Non-Invasive Blood Flow Measurem ents Using … · Non-Invasive Blood Flow Measurem ents Using...

of 8 /8
Non-Invasive Blood Flow Measurements Using Ultrasound Modulated Diffused Light N. Racheli a , A. Ron a , Y. Metzger a , I. Breskin a , G. Enden b , M. Balberg a , R. Shechter a1 a Ornim Medical Ltd, Israel b Biomedical Engineering Department, Ben- Gurion University, Israel ABSTRACT Adequate capillary blood flow is a critical parameter for tissue vitality. We present a novel non-invasive method for measuring blood flow based on the acousto-optic effect, using ultrasound modulated diffused light. The benefits of the presented method are: deep tissue sampling (> 1cm), continuous real time measurement, simplicity of apparatus and ease of operation. We demonstrate the ability of the method to measure flow of scattering fluid using a calibrated flow phantom model. Fluid flow was generated by a calibrated syringe pump and the phantom’s sampled volume contained millimeter size flow channels. Results demonstrate linear dependence of flow as measured by the presented technique (CFI) to actual flow values with R 2 =0.91 in the range of 0 to 2 ml/min, and a linear correlation to simultaneous readings of a laser Doppler probe from the same phantom. This data demonstrates that CFI readings provide a non-invasive platform form measuring tissue microcirculatory blood flow. Keyword List: Ultrasound Modulated Light; Capillary Blood Flow; Phantom; Near Infrared; Acousto Optic Effect; 1. Introduction Measuring capillary blood flow is critical for determining tissue vitality (1). Several optical methods for measuring flow already exist. All these methods utilize Near Infrared (NIR) light due to its low absorption in tissue which allows a large penetration depth of up to several centimeters (2) . The two mostly used methods are Laser Doppler (LD) and Diffuse Correlation Spectroscopy (DCS). Both of these methods are based on detecting a change in the phase or frequency of a coherent laser beam that irradiates the tissue. In DCS, the tissue is irradiated with a laser beam and the resulting speckle pattern is detected using photon counting detectors (3) .Flow is extracted by measuring the decorrelation time of the detected light, and fitting it to a theoretical model. In DCS, the measured depth is roughly proportional to the source-detector separation. In LD based methods, a laser beam is directed onto the tissue and is sometimes combined with a reference beam to obtain the Doppler shift caused by the flow (4) . The drawbacks of this method are: high sensitivity to movements, it is limited to shallow (few millimeters) flows and gives only relative measurements. In order to obtain flow measurements at deeper tissue depth, the LD probe is inserted into the tissue. In this work a novel non-invasive method for measuring blood flow based on the acousto-optic effect is presented. For this method, the sampled tissue is illuminated with a highly coherent light source. A slowly travelling ultrasound (US) beam modulates the light through the acousto-optic effect. The light exiting the tissue at a certain distance produces a speckle pattern which is detected with a single photo-detector. The detected light has a spectral component at the US frequency. Blood flow within the sampled volume affects the photons’ temporal correlation and therefore the amplitude of the spectral component at the US frequency decreases as flow increases while the spectral width around the ultrasound frequency broadens. 1 Correspondence to R. Shechter, Ph.D.: E-mail [email protected] Photons Plus Ultrasound: Imaging and Sensing 2012, edited by Alexander A. Oraevsky, Lihong V. Wang, Proc. of SPIE Vol. 8223, 82232A · © 2012 SPIE · CCC code: 1605-7422/12/$18 · doi: 10.1117/12.906342 Proc. of SPIE Vol. 8223 82232A-1 Downloaded from SPIE Digital Library on 23 Mar 2012 to 132.64.36.190. Terms of Use: http://spiedl.org/terms

Embed Size (px)

Transcript of Non-Invasive Blood Flow Measurem ents Using … · Non-Invasive Blood Flow Measurem ents Using...

