Non-Invasive Non-Invasive Assessment of Assessment of Prosthetic Valves Prosthetic Valves November 21, 2007 November 21, 2007 Alex Morss Alex Morss

Transcript of Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November...

Page 1: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.

Non-Invasive Non-Invasive Assessment of Assessment of

Prosthetic ValvesProsthetic Valves

November 21, 2007November 21, 2007

Alex MorssAlex Morss

Page 2: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.

Terms and definitionsTerms and definitions

Prosthetic valves: manmade Prosthetic valves: manmade replacements of heart valves, either replacements of heart valves, either mechanical (nonorganic) or bioprosthetic mechanical (nonorganic) or bioprosthetic (organic materials used in construction, (organic materials used in construction, i.e. animal or human parts)i.e. animal or human parts)

Assessment: visual evaluationAssessment: visual evaluation Non-invasive: taking pictures of hidden Non-invasive: taking pictures of hidden

things without using sharp objects (i.e., things without using sharp objects (i.e., essentially nonviolent means of seeing essentially nonviolent means of seeing them undressed)them undressed)

Page 3: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.

Main Characters Main Characters (Protagonists)(Protagonists)

Mechanical ValvesMechanical Valves Ball-cage Ball-cage Tilting discTilting disc BileafletBileaflet

Bioprosthetic ValvesBioprosthetic Valves StentedStented Unstented (aortic root)Unstented (aortic root) Homograft, allograftHomograft, allograft

Page 4: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.

The AntagonistsThe Antagonists

Endocarditis +/- vegetationsEndocarditis +/- vegetations Paravalvular leak (+/- dehiscence)Paravalvular leak (+/- dehiscence) ThrombosisThrombosis Pannus formationPannus formation Degeneration, +/- calcific stenosis or Degeneration, +/- calcific stenosis or

leaflet tearleaflet tear Strut fracture, disc embolizationStrut fracture, disc embolization Unseating of valveUnseating of valve

Page 5: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.

The ArsenalThe Arsenal

Roentgenography (X-Roentgenography (X-Ray)Ray)

FluoroscopyFluoroscopy Cardiac MRI (?)Cardiac MRI (?) EchocardiographyEchocardiography

TTE will often best allow TTE will often best allow optimal Doppler anglesoptimal Doppler angles

TEE will often best TEE will often best allow optimal direct allow optimal direct visualizationvisualization

IntangiblesIntangibles

Page 6: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.

The Main CharactersThe Main Characters Mechanical ValvesMechanical Valves

Ball-cage (Starr-Ball-cage (Starr-Edwards)Edwards)

Tilted-disc (Bjork-Tilted-disc (Bjork-Shiley, Medtronic-Hall)Shiley, Medtronic-Hall)

* Note: the convexoconcave * Note: the convexoconcave version of the Bjork-Shiley version of the Bjork-Shiley valve earned a bad name due valve earned a bad name due to cases of strut fracture and to cases of strut fracture and disc embolizationdisc embolization

Bileaflet (St. Jude, Bileaflet (St. Jude, Carbomedics)Carbomedics)

All valves are sized by All valves are sized by diameter, 19-33 mmdiameter, 19-33 mm

O’Neill NEJM 1995

Page 7: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.

Main Characters, Part Main Characters, Part DeuxDeux

Bioprosthetic Bioprosthetic ValvesValves Stented: Carpentier-Stented: Carpentier-

Edwards, Hancock, Edwards, Hancock, Ionescu-Shiley, St. Ionescu-Shiley, St. Jude MosaicJude Mosaic

Stentless: Biocor.Stentless: Biocor. Homografts/Homografts/

autografts: may not autografts: may not be able to detect be able to detect noninvasivelynoninvasively

Also sized by Also sized by diameter, 19-33 mmdiameter, 19-33 mm

Hancock

Mosaic

Stentless porcine

Page 8: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.

Unveiling the arsenalUnveiling the arsenalOld School: RoentgenographyOld School: Roentgenography

Mechanical and many Mechanical and many stented bioprosthetic stented bioprosthetic valves are radiopaque, valves are radiopaque, allowing allowing determination of valve determination of valve type and position on type and position on chest X-ray.chest X-ray.

May be used to assess May be used to assess for device fracture in for device fracture in some casessome cases Starr-Edwards valve seen in aortic position

in 1965. From Nery, Heart 2004

Page 9: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.

Up a Notch: FluoroscopyUp a Notch: Fluoroscopy Best methodology to Best methodology to

assess mechanical assess mechanical leaflet motion due to leaflet motion due to outstanding spatial outstanding spatial and temporal and temporal resolution.resolution.

