report diagram

The main grid lists major cylinder data. In addition to the cylinder values these statistic values are displayed:

Mean: Mean value of all cylinder values

Minimum: The minimum value of all cylinder values

Maximum: The maximum value of all cylinder values

Max. Dev.: The maximum deviation. (Maximum – Mean) or (Minimum – Mean) whichever absolute

value is larger

Max Dev. %: The maximum percentage Deviation (absolute value)

Dev. Limit: The allowable deviation limit. Cylinders with deviations exceeding this limit are

highlighted.

The columns in the main grid and the number and titles of the customizable data fields can be defined. Click the

“Configure” button. The configuration dialog (Figure 47) is shown.

Fields with grey background are calculated by the software,

blue fields can be input by the user.

Engine data

customizable data fields

Input deviation limit. All cylinders with deviations

exceeding this limit are

highlighted

Main grid

Press “IPE” button for direct data transfer to IPE software

(optional)

Page 34: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 34 of 70

Figure 47: Engine report configuration dialog

The configuration can be saved and loaded to or from a template file. When a report configuration is saved as a

template file it can also be used in the engine configuration dialog when downloading new measurements from

the EPM-XPplus device (see section 9.3).

9.5.5 Balance diagrams

Figure 48: Balance diagram

Add additional fields with

custom titles to the report

Select which columns you want

to have in the main grid

Load and store a report configuration

from or to a template file

Selection for indication of Pmax, Pcomp, IMEP and Ipower

Page 35: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 35 of 70

9.5.6 Trending

Figure 49: Overview of bush puttons

Figure 50: Selection criteria for measurement files

Figure 51: Selection of measurement files

Select data files from the same engine

Selection of 2- or 4-stroke engine

Selection of measurement files

Press “Open“

Page 36: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 36 of 70

Figure 52: Selection criteria for values and cylinder

Figure 53: Display of trending function

Selection of start- and stop date

Selection of data files with similar outputs

Deviation of data files

Selection of values

Selection of cylinder

Press “Show values”

Page 37: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 37 of 70

9.5.7 Save to ASCII format

Press “File” (Figure 54) in the menu bar and select “Export CSV” for storing data in ASCII format. The user

can select data for export to “EPM-XP csv-format” or as “Default csv-format” (Figure 55). Customers which

have an expert system of ABB/Tekomar should select “EPM-XP csv-format” for data export/import.

Figure 54: File entry in menu bar

Figure 55: CSV file selection

9.5.8 Change Engine data

Press “Edit” in the menu bar and select “Engine” to change engine settings for the current measurement file. The

engine configuration dialog (Figure 56) appears:

Figure 56: Engine Configuration Dialog

Page 38: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 38 of 70

9.5.9 Engine Database Editor

The software is supplied with a database of pre-defined engine models from several manufacturers. To add new

engine models to the database or to modify or delete models from the database, open the engine database editor

(Figure 57). Press “Tools” in the menu bar and select “Engine Database Editor”.

Figure 57: Engine Database Editor

9.5.10 Installation Template Editor

To reduce the amount of data that needs to be input when downloading a new measurement from the EPM-XPplus

device it is possible to create engine installation templates. Press “Tools” in the menu bar and select “Installation

Template Editor”. The engine installation template editor dialog box (Figure 58) opens.

Figure 58: Installation Template Editor

File Name Specification:

It is possible to specify a file name structure how the measurement files should be named. For this the following

placeholders are available:

%S : Ship/Plant name %L : Location %E : Engine ID

%Y : Year %M : Month %D : Day

%h : Hour %m : Minute %s : Second

These placeholders are replaced during the download of a measurement with actual values.

Fix texts within these fields are also possible.

During download of measured data to the pc, select the template from the drop down list in the engine

configuration dialog (Figure 59):

Insert a name for a new engine model and

input the engine data into the data fields. Click save to store the engine into the

database.

-- or --

use the drop-down list to choose an existing engine model from the database

to edit engine parameters or to remove it

from the database.

Insert a name for a new engine template and input the engine data into the data fields. Click save to store the

template.

