Compact Urine Analyzer
Operating Manual
1 AUTION MICRO
This manual contains important information on thefunctions of the AUTION MICRO PU-4210.
Supplied by : ARKRAY, Inc.distributed by : A. Menarini I.F.R. srl
Diagnostic Division
This manual has been issued by: ARKRAY, Inc.Read carefully prior to starting up the instrument.It is recommended to retain this manual for futureuse.
This product conforms to European Directive98/79/EC.
PremiseAUTION MICRO PU-4210 OPERATING MANUALDOC. No. OM-28373 REV.5 06/05
e
IntroductionThis manual contains outlines, instructions for the
operation, maintenance and troubleshooting for the
AUTION MICRO PU-4210. Read this manual carefully
before use the Analyzer in order to operate it correctly.
After you read the manual, keep it near the Analyzer to
refer to it when necessary.
● This instrument uses urine as a sample. Urine may becontaminated by pathogenic microbes that causeinfectious diseases. Always be careful when handlingurine samples. If operated incorrectly, there is a dangerof infection to you and those around you by pathogenicmicrobes.
● Never touch places where samples may be contactedwith bare hands. When you perform maintenance,always wear protective gloves to prevent exposure topathogenic microbes.
● Separate used Test Strips, etc. from general wasteand dispose of them according to local regulations onbiohazardous waste.
● This instrument may becomeinfectious in the course of use.Discard the product in accordancewith local regulations for biohazardouswaste.
AUTION MICRO 2
AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.5 06/05
e
3 AUTION MICRO
Pay careful attention to these caution marks in this Operating Manual.
Caution Marks
To prevent infection of operator or others from pathogenic
microbes, follow the instructions described.
To prevent injuries to operator or others, or material
damages, follow the instructions described.
Failure to follow the instructions may lead to incorrect
results.
To prevent damage to the unit and assist you in
making best use of the capabilities of the Analyzer,
additional explanations and notes are provided herein.
Regarding accidents resulting injury.
Regarding damage and performance of Analyzer.
Reference information on operation, additional
explanations and related functions are provided herein.
Important
Note
Reference
AUTION MICRO PU-4210 OPERATING MANUALDOC. No. OM-28373 REV.5 06/05
e
AUTION MICRO 4
■For installation, read thefollowing precautions carefully
and always follow safety
measures.
●Place the Analyzer on a stable and levelsurface, free of vibration.
●Select a room in which to set up theAnalyzer where the temperature can becontrolled between 10℃ and 30℃, andhumidity in a range of 20 to 80 %.
●Do NOT place the Analyzer where it maybe affected by magnetic materials includingmagnets or speakers.
●Do NOT place the Analyzer where it maybe affected by chemicals, corrosive gasesand electronic noise.
●Do NOT place the Analyzer where it maybe affected by direct sunlight or wind.
Precautions for Installation AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01
AUTION MICRO PU-4210 OPERATING MANUALDOC. No. OM-28373 REV.0 04/01
5 AUTION MICRO
●When the instrument is not in use, unplugthe AC adapter.
●If an abnormal sound, odor or smoke isfound, turn the power off immediately andconsult your distributor after making surethat the abnormal situation is not continuing.
●Do NOT drop or hit the Analyzer.
●Should the Analyzer stop functioning,contact your distributor for repairs. Do NOTattempt to repair it your self. Doing so mayresult in damaging the Analyzer or injury.
●Do NOT place the Analyzer where it maybe affected by water. Samples may causedamage to the Analyzer.
●Plug in an exclusive AC adapter.
