Genius TM 2 Thermometer Checker/Calibrator Operating Manual F/C˚ F/C˚ Checker / Calibrator For use with Model 303000 Tympanic Thermometers

Transcript of Operating Manual Genius 2 - Medtronic · ensure that the calibration tolerances are correctly...

GeniusTM 2

Thermometer Checker/Calibrator

Operating Manual

F/C˚

F/C˚

Checker / Calibrator

For use with Model 303000 Tympanic Thermometers

Table of Contents

Page

Section I - Functional Description................................................................................................................................................... 1

Section II - Precautions and Warnings........................................................................................................................................... 2

Section III - Icon Identification........................................................................................................................................................ 3

Section IV - Required Equipment.................................................................................................................................................... 4

Section V - System Setup.................................................................................................................................................................. 4

Section VI - Operational Procedure................................................................................................................................................. 5

ApplyPower.................................................................................................................................................................................. 5

Start-upScreen............................................................................................................................................................................. 5

ReminderScreen........................................................................................................................................................................... 5

MainMenu.................................................................................................................................................................................... 5

Preferences.................................................................................................................................................................................... 5

LanguageSelection....................................................................................................................................................................... 5

SetUser/SiteInformation.............................................................................................................................................................. 5

SetDateandTime......................................................................................................................................................................... 6

CheckCalibration........................................................................................................................................................................... 6

Section VII - Error Messages............................................................................................................................................................. 8

Section VIII - Checker/Calibrator Specifications............................................................................................................................ 9

Section IX - Cleaning Instructions................................................................................................................................................... 9

Section X - Troubleshooting........................................................................................................................................................... 10

Section XI - Customer Service........................................................................................................................................................ 11

Section XII - Warranty..................................................................................................................................................................... 11

Section XIII - Electromagnetic Conformity Declaration............................................................................................................. 12

ThisproductcontainssoftwaresolelyownedbyCovidien.Covidiengrantstheuseranon-exclusive,limitedlicensetousethesoftwareaccordingtotheoperatinginstructions.AcopyofthelicensecanbeobtainedfromCovidien.

en

GeniusTM2Checker/Calibrator Table of Contents1

Section I - Functional Description

ThismanualdescribestheoperationoftheGenius™2Checker/Calibrator.ThisdevicehasbeendevelopedtochecktheaccuracyoftheGenius™2TympanicThermometerandautomaticallyrecalibratethethermometer,ifnecessary.Inaddition,theGenius2Checker/CalibratorhasthecapabilitytogenerateatestreportforeachthermometertestedandsaveittoaUSBflashdrive.TheGenius2Checker/CalibratorwillworkwithallGenius™2Thermometersoftwarerevisions.

TheGenius2Checker/Calibratorcontainstwoindependentlycontrolledinfraredcalibrationtargetsthataresimilartofactorycalibrationtargets.These“blackbody”targetsaredesignedtohaveefficientradiativeheattransferandproduceinfraredradiationthatapproximatesthehumanearattemperaturesof32.2°C(90.0°F)and40.6°C(105.0°F).

F/C˚

F/C˚

1 2 3 4

8 910

5

6

7

11

F 3.15A 250V

Checker / Calibrator

Figure 1 - Front Panel Layout

Legend

1 PowerConnector 7 HighTemperatureTarget2 PowerSwitch 8 HighTemperatureTargetLED3 FuseHolder 9 LCDDisplay4 LowTemperatureTargetLED 10 EncoderKnob5 LowTemperatureTarget 11 USBport6 Genius2ThermometerConnector

GeniusTM2Checker/Calibrator Table of Contents2

Section II - Precautions and Warnings

• TheGenius2Checker/Calibratorisdesignedtobeoperatedasspecifiedinthismanual.ProtectionprovidedbythesafetydevicesintheGenius2Checker/Calibratormaybeimpairedifthedeviceisnotusedinaccordancewiththedirectionscontainedinthismanual.

• TheGenius2Checker/Calibratorenclosureisnotwaterproof.Avoidspillingliquidofanykindontothefrontpanelofthedevice.Likewise,theGenius2Checker/CalibratorispoweredfromanACpowersource.Avoidthepotentialforcontactbetweenliquidandanyexternalsurfaceofthedevicewhileconnectedtothemainssource.

• Thesurfacequalityoftheinteriorofthereferenceblackbodytargetsiscriticaltotheiroperationasefficientinfraredradiators.Careshouldbeexercisedwhenworkingneartheentrancetotheblackbodytargetstoensurethatnocontactismadewiththesurfaceinsidetherubberflaps.

• Intheeventofablownfuse,replacethefusewithafusethatexactlymatchestheratedvoltage,current,andformfactorasspecifiedintheGenius2Checker/CalibratorSpecificationssectionofthismanual.MakesurethedeviceisdisconnectedfromtheACpowersourcebeforechangingablownfuse.