-- or --

use the drop-down list to choose an existing template to

edit or remove it.

Select an engine model from the drop down list or input the engine parameters manually.

If report templates are defined, choose a report template

(optional). If none is selected, a default engine report

will be created.

Choose a base file path, where to store the measurement files for this engine

It is possible to specify a file name structure how the

measurement files should be named. See explanation below.

Page 39: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 39 of 70

Figure 59: Engine Configuration Dialog

9.5.11 Solver configuration

The solver configuration allows an additional method of TDC calculation. Press “Options” in the menu bar and

select Solver Configuration. The dialog (Figure 60) allows two kind of solvers for TDC calculation. Select the

appropriate solver and close the dialog with “OK”. The selected solver will be used for the next time a new

measurement is loaded from the EPM-XPplus device.

To recalculate the TDC for the actual opened measurement file check “Recalculate TDC” and press “OK”

button.

Figure 60: “Solver Configuration” Dialog

9.5.12 Color setup

The color setup allows to customize the cylinder colors. Press “Options” in the menu bar and select Color Setup.

The Color Setup dialog opens (Figure 61).

Figure 61: “Color Setup” Dialog

Press a cylinder cutton and select a new

color from the color chooser dialog

Select the template from the drop down list.

All other data fields are filled with the values from the

template.

Page 40: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 40 of 70

9.5.13 Optimize TDC calculation

For manual optimization of TDC calculation the user has the possibility to shift the TDC by use of arrow

functions (Figure 62)

Figure 62: Cylinder pressure curve incl. motored curve

Figure 63: Shifted Cylinder pressure curve incl. motored curve

Arrow function to

shift curve and

motored curve

Page 41: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 41 of 70

9.5.14 Save as reference curve

Press button “Save REF” (Figure 64) to save the currently selected curve as a reference curve.

Figure 64: Button “Save REF”

Figure 65: Save the reference on hard disc of PC

9.5.15 Open a reference file

Press button “Open REF” (Figure 66) to open a previously saved curve as a reference curve or a measurement

file as a reference file.

Figure 66: Button “Open REF”

Figure 67: Selection of reference curve on hard disc of PC

Page 42: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 42 of 70

Figure 68: Display of selected reference curve in comparison to a cylinder curve

Figure 69: Close reference curve

9.5.16 Delete measurements on EPM-XPplus

Press button “Del.Meas.” (Figure 70) to delete all measurements on EPM-XPplus.

Figure 70: Button “Del.Meas”

Page 43: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 43 of 70

9.1 Online measurements on engines with EPM-XPplus

For online measurements on an indicator cock of a selected cylinder the EPM-XPplus must be connected via USB

cable to notebook.

9.1.1 Preparation of EPM-XPplus for online measurements on 2-stroke engines

Open indicator cock and close ( ). Install sensor with adaptor on

selected cylinder and open indicator

cock (Figure 72). Start measure on EPM-XPplus and connect EPM-XPplus to notebook via USB cable (Figure 73).

Please select engine type 2-or 4 stroke on EPM-XPplus and press by arrow file the selected cylinder.

Figure 73: EPM-XPplus connected to notebook

Figure 71: Indicator cock Figure 72: Adaptor installed on indicator cock

Figure 72: Adaptor installed on indicator cock

Page 44: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 44 of 70

9.1.2 Preparation of EPM-XPplus for online measurements on 4-stroke engines

WARNING

For online measurements with EPM-XPplus on 4-stroke engines it is not allowed to use standard Thompson adaptor because of extreme high temperature. The user must use a special cooling adaptor for online measurements on 4-strokes

Please disconnect sensor from adaptor by using tools with wrench size 19 mm and 27 mm (Figure 75). Before

disconnection of sensor from adaptor spring must be disconnected from adaptor (Figure 74). with hexagon

socket screw key (2.5mm)

Figure 74: Disconnection of spring Figure 75: Disconnection of sensor from adaptor

Please connect sensor to special cooling adaptor and use wrench size 19 mm and 27 mm (Figure 76). The

tightening torque should be 25 Nm.

