Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst...

64
www.EmersonProcess.com Instruction Manual Thermo Flame Ionization Detector FID Analyzer or Analyzer Module (combined with NGA 2000 Platform, MLT or CAT 200) 2 nd Edition 04/2008 Instruction Manual 90003806 04/2008

Transcript of Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst...

Page 1: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

www.EmersonProcess.com

Instruction ManualThermo Flame Ionization DetectorFID Analyzer or Analyzer Module (combined with NGA2000 Platform, MLT or CAT 200)

2nd Edition 04/2008

Instruction Manual9000380604/2008

Page 2: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

TFID Thermo Flame Ionization Detector Instruction Manual90003806

04/2008

Emerson Process ManagementGmbH & Co. OHGIndustriestrasse 1D-63594 HasselrothGermanyT +49 (0) 6055 884-0F +49 (0) 6055 884-209Internet: www.EmersonProcess.com

ESSENTIAL INSTRUCTIONSREAD THIS PAGE BEFORE PROCEEDING!

Emerson Process Management (Rosemount Analytical) designs, manufactures and testsits products to meet many national and international standards. Because these instrumentsare sophisticated technical products, you MUST properly install, use, and maintainthem to ensure they continue to operate within their normal specifications. The followinginstructions MUST be adhered to and integrated into your safety program when installing,using and maintaining Emerson Process Management (Rosemount Analytical) products.Failure to follow the proper instructions may cause any one of the following situations tooccur: Loss of life; personal injury; property damage; damage to this instrument; and warrantyinvalidation.

• Read all instructions prior to installing, operating, and servicing the product.

• If you do not understand any of the instructions, contact your Emerson ProcessManagement (Rosemount Analytical) representative for clarification.

• Follow all warnings, cautions, and instructions marked on and supplied with the product.

• Inform and educate your personnel in the proper installation, operation, andmaintenance of the product.

• Install your equipment as specified in the Installation Instructions of the appropriateInstruction Manual and per applicable local and national codes. Connect all productsto the proper electrical and pressure sources.

• To ensure proper performance, use qualified personnel to install, operate, update, program,and maintain the product.

• When replacement parts are required, ensure that qualified people use replacement partsspecified by Emerson Process Management (Rosemount Analytical). Unauthorized partsand procedures can affect the product’s performance, place the safe operation of yourprocess at risk, and VOID YOUR WARRANTY. Look-alike substitutions may result in fire,electrical hazards, or improper operation.

• Ensure that all equipment doors are closed and protective covers are in place, exceptwhen maintenance is being performed by qualified persons, to prevent electricalshock and personal injury.

The information contained in this document is subject to change without notice. Misprintsreserved.

1st Edition 06/1998 2nd Edition 04/2008 © 2008 by Emerson Process Management

Page 3: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

CONTENTS

I90003806(1) NGA-TFID e 16.06.98

Rosemount Analytical

Table of Contents

SAFETY SUMMARY S - 1

General S - 2

Gases and Gas Conditioning (Sample Handling) S - 3

Supply Voltage S - 4

Analyzer specific notes for the user S - 5

Additional notes for service / maintenance S - 6

Electrostatic Discharge S - 7

PREFACE P - 1

DESCRIPTION

1. Technical Description 1 - 1

1.1 Front View 1 - 1

1.2 Rear Panel 1 - 3

1.3 Internal Construction 1 - 4

1.3.1 Thermo-FID Assembly 1 - 5

1.3.2 Printed Circuit Boards 1 - 6

1.4 Applications 1 - 8

2. Theory of Technology and Operation 2 - 1

2.1 Measuring Principle 2 - 1

2.2 TFID Mode of Operation 2 - 3

2.3 Sample Gas Transport 2 - 4

3. (vacant)

4. (vacant)

Page 4: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

CONTENTS

II 90003806(1) NGA-TFID e 16.06.98

Rosemount Analytical

OPERATION

5. Preparation 5 - 1

5.1 Installation Site 5 - 2

5.2 Gas Conditioning (Sample Handling) 5 - 3

5.3 Gases 5 - 4

5.3.1 Supply Gases 5 - 4

5.3.2 Gas Connections 5 - 5

6. Switching On 6 - 1

7. Switching Off 7 - 1

8.

- (vacant)

14.

MAINTENANCE

15. Opening the Housing 15 - 1

15.1 Housing Cover 15 - 1

15.2 Front Panel 15 - 1

16.

- (vacant)

18.

19. Cleaning of Housing Outside 19 - 1

Page 5: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

CONTENTS

III90003806(1) NGA-TFID e 16.06.98

Rosemount Analytical

TECHNICAL DATA 20 - 1

20.1 Housing 20 - 1

20.2 Ambient Conditions 20 - 2

20.3 General Specifications 20 - 3

20.4 Voltage supply 20 - 3

20.4.1 Electrical Safety 20 - 4

20.5 Response Factors 20 - 4

SUPPLEMEMT

21. Pin Assignments 21 - 1

21.1 230 or 115 Vac (Input and Output) 21 - 1

21.2 Pt 100 (external heater) 21 - 1

21.3 Option SIO (Standard Input/Output Module) 21 - 2

21.3.1 Analog Signal Outputs 21 - 2

21.3.2 Relay Outputs / Serial Interfaces 21 - 2

21.4 Option DIO (Digital Input/Output Module) 21 - 3

21.4.1 Wiring with external actors 21 - 4

21.5 COM - Ports (for service use only) 21 - 6

INDEX I - 1

List of Figures I - 4

Page 6: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

CONTENTS

IV 90003806(1) NGA-TFID e 16.06.98

Rosemount Analytical

Page 7: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

SAFETY SUMMARY

S - 190003806(1) NGA-TFID e 16.06.98

Rosemount Analytical

Safety Summary

Outside and/or inside TFID or at operation manual resp. different symbols gives you a hint to

special sources of danger.

Source of danger !See Operation Manual!

High Voltage !

Electrostatic Discharge (ESD) !

Explosives !

Hot components !

Toxic !

Risk to health !

Analyzer specific notes for the user !

In operation manual we will give partly additional informations to these symbols.

Strictly follow these instructions please !

GENERAL

Page 8: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

SAFETY SUMMARY

S - 2 90003806(1) NGA-TFID e 16.06.98

Rosemount AnalyticalGENERAL

1. General

The following general safety precautions must be observed during all phases of operation,

service and repair of this instrument !

Failure to comply with these precautions or with specific warnings elsewhere in this manual

violates safety standards of design, manufacture and intended use of this instrument !

Failure to comply with these precautions may lead to personal injury and damage to this

instrument !

Fisher-Rosemount GmbH & Co. does not take responsibility (liability) for the customer´s

failure to comply with these requirements !

Do not attempt internal service or adjustment unless other person, capable of rendering first

aid and resuscitation, is present !

Because of the danger of introducing additional hazards, do not perform any unauthorized

modification to the instrument !

Return the instrument to a Fisher-Rosemount Sales and Service office for service or repair

to ensure that safety features are maintained !

Instruments which appear damaged or defective should be made inoperative and secured

against unintended operation until they can be repaired by qualified service personnel.

Page 9: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

SAFETY SUMMARY

S - 390003806(1) NGA-TFID e 16.06.98

Rosemount AnalyticalGENERAL / GASES AND GAS CONDITIONING (SAMPLE HANDLING)

Operating personnel must not remove instrument covers !Component replacement and internal adjustments must be made by qualifiedservice personnel only !

Read all operation manuals before attempting to operate with the instrument !Be sure to observe the additional notes, safety precautions and warnings givenin the individual operation manuals (e. g. platform, analyzer modules andI/O-modules) !

Do not operate the instrument in the presence of flammable gases or explosiveatmosphere without supplementary protective measures !

At heated components there could be exist hot components !

If any moisture or objects come into the internal of the instrument disconnectpower and all supply gases.

2. Gases and Gas Conditioning (Sample Handling)

Be sure to observe the safety regulations for the respective gases(sample gas and test gases / span gases) and the gas bottles !

All pneumatic supplies and mains, especially the hydrogen mains, have to betight ! All gases used must have the right input pressure !

Inflammable or explosive gas mixtures must not be purged into the instrumentwithout supplementary protective measures !

To avoid a danger to the operators by explosive, toxic or unhealthy gascomponents, first purge the gas lines with ambient air or nitrogen (N

2) before

cleaning or exchange parts of the gas paths.

Page 10: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

SAFETY SUMMARY

S - 4 90003806(1) NGA-TFID e 16.06.98

Rosemount AnalyticalSUPPLY VOLTAGE

3. Supply Voltage

The socket outlet shall be installed near the equipment and shall be easilyaccessible to disconnect the device from the socket outlet.

Verify before operation whether the line voltage stated on the label of theinstrument ore power supply agrees with that of your mains line!

Do not put any object on the supply voltage cable. Install the line voltage cablein such a manner that nobody can step on it or fall over it.