• Thisdeviceshouldnotbeusedinthepresenceofflammableanesthetics.Thereisriskofanexplosioninthepresenceofthese anesthetics.

• Therearenouser-serviceablecomponentsinthisdevice.Donotremovethefrontpanel.Referservicingtoqualifiedservice personnel.

• Useonlythesuppliedpowermodulewiththisdevice.• WhenpromptedtoinserttheGenius2Thermometerintooneoftheblackbodytargets,thereisa60secondwindowoftimeto

accomplishtheinsertion.Ifthetimedelayexceedsthisperiod,theprocedurewillbeterminated.• Priortochecking/calibratingaGenius2Thermometer,ensurethatthethermometerlensisclean.Effectivecleaningwill

ensurethatthecalibrationtolerancesarecorrectlyverified.RefertothecleaninginstructionsintheGenius2Thermometeroperating manual.

• WheninstallingaGenius2probecoverontothetipoftheGenius2Thermometer,useonlycoversthatareinacassetteandnotreleased.Trytoavoidallhandcontactwiththeprobecoversinordertopreventinaccuratereadings.Donotreuseprobecovers.UseonlyGenius2probecovers.

• LettheGenius2Checker/Calibratorwarmupafterpower-upforatleast15minutesbeforeuse.MakesurethattheGenius2Thermometers,probecoversandtheGenius2Checker/Calibratorhavehadenoughtimetoequilibratetoroomtemperaturebeforeproceeding.Allowalongerwarmupperiodifthedeviceorthermometerswereattheextremelimitsofallowedambienttemperature.TheportabilityoftheGenius2Checker/Calibratorwillallowforrelocationofthedevicetoareasthatmeettheambienttemperaturerequirements.

• TomaintainElectromagneticCompatibility(EMC)conformance,useaUSBflashdrivethatisCEmarkedandmeetsEN55022 standard.

• DonotputorresttheGenius2ThermometeronthesurfaceoftheGenius2Checker/Calibratoratanytime.Thethermalsensorsinsidethethermometerprobetipwilldetecttheheattransferredfromtheblackbodytargetsthroughthefrontpanelsurface.Thiswillcauseinstabilityofthethermometersensors,whichwillproduceerroneoustemperaturereadings.

• TheGenius2Checker/Calibratormayonlybeusedindoors,inanareafreefromdraftsandwideambienttemperatureswings.• TheUSBportonthefrontpanelisintendedtobeusedwithonlyaUSBFlashmemorydrive.NootherUSBdevicesshouldbe

insertedintotheUSBport.TheUSBflashmemorydriveshouldbeinstalledpriortodevicepower-up.• IntheeventthattheGenius2Checker/Calibratorisdropped,returntheunittothefactoryforre-calibration.

GeniusTM2Checker/Calibrator Table of Contents3

Section III - Icon Identification

SN SerialNumber

CE MARK SYMBOLS

NON-CE MARK SYMBOLS

ELECTRONIC SYMBOLS AND CERTIFIED TESTING MARKS

CorrugatedRecycles

STERILE

Keep AwayFrom Sunlight Keep Dry

Single Use

LOT

0123

FRAGILETHIS SIDE UP

DEHP – FREE

40°

Avoid TemperatureExtremes

Do not use if package is opened or damaged.

STERILE R

ASTERILE

EOSTERILE

STERILE

STERILE R

ASTERILE

EOSTERILE

Humidity LimitationAtmospheric pressure limitation

Use By

C US

LATEX – FREE

Caution, consultaccompanying

documents

ConsultInstructions

For Use

MRMR UNSAFE NON-PYROGENIC

PROTECT FROM HEAT ANDRADIOACTIVE SOURCES

NONSTERILE

2STERILIZE

DO NOT RESTERILIZE

POWDER – FREE

POWDER

Manufacturer

ConsultAccompanyingDocuments

CE MARK SYMBOLS

NON-CE MARK SYMBOLS

ELECTRONIC SYMBOLS AND CERTIFIED TESTING MARKS

CorrugatedRecycles

STERILE

Keep AwayFrom Sunlight Keep Dry

Single Use

LOT

0123

FRAGILETHIS SIDE UP

DEHP – FREE

40°

Avoid TemperatureExtremes

Do not use if package is opened or damaged.