Figure 76: Connection of HTT-04 sensor on special cooling adaptor

19 mm wrench size tool

27 mm wrench size tool

Use with hexagon

socket screw

key(2,5 mm) for

disconnection of

spring from adaptor

Page 45: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 45 of 70

Open indicator cock and close (Figure 77). Install sensor with adaptor on the selected cylinder and open

indicator cock (Figure 78). Connect EPM-XPplus to notebook or PC via USB cable (Figure 79).

Select engine type 2-or 4 stroke on EPM-XPplus and use the arrow buttons to set the cylinder number.

Figure 77: Indicator cock Figure 78: Adaptor installed on indicator cock

Figure 79: EPM-XPplus connected via USB cable to notebook

Page 46: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 46 of 70

9.1.3 Preparation of EPM-XPplus for online measurements via Ethernet cable

In case the notebook or PC in installed more than 2m distance to EPM-XPplus it is necessary to use Ethernet cable

with max. cable length 50m incl. USB/Ethernet converters (Figure 80 and Figure 81). Please use shielded Eternet

cable (see chapter 14)

Figure 80: Overview of online measurement with EPM-XPplus

Figure 81: Overview of Ethernet connection to EPM-XPplus

EPM-XPplus

USB cable (1m)

Ethernet cable

(shielded)

Ethernet-USB extender incl.

0,2m USB cable

Ethernet-USB extender

Notebook PC

1.

2.

3.

4.

5.

Page 47: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 47 of 70

Open indicator cock and close (Figure 77). Install sensor with adaptor on selected cylinder and open indicator

cock.

Please do the following procedure:

1. Start measure on EPM-XPplus and select engine type by arrow function

2. Connect USB cable (1 m) to EPM-XPplus

3. Connect USB/Ethernet converter to USB cable (1m) and to Ethernet cable

4. Connect USB/Ethernet converter incl. USB cable (0,2m) to Ethernet cable

5. Connect USB/Ethernet converter incl. USB cable (0,2m) to PC

USB-indication should be indicated on display (see fig.76).

Figure 82: Indication on display for successful connection to PC

9.1.4 Visualization Software

Start by double click visualisation software and press “Connect” (Figure 83).

Press “Connect”

Figure 83: Connect button

When the connections is established, the connect button shows an activated symbol and the caption changes to

“Disconnect” (Figure 84Figure 29).

Figure 84: Connect button with connection established

Page 48: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 48 of 70

9.1.5 Screen page “Online”

Switch to the “Online” tab for pressure curve visualisation (Figure 85)

Figure 85: Indication of online pressure curve

During longer measurements it can occur that the measuing element gets too hot. This is indicated by a warning

message (Figure 86). To prevent damages to the EPM-XPplus device close the indicator cock and disconnect

Thomson adaptor until the temperature is below the admissible limit again.

Figure 86: Example for over temperature on measuring cell of sensor

Temperature of measuring element during measurement

Input of scavenge air pressure

Indication of engine cycle

and engine speed

Mean value of peak pressure over 100 cycles

Adjustment of TDC position

Page 49: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 49 of 70

WARNING

In case of temperature indication of more than 300 °C please close indicator cock immediately and disconnect Thompson adaptor from indicator cock for cooling down adaptor.

9.1.5.1 Save pressure curve

It is possible to save pressure curves for later investigations.

Press button “Pressure curve save” (Figure 87).

Figure 87: Button “Pressure curve save”

A dialog to select the folder where to store the file appears (Figure 88).

Figure 88: Register for saving the file

By default, the file will be saved under Press_year-month-date in the register

C:\Users\name\documents\IMES_EPMXPplus_Data\press-year-month-day

Page 50: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 50 of 70

Please press button “Stop Saving” for to stop saving pressure curves (Figure 89).

Figure 89: Indication of online pressure curve during saving process

9.1.5.2 Save Working cycle

The work cycle dialog allows to view and record the development of peak pressures of consecutive working

cycles.

