NIRS – Getting started
US Dairy Forage Research Center2004 Consortium Annual Meeting
![Page 2: NIRS – Getting started · History 1800 The First NIR Spectrum recorded (Herschel) 1950-1960 Potential of NIR quantitative analysis was recognized (Kaye) 1960s Research program at](https://reader036.fdocuments.net/reader036/viewer/2022070913/5fb4d94e9fe8f10ff74541e3/html5/thumbnails/2.jpg)
Terminology
NIR = Near Infrared
or
NIR = Near Infrared Reflectance
NIT = Near Infrared Transmittance
![Page 3: NIRS – Getting started · History 1800 The First NIR Spectrum recorded (Herschel) 1950-1960 Potential of NIR quantitative analysis was recognized (Kaye) 1960s Research program at](https://reader036.fdocuments.net/reader036/viewer/2022070913/5fb4d94e9fe8f10ff74541e3/html5/thumbnails/3.jpg)
History1800 The First NIR Spectrum recorded (Herschel)1950-1960 Potential of NIR quantitative analysis was recognized (Kaye) 1960s Research program at USDA (Norris) for NIR analysis of
agricultural commodities1971 First Generation of NIR Instruments - Reflectance 1975 Second Generation of NIR Instruments - More Stable
Electronics and improved Optics1977-1978 Third Generation of NIR Instruments - Microprocessors, Small
Computers1982-1983 Fourth Generation of NIR Instruments – Scanning
Monochromators1980s Principal components calibration techniques introduced 1987 Fifth Generation of NIR Instruments - Transmittance1990s Nonlinear calibration methods (Local, ANN)
![Page 4: NIRS – Getting started · History 1800 The First NIR Spectrum recorded (Herschel) 1950-1960 Potential of NIR quantitative analysis was recognized (Kaye) 1960s Research program at](https://reader036.fdocuments.net/reader036/viewer/2022070913/5fb4d94e9fe8f10ff74541e3/html5/thumbnails/4.jpg)
NIR MIR FIR
Gamma X-rays UV Visibile Infrared Microwave Radio
Near Infrared (NIR) Overtones, Combination 700-2500
Middle Infrared (MIR) Fundamental vibration 2500-5x104
Far Infrared (FIR) Rotation 5x104-106
NIT NIR
700 1100 2500
Region Characteristic wavelenght
Range (nm)
![Page 5: NIRS – Getting started · History 1800 The First NIR Spectrum recorded (Herschel) 1950-1960 Potential of NIR quantitative analysis was recognized (Kaye) 1960s Research program at](https://reader036.fdocuments.net/reader036/viewer/2022070913/5fb4d94e9fe8f10ff74541e3/html5/thumbnails/5.jpg)
Trnsmission/Reflectance
Trasmission (NIT) Reflectance (NIR)
ITEC169A ITEC168A
Source of light
Detector
Source of light
Detector
![Page 6: NIRS – Getting started · History 1800 The First NIR Spectrum recorded (Herschel) 1950-1960 Potential of NIR quantitative analysis was recognized (Kaye) 1960s Research program at](https://reader036.fdocuments.net/reader036/viewer/2022070913/5fb4d94e9fe8f10ff74541e3/html5/thumbnails/6.jpg)
Absorption units
RAUUnitsAbsorbance 1log)(_ =
201.0
==
AUR
0IaI .1
bI .1
RAbsorptionIII ba
−=
+=
1
)Reflect_(R0
11
![Page 7: NIRS – Getting started · History 1800 The First NIR Spectrum recorded (Herschel) 1950-1960 Potential of NIR quantitative analysis was recognized (Kaye) 1960s Research program at](https://reader036.fdocuments.net/reader036/viewer/2022070913/5fb4d94e9fe8f10ff74541e3/html5/thumbnails/7.jpg)
