Operator Manual - ELGA LabWater

35
WATER PURIFICATION Operator Manual ® MANU41319 VERSION D 08/20 Part No. LA824 DRAFT 07/8/2020

Transcript of Operator Manual - ELGA LabWater

Page 1: Operator Manual - ELGA LabWater

WATER PURIFICATION

Operator Manual

®

MANU41319 VERSION D 08/20Part No. LA824

DRAFT

07/8/2

020

Page 2: Operator Manual - ELGA LabWater

YOUR GUIDE TO THE CHORUS

Page 3: Operator Manual - ELGA LabWater

Quick Start Guide Page 1 - 151. Manual Introduction Page 161.1 Health and Safety Page 161.2 Product Model Page 161.3 Use of this Manual Page 161.4 Installation Page 161.5 Environment Page 161.6 Commissioning Page 161.7 Customer Support Page 162. Guide to Your PURELAB® Dispenser Page 17

3. Main Display Settings and Button Operation Page 184. Product Consumables Page 195. Register Your Product Page 196. Quick Reference Guide Page 20 - 216.1 Manual Dispense Page 206.2 Continuous Dispense Page 206.3 Auto Volume Dispense Page 206.4 Deactivating - Auto Volume Dispense Page 216.5 ON - OFF, Main Menu and Accept Page 217. Maintenance Page 227.1 Replacing the Point-of-Use Filters LC134 / LC145 / LC197 Page 228. Customize Operation Page 238.1 Set Alarm Points Page 238.2 Advanced Data Logging Page 238.3 Software Update Page 239. Troubleshooting Page 24 - 2510. Technical Specifications Page 2611. Warranty / Conditions of Sale Page 27

Copyright Note

The information contained in this document is the property of VWS (UK) Ltd, trading as ELGA VEOLIA , and is supplied without liability for errors or omissions.

No part of this document may be reproduced or used except as authorized by contract or other written permission from VWS (UK) Ltd.

The copyright and all restrictions on reproducing and use apply to all media in which this information may be placed.

VWS (UK) Ltd. Pursue a policy of continual product improvement and reserve the right to alter without notice the specification, design,

price or conditions of supply of any product or service.

© VWS (UK) Ltd. 2020 - All rights reserved. ELGA and PURELAB are registered trademarks of VWS (UK) Ltd.

CONTENT

Page 4: Operator Manual - ELGA LabWater

Environment and Positioning

Control of Substances Hazardous to Health (COSHH)

Personal Protective Equipment (PPE)

Electricity

Clean dry indoor, temp 5 - 40°c, humidity max 80%, non-condensing.Caution!

Do not connect to a pressurised water supply (potable or pretreated).Failure to do so could result in damage to the installation.

WARNING!

The appliance coupler (mains lead) or power supply connected to the rear of the unit can be removed to isolate the power supply. If access to this is restricted then it is recommended that access to the mains socket is easily available to disconnect the power supply.

Please follow the safety information detailed below for the Quick Start Guide and Operator Manual.

WARNINGS ARE GIVEN WHERE FAILING TO OBSERVE THE INSTRUCTIONS COULD RESULT IN INJURY OR FATALITY.

Cautions are given where failure to observe the instructions could result in damage to the equipment, associatedequipment and processes.

Material safety data sheets covering the consumables are available upon request. Note: Ensure the consumables are disposed of in accordance with local regulations.

Caution!PURELAB Dispenser is not designed for use in fume cupboards where chemicals could damage the product.

Caution!Installed on a flat level worktop, PURELAB Dispenser can also be securely fitted to a worktop using themounting bracket provided, please see Quick Start Guide - section 2 for details.

WARNING!POSITION THE POWER SUPPLY SO THAT IT CANNOT COME INTO CONTACT WITH WATER.

MAINTENANCE MUST BE CARRIED OUT WITH PROPER PROTECTIVE EQUIPMENT THAT INCLUDES STERILE LATEX NITRILE GLOVES CAT 3.

WARNING!

ONLY USE THE APPLIANCE COUPLER (MAINS LEAD) AND POWER SUPPLY PROVIDED, USE OF THESE WILLENSURE ADEQUATE EARTH PROTECTION. DISCONNECT ELECTRICAL MAINS POWER SUPPLY BEFORE ANYMAINTENANCE WORK IS STARTED. IF THE EQUIPMENT IS USED IN A MANNER NOT SPECIFIED BY ELGA VEOLIA , THE PROTECTION PROVIDED BY THE EQUIPMENT MAY BE IMPAIRED.

WARNING!

HEALTH AND SAFETY NOTICES

®

®

Page 5: Operator Manual - ELGA LabWater

Items not supplied

EN14126

EN388 & EN374Tape Measure Plastic Pipe Cutters

Tube Keys GogglesØ3mmDrill BitDrill

Phillips Screw Driver

Power

Handset Labels

COMMS

1 X

1 X

SUPPLIED PARTS

HEALTH AND SAFETY NOTICES

Page 6: Operator Manual - ELGA LabWater

QUICK START GUIDE ENGLISH

WELCOME

HEALTH & SAFETY

Thank you for purchasing a PURELAB Dispenser, please read the Health and Safety notice before proceeding with this installation.

Dispenser Handset

LCD Display

Screen

FlexibleDispenser Hose

Auto Volume Dispense Button

Dispense Button

Water Purity Display

Scrolling information bar -bar indicating the status.

Auto Volume or Reservoir Volume

Up Button

Accept Button

Down Button

Dispenser Tip

Worktop Mounting

Bracket

Handset Cradle

Height Adjustment

Button

Weighted Base

Please follow the safety information detailed in the front of the Quick Start Guide. Failure to observe these instructions could result in damage to the equipment and associated equipment resulting injury or death.

ELGA VEOLIA Tel: +44 (0) 203 567 7300 Email: [email protected] Website: www.elgalabwater.comUnit 10 Lane End Industrial Park, Lane End, High Wycombe, HP14 3BY

ELGA VEOLIA is the global laboratory water brand name of Veolia. PURELAB is an ELGA trademark and technology.Owning to a policy of continual improvement, we reserve the right to amend the specifications given in this document. 2020 ELGA VEOLIA / VWS (UK) Ltd. All rights reserved.