Protect the line voltage and signal cables for damage and moisture ingress.

This product is a Safety Class 1 instruments

This product is provided with a protective earth terminal.To prevent shock hazard, the instrument chassis and cabinet must be connectedto an electrical ground. The instrument must be connected to the AC powersupply mains through a three-conductor power cable, with the third wire firmlyconnected to an electrical ground (safety ground) at the power outlet.If the instrument is to be energized via an external power supply, that goes for thepower supply too.Any interruption of the protective (grounding) conductor or disconnection of theprotective earth terminal will cause a potential shock hazard that could result inpersonal injury. Deliberate disconnection is inadmissible / prohibited !

Page 11: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

SAFETY SUMMARY

S - 590003806(1) NGA-TFID e 16.06.98

Rosemount Analytical

4. Analyzer specific notes for the user

First of all operations unscrew transfer safety lock (knurled-head screws)for the optional sucking pump of the analyzer (Item 5. of operation manual) !

The installation site for the instrument has to be dry and remain above freezingpoin t at all times.The instrument must be exposed neither to direct sunlight nor to strong sourcesof heat. Be sure to observe the permissible ambient temperature !For outdoor sites, we recommend to install the instrument in a protective cabinet.At least, the instrument has to be protected against rain (e.g., shelter).

Free flow of air into and out of the equipment must not be hindered bynearby objects or walls !

For rack installation be sure, that the TFID has space on top and on bottom ofmin. 1/2 HU to avoid overheating !

Do not interchange gas inlets and gas outlets !All gases have to be supplied to the analyzer as conditioned gases !When the instrument is used with corrosive gases, it is to be verified that thereare no gas components which may damage the gas path components.

The exhaust gas lines have to be mounted in a declining, descending,pressureless and frost-free and according to the valid emission legislation !

In case it is necessary to open the gas paths, close the analyzersgas connections with PVC caps immediatly !

Use only from our factory optional delivered cables or equivalent shielded cablesto be in agreement with the CE - conformity.The customer has to guarantee, that the shield is be connected bothsided.Shield and connectors housing have to be connected conductive.Sub.-min.-D-plugs/sockets have to be screwed to the analyzer.

By using of optional delivering terminal strip adapters the analyzer is not be inagreement with the CE - conformity. In this case CE - conformity is to be declaredby customer as “manufacturer of system”.

ANALYZER SPECIFIC NOTES FOR USER

Page 12: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

SAFETY SUMMARY

S - 6 90003806(1) NGA-TFID e 16.06.98

Rosemount Analytical

5. Additional notes for service / maintenance

Operating personnel must not remove instrument covers !Component replacement and internal adjustments must be made by qualifiedservice personnel only !

Always disconnect power, discharge circuits and remove external voltagesources before troubleshooting, repair or replacement of any component !

Any work inside the instrument without switching off the power must beperformed by a specialist, who is familiar with the related danger, only !

In case of exchanging fuses the customer has to be certain that fuses of specifiedtype and rated current are used. It is prohibited to use repaired fuses or defectivefuse holders or to short-circuit fuse carriers (fire hazard).

In case of exchanging optional sucking pump, verify whether the line voltagestated on the label of the TFID agrees with that of the sucking pump !

At component replacement or installation the RF shielding contacts must not bebended !

ADDITIONAL NOTES FOR SERVICE / MAINTENANCE

Page 13: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

SAFETY SUMMARY

S - 790003806(1) NGA-TFID e 16.06.98

Rosemount Analytical

5.1 Electrostatic Discharge

The electronic parts of the analyzer can be irreparably damaged if exposed to electrostatic

discharge (ESD).

The instrument is ESD protected when the covers have been secured and safety precautions

observed. When the housing is open, the internal components are not ESD protected anymore.

Although the electronic parts are reasonable safe to handle, you should be aware of the following

considerations:

Best ESD example is when you walked across a carpet and then touched an electrical grounded

metal doorknob. The tiny spark which has jumped is the result of electrostatic discharge (ESD).

You prevent ESD by doing the following:

Remove the charge from your body before opening the housing and maintain during work with

opened housing, that no electrostatic charge can be built up.

Ideally you are opening the housing and working at an ESD - protecting workstation.

Here you can wear a wrist trap.

However, if you do not have such a workstation, be sure to do the following procedure exactly:

Discharge the electric charge from your body. Do this by touching a device that is grounded

electrically (any device that has a three - prong plug is grounded electrically when it is plugged into

a power receptacle).

This should be done several times during the operation with opened housing (especially after

leaving the service site because the movement on a low conducting floors or in the air might cause

additional ESDs).

ELECTROSTATIC DISCHARGE

Page 14: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

SAFETY SUMMARY

S - 8 90003806(1) NGA-TFID e 16.06.98

Rosemount Analytical

Page 15: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

P - 1

PREFACE

90003806(1) NGA-TFID e 16.06.98

Rosemount Analytical

Preface

The Thermo-FID (TFID) as member of the NGA 2000 familiy has all the advantages of a modular

analyzer system. A TFID Analyzer Module (AM) as “blind” analysis unit it can be part of a multiple

NGA analyzer system or special customer-developed control unit. TFID with front panel display/

operator interface is a stand-alone analyzer or central interface (control module) for multiple

analyzer modules.

The TFID is used in several areas of industry, safety of environment, research and develop-ment.

Typical applications are the LEL control, the emission and ambient air control or the exhaust gas

analysis in the chemistry and combustion engine development. But you can also use the TFID to

optimize a process or to control any limits at operating position.

A lot of auxiliary functions are available by the micro electronic of the TFID. The integrated CPU

allows the operation via menu, the automatic self control and error diagnostics. The display shows

all instruction notes and auxiliary reports in plain language. The very small currents of the FID

detector are always evaluated in the optimum range by the processor controlled dynamic amplifier.

The amplified signals will be digitalized directly at the detector and sent out to the CPU. So, a

falsification of the small detector signal will become impossible and the high resistant screened

cable will become unnecessary.

Technical Properties of Measurement

Integrated licensed backfiring flame trap of the “Ex s IIC class”

Electronic control of the sample gas flow

Electronic control of the accelerating voltage

Flame control with automatic ignition

Electronic control of the ignition power

Automatic calibration of the flame size

Self calibration without any other valves or components

Large range of linearization without switching of the coupling

FID amplifier and A/D transformer are connected with the detector directly wireless

Integrated thermic catalyst for burning air, zero gas and stripper

Determination of organic carbon (org. C) in superheat vapor or condensate

Determination of organically bound carbon down to LEL

Determination of organically bound carbon according to “17th BImSchV” (in progress)

Page 16: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

P - 2

PREFACE

90003806(1) NGA-TFID e 16.06.98

Rosemount Analytical

Page 17: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

TECHNICAL DESCRIPTION

1 - 190003806(1) NGA-TFID e 22.06.98

Rosemount Analytical

1. Technical Description

All components of analyzer or analyzer module are incorporated into a 1/1 19" housing.The housing is available as rack mounting (R) or as table-top (T) versions. For analyzer modulesthere is mounted a blind plate instead of an operation front panel.

The Flame Ionization Detector (FID) is designed to continuously determine concentrations ofhydrocarbons.

The TFID offers the following Advantages:a) The instrument is designed modular according to function groups resulting in:

Easy troubleshooting Fast error removal High availability

b) The analyzer is sucking automatically sample gas via a pump system being installed behindthe detector.

c) The sample gas is sucked direct into the detector without using a superposed sample gaspump. Therefrom: No heated sample gas pump No additional heating of sample gas inlet (or heated bypass as option) No “cold” connections leading to condensation; in that way avoiding corrosion and

signal falsification Fast response (to achieve values)

1.1 Front View

The front panel of the analyzer is the operating front panel (see Fig. 1-1).Measured values and the entire operating procedure are displayed on a LC display. The operationand programming of the instrument is performed by using the four cursor keys, the ENTER keyand the five soft keys

For analyzer modules there is mounted a blind plate instead of an operation front panel.

Page 18: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

TECHNICAL DESCRIPTION

1 - 2 90003806(1) NGA-TFID e 16.06.98

Rosemount Analytical

1234567890123456789012345678901234567890123456789012345678901234567890123456789012345678901234567890123456789012345678901234567890123456789012345678901234567890123456789012345678901234567890123456789012345678901234567890123456789012345678901234567890123456789012345678901234567890123456789012345678901234567890123456789012345678901234567890123456789012345678901234567890123456789012345678901234567890

Fastening screws for rack mounting or carrying-strap bracket

ENTER key

LC display Cursor keys

Soft keys(function is depending on Software Menu)

Fastening screws for rack mounting or carrying-strap bracket

Fig. 1-1: Operation front panel, Front view

FRONT PANEL

Fig. 1-2: TFID, Front view

optional Fan(together with internal sucking pump option)

Blind plate oroperation front panel

Page 19: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

TECHNICAL DESCRIPTION

1 - 390003806(1) NGA-TFID e 16.06.98

Rosemount AnalyticalREAR PANEL

1.2 Rear Panel

On the rear panel the connector for 230 or 115 Vac supply, the gas connections, the networkconnections and the connectors of optional PCBs (see optional, separate Operation Manuals) areaccommodated.

