M E A S U R E M E N T S Y S T E M S
NUFLO ™
Scanner® 2000microEFM
The Scanner 2000 microEFM is among the most versatile flow computers on the market. The device can operate alone to fulfill a variety of measurement applications or be networked with other devices to provide a comprehensive flow management SCADA solution.
Key Features & Functions• Autonomous operation for
1 year (typical) via integrallithium battery pack*
• Measures •• Industrial and Hydrocarbon
Gases•• Liquids•• Saturated Steam
• Senses•• Differential Pressure•• Pressure•• Temperature•• Flowmeter Pulses
• Computes•• Standard Volume•• Mass•• Energy
• Logs, indicates andcommunicates any measured or computed variable
• Downloads historical data athigh speeds for analysis and data export using full-featuredWindows-based software
• Certified internationally formeasurement and electricalsafety•• ATEX/IECEx•• CSA•• Measurement Canada•• GOST
• Supports a variety of protocols•• RTU Modbus®
•• Enron Modbus®
•• FOUNDATION™ Fieldbus
* Actual battery life is dependent on calculation,sampling, and display update frequency.
Mounts directly to turbine meters, orifice meters and cone meters
Flow Computer
Data Logger
Multi-variable Transmitter
PID Controller
FOUNDATION™ Fieldbus CommunicationsThe Scanner 2000 for FOUNDATION™ fieldbus is an a CSA explosionproof andATEX intrinsically safe device that is approved by the Fieldbus Foundation™.An integral fieldbus module converts Scanner 2000 Modbus® data to fieldbussignals. Additional inputs, flow volumes and calculations can be read by afieldbus host and recorded. Power and communications are supplied by afieldbus network, and the integral lithium battery provides backup power.
2
PID Control*When purchased with the PID control option, the Scanner 2000 can be used to control processvariables such as static pressure, differentialpressure, temperature, and flow rate. A 4-to-20mA output is configured to regulate a controlvalve or adjustable speed drive (ASD), and controlparameters are tuned with the easy-to-use toolsthat are built into the standard Scanner 2000software platform, ModWorX Pro. Users canconfigure the Scanner 2000 to provide PID controlfor a single parameter, or opt for PID control offlow rate with a secondary pressure control.
See our Feature Profile on PID Control on the Cameron website for more information.
* Not available with FOUNDATION™ fieldbus model
The Scanner 2000 can operate autonomously with an integralbattery for a year or longer. Using only its integrated sensors,the Scanner 2000 can measure differential pressure, pressure,temperature and flow meter pulses. The device can scale and calculate flow, indicate and record data, which can bedownloaded to a user’s PC via a USB connection. Alternatively,the integral keypad can be used to configure basic parametersand access historical data.
The instrument can be factory-installed on Cameron gas andliquid turbine meters, orifice meters, and cone meters orshipped separately for connection to other primary flowsensors. The Scanner 2000 can measure the standard volume,mass and energy flows of saturated steam and many types of gasses and liquids with custody transfer precision.
In its most basic form, the Scanner 2000 replicates the Barton 202 flow recorder, the Barton 242 pressure andtemperature recorder, and the Barton 200 and 227 differential pressure indicators.
With the addition of external connections, the Scanner 2000can also act as a multi-variable transmitter, RTU and PIDcontroller. This additional information can be transferred toand from the 2000 by analog or pulse (frequency), Modbus®
or FOUNDATION™ fieldbus serial communications.
A single Scanner 2000 is powerful enough to measure the gas,oil and water flow from a 2- or 3-phase separator or computethe flow from an ultrasonic gas flow meter. The Scanner 2000is also versatile enough to act as a simple two-stream turbinetotalizer or provide a fieldbus connection to a positivedisplacement or turbine meter.