®

COMMS Port

COMMS Port

Port 5: ReturnPower

24V DC Power

24V DCPort 1:

Inlet

®

® ®

PURELAB® Dispenser Operator Manual MANU41319 Version D 08/20 Page 1

Page 7: Operator Manual - ELGA LabWater

Quest Port 1: Feedwater Inlet

Quest Port 2: Overflow

Quest Port 3: Drain

Quest Port 6: Dispenser Return

Quest Port 7: Dispenser Outlet

Dispenser Port 1: Inlet

Dispenser Port 5: Return

Example 1: 1 X PURELAB Quest RO/DI/UV or RO/DI - 1 X PURELAB Dispenser LA824 (Max 1 Dispenser) - LA648 Boost Pump®

SYSTEM CONFIGURATION - PURELAB QUEST

COMMS

Key

Power

Water

15mm Valve

8mm Flow Bend

8mm Tube Clip

Regulator LA512

Optional - LA648 Boost Pump - for feedwater below <2.0 bar

Air Break Device

27165

Max Distance 5000mm

Inlet Outlet

Quest

Dispenser Type 1

Type 1

LA648

MAX 1M

Potable Feedwater Connection

Quest Port 1: Feedwater Inlet

Quest Port 2: Overflow

Quest Port 3: Drain

Quest Port 6: Dispenser Return

Quest Port 7: Dispenser Outlet

Dispenser Port 1: Inlet

Dispenser Port 5: Return

6

5

7

Example 2: 1 X PURELAB Quest RO/DI/UV or RO/DI - 1 X PURELAB Dispenser LA824 - Wall Mounting Bracket LA735 ®®

Wall Mounting Bracket LA735 (INST39088)

Air Break Device 2

Max Distance 5000mm Quest

Type 1

Dispenser Type 1

Potable Feedwater Connection

®

Existing system configuration: If you are connecting a PURELAB Dispenser to an existing system configuration, please ensure all of your products

are up to date and using the latest software. To find the software version of your existing system, simply turn the power off by the wall and back on

again and the current software version number will be displayed in the bottom right-hand corner of the screen. If your software version number is

older than the one listed on the website at www.elgalabwater.com/customize, then please perform a software update,

see section 8.3 'Software Update' in this operator manual.

®

PURELAB® Dispenser Operator Manual MANU41319 Version D 08/20 Page 2

Page 8: Operator Manual - ELGA LabWater

SYSTEM CONFIGURATION - PURELAB CHORUS 1 ANR/LSC/GSC

Chorus Port 1: Feedwater Inlet

Chorus Port 3: Drain

Chorus Port 4: Product Outlet

Chorus Port 5: Return

Chorus Port 6: Power

Chorus Port 7: Comms

Chorus Port 8: Power

Chorus Port 9: Comms

Chorus Reservoir Port 1: Inlet

Chorus Reservoir Port 4a: Outlet

Chorus Reservoir Port 4b: Outlet

Chorus Reservoir J1: Comms

Chorus Reservoir J2: Comms

Dispenser Port 1: Inlet

Dispenser Port 5: Return

Halo Dispenser Port 1: Inlet

Halo Dispenser Port 5: Return

Halo Dispenser Port 6: Power

Halo Dispenser Port 7: Comms

Halo Dispenser Port 8: Power

Halo Dispenser Port 9: Comms

Example 1: 2 X PURELAB Chorus 3 RO or 2 X PURELAB Chorus 2 RO/DI - 1 X PURELAB Chorus 1 - 3 X PURELAB Dispensers LA824

(Max 4 Dispensers) - LA759 60L Reservoir or LA760 100L Reservoir

® ® ® ®

3

4

6

9

7

777

9

5555

Air Break DeviceAir Break Device

Chorus 3 RO or Chorus 2 RO/DI fills an external reservoir with purified Type 3 or Type 2 water, which then feeds a Chorus 1 to produce Type 1 water.

New system configuration:

1. Commission Chorus 3 RO or Chorus 2 RO/DI systems first, directing reservoir port: 4a to drain during this procedure.

(Please refer to operator manual Chorus 3 RO MANU39996 or Chorus 2 RO/DI MANU40003 for commissioning instructions)

2. Once the reservoir is full, open reservoir port:4a manual valve to Chorus 1 port 1: Feedwater inlet and continue to commission Chorus 1

with the PURELAB Dispensers connected.

3. Chorus 1 is set in pre-commisioning mode and will automatically direct water to drain during this procedure.

(Please refer to operator manual Chorus 1 MANU39998 for commissioning instructions)

Existing system configuration: If you are connecting a PURELAB Dispenser to an existing system configuration, please ensure all of your products

are up to date and using the latest software. To find the software version of your existing system, simply turn the power off by the wall and back on

again and the current software version number will be displayed in the bottom right-hand corner of the screen. If your software version number is

older than the one listed on the website at www.elgalabwater.com/customize, then please perform a software update,

see section 8.3 'Software Update' in this operator manual.