Fig. 1-3: TFID, Rear view

Network connection(RJ 45 socket)

Input 230 or 115 Vac Gas connections(Item 5.3)

optional PCB[SIO/DIO e.g.]

Connection Pump Controlfor optional internal sucking pump

(Sub. Min. D. 9 pin)

Gas connections(optional internalsucking pump)

Output 230 or 115 Vac(max. 4 A)

Page 20: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

TECHNICAL DESCRIPTION

1 - 4 90003806(1) NGA-TFID e 22.06.98

Rosemount AnalyticalINTERNAL CONSTRUCTION

1.3 Internal Construction

By frontal view, the electronic unit with interconnection PCB and other PCBs is located on the right.The analysis assembly and other parts are located on the left.

analysis assembly internal power supply optional I/O-boards

ACU PCB

sucking pump(option)

front panel

Fig. 1-4: TFID, internal construction

PCB LEM 02(network in)

Page 21: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

TECHNICAL DESCRIPTION

1 - 590003806(1) NGA-TFID e 22.06.98

Rosemount Analytical

1.3.1 Thermo-FID Assembly

a) Detector Unit

including the following component parts: Detector block Heating unit and temperature control Flame control Ignition module with ignitor and ignition current control Flame arrestors and sample gas orifice, exchangeable without demounting the

detector Accelerating voltage module with control unit FID preamplifier incl. AC/DC converter Air jet injectors for operation with instrument air

(as option TFID is using alternitavely a diaphragm pump for sample gas transportwhich is not located in the detector block but on rear side of TFID or alternativelyoutside).

b) Gas Distribution Block

Consisting of the following Parts: Gas distribution block with stream limitating orifices for fuel gas (hydrogen, H2/He),

burner air, span gas, external zero gas, sample gas and sucking pump (option) Differential pressure sensors with preamplifier for burner and sample gas pump

or sucking pump (option), sample gas pressure and fuel gas pressure (H2, H2/He) Electronic pressure sensors for instrument air and span gas Electronic controlling valve for fuel gas (hydrogen, H

2/He), 2 controlling valves

for the air jet injectors or sucking pump (option), zero and span gas valve

c) Internal Gas Supply (Options)

Consisting of the following parts: Zero gas pump, sample gas pump and catalyst (for burner air in the detector device)

The complete pneumatic constructional units for detector supply as well as thesensors for determination of sample gas flow do come into contact only withclean instrumental air thus showing a very good wear charactiristic.

d) Electronic Unit CPU with bus board Power supply 230 Vac or 115Vac and 230 Vac or 115Vac/ 4A power output digital/analog interface board for internal supply of detector block,

gas distribution block, catalyst, and internal pumps.

INTERNAL CONSTRUCTION

Page 22: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

TECHNICAL DESCRIPTION

1 - 6 90003806(1) NGA-TFID e 16.06.98

Rosemount Analytical

1.3.2 Printed Circuit Boards

All necessary PCBs are pushed in into a cardcage (see Fig. 1-4 and 1-5).

optional PCB(e.g. SIO / DIO)

TFID Rear Panel

Interconnection Board(ICB)

Network Inlet Module(LEM)

Cardcage

Controller Board (ACU)

TFID Front Side

Fig. 1-5: Cardcage TFID, Top View

ICB

ICB is an interconnection board consisting of six 64-pin ICB bus slots to accommodate printedcircuit boards (PCB of Euro standard format).

PRINTED CIRCUIT BOARDS

Page 23: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

TECHNICAL DESCRIPTION

1 - 790003806(1) NGA-TFID e 16.06.98

Rosemount Analytical

ACU

The ACU 01 card is the main part of the TFID.It is a one-board CPU card.All user data are safe loaded via a battery buffer, even when the voltage supply is disconnected,.

This card contains the complete software for controlling and operating the analyzer module.For MLT analyzers (with operation front panel) additional the Control Unit Software [functions ofplatform (network managment, control of LCD display e. g.)] is included.

The board contains the following function blocks:

32 / 16-bit micro controller

Floating point co-processor

dependent on TFID version different RAM / Flash-EEPROM

Real-time Clock withCalendar functionAlarm function

Watchdog function

Serial interface RS 232 C

A network interface with ECHELON chip

System bus:Parallel bus A6 : D812 TPU wiringnetwork busSynchron serial bus

Buffered parallel interface for LCD control

Local bus interface (e.g. storage extension)

PRINTED CIRCUIT BOARDS

Page 24: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

TECHNICAL DESCRIPTION

1 - 8 90003806(1) NGA-TFID e 22.06.98

Rosemount Analytical

1.4 Applications

The TFID is designed for continuous monitoring of total hydrocarbon concentrations or otherburnable gases in air, flue gas, process and pure gas.

The TFID has a high dynamic linear range over several decades and is a fast response analyzerwith smaller than 1,5 sec. (for ranges > 10 ppm) or 2,5 sec (range: 4 ... 10 ppm).

Applications:

Internal Combustion Engine Emissions (ICEE) Ambient Air Monitoring Continuous Emissions Monitoring Systems (CEMS) Detection of TRK and MAK Values (human protection) Monitoring of Lower Explosion Limits (LEL) Production Plants Fuel Stations Working Places Emission Sources of Production Plants Stack Gas Monitoring of Waste Incinerator Plants Charcoal Stack Gas Cleaning Systems (Inlet and Emissions) Process of Afterburner Plants

APPLICATIONS

Page 25: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

THEORY OF TECHNOLOGY AND OPERATION

2 - 190003806(1) NGA-TFID e 16.06.98

Rosemount AnalyticalMEASURING PRINCIPLE

2. Theory o f Technology and Operation

2.1 Measuring Principle

The Flame Ionization Detector measuring principle is based on ionization of hydrocarbons in a

hydrogen flame (or hydrogen/ helium mixture ).

Since a pure hydrogen flame creates only a very small basic ionization the burning of

hydrocarbons in ppm range results in ionization currents being decades higher than pure

hydrogen.

Thus a constant sample gas stream and a constant hydrogen or H2/He stream are mixed and fed

to the detector over a burner nozzle being put on a negative potential (approx. - 80 Vdc to

- 400 Vdc). In the burner chamber of the detector the mixture is burned by adding hydrocarbon

free air in relation of approx. 10 : 1.

The resulting ions are collected on a polarized electrode, amplified and put to display.

Page 26: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

THEORY OF TECHNOLOGY AND OPERATION

2 - 2 90003806(1) NGA-TFID e 16.06.98

Rosemount Analytical

Fig. 2-1: Measuring Principle for TFID Measurement

Sample Gas

Burner Air

Hydrogen

or H2/He

Accelerating

Voltage

Burner

Data

Acquisition

Detector

Ion Current

Flame

CPU

Control / Data Acquisition

MEASURING PRINCIPLE

Page 27: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

THEORY OF TECHNOLOGY AND OPERATION

2 - 390003806(1) NGA-TFID e 22.06.98

Rosemount Analytical

2.2 TFID Mode of Operation

The burning of organic substances in a hydrogen flame will produce electrical charged particles

(ions). A negative acceleration voltage at the burner nozzle will discharge these ions to the

electrode of the current amplifier. The high sensitive amplifier will evaluate the current of this

process. The intensity of this current depends on the concentration of the carbon atoms in the

sample gas. Therefore it is possible to determine the measured value of the organic carbon

(organic C) directly.

A flame is produced in a heated burner chamber by burning pure hydrogen and added clean air.

Besides a part of the sample gas will be fed to the flame. A constant sample mass flow is absolutely

necessary to determine accurate and precise the concentration of the organic carbon. It is also

necessary that deviations of the sample gas pressure may not change the pressure proportions

in the burner chamber and the sample mass flow to the flame.

A constant mass flow over a long period is achieved by special transport of the sample gas and

the supply gases. Even big pressure deviations of the sample gas against atmosphere will have

no significant influence to the signal. Two independent electronic differential pressure gauges

keep constant the differential pressure of the constant sample mass flow to the flame by adding

separate air. The differential pressure system will always be kept on a constant pressure difference

against atmosphere. Thus, neither deviations of air pressure nor differences in altitudes will have

any influence to the measured values.

Instead of a conventional heated sample gas pump a heated air jet injector is inserted each behind

the burning chamber resp. behind the flame arrestor. As option it is also possible, that an unheated

diaphragm pump is used instead of the heated air jet injectors.

The suction pump can be placed either on the rear side of TFID or external.

The sample gas has no direct contact with the used control fittings. Even the electronic sample

gas flow control happens without a flow meter and without direct contact with the sample gas.