Press button “Working cycle save”. (Figure 90)

Figure 90: Button “Working cycle save”

The working cycle dialog appears (Figure 91).

Page 51: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 51 of 70

Figure 91: Indication of online Pmax data

To record working cycle data to a file press button “Record” (Figure 92).

Figure 92: Record button

A dialog to select the folder where to store the file appears (Figure 93).

Figure 93: Register for saving the file

By default, the file will be saved under “WorkCycl_year_month-date” in the register

C:\Users\name\documents\IMES_EPMXPplus_DATA\WorkCycl_year-month-day.

Temperature on measuring cell

Engine cycle setting and selected

cylinder on EPM-XPplus

Page 52: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 52 of 70

During recording the number of recorded cycles is shown.

To stop recording, press button “Stop recording” (Figure 94).

Figure 94: Indication of Pmax data during recording process

During longer measurements it can occur that the measuing element gets too hot. This is indicated by a warning

message (Figure 95Figure 86). To prevent damages to the EPM-XPplus device close the indicator cock and

disconnect Thomson adaptor until the temperature is below the admissible limit again.

Figure 95: Example for over temperature on measuring cell of sensor

WARNING

In case of temperature indication of more than 300 °C please close indicator cock immediately and disconnect Thompson adaptor from indicator cock for cooling down adaptor.

“Stop recording” button Indication of number of recorded cycles

Page 53: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 53 of 70

9.1.6 Screen page “Offline”

Select screen page “Offline” (fig.90)

Figure 96: Screen page “Offline”

9.1.6.1 Open saved pressure curves

Press button “100 Pressure curve open”. (Figure 97)

The file will be opened under “Press_year_month-date” in the register

C:\Users\name\documents\IMES_EPMXPplus_DATA\WorkCycl_year_month_data (Figure 98)

Figure 97: Button “100 Pressure curve open”

Figure 98: Register of saved file

Page 54: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 54 of 70

Buttons for cycle selection and indication

Figure 99: Indication of pressure cycles

9.1.6.2 Open saved working cycles

Click on button “Working cycle open” (Figure 100)

Figure 100: Button “Working cycle open”

Select “Open file” (Figure 101)

Figure 101: Screen page “working cycle offline”

Press button “Working cycle open”. (Figure 101)

The file will be opened under “WorkCycl_year_month-date” in the register

C:\Users\name\documents\IMES_EPMXPplus_DATA\WorkCycl_year_month_data (Figure 102)

Indication of selected cycle no.

Page 55: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 55 of 70

Figure 102: Register of saved working cycle data file

Please select start- and end time and press “Load data”

Figure 103: Selection of saved Pmax working cycle

Figure 104: Indication of selected Pmax data

Page 56: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 56 of 70

9.2 Update firmware to EPM-XPplus from visualisation software

In case a new firmware update for EPM-XPplus is available. The user has the possibility to update the EPM-XPplus

to new firmware.

The user has to do the following.

1. Press “On” at EPM-XPplus

2. Connect EPM-XPplus to PC by USB cable

3. Start visualisation software and press “Connect” (Figure 105)

4. Select “Tools” (Figure 106) and select “Firmware Update” (Figure 107)

5. Load firmware (Figure 108) from register C:\Users\name\documents\

6. Update firmware

7. Press On/OFF at EPM-XP for to check actual firmware release

8. Press 3s “2/4” button and select “device info (Figure 109 + Figure 110 + Figure 111)

Figure 105: Button “Connect” Figure 106: Selection “Tools”

Figure 107: Selection “Firmware” Figure 108: Load-and update firmware

9.2.1 Check actual firmware version

Press “On/Off” EPM-XP plus and press 3s button “2/4” the main menu will appear (Figure 109)

Figure 109: Indication of main menu

Select by arrow function “device info” and select with button “2/4” (Figure 110)

Page 57: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 57 of 70

Figure 110: Indication of device info

Select with button “2/4” to indicate firmware version (Figure 111)

Figure 111: Indication of actual firmware

10 Accuracy check The EPM-XPplus is adjusted at IMES workshop at different pressure values (0…300 bar) and temperature values

(50°C- 200°C). The values are described at calibration certificate (Figure 112). A calibration certificate will be

delivered with each EPM-XPplus unit.