  • Non-Invasive Blood Flow Measurements Using Ultrasound Modulated Diffused Light

    N. Rachelia, A. Rona, Y. Metzgera, I. Breskina, G. Endenb, M. Balberga, R. Shechtera1

    a Ornim Medical Ltd, Israel

    b Biomedical Engineering Department, Ben- Gurion University, Israel

    ABSTRACT Adequate capillary blood flow is a critical parameter for tissue vitality. We present a novel non-invasive method for measuring blood flow based on the acousto-optic effect, using ultrasound modulated diffused light. The benefits of the presented method are: deep tissue sampling (> 1cm), continuous real time measurement, simplicity of apparatus and ease of operation. We demonstrate the ability of the method to measure flow of scattering fluid using a calibrated flow phantom model. Fluid flow was generated by a calibrated syringe pump and the phantoms sampled volume contained millimeter size flow channels. Results demonstrate linear dependence of flow as measured by the presented technique (CFI) to actual flow values with R2=0.91 in the range of 0 to 2 ml/min, and a linear correlation to simultaneous readings of a laser Doppler probe from the same phantom. This data demonstrates that CFI readings provide a non-invasive platform form measuring tissue microcirculatory blood flow. Keyword List: Ultrasound Modulated Light; Capillary Blood Flow; Phantom; Near Infrared; Acousto Optic Effect;

    1. Introduction Measuring capillary blood flow is critical for determining tissue vitality (1). Several optical methods for measuring flow already exist. All these methods utilize Near Infrared (NIR) light due to its low absorption in tissue which allows a large penetration depth of up to several centimeters (2). The two mostly used methods are Laser Doppler (LD) and Diffuse Correlation Spectroscopy (DCS). Both of these methods are based on detecting a change in the phase or frequency of a coherent laser beam that irradiates the tissue. In DCS, the tissue is irradiated with a laser beam and the resulting speckle pattern is detected using photon counting detectors (3).Flow is extracted by measuring the decorrelation time of the detected light, and fitting it to a theoretical model. In DCS, the measured depth is roughly proportional to the source-detector separation. In LD based methods, a laser beam is directed onto the tissue and is sometimes combined with a reference beam to obtain the Doppler shift caused by the flow (4). The drawbacks of this method are: high sensitivity to movements, it is limited to shallow (few millimeters) flows and gives only relative measurements. In order to obtain flow measurements at deeper tissue depth, the LD probe is inserted into the tissue. In this work a novel non-invasive method for measuring blood flow based on the acousto-optic effect is presented. For this method, the sampled tissue is illuminated with a highly coherent light source. A slowly travelling ultrasound (US) beam modulates the light through the acousto-optic effect. The light exiting the tissue at a certain distance produces a speckle pattern which is detected with a single photo-detector. The detected light has a spectral component at the US frequency. Blood flow within the sampled volume affects the photons temporal correlation and therefore the amplitude of the spectral component at the US frequency decreases as flow increases while the spectral width around the ultrasound frequency broadens.

    1Correspondence to R. Shechter, Ph.D.: E-mail [email protected]

    Photons Plus Ultrasound: Imaging and Sensing 2012, edited by Alexander A. Oraevsky, Lihong V. Wang,Proc. of SPIE Vol. 8223, 82232A 2012 SPIE CCC code: 1605-7422/12/$18 doi: 10.1117/12.906342

    Proc. of SPIE Vol. 8223 82232A-1

    Downloaded from SPIE Digital Library on 23 Mar 2012 to 132.64.36.190. Terms of Use: http://spiedl.org/terms

  • When a lighttrajectories ofthe peak of thWhen the lighdestructively In biological temperature. substantial retrajectories ofLight that traacousto-opticthe US freque An example olight as it travin Figure 1B.the tagged limedium) as s

    Instead of senautocorrelatiotemporal resosignal is Re( Where I(t) is sound velocitlength was 4measured on

    t beam entersf photons thathe photons diht source is coat the detecto

    tissues, the mTherefore, th

    elative to thof the photons avels through c effect. Thereency (7) in add

    of a short traivels through t. The time-basight intensity shown in Figu

    Figure 1

    c) T

    nding a shorton is used. Suolution. The sg(t)) . The a

    the detected ty in the medsec.The totaa synthetic ph

    Fi

    a

    a semi-infinit exit the mediistribution alonoherent, diffe

    or plane, creati

    medium througe speckle patt

    ose induced bthat reach thea medium th

    efore, the specdition to the ra

    in of US pulsethe different msed cross corrwhich corres

    ure 1C. We den

    a) The transmThe temporal cro

    t train of USuch a series pseries is transamplitude of th

    light intensitydium. In the pal series lengthhantom model

    gure 2 An exa

    ite turbid medium at a distanng the axis perent photons ting numerous

    gh which lighttern is varyingby temperature detector decrhat is irradiateckle pattern o

    andom speckle

    es introduced medium layersrelation betwesponds to a cnote this resul

    mitted pulse trainoss correlation

    pulses, a codprovides high

    sformed into ahe UTL signa UTL()y at time t. Th

    presented workh was 5msec.l is presented

    ample of a UTL

    0 50

    2

    4

    6

    8

    10

    12x 10

    4

    ampl

    itude

    [A.U

    ]

    dium, photonsnce d from theerpendicular tothat travel thr bright and da

    t travels is in cg as a functiore, e.g. due treases and resued with an USobtained frome variations.

    into the tissues. The measureen the US pucross section lting profile a

    n. B) The detecbetween the tra

    ded series G(th SNR due toa complex serl at delay is )=tI(t)ghe time cank, the carrier An example in Figure 2.