Because of this construction the Thermo-FID has a very small dead volume. As consequence the

Thermo-FID has very short response times to determine the measured values. The compact

detector block is heated between 120 °C and 160 °C (up to 210 °C as option) being pre-selective

to avoid condensation by burning and corrosion of the materials.

TFID MODE OF OPERATION

Page 28: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

THEORY OF TECHNOLOGY AND OPERATION

2 - 4 90003806(1) NGA-TFID e 22.06.98

Rosemount AnalyticalSAMPLE GAS TRANSPORT

2.3 Sample Gas Transport

The sample gas transport through the detector is realized by two independent pumps as standard.

2 heated air jet injectors and as option one unheated diaphragm pump sucking before or behind

the burner chambers are used.

The pumps create before and behind the burner chamber an underpressure being kept constant

via 2 independent differential pressure gauges to different pressures.

Between the differential pressure gauges and before the burner chamber an orifice is placed

keeping the sample mass flow to the burner nozzle constant.

Using this principle of sample gas transport a heated pump is not necessary.

Pressure deviations of about -300 mbar to + 600 mbar against atmosphere do not have any

significant influence on the result. With a optional sucking pump -200 mbar to + 600 mbar against

atmosphere is possible. This means very low response times and no additional heating of the

sample inlet.

The volume of the sucked sample gas is about 10 to 25 l/h (25 l/h as standard). If very small dead

times are requested, the sucked sample gas can be increased to approx. 80 - 100 l/h as option.

At detector sample gas inlet a flame trap is located which has detector temperature.

Behind the heated flame trap working also as sample gas filter an electronic flow control is placed

being not in direct contact with the sample gas. This is avoiding contamination.

All materials being in contact with sample gas are heated to 120 - 210 °C (pre-selective) and high

corrosion resistant.

Page 29: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

THEORY OF TECHNOLOGY AND OPERATION

2 - 590003806(1) NGA-TFID e 22.06.98

Rosemount Analytical

Fig. 2-2: TFID, Flow / Function Schemstic

Page 30: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

THEORY OF TECHNOLOGY AND OPERATION

2 - 6 90003806(1) NGA-TFID e 16.06.98

Rosemount Analytical

Page 31: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

PREPARATION

5 - 190003806(1) NGA-TFID e 16.06.98

Rosemount Analytical

5. Preparation

Please check the packing and its contents immediately upon arrival.

If any item is damageg or lost you are kindly requested to notify the forwarder to undertake a

damage survey and report the loss or damage to us immediately.

First of all operations unscrew transfer safety lock (both knurled-head screws)

for the optional sucking pump on bottom side of the analyzer/module (Fig. 5-1) !

For transport of equipment the transfer safety lock absolutely have to be locked!

Fig. 5-1: Transfer safety lock(housing side view, detail sketch)

Transfer safety lock (two knurled-head screws)

Front panel

Page 32: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

PREPARATION

5 - 2 90003806(1) NGA-TFID e 16.06.98

Rosemount Analytical

5.1 Installation Site

Be sure to observe the additional notes, safety precautions and warnings given

in the individual manuals (platform, modules, I/O boards etc.) !

The analyzer/module must not operate in explosive atmosphere without

supplementary protective measures !

Free flow of air into and out of the TFID (ventilation slits) must not be

hindered by nearby objects or walls !

For rack installation be sure, that the TFID has space on top and on bottom of

min. 1/2 HU to avoid overheating !

The installation site for the TFID has to be dry and remain above freezing

point at all times. The TFID must be exposed neither to direct sunlight nor

to strong sources of heat.

Be sure to observe the permissible ambient temperatures (c.f. Item 20: Technical

Data). For outdoor installation, we recommend to install the TFID in a protective

cabinet. At least, the TFID has to be protected against rain (e.g., shelter).

For Installation site please take care about safe way to reach the analyzer.

The equipment has to be installed as near as possible to the sample point , in order to avoid

low response time caused by long sample gas lines.

INSTALLATION SITE

Page 33: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

PREPARATION

5 - 390003806(1) NGA-TFID e 16.06.98

Rosemount Analytical

5.2 Gas Conditioning (Sample Handling)

The conditioning of the sample gas is of greatest importance for the successful operation of any

analyzer according to extractive method.

All gases have to be supplied to the TFID as conditionned gases !

When the instrument is used with corrosive gases, it is to be verified that there

are no gas components which may damage the gas path components.

The gas has to fulfill the following conditions:

It must be

free of condensable constituents being not compatible with detector temperature

free of dust

free of aggressive constituents which are not compatible with the material of the gas

paths.

have temperatures and pressures which are within the specifications stated in “Technical

Data” of this manual.

Inflammable or explosive gas mixtures may not be introduced into the TFID

without supplementary protective measures !

When analysing vapours, the dewpoint of the sample gas has to be at least 10 °C below the

detector temperature in order to avoid the precipitation of condensate in the gas paths.

Suitable gas conditionning hardware may be supplied or recommended for specific analytical

problems and operating conditions.

GAS CONDITIONING (SAMPLE HANDLING)

Page 34: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

PREPARATION

5 - 4 90003806(1) NGA-TFID e 22.06.98

Rosemount Analytical

5.3 Gases

Be sure to observe the safety regulations for the respective gases

(sample gas and test gases / span gases) and the gas bottles !

The test gas containers have to be set up according to the current legislation.

5.3.1 Supply Gases

Burner Air:Ambient air via internal catalystOption: For small ranges (< 100 ppm) hydrocarbon free synthetic air is

recommended.(separate input as option, secondary pressure: 2,000 hPa = 14.5 psig)

Fuel Gas (Burner Gas):Hydrogen 5.0 from pressure gas cylinders via integrated flow controller and shut-offcontrol valve incl. pressure and leakage control.Option: Mixture with 40 % Hydrogen/ 60 % Helium, quality 5.0, pre-selectiveOption: Gas cylinder pressure reducers with supply and leakage control

(secondary pressure: 2,000 hPa = 14.5 psig)

Zero Gas:Ambient air via internal thermal catalystZero gas supply via integrated zero gas pump and solenoid valveOption: External zero gas supply from zero gas cylinder via pressure regulator

(secondary pressure: 3,000 hPa = 29 psig)

Span Gas:Span gas supply via integrated solenoid valveIntegrated flow limiting device and pressure controlOption: Propane (butane or methane) in synthetic air or nitrogen from calibration

gas cylinder via regulator (secondary pressure: 3,000 hPa = 29 psig)

Compressed Air Supply:For Thermo-FID operation with air jet injectors instrument air or oil, dust and water freeair (dew point - 20 °C) with a inlet pressure of minimum 5,000 hPa (58 psig) has to besupplied. An integrated pressure sensor controls the pressure of the compressed airsupply.Option: Air compressor with dust filter and moisture separator

(secondary pressure: 5,000 hPa = 58 psig)Option: Suction diaphragm pump instead of air injectors and instrument air to be

installed in TFID or external and flow limiting device

SUPPLY GASES CONNECTIONS

Page 35: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

PREPARATION

5 - 590003806(1) NGA-TFID e 22.06.98

Rosemount Analytical

5.3.2 Gas Connections

The installed gas connections are specific to the TFID. All fittings are clearly marked.

The fittings are located on the rear panel of the instrument.

The exhaust gas lines have to be mounted in a declining, pressureless

and frost-free way and according to the valid emission legislation !

Do not interchange gas inlets and gas outlets !

All pneumatic supplies and mains, especially the hydrogen mains, have to be

tight ! All gases used must have the right input pressure !

The following standard gas connections are installed (Fig. 5-2 and 5-3):

Fuel gas (hydrogen 5.0) or H2/He (40/60) Ø 6/4 mm tube fitting, ss

Burner air, external (option) Ø 6/4 mm tube fitting, ss

Compressed air supply Ø 6/4 mm tube fitting, ss

Sample gas in Ø 6/4 mm tube fitting, ss

Zero gas, external (option) Ø 6/4 mm tube fitting, ss

Span gas Ø 6/4 mm tube fitting, ss

Calibration gas out (to Cal. gas in) tube connection, ss

Calibration gas in (at sample in) tube connection, ss

Exhaust (optional to sucking pump) Ø 6/4 mm tube fitting, ss

Exhaust “In” (option sucking pump, from exhaust) Ø 6/4 mm tube fitting, ss

Exhaust “Out” (option sucking pump) Ø 6/4 mm tube fitting, ss

Instead of Ø 6/4 mm tube fittings 1/4" O.D. tube fittings are available as option.

For option sucking pump the gas inlet marked “air press.” is equipped with a dust filter instead of

a gas fitting.