Precautions for Installation
AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01
AUTION MICRO 6
AUTION MICRO(PU-4210)Operating Manual
Premise
Introduction
Caution Marks
Precautions for Installation
Table of Contents
Outline of AUTION MICRO
Shipping Carton
Parts Description and Functions
Display Screen and Key Arrangement
Flow of Measurement
Cautions
Start Up
Setting Up Parameters
Normal Measurement
Check Measurement
Auxiliary Function
Special Parameters
Printer
Printing Item
Maintenance
Troubleshooting
…………………………………………………………1
……………………………………………………2
…………………………………………………3
…………………………………4
……………………………………………6
…………………………………7
………………………………………………8
……………………………9
……………………10
………………………………………11
………………………………………………………12
………………………………………………………14
……………………………………15
………………………………………16
………………………………………19
……………………………………………21
…………………………………………26
…………………………………………………………27
…………………………………………………28
…………………………………………………29
………………………………………………30
Introduction
Measurement
Printing
Maintenance and
Troubleshooting
Table of Contents
AUTION MICRO PU-4210 OPERATING MANUALDOC. No. OM-28373 REV.0 04/01
7 AUTION MICRO
The AUTION MICRO (PU-4210) is a small portable urine-analyzing device
which anyone can use anytime, anywhere. Since the Printer is removed
from the Analyzer, you can operate it by a minimal key operation.
The device is equipped with transmission capability to adapt to
systematization.
Characteristics
Minimum MaintenanceThe only necessary maintenance item is cleaning the Strip Holder.
Long Life Light SourceUse of LED and short-time lighting enables a longer light source life.
Temperature Correction FunctionAn automatic correction function, which is provided by the internal thermalsensor, solves any errors caused by temperature changes in the environment.
Power SourceUse a dry battery or a AC adapter when using the Analyzer alone, and use theAC adapter when using it with a Printer.
Regent SelectionAUTION Sticks 10EA and Uriflet 9UB are available to use.
100 memoriesMaximum 100 measurement results can be stored.When more than 100 samples have been measured, older data will be automaticallydeleted in order.
Outline of PocketChem UA Introduction
AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01
AUTION MICRO 8
Analyzer
OptionalAccessories
OptionalConsumables
RS-232C Cable Thermal Printer Paper Check Strip Set
Strip Holder
Printer
Shipping carton of AUTION MICRO (PU-4210) contains the following.
Make sure that all of the following are included:
Shipping Carton Introduction
AC Adapter Check Strip Set Thermal Printer Paper Operating Manual(this booklet)
Compact Urine Analyzer
Operating Manual
AUTION MICRO PU-4210 OPERATING MANUALDOC. No. OM-28373 REV.0 04/01
9 AUTION MICRO
Front
Bottom (Analyzer) Rear (Analyzer)
8
21
3
74
5
9
6
Name
① Display Screen
② Operator Panel
③ Optical System
④ Power Input Terminal
⑤ Connection Terminal
⑥ DATA OUT Terminal
⑦ Battery Box
⑧ FEED key
⑨ Lock Release Button
Function
Displays measurement results and error codes
For setting conditions and starting measurement
Measures samples
For connection of exclusive AC adapter
For electrical connection to the Printer
Terminal for connection to a PC or other devices
Place batteries here
Feeds printer paper
Unlocks connection between the Analyzer and Printer
Parts Description and Functions Introduction
Parts description in this Operating Manual is as follows:
AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01
AUTION MICRO 10
Display Screen and Key Arrangement Introduction
4. Delete1. Memory2. ID
3. No.
11. Test Strip Icon
12. Waiting Icon 13. Transmission Icon
14. Printer Icon
15. Battery Icon
10. figure display13 digits
9. Item Name
7. Test Strip
6. Time
LCD
8. Semiquantitative Symbol
Memory
ID
No.
Delete
Date
Time
Test Strip
Semiquantitative Symbol
Item Name
Figure display 13 digits
Test Strip Icon
Waiting Icon
Transmission Icon
Printer Icon
Battery Icon
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
Displayed when the data is redisplayed.
Displayed (Blinks during ID setting) when ID is displayed.
Displayed (Blinks during No. setting) when No. is displayed.
Blinks when the memory is deleted. For deletion, enter [ENT].
Light is on (Blinks during Date setting) when the Date is displayed.