Potable Mains Feedwater Connection

Air Break Device

4J1 J2

4a

4b

43 3

8 66

Reservoir Overflow

COMMS

Key

Power

Water

Valve

15mm Valve

8mm Flow Bend

8mm Tube Clip

Regulator LA512

8mm Equal T Dispenser Type 1 Dispenser Type 1 Dispenser Type 1

Chorus 3 RO or 2 RO/DI Chorus 3 RO or 2 RO/DI Reservoir 60L or 100L

Chorus 1

Type 3 or 2

Max Distance 5000mm Max Distance 5000mm

®

Outlet to Application

Air Break Device

PURELAB® Dispenser Operator Manual MANU41319 Version D 08/20 Page 3

®

Page 9: Operator Manual - ELGA LabWater

Air Break Device

SYSTEM CONFIGURATION - PURELAB CHORUS 1 ANR/LSC/GSC

Example 2: 1 X PURELAB Chorus 3 RO or 1 X PURELAB Chorus 2 RO/DI - 1 X PURELAB Chorus 1 - 2 X PURELAB Dispensers LA824 -

LA757 15L Reservoir or LA758 30L Reservoir - Halo Dispenser Wall Mounting Bracket LA768 - Chorus Reservoir Bracket LA770

1 X PURELAB Halo Dispenser LA754 or LA755

Example 3: 1 X PURELAB Chorus 3 RO or 1 X PURELAB Chorus 2 RO/DI - 1 X PURELAB Chorus 1 - 1 X PURELAB Chorus Halo Dispenser LA754 or

LA755 - 1 X PURELAB Dispenser LA824 - LA757 15L Reservoir or LA758 30L Reservoir

® ® ® ®

Air Break Device

Air Break Device

Air Break Device

7 J2

5

1Halo 5

Halo 9 Halo 7

Halo 6 Halo 1 Halo 8

5 5

3

5

347

J1

J2

J11

6

6

48

9

71

4a

53

34

89

5

6

1

'See Chorus 1 system configuration notes'

'See Chorus 1 system configuration notes'

4a

Dispenser Type 1 Dispenser Type 1

Dispenser Type 1 Dispenser Type 1

Chorus 3 RO or 2 RO/DI

Chorus 3 RO or 2 RO/DI*

Type 3 or 2

Chorus 1 Reservoir 15L or 30L

Reservoir 15L or 30L Reservoir Overflow

Chorus 1

Type 3 or 2

Type 1 1

6

47

® ®® ® ®

Adding An

Additional Dispenser

Max Distance 5000mm

Max Distance 5000mm Max Distance 5000mm

Reservoir Overflow

Chorus 3 RO or 2 RO/DI

Halo 1Halo 9 Halo 8 Halo 5

Halo 7

Potable Mains Feedwater Connection

®

Potable Mains

Feedwater Connection

®

Basic or AdvancedHalo Dispenser

Basic or AdvancedHalo Dispenser

Type 1

PURELAB® Dispenser Operator Manual MANU41319 Version D 08/20 Page 4

Page 10: Operator Manual - ELGA LabWater

SYSTEM CONFIGURATION - PURELAB CHORUS 1 ANR/LSC/GSC

Example 4: 1 X PURELAB Chorus 3 RO or 1 X PURELAB Chorus 2 RO/DI - 1 X PURELAB Chorus 1 - 2 X PURELAB Dispensers LA824 -

LA757 15L Reservoir or LA758 30L Reservoir - Chorus Wall Mounting Bracket LA795 - Chorus Reservoir Bracket LA770

Example 5: 1 X PURELAB Chorus 3 RO or 1 X PURELAB Chorus 2 RO/DI - 1 X PURELAB Chorus 1 - 1 X PURELAB Chorus Halo Dispenser LA754 or

LA755 or LA756 - 1 X PURELAB Dispenser LA824 - LA757 15L Reservoir or LA758 30L Reservoir

® ® ® ®

Air Break Device

Air Break Device

Air Break Device

Air Break Device

7 J2

5

1Halo 9Halo 8 Halo 1

5

3

5

347

J1

6

3

5489

1

67

4aJ2

1

345

9

68

1

'See Chorus 1 system configuration notes'

'See Chorus 1 system configuration notes'

4a

Dispenser Type 1

Flexible Halo Dispenser Type 1

Dispenser Type 1 Dispenser Type 1

Chorus 3 RO or 2 RO/DI

Chorus 3 RO or 2 RO/DI*

Type 3 or 2

Chorus 1 Reservoir 15L or 30L

Reservoir Overflow

Chorus 1 Chorus 1

6

47

® ® ® ®

Max Distance 5000mm

Max Distance 5000mm

Reservoir Overflow

Chorus 3 RO or 2 RO/DI

Halo 5Halo 7

Potable Mains Feedwater Connection

Potable Mains

Feedwater Connection

Reservoir 15L or 30L

Type 3 or 2

J1

Halo 6

®

PURELAB® Dispenser Operator Manual MANU41319 Version D 08/20 Page 5

Page 11: Operator Manual - ELGA LabWater

SYSTEM CONFIGURATION - PURELAB CHORUS 1 COMPLETE AND 2+ NON EDI

Chorus Port 1: Feedwater Inlet

Chorus Port 3: Drain

Chorus Port 4: Product Outlet

Chorus Port 5: Return

Chorus Port 6: Power

Chorus Port 7: Comms

Chorus Reservoir Port 1: Inlet

Chorus Reservoir Port 4a: Outlet

Chorus Reservoir Port 4b: Outlet

Chorus Reservoir J1: Comms

Chorus Reservoir J2: Comms

Dispenser Port 1: Inlet

Dispenser Port 5: Return

Halo Dispenser Port 1: Inlet

Halo Dispenser Port 5: Return

Halo Dispenser Port 6: Power

Halo Dispenser Port 7: Comms

Halo Dispenser Port 8: Power

Halo Dispenser Port 9: Comms

Example 1: 1 X PURELAB Chorus 1 Complete or PURELAB Chorus 2+ (Non EDI) - 3 X PURELAB Dispensers LA824 (Max 3 Dispensers) -

LA759 60L Reservoir or LA760 100L Reservoir

® ®

Chorus 1 Complete and 2+ recirculate purified water throught the reservoir for repeated exposure to the purifying technologies.

New system configuration:

1. Remove and relocate Non-Return-Valve before starting the commissiong procedure.

(Please refer to operator manual MANU40932 section 7.2 for details)

2. Commission PURELAB Chorus 1 Complete or PURELAB Chorus 2+ system with the dispensers connected, redirecting reservoir port: 4b to drain

during this procedure (Please refer to operator manual MANU40932 section 5.5 for commissioning instructions).