The EPM-XPplus has a very stable and long life cylinder pressure sensor type HTT. We emphasize to check

calibration values of EPM-XPplus on static pressure calibrator by hydraulic oil once a year.

Figure 112: Calibration certificate of EPM-XPplus unit

1545-10-0005

Page 58: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 58 of 70

10.1 Check on pressure calibrator

We emphasize to mount thompson adaptor of EPM-XPplus direct on a pressure plate (Figure 113) which is

connected by hydraulic hose to pressure generator.

Figure 113: Thompson Adaptor mounted on pressure plate

We emphasize to check EPM-XPplus at most frequently measured Pmax-value (Figure 114) on engine.

For example pressure values at 0 bar and between 50 bar – and 200 bar (Figure 115).

Figure 114: Example for indication of static pressure @ 0 bar

NOTE

Remarks: The pressure plate (product-no.: IW-1574) is not in the scope of supply of EPM-XPplus and can be ordered additional by IMES

Pressure connection thread G1/4“ to

hydraulic hose to pressure generator

Page 59: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 59 of 70

Figure 115: Example for indication of static pressure @ 50 bar

Static pressure checking procedure on EPM-XPplus:

a) Switch – On by pressing operating key “On/Off “ on EPM-XPplus

b) Press 3s button “2/4” for to enter into main menu (Figure 116)

Figure 116: Indication of main menu

c) Select by arrow function “Measuring mode” (Figure 117) and confirm with button “2/4”

Figure 117: Indication of measuring mode

d) Select “Sensor test” (Figure 118) and confirm with button “2/4”

Figure 118: Indication of sensor test

e) Confirm with button “2/4” and change to “Exit” by arrow function and confirm with button “2/4”. The

following indication @ 0 bar will be displayed after 3-5 s on display (Figure 119)

Figure 119: Indication @ 0 bar on display

Page 60: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 60 of 70

f) Generate pressure calibrator for example 50 bar and compare displayed pressure value on EPM-XPplus

with indicated pressure value on pressure generator (Figure 120 and Figure 121)

Figure 120: Indication @ 50 bar Figure 121: Indication on pressure pump

10.1 Example of static pressure indication

Display on pressure generator [bar] Display on EPM-XPplus Dev. in [bar]

0 0 0

100 99 1

200 198 2

Table 4: Static pressure comparison

NOTE

The max. acceptable difference between adjusted pressure at pressure calibrator by ambient (20-25°C) temperature and displayed static pressure at EPM-XPplus is +/- 2.5 bar. By increasing pressure difference we emphasize to send the unit to IMES workshop for new adjustment and calibration.

Page 61: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 61 of 70

11 Cleaning

11.1 Periodically cleaning

Clean gas channel of Thompson adaptor by compressed air periodically (see Figure 122)

Figure 122: Periodical cleaning of gas channel

WARNING

Before starting cleaning procedure, the temperature of complete adaptor must have ambient temperature. It is not allowed to start cleaning process short after measurement on engine.

11.2 Cleaning in case of hard deposit

For elimination of hard deposit please disconnect sensor from adaptor by using tools with wrench size 19 mm

and 27 mm (Figure 124). Before disconnection of sensor from adaptor spring must be disconnected from adaptor

(Figure 123) with hexagon socket screw key (2.5mm)

Figure 123: Disconnection of spring Figure 124: Disconnection of sensor from adaptor

27 mm wrench size tool

19 mm wrench size tool Use hexagon socket

screw key (2,5 mm)

for disconnection of

spring from adaptor

Page 62: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 62 of 70

11.3 Cleaning procedure of adaptor

Please use a drill with 2.5 mm diameter (Figure 125) for to clean gas channel of adaptor.