GAS CONNECTIONS

Page 36: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

PREPARATION

5 - 6 90003806(1) NGA-TFID e 22.06.98

Rosemount AnalyticalGAS CONNECTIONS

Fig. 5-2: TFID, gas connections (drawing)

Exhaust out(option sucking pump)

Cal.-Gas out

(to Cal.-Gas in)

Span-Gas

in

Cal.-Gas in

(from

Cal.-Gas out)

Sample-Gas

in

H2 or H

2/He

in

Zero-Gas external in

(option)

Exhaust(optional to sucking

pump)

Exhaust in(option sucking pump)

Fig. 5-3: TFID, gas connections (photo)

Compressed air supply in(dust filter for option sucking pump)

Burner air external in

(option)

Page 37: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

SWITCHING ON

6 - 190003806(1) NGA-TFID e 16.06.98

Rosemount Analytical

6. Switching On

Be sure to observe the safety precautions and warnings !

Be sure to observe the additional notes, safety precautions and warnings given

in the individual manuals (platform, modules, I/O boards etc.) !

Once the instrument has been correctly assembled and installed in accordance with the general

instructions given in section 5., the equipment is ready for operation.

Fig. 6-1: TFID, Voltage supply

Plug POWER

(230 or 115 Vac IN)

Network connection

(RJ 45 socket)

Terminal strips

(Pt 100)

Mains Switch 230 or 115 Vac (max. 4 A) OUT

(option external heater)

Page 38: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

SWITCHING ON

6 - 2 90003806(1) NGA-TFID e 16.06.98

Rosemount Analytical

The equipment has an internal power supply and is specified for an operating voltage of 230 Vac

or 115 Vac resp., 50-60 Hz.

For supply of external components (external heater) there is built in a 230 Vac or 115 Vac outlet

plug (female, max. 4 A). For temperature control of the external heater there are terminal strips

to connect the thermo element of the external heater (Pt 100) to the TFID (Fig. 6-1).

Be sure, that all gas supply valves are closed !.

Connect network to other analyzers / modules (see Fig. 6-1).

Connect mains line and internal power supply (see Fig. 6-1, Plug POWER).

Verify before operation whether the line voltage stated on the label of the

TFID agrees with that of your mains line!

Optional connect external heater to voltage supply and Pt-100 to terminal strips (Fig. 6-1).

Switchin On the TFID via mains switch (Fig. 6-1, switch to its "I" position).

For analyzers only:

The analyzer must be switched on only after switching on all modules connected

to the network.

Upon switching on, the TFID will perform a self-diagnostic test routine.

After switching on, the software integrates automatically all the individual modules, which are

incorporated in the TFID via network and have been previously switched on, in the system.

You will find the “Single menu display” and raw values on the display and as secondary value

the message „WarmUp“. If the other secondary values show “Failures”, “Maintenance-

Requests” and “Function-Control” you can see “YES” for each of them.

For further informations about the displays while switching on look at the respective software

manual.

Warming-up after switching on takes about 45 minutes,

depending on the installed detectors and themostatic control !

Page 39: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

SWITCHING ON

6 - 390003806(1) NGA-TFID e 16.06.98

Rosemount Analytical

Put in the span gas concentration (Vol.-% or ppm) of the span gas certificate, the number

of carbon atoms, the name of the span gas and the response factor (see Item 20.5) !

The correctness of these values is very important for the automatic calibration !

(The standard concentration of the span gas is approx. 500 ppm propane in nitrogen.)

Now, you can start the pneumatic operations:

Open the supply gas valves and set up the supply gas pressure values !

The calculator will start with the automatic startup procedure after the catalyst and the

detector have reached the necessary temperature.

The following procedures will run automatically:

Test of the temperatures

Test of the control sensors

Calibration of the differential pressures

If all inputs are working correct the startup procedure will continue automatically with:

Switch on of the sample gas system - pump will start automatically

Control of the fuel gas (hydrogen or H2/He) pressure

Set up of the fuel gas (hydrogen, H2/He) flow rate for the flame

After fuel gas, zero and span gas pressure are correct and warm-up is finished the

messages “Failure” and “Maintenance-Request” will disappear

Ignition of the flame

After a successful ignition of the flame the automatic internal calibration will start.

Page 40: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

SWITCHING ON

6 - 4 90003806(1) NGA-TFID e 16.06.98

Rosemount Analytical

Automatic Internal Calibration

Operation will switch from “WarmUp” to “Calibration” and Function Control shows “YES”.

When the procedure is finished Operation shows “Ready” and Function Control “NO”.

The internal automatic Calibration procedure will be:

Purging with zero gas

After the purging time: Zero calibration.

Purging with span gas

After the purging time: Span calibration.

Purging free with zero gas

Switching to sample gas

The calculator will switch to the measurement cycle if the current calibration curve agrees with

the reference calibration curve of the initial calibration.

The first startup may contain some mistakes:

Maintenance request will be set if the deviation from the initial calibration curve is more than

entered value

The deviation from the initial calibration curve during the first startup of the system happens

normally only for small ranges. This effect will vanish after a certain operation time.

Notes:

You have to use nitrogen as zero gas for ranges <20 mg/m3 if the concentration of hydrocarbons

in the environment air is more than 10 mg/m3 and cannot be transformed catalytic to CO2.

The span gas rate always has to be greater than the sample gas rate to avoid dilution effects with

the air. Normally, you can achieve it with a span gas pressure of 3 bar abs.

A suitable span gas is propane in nitrogen or in synthetical air. The span gas concentration should

be 60 to 80% of the required measurement range. It can be necessary to calculate “response factors”

because the measurement signal of a flame ionization detection may differ for different hydrocar-

bons.

Before start of measurement we recommend entry of the desired TFID parameters and to zero

and span the TFID.

Page 41: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

7 - 1

SWITCHING OFF

90003806(1) NGA-TFID e 16.06.98

Rosemount Analytical

7. Switching Off

Before switching off the analyzer, we recommend first purging all the gas lines for about 5 minutes

with zeroing gas (N2) or adequate conditioned air. The full procedure for shutting down is as follows:

Admit zeroing gas at the respective gas inlet fitting.

Set the presasure resp. gas flow rate to allowable rate.

After 5 minutes have elapsed:

Switch Off the TFID via mains switch (Fig. 6-1, switch to its "0" position).

Shut Off the gas supply.

Close all gas line fittings immediately.

Page 42: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

7 - 2

SWITCHING OFF

90003806(1) NGA-TFID e 16.06.98

Rosemount Analytical

Page 43: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

OPENING OF HOUSING

90003806(1) NGA-TFID e 16.06.98 15 - 1

Rosemount Analytical

15. Opening the Housing

The housing must be opened for checking the electrical connections and for replacement orcleaning of any of the components of the analyzer.

Be sure to observe Item 5. of the safety measures !

Switch Off the TFID via mains switch (Fig. 6-1, switch to its "0" position) and disconnect allvoltage supplies.

15.1 Housing Cover

Unscrew the 8 fastening screws on top and pull the cover straight up.

Closing of the housing is performed in reverse order.

15.2 Front Panel

Unscrew the 4 or 6 fastening screws for respective panel (Fig. 15-1) and remove respectivefront panel to the front carefully.

Optional components are mounted at front panel !

Closing of the housing is performed in reverse order.

Do not squeeze cables or gas lines !

Fig. 15-1: TFID (Fastening screws front panels)

Fastening screws, operation front panel or blind plate

Fastening screws, blind plate with fan Fastening screws, operation front panel or blind plate

1234567890123456789012345678901234567890123456789012345678901234567890123456789012345678901234567890123456789012345678901234567890123456789012345678901234567890123456789012345678901234567890123456789012345678901234567890123456789012345678901234567890123456789012345678901234567890123456789012345678901234567890123456789012345678901234567890123456789012345678901234567890123456789012345678901234567890

Page 44: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

OPENING OF HOUSING

90003806(1) NGA-TFID e 16.06.9815 - 2

Rosemount Analytical

Page 45: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

CLEANING OF HOUSING OUTSIDE

19 - 190003806(1) NGA-TFID e 16.06.98

Rosemount Analytical

19. Cleaning of Housing Outside

For cleaning TFID housing outside, you need a soft, fluff free cloth and all purpose detergent.

Switch Off the TFID via mains switch (Fig. 6-1, switch to its "0" position).

To avoid a danger to the operators by explosive, toxic or unhealthy gas

components, first purge the gas lines with ambient air or nitrogen (N2) before

cleaning or exchange parts of the gas paths.

If it is necessary to disconnect the gas connections, the gas line fittings of the

TFID have to be closed with PVC-caps before cleaning !

Moisten of the soft, fluff free cloth with the cleaning solution(mixture of 3 parts water, 1 part all purpose detergent max.).

Be sure to use a moisted, but not wet, cloth only !

Be sure, that non liquid can drop into the housing inside !

Cleaning of the TFID housing outside with the moisted cloth.

If required, rub off the housing with a dry cloth afterwards.

