Wet Gas Flow Measurement with Conditioning Orifice Meter Flow
Orifice Meter Diagnostics – Theory, Laboratory Tests...
Transcript of Orifice Meter Diagnostics – Theory, Laboratory Tests...
![Page 1: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/1.jpg)
Orifice Meter Diagnostics– Theory, Laboratory Tests,
Operator Interface& Field Results
Kim LewisDP Diagnostics
by
![Page 2: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/2.jpg)
![Page 3: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/3.jpg)
Introduction
• Orifice meters are as popular as they are simple,generally inexpensive & reliable, however…
there is a long held axiom throughout industry thatorifice meters have no internal diagnostics.
• A flow meter diagnostic system is desirable forseveral reasons:
– guarantees meter performance thereforereducing exposure to measurement error,
![Page 4: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/4.jpg)
Introduction (cont.)
– allows condition based maintenance (CBM)instead of routine scheduled maintenance (RSM):
o avoiding unnecessary risky routine scheduledmaintenance on HP systems!
… if it ain’t broke don’t try and fix it,
- there is always the risk of loss of containment.
– CBM makes technicians in the field hugely moreproductive & efficient. They go to a meter with aproblem every time, instead of blindly searchingfor problems!
![Page 5: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/5.jpg)
Introduction (cont.)
– CBM decreases the Safety Factor for Technicians
• Less Time Behind the Windshield
• Reduces the Exposure to Hazardous Conditions
– Unnecessary to Break Containment during theInspection of the meter
– Reduced Exposure to H2S Hazards
![Page 6: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/6.jpg)
Traditional DP Orifice Meter
flow
P2P1
DP
![Page 7: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/7.jpg)
Add a downstream tapping
flow
P2 P3P1
DP
![Page 8: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/8.jpg)
Two Extra DP Transmitters
flow
P2 P3P1
DP
DPr
DPppl
![Page 9: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/9.jpg)
DPt = DPr + DPppl
![Page 10: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/10.jpg)
The DP Diagnostics method:
![Page 11: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/11.jpg)
Three Flowrate Predictions
flow
P2 P3P1
DPt
DPr
DPppl
mr = kr DPr
mppl = kppl DPppl
mt = k DPt
![Page 12: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/12.jpg)
Three DP Pairings plus DP Ratios
flow
P2 P3P1
DPt
DPr
DPppl
DPppl / DPt = PLR
DPr / DPt = PRR
DPr / DPppl = RPR
![Page 13: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/13.jpg)
3 DP’s produces 3 DP pairs & 3 DP ratios:
i.e. “PLR” “PRR” “RPR”
• DP ratios are set characteristics of the meter.Hence, we can compare each “read” value to theexpected value.
• All orifice meter diagnostic characteristics arederivable from the standards.
![Page 14: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/14.jpg)
4”, 0.5 beta ratioorifice plate data
•CEESI Double Chamberwith Daniel Plate, NaturalGas, 112 pts
• 2008: Daniel PaddlePlates, Air, 44 pts
• 2009: Yokogawa PaddlePlates, Air, 124 pts
![Page 15: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/15.jpg)
![Page 16: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/16.jpg)
![Page 17: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/17.jpg)
Sample Actual Baseline Result, in a NDB:
![Page 18: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/18.jpg)
Orifice Meter Malfunction Examples
• 4”, 0.5 β paddle plate orifice meters with air flow were given various malfunctions at CEESI.
• Each example shows the scale of the induced bias &the corresponding diagnostic response.
• To put the induced bias in industrial context themonetary value of such a bias for a mid-rangenatural gas flow through such a meter is also given.
1 MMSCF is approx $3,500 /day
![Page 19: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/19.jpg)
![Page 20: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/20.jpg)
4”, 0.5 β Orifice Meter with Error in Inlet Diameter Keypad Entry
+1.2% bias > +$13.3K / quarter
![Page 21: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/21.jpg)
4”, 0.5 β Orifice Meter with Error in Orifice Diameter Keypad Entry
-2.6% bias > -$28.5K / quarter
![Page 22: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/22.jpg)
4”, 0.5 β Orifice Meter Buckled Plate
-7% bias > -$78K / quarter
![Page 23: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/23.jpg)
4”, 0.5 β Orifice Meter Contaminated Plate
-3.5% bias > -$39K / quarter
![Page 24: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/24.jpg)
4”, 0.5 β Orifice Meter Worn Edge Plate
-5% > -$55K /quarter
![Page 25: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/25.jpg)
4”, 0.5 β Orifice Meter Backwards Plate
-15% > -$167K / quarter
![Page 26: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/26.jpg)
4”, 0.4 β Orifice Meter with Trapped Debris
+118% > +$1.3e6 / quarter
![Page 27: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/27.jpg)
Disturbed Flow 4”, 0.5 β Orifice Meter – HMOP at 2D upstream
-5.8% > -$64K / quarter
![Page 28: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/28.jpg)
4”, 0.62 β Orifice Meter with Wet Natural Gas Flow (13.7 MMSCFD)
P = 42.6 Bar
Gas Flow = 3.3 kg/s
Liquid Flow = 0.4 kg/s
GVF = 98.9%
Meter over-reads gas flow rate by +4%
+4% > +$172K / quarter
![Page 29: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/29.jpg)
Field Data
Flow Computer
MeteringSupervisory PC
+ DiagnosticsSoftware
Additional I/O
MeteringSupervisory PC
Diagnostics PC
Applying the DP Meter Diagnostics
Swinton Technology has developed the DPDconcept into a product called “Prognosis”.