May be used to May be used to assess stability of assess stability of valve ring with the valve ring with the cardiac cyclecardiac cycle

May optimally May optimally position angle to best position angle to best assess subtle fractureassess subtle fracture

O’Neill NEJM 1995

Page 10: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.

Disc embolizationDisc embolization

O’Neill NEJM 1995

disc

Page 11: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.

Cardiac MRICardiac MRI

Cardiac MRI may Cardiac MRI may visualize mechanical visualize mechanical valves, but lacks the valves, but lacks the temporal resolution temporal resolution and Doppler and Doppler capablities of capablities of echocardiographyechocardiography

May show gross May show gross valve position, valve position, function, and function, and regurgitationregurgitation

Page 12: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.

More cardiac MRIMore cardiac MRI

Page 13: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.
Page 14: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.

EchocardiographyEchocardiography TransthoracicTransthoracic

Allows assessment of valve area and regurgitation Allows assessment of valve area and regurgitation via Doppler, which is generally adequate to via Doppler, which is generally adequate to exclude significant obstructive or regurgitant exclude significant obstructive or regurgitant change. Flow velocity is the crucial measurement. change. Flow velocity is the crucial measurement.

Inadequate to assess infection or small structural Inadequate to assess infection or small structural changes (e.g. strut fracture, small vegetation, changes (e.g. strut fracture, small vegetation, paravalvular leak)paravalvular leak)

TransesophagealTransesophageal Ideal for visual inspection of valve apparatus and Ideal for visual inspection of valve apparatus and

seating; may not accurately quantify valve flow seating; may not accurately quantify valve flow velocities. May directly measure aortic valve area velocities. May directly measure aortic valve area via planimetryvia planimetry

Page 15: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.

Echo by valve positionEcho by valve position AorticAortic

Accurate TTE assessment relies on accurate Doppler Accurate TTE assessment relies on accurate Doppler assessments in multiple viewsassessments in multiple views

Often many TTE views partially obscured by shadowing. Often many TTE views partially obscured by shadowing. Often TEE required to view leaflets Often TEE required to view leaflets

MitralMitral Among the best positions for TTE visualization, usually Among the best positions for TTE visualization, usually

able to see leaflets via apical viewsable to see leaflets via apical views TricuspidTricuspid

Also usually adequately visualized by TTE directly and Also usually adequately visualized by TTE directly and via Dopplervia Doppler

PulmonicPulmonic Rarest position for valve replacement. Difficult to Rarest position for valve replacement. Difficult to

visualize for both TTE and TEE, no clear advantagevisualize for both TTE and TEE, no clear advantage

Page 16: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.

Aortic ProsthesesAortic Prostheses

Focus on Doppler imaging of aortic outflows to Focus on Doppler imaging of aortic outflows to determine mean and peak gradientsdetermine mean and peak gradients

Can identify prosthesis type by direct Can identify prosthesis type by direct visualizationvisualization

As with all prostheses, need to know their SIZE As with all prostheses, need to know their SIZE to allow assessment of normal vs. to allow assessment of normal vs. pathologically increased transvalvular pathologically increased transvalvular gradient. gradient.

Size varies from 19-29 mm in diameterSize varies from 19-29 mm in diameter Normal gradients for each valve type and size Normal gradients for each valve type and size

may be found on reference tablesmay be found on reference tables

Page 17: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.

Normal aortic bileaflet Normal aortic bileaflet valvevalve

Page 18: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.
Page 19: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.
Page 20: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.
Page 21: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.

Peak gradient 20mmHg, mean 12 mmHgPeak gradient 20mmHg, mean 12 mmHg Normal bileaflet gradients are dependent Normal bileaflet gradients are dependent

on valve sizeon valve size

Page 22: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.

Normal mechanical aortic Normal mechanical aortic valve peak gradientsvalve peak gradients

BileafletBileaflet 19mm: 33 +/- 1119mm: 33 +/- 11 29mm: 13 +/- 529mm: 13 +/- 5

Tilting discTilting disc 19mm: 4619mm: 46 29mm: 12+/- 829mm: 12+/- 8

Ball-cageBall-cage 23mm: 33+/-1323mm: 33+/-13 29mm: 29+/-929mm: 29+/-9

Full tables are available in echo texts or on fellowship echo website

Feigenbaum 2005

Approximate flow velocities:

19mm 2.9 m/sec

29 mm 1.9 m/sec

(+/- 0.5 m/sec)

Page 23: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.

Normal aortic Normal aortic bioprosthetic valvebioprosthetic valve

Page 24: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.
Page 25: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.
Page 26: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.