Existing system configuration: If you are connecting a PURELAB Dispenser to an existing system configuration, please ensure all of your products

are up to date and using the latest software. To find the software version of your existing system, simply turn the power off by the wall and back on

again and the current software version number will be displayed in the bottom right-hand corner of the screen. If your software version number is

older than the one listed on the website at www.elgalabwater.com/customize, then please perform a software update,

see section 8.3 'Software Update' in this operator manual.

® ®

COMMSKey

Power

Water

Valve

Non-Return-Valve

15mm Valve

8mm Flow Bend

8mm Tube Clip

Regulator LA512

®

Max Distance 5000mm Max Distance 5000mm

NRV

Air Break Device

Outlet to Application

Reservoir Overflow

11 55 567

4a

J2

1

4b

3

45

Max Distance 5000mm

Adding An

Additional Dispenser

Chorus 1 Complete or 2+ (Non EDI)

Chorus Reservoir 60L or 100L

1

1

Dispenser Type 1 or Type 2+ Dispenser Type 1 or Type 2+ Dispenser Type 1 or Type 2+

Air Break Device

Potable Mains

Feedwater Connection

J1

®

PURELAB® Dispenser Operator Manual MANU41319 Version D 08/20 Page 6

Page 12: Operator Manual - ELGA LabWater

SYSTEM CONFIGURATION - PURELAB CHORUS 1 COMPLETE AND 2+ NON EDI

Example 2: 1 X PURELAB Chorus 1 Complete or PURELAB Chorus 2+ (Non EDI) - 1 X PURELAB Dispenser LA824 -

LA757 15L Reservoir or LA758 30L Reservoir - 2 X PURELAB Chorus Halo Dispenser LA754 or LA755 or LA756

® ® ®

'See Chorus 1 Complete and 2+ (Non EDI) system configuration notes'

Potable Mains FeedwaterConnection

Reservoir Overflow

Air Break Device

1

3

4b

4a

46

5

1

7

Chorus Reservoir 15L or 30L

1

5

Max Distance 5000mm Max Distance 5000mm Max Distance 5000mm

Dispenser Type 1 or Type 2+Flexible Halo Dispenser Type 1 or Type 2+

Outlet to Application

NRVJ1J2

Halo 1 Halo 1

Halo 8

Halo 5Halo 6

Halo 5Halo 6

Halo 7 Halo 7 Halo 9 Halo 9

®

Basic or Advanced Halo Dispenser Type 1 or Type 2+

Chorus 1 Complete or 2+ (Non EDI)

PURELAB® Dispenser Operator Manual MANU41319 Version D 08/20 Page 7

Page 13: Operator Manual - ELGA LabWater

SYSTEM CONFIGURATION - PURELAB CHORUS 2+ EDI

Chorus Port 1: Feedwater Inlet

Chorus Port 2: EDI Drain Chorus Port 3: Drain

Chorus Port 4: Product Outlet

Chorus Port 5: Return

Chorus Port 6: Power

Chorus Port 7: Comms

Chorus Reservoir Port 1: Inlet

Chorus Reservoir Port 4a: Outlet

Chorus Reservoir Port 4b: Outlet

Chorus Reservoir J1: Comms

Chorus Reservoir J2: Comms

Dispenser Port 1: Inlet

Dispenser Port 5: Return

Halo Dispenser Port 1: Inlet

Halo Dispenser Port 5: Return

Halo Dispenser Port 6: Power

Halo Dispenser Port 7: Comms

Halo Dispenser Port 8: Power

Halo Dispenser Port 9: Comms

Example 1: 1 X PURELAB Chorus 2+ (EDI) - 3 X PURELAB Dispensers (Max 3 Dispensers) - LA759 60L Reservoir or LA760 100L Reservoir ®

®

Chorus 2+ EDI recirculate purified water throught the reservoir for repeated exposure to the purifying technologies.

New System configuration:

1. Remove and relocate Non-Return-Valve before starting the commissiong procedure.

(Please refer to operator manual MANU40932 section 7.2 for details)

2. Commission PURELAB Chorus 2+ EDI with the PURELAB Dispensers connected, water will be automatically redirected to drain through an internal

flush valve (V2) during this procedure. (Please refer to operator manual MANU40932 section 5.6 for commissioning instructions)

3. Once commissioning is completed the internal EDI filter module will need 'bleeding' before purified water can be dispensed from the Chorus 2+ tap

or dispensers. (Please refer to operator manual MANU40932 section 5.6 for instructions)

Existing system configuration: If you are connecting a PURELAB Dispenser to an existing system configuration, please ensure all of your products

are up to date and using the latest software. To find the software version of your existing system, simply turn the power off by the wall and back on

again and the current software version number will be displayed in the bottom right-hand corner of the screen. If your software version number is

older than the one listed on the website at www.elgalabwater.com/customize, then please perform a software update,

see section 8.3 'Software Update' in this operator manual.

® ®

Max Distance 5000mm Max Distance 5000mm

NRV

Air Break Device

Outlet to Application

Reservoir Overflow

11 55 567

4a

J2

1

4b

3

45

Max Distance 5000mm

Adding An

Additional Dispenser

Chorus 2+ EDI

Chorus Reservoir 60L or 100L

1

1

Dispenser Type 2+ Dispenser Type 2+ Dispenser Type 2+

Air Break Device

Potable Mains

Feedwater Connection

J1

COMMSKey

Power

Water

Valve

Non-Return-Valve

15mm Valve

8mm Flow Bend

8mm Tube Clip

Regulator LA512

®

2

PURELAB® Dispenser Operator Manual MANU41319 Version D 08/20 Page 8

Page 14: Operator Manual - ELGA LabWater

SYSTEM CONFIGURATION - PURELAB CHORUS 2+ EDI

Example 2: 1 X PURELAB Chorus 2+ (EDI) - 1 X PURELAB Dispenser LA824 - LA757 15L Reservoir or LA758 30L Reservoir -

2 X PURELAB Chorus Halo Dispenser LA754 or LA755 or LA756

®

®

®

'See Chorus 2+ (EDI) system configuration notes'

