1IHE Workshop – June 2006 What IHE Delivers
Kevin O’DonnellKevin O’Donnell
Toshiba Medical SystemsToshiba Medical Systems
NM ImageNM Image
![Page 2: Nuclear Medicine Image Profile](https://reader035.fdocuments.net/reader035/viewer/2022081401/558d3da6d8b42a0b318b465c/html5/thumbnails/2.jpg)
2IHE Workshop – June 2006IHE Workshop – June 2006
NM ImageNM Image The ProblemThe Problem
DICOM NM objects are different and support in products is inconsistent
NM cardiac display needs are different and support in products is inconsistent
Clinical Analysis Programs are used extensively (particularly for Cardiology), but no support for distribution and display of those Result Screens
NM Users are not adept at getting these requirements accurately included in RFPs
Vendors (especially of general purpose systems) are not adept at accurately representing their support for such features
![Page 3: Nuclear Medicine Image Profile](https://reader035.fdocuments.net/reader035/viewer/2022081401/558d3da6d8b42a0b318b465c/html5/thumbnails/3.jpg)
3IHE Workshop – June 2006IHE Workshop – June 2006
General NM Viewing SuboptimalGeneral NM Viewing Suboptimal
NM PACS
![Page 4: Nuclear Medicine Image Profile](https://reader035.fdocuments.net/reader035/viewer/2022081401/558d3da6d8b42a0b318b465c/html5/thumbnails/4.jpg)
4IHE Workshop – June 2006IHE Workshop – June 2006
General NM Viewing SuboptimalGeneral NM Viewing Suboptimal
NM PACS
![Page 5: Nuclear Medicine Image Profile](https://reader035.fdocuments.net/reader035/viewer/2022081401/558d3da6d8b42a0b318b465c/html5/thumbnails/5.jpg)
5IHE Workshop – June 2006IHE Workshop – June 2006
General NM Viewing SuboptimalGeneral NM Viewing Suboptimal
Click and move mouse down to adjust window level, right to adjust window center
![Page 6: Nuclear Medicine Image Profile](https://reader035.fdocuments.net/reader035/viewer/2022081401/558d3da6d8b42a0b318b465c/html5/thumbnails/6.jpg)
6IHE Workshop – June 2006IHE Workshop – June 2006
Cardiac Viewing SuboptimalCardiac Viewing Suboptimal
If sent as Short axis Images not
displayed in a useful format.
If sent as screen snapshots(ie, static Secondary Capture) Can’t separately adjust stress and rest images Can’t view gated cine
![Page 7: Nuclear Medicine Image Profile](https://reader035.fdocuments.net/reader035/viewer/2022081401/558d3da6d8b42a0b318b465c/html5/thumbnails/7.jpg)
7IHE Workshop – June 2006IHE Workshop – June 2006
New in 2006 !New in 2006 !
Cardiac Option Better addresses cardiac needs
Minor restructuring of remaining profile General NM option MPR option
(Review option retired) (Comparison display eliminated)
![Page 8: Nuclear Medicine Image Profile](https://reader035.fdocuments.net/reader035/viewer/2022081401/558d3da6d8b42a0b318b465c/html5/thumbnails/8.jpg)
8IHE Workshop – June 2006IHE Workshop – June 2006
NM Image NM Image Abstract / ScopeAbstract / Scope
Content Profile for Users to request and Vendors to claim
No New Actors or Transactions Additional requirements on existing actors and transactions Total of 4 options
New Informative Text in Vol. 1 Clarifications of NM Workflow mapping to SWF Explanation of NM Data Explanations of NM Clinical Display concepts
New Normative Text in Vol. 2 < See Following Slides >
![Page 9: Nuclear Medicine Image Profile](https://reader035.fdocuments.net/reader035/viewer/2022081401/558d3da6d8b42a0b318b465c/html5/thumbnails/9.jpg)
9IHE Workshop – June 2006IHE Workshop – June 2006
NM ImageNM ImageValue PropositionValue Proposition
Use general Radiology Infrastructure to:
Satisfy Basic NM Data Needs Access to NM Images Access to usable presentations of NM Images Access to Clinical Processing Results
Satisfy NM Workflow Needs Map Cardiac and Multi-phase Studies to SWF Improved Query
![Page 10: Nuclear Medicine Image Profile](https://reader035.fdocuments.net/reader035/viewer/2022081401/558d3da6d8b42a0b318b465c/html5/thumbnails/10.jpg)
10IHE Workshop – June 2006IHE Workshop – June 2006
NM ImageNM ImageTransaction DiagramTransaction Diagram
![Page 11: Nuclear Medicine Image Profile](https://reader035.fdocuments.net/reader035/viewer/2022081401/558d3da6d8b42a0b318b465c/html5/thumbnails/11.jpg)
