New Standards in the Analysis of Archaeological Metalwork ... of... · New Standards in the...

1
Clayton Meredith 1 , Monica Tromp 1,2 , David Peterson, Ph.D. 1,2 , John Dudgeon, Ph.D. 1,2 , Aram Gevorkyan, Ph.D. 3 , Khachatur Meliksean, Ph.D. 4 (1) Anthropology Department, Idaho State University, (2) Center for Archaeology, Materials and Applied Spectroscopy (CAMAS), Idaho State University, (3) Instute of Archaeology and Ethnography, Naonal Academy of Sciences of the Republic of Armenia; (4) Instute of Geological Sciences, NAS RA New Standards in the Analysis of Archaeological Metalwork Using LA-ICP-MS: A Case Study from the South Caucasus Archaeometallurgy Project Results and ICP-MS and OES Comparison Researchers have found problems in integrang OES data with data collected with newer techniques (Pollard et al. 2007: 47-48). This is par- cularly difficult in the archaeometallurgy of Eurasia (Chernykh 1992) since nearly all published composional data are from OES. Metallurgi- cal groups (Table 2) were assigned to both OES and LA-ICP-MS data. 28 of the 47 samples that had both data sets could assigned the same metallurgical group, 19 could not. There are two possible explanaons for this discrepancy. First, the OES data was not normalized to copper, as was the case with the LA-ICP-MS result. Tin and arsenic values were generally higher for LA-ICP-MS than OES analysis of the same samples, which may be due to the non-normalized values reported for OES. The normalizaon of values for n and arsenic makes LA-ICP-MS more reliable. Second, variaon in the lead values may also be due to segregaon within the sample, since lead is insoluble in copper. Figure 8 is a ternary plot of three major variable elements (arsenic, n and lead) detected by LA-ICP-MS. We grouped the values by object class (“Ornament” and “Implement”) and observed a general paern of ornamental objects being high in n, while implements were generally high in arsenic, but there were sev- eral cases of implements also showing a greater variability overall. When lead does occur, it is generally in similar proporons to arsenic and n. This relaon- ship can be seen in objects with low arsenic and n also having low lead. Table 2. Metallurgi- cal groups (elements at concentraons greater than 1 %). Figure 8. We arrived at these values by subtracng arsenic, n and lead from all other elements, and then subtracng that re- mainder from 100 and mulplying that number by the concen- traon x100. Values for these elements have the following rang- es: Arsenic 0.04—4.5 %, Tin 0.01—25.9 %, Lead 0.02—3.9%. Object Type Implement Ornament Outcome and Conclusions Our analysis of 66 copper and bronze arfacts shows that LA- ICP-MS presents a new direcon for metallurgical analysis in a post-OES analycal environment, given the speed at which it can be performed and the increased specificity of sampling. LA -ICP-MS has many advantages over OES including: 1) greatly enhanced detecon limits (Table 1), 2) analysis with physical standards to enhance replicability and 3) greater accuracy of experimental data with normalized results. Further research should be directed towards integraon with ore analysis for source idenficaon and in depth invesgaon of recycling by comparison with assemblages where endemic recycling is less likely, such as EBA metalwork. Though microstructural hetero- geneity is characterisc of ancient copper and bronze, it may be migated in LA-ICP-MS analysis through selecon of a large area of analysis. Acknowledgments This research was supported by American Councils for Internaonal Educaon through an NEH Internaonal Collaborave Research Fellowship to David Peterson (2009-2011), as well as funding from the Naonal Science Foundaon (BCS 0821783 and OPP 0722771) and the Idaho State University Office of Research, Faculty Research Commiee, and Humanies and Social Sciences Research Com- miee. Dr. Laure Dussubieux (Field Museum) graciously provided copper and bronze standards for the LA-ICP-MS analysis. Thanks to Dr. Philip Kohl and Dr. Ruben Badalyan for use of materials from their excavaons at Horom. Figure 9. EDS map of large scale microstructure variaons displaying dendric inclusions characterisc of cast bronze (leſt) and of a corroded surface of bronze sample where oxidaon has obscured the microstructure (right). Materials and Methods Sample Preparaon Porons of 74 copper and bronze arfacts were embedded in epoxy, then sanded and pol- ished to ensure that they were free of residue and could be oriented perpendicular to the la- ser beam. Before LA-ICP-MS analysis, imaging and analysis was conducted by Scanning Elec- tron Microscopy with Energy Dispersive Spectroscopy (SEM-EDS) using a FEI Quanta 200F field emission gun SEM with a Bruker Quantax 200 