Potable Mains FeedwaterConnection

Reservoir Overflow

Air Break Device

1

32

4b

4a

465

1

7

Chorus 2+ EDI

Chorus Reservoir 15L or 30L

1

5

Max Distance 5000mm Max Distance 5000mm Max Distance 5000mm

Dispenser Type 2+Basic or Advanced Halo Dispenser Type 2+Flexible Halo Dispenser Type 2+

Outlet to Application

NRVJ1J2

Halo 1 Halo 1

Halo 8

Halo 5Halo 6

Halo 5Halo 6

Halo 7 Halo 7 Halo 9 Halo 9

PURELAB® Dispenser Operator Manual MANU41319 Version D 08/20 Page 9

Page 15: Operator Manual - ELGA LabWater

SET-UP ENGLISH

HOW TO ASSEMBLE AND MAKE HEIGHT ADJUSTMENTS 1

(Fig.1) (Fig.2)

a) Insert the handset set into the cradle as shown. (Fig.1)

c) Press and hold the height adjustment button to slide

the handset arm up and down. (Fig.3)

b) Rotate the handset. (Fig.2)

d) When you have found the perfect height, release the hight adjustment

button to lock into position. (Fig.4)

(Fig.3) (Fig.4)

MAX 180 ˚

PURELAB® Dispenser Operator Manual MANU41319 Version D 08/20 Page 10

Page 16: Operator Manual - ELGA LabWater

SET - UP

REMOVING THE SUPPLIED WORKTOP MOUNTING BRACKET

OPTIONAL WORKTOP MOUNTING BRACKET2

3

(Fig.5)

(Fig.6)

a) Undo the fixing screws screw from the base of the PURELAB Dispenser to remove the mounting bracket.

(Fig.6)

This is an optional attachment, please skip to section 7 in the Quick Start Guide if you do not need to fit this sercuring bracket.

Items not supplied

EN388 & EN374Phillips Screw DriverØ3mm Drill BitDrill Goggles

Caution! Please read the Health and Safety Notices in the

front of the Quick Start Guide before proceeding

with this installation.

Caution! This bracket is only suitable for wooden worktops.

PURELAB® Dispenser Operator Manual MANU41319 Version D 08/20 Page 11

Page 17: Operator Manual - ELGA LabWater

SET - UP

MARK AND DRILLING THE HOLES REASSEMBLE THE BRACKET

FIXING THE PURELAB DISPENSER TO THE WORKTOP

4 5

6

a) Mark and drill two holes into the wooden worktop to a depth of

30mm using an Ø3mm pilot drill bit. (Fig.7)

Ø3mm

a) Reassemble the mounting bracket back onto the base

as shown above. (Fig.8)

a) Insert the fixing screws into the screw cap washers. (Fig.9)

Screw cap

Fixing Screw

Screw cap washer

b) Fix the screws into the pre drilled holes and push the screw caps over the screw heads. (Fig.9) Bracket installations completed

(Fig.7)

(Fig.8)

(Fig.9)

PURELAB® Dispenser Operator Manual MANU41319 Version D 08/20 Page 12

Page 18: Operator Manual - ELGA LabWater

SET - UP

CONNECT THE WATER INLET

CONNECT THE WATER RETURN

7

8

(Fig.12) (Fig.13)

a) Remove the transit plug from Port 5 water Return connection.

(Fig.12)

b) Using the tube supplied, firmly push one end securely into the

Port 5 water Return connection. Use the flow bends provided to

mould the tube into position. (Fig.13)

b) Using the tube supplied, firmly push one end securely into

Port 1 inlet connection. Use the flow bends provided to mould

the tube into position. (Fig.11)

Push in collet

Remove Transit Plug

a) Remove the transit plug from Port 1 water inlet connection.

(Fig.10)

Push in collet

Remove Transit Plug

(Fig.10) (Fig.11)

Flow Bends

Transit Plug

PURELAB® Dispenser Operator Manual MANU41319 Version D 08/20 Page 13

Page 19: Operator Manual - ELGA LabWater

Comms cable from main

product or reservoir port

Comms cable from main

product or reservoir port

Power supply

(Power supply is supplied with main product)

SET - UP

CONNECT THE COMMUNICATIONS CABLES

CONNECT THE POWER CABLES

9

10

a) Link the comms cables as shown above,

connections may vary depending on your

system configuration. (Fig.14)

a) Link the power cables as shown above,

connections may vary depending on your

system configuration. (Fig.15)

Caution!

Only use the power supply provided by ELGA VEOLIA.

3

2

1

(Fig.14)

(Fig.15)

PURELAB® Dispenser Operator Manual MANU41319 Version D 08/20 Page 14

Page 20: Operator Manual - ELGA LabWater

SET - UP

11

12

SET LANGUAGE

FOLLOW THE ON SCREEN INSTRUCTIONS

a) Press up and down buttons to select your language. (Fig.16)

(Language section will only appear on screen if you are setting up a new system configuration.)

(The Commissioning procedure will only appear on screen if you are setting up a new system configuration. When connecting to an existing system configuration, wait for the purity to be displayed before dispensing)

a) For new system system configurations,

follow the on screen commissioning

instructions. (Fig.18)

b) After completing the commissioning

procedure, please wait for the

reservoir to fill and reach full purity

before dispensing any water. (Fig.19)

(Water Purity Alarm will be active

during the commissioning)

C) When full purity is reached reached on your

Type 1 or Type 2 system configuration

you can dispense purified water from the

PURELAB Dispenser. (Fig.20)

Type 1 or Type 2 handset label.

b) Press accept to confirm. (Fig.17)

(Fig.16)

(Fig.18)

(Fig.19) (Fig.20)

(Fig.17)

PURELAB® Dispenser Operator Manual MANU41319 Version D 08/20 Page 15

Page 21: Operator Manual - ELGA LabWater

PURELAB® Dispenser Operator Manual MANU41319 Version D 08/20 Page 16

MANUAL INTRODUCTION

1.1 Health & Safety

Please ensure you read the Health & Safety notes in the Quick Start Guide.