Light is on (Blinks during Time setting) when the Time is displayed.
Displayed when the Test Strip type is displayed. Blinks during Test Strip type setting.
Displayed according to measurement results when measurement results are displayed or redisplayed.
Displayed according to test items when measurement results are displayed or redisplayed.
Measurement No., ID, Date, Time, Test Strip name, etc. are displayed. (13 digits)
Blinks when the Test Strip is dipped or the Test Strip Holder is moved.
Displayed from the measurement start to finish.
Displayed for the external output.
Displayed when printing is performed.
Displayed in three different signs depending on the degree of exhaustion.
5. Date
1. Function Key or Numeric Key
2. Start Key, Enter Key or Power Switch
3. Stop Key or Hyphen Key
Key Arrangement Specification
AUTION MICRO PU-4210 OPERATING MANUALDOC. No. OM-28373 REV.5 06/05
11 AUTION MICRO
Flow chart of measurementThe following flow chart explains a measurement procedure.
Before measurement, understand the whole procedure.
Turn on the main power.
Start Measurement
Dispose of Test Strip.
Dip the Test Strip in the sample for 2 seconds in accordance with the alarm.
Remove the excess sample.
Set the Test Strip onto the Strip Holder.
The Strip Holder moves inside of the Analyzer and measurement starts.
Measurement Results are displayed and the Strip Holder comes out from the Analyzer.
Enter Measurement No. and Patient ID.Select Test Strip Type
Continue measurement
Finish measurement
Clean the Strip Holder, and turn off the main power.
Flow of Measurement Measurement
Wear protective gloves to prevent exposure to pathogenic
microbes.
Separate used test papers, etc. from general waste anddispose of them according to local regulations onbiohazardous waste.
e
AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01
AUTION MICRO 12
Wear protective gloves to prevent exposure to pathogenic microbes.
Use a fresh urine sample. If the sample urine is not measured within anhour after collection, cap it tightly and keep it refrigerated. Be sure toallow it to reach to room temperature before measurement.
Mix the sample thoroughly before measurement. Do NOT centrifuge them.
Prepare an adequate amount of samples so that the whole pad portion ofthe Test Strip can be immersed.
Do NOT add any preservative, disinfectant, or detergent to the sample.
Keep the sample out of direct sunlight.
If samples containing ascorbic acid are measured, the measured valuesof glucose and occult blood are occasionally lower than the real values.
If drug-administered urine or visually observed hematuria are measured,correct measurement results sometimes cannot be obtained.
●
●
●
●
●
●
●
●
Handling Samples
Cautions Measurement
AUTION MICRO PU-4210 OPERATING MANUALDOC. No. OM-28373 REV.0 04/01
Do NOT use Test Strips of other manufacturers.
Use only our Test Strips,“AUTION Sticks, Uriflet”.Read the attached manuals of the Test Strips.
Do NOT use expired Test Strips.
Do NOT use expired Test Strips and/or Test Strips if the reagent padsshow any signs of discoloration, deformation or deterioration.
Prepare Test Strips just before the measurement.
Take out only the necessary number of Test Strips to use and close thecap. If Test Strips are left out for a long time, they may absorb humidityor collect dust, causing incorrect measurement results.
Do NOT touch the reagent pad of the Test Strip.
Do NOT touch the reagent pad of the Test Strip, otherwise correctmeasurement results may not be obtained.
Set Test Strip type.
Set the Test Strip type before measurement. If different Test Strips areused, correct measurement results may not be obtained.
Keep the desiccant.
Keep the desiccant contained in the Test Strip container until all TestStrips are used.
●
●
●
●
●
●
13 AUTION MICRO
Handling Test Strips
Cautions Measurement
AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.5 06/05
AUTION MICRO 14
1. Preparation.When using the Printer, make sure that the
Printer is connected to the Analyzer correctly
and paper is loaded. If there is a red strip on
the both sides of the paper, replace the roll.