NIR absorption
- CH
- OH
- NH
- CH
- OH
- NH
Absorbance is mainly due to Hydrogen bonds
![Page 8: NIRS – Getting started · History 1800 The First NIR Spectrum recorded (Herschel) 1950-1960 Potential of NIR quantitative analysis was recognized (Kaye) 1960s Research program at](https://reader036.fdocuments.net/reader036/viewer/2022070913/5fb4d94e9fe8f10ff74541e3/html5/thumbnails/8.jpg)
![Page 9: NIRS – Getting started · History 1800 The First NIR Spectrum recorded (Herschel) 1950-1960 Potential of NIR quantitative analysis was recognized (Kaye) 1960s Research program at](https://reader036.fdocuments.net/reader036/viewer/2022070913/5fb4d94e9fe8f10ff74541e3/html5/thumbnails/9.jpg)
![Page 10: NIRS – Getting started · History 1800 The First NIR Spectrum recorded (Herschel) 1950-1960 Potential of NIR quantitative analysis was recognized (Kaye) 1960s Research program at](https://reader036.fdocuments.net/reader036/viewer/2022070913/5fb4d94e9fe8f10ff74541e3/html5/thumbnails/10.jpg)
Light - Matter interaction
Absorption
Reflection
Trasmission
![Page 11: NIRS – Getting started · History 1800 The First NIR Spectrum recorded (Herschel) 1950-1960 Potential of NIR quantitative analysis was recognized (Kaye) 1960s Research program at](https://reader036.fdocuments.net/reader036/viewer/2022070913/5fb4d94e9fe8f10ff74541e3/html5/thumbnails/11.jpg)
Light - Matter interaction
Wavelength
Trasmission Absorption
![Page 12: NIRS – Getting started · History 1800 The First NIR Spectrum recorded (Herschel) 1950-1960 Potential of NIR quantitative analysis was recognized (Kaye) 1960s Research program at](https://reader036.fdocuments.net/reader036/viewer/2022070913/5fb4d94e9fe8f10ff74541e3/html5/thumbnails/12.jpg)
Undried Unground
Dried Ground
Dried Unground
![Page 13: NIRS – Getting started · History 1800 The First NIR Spectrum recorded (Herschel) 1950-1960 Potential of NIR quantitative analysis was recognized (Kaye) 1960s Research program at](https://reader036.fdocuments.net/reader036/viewer/2022070913/5fb4d94e9fe8f10ff74541e3/html5/thumbnails/13.jpg)
Foss NIRSystem 5000-6500
Lab instrument
The most used instrument in agriculture
There are severalForage and /or FeedNetwork based on thisinstrument
![Page 14: NIRS – Getting started · History 1800 The First NIR Spectrum recorded (Herschel) 1950-1960 Potential of NIR quantitative analysis was recognized (Kaye) 1960s Research program at](https://reader036.fdocuments.net/reader036/viewer/2022070913/5fb4d94e9fe8f10ff74541e3/html5/thumbnails/14.jpg)
Foss NIRSystem 5000-6500Grating
light
Mirror
Detectors Sample
•Moving parts•Pre-dispersive•PbS detectors•Slow scanning (1 sec)•Highly accurate•400-1098; 1100-2498nm•Every 2nm
![Page 15: NIRS – Getting started · History 1800 The First NIR Spectrum recorded (Herschel) 1950-1960 Potential of NIR quantitative analysis was recognized (Kaye) 1960s Research program at](https://reader036.fdocuments.net/reader036/viewer/2022070913/5fb4d94e9fe8f10ff74541e3/html5/thumbnails/15.jpg)