For more information about Scanner 2000 applications, see the following publications:
• Scanner 2000 microEFM Solutions
• Scanner 2200 EFM Solutions
• Scanner 2000 microEFM Well Testing Solution
• Feature Profile on PID Control
3
M E A S U R E M E N T S Y S T E M S
Calculations • Flow rate (natural gas, steam, liquid)
•• AGA-3•• AGA-7•• Compensated Liquid Turbine•• ISO 5167•• Cone•• Averaging Pitot Tube (Annubar®)
• Fluid properties•• AGA-8-92 (Detail and Gross)•• IF-97 (Steam)•• Generic Liquid
(Water or Emulsions)•• API-2540 Liquid (Crude Oil, Jet
Fuel, Gasoline, Fuel Oils, Lube Oil)
• Wet correction (Steam)•• James (Orifice)•• Chisholm (Orifice)•• Steven (Cone)
Communications /Archive Retrieval • Modbus® (RTU) with two on-board
RS-485 communications slave ports•• COM 1 and COM 2 baud rates:
300 to 38.4K
• FOUNDATION™ fieldbus with one on-board RS-485 communications slaveport•• COM 1 baud rate: 300 to 38.4K
• USB communications with optional adapter
• Enron Modbus® compliantdownloads
• User-definable block reads allows thegrouping of up to 25 floating pointvalues for faster data transfer whenused with a SCADA system
• Full archive download inapproximately 3 minutes with main board only (6 minutes withexpansion board option)
Inputs Turbine Meter Inputs 1 and 2(Expansion Board Required for Turbine Input 2*)
• Configurable sensitivity adjustment (20 mV to 200 mV, peak to peak)
• Frequency range: 0 to 3500 Hz
• Input amplitude: 20 mV to 3000 mV,peak to peak
• Turbine Input 2 cannot be usedsimultaneously with a pulse input
Pulse Input(Expansion Board Required*)
• Accepts a signal from a turbine meteror PD meter
• Optically isolated
• Input: 3 to 30 VDC or contact closure
• Cannot be used simultaneously with Turbine Input 2
Analog Inputs 1 and 2(Expansion Board Required*)
• 3-wire sensor interface
• Sensor power same as external power supply for main board (6 to 30 VDC)
• Accuracy: 0.1% of full scale
• Temperature effect: 0.25% of fullscale over operating temperaturerange of -40°F to 158°F (-40°C to 70°C)
• Resolution: 20 bits
• User-adjustable sample time and damping
Process Temperature Input
• 100-ohm platinum RTD with 2-wire,3-wire, or 4-wire interface
• Sensing Range: -40°F to 800°F (-40°C to 427°C)
• Accuracy: 0.2°C (0.36°F) over sensingrange at calibrated temperature
• Temperature effect (Fahrenheit): 0.54°F over operating range of -40°F to 158°F
• Temperature effect (Celsius): 0.3°C over operating range of -40°C to 70°C
• Resolution: 24 bits
• User-adjustable sample time and damping
* Not available with FOUNDATION™fieldbus model
ApprovalsExplosion-proof Package
• CE approved
• Complies with EMC Directive2004/108/EC
• ATEX and IECEx certified, II 2 GD Ex d IIC T6 IP68 (-40°C to 70°C)
• CSA certified for US and Canada Class I, Div. 1, Groups B, C, D, Class I, Div. 2, Groups A,B,C,D (non-sparking), Type 4 enclosure, ANSI12.27.01 single seal (0 to 3000 psi)
• Measurement Canada certified for custody transfer, Approval No. AG-0557C
• GOST-R/GOST-K certified
• ASME pressure vessel codecompliant (0 to 3000 psi); CRN 0F10472.5C
• Fieldbus Foundation™ approved
Intrinsically Safe Package
• CE approved
• ATEX certified, II 2G Ex ia IIB T4 IP66(-40° to 78°C)
• Fieldbus Foundation™ approved
Display• Two-line LCD with easy-to-read
alphanumeric characters•• 8-digit display of values (top line) •• 6-digit display identifies each
scrolling parameter and itsengineering unit (bottom line)
• View up to 12 user-definedparameters
• View daily log data (99 days)
• User-selectable units ofmeasurement
• Character height - 0.3 in.
• Adjustable contrast and update period
Power• Lithium battery for autonomous
operation
• External power supply (6-30 VDC)optional with internal battery backup
• Fieldbus network power (Scanner 2000 FOUNDATION™ fieldbusmodel only)
4
TEST ACCESS
CONFIGURATION:Move between menusand menu selections
OPERATION:View next parameter
OPERATION:View daily logs
CONFIGURATION:Change digits andother menu selections
CONFIGURATION:Save configurationsettings
OPERATION:Save totals
PRESS +
simultaneously to view time/date, temperature, and battery voltage
PRESS +
simultaneously to accessConfiguration menu
Keypad ConfigurationWith the three-button keypad, changesto basic parameters can be made quickly and easily simply by removing theenclosure lid (computer software is notrequired). The slave address, baud rate,date and time, and orifice plate size canall be configured from the keypad.