2. Turn on the power.Connect the exclusive AC adapter with the
power input terminal on the rear of the device
(while using the AC adapter, battery is not
consumed). When only the Analyzer is
operated with batteries, insert batteries into
the battery box at the bottom of the Analyzer.
Press [START-ENT] on the operator panel to
turn on the power. After initialization, the
Ready Screen appears.
There are two screens for the measurement-standby stage.(Ready Screen).
In either case, press [START] to start measurement.
Make sure that the Strip Holder is set up correctly.
Start Up Measurement
Do NOT touch the optical system.
Doing so may result in injury.
①
①
②
③
Ready Screen 1The latest measurement
result is displayed on
the LCD.
Ready Screen 2Parameters for the next
measurement are
displayed.
(Measurement No., ID,
Test Strips, Measurement
Item for Test Strip)
e
AUTION MICRO PU-4210 OPERATING MANUALDOC. No. OM-28373 REV.0 04/01
1. Setting up the measurement number.Press [No.] from the Ready Screen.
Measurement No. setting mode appears and“No.”and
the cursor blinks.
Enter any numbers using the numeric keys.(Max. 4 digits)
Press [ENT] to confirm the entry. Press [STOP] to cancel,
and the Ready Screen is restored.
Measurement No. can be set arbitrarily.Once the Measurement No. is set, the number increaseswith every measurement.
2. Setting up of Patient ID.Enter [ID] from the Ready Screen. The patient ID setting
mode appears and“ID”and the cursor blinks.
Enter any numbers using the numeric keys.
(Max. 13 digits including hyphens)
Press [ENT] to confirm the entry. The Ready Screen is
restored.
1: [STOP] functions as a hyphen key.To cancel the entry, press [ENT] once before re-entry.
2: When the patient ID is entered, the Ready or Measurementscreen displays Measurement No. and Patient IDalternately.
3: Patient ID is included in the printing results and externaltransmission. (When not entered, -----------is displayed.)
3. Setting of Test Strip Type.Press [STRIP] from the Ready Screen, Test Strip type
selection mode appears and“STRIP”blinks, and theTest Strip type and Item are displayed.
To change the Test Strip type and Item, press [STRIP].
(10EA→9Ub(9UB)→C(CHECK)→10EA)Press [ENT] to confirm the entry.
Press [STOP] to cancel, and the Ready Screen is restored.
15 AUTION MICRO
Setting Up Parameters Measurement
①
②
③
Reference
①
②
③
Reference
①
②
③
AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.5 06/05
AUTION MICRO 16
1. Prepare sample.Prepare sample in a sample cup.
Make sure that sample isenough to cover the whole reagent pads.
Stir sample well.
Do NOT use a centrifuge for mixing.
Prepare tissue paper to wipe off excess urine.
2. Prepare the Test Strip.Take out only the necessary number of strips
from the Test Strip container.
Close the cap.
Read the attached manualsfor the Test Strip.
Reagent Pads
Close the lid tightly
Normal Measurement Measurement
Wear protective gloves to prevent exposure to pathogenic
microbes.
Important
①
②
③
Important
①
②
Important
e
AUTION MICRO PU-4210 OPERATING MANUALDOC. No. OM-28373 REV.0 04/01
3. Press [START] to start Normal Measurement.Test Strip Icon blinks and the timer starts to count
down the timing of dipping of the Test Strip from [3
sec]. (A timing buzzer also sounds in accordance with
the count down.)
To stop the measurement,keep pressing [STOP] for 2-3 seconds during measurement. Do NOT move the Analyzer duringmeasurement. If impact or shock is given to the Analyzer,measurement cannot be done correctly. The Test Stripmay become jammed in the Analyzer.
4. Dip a Test Strip in the sample.Following the indicator timing (--sec), a timing buzzer
sounds for 2 seconds.
Keep dipping the Test Strip while the buzzer is
sounding.