    L curve measur

    10 15 20depth [mm]

    b

    s scatter manye source are wo the source-drough differenark spots calle

    constant motioon of time. Wo blood flowults in a decreS beam is "ta

    m such a mediu

    e is plotted inred light intenulse train and d

    of the bananUTL curve

    ted light intensansmitted US se

    (t) comprisingo its relativelyries g(t)=Gdefined by: g(t+)2

    n be transformfrequency f oof A UTL cu

    red on a synthet

    25 30

    y times beforwithin a banandetector axis isnt trajectories ed speckles (6).

    on due to cellhen the move

    w, the temporease of specklegged" by the um will have

    n Figure 1A .Tnsity by the dedetected lightna (for a giv

    (Ultrasound

    ity I(t) averagedequence and the

    g of 1 and y long lengthG(t)e 2 f t and

    med into depthof the series wurve obtained

    tic phantom mo

    c

    re exiting the na shape. Ts around d/2 (5interfere cons.

    s or large moements withinral correlatione contrast (6). acoustic wava modulated

    This US train etector is of tht provides a deven speed ofTagged Light

    d over 10,000 te detected light

    -1 bits havh while maintd the transmitt

    h by multiplyiwas 1MHz and with the cod

    odel

    c

    medium. TheThe location o5). structively and

    olecules finiten the tissue aren between the

    ve through thecomponent a

    modulates thehe form shownepth profile of sound in thet curve).

    traces. .

    ving a narrowtaining a highted ultrasound

    ( 1 )

    ing it with thend the code bided series G(t

    e f

    d

    e e e

    e at

    e n f e

    w h d

    e it )

    Proc. of SPIE Vol. 8223 82232A-2

    Downloaded from SPIE Digital Library on 23 Mar 2012 to 132.64.36.190. Terms of Use: http://spiedl.org/terms

  • As mentioned previously, the speckle contrast decreases proportionally to the motion of scatterers in the medium. Therefore, the amplitude of the cross correlation between the US pulse and the intensity of the light, or the amplitude of the UTL curve, is expected to decrease as flow increases.

    2. Materials and methods Experiments were performed on a tissue-mimicking phantom model in order to demonstrate the detection of changes in flow using the proposed method. A tissue mimicking phantom that encapsulates millimeter size flow channels was fabricated and the flow rate of scattering fluid was varied with a syringe pump. The experimental setup used for these experiments is explained in detail in the following sections.

    2.1. Tissue-mimicking Phantom structure A phantom model that mimics blood flow in the tissue was designed. The phantom is made of UltraFlex (Douglas & Sturgess Inc) which is a synthetic polymer matrix soaked with oil. TiO2 particles (0.1% by weight) were added as light scattering agents. The optical and acoustic properties of the phantom are similar to those of tissue as listed in Table 1.

    Table 1- optical and acoustic properties of the phantom and the tissue (2;11) Property Tissue muscle/brain Phantom Light Effective decay coefficient 2.17/2.12 cm-1 2.2 0.2cm-1 Sound Velocity 1.5 105 cm/s 1.43 105 cm/s Acoustic impedance 150-170 Kg/cm2s 149 Kg/cm2s

    Phantom preparation TiO2 powder was mixed with mineral oil and the suspension was inserted into an ultrasonic bath to disintegrate aggregates. UltraFlex was heated to 200oC and stirred with the TiO2 suspension using a magnetic stirrer. The mixture was then poured into a special rectangular mold with 20 parallel fibers (outer diameter of 1mm) serving as a temporary support structure for the channels. After solidification of the phantom, the fibers were pulled out to create empty channels (Figure 3).

    Figure 3 a) A 3D scheme of the phantom and the measuring probe. The optic fibers and US transducer are within the probe. b) the Phantom cross section.