STERILE R

ASTERILE

EOSTERILE

STERILE

STERILE R

ASTERILE

EOSTERILE

Humidity LimitationAtmospheric pressure limitation

Use By

C US

LATEX – FREE

Caution, consultaccompanying

documents

ConsultInstructions

For Use

MRMR UNSAFE NON-PYROGENIC

PROTECT FROM HEAT ANDRADIOACTIVE SOURCES

NONSTERILE

2STERILIZE

DO NOT RESTERILIZE

POWDER – FREE

POWDER

DateofManufacture

StandbyPower DoubleorReinforcedInsulationProtection

ReplacewithSameRatedFuse CEMark

USBPort DisposeofasElectricalandElectronicWaste

F/C˚

HighTemperatureTarget

LISTED(LABORATORY EQUIPMENT)

46EA

ULListingInformation

F/C˚

LowTemperatureTarget

Caution:IndoorUseOnly

Genius2ThermometerConnection

GeniusTM2Checker/Calibrator Table of Contents4

Section IV - Required Equipment

1. Genius2Checker/Calibrator 2. PowermodulewithACadapter 3. Probecable 4. Operator’sManual 5. Genius2Thermometerstobetested 6. Genius2disposableprobecovers 7. SmallPhillipsscrewdriver

Section V - System Setup

The following actions should precede use of the Genius 2 Checker/Calibrator: 1.ConnectthecorrectACplugadapterforyourcountrytothepowermodule. 2.Connectthepowermodulecabletothepowerjackonthefrontpanel. 3.ConnectthepowermoduletoanACpoweroutlet. 4.Turntheunitonviatheswitchonthefrontpanel. 5.LettheGenius2Checker/Calibratorunitwarmupforatleast15minutes. 6.SetasideacassettewithnewGenius2probecovers. 7.LettheGenius2Thermometersequilibratetotheroomtemperature. 8.PreparetheGenius2Thermometersbythoroughlycleaningthelensespriortouse.RefertotheGenius2Thermometeroperating

manualforcleaninginstructions. 9.RemovethebatterycompartmentaccesspanelbyusingasmallPhillipsscrewdriver.RemoveallthreeAAAbatteries. 10.ConnecttheprobecableroundconnectortothefrontpaneloftheGenius2Checker/Calibrator. 11. Connecttheotherendoftheprobeconnectorintothebatterycompartment.Theconnectorwillsnapintoplacewherethe batterieswere.RefertoFigure2.

Figure 2 - Insertion of Probe Connector into Genius 2 Battery Compartment

GeniusTM2Checker/Calibrator Table of Contents5

Section VI - Operational Procedure

ThefollowingprocedurewilldescribethestepsrequiredtooperatetheGenius2Checker/Calibrator.Inordertofacilitatefamiliaritywiththeproduct,itwouldbeadvisabletousethissectionofthemanualduringthecalibrationcheckofthefirstfewthermometers.Installation of USB Flash Drive (Optional) InstallaUSBflashdriveintotheUSBportlocatedonthefrontpanelifareportfilewrittentotheUSBflashdriveisdesired.IftheUSBflashdriveisinstalledafterperformingacalibrationcheck,anerrormessagemayoccurwhenattemptingtowritethereportfileiftheoperatordoesnotwaitfortheUSBflashdrivetocompleteinitialization.IftheGenius2Checker/Calibratorispoweredoff,allinternalmemory(notyetsavedtotheUSBflashdrive)willbelost.

Apply Power MakesuretheAC-adapterendofthepowermoduleispluggedintoagroundedoutlet.Connectthepowermoduletothefrontpanelpowerconnector.ApplypowertotheGenius2Checker/Calibratorbyusingthepowerswitchonthefrontpanel.

Start-up Screen ThefirstscreenontheLCDdisplayistheInitializationScreen,whichdisplaysthesoftwarerevisionanddateoftheGenius2Checker/Calibrator.Ifthereisaninternalproblemduringinitialization,anErrorscreenwillbedisplayed(SeeSectionVII).

Reminder Screen ThenextscreenontheLCDdisplayisaremindertolettheGenius2Checker/Calibratorwarm-upforatleast15minutesafterapplyingpower.Thiswilllettheinfraredblackbodytargetswarmuptothepropertemperatures.Toexitthisscreen,presstheencoderknob.

Main Menu TheMainMenuscreenisdisplayednext.Thisscreenhasfourmenuoptions:

CheckCalibration Preferences SetUser/SiteInformation SetTime/DateToenteranymenuoption,usetheencoderknobtoscrolloverandhighlightthedesiredmenuoption,thenpresstheencoderknob.

Preferences ThePreferencesmenuisusedtosetupthelanguageoption.PresstheencoderknobtoentertheLanguageSelectionscreen.

Language Selection TheLanguageSelectionscreenliststheavailablelanguagesfortheLCDscreentext.Toselectalanguage,scrollovertheselectionanditwillbehighlighted.PresstheencoderknobtocompletetheselectionandreturntotheMainMenuscreen.Theselectedlanguagewillbestoredinnon-volatilememoryandwillbeusednexttimetheunitispoweredon.

Set User/Site Information TheSetUser/SiteInformationscreenisusedtoentertheoperator’snameandtheorganizationand/orsiteinformation.Thisinformationwillbewrittentothetestreportfile.