Zeiss - Corona NIR 45• NO moving parts• Post-dispersive• InGaAs detector• 128 diodes• Fast scanning (10 ms)• Very accurate• 900-1700nm• Every 6nm
![Page 16: NIRS – Getting started · History 1800 The First NIR Spectrum recorded (Herschel) 1950-1960 Potential of NIR quantitative analysis was recognized (Kaye) 1960s Research program at](https://reader036.fdocuments.net/reader036/viewer/2022070913/5fb4d94e9fe8f10ff74541e3/html5/thumbnails/16.jpg)
•Diagnostics procedures:They are required to make sure all of the parts are working properly•Weekly diagnostics•Daily check cell
•Standardization procedures•Sealed cups with different products•Repeated periodically (once a year)
Foss NIRSystem 5000-6500
![Page 17: NIRS – Getting started · History 1800 The First NIR Spectrum recorded (Herschel) 1950-1960 Potential of NIR quantitative analysis was recognized (Kaye) 1960s Research program at](https://reader036.fdocuments.net/reader036/viewer/2022070913/5fb4d94e9fe8f10ff74541e3/html5/thumbnails/17.jpg)
Zeiss - Corona NIR 45
•Diagnostics procedures:There are no diagnostics. No movingparts reduce variation in performances•Daily check cell? We think it would benecessary to ensure the instrument isworking properly
•Standardization procedures•There are no procedures•Solved by calibration
![Page 18: NIRS – Getting started · History 1800 The First NIR Spectrum recorded (Herschel) 1950-1960 Potential of NIR quantitative analysis was recognized (Kaye) 1960s Research program at](https://reader036.fdocuments.net/reader036/viewer/2022070913/5fb4d94e9fe8f10ff74541e3/html5/thumbnails/18.jpg)
Spectra
Reference data
Calibration Model (equation)
WhatWhat isis a a CalibrationCalibration??
![Page 19: NIRS – Getting started · History 1800 The First NIR Spectrum recorded (Herschel) 1950-1960 Potential of NIR quantitative analysis was recognized (Kaye) 1960s Research program at](https://reader036.fdocuments.net/reader036/viewer/2022070913/5fb4d94e9fe8f10ff74541e3/html5/thumbnails/19.jpg)
Spectrum of anunknown sample Calibration
Prediction
Use of a CalibrationUse of a Calibration
![Page 20: NIRS – Getting started · History 1800 The First NIR Spectrum recorded (Herschel) 1950-1960 Potential of NIR quantitative analysis was recognized (Kaye) 1960s Research program at](https://reader036.fdocuments.net/reader036/viewer/2022070913/5fb4d94e9fe8f10ff74541e3/html5/thumbnails/20.jpg)
Sample
SPECTRUM
0.04 0.09 0.14 0.16 0.26 0.54 0.55 .059
DRY MATTER AS IS GH TPROTEIN 51.42 46.28 1.25 2.13 FAT 1.96 1.76 1.25 2.13 FIBER 5.4 4.86 1.25 2.13 DM 100.00 90.00 1.25 2.13 MOISTURE 0.0 10.00
Routine Analysis
![Page 21: NIRS – Getting started · History 1800 The First NIR Spectrum recorded (Herschel) 1950-1960 Potential of NIR quantitative analysis was recognized (Kaye) 1960s Research program at](https://reader036.fdocuments.net/reader036/viewer/2022070913/5fb4d94e9fe8f10ff74541e3/html5/thumbnails/21.jpg)
Y1 Y2 Y3…12.8 97.4 43.211.5 96.2 45.6
10.7 94.3 48.0
10.5 96.7 45.2
9.7 98.3 47.9
8.4 96.5 44.1
7.5 95.0 46.77.2 93.2 41.4
Regression
Y=b0+b1x1+b2x2+…+bnxn
Simple, multiple, multivariate, neural network
What is Calibration?