![Page 30: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/30.jpg)
ATMOS Training Facility with OrificeMeter & Diagnostic System (Prognosis)
![Page 31: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/31.jpg)
2”, 0.54 β Orifice Meter, 147 psig, 410 MSCFD
![Page 32: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/32.jpg)
An Over-Ranged (‘Saturated’) DP Transmitter147.5 psig, 412 MSCFD, Actual DPt 25.8 “WC,
Read Value 23.0”WC, Induced Error -5.6%
![Page 33: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/33.jpg)
Unseated Orifice Plate in Carrier
![Page 34: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/34.jpg)
Unseated Orifice Plate in Carrier417.5psig, 471 MSCFD, Induced Error -15%
![Page 35: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/35.jpg)
![Page 36: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/36.jpg)
Pipe ID 13.738”d=8.195” (0.596 beta)
201.5 MMSCFD ≈ $705K/day
BP CATS, May 2010
![Page 37: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/37.jpg)
Baseline Results at BP CATS (as found)
![Page 38: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/38.jpg)
BP CATS Reversed Orifice Plate(-15% flow rate prediction error)
-15% > -$105K/day
or > -$9.5 million / quarter
![Page 39: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/39.jpg)
BP CATS - Saturated DP Transmitter(-6.2% flow rate prediction error)
•Tradition DP transmitterSpanned 15.0kPa / 60”WC
•DPt = 17.5kPa / 70”WCa -7.4% flowrate error
•Also with 3 DP transmitters:
-7.4% > -$4.6 million / quarter
kPakPa 3.112.60.15
![Page 40: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/40.jpg)
BP CATS - Worn Orifice Edge
• Plate deliberately worn, -2% error induced
-2% > -$1.25 million / quarter
![Page 41: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/41.jpg)
Efficient Audits
Improved Efficiency& Reduced CarbonFootprint
BP’s & CoP’s Observed Benefits
Assuranceof Complianceand Accuracy
Less RoutineInspections in RemoteOperation
Less Safety Risk& less logistics
![Page 42: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/42.jpg)
Orifice Meter Audit 1 – Wet Gas Flow Found
![Page 43: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/43.jpg)
Orifice Meter Audit 2 – Serviceable Meter
![Page 44: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/44.jpg)
Orifice Meter Audit 3 –Contamination Found
![Page 45: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/45.jpg)
CoP Franklyn Gas Station 10” Orifice Meter
![Page 46: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/46.jpg)
Correct Meter Operation
![Page 47: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/47.jpg)
Backwards Plate
![Page 48: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/48.jpg)
Faulty DPt Reading– Leaking 5 Way Manifold Valve
![Page 49: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/49.jpg)
Erroneous Orifice Diameter4.25” entered instead of 4.5”
![Page 50: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/50.jpg)
ConocoPhilips / Rosneft “Polar Lights” Field
![Page 51: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/51.jpg)
![Page 52: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/52.jpg)
Metering Stream thought to be a Pristine Meter Run
![Page 53: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/53.jpg)
![Page 54: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/54.jpg)
Trending Case Study- Wet Gas Flow Metering
• The diagnostics showtrending in meterperformance, e.g. theycan show a problemgetting better or worse,
e.g. a wearing sharp edge,contamination build up,DP transmitter drift,wetness of a gas…
![Page 55: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/55.jpg)
4”, 0.683β Orifice Meter72 Bar (a), 40C, 15 MMSCFD
![Page 56: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/56.jpg)
CoP Jasmine 12”, 0.636β Orifice Meter Test Separator Gas Outlet
![Page 57: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/57.jpg)
![Page 58: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/58.jpg)
Conclusions
• DP Diagnostics has developed and proven anorifice meter diagnostic tool called “Prognosis”.
• Prognosis can:
– assure correct orifice meter operation
– reduce exposure to mis-measurements
– increase the productivity of field technicians
– limit risk by limiting maintenance to orificemeters that are shown to require it only!
![Page 59: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/59.jpg)
• Prognosis has been laboratory tested by:
– CEESI (Colorado & Iowa)
– NEL (UK) confidential
– ATMOS (Texas)
– CIATEQ (Mexico)
• Prognosis has been field tested at:
– BP CATS
– CoP TGT
– CoP Franklyn
– UK National Grid (Corby)
![Page 60: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/60.jpg)
• In the 5th year on the market Prognosis is alreadybeen procured by:
BP, ConocoPhillips, Centrica, Petronas, Yanpet(Exxon/Saudi Aramco), E.ON Ruhrgas,UK National Grid, GDF Suez, Perenco, Talisman,EnCana, Nova Chemical, Maersk + others.
• The concept is now mentioned in:
– the UK governments DECC “Guidelines forPetroleum Measurement”
– The Canadian regulators “Directive 17”
![Page 61: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/61.jpg)
PROGNOSIS END USERS
![Page 62: Orifice Meter Diagnostics – Theory, Laboratory Tests ...rmmsociety.org/wp-content/uploads/2015/09/Orifice-Meter...Orifice Meter Malfunction Examples •4”, 0.5 βpaddle plate orifice](https://reader034.fdocuments.net/reader034/viewer/2022051407/5aeaa0d67f8b9a66258c4e2a/html5/thumbnails/62.jpg)
Thank You.