1.2 Product Model

This Operator manual has been prepared for PURELAB® Dispenser product model:

• PURELAB® Dispenser LA824 1.3 Use of this Manual

This manual guides you through the operation and maintenance of PURELAB® Dispenser allowing you to obtain a guaranteed supply of purified water to meet your requirements. 1.4 Installation

Quick Start Guide shown in the front of this manual shows you how to install PURELAB® Dispenser.

Quick Reference Guide shows you 'day to day' functions while using the PURELAB® Dispenser.

1.5 Environment

PURELAB® Dispenser should be installed on a flat, level surface, in a clean, dry environment it can also be mounted to a worktop using the purpose designed bracket, See Quick Start Guide section 2 for details 1.6 Commissioning

Please refer to the main product operator manuals for commissioning a new system configuration.

1.7 Customer Support

If you need help with your product, please call your local ELGA Veolia® representative. For the address of the nearest ELGA Veolia® Sales and Service office visit our website.

www.elgalabwater.com or contact ELGA Veolia at: Email: [email protected]

Existing system configuration: If you are connecting a PURELAB® Dispenser to an existing system configuration, please ensure all of

your products are up to date and using the latest software. To find the software version of your existing system, simply turn the power off

by the wall and back on again and the current software version number will be displayed in the bottom right-hand corner of the screen.

If your software version number is older than the one listed on the website at www.elgalabwater.com/customize, then please perform a

software update, see section 8.3 'Software Update' in this operator manual.

1. MANUAL INTRODUCTION

Page 22: Operator Manual - ELGA LabWater

PURELAB® Dispenser Operator Manual MANU41319 Version D 08/20 Page 17

YOUR GUIDE TO THE CHORUS 2. GUIDE TO YOUR PURELAB DISPENSER

Dispenser Handset

LCD Display

Screen FlexibleDispenser Tube

Auto Volume Button

Dispense Button

Water Purity Display

Scrolling information bar -bar indicating the status.

Auto Volume or Reservoir Volume

Up Button

Accept Button

Down Button

Dispenser Tip

Worktop Mounting

Bracket

Handset Cradle

Height Adjustment

Button

Weighted Base

COMMS Port

COMMS Port

Port 5: ReturnPower

24V DCPower

24V DC

Port 1:

Inlet

Page 23: Operator Manual - ELGA LabWater

Main Display

Water Purity Display Megohm - 1 to 18.2 MΩ

(When the water purity falls below the default setting, numbers will flash in red indicating that the purity alarm has been activated)

Water Purity Display Micro Siemens

(This feature can be activated in the main menu options tabs)

Scrolling Information Bar Alarms and product information are displayed in this scroll bar.

Reservoir Volume Reservoir levels and fill status are shown on the display. (Reservoir symbol flashes red indicating that the reservoir is almost empty)

Auto Volume Dispense This feature is activated by pressing the Auto Volume Dispense button, volumes are displayed in incremental order and changed when the up and down buttons are pressed. Pressing accept button start the dispense.

Process symbol Shows that the product is perfoming an operation.

Button Operation

Scroll Up Button Press to scroll up on a menu or volume increments increase in a feature.

Scroll Down Button Press to scroll down on a menu or volume increments decrease in a feature.

(Press and hold to exit data logging)

Dispense Button Press for manual dispense

Auto Volume Dispense Button Press once to activate and again to deactivate the Auto Volume Dispense feature.

On some products pressing and holding the Auto Volume Dispense button will initiate a reservoir refill cycle.

Accept Button Press the accept button to enter.

Press and hold the accept button to enter the main menu options tabs.

PURELAB® Dispenser Operator Manual MANU41319 Version D 08/20 Page 18

YOUR GUIDE TO THE CHORUS 3. MAIN DISPLAY SETTINGS AND BUTTON OPERATION

,

100ml

7L

*Volumes depend on your system configuration.

Main Menu options tabsScreensaver /Sleep mode *Options tabs will be driven from the main product. Handset ButtonsHome Screen

Scroll Bar info

Reservoir Level

Water Purity Display

Page 24: Operator Manual - ELGA LabWater

* Consumable life is an estimate only, and will depend on the application.

4. PRODUCT CONSUMABLES

POU Filters LC134LC145

POU FilterLC197

(See operator manual maintenance section 7 'Replacing the POU filter LC134 / LC145 / LC197 for details)

5.1 Registering a Product

Taking the time now to register will mean that we can provide a better service to you in the future. We can contact you about product information and service updates.

Why Register your product?

• Validated your product warranty

• Proof of product registration

• Receive software and service updates How can I quickly register?

The model number and serial number can be found on the ratings plate, see below:

* Model number

** Serial number

Enter the model and serial number into the online form at: https://www.elgalabwater.com/support/register-a-product

ELGA Veolia is the global laboratory water brand of

Tel: +44 (0) 203 567 7300, Web: www.elgalabwater.com

Maximum PressureInlet (bar/psi)Working (bar/psi)

REFER TO OPERATING INSTRUCTIONSSU REFERER AU MANUEL D'OPERATIONIN DER BEDIENUNGSANLEITUNG NACHSEHEN

Serial No.

MADE IN THE UK

MODEL *

**

5. REGISTER YOUR PRODUCT

Part No. Description Typical Service Life* Max. Shelf Life

LC134 Point-Of-Use 0.2µm Micro Filter (POU) 3 Months 2 Years

LC145 Point-Of-Use 0.2µm Micro Filter (POU) 3 Months 2 Years

LC197 Point-Of-Use Bio Filter (POU) 3 Months 2 Years

PURELAB® Dispenser Operator Manual MANU41319 Version D 08/20 Page 19

Page 25: Operator Manual - ELGA LabWater

6.1 Manual Dispense 6.2 Continuous Dispense

6.3 Auto Volume Dispense

6. QUICK REFERENCE GUIDE

Press and hold the dispense button at anytime for manual dispense, release to stop dispensing.