Page 46: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

CLEANING OF HOUSING OUTSIDE

19 - 2 90003806(1) NGA-TFID e 16.06.98

Rosemount Analytical

Page 47: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

TECHNICAL DATA

20 - 190003806(1) NGA-TFID e 22.06.98

Rosemount Analytical

20. Technical Data

Certifications EN 50081-1, EN 50082-2, EN 61010-1

(in progress) NAMUR, CSA NRTL/C, C-Tick, PAC, IS

Suitability tests: TÜV (in progress):TI Air, 17th BImSchV

20.1 Housing

Gas connections:Compressed air supply Ø 6/4 mm tube fitting, ss

(pressure max. 5,000 hPa / 58 psig)Hydrogen (qualitiy 5.0) or H2/He (40/60) Ø 6/4 mm tube fitting, ss

(pressure max. 2,000 hPa / 14,5 psig)Burner air external (option, Ø 6/4 mm tube fitting, sshydrocarbon free synthetic air) (pressure max. 2,000 hPa / 14,5 psig)Exhaust (optional to sucking pump) Ø 6/4 mm tube fitting, ssCal.-Gas out tube connection, ssCal.-Gas in (at sample in) tube connection, ssSample-Gas in Ø 6/4 mm tube fitting, ss

(pressure max. 700 - 1,600 hPa for air injector,max. 800-1,600 hPa for sucking pump option)

Zero-Gas external (option) Ø 6/4 mm tube fitting, ss(pressure max. 3,000 hPa / 29 psig)

Span-Gas Ø 6/4 mm tube fitting, ss(pressure max. 3,000 hPa / 29 psig)

Exhaust “In” (option sucking pump, from exhaust) Ø 6/4 mm tube fitting, ssExhaust “Out” (option sucking pump) Ø 6/4 mm tube fitting, ss

Instead of Ø 6/4 mm tube fittings 1/4" O.D. tube fittings are available as option.

Housing dimensions see dimensional sktech (Fig. 20-1)

Weight (depending on configuration) approx. 10 - 15 kg

Protection class IP 20 (according to DIN standard 40050)

HOUSING

Page 48: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

TECHNICAL DATA

20 - 2 90003806(1) NGA-TFID e 16.06.98

Rosemount Analytical

20.2 Ambient Conditions

Permissible ambient temperature + 5 °C to + 40 °C

Humidity (non condensing) < 90 % rel. humidity at + 20 °C< 70 % rel. humidity at + 40 °C

Rain / Drop and splash water The TFID must not be exposed to rain ordrop/splash water

Explosive atmosphere The TFID must not be operated in explosiveatmosphere without supplementaryprotective measures

Ventilation Free flow of air into and out of the TFID(ventilation slits) must not be hindered bynearby objects or walls !

Altitude 0-2000 m (above sea level)

DIMENSIONS TFID / AMBIENT CONDITIONS

Front view

Top view

448.8

465.1

483

57.1

132.

5

140

appr

ox. 5

37

appr

ox. 4

88

Fig. 20-1: Dimensional sketch TFID [all dimensions in mm]

123456789123456789123456789123456789123456789123456789123456789123456789

Page 49: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

TECHNICAL DATA

20 - 390003806(1) NGA-TFID e 22.06.98

Rosemount AnalyticalGENERAL SPECIFICATIONS / VOLTAGE SUPPLY

20.3 General Specifications

Measuring component organic CMeasuring ranges 0 - 1 ... 100,000 mg org. CRanges selectable Linear reference curveUnits selectable ppm, mg/m3, g/m3, Vol.-%, %LELDetection limit < 0.1 mg org. C/m3

90%-response time at sample gas input < 1,5 sec. (for ranges > 10 ppm)< 2,5 sec (range: 4 ... 10 ppm)

Detector temperature pre-selectable 120 °C - 160 °C (up to 210 °C as option)Auxiliary heating 60 °C - 210 °CThermo element of auxiliary heating Pt 100

Gas flowsample gas self sucking (at 1013 hPa) approx. 25 l/h or approx. 90 l/hhydrogen approx. 40 ml/min.hydrogen / helium (option) approx. 160 ml/min.span gas (zero gas as option) approx. 130 Nl/hburner air (option) approx. 25 l/h

20.4 Voltage supply

Input: plug, maleNominal voltage 230 or 115 Vac, 50 / 60 HzInput power nom. 300 VA / max. 1,000 VAFuse (external) T10A/230V

Output for external heater plug, femaleOutput voltage 230 or 115 Vac, 50 / 60 Hz, max 4 A

Output power 920 VA (230 Vac) / 460 VA (115 Vac)Fuse (external) T4A/230V

Detector power (Fuse, internal) 150 VA (T2A/230V)Catalyst power (Fuse, internal) 100 VA (T1A/230V)Power supply, internal Fuse T4A/230V

Page 50: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

TECHNICAL DATA

20 - 4 90003806(1) NGA-TFID e 16.06.98

Rosemount AnalyticalELECTRICAL SAFETY / RESPONSE FACTORS

20.4.1 Electrical Safety

Over-voltage category II

Pollution degree 2Safety Class 1

all I/O´s SELV voltageoptically isolated to electrical supply

20.5 Response Factors

Substance Response Factor

n-butane 1.00Methane 1.06Propane 1.00n-hexane 0.83Iso-octane 0.97Ethine 1.28Propene 0.97Methanol 0.87Ethanol 0.84Acetic Acid 1.08Methyl-acetate 0.58Benzene 0.93Ethyl-benzene 1.00p-Xylene 1.02Di-chlorine-methane 0.87Tri-chlorine-ethene 1.08Terta-chlorine-ethene 1.19Chloroform 0.75Chlor-benzene 1.09n-heptane 1.03Cyclo-hexane 0.962-Propanol 0.76Toluene 1.01Acetone 0.77Ethyl-acetate 0.76Iso-butyl-acetate 0.95

Page 51: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

PIN ASSIGNMENTS

21 - 190003806(1) NGA-TFID e 16.06.98

Rosemount Analytical

21. Pin Assignments

Use only from our factory optional delivered cables or equivalent shielded cables

to be in agreement with the CE - conformity.

The customer has to guarantee, that the shield is be connected bothsided.

Shield and connectors housing have to be connected conductive.

Sub.-min.-D-plugs/sockets have to be screwed to the analyzer.

21.1 230 or 115 Vac (Input and Output)

21.2 Pt 100 (external heater)

Fig. 21-1: Pin assignments 230/115 V ac Input/Output

Fig. 21-2: Pin assignments Pt-100 (external heater)

VOLTAGE INPUT

Input: Output (external heater):max. 1000 VA max. 4 Anom. 300 VA

L

N

PE

12

Page 52: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

PIN ASSIGNMENTS

21 - 2 90003806(1) NGA-TFID e 16.06.98

Rosemount Analytical

21.3 Option SIO (Standard Input/Output Module)

21.3.1 Analog Signal Outputs

1 ± 11 V DC (burden > 2 kΩ), Output 12 ± 22 mA, (burden 500 Ω), Output 13 ± 11 V DC (burden > 2 kΩ), Output 24 ± 22 mA, (burden 500 Ω), Output 25 ⊥⊥⊥⊥⊥ (V DC), Outputs 1 + 26 ⊥⊥⊥⊥⊥ (mA), Outputs 1 + 27 ± 11 V DC (burden > 2 kΩ), Output 3 (Option)8 ± 22 mA, (burden 500 Ω), Output 3 (Option)9 ± 11 V DC (burden > 2 kΩ), Output 4 (Option)10 ± 22 mA, (burden 500 Ω), Output 4 (Option)11 ⊥⊥⊥⊥⊥ (V DC), Outputs 3 + 4 (Option)12 ⊥⊥⊥⊥⊥ (mA), Outputs 3 + 4 (Option)13 FE14 ± 11 V DC (burden > 2 kΩ), Output 5 (Option)15 ± 22 mA, (burden 500 Ω), Output 5 (Option)16 ± 11 V DC (burden > 2 kΩ), Output 6 (Option)17 ± 22 mA, (burden 500 Ω), Output 6 (Option)18 ⊥⊥⊥⊥⊥ (V DC), Outputs 5 + 6 (Option)19 ⊥⊥⊥⊥⊥ (mA), Outputs 5 + 6 (Option)20 ± 11 V DC (burden > 2 kΩ), Output 7 (Option)21 ± 22 mA, (burden 500 Ω), Output 7 (Option)22 ± 11 V DC (burden > 2 kΩ), Output 8 (Option)23 ± 22 mA, (burden 500 Ω), Output 8 (Option)24 ⊥⊥⊥⊥⊥ (V DC), Outputs 7 + 8 (Option)25 ⊥⊥⊥⊥⊥ (mA), Outputs 7 + 8 (Option)

21.3.2 Relay Outputs / Serial Interfaces

Fig. 21-3: Pin assignments Socket Analog Signal Outputs (Option SIO)

Fig. 21-4a: Pin assignments Socket Relay Outputs / RS 232 Serial Interface (Option SIO)

Fig. 21-4b: Pin assignments Socket Relay Outputs / RS 485 Serial Interface (Option SIO)