Dip the Test Strip properly for 2seconds following the sound of thetiming buzzer. Shorter dipping timemay result in insufficient coloringand longer dipping time may causedischarging of the reagent.
5. Remove the excess sample.When the timing buzzer stops, pull out the Test
Strip and remove the excess sample with tissue
paper.
17 AUTION MICRO
Do NOT dip the black marker
Normal Measurement Measurement
Reference
Important
Important
AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01
AUTION MICRO 18
6. Set the Test Strip.While the Test Strip Icon is blinking, place the Test
Strip onto the Strip Holder.
A few seconds before the Strip Holder moves inside of the Analyzer, the Test Strip Iconstops blinking. When the StripHolder is set inside the Analyzer,the Test Strip Icon disappears.If the Test Strip is not correctlyset on the Strip Holder, correct measurement results cannot beobtained.
7. Measurement.The Strip Holder is pulled into the Analyzer
and the measurement starts.
8. Finish Measurement.When measurement is completed,the
measurement result is displayed and the
Strip Holder comes out from the Analyzer.
Discard the used Test Strip, and when
finishing the whole measurement process,
remove the Strip Holder from the Analyzer
and clean it.
9. Turn off the power.Keep pressing [START] for 2-3 seconds to turn off the power (the Strip Holder is
pulled into the Analyzer automatically).
When no key entry is made for more than 3 minutes, the power automatically turns off.
Normal Measurement Measurement
Reference
Important
Make sure that the tip of Test Strip
reaches to the left end of the holder.
Reference
AUTION MICRO PU-4210 OPERATING MANUALDOC. No. OM-28373 REV.5 06/05
For installation, read the following precautions carefully and always
follow safety measures.
1. Clean the Strip Holder.Wash the Strip Holder and dry it completely with
the soft cloth. See“Maintenance”.
Make sure the holder is clean,otherwise dirt may adhere tothe Check Strip and damageit.
2. Set the Test Strip Type.Set the Test Strip Type to C (Check).
See“Setting Up Parameters”.
3. Prepare a Check Strip.Remove 1 Check Strip from the Check Strip
container.
Do NOT touch the gray-colored part of Check Strip.
19 AUTION MICRO
Check Strip(Gray)
Semiquantitative Value Label
Check Measurement Measurement
Wear protective gloves to prevent exposure to pathogenic
microbes.
Note
Important
e
AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01
AUTION MICRO 20
4. Press [START] to start Check
Measurement.While the Test Strip Icon is blinking, place the
Check Strip onto the Strip Holder.
A few seconds before the StripHolder moves inside of the Analyzer, the Test Strip Icon stops blinking.When the Strip Holder is setinside the Analyzer, the Test Strip Icon disappears.
5. Confirm the results.Compare the Check Measurement results with the ranges of semiquantitative
values attached to the Check Strips container.
●When the Check Measurement results are within the range:The Analyzer is performing correctly. End Check Measurement.
●When the Check Measurement results are out of the range:Repeat the Check Measurement with the other Check Strip.
6. Repeat the Check Measurement with
the other Check Strip.
●If the second Check Measurement results are within the range:The Check Strip used for the first measurement was defective.
Do NOT use it again.
●If the second Check Measurement results are still out of the range:Contact your distributor.
Check Measurement Measurement
Reference
AUTION MICRO PU-4210 OPERATING MANUALDOC. No. OM-28373 REV.0 04/01
1. Calendar Display and Setting.Press [DATE] from the Ready Screen,the current
date is displayed. (The Ready Screen is restored
in 2-3 seconds automatically.)
Press [DATE] again while the date is displayed,
the Date Setting Screen appears and“DATE”and cursor blink.
Enter the current date.
Press [ENT] to confirm the entry. Press [STOP]
to cancel, and the Ready Screen is restored.
2. Time Display and Setting.Press [TIME] from the Ready Screen, current
time is displayed.
(The Ready Screen is restored in 2-3 seconds
automatically).