11IHE Workshop – June 2006IHE Workshop – June 2006
NM ImageNM ImageStandards UsedStandards Used
DICOM NM Image Object
DICOM Multiframe Secondary Capture Dynamic Results Collection of Static Results
DICOM Secondary Capture Static Results
ACC NM Cardiac Display format
![Page 12: Nuclear Medicine Image Profile](https://reader035.fdocuments.net/reader035/viewer/2022081401/558d3da6d8b42a0b318b465c/html5/thumbnails/12.jpg)
12IHE Workshop – June 2006IHE Workshop – June 2006
NM Image ActorsNM Image ActorsAcquisition Modalities Create NM objects as specified (Series, attribute & vector handling) Include specific codes in Cardiac studies
Evidence Creators Create NM objects as specified (Series, attribute & vector handling) Include specific codes in Cardiac studies
Image Manager/Archives Store/Retrieve NM & Secondary Capture (SC & MFSC) Objects Support Query on Study Description
Image Displays Display NM and Secondary Capture (SC & MFSC) Images Support specified display capabilities
![Page 13: Nuclear Medicine Image Profile](https://reader035.fdocuments.net/reader035/viewer/2022081401/558d3da6d8b42a0b318b465c/html5/thumbnails/13.jpg)
13IHE Workshop – June 2006IHE Workshop – June 2006
Image DisplayImage Display
Display NM and Result-Screen Images
Support specified NM essential display capabilities
Additional attribute handling e.g. Patient State, Image Orientation, Detector Sequence, View
Code Sequence
![Page 14: Nuclear Medicine Image Profile](https://reader035.fdocuments.net/reader035/viewer/2022081401/558d3da6d8b42a0b318b465c/html5/thumbnails/14.jpg)
14IHE Workshop – June 2006IHE Workshop – June 2006
Options: Image DisplaysOptions: Image Displays
General NM Option Whole body display
Cardiac NM Option ACC NM Cardiac display
Displays must support at least one of these options!
![Page 15: Nuclear Medicine Image Profile](https://reader035.fdocuments.net/reader035/viewer/2022081401/558d3da6d8b42a0b318b465c/html5/thumbnails/15.jpg)
15IHE Workshop – June 2006IHE Workshop – June 2006
Handle multi-frame / multi-vector images
Multiple simultaneous cines
Grayscale and color result screens
Dynamic and static result screens
Original resolution and scaled
Separate adjustment of upper and lower window levels
Apply local color tables to grayscale screens
Add new local color tables to system
Change intensity of grayscale result screens
Display Series Description tag
Image DisplayImage Display
![Page 16: Nuclear Medicine Image Profile](https://reader035.fdocuments.net/reader035/viewer/2022081401/558d3da6d8b42a0b318b465c/html5/thumbnails/16.jpg)
16IHE Workshop – June 2006IHE Workshop – June 2006
NM Image FormatNM Image Format
Understood by dedicated NM systems
Often Misunderstood by general storage and display systems
Multiframe Dynamic
Anterior Posterior Multi-energy
NM Image
Image vectors
![Page 17: Nuclear Medicine Image Profile](https://reader035.fdocuments.net/reader035/viewer/2022081401/558d3da6d8b42a0b318b465c/html5/thumbnails/17.jpg)
17IHE Workshop – June 2006IHE Workshop – June 2006
Image DisplayImage Display
Scale “intelligently” to avoid presenting postage-stamp size NM images
Allow user to select different vectors/framesets of complex multi-frame images
Allow user to separately adjust intensity for different framesets (eg, phases) in the same “image”
![Page 18: Nuclear Medicine Image Profile](https://reader035.fdocuments.net/reader035/viewer/2022081401/558d3da6d8b42a0b318b465c/html5/thumbnails/18.jpg)
18IHE Workshop – June 2006IHE Workshop – June 2006
Display FormatsDisplay Formats
Grid display Frames from one
image set
A32
…
…………………
……………………
…………………A9
A8 A7 A6A5 A4 A3 A2 A1
C1 C1
B2B1 A2 A1 Fit display Frames from multiple image
sets
![Page 19: Nuclear Medicine Image Profile](https://reader035.fdocuments.net/reader035/viewer/2022081401/558d3da6d8b42a0b318b465c/html5/thumbnails/19.jpg)
19IHE Workshop – June 2006IHE Workshop – June 2006
ACC NM Cardiac DisplayACC NM Cardiac Display
Generate long axis from short axis images
Display of 2 image sets (Stress and Rest)
Use patient state tags to set display order
Follow ACC guidelines for display
Controls for adjusting stress and rest Image alignment Window upper and lower levels.