Press the dispense button and accept button together to start continuous dispense.Press dispense button to exit this feature.

Press Auto Volume dispensebutton at anytime to open this feature.

Press up and down buttons, to scroll throught the volumes.

Note:

Water volumes

may vary see the

technical

specifications in your

main product.

To confirm press dispense button to start dispensing the water.

PURELAB® Dispenser Operator Manual MANU41319 Version D 08/20 Page 20

7.0L

100ml

Page 26: Operator Manual - ELGA LabWater

6.4 Deactivating Auto Volume Dispense

6.5 ON - OFF, Menu and Accept

YOUR GUIDE TO THE CHORUS 6. QUICK REFERENCE GUIDE

Press Auto Volume Dispense button once to deactivate.

Press and hold accept button for 2 seconds to enter the Main Menu.

Press up and down buttons to scoll throught the main menu options tabs, press accept button to confirm your selection.

0202

7.0L

100ml

PURELAB® Dispenser Operator Manual MANU41319 Version D 08/20 Page 21

Note: Menu options tabs

are driven from your main

product.

Page 27: Operator Manual - ELGA LabWater

7.1 Replacing the Point-Of-Use Filter LC134 / LC145 / LC197

Step 1 - Removing the Dispense Tip

1. Unscrew the dispense tip from underneath the handset (Fig.1).2. Do not dispose of the dispense tip, this part is required for the sanitization procedure (Fig.2).

Step 2 - Unpacking the LC134 / LC145 / LC197 Point-Of-Use Filter and Installing

1. Unpack the new POU filter from it's outer and inner packaging, remove the cap (Fig.3).2. Coupler is not required for PURELAB® Dispenser, screw in the Point-of-Use filter (Fig.4).

3. Slightly open both bleed valves, before dispensing any water to release any trapped air (Fig.5).4. Press the dispense button, drawing water through and close both bleed valves (Fig.6).5. Installation is completed, normal operation can resume.

YOUR GUIDE TO THE CHORUS 7. MAINTENANCE

Dispense Tip

Only use 1 X o-ring

Open Close

Completed

(Fig.1)

(Fig.5) (Fig.6)

(Fig.2)

Coupler (not required)

Point-Of-Use LC134 / LC145 / LC197

Bleed Valve

M8mm Thread

Bleed Valve

Cap

(Fig.3)

PURELAB® Dispenser Operator Manual MANU41319 Version D 08/20 Page 22

(Fig.4)

Page 28: Operator Manual - ELGA LabWater

8. CUSTOMIZE OPERATION

PURELAB® Dispenser Operator Manual MANU41319 Version D 08/20 Page 23

8.1 Set Alarm Points

1. Hold and press accept to enter the main menu (Fig.1).2. Scroll + and select Set Alarm Points, press accept to confirm (Fig.2). 3. Scroll + and select a feature, press accept to confirm (Fig.3). 4. Adjust the valve by scrolling + , press accept to save (Fig.4). 5. Normal operation will now resume. (Fig.5).

8.2 Advanced Data Logging

1. Insert a USB stick into the USB port on the main product (Fig.1). 2. Scroll + and select 'Start Data Logging', press accept to confirm (Fig.2).3. Scroll + and select an interval data recording time, press accept to confirm (Fig.3).4. To finish recording press the down and remove the USB. (Fig.4). 5. Downloaded Data files can be viewed on your PC or Mac.

8.3 Software Update

1. Download the latest software version onto a blank USB at: www.elgalabwater.com/customize2. Insert the USB, scroll + and select 'Software Update', press accept to confirm (Fig.2).3. Progress will be displayed on screen (Fig.3).4. Once completed the system configuration will power down, hold and press accept to power ON (Fig.4). 5. Software update is now complete, normal operation will resume (Fig.5).

04:58USB Port on the

main product

USB Port on the

main productProduct software version number location (example displayed)

(Fig.1) (Fig.2) (Fig.3) (Fig.4) (Fig.5)

Page 29: Operator Manual - ELGA LabWater

9. TROUBLESHOOTING

PURELAB® Dispenser Operator Manual MANU41319 Version D 08/20 Page 24

Issues Action

Blank Display Screen

1. Press any button to wake from sleep mode.

or

2. Check the power supply is connected to the product, ensuring that all connections are securly fitted. (Turn the power OFF/ON)

or

3. Check the plug socket fuse, replace if nessary.

Unable to Enter the Features in the Main Menu

1. Check the product is turned ON.

or

2. Make sure Auto volume dispense is disabled. (See section 6.3 - Quick Reference Guide)

Incorrect Volume of Water 1. Select Auto Volume Calibration in the main menu.

Message Displayed - Sanitization Reminder

1. Complete a Sanitization Procedure.

(See main product operator manual for instructions)

Message Displayed -

UV Lamp Change Reminder

* Some main products do not have a UV Lamp so

this message may not appear on your handset.

1. Complete a UV lamp change.

(See main product operator manual for instructions)

2. Reset UV lamp reminder.

(See main product operator manual for instructions)

Message Displayed - High Temperature Water Alarm

1. Adjust the alarm points settings via Main Menu.

or

2. Check the feedwater temperature.

3. Press the dispense button to draw some water through the system configuration, switch OFF and restart.

Reduced Dispensing Flow Rate

1. Check the mesh feedwater inlet filter for blockages. (See main product operator manual for instructions)

or

2. Replace the Point-Of-Use Filter.

or

3. Check the feedwater valves are fully opened.

Page 30: Operator Manual - ELGA LabWater

Issues Action

Message Displayed - Dispense Water Purity Alarm

1. Check the alarm points settings via Main Menu.

or

2. Replace the Purification Pack and wait for the cycle to complete. (See main product operator manual for instructions)

or

3. Replace the UV lamp.

* Some main products do not have a UV Lamp

(See main product operator manual for instructions)

(Water Purity Alarm is active during commissioning procudure,

see quick start guide for details)

Message Displayed -

System configuration error

1. check connections, ensure that the configuration is valid.