    Phantom assembly The channels are connected to silicon tubes (each channel separately) while each row of channels is connected to a separate valve, as seen in Figure 4. Using these valves flow to each row can be controlled (turned on/off) separately.

    2mm4.5mm

    20 x 1mm

    a b

    Proc. of SPIE Vol. 8223 82232A-3

    Downloaded from SPIE Digital Library on 23 Mar 2012 to 132.64.36.190. Terms of Use: http://spiedl.org/terms

  • Figure 4 Thsilicon tube

    2.2. Blood-mThe fluid thatsuspension odiameter rangspheres of dif

    2.3. Flow mA calibrated cThe syringe pusing a design(8). The experthe mean velo

    2.4. ExperimA schematic through an odeliver it to a The US transsequence intmanufactures

    The US sequtransducer weacoustic beamthe liquid velwere chosen the phantom analysis of thA/D and all o

    hree different ves connected to

    mimicking flut flows throug

    of polystyrenege of 0.4-0.9fferent sizes.

    mechanisms computer-conpump enables nated Lab Vie

    rimental meanocities in the m

    mental setup diagram of th

    optical fiber (6a photo-detecto

    sducer was plato the mediums by Ornim me

    uence was gere placed sucm was 15mm,locity was kein the range owas collected

    he data was peother hardware

    views on the flowchannels. c) A

    uid gh the channele spheres (70m) was chos

    ntrolled syringaccurate and

    ew program. Pn velocities (camicro-vascula

    assembly he experimen62.5m, NA or (bundle of

    aced between m. The two edical.

    enerated usinch that the flo, and the centpt constant fo

    of 0 to 2.5 mmd into a drain rerformed usine parts are a p

    a w phantom. a) A

    A general view o

    ls should have0% purified Hsen. The scatt

    ge pump (Chemreproducible g

    Parameters usealculated by acar bed (0.2-2.5

    ntal set-up is p0.27). A seco12*200m, N

    the receivingoptical fiber

    Figure 5

    ng a computeow channels atral ultrasoundor all experimm/sec to matchreservoir. The

    ng MATLAB art of Ornim

    A general viewof the valves wh

    e scattering opH2O; 10% Gltering propert

    mix, Model Ingeneration of ed in the expeccounting for

    5 mm/s for cap

    presented in Fond optic fibe

    NA=0.22).

    g (Rx) and tranrs and the U

    The experime

    er controlled are within thed frequency w

    ments accordinh the mean floe photodetectosoftware as ds monitor.

    b w of the phantomhich controls th

    ptical propertiycerol; 20% ties of this so

    nfusion 100) wdesired flow r

    eriments correthe known cr

    pillary flow (9;

    Figure 5. A lier was used t

    nsmitting (Tx) US transducer

    ental setup

    function gene sampled voluwas 1MHz. Seng to the activow velocity inors signals wdescribed in 3

    m with a probe he flow into the

    ies similar to tSpherotechInc

    olution can be

    was used to gerates. The syr

    esponded to avross sections o; 8).

    ight beam wato collect the

    optical fibersr were assem

    nerator. The ume of the syeveral flow vave channels cn small blood

    were sampled w. The photo-d

    c

    on top. b) A clodifferent rows

    those of bloodc Polystyrene e precisely co

    enerate variouringe pump wavailable physioof the channel

    as delivered toexiting scatte

    s and transmitmbled togethe

    optical fibersystem. The dalues were gecross section.vessels (9; 8). Twith an A/D cdetector, funct

    ose look at the of channels.

    d. Therefore, aspheres with

    ontrolled using

    us flow rates. as controlled ological data s), matched

    o the phantomered light and

    tted a pressureer in a probe

    s and the USiameter of theenerated while These values

    The flow fromconverter. Thetion generator

    a h g

    m d

    e e

    S e e s

    m e r,

    Proc. of SPIE Vol. 8223 82232A-4

    Downloaded from SPIE Digital Library on 23 Mar 2012 to 132.64.36.190. Terms of Use: http://spiedl.org/terms

  • 2.5. Experimental procedure Experiments were carried out with different flow rates. In each session a chosen flow rate was set and fluid flow was kept constant for 2 minutes. Six experiments were performed to test the sensitivity to fluid flow in deeper channels. In the first experiment all channels were activated. In the second experiment, the shallowest row of channels (4.5mm depth) was activated and in the third experiment the central row of channels (8.5mm depth) was activated.