6

9

1

5

6

9

1

5

14

25

1

13

OPTION SIO

1 ⊥⊥⊥⊥⊥ (GND)2 RxD3 TxD4 not used5 ⊥⊥⊥⊥⊥ (GND)6 Relay Contact 1 (max. 30 V / 1 A / 30 W)7 Relay Contact 2 (max. 30 V / 1 A / 30 W)8 Relay Contact 3 (max. 30 V / 1 A / 30 W)9 Relay Contacts (Common)

1 ⊥⊥⊥⊥⊥ (GND)2 RxD-3 RxD+4 TxD+5 TxD-6 Relay Contact 1 (max. 30 V / 1 A / 30 W)7 Relay Contact 2 (max. 30 V / 1 A / 30 W)8 Relay Contact 3 (max. 30 V / 1 A / 30 W)9 Relay Contacts (Common)

Page 53: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

PIN ASSIGNMENTS

21 - 390003806(1) NGA-TFID e 16.06.98

Rosemount Analytical

21.4 Option DIO (Digital Input/Output Module)

1 (5 - max. 30 VDC / max. 500 mA) Output 12 (5 - max. 30 VDC / max. 500 mA) Output 23 (5 - max. 30 VDC / max. 500 mA) Output 34 (5 - max. 30 VDC / max. 500 mA) Output 45 (5 - max. 30 VDC / max. 500 mA) Output 56 (5 - max. 30 VDC / max. 500 mA) Output 67 (5 - max. 30 VDC / max. 500 mA) Output 78 (5 - max. 30 VDC / max. 500 mA) Output 89 (5 - max. 30 VDC / max. 500 mA) Output 910 (5 - max. 30 VDC / max. 500 mA) Output 1011 (5 - max. 30 VDC / max. 500 mA) Output 1112 (5 - max. 30 VDC / max. 500 mA) Output 1213 (5 - max. 30 VDC / max. 500 mA) Output 1314 (5 - max. 30 VDC / max. 500 mA) Output 1415 (5 - max. 30 VDC / max. 500 mA) Output 1516 (5 - max. 30 VDC / max. 500 mA) Output 1617 (5 - max. 30 VDC / max. 500 mA) Output 1718 (5 - max. 30 VDC / max. 500 mA) Output 1819 (5 - max. 30 VDC / max. 500 mA) Output 1920 (5 - max. 30 VDC / max. 500 mA) Output 2021 (5 - max. 30 VDC / max. 500 mA) Output 2122 (5 - max. 30 VDC / max. 500 mA) Output 2223 (5 - max. 30 VDC / max. 500 mA) Output 2324 (5 - max. 30 VDC / max. 500 mA) Output 2425 (5 - max. 30 VDC / 2,2 mA) Input 126 (5 - max. 30 VDC / 2,2 mA) Input 227 (5 - max. 30 VDC / 2,2 mA) Input 328 (5 - max. 30 VDC / 2,2 mA) Input 429 (5 - max. 30 VDC / 2,2 mA) Input 530 (5 - max. 30 VDC / 2,2 mA) Input 631 (5 - max. 30 VDC / 2,2 mA) Input 732 (5 - max. 30 VDC / 2,2 mA) Input 833 ⊥⊥⊥⊥⊥ (Outputs 1... 8 und Inputs)34 ⊥⊥⊥⊥⊥ (Outputs 9 ... 15)35 ⊥⊥⊥⊥⊥ (Outputs 16 ... 24)36 Summary Failure: Short circuit37 Failure: Timeout or Reset

201

3719

Fig. 21-5: Pin assignments Socket Digital Inputs/Outputs (Option DIO)

OPTION DIO

Page 54: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

PIN ASSIGNMENTS

21 - 4 90003806(1) NGA-TFID e 16.06.98

Rosemount Analytical

21.4.1 Wiring with external actors

To connect external actors, observe following rules.

Problems f or switc hing off inductive loads:

Switching off an inductive consumer (relay, solenoid valve, etc.) can generate high voltage (up toseveral 100 Volt) at the connections of the coil. This voltage impuls circulates via connected linesand could impair the function of near by equipments. Additional the outputs of DIO PCB couldbe become wrecked.

Fig. 21-6: Option DIO, Problems for switching off inductive loads

U [V]

t

OPTION DIO

Page 55: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

PIN ASSIGNMENTS

21 - 590003806(1) NGA-TFID e 16.06.98

Rosemount Analytical

Solution of pr oblem:

Parallel connection of a silicium diode to the connections of inductive loading can short circuit thisvoltage impuls to itself. This prevents circulation of voltage impuls via connected lines.

The diode should have following features:Blocking voltage > 50 VForeward current > 1 ADiode type 1N4000 (e.g.) fits this requirements.

Connect cathode of diode to postive connection of consumer (see Fig. 21-7). Connect anode ofdiode to negative connection of consumer.

Fig. 21-7: Option DIO, Solution of problem

U [V]

t

OPTION DIO

Page 56: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

PIN ASSIGNMENTS

21 - 6 90003806(1) NGA-TFID e 16.06.98

Rosemount Analytical

21.5 COM - Ports (for service use only)

Fig. 21-8: Pin assignments Socket COM PortsCOM 1 / RS 232 Serial Interface (Data Transfer)

COM 2 / RS 232 Serial Interface (external Data Input)

6

9

1

5

1 DCD2 RXD3 TXD4 DTR5 ⊥⊥⊥⊥⊥ (GND)6 DSR7 RTS8 CTS9 RI

COM-PORTS (FOR SERVICE ONLY)

Page 57: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

INDEX

90003806(1) NGA-TFID e 16.06.98 I - 1

Rosemount Analytical

Index

G

Gas Conditioning (Sample Handling) 5 - 3

Safety Measures S - 3

Gas Connections 5 - 5, 20 - 1

Gas Flow 20 - 3

Gases 5 - 4

Gas Connections 5 - 5, 20 - 1

Gas Flow 20 - 3

Permissible Pressure 20 - 1

Preparation 5 - 4

Supply Gases 5 - 4

Burner Air 5 - 4

Burner Gas (Fuel Gas) 5 - 4

Compressed Air Supply 5 - 4

Fuel Gas (Burner Gas) 5 - 4

Gas Flow 20 - 3

Permissible Pressure 20 - 1

Span Gas 5 - 4

Zero Gas 5 - 4

General Specifications 20 - 3

H

Housing

Front view 1 - 1, 1 - 2

Operation front panel 1 - 2

Rear panel 1 - 3

Housing Dimensions 20 - 2

I

Installation Site 5 - 2, 20 - 2

Safety Measures S - 1

Internal Construction 1 - 4

Printed Circuit Boards 1 - 6

K

Keys

Cursor Keys 1 - 1

ENTER Keys 1 - 1

Soft Keys 1 - 1

L

LC display 1 - 1

A

Additional notes for service / maintenance S - 6

Electrostatic Discharge S - 7

Ambiente Conditions 20 - 2

Analyzer specific notes for the user S - 5

Applications 1 - 8

B

Burner Air 5 - 4

Burner Gas (Fuel Gas) 5 - 4

C

Certifications 20 - 1

Cleaning of housing outside 19 - 1

COM - Ports (option for service use only) 21 - 6

Pin Assignments 21 - 6

Compressed Air Supply 5 - 4

Cursor Keys 1 - 1

D

Display / Keys

Cursor Keys 1 - 1

ENTER Key 1 - 1

LC display 1 - 1

Soft Keys 1 - 1

E

Electrical Safety 20 - 4

Safety Class 20 - 4

Electrostatic Discharge S - 7

ENTER Key 1 - 1

F

Front view 1 - 1, 1 - 2

Operation front panel 1 - 2

Fuel Gas (Burner Gas) 5 - 4

Page 58: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

INDEX

90003806(1) NGA-TFID e 16.06.98I - 2

Rosemount Analytical

M

Maintenance

Cleaning of housing outside 19 - 1

Opening the Housing 15 - 1

Front Panel 15 - 1

Housing Cover 15 - 1

Safety Measures S - 1

Additional notes for service / maintenance S - 6

Analyzer specific notes for the user S - 5

Electrostatic Discharge S - 7

Gases and Gas Conditioning (Sample Handling) S - 3

General S - 2

Supply Voltage S - 4

Measurement

Gas Conditioning (Sample Handling) 5 - 3

Preparation 5 - 1

Gas Conditioning (Sample Handling) 5 - 3

Gases 5 - 4

Installation Site 5 - 2

Measuring Principle 2 - 1

Mode of Operation 2 - 3

O

Opening the Housing 15 - 1

Front Panel 15 - 1

Housing Cover 15 - 1

Operating voltage 6 - 2

Operation

Keys

Cursor Keys 1 - 1

ENTER Key 1 - 1

Soft Keys 1 - 1

Preparation 5 - 1

Gas Conditioning (Sample Handling) 5 - 3

Gas Flow 20 - 3

Gases 5 - 4

Installation Site 5 - 2

Permissible Gas Pressures 20 - 1

Safety Measures S - 1

Switching Off 7 - 1

Switching On 6 - 1

Safety Measures S - 1

Operation front panel 1 - 2

Options

COM - Ports (for service use only)