SDD EDS system (Figure 7 and 9). LA-ICP-MS Analysis LA-ICP-MS was performed on 66 copper-based samples. The remaining eight samples pre- pared for this project are lead-based and are not considered here. Samples were ablated us- ing a New Wave UP-213 (213 nm wavelength) laser ablaon system in imaged aperture mode. Ablated parculate material was analyzed by a Thermo X Series 2 Inducvely Coupled Plasma-Mass Spectrometer. Abla- on of each sample was performed using an 80 μm beam, traveling at 50 μm/s over three separate 500 x 500 μm raster paerns (Figure 5). Experiment Standardizaon and Calibraon A total of 10 analytes were considered: 59Co, 62Ni, 75As, 109Ag, 117Sn, 121Sb, 125Te, 197Au, 204Pb, 209Bi. In addion 27Al, 55Mn, 57Fe, 67Zn, 82Se, 139La, 140Ce, 153Eu, 172Yb were collected but subsequently removed from the analysis due to poor detecon or non-detecon (Table 1). 65Cu was collected, but due to our internal standardizaon method we did not use the LA- ICP-MS data. A combinaon of copper and bronze standards (B10, B12, 51.13-4, 71.32-4 and SRM-494) were analyzed to effec- vely bracket the unknown samples for calibraon. The basic approach to data calibraon is a slight modificaon of the ap- proach described previously by Neff and Dudgeon (2006). ICP-MS data calibraon was performed by fing standardized con- centraons (in rao to copper) in the standards to standardized counts (raos of raw counts to raw copper counts in the stand- ard). We adopted a procedure to measure the copper concentraon at each ablaon site using SEM-EDS for incorporaon into the LA-ICP-MS data calculaons. Replicate EDS measures of copper were averaged and the mean value was used in the internal standard correcon/calibraon of the unknowns. Opcal Emission Spectrometry (OES) The samples were taken by Dr. Aram Gevorkyan for OES analysis in the early 1990s. He ana- lyzed 47 samples from the Horom site (six from LBA II graves, 41 from Iron I burials) at the Instute for Archaeology and Ethnography in Yerevan, Armenia. The results are unpublished and had to be transcribed from photographs of handwrien records for this study (Figure 6). In these analyses, copper concentraons were not quanfied and values for other elements were not normalized. Copper was instead idenfied as the “base” material. Maximum non- normalized detecon limits were 0.003-0.001 percent. Microstructural Variaon Elements commonly found as trace elements, impuries and alloy constuents in copper and bronze may segregate as cast metal cools, as in the case of lead (Pb), or may form differ- ent phases with copper, as arsenic (As) and n (Sn) will do (Figure 7 and 9). This structural variaon can be seen under high magnificaon. SEM-EDS adds the capacity for chemical analysis of the phases that are present, and can be used to determine whether a sample ar- ea targeted for laser ablaon is characterisc of the overall structure. SEM-EDS maps demonstrate the potenal for idenfying regions of sample heterogeneity (Figure 7 and 9) within the raster paerns (Figure 5) created by LA-ICP-MS. Concerns about microstructural variaon for LA-ICP-MS can be migated if idenfied beforehand by SEM-EDS. Figure 5. SEM image of laser ablaon raster paern. Figure 6. Original handwrien record of OES analysis. Figure 7. EDS map of small scale microstructure with n. Table 1 LA-ICP-MS experimental detecon limits. Background Images The background images of the ceremonial standard from Lchashen (leſt), the Karashamb goblet (center) and the representaon of the solar system from the Sevan Basin (right) are from Les Dossiers d’Archeolo- gie 321 (22, 23, 43). 27Al 55Mn 57Fe 59Co 62Ni 67Zn 75As 82Se 109Ag 117Sn 121Sb 139La 140Ce 153Eu 172Yb 197Au 204Pb 209Bi Detecon Limit (ppm) 1338.02 133.08 732.98 12.93 350.39 1166.77 1312.58 2.61 47.38 5671.30 136.24 5.71 0.08 0.00 0.00 1.44 1458.83 22.29 Figure 2. Locaon of the Horom fortress and necropolis within Armenia. References Chernykh, E. N. (1992) Ancient Metallurgy in the USSR. Cambridge, England: Cambridge University Press. Djindjian, F. (2007) Une histoire de l’Areménie, des origins aux débuts de IV° siècle. Les Dossiers d’Archeologie 321: 6-19. Neff, H. and J. V. Dudgeon (2006) LA-ICP-MS Analysis of Ceramics and Ceramic Raw Materials from the Gila River Indian Community, Arizona. Research Report: Instute for Integrated Research in Ma- terials, Environments, and Society, California State University, Long Beach. Pollard, M., C. Ba, B. Stearn, and S. Young (2007) Analycal Chemistry in Archaeology. New York: Cambridge University Press. Figure 1. The Early Iron Age fortress at Horom (Djindjian 2007: 14). Figure 4. Ornament from the sampled objects. Figure 3. Bronze point from the sampled objects. Discussion The results obtained suggest a process of alloying