Press [TIME] again while the time is displayed,
the Time Setting Screen appears and [TIME]
and cursor blink.
Enter the current time.
Press [ENT] to confirm the entry.
Press [STOP] to cancel, and the Ready Screen
is restored.
21 AUTION MICRO
Auxiliary Function Measurement
①
②
③
④
①
②
③
④
AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01
AUTION MICRO 22
3. Redisplay of Measurement
Results.
Press [MEMORY▲] or [MEMORY▼] from theReady screen, the latest data is redisplayed
([MEMORY] is displayed at the top left of the
screen).
Press [MEMORY▼] again, one previous data isdisplayed.
Repeat this procedure, if necessary.
Press [MEMORY▲], the next data is displayed.
Press [STOP], the Ready Screen is restored.
If No., ID, DATE and TIME are pressed during Redisplay ofdata, Measurement No.,Patient ID, Measurement Dateand Measurement Time on theRedisplayed data are displayed.(Measurement No. is displayedas default. This setting isreflected in all data during aseries of redisplay operation.)
Auxiliary Function Measurement
Reference
①
②
③
④
AUTION MICRO PU-4210 OPERATING MANUALDOC. No. OM-28373 REV.0 04/01
4. Delete Data.A: Delete Single Data.
Press [DELETE] on the Redisplay screen of
measurement results,“DELETE”blinks.(The measurement results are also displayed.)
Press [ENT] to delete the displayed measurement
data and the next data is displayed.
To cancel it, press [STOP] and the Redisplay
screen is restored.
B: Delete All Data.Keep pressing [DELETE] on the Redisplay screen
of the measurement results for 2-3 seconds,
“DELETE”blinks and“ALL”is displayed.(Other displayed items disappear.)
Press [ENT] to delete all measurement data in the
memory.
To cancel it, press [STOP] and the Redisplay
screen is restored.
5. Retransmit.A: Retransmit Single Data.
Press [SEND] on the Redisplay screen of the
measurement results, the Transmission Icon blinks.
(The measurement Results are also displayed.)
Press [ENT] to transmit the displayed measurement
data. To cancel it, press [STOP] and the Redisplay
screen is restored.
23 AUTION MICRO
Auxiliary Function Measurement
①
②
①
②
①
②
AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01
AUTION MICRO 24
B: Retransmit All Data.Keep pressing [SEND] on the Redisplay screen of
the Measurement Results for 2-3 seconds, the
Transmission Icon blinks and“ALL”is displayed.
(Other displayed items disappear.)
Press [ENT] to transmit all measurement data in
the memory from the latest data.
To cancel it, press [STOP] and the Redisplay screen
is restored.
To stop transmission, press [STOP] during transmission.
6. Reprint.
(when the Printer is connected.)A: Reprint Single Data.
Press [PRINT] on the Redisplay screen of the
measurement results, the Printer Icon blinks.
(The measurement results are also displayed.)
Press [ENT] to print the displayed measurement
data. To cancel it, press [STOP] and the Redisplay
screen is restored.
B: Reprint All Data.Keep pressing [PRINT] on the Redisplay screen of
the measurement results for 2-3 seconds, the
Printer Icon blinks and [All] is displayed.
(Other displayed items disappear.)
Auxiliary Function Measurement
Reference
①
②
①
①
②
AUTION MICRO PU-4210 OPERATING MANUALDOC. No. OM-28373 REV.01 01/01
25 AUTION MICRO
Press [ENT] to print all measurement data in
memory from the latest data. To cancel it, press
[STOP] and the Redisplay screen is restored.
To stop printing, press [STOP] during printing.To print the latest data,press[PRINT] on the Ready Screen.
7. Battery Icon.A battery icon at the lower right of the analyzer
display tells you the timing of battery exchange.
Follow the instructions below, and replace the
battery to a new one if necessary.
The supplied AC adapter can be used as well.
The battery is full with power.
Although measurement can be continued
with this battery, the power is getting
short. Prepare a new battery.