Figure 125: Drill 2.5 mm Figure 126: Drill in gas channel of adaptor for cleaning

11.4 Cleaning procedure of cylinder pressure sensor

Please use the special drill tool (Figure 127). and clean gas channel of HTT cylinder pressure sensor. The length

of gas channel is 29 mm. The max. entering length of special drill tool into gas channel is 28 mm which is

marked. Do not try to drill more than 28 mm into gas channel!

Figure 127: Special drill tool Figure 128: Special drill tool in gas channel of sensor

Page 63: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 63 of 70

12 Basic check for fault finding

Problem Indication on EPM-XP/CAT Action

No display indication after switch

“On”

Charge battery by connecting

to USB port from EPM-XPplus

to PC

Over temperature indication

Indication after temperature of

measuring cell > 300 °C

Disconnect sensor from

indicator cock and wait 5-10

minutes

Sensor defect

Sensor is defect.

Please send EPM-XPplus unit

to IMES workshop

Table 5: Basic check for fault finding

13 Nomenclature

Item Description

RPM Engine speed

Pmax Maximum pressure [bar]

Pcomp Compression pressure [bar]

dp/ca Maximum rise of pressure curve [bar/ca]

Cycles Number of working cycles

IMEP Mean indicated pressure [bar]

Ipower Indicated power [kW] Table 6: Nomenclature

Page 64: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 64 of 70

14 Option

14.1 IMES IPE Performance evaluation software

- Order no. IW-8050 License for one 2-stroke engine and four 4-stroke engines (same type).

Figure 129: Example IPE software

The IMES performance evaluation software (IPE) is designed to facilitate the collection, evaluation,

management and comparison of engine performance data for marine diesel engines. It loads recorded

cylinder pressure data from EPM-XP plus directly. Additional required information and parameters have to

be entered so the programme can do ISO corrections and compare against new-engine performance

benchmarks. Performance graphs and reports give a quick status of an engine and suggest actions to take

for optimising engine conditions.

Page 65: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 65 of 70

14.2 Ethernet cable 20m

- Order no. IW-8130

Figure 130: Ethernet cable

Page 66: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 66 of 70

14.3 USB Extender for connection to Ethernet cable

- Order no. IW-1559

Figure 131: USB-Ethernet Extender

Page 67: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 67 of 70

14.4 Thompson adaptor for 4-stroke continuous online measurement

- Order no. IW-1222

Figure 132: Thompson adaptor for 4-stroke online measurement

Page 68: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 68 of 70

14.5 Pressure pump coupling for Thompson adaptor

- Order no. IW-1574

Figure 133: Pressure pump coupling

Pressure connection thread G1/4“ to hydraulic house

to pressure generator

Pressure connection thread W27x1/10 for Thompson

adaptor

Page 69: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 69 of 70

14.6 Thread cleaner for Thompson thread

- Order no. IW-1571

Figure 134: Thread cleaner for Thompson thread

Page 70: Original operating instruction Electronic Analyser

Operation Manual EPM-XPplus

Software Release 1.3.0.2 25.09.2020

Page 70 of 70

15 EU Declaration of Conformity In accordance with EN ISO 17050-1:2010

Manufacturer: IMES GmbH

Dr.-Herbert-Kittel-Str.2

D- 87600 Kaufbeuern

Name: Electronic Indicator

Type: EPM-XPPLUS

The object of the declaration described above is in conformity with the relevant EU

harmonisation legislation:

2014/30/EU Electromagnetic Compatibility Directive

2014/35/EU Low Voltage Directive

2011/65/EU Restrictions of Hazardous Substances Directive

Conformity is shown by compliance with the applicable requirements of the following

documents:

DIN EN 61326, including

DIN EN 61000-4-2

DIN EN 61000-4-3

DIN EN 61000-4-4

DIN EN 61000-4-5

DIN EN 61000-4-6

DIN EN 61000-4-11

Signed for and in behalf of:

IMES GmbH

Dr.-Herbert-Kittel-Str.2

D-87600 Kaufbeuren

Tel. +49(0)8341-966173 0

Fax.+49(0)8341-96617318

Kaufbeuren, 02nd. April 2019 Stefan Neumann

(Managing director)