![Page 22: NIRS – Getting started · History 1800 The First NIR Spectrum recorded (Herschel) 1950-1960 Potential of NIR quantitative analysis was recognized (Kaye) 1960s Research program at](https://reader036.fdocuments.net/reader036/viewer/2022070913/5fb4d94e9fe8f10ff74541e3/html5/thumbnails/22.jpg)
Simple and multiple Calibration
![Page 23: NIRS – Getting started · History 1800 The First NIR Spectrum recorded (Herschel) 1950-1960 Potential of NIR quantitative analysis was recognized (Kaye) 1960s Research program at](https://reader036.fdocuments.net/reader036/viewer/2022070913/5fb4d94e9fe8f10ff74541e3/html5/thumbnails/23.jpg)
Simple and multiple Calibration
![Page 24: NIRS – Getting started · History 1800 The First NIR Spectrum recorded (Herschel) 1950-1960 Potential of NIR quantitative analysis was recognized (Kaye) 1960s Research program at](https://reader036.fdocuments.net/reader036/viewer/2022070913/5fb4d94e9fe8f10ff74541e3/html5/thumbnails/24.jpg)
Population structure
Chemical variation
Climatic areas
Cultivars/hybrids
Years of production
Harvesting methods
Botanical composition
![Page 25: NIRS – Getting started · History 1800 The First NIR Spectrum recorded (Herschel) 1950-1960 Potential of NIR quantitative analysis was recognized (Kaye) 1960s Research program at](https://reader036.fdocuments.net/reader036/viewer/2022070913/5fb4d94e9fe8f10ff74541e3/html5/thumbnails/25.jpg)
• Our goal is to build an homogenueousdistribution of samples which covers the variability of the product
•In commercial labs it is difficult to know the origin and history of samples analyzed.
• PCA and scores are the tools to use building a population based on NIR spectra.
Population structure: How to built it
![Page 26: NIRS – Getting started · History 1800 The First NIR Spectrum recorded (Herschel) 1950-1960 Potential of NIR quantitative analysis was recognized (Kaye) 1960s Research program at](https://reader036.fdocuments.net/reader036/viewer/2022070913/5fb4d94e9fe8f10ff74541e3/html5/thumbnails/26.jpg)
Principal components and population structure
![Page 27: NIRS – Getting started · History 1800 The First NIR Spectrum recorded (Herschel) 1950-1960 Potential of NIR quantitative analysis was recognized (Kaye) 1960s Research program at](https://reader036.fdocuments.net/reader036/viewer/2022070913/5fb4d94e9fe8f10ff74541e3/html5/thumbnails/27.jpg)
Global H number = distance from the population average (center) (maximum value=3.0)
Neighborhood H number = distance from the closest sample (value=0.6)
Population structure:How to built it
![Page 28: NIRS – Getting started · History 1800 The First NIR Spectrum recorded (Herschel) 1950-1960 Potential of NIR quantitative analysis was recognized (Kaye) 1960s Research program at](https://reader036.fdocuments.net/reader036/viewer/2022070913/5fb4d94e9fe8f10ff74541e3/html5/thumbnails/28.jpg)
NH outlier
GH >3.0 NH >0.6
GH outlier
Determination of outliers samples
![Page 29: NIRS – Getting started · History 1800 The First NIR Spectrum recorded (Herschel) 1950-1960 Potential of NIR quantitative analysis was recognized (Kaye) 1960s Research program at](https://reader036.fdocuments.net/reader036/viewer/2022070913/5fb4d94e9fe8f10ff74541e3/html5/thumbnails/29.jpg)
Building a Calibration
NH and GH **Unrealiable predictions
![Page 30: NIRS – Getting started · History 1800 The First NIR Spectrum recorded (Herschel) 1950-1960 Potential of NIR quantitative analysis was recognized (Kaye) 1960s Research program at](https://reader036.fdocuments.net/reader036/viewer/2022070913/5fb4d94e9fe8f10ff74541e3/html5/thumbnails/30.jpg)
Wha’s next in NIRS
New hardware
New calibration procedures
New standardization procedures
Undried, unground feeds and forages
![Page 31: NIRS – Getting started · History 1800 The First NIR Spectrum recorded (Herschel) 1950-1960 Potential of NIR quantitative analysis was recognized (Kaye) 1960s Research program at](https://reader036.fdocuments.net/reader036/viewer/2022070913/5fb4d94e9fe8f10ff74541e3/html5/thumbnails/31.jpg)
Undried unground analysis