Toentertheinformation,usetheencoderknobtoscrollthroughthealphabetforagivenletterpositionintheoperatornameorsiteinformation.Whenthecorrectletterisshown,presstheencoderknobtoselecttheletterandadvancetothenextletterposition.Toenteraspace,justpresstheencoderknobwhenaspaceorblankletterisshowninthatletterposition.Theusernamecanbeupto20letterslong.Usetheencoderknobtoscrolltotheorganizationfieldandenteranameforthesiteinasimilarmanner.Toexitthemenu,scrolltotheDONEtextandpresstheencoderknob.

GeniusTM2Checker/Calibrator Table of Contents6

Set Date and Time TheSetDateandTimescreenmenuoptionisusedtosetupthedateandtimethatisusedonthetestreport.Thetimeanddateareinternallystoredandautomaticallyupdatedwhentheunitispoweredup.

Toentertime,usetheencoderknobtoselectthehoursandminutes.Note:thehoursissetupusinga24hourclockformat,where1PMisenteredas13:00etc.Whenthevalueiscorrect,pushtheencoderknobtomovetothenextfieldofthetimeandthentothedatefield.ThedatefieldissetupusingaMMDDYYYYformat.Afterthelastfieldofthedateisentered,thenextencoderpushwillexitthismenuandreturntotheMainMenu.Note:thetimeanddateareshownatthebottomoftheMainMenu.

Check Calibration ThismenuoptionperformsthecalibrationcheckofaGenius2Thermometer,andifrequired,automaticallyrecalibratesthethermometerusingaseriesofstep-by-stepscreens.Beforeenteringthismenuoption,makesurethattherequiredequipmentlistedinSectionIVisavailable.Thisincludesacassettewithunused(new)probecovers.

Step 1: Warmupofinfraredblackbodytargets.Thisscreenisdisplayedwhilethetargetsarecheckedforpropertemperature.BoththeLowTargetandtheHighTargetarechecked,andwhentheyareatthecorrecttemperature,thescreenwillchangeto“OK”foreachtarget.Thereisaprompttopresstheencoderknobtocontinuetothenextscreen.Ifeithertargetcannotreachthecorrecttemperature,anErrormessagescreenwillbedisplayed.Ifthishappens,checktobesuretheGenius2Checker/Calibratoriswithinthecorrectambient(room)temperaturerange,andhasbeengivenatleast15minutesofwarmuptimesincepowerwasswitchedon.

Step 2: InspecttheGenius2Thermometerthatwillbetested.Ensurethatthethermometerlensisnotscratchedandisfreefromearwaxand/orothercontaminants.Ifthethermometerlensis“dirty”,cleanthelensaccordingtotheinstructionsintheGenius2Thermometeroperatingmanual.Oncethethermometerlensisclean,connecttheGenius2ThermometertotheChecker/CalibratorbypluggingtheroundendofthethermometerinterfacecableintotheroundconnectorsocketontheGenius2Checker/Calibratorfrontpanel(refertoFigure3).OpenthebatterydoorofthethermometerbyusingasmallPhillipsscrewdriver.Removethebatteriesandsetaside.Connecttheotherendofthethermometerprobecableintothebatterycompartment,makingsuretheconnector“snaps”intoplacesecurely.RefertoFigure2inSectionV.

Note: Anewprobecovermustbeusedpriortoeverytargetinsertion.Fingeroilsonpreviouslyusedprobecoversorre-usedprobecoversthathavebeenheatedbyGenius2Checker/Calibratortargetcancauseerroneousreadingswhichmayresultinafailedcalibrationcheckorrecalibration.Therefore,itisextremelyimportantthatanewprobecoverisusedeachtime.

Step 3: Presstheencoderknobtocontinue,theninstallanewGenius2probecoverontotheGenius2Thermometer.Makesuretouseacoverthatisheldinthecassette.Oncethecoverisonthethermometer,donotpointtheprobetipatanyobject,includinghands,fingersorLCDdisplay,asthiswillcauseaninaccuratetemperaturereading.Inspecttheprobecovertomakesureitisfullyseated(nospacebetweenthecoverandthetipbase)andnoholes,tears,orwrinklesarepresentintheplasticfilm.Whenthecoverison,inserttheprobetipintotheLowtemperaturetarget,whichwillhaveablinkingLEDindicatornexttoit.Makesurethattheprobetipisfullyinsertedintothetargetwell.RefertoFigure3forproperinsertion.Inordertoavoidatime-outerror,insertthecoveredprobetipintothetargetwithin60secondsfromthetimetheencoderknobispressed.