    3. Data analysis and results

    3.1. Extracting the Flow Index from the UTL curve As a first step, the UTL curve (Figure 6) is normalized to the average light intensity (DC value) to account for the effect of the light intensity on the amplitude of the cross correlation. Next, the Flow Index (CFI) is calculated from an interest range k to k+N of the normalized UTL curved:

    ( ) = ( ) ( ) ( 2 ) where t is the discrete recording time.

    Figure 6 An example of a UTL curve. The solid line over the dashed UTL curve is the interest range from k to k+N

    The chosen interest range is the range over which the UTL curve is most sensitive to flow variations. To find this range, a linear regression between CFI and the real velocity rates was calculated for different interest ranges for the configuration with all channels active. The interest range length (N) was chosen to be 5mm.

    3.2. Experimental results Experiments were performed with a sequence of ten flow rates in the range of 0 to 450 ml/min. The same sequence was repeated while different channels were activated, as mentioned in section 2.5. As expected, the amplitude of the UTL curves decreased as the flow rate through channel increased. In Figure 7 the average UTL curves for different flow rates are presented. The most sensitive interest range was in the range of 8 to 13 mm. This interest range was used for all future CFI calculations.

    0 5 10 15 20 25 300

    2

    4

    6

    8

    10

    12x 104

    depth [mm]

    ampl

    itude

    [A.U

    ]

    N

    K

    Proc. of SPIE Vol. 8223 82232A-5

    Downloaded from SPIE Digital Library on 23 Mar 2012 to 132.64.36.190. Terms of Use: http://spiedl.org/terms

  • Figure 7 The UTL curves averaged over 2 min periods for three different flow rates.

    After choosing the most sensitive interest range, CFI was calculated over 2 minutes for each flow rate. Abnormal UTL frames were rejected before calculating the CFI at time points where the DC light intensity exceeded 2.5 standard deviations from its mean value. As a robust estimator, the median value of CFI was calculated. The results for three configurations of active channels: flow in all channels, flow only in upper row of channels (row a) and flow only in central row of channels (row c) are plotted in Figure 8. The CFI linearly depends on the flow velocity. Different slopes were obtained for different depths since light decrease exponentially in the phantom and thus, the portion of photons scattered from moving particles relative to the overall light intensity decreases as the depth of flow increases.

    Figure 8 - CFI vs. Flow Velocity for three channels configuration.

    For evaluation of CFI, a comparison to Laser Doppler was performed. A Moor Instrument Laser Doppler DRT4, was used with a DP3b probe. The probe was positioned at the distance of approximately 1cm from the prototype probe. For each flow level generated by the syringe pump we obtained an averaged reading of the Laser Doppler system and Ornims system. The correlation between the laser Dopplers flow index and CFI is presented in Figure 9. The correlation coefficient (r2) between CFI and Laser Dopplers Flow Index was 0.94.

    0 5 10 15 20 25 30-2

    0

    2

    4

    6

    8

    10

    12x 10

    4

    depth [mm]

    ampl

    itude

    [A.U

    ]

    0 l/min 750 l/min1750 l/min

    0 0.5 1 1.5 2 2.5-0.5

    0

    0.5

    1

    1.5

    2

    2.5

    3

    3.5

    4

    4.5

    velocity [mm/sec]

    CFI

    [A.U

    ]

    All rowsRow cRow a

    Proc. of SPIE Vol. 8223 82232A-6

    Downloaded from SPIE Digital Library on 23 Mar 2012 to 132.64.36.190. Terms of Use: http://spiedl.org/terms

  • Figure 9 - CFI vs. Laser Dopplers Flow Index

    In order to determine the effect of the concentration of scattering particles on CFI the scattering fluid described in section 2.2 was diluted with purified water to reduce the concentration of scattering particles from 20% to 13%. As CFIs principles are similar to laser Dopplers theory, CFI depends on both the scatterers velocity and concentration. For two trials, where all channels were activated, the velocity values were multiplied by the concentration, and the data was combined together, as presented in Figure 10. The regression of combined results shows good linearity with r2 of 0.87.

    Figure 10 - CFI vs. Flow Velocity times concentration for the two scattering particles concentration.

    4. Discussion This study demonstrates a very good linear correlation between the calculated Flow Index (CFI), defined by the UTL curves, and the actual flow rates in an artificial flow phantom. In the experimental setup presented herein, the velocity is linearly proportional to the flow rates. Therefore, this data also shows a linear correlation between CFI to velocity, very similar to that obtained with Laser Doppler, in addition to a good correlation to Laser Doppler readings from the same phantom. Accounting for variations in the concentration of the scattering elements, we show that the calculated CFI depends linearly on the product of concentration*velocity, as is the case for Laser Doppler measurements (10). This preliminary data shows that CFI depends on the flow of scattering particles through the measured volume, similar to the flux quantity measured by Laser Doppler.