Peak gradient 17mm- need valve size and ideally baseline Peak gradient 17mm- need valve size and ideally baseline gradient at time of valve implantation to assess for normal gradient at time of valve implantation to assess for normal valuevalue

May also use continuity equation with measurement of May also use continuity equation with measurement of LVOT to calculate effective aortic valve areaLVOT to calculate effective aortic valve area

Page 27: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.

Tilting disc AVRTilting disc AVR

Page 28: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.
Page 29: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.
Page 30: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.
Page 31: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.
Page 32: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.

Ball-cage AVRBall-cage AVR

Page 33: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.
Page 34: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.
Page 35: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.
Page 36: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.

Name the AVR typeName the AVR type

Page 37: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.
Page 38: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.
Page 39: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.

Aortic bioprosthetic valve with paravalvular Aortic bioprosthetic valve with paravalvular leak, valvular regurgitationleak, valvular regurgitation

Page 40: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.

What’s wrong?What’s wrong?

Page 41: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.
Page 42: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.
Page 43: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.
Page 44: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.
Page 45: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.
Page 46: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.

Tilting disc valve with thrombosis causing partial obstruction

Talley, Can J Card 1986

Page 47: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.

Mitral prosthetic valvesMitral prosthetic valves

Better visualization on TTEBetter visualization on TTE Much lower normal gradients than Much lower normal gradients than

aortic valves due to lower flow aortic valves due to lower flow velocities and larger size overallvelocities and larger size overall

Sizes generally vary from 25-33mm Sizes generally vary from 25-33mm in diameterin diameter

Page 48: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.

Normal bioprosthetic Normal bioprosthetic MVRMVR

Page 49: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.
Page 50: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.
Page 51: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.

Peak mitral velocity of 1.30 m/sec, peak gradient 8 mmHg, Peak mitral velocity of 1.30 m/sec, peak gradient 8 mmHg, mean 5mmHg (normal ~ 5 depending on valve size)mean 5mmHg (normal ~ 5 depending on valve size)

Page 52: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.

Bileaflet MVRBileaflet MVR

Page 53: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.
Page 54: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.
Page 55: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.

Appropriate regurgitation shown- 2 jets in bileaflet valveAppropriate regurgitation shown- 2 jets in bileaflet valve By design in mechanical valves to minimize thrombosis By design in mechanical valves to minimize thrombosis

riskrisk

Page 56: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.
Page 57: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.

Ball-cage MVRBall-cage MVR

Page 58: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.
Page 59: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.
Page 60: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.

Cage-ball valves have the highest nomalvalve gradients due to design

Page 61: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.
Page 62: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.
Page 63: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.

Mean gradient 9 mmHgMean gradient 9 mmHg Standard mean gradient for ball-cage Standard mean gradient for ball-cage

MVR is 5-7 +/- 3, depending on sizeMVR is 5-7 +/- 3, depending on size

Page 64: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.

Mitral endocarditisMitral endocarditis

Kort, JASE 2006 (from Gelfand)

Page 65: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.

Mitral unpleasantries, Mitral unpleasantries, part IIpart II

Page 66: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.
Page 67: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.
Page 68: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.
Page 69: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.
Page 70: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.

A league of its ownA league of its own

Page 71: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.
Page 72: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.
Page 73: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.
Page 74: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.
Page 75: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.

SummarySummary Valve type and position often easily determined on CXRValve type and position often easily determined on CXR Fluoroscopy optimal for assessing mechanical valve Fluoroscopy optimal for assessing mechanical valve

leaflet motion, unable to see bioprosthetic leafletsleaflet motion, unable to see bioprosthetic leaflets TTE may identify flow velocities and gross structure, TTE may identify flow velocities and gross structure,

best suited for MVR and exclusion of obstruction due to best suited for MVR and exclusion of obstruction due to good Doppler views. Doppler velocities and valve good Doppler views. Doppler velocities and valve gradients are likely better than calculation of valve gradients are likely better than calculation of valve area due fewer variablesarea due fewer variables

TEE allows better visualization of smaller structural TEE allows better visualization of smaller structural changes, vegetations, and paravalvular leaks. It is the changes, vegetations, and paravalvular leaks. It is the study of choice for concerns of endocarditis in any study of choice for concerns of endocarditis in any prosthetic valve.prosthetic valve.

CMR limited role and not generally used to assess CMR limited role and not generally used to assess prosthetic valves.prosthetic valves.

Page 76: Non-Invasive Assessment of Prosthetic Valves Non-Invasive Assessment of Prosthetic Valves November 21, 2007 Alex Morss.

Thank you!