GeniusTM2Checker/Calibrator Table of Contents7

F/C˚

F/C˚

Checker / Calibrator

Figure 3 - Proper Insertion of Genius 2 Probe Tip into Target

Step 4: Theinsertionoftheprobetipintothetargetwillautomaticallybedetectedandthescreendisplaywillchangeto“PressGenius2Thermometerscanbutton.”PressthescanbuttonontheGenius2ThermometerwhileitisinsertedintotheLowtemperaturetarget.TheGenius2Thermometerwilloutput3shortbeepsasittakesthetemperatureoftheblackbodytarget.Afterthetemperatureistaken,thenextscreenwilldisplay.

Step 5: WithdrawthethermometerfromtheLowtemperaturetargetandejecttheprobecover.Thescreendisplaywillchangeto“InstallanewprobecoverandinsertprobeintotheHightarget.”Again,useanewGenius2probecoverfromthecassetteandbecarefulnottopointtheprobetipatanyobject,includinghands,fingersorLCDdisplay.Inspecttheprobecoverstomakesureitisfullyseated(nospacebetweenthecoverandthetipbase)andnoholes,tears,orwrinklesarepresentintheplasticfilm.Whenthecoverison,inserttheprobetipintotheHightemperaturetarget,whichwillhaveablinkingLEDindicatornexttoit.Makesurethattheprobetipisfullyinsertedintothetargetwell.

Step 6: TheinsertionoftheprobetipintotheHightargetwillautomaticallybedetectedandthescreendisplaywillchangeto“PressGenius2Thermometerscanbutton.”PressthescanbuttonontheGenius2ThermometerwhileitisinsertedintotheHightemperaturetarget.TheGenius2Thermometerwilloutput3shortbeepsasittakesthetemperatureoftheblackbodytarget.Afterthetemperatureistaken,thenextscreenwilldisplay.Step 7: IftheGenius2thermometeriswithinspecifiedaccuracylimits,aResultsscreenwillbedisplayedshowingthethermometerserialnumberalongwithanindicationthatthethermometerpassedcalibrationcheck.IftheGenius2thermometerisfoundtobeoutsidethespecifiedaccuracylimits,thedisplaywillindicatetotheuserthattheGenius2Checker/CalibratorisenteringintoacalibrationsequencethatrequirestheusertorepeatSteps3through6threemoretimesinordertogatherrecalibrationdata.FollowingthelastpassthroughSteps3through6,theGenius2Checker/CalibratorwillattempttorecalibratetheGenius2thermometer.Iftherecalibrationissuccessful,aResultsscreenwillbedisplayedshowingthethermometerserialnumberalongwithanindicationthatthethermometerpassedcalibration.Iftherecalibrationisunsuccessfulorthethermometerisunabletobecalibrated,aResultsscreenwillbedisplayedshowingthethermometerserialnumberalongwithanindicationthatthethermometerfailedcalibration.

GeniusTM2Checker/Calibrator Table of Contents8

Step 8: Afterthepass/failResultsscreenhasbeendisplayed,presstheencoderknobtodisplayasecondResultsscreencontainingthetargettemperatures,theGenius2readings,andthetargetversusthermometervariances.Step 9: Thenextscreenwilldisplay“WritereporttoUSBflashdrive?”anddisplayYesandNobelow.Ifareportisdesired,insertaflashdriveintothefrontpanelUSBport.Note:allowabout30secondsafterinsertionfortheUSBflashdrivetoinitialize.UsetheencoderknobtoscrolltoeitherYesorNoandpresstheencoderknobtomakeaselection.IfNoisselected,thescreenreturnstotheMainMenu.IfYesisselected,areportiswrittentotheUSBflashdrive,usingthethermometerserialnumberforfilenameand.TXTasthefileextension.Thefilecontainsthetestresultsaswellasotherinformationsuchasserialnumber,date,time,username,andorganization.ThisfileisasimpleASCIItextfileandisonlyavailableinEnglish.ThisfilecanbereadbyMicrosoftWindows®operatingsystemandMicrosoftWindows®notepadandsenttoaprinter.Afterthefileiswritten,thedisplaywillshow“Pushknobtocontinue.”PresstheencoderknobtoreturntotheMainMenu.Step 10: Aftercompletingacalibrationcheck/recalibration,removetheprobeconnectorfromthebatterycompartment,usingarockingmotiontofreetheconnector.Thenreinstallthe3AAAbatteries,observingtheproperpolarityorientation.Finally,reinstallthebatterycompartmentcoverandtightenthescrew.Step 11: RemoveUSBflashdrivebeforeclosingthelidofthecase.ThelidshouldbeclosedwhentheGenius2Checker/Calibratorisnotinuse.

Section VII - Error Messages

Theerrormessageswillhaveanumberassociatedwiththem,suchasError1,etc.Thefollowingisanexplanationoftheerrors:

Error 1: Internal Error ThiserrorisanindicationofaproblemwithintheGenius2Checker/Calibrator.Ifthiserrorpersistsaftercyclingpoweroffandbackon,returntheunitforfactoryservicing.