    0 100 200 300 400 500 600-0.5

    0

    0.5

    1

    1.5

    2

    2.5

    3

    3.5

    Laser Doppler Flow Index [A.U]

    CFI

    [A.U

    ]

    0 0.005 0.01 0.015 0.02 0.0250

    0.5

    1

    1.5

    2

    2.5

    3

    3.5

    4

    velocity * concentration [mm/sec*g/ml]

    CFI

    [A.U

    ]

    High concentrationLow concentration

    Proc. of SPIE Vol. 8223 82232A-7

    Downloaded from SPIE Digital Library on 23 Mar 2012 to 132.64.36.190. Terms of Use: http://spiedl.org/terms

  • The described experimental setup demonstrates the linear dependence of CFI on flow of scattering particles through the channels, but is not used to calibrate the CFI dependence on flow in live tissue. This is due to the different decorrelation time of the phantoms solid matrix and due to the different coupling efficiency between the probe and the phantom relative to those of tissue. For clinical application CFI depends on several parameters, including the US coupling efficiency between the transducer and the skin, the decorrelation time of the tissue and on the concentration of moving particles inside the measured volume.

    5. Conclusions A novel method for continuously and non invasively measuring flow in deep tissue based on ultrasound modulated diffused light is presented. The data demonstrates a linear correlation of CFI readings to flow rates in channels that are deeper than 1cm in a synthetic phantom. The CFI readings are shown to depend on the concentration of scattering centers within the channels. A very good correlation to Laser Doppler readings from the same phantom is also demonstrated. CFI readings provide a non-invasive platform form measuring tissue microcirculatory blood flow.

    6. Bibliography 1. Louis, Jean Vincent and De Backer, Daniel. Microvascular dysfunction as a cause of organ dysfunction in severe sepsis, Critical Care , pp. s9-s12 (2005). 2. Welch, Ashley J., Prahl, Scott A. and Wai-Fung, Cheong. A Review of the Optical Properties of Biological Tissues, IEEE Journal Of Quantum Electronics, VOL. 26. No. 12, December, pp. 2166-2185 (1990) 3. Cheung, Cecil, et al. In vivo cerebrovascular measurement combining diffuse near-infrared absorption and correlation spectroscopies, Phys. Med. Biol. 46, pp. 20532065 (2001) 4. Fredriksson I., Fors C. and Johansson J., "Laser Doppler Flowmetry a Theoretical Framework," Department of Biomedical Engineering, Linkping University, www.imt.liu.se/bit/ldf/ldfmain.html (2007) 5. Guoqiang Yu, Turgut Durduran, Gwen Lech, Chao Zhou, Britton Chance, Emile R. Mohler III, and Arjun G. Yodh, J. Time-dependent blood flow and oxygenation in human skeletal muscles measured with noninvasive near-infrared diffuse optical spectroscopie Biomed. Opt. 10, 024027 (2005) 6. Yodh, Arjun G. and Boas, David A. Spatially varying dynamical properties of turbid media probed with diffusing temporal light correlation, J. Opt. Soc. Am. A/ Vol. 14, No. 1/January, pp. 192-215 (1997) 7. Uzgiris, E., et al. Ultrasonic tagging of light: Theory, Applied Physical Sciences, Vol 95, pp. 14015-14019 (1998) 8. Guyton, Arthur C. and Hall, Jhon E. [Medical Physiology] Elsevier Health Sciences , pp. 150-158. (2005) 9. Schaller, B. Physiology of cerebral venous blood flow: from experimental data in animals to normal function in humans, Brain Research Reviews 46, pp. 243-260. (2004) 10. Nilsson, E. Gert., Tenland, Torsten and Ake, P. Oberg. Evaluation of a Laser Doppler Flowmeter for Measurement of Tissue Blood Flow, IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, pp. 597-604. (1980) 11. Christensen, Douglas A. [Ultrasonic Bioinstrumentation] Wiley , (1988)

    Proc. of SPIE Vol. 8223 82232A-8

    Downloaded from SPIE Digital Library on 23 Mar 2012 to 132.64.36.190. Terms of Use: http://spiedl.org/terms