that was shaped by technical and aesthec preferences based on the materiality of copper and bronze. Arsenic improves the fluidity of bronzes and reduces gas emied by molten copper, allowing for molding of ornate designs. In concentraons of up to about 5.0 wt %, arsenic improves the malleability of copper, but higher concentraons result in a more brile mate- rial. Most of the implements analyzed in this study are cast points (Figure 3) and the use of copper-arsenic alloys enhanced the casng of these objects. While lead also enhances malleability and casng, it segregates in molten copper and through praccal experience, the met- alworkers that made the Horom assemblages probably knew that high concentraons could result in weaker products. The tendency for lead to vary with arsenic and n in the ternary plot (Figure 8) suggests recycling has taken place and was driving down levels of alloy constuents in the local metal pool. The associaon of n bronze with ornaments could indicate that the alloy was valued for its aesthec properes. However, high n concentraons also increase fluidity of copper when molten, and its malleability when solid. While they share some tech- nical properes, the selecon of high n for ornaments and high arsenic for implements appears to have been a maer of aesthec prefer- ence based on visible differences (i.e., n gives copper a golden hue while arsenic lightens and gives it a silvery nt). Objecves Dr. David Peterson, Dr. John Dudgeon and students at Idaho State University conducted Laser Ablaon Inducvely Coupled Plasma Mass Spectrometry (LA-ICP-MS) on 74 metal arfacts from burials dang to the Late Bronze Age II (LBA II, 13th-12th century BCE) and Early Iron Age (Iron I, 11th-9th century BCE) at the Horom necropolis in northern Armenia. This study had three goals: 1. To explore the effecveness of LA-ICP-MS for analyzing systemac differences in the composion of copper and bronze arfacts both within and between assemblages, in order to characterize metal technologies and to aid in the discrimi- naon of copper sources, 2. To compare LA-ICP-MS results with earlier Opcal Emission Spectrometry (OES) results for 47 of the same samples, 3. To improve the efficiency and economy of archaeometallurgical research on materials from Armenia by ulizing facilies at Idaho State University in the Center for Archaeology, Materials and Applied Spectroscopy laboratory (CAMAS). The analyses were performed with cerfied physical standards (4 bronze, 1 copper). Hence our play on words in the tle, in stang that we’ve added ‘new standards’ to analysis of ancient Eurasian metalwork. Background Samples from prehistoric copper and bronze arfacts from Armenia were collected in 2009 and 2010 by the South Caucasus Archaeometallurgy Project, a collaboraon between researchers from the Naonal Academy of Sciences of the Republic of Armenia and Idaho State University. All of the sampled arfacts date to the LBA II and Iron I periods. Archaeometallurgy has been an important part of Armenian archaeology since the 1960s, but has suffered a sharp decrease in funding since the late 1980s. This research previously had centered on OES conducted by Dr. Aram Gevorkyan. Since 2000, Dr. Khachatur Me- liksean has conducted XRF, INAA and MC-ICP-MS analysis of metal arfacts from Armenia at the Curt Engelhorn Center for Archaeometry (Mannheim). Armenia is in the South Caucasus Mountains region, where copper metalwork appeared in the 6th millennium BC. Copper ore deposits were exploited there by the early 4th or perhaps 5th mil- lennium BC. By the Early Bronze Age (ca. 3750/3500–2500 BC) evidence is found for regular pro- ducon of copper and arsenic bronze. From the Early Bronze Age onward, early metal producers in the Caucasus were part of geographically broad networks in which metal, arfacts and tech- nical knowledge circulated (Chernykh 1992). At first, much of this occurred in the context of small villages, while the Middle Bronze Age saw an increase in the consumpon of metalwork in lavish mound burials like those at Karashamb (background, center). By the Late Bronze II and Iron I peri- ods, the consumpon of copper and bronze increased even further in necropoli such as Ark, Lori Berd, and Horom. While in many cases iron came to replace copper and bronze for implements, it was sll used in ornaments (Figure 4) and symbolic items such as representaons of the solar sys- tem and ceremonial standards (background, right and leſt respecvely). Iron did not replace bronze, but contributed to reorientaons in technology, while new alloys like brass and anmony bronze were developed as a complement to arsenic bronze and n bronze, which had been used since the EBA. Increasing Tin (Sn) Increasing Arsenic (As) Increasing Lead (Pb)