Error 2: Unable to communicate with Genius 2 Thermometer. ThiserrorisanindicationofeitheralackofcommunicationoramiscommunicationbetweentheGenius2ThermometerandtheGenius2Checker/Calibrator.Checkbothendsofthethermometerprobecable,especiallytheendthatconnectsinsidethebatterycompartment.Thepinsinthethermometerprobecablecanlosespringactioniftheyarebentordamaged.Afterinspectingtheprobecable,re-attemptthecalibrationcheck.Iftheproblempersists,tryadifferentGenius2Thermometer,ifavailable.Ifthisdoesnotresolvetheproblem,returntheunitforfactoryservicing,alongwiththethermometercableandtheGenius2Thermometerthatindicatedthiserror.

Error 3: TimeoutThiserrorisanindicationthattheoperatortooktoolongtoperformanoperation.ThiserrormightbeseenwhenconnectingtotheGenius2Thermometer,orwhileinstallingaprobecoveronthethermometer.Thetimeoutissetforoneminute.Tryrepeatingtheoperationthatgavethiserror.

Error 4: Target Out of Range Thiserrorisanindicationthatoneoftheblackbodytargetsisnotwithintherangeoftemperatureallowed.Checkthattheunitisbeingoperatedwithintheallowedambienttemperatureoperatingrange.RefertoSectionVIII.Ifoperatingwithinthespecifiedambientconditionsdoesnoteliminatetheerror,theblackbodytarget(s)maybedamaged.Returntheunitforfactoryservicing.

Error 5: USB Flash Drive Write Error ThiserrorisanindicationofaproblemwhenwritingthereportfiletoaUSBflashdrive.CheckthatthewriteprotectswitchontheUSBdriveisinthe“off”position.Makesurethatthereissufficientfreespaceonthedrive,andthatsecurityorencryptionfeaturesarenotinuse.

Error 6: Ambient Temperature Out of RangeThiserrorisanindicationthattheambienttemperatureisnotwithinthespecifiedrangeforperforminganaccuraterecalibration.Checkthattheunitisbeingoperatedwithintheallowedambienttemperatureoperatingrange.RefertoSectionVIII.

GeniusTM2Checker/Calibrator Table of Contents9

Error 7: Wrong TargetThiserrorisanindicationthattheGenius2Thermometerhasbeeninsertedintowrongtargetorwasinsertedintothetargetprematurely.Pleasetakecaretofollowtheinstructionsonthescreenandnotperformstepspriortotheinstructionsbeingdisplayed.

TheErrorscreenwillbedisplayedfor60secondsbeforereturningtotheMainMenu.ToexittheErrorscreensooner,presstheencoderknob.

Section VIII - Checker/Calibrator Specifications

Temperature Set PointsLow Target . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32.2°C+/-0.3°C(90.0°F+/-0.5°F)High Target . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40.6°C+/-0.3°C(105.0°F+/-0.5°F)

Target Temperature Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .InitialLimit+/-0.03°C(0.05°F) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Postyear1+/-0.06°C(0.10°F)

Genius 2 Thermometer Accuracy After Recalibration*Readings between 36. 0 °C and 39.0°C(96.8 °F and 102.2 °F). . . . . . . . . . . . . . . . . . . . . . . . +/-0.2°C(+/-0.4°F)Readings less than 36. 0 °C (96.8 °F) or greater than 39.0°C(102.2 °F). . . . . . . . . . . . . . . . . . . . +/-0.3°C(+/-0.5°F)

Warm-up Time Before Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15minutes(fromnon-extremeambient)Ambient Temperature Operating Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21.1-26.7°C(70-80°F)Relative Humidity Operating Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50+/-20%,non-condensingAltitude Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Upto2000meters(6563feet)Storage Temperature Range . . . . . . . . . . . . . . . . . . . . . . . . . -25°C-55°C(-13°F-131°F)85%RHnon-condensing

Approximate SizeLength . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .31.1cm(12.25”)Depth . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15.9cm(6.25”)Width . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26.0cm(10.25”)Weight. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.3kg(7.2lbs)Power Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100-240V0.8-0.4A47-63HzFuse Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . SchurterFSF5x20mmFastBlow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .LowBreakingCapacity250V,3.15ARecommended Checker/Calibrator Calibration Interval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .OneYearESD Immunity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .MeetsIEC61000-4-2Non-continuousCriteriaCPollution Degree . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2Installation Category. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . II . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IndoorUseOnly*PostrecalibrationaccuracyusingtheGenius2Checker/Calibratormaynotnecessarilybeequivalenttofactorycalibration

Section IX - Cleaning Instructions

Caution: Do not immerse the checker/calibrator, cables, or power cord in water or other cleaning solution; clean using a damp (not wet) cloth. Failure to follow cleaning procedures described herein could result in hazards to users.