(follow the system configurations - Quick Start Guide)

or

2. Update your main product software via the USB port.

(See Customize Operation - section 9 Customize Operation)

YOUR GUIDE TO THE CHORUS 9. TROUBLESHOOTING

PURELAB® Dispenser Operator Manual MANU41319 Version D 08/20 Page 25

Page 31: Operator Manual - ELGA LabWater

10. TECHNICAL SPECIFICATIONS

PURELAB® Dispenser Operator Manual MANU41319 Version D 08/20 Page 26

PURELAB Dispenser - LA824

Dimensions and Weights

Dispatch Weight 7.7 KG

Operational Weight 4.5 KG

Installation Free-standing on a worktop or secured to a worktop using the mounting bracket provided (see quick start guide for details)

Electrical Requirements

Mains Input 100-240 VAC, 50-60 Hz

Power Required 24 V DC

Power Consumption 120 VA

Noise Output dBA - < 40

Water Connections

Inlet 8mm OD Tube

Return 8mm OD Tube

Flow Bends 8mm (Provided in the basic installation kit with the main product)

580

mm

375mm264mm

Page 32: Operator Manual - ELGA LabWater

YOUR GUIDE TO THE CHORUS 11. WARRANTY / CONDITIONS OF SALE

PURELAB® Dispenser Operator Manual MANU41319 Version D 08/20 Page 27

ELGA VEOLIA® is a trading name of VWS (UK) Ltd.

General Limited Warranty

VWS (UK) Ltd. warrants the products manufactured by it against defects in materials and workmanship when used in accordance with applicable instructions for a period of one year from the date of shipment for the products. VWS (UK) Ltd. MAKES NO OTHER WARRANTY, EXPRESSED OR IMPLIED. THERE IS NO WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.

The warranty provided herein and the data, specifications and descriptions of the VWS (UK) Ltd. products appearing in VWS (UK) Ltd. published catalogues and product literature may not be altered except by express written agreement signed by an officer of VWS (UK) Ltd. Representations, oral or written, which are inconsistent with this warranty or such publications are not authorized and, if given, should not be relied upon.

In the event of a breach of the foregoing warranty, VWS (UK) Ltd. sole obligation shall be to repair or replace, at its option, any product or part thereof that proves to be defective in materials or workmanship within the warranty period, provided the customer notifies VWS (UK) Ltd. promptly of any such defect. The exclusive remedy provided herein shall not be deemed to have failed of its essential purpose so long as VWS (UK) Ltd. is willing and able to repair or replace any nonconforming VWS (UK) Ltd. product or part. VWS (UK) shall not be liable for consequential, incidental, special or any other indirect damages resulting from economic loss or property damage sustained by any customer from the use of its products.

Water System Limited Warranty

VWS (UK) Ltd. warrants the water systems manufactured by it, BUT EXCLUDING MEMBRANES AND PURIFICATION PACKS, against defects in materials and workmanship when used in accordance with the applicable instructions and within the operating conditions specified for the systems for a period of one year from the earlier of:

the date of installation, or the 120th day following the date of shipment.

VWS (UK) LTD. MAKES NO OTHER WARRANTY, EXPRESSED OR IMPLIED. THERE IS NO WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. The warranty provided herein and the data, specifications and descriptions of the VWS (UK) Ltd. systems appearing in VWS (UK) Ltd. published catalogues and product literature may not be altered except by express written agreement signed by an officer VWS (UK) Ltd. Representations, oral or written, which are inconsistent with this warranty or such publications are not authorised and, if given, should not be relied upon.

In the event of a breach of the foregoing warranty, VWS (UK) Ltd. sole obligation shall be to repair or replace, at its option, any product or part thereof that proves to be defective in materials or workmanship within the warranty period, provided the customer notifies VWS (UK) Ltd. promptly of any such defect. The cost of labour for the first ninety (90) days of the above warranty period is included in the warranty; thereafter, labour cost shall be at the customer’s expense. The exclusive remedy provided herein shall not be deemed to have failed of its essential purpose so long as VWS (UK) Ltd. is willing and able to repair or replace any nonconforming VWS (UK) Ltd. system or component part. VWS (UK) Ltd. Ltd. shall not be liable for consequential, incidental, special or any other indirect damages resulting from economic loss or property damage sustained by any customer from the use of its process systems.

Products or components manufactured by companies other than VWS (UK) Ltd. or its affiliates

(“Non - VWS (UK) Ltd. products”) are covered by the warranty, if any, extended by the Product manufacturer. VWS (UK) Ltd. hereby assigns to the purchaser any such warranty; however VWS (UK) LTD. EXPRESSLY DISCLAIMS ANY WARRANTY WHETHER EXPRESSED OR IMPLIED, THAT THE NON - VWS (UK) Ltd. PRODUCTS ARE MERCHANTABLE OR FIT FOR A PARTICULAR PURPOSE.

NOTICE

VWS (UK) Ltd. is constantly striving to improve its products and services. Consequently, the information in this document is subject to change without notice and should not be construed as a commitment by VWS (UK) Ltd.

Also VWS (UK) Ltd. assumes no responsibility for any errors that may appear in this document. This manual is believed to be complete and accurate at the time of publication. In no event shall VWS (UK) Ltd. be liable for incidental or consequential damages in connection with or arising from the use of this manual.

VWS (UK) Ltd. warrants its products against defects in materials and workmanship as described in the Warranty statement on the preceding pages.

Page 33: Operator Manual - ELGA LabWater
Page 34: Operator Manual - ELGA LabWater

ELGA VEOLIA is the global laboratory water brand name of Veolia. The information contained in this document is the

property of VWS (UK) Ltd, trading as ELGA Veolia, and is supplied without liability for errors or omissions. © VWS (UK) Ltd.

2020 - All right reserved. ELGA® , PURELAB® are registered trademarks of VWS (UK) Ltd.

Tel: +44 203 567 7300 Email: [email protected] Web: www.elgalabwater.com

The Labwater Specialists