Pin Assignments 21 - 6

DIO (Digital I/O)

Pin Assignments 21 - 3

Wiring 21 - 4

Pt 100 (external heater)

Pin Assignments 21 - 1

SIO (standard I/O)

Pin Assignments 21 - 2

P

Permissible Ambient Temperature 5 - 2, 20 - 2

Pin Assignments 21 - 1

230 or 115 Vac (Input and Output) 21 - 1

COM - Ports (option for service use only) 21 - 6

Option DIO (Digital I/O) 21 - 3

Wiring 21 - 4

Option SIO (standard I/O) 21 - 2

Analog Signal Outputs 21 - 2

Relay Outputs / Serial Interfaces 21 - 2

Pt 100 (external heater) 21 - 1

Preface P - 1

Technical Properties of Measurement P - 1

Preparation 5 - 1

Gas Conditioning (Sample Handling) 5 - 3

Gas Connections 5 - 5

Gases 5 - 4

Gas Connections 5 - 5

Gas Flow 20 - 3

Permissible Pressure 20 - 1

Supply Gases 5 - 4

Installation Site 5 - 2

Safety Measures S - 1

Additional notes for service / maintenance S - 6

Analyzer specific notes for the user S - 5

Gases and Gas Conditioning (Sample Handling) S - 3

General S - 2

Supply Voltage S - 4

Supply Gases 5 - 4

Burner Air 5 - 4

Burner Gas (Fuel Gas) 5 - 4

Compressed Air Supply 5 - 4

Fuel Gas (Burner Gas) 5 - 4

Gas Flow 20 - 3

Permissible Pressure 20 - 1

Span Gas 5 - 4

Zero Gas 5 - 4

Transfer safety lock 5 - 1

Printed Circuit Boards 1 - 6

Protection class 20 - 1

Pt 100 (external heater) 21 - 1

R

Rear Panel 1 - 3

Response time (t90) 1 - 8, 20 - 3

Page 59: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

INDEX

90003806(1) NGA-TFID e 16.06.98 I - 3

Rosemount Analytical

S

Safety Class 20 - 4

Safety Measures S - 1

Additional notes for service / maintenance S - 6

Electrostatic Discharge S - 7

Analyzer specific notes for the user S - 5

Electrostatic Discharge S - 7

Gases and Gas Conditioning (Sample Handling) S - 3

General S - 2

Supply Voltage S - 4

Sample Gas Transport 2 - 4

Soft Keys 1 - 1

Span Gas 5 - 4

Specifications 20 - 3

Suitability tests 20 - 1

Supply Gases 5 - 4

Burner Air 5 - 4

Burner Gas (Fuel Gas) 5 - 4

Compressed Air Supply 5 - 4

Fuel Gas (Burner Gas) 5 - 4

Gas Flow 20 - 3

Permissible Pressure 20 - 1

Span Gas 5 - 4

Zero Gas 5 - 4

Switching Off 7 - 1

Switching On 6 - 1

Automatic Internal Calibration 6 - 4

Operating voltage 6 - 2

Pin Assignments 21 - 1

230 or 115 Vac (Input and Output) 21 - 1

Pneumatic operations 6 - 3

Safety Measures S - 1

Additional notes for service / maintenance S - 6

Analyzer specific notes for the user S - 5

Gases and Gas Conditioning (Sample Handling) S - 3

General S - 2

Supply Voltage S - 4

Voltage Supply 20 - 3

T

t90 (response) time 1 - 8, 20 - 3

Technical Data 20 - 1

Ambiente Conditions 20 - 2

Certifications 20 - 1

Electrical Safety 20 - 4

Safety Class 20 - 4

Gas connections 20 - 1

General Specifications

Response time (t90) 1 - 8, 20 - 3

t90 (response) time 1 - 8, 20 - 3

Housing Dimensions 20 - 2

Protection class 20 - 1

Response time (t90) 20 - 3

Specifications 20 - 3

Suitability tests 20 - 1

t90 (response) time 20 - 3

Voltage supply 20 - 3

Weight 20 - 1

Technical description 1 - 1

Applications 1 - 8

Front view 1 - 1

Internal construction 1 - 4

Operation front panel 1 - 2

Printed Circuit Boards 1 - 6

Rear view 1 - 3

Theory of Technology and Operation 2 - 1

Measuring Principle 2 - 1

Mode of Operation 2 - 3

Sample Gas Transport 2 - 4

Transfer safety lock 5 - 1

Troubleshooting

Opening the Housing 15 - 1

Front Panel 15 - 1

Housing Cover 15 - 1

Safety Measures S - 1

Additional notes for service / maintenance S - 6

Analyzer specific notes for the user S - 5

Electrostatic Discharge S - 7

Gases and Gas Conditioning (Sample Handling) S - 3

General S - 2

Supply Voltage S - 4

V

Voltage Supply 6 - 2, 20 - 3

Pin Assignments

230 or 115 Vac (Input and Output) 21 - 1

W

Weight 20 - 1

Z

Zero Gas 5 - 4

Page 60: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

INDEX

90003806(1) NGA-TFID e 16.06.98I - 4

Rosemount AnalyticalLIST OF FIGURES

List of Figures

Fig. Title Page

Fig. 1-1: Operation front panel, Front view 1 - 2Fig. 1-2: TFID, Front view 1 - 2Fig. 1-3: TFID, Rear view 1 - 3Fig. 1-4: TFID, internal construction 1 - 4Fig. 1-5: Cardcage TFID, Top View 1 - 6

Fig. 2-1: Measuring Principle for TFID Measurement 2 - 2Fig. 2-2: TFID, Flow / Function Schemstic 2 - 5

Fig. 5-1: Transfer safety locke (housing side view, detail sketch) 5 - 1Fig. 5-2: TFID, gas connections (drawing) 5 - 6Fig. 5-3: TFID, gas connections (photo 5 - 6

Fig. 6-1: TFID, Voltage supply 6 - 1

Fig. 15-1: TFID (Fastening screws front panels) 15 - 1

Fig. 20-1: Dimensional sketch TFID [all dimensions in mm] 20 - 2

Fig. 21-1: Pin assignments 230 or 115 Vac (Input and Output) 21 - 1Fig. 21-2: Pin assignments Pt-100 (external heater) 21 - 1Fig. 21-3: Pin assignments Socket Analog Signal Outputs (Option SIO) 21 - 2

Fig. 21-4a: Pin assignments Socket Relay Outputs / RS 232 Serial Interface (Option SIO) 21 - 2Fig. 21-4b: Pin assignments Socket Relay Outputs / RS 485 Serial Interface (Option SIO) 21 - 2Fig. 21-5: Pin assignments Socket Digital Inputs/Outputs (Option DIO) 21 - 3Fig. 21-6: Option DIO, Problems for switching off inductive loads 21 - 4Fig. 21-7: Option DIO, Solution of problem 21 - 5Fig. 21-8: Pin assignments Socket COM Ports 21 - 6

Page 61: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

NGA 2000 TFID HardwareInstruction Manual

9000380604/2008

APPENDIX Declaration of Conformity

Page 62: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected
Page 63: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

NGA 2000 TFID HardwareInstruction Manual

9000380604/2008

Page 64: Operation Manual NGA 2000 - TFID - Emerson · F or outdoor site s, we recommend to install the inst rument in a protecti v e cabinet. At least, the instrument has to be protected

Instruction Manual9000380604/2008 NGA 2000 TFID Hardware

© Emerson Process Management GmbH & Co. OHG 2007

WORLD HEADQUARTERSROSEMOUNT ANALYTICAL EUROPEEmerson Process ManagementGmbH & Co. OHGIndustriestrasse 163594 HasselrothGermanyT 49 6055 884 0F 49 6055 884209

Emerson Process ManagementRosemount Analytical Inc.6565 P Davis Industrial ParkwaySolon, OH 44139 USAT 440.914.1261Toll Free in US and Canada 800.433.6076F 440.914.1271e-mail: [email protected]

www.raihome.com

GAS CHROMATOGRAPHY CENTERAND LATIN AMERICAEmerson Process ManagementRosemount Analytical Inc.11100 Brittmoore Park Drive

Houston, TX 77041

T 713 467 6000

F 713 827 3329

EUROPE, MIDDLE EAST AND AFRICAEmerson Process ManagementShared Services LimitedHeath PlaceBognor RegisWest Sussex PO22 9SHEnglandT 44 1243 863121F 44 1243 845354

ASIA-PACIFICEmerson Process ManagementAsia Pacific Private Limited1 Pandan CrescentSingapore 128461Republic of SingaporeT 65 6 777 8211F 65 6 777 0947e-mail: [email protected]