AswithanyACpoweredelectricaldevice,caremustbetakentopreventliquidfromenteringtheChecker/Calibratortoavoidelectricalshockhazard,firehazard,ordamagetoelectricalcomponents.

GeniusTM2Checker/Calibrator Table of Contents10

General Cleaning InstructionsCleaningoftheGenius2Checker/Calibratormaybeperformedasfollows:

• First, unplug theGenius2Checker/Calibratorfrom power source. Never clean the device while connected to the mains supply.

• Amilddetergentshouldbeusedforgeneralcleaningofoutsidesurfaces.Also,aclothdampedwithisopropylalcoholcanbeusedtocleantheunit.Do not useabrasivesorsolvents.Avoidexcessmoisturearoundthetargetassembliesandtheelectricalconnectors.

• Donotattempttocleanthetargetbelowtherubberflaps.• Donotcleanthetargetnestsorflapswithalcohol.

IfthereisanydoubtaboutthecompatibilityofacleaningagentwithpartofthisequipmentoritsmaterialscontactCovidiencustomerservice(SeeSectionXI).

Section X - Troubleshooting

Some of the common problems are as follows:

Problem Probable Cause Corrections

Genius2Checker/Calibratorwillnotpowerup

Genius2Checker/CalibratornotreceivingACpower CheckACpowerconnections

Blownfuse Replacefuse

TestreportdoesnotwritetoUSBflashdrive

USBflashdrivenotformatted UsePCorlaptoptoformatflashdrive

USBflashdrivewriteprotected UsePCorlaptoptoturnoffwriteprotection

USBflashdrivedoesnotcontainenoughfreememoryspace UsePCorlaptoptodeleteunnecessaryitemsorusedifferentUSBflashdrive

USBflashdriveincompatiblewithGenius2Checker/Calibrator UsedifferentUSBflashdrive

Genius2Checker/CalibratorhasnothadenoughtimetodetectthenewlyinsertedUSBflashdrive

Retrywriteprocess

MultipleGenius2thermometersfailrecalibration

Probecoverstuckinoneoftheblackbodytargets Removeprobecoverfromblackbodytargetusingsmallneedle-nosepliersorhemostats.Donotscratchtheinteriorsurfaceofthetarget.

Dirtyprobecover Useanewprobecoverforeachreading.Donotre-useaprobecover.Probecoversshouldneverbetouchedbyhand.

Dirtythermometerlens CleanthethermometerlenspertheinstructionsintheGenius2operatingmanual.

Genius2thermometernotseatedcorrectly EnsurethattheGenius2thermometeriscompletelyseatedinthetargetandattheproperangle.

Genius2thermometers,probecovers,and/orchecker/calibratornotatroomtemperature.

EnsurethattheGenius2thermometers,probecovers,and/orchecker/calibratorhavehadenoughtimetoequilibratetoroomtemperaturebeforeproceeding.

Error6repeatedlyappearswhiletestingthesameGenius2thermometer

PotentialpowerconsumptionissuewithGenius2thermometer Repeatcalibrationcheck/recalibrationwithadifferentGenius2thermometerandreturnproblemthermometerforfactoryservicing.

Genius2thermometerseemstoreadhighorlowfollowingasuccessfulrecalibration

Probecoverreusedduringcalibrationcheck/recalibration Repeatcalibrationcheck/recalibrationusinganewprobecoverpriortoeverytargetinsertionduringtheprocess.

GeniusTM2Checker/Calibrator Table of Contents11

Section XI - Customer Service

TheGenius2Checker/Calibratorrequiresfactorycalibrationandinspectiononceayear.NotifyandreturnthedevicetoCovidienonorbeforethecalibrationdatefortheChecker/Calibrator.TheChecker/Calibratorwilldisplayamessageonemonthpriortothecalibrationdateasareminder.

Caution:There are no serviceable parts inside, return to factory for service.

Intheeventthatitisnecessarytoreturnaunitforrepair,pleaseobservethefollowing:

1. ContactCovidienasshownbelow,oryourlocalcustomerservicerepresentativeforanAuthorizedReturnNumberandshipping instructions.

2. Shipinsuredparceltoyourlocalservicecontactortheappropriatelocationbelow.

Parts ListingPleasecontactyourlocalcustomerservicecenterorsalesrepresentativeforthepartslistedbelow.