Outputs
Digital Output
• Configurable as pulse output or alarm output
• Solid-state relay
• Output rating: 60 mA max @ 30 VDC
• Pulse output•• Configurable pulse duration•• Maximum frequency: 50 Hz•• Configurable pulse representation
(1 pulse = 1 MCF)•• Based on any accumulator
(flow run or turbine inputs)
• Alarm output•• Low/high•• Out-of-range•• Status/diagnostic•• Latched/unlatched•• Normally open/normally closed
Analog Output(Expansion Board Required*)
• 4-20 mA
• Accuracy: 0.1% of full scale @ 25°C(77°F), 50 PPM/°C (27.8 PPM/°F)temperature drift
• Represents any measured variable(e.g., differential pressure) orcalculated parameter (e.g., flow rate)
• Regulates control valve in PID control applications
• Optically isolated
• Resolution: 16 bits
* Not available with FOUNDATION™fieldbus model
5
M E A S U R E M E N T S Y S T E M S
Memory• Non-volatile memory for
configuration and log data•• 256 KB (512 KB with expansion
board option)•• Data stored for 10 years without
power
MVT• Provides linearized digital data
•• Static pressure•• Differential pressure
• Available with bottom ports (gas measurement) or side ports (liquid or steam measurement)
• NACE-compliant units also available
• User-adjustable sample time anddamping
MVT Accuracy• Stability: Long-term drift is less than
±0.05% of URL per year over a 5-year period
• Differential pressure: ±0.05% ofspan •• Effect on differential pressure for
a 1000-psi change in pressure
• Zero shift: ±0.05% of URL
• Span shift: ±0.01% of reading
• Static pressure: ±0.05% of span
• Temperature performance: 0.25% offull scale over full operatingtemperature range
• Resolution: 24 bits
Environmental Operating Temperature Range
• -40°F to +158°F (-40°C to +70°C)
• LCD contrast is reduced below -22°F (-30°C)
Audit Trail• Daily records: 768 (>2 years)
• Interval records: 2304 (>3 months of 1-hour intervals); 6392 (>8 months of 1-hour intervals) with expansion board option•• Adjustable from 5 seconds to
12 hours
• Event/alarm records: 1152
• Records up to 16 user-defined parameters
Interface Software• Provided at no charge
• Easy to use
• Real-time data polling
• Complete configuration•• Configuration upload tool for
configuring multiple units
• Multi-level security
• Field calibration •• 1 to 12 calibration points
for each parameter•• Three methods: multi-point,
set zero point, and verify•• Inputs are automatically locked
during calibration
• Maintenance•• Change plate•• Change cone
(linearization: 1 to 12 points)•• Change averaging pitot tube•• Change gas composition•• Change steam properties•• Change liquid parameters•• Change flow coefficients•• Change K-factor
(linearization: 1 to 12 points)•• Change turbine flowmeter
• Archive data downloads •• Configurable downloads of
“all” or “new” records•• Download types: daily, interval,
and event/alarm records•• Downloads are automatically
saved in uneditable binary (SDF) files
•• Exports to .xls, .csv, .rtf, .html, Flow-Cal® and PGAS® formats
• Reporting•• Daily logs (table or trend graph)•• Interval logs (table or trend graph)•• Event/alarm logs•• Configuration settings•• Calibration settings•• Snapshot of current status data
and calculated parameters
• Online documentation•• Quick Start guides•• Hardware manual•• Software manual
MVT Pressure Ranges
Static Maximum Pressure Differential Overrange
/SWP Pressure Pressure(PSIA) (In. H2O) (PSIA)
0100 030 150
0300 200* 450840 450
0500 030** 750200* 750
1500 200* 2250300 2250400* 2250840 2250
3000*** 200* 4500300 4500400* 4500840 4500
5300*** 200* 7420300 7420400 7420840 7420
* These ranges are stocked for quicker delivery.** Available on request.*** Not evaluated for Measurement Canada
approval
Four menussimplify navigation
Dedicated button for quick and easy downloads
Documentation for quick reference
Calibrate Inputs:Differential PressureStatic PressureProcess Temperature
Configure:SYSTEM PARAMETERS
Device IDDate and TimeCommunicationsSecurityDisplayArchiveAlarmModbus Registers
FLOW RUN SETUP
TURBINE SETUP
INPUTSDifferential PressureStatic PressureProcess Temperature
OUTPUTSDigital
Calibrate, Verify and Set Zero Offset selectiionsavailable for each parameter.
Maintain Flow Run:Change Orifice PlateChange ConeChange Gas CompositionChange Steam PropertiesChange Flow CoefficientsChange Avg. Pitot Tube
Maintain Turbine:Change Turbine FlowmeterChange K-Factors
The ModWorX Prosoftware interface offers easy access to the most commonly usedfunctions from one main display screen.
Task-based menus help operators perform their jobs quickly and easily.
EUROPE,MIDDLE EAST
& AFRICA
M E A S U R E M E N T S Y S T E M S
U S A • C A N A D A • U K • C H I N A • U A E • A L G E R I A • M A L A Y S I A • I N D I A • R U S S I A w w w. c - a - m . c o m / f l o
ASIAPACIFIC
1.800.654.3760ms-us @ c-a-m.com
NORTHAMERICA
281.582.9500HOUSTONHEAD OFFICE
M E A S U R E M E N T S Y S T E M S
EFM-2000 NF00097 1112 © Copyright 2011 Cameron International Corp. All Rights Reserved.