This icon sometimeschanges to the next icon(blinking) even if enoughpower remains in.
The battery runs out.
To continue measurement, replace the
battery to a new one.
Auxiliary Function Measurement
②
Reference
(Blinking)
①
② ③
④
①
②③
④
Note
AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01
The following special parameters are available to set with the PU-4210.
(The parameters are stored until the settings are changed.)
●Orientation of the Test Strip set.Setting of Test Strip from the right side or from the left side of the Analyzer
(default is set to the right side).
To change the Test Strip orientation, remove the StripHolder from the Analyzer beforehand.
●Date Order.Choose Date order among Y-M-D, M-D-Y, and D-M-Y.
(The default setting is Y-M-D).
Setting is reflected on the screen display and printing. (Fixed for external output.)
●Setting of measurement units and symbols for semi-quantitative ranks.Set the measurement units for the measurement results (SI or usual) and format
of semi-quantitative ranks ( +++ or +n ) ( default is set to usual, +++).
Setting is reflected on the printing and external output. (Fixed for the screen display.)
1. Operation. Turn on the power while pressing [STOP].
(Keep pressing [STOP] during initialization.)
After initialization, the entry screen for 3 digits
numbers is displayed and the parameter setting
mode appears.
The numeral of each digit corresponds to the
following. Input the numerals for the desired setting.
In this case, if numerals other than the designated
ones are input, they are not accepted.
Hundreds 1:Right setting 2:Left settingTens 1:Y-M-D 2:M-D-Y 3:D-M-Y
Ones 1:mg/dl,+++ 2:mg/dl,+n 3:mmol/l,+++ 4:mmol/l,+n
Press [ENT] after the entry to change the settings.
The Ready Screen is restored.
e.g.) Change from the default (Right setting, Y-M-D, mg/dl,+++)
●When“232”is entered,(Left setting, D-M-Y, mg/dl, +n) is set. Then the Ready Screen is restored.
●When“063”is entered,(Right setting, Y-M-D, mmol/l, +++) is set. Then the Ready Screen is restored.
AUTION MICRO 26
previous setting
Special Parameters Measurement
Note
①
②
③
④
AUTION MICRO PU-4210 OPERATING MANUALDOC. No. OM-28373 REV.0 04/01
27 AUTION MICRO
1. Attachment or Removal of
Printer.When connecting the Printer to the Analyzer,
insert the hook of the Printer in the hole at the
bottom of the Analyzer, and then slide it firmly
enough to achieve a secure fit.
When removing the Printer, slide the Analyzer
to the front while pressing the lock release button of the Printer.
Before you attach or remove the Printer, make sure to turnoff the power and pull out the AC adapter of the Printer.
2. Paper Jam.When paper is jammed, press [FEED] of the
Printer to remove the paper.
3. Loading of Paper.After confirming the face and the back of the
paper, put the thermal printer paper into the
insertion position, and then the paper will be
fed automatically.
When inserting a new thermal Printer paper,cut the edge with scissorsto make it even, and set it inthe position.
4. Error Comments.When an error occurs, error codes and error comments are printed.
(Only error codes are displayed on the screen.)
※See“Troubleshooting”.
lock release button
FEED
Printer Printing
Note
Note
AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01
AUTION MICRO 28
Abnormal Mark
Marks for abnormal values(Normal:Blank)
Measurement Item
Measurement No.
Patient IDDate・Time・ Type of Test Strip・Temperature in Instrument
Measurement Results
Ranks (Semi-quantitative symbols)
Concentration values
Measurement Units
Abnormal Mark
Measurement No.
Patient ID
Date, Time
Type of Strip
Temperature in
Instrument
Measurement
Results
When the measurement results is abnormal (Rank is more than“+1”.),
sample abnormal mark‘●’is displayed.
4 digits
ID is entered using keys. When there is no ID, [------------------]is printed.