DescriptionGenius2Checker/Calibrator. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .303097RefurbishedGenius2Checker/Calibrator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .303096

Section XII - Warranty

CovidienwarrantstheGenius2Checker/Calibratortobefreeofdefectsinmaterialsandworkmanshipundernormaluseandserviceforaperiodof1yearfromthedateofdeliverybyCovidientothefirstpurchaser.Liabilityhereunderislimitedtorepairorreplacement,atCovidien’soption,atCovidienservicefacilityforanyproduct,whichshallundernormaluseandservice,appeartoCovidientohavebeendefectiveinmaterialorworkmanship.Thiswarrantyshallnotapplyto,andCovidienshallnotberesponsiblefor,anylossarisinginconnectionwiththepurchaseofanyproductthathasbeenrepairedbyanyoneotherthanCovidienoritsauthorizedrepresentative,orwhichhasbeensubjecttomisuse,neglect,oraccident,orwhichhasbeenusedotherwisethaninaccordancewiththeinstructionsfurnishedbyCovidien.Covidienneitherassumesnorauthorizesanyrepresentativeorotherpersontoassumeforitanyliabilityotherthanasexpresslysetforthherein.

Covidien disclaims all other warranties, expressed or implied, including any implied warranty of merchantability or fitness for a particular purpose or application other than as expressly set forth in the product labeling. In no event will Covidien be liable for any incidental, indirect, or consequential damages in connection with the purchase or use of this product.

UnitedStatesCovidien5920LongbowDriveBoulder,CO803011-800-448-0190

CanadaCovidien7300TransCanadaHighwayPointe-Claire,QCH9R1C7,Canada1-877-664-8926

OutsideUS&CanadaCovidienUnit2TalismanBusinessCenter,LondonRoadBicester,OX266HR,UK1+44-1869-32806

GeniusTM2Checker/Calibrator Table of Contents12

Section XIII - Electromagnetic Conformity Declaration

TheGenius2Checker/CalibratorhasbeenbuiltandtestedaccordingtoIEC/EN61010-1andEN61326-1standards

TheChecker/Calibratorisintendedforuseintheelectromagneticenvironmentspecifiedbelow.Theuseroftheequipmentshouldassurethatitisusedinsuchanenvironment.

Emissions Test Compliance Electromagnetic Environment -Guidance

ConductedandRadiatedEmissions(EN61326-1)

Group1,ClassB TheChecker/Calibratoremissionsareverylowandarenotlikelytocauseinterferenceinnearbyelectronicequipment.

HarmonicCurrent(EN61000-3-2)

ClassA

VoltageFluctuationandFlicker(EN61000-3-3)

Complies

RadiatedDisturbanceImmunity(EN61000-4-3)

Complies

ConductedDisturbanceImmunity(EN61000-4-6)

Complies

PowerFrequencyMagneticFieldImmunity(EN61000-4-8)

Complies

VoltageDipsandInterruptsImmunity(EN61000-4-11)

Complies

ElectricalFastTransient/BurstsImmunity(EN61000-4-4)

Complies

SurgeImmunity(EN61000-4-5) Complies

ElectrostaticDischarge(EN61000-4-2)

MeetsCriteriaC,Non-Continuous

Manual No. 1038307

COVIDIEN, COVIDIEN with logo and Covidien logo are U.S. and internationally registered trademarks of Covidien AG.

Microsoft and Windows are either registered trademarks or trademarks of Microsoft Corporation in the United States and other countries.

Other brands are trademarks of a Covidien company.

© 2011 Covidien.

Made in USA. Covidien llc, 15 Hampshire Street, Mansfield, MA 02048 USA.

Covidien Ireland Limited, IDA Business & Technology Park, Tullamore.

www.covidien.com 1-800-448-0190 [T]

REV 12/2011

Identification of a substance that is not contained or present within the product or packaging.Identification d’une substance non présente dans le produit ou l’emballage.Identifizierung einer Substanz, die weder im Produkt noch in der Verpackung enthalten bzw. vorhanden ist.Identificazione di una sostanza che non è contenuta né presente nel prodotto o nella confezione.Identificación de una sustancia no contenida o que no esté presente dentro del producto o embalaje.Identifiering av ett ämne som inte ingår eller förekommer i produkten eller förpackningen.Identificatie van een stof die niet in het product of de verpakking is vervat of aanwezig is.Identificação de uma substância não contida ou não existente no produto ou embalagem.Tuotteessa tai pakkauksessa olemattoman aineen tunnistus.Identifikation af et stof, der ikke er indeholdt eller til stede i produktet eller emballagen.Αναγνώριση ουσίας που δεν περιέχεται ή δεν υπάρχει στο προϊόν ή στη συσκευασία.Identifikace látky, která není obsažena nebo přítomna v produktu nebo obalu.Olyan anyag azonosítása, amelyet sem a termék, sem a csomagolás nem tartalmaz, vagy amely azokban nincs jelen.Определение вещества, не содержащегося или не присутствующего в продукте и упаковке.Identyfikacja substancji nie zawartej i nieobecnej w produkcie lub jego opakowaniu.Ürün veya ambalajında bulunmayan veya var olmayan bir madde tanımlaması.Identifikasjon av et stoff som ikke er til stede i produktet eller emballasjen.