![Page 20: Nuclear Medicine Image Profile](https://reader035.fdocuments.net/reader035/viewer/2022081401/558d3da6d8b42a0b318b465c/html5/thumbnails/20.jpg)
20IHE Workshop – June 2006IHE Workshop – June 2006
ACC NM Display ExampleACC NM Display ExampleShort axis
Horizontal Long
Vertical Long
![Page 21: Nuclear Medicine Image Profile](https://reader035.fdocuments.net/reader035/viewer/2022081401/558d3da6d8b42a0b318b465c/html5/thumbnails/21.jpg)
21IHE Workshop – June 2006IHE Workshop – June 2006
ACC NM Display ExampleACC NM Display ExampleShort
Horizontal
Vertical
Stress
Rest
Stress
Rest
Stress
Rest
Stress
Rest
![Page 22: Nuclear Medicine Image Profile](https://reader035.fdocuments.net/reader035/viewer/2022081401/558d3da6d8b42a0b318b465c/html5/thumbnails/22.jpg)
22IHE Workshop – June 2006IHE Workshop – June 2006
Another ACC Display ExampleAnother ACC Display Example
Different Layout
Same stress and rest views
![Page 23: Nuclear Medicine Image Profile](https://reader035.fdocuments.net/reader035/viewer/2022081401/558d3da6d8b42a0b318b465c/html5/thumbnails/23.jpg)
23IHE Workshop – June 2006IHE Workshop – June 2006
MPR Option: Image DisplaysMPR Option: Image Displays
Adds Image Display requirements to support Review of SPECT NM Images with Multi-planar reconstruction (MPR) Transaxial Sagittal Coronal
![Page 24: Nuclear Medicine Image Profile](https://reader035.fdocuments.net/reader035/viewer/2022081401/558d3da6d8b42a0b318b465c/html5/thumbnails/24.jpg)
24IHE Workshop – June 2006IHE Workshop – June 2006
Result Screen Export OptionResult Screen Export Option
![Page 25: Nuclear Medicine Image Profile](https://reader035.fdocuments.net/reader035/viewer/2022081401/558d3da6d8b42a0b318b465c/html5/thumbnails/25.jpg)
25IHE Workshop – June 2006IHE Workshop – June 2006
Result Screen Export Option:Result Screen Export Option:Evidence CreatorsEvidence Creators
Render Dynamic Result Screens Based on the following DICOM Objects:
• Multi-Frame Grayscale Byte Secondary Capture• Multi-Frame True Color Secondary Capture
Creating multiple single frame objects is NOT sufficient.Render Static Result Screens Using single frame DICOM Secondary Captures is sufficient Using Multi-frame to render single or several related static result
screens is recommended• provides better image comment/label attributes• provides implicit display order• displayed by stepping, not cine
Must generate appropriate Secondary Capture objects from any Result Screens created.
![Page 26: Nuclear Medicine Image Profile](https://reader035.fdocuments.net/reader035/viewer/2022081401/558d3da6d8b42a0b318b465c/html5/thumbnails/26.jpg)
26IHE Workshop – June 2006IHE Workshop – June 2006
Points of Interest - NMPoints of Interest - NM
Appendix E, Volume 1 (Informative) New to Nuclear Medicine? Trouble matching NM to Scheduled Workflow? Start Here.
E.2 NM Workflow – What happens in NM procedures E.3 NM Worklists – How to schedule/model it E.4 NM Data – What to expect in the DICOM E.5 NM Display – How it should be presented
Many subtleties here
![Page 27: Nuclear Medicine Image Profile](https://reader035.fdocuments.net/reader035/viewer/2022081401/558d3da6d8b42a0b318b465c/html5/thumbnails/27.jpg)
27IHE Workshop – June 2006IHE Workshop – June 2006