There are 3 printing patterns of the date
YMD, MDY, DMY
Example 2001-07-05 09:53
07-05-2001 09:53
05-07-2001 09:53
The Test Strip Type is displayed.
Example 9UB
Temperature in the instrument is printed in ℃
Example 29℃
Abnormal Mark / When an abnormal measurement result (rank is over “+1”) is recognized,
‘*’ is printed before the name of item.
Measurement Item / GLU, URO, PRO, KET, BLD, BIL, NIT, LEU, pH, S.G.
Ranks (Semi-quantitative symbols), Concentration,
Measurement Units.
Printing Item Printing
AUTION MICRO PU-4210 OPERATING MANUALDOC. No. OM-28373 REV.5 06/05
Clean the Strip Holder after finishing measurement at the end of the day.
29 AUTION MICRO
1. Slide the Strip Holder to remove. 2. Wash the Strip Holder with
detergent or alcohol and then rinse
it with tap water. Especially, clean
the groove with special care.
3. Dry the Strip Holder completely
with a soft cloth.
4. Insert the Strip Holder completely
by using grooves of the backside of
the Strip Holder.
Maintenance Troubleshooting and Maintenance
Wear protective gloves to prevent exposure to pathogenic
microbes.
Discard used cotton swabs and used gloves according to
local regulations for biohazardous waste.
e
AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01
AUTION MICRO 30
When measurement and/or printing trouble occurs, the Analyzer displays
error codes. Perform the appropriate remedy described in the following
chart. If measurement is still impossible, contact your distributor.
Optical System does not function correctly.
E001Optical Movement
Make sure that the Strip Holder and Test Strip are set properly and turn on the power.
Lamp intensity of LED is too strong.
E002Lamp Intensity High
Clean the White Plate of the Strip Holder with a soft cloth moistened with water or alcohol and turn on the power.
Lamp intensity of LED is too weak.
E003Lamp Intensity Low
Clean the White Plate of the Strip Holder with a soft cloth moistened with water or alcohol and turn on the power.
No Printer PaperE004 Load printer paper.(Press [STOP] to restore the Ready Screen.)
Printer does not function correctly.
E005 Make sure that the Analyzer and the printer are connected correctly. (Press [STOP] to restore the Ready Screen.)
The Test Strip is not inserted correctly.
E006Strip Error
Insert the Test Strip and Strip Holder correctly and start measurement again.(Press [STOP] to restore the Ready Screen.)
Error CodeError Comment Cause Remedy
Test Strip type is not judged.
Important
Troubleshooting Troubleshooting and Maintenance
84-00259E
SampleTest StripMeasurement ItemMeasurement Method
Measurement WavelengthSample Supply Reaction TimeCalibration FunctionDisplayOperator Panel Memory CapacityThermal Correction FunctionSG Correction FunctionExternal OutputMeasurement Environment Condition DimensionsWeight Power Supply
1234
56789
101112131415161718
UrineAUTION Sticks 10EA, Uriflet 9UBGLU, URO, PRO, KET, BLD, BIL, NIT, LEU, pH, S.G. Test stripDual or Single (BLD only) Wavelength Reflectance MethodTriple Wavelength, Individual LED (565, 632, 755nm)Manual DippingApprox. 45 seconds/testAutomatic Calibration by CAL padLCD12 keys100 testsAutomatic Correction Function (10-30℃) Automatic Correction according to pH measurement valueConformed to RS-232CTemperature:10~30℃, Humidity:20~80%RH 125(W)×87(D)×31(H) mmApproximately 300gSize AA alkali cells (×2) (Or AC adapter)
No. Items Contents
PrinterPrinter PaperDimensionsWeightPower Supply
12345
36 digits of half-sized characters, Line Thermal PrinterThermal Printer Paper 58mm (Width)×φ26mm 125(W)×163(D)×38(H) mmApproximately 300g AC adapter (Approximately 20W)
No. Items Contents
Printer
AnalyzerSpecifications