Optimal Device Performance Guide ETHICON ENDO-SURGERY™ Linear Cutter (NTLC)

Transcript of Optimal Device Performance Guide - Global Preferences · Optimal Device Performance Guide ETHICON...

Page 1: Optimal Device Performance Guide - Global Preferences · Optimal Device Performance Guide ETHICON ENDO-SURGERY™ Linear Cutter (NTLC) 2 Two-sided firing Intermediate closure ETHICON

Optimal Device Performance GuideETHICON ENDO-SURGERY™ Linear Cutter (NTLC)

Page 2: Optimal Device Performance Guide - Global Preferences · Optimal Device Performance Guide ETHICON ENDO-SURGERY™ Linear Cutter (NTLC) 2 Two-sided firing Intermediate closure ETHICON

2

Two-sided firing

Intermediate closure

ETHICON ENDO-SURGERY™ Linear Cutter (NTLC)

Non-slip grip surface

Staple height selector

6-row cartridge/3-D surgical staple

Product codes: NTLC55, NTLC75Reload codes: SR55, SR75

Page 3: Optimal Device Performance Guide - Global Preferences · Optimal Device Performance Guide ETHICON ENDO-SURGERY™ Linear Cutter (NTLC) 2 Two-sided firing Intermediate closure ETHICON

3

Cartridge loading

1. Linear cutter and cartridge are packaged separately. Keep the retaining cap on the cartridge.

3. Device alignment/locking lever should be completely open before loading a new cartridge.

5. Remove safety retaining cap from the cartridge before firing. Leave device fully open until ready to be placed on the tissue.

This device is indicated for 12 firings.

2. Ensure that the firing knob is in the proper position before loading.

4. Load the cartridge with the staple retaining cap on from the nose of the device first, otherwise, it may lock out the new cartridge.

Note: Do not remove staple retaining cap from staged reloads until ready to install. This prevents the possibility of nuisance lock-outs and firing issues due to mixing of used and unused reloads.

Page 4: Optimal Device Performance Guide - Global Preferences · Optimal Device Performance Guide ETHICON ENDO-SURGERY™ Linear Cutter (NTLC) 2 Two-sided firing Intermediate closure ETHICON

4

Assembly

Staple height selection

1. When assembling the device, work proximal to distal to ensure the channel post is aligned within the plastic casing of the cartridge.

1. Ensure the device is fully in the open or intermediate locking position.

2. A desired staple height must be selected before the device is closed or firing sequence is initiated.

Staple height selection can be changed after closing the device, as long as firing knob has not been moved.

Page 5: Optimal Device Performance Guide - Global Preferences · Optimal Device Performance Guide ETHICON ENDO-SURGERY™ Linear Cutter (NTLC) 2 Two-sided firing Intermediate closure ETHICON

5

Firing

1. Surgeons may use their thumb, palm of their hand, or thumb and forefinger to fire the linear cutter.

2. During subsequent firing: After completing one firing sequence, return knob to the original pre-firing “RETURN KNOB HERE” position, aligning the two white arrows.

Once the surgeon has moved the firing knob to either side, they are committed to firing on that side.

Cartridge unloading

1. Open device holding handle. Use cartridge shoulder to remove cartridge.

2. Swish after each firing (without cartridge).

Page 6: Optimal Device Performance Guide - Global Preferences · Optimal Device Performance Guide ETHICON ENDO-SURGERY™ Linear Cutter (NTLC) 2 Two-sided firing Intermediate closure ETHICON

6

For complete product information, including full steps for use, indications, contraindications, warnings and precautions, please see the Instructions for Use.

© 2015 Ethicon, Inc.. All Rights Reserved. 041341-151001

Determining a cartridge lockout

1. Look at the yellow bar on the cartridge fork.

If you see the yellow bar laterally, the cartridge is unlocked and is ready to be fired.

2. If you do not see the yellow bar, the cartridge has been used or is locked out, and it needs a new reload.