Transcript of New Standards in the Analysis of Archaeological Metalwork ... of... · New Standards in the...

Page 1: New Standards in the Analysis of Archaeological Metalwork ... of... · New Standards in the Analysis of Archaeological Metalwork Using LA-IP-MS: A ase Study from the South aucasus

Clayton Meredith1, Monica Tromp1,2, David Peterson, Ph.D.1,2, John Dudgeon, Ph.D.1,2, Aram Gevorkyan, Ph.D.3, Khachatur Meliksetian, Ph.D.4

(1) Anthropology Department, Idaho State University, (2) Center for Archaeology, Materials and Applied Spectroscopy (CAMAS), Idaho State University, (3) Institute of Archaeology and Ethnography, National Academy of Sciences of the Republic of Armenia; (4) Institute of Geological Sciences, NAS RA

New Standards in the Analysis of Archaeological Metalwork Using LA-ICP-MS: A Case Study from the South Caucasus Archaeometallurgy Project

Results and ICP-MS and OES Comparison

Researchers have found problems in integrating OES data with data collected with newer techniques (Pollard et al. 2007: 47-48). This is par-

ticularly difficult in the archaeometallurgy of Eurasia (Chernykh 1992) since nearly all published compositional data are from OES. Metallurgi-

cal groups (Table 2) were assigned to both OES and LA-ICP-MS data. 28 of the 47

samples that had both data sets could assigned the same metallurgical group, 19

could not. There are two possible explanations for this discrepancy. First, the OES

data was not normalized to copper, as was the case with the LA-ICP-MS result. Tin

and arsenic values were generally higher for LA-ICP-MS than OES analysis of the

same samples, which may be due to the non-normalized values reported for OES.

The normalization of values for tin and arsenic makes LA-ICP-MS more reliable.

Second, variation in the lead values may also be due to segregation within the

sample, since lead is insoluble in copper.

Figure 8 is a ternary plot of three major variable elements (arsenic, tin and

lead) detected by LA-ICP-MS. We grouped the values by object class (“Ornament”

and “Implement”) and observed a general pattern of ornamental objects being

high in tin, while implements were generally high in arsenic, but there were sev-

eral cases of implements also showing a greater variability overall. When lead

does occur, it is generally in similar proportions to arsenic and tin. This relation-

ship can be seen in objects with low arsenic and tin also having low lead.

Table 2. Metallurgi-cal groups (elements at concentrations greater than 1 %). Figure 8. We arrived at these values by subtracting arsenic, tin

and lead from all other elements, and then subtracting that re-mainder from 100 and multiplying that number by the concen-tration x100. Values for these elements have the following rang-es: Arsenic 0.04—4.5 %, Tin 0.01—25.9 %, Lead 0.02—3.9%.

Object Type Implement Ornament

Outcome and Conclusions Our analysis of 66 copper and bronze artifacts shows that LA-

ICP-MS presents a new direction for metallurgical analysis in a

post-OES analytical environment, given the speed at which it

can be performed and the increased specificity of sampling. LA

-ICP-MS has many advantages over OES including: 1) greatly

enhanced detection limits (Table 1), 2) analysis with physical

standards to enhance replicability and 3) greater accuracy of

experimental data with normalized results. Further research

should be directed towards integration with ore analysis for

source identification and in depth investigation of recycling by

comparison with assemblages where endemic recycling is less

likely, such as EBA metalwork. Though microstructural hetero-

geneity is characteristic of ancient copper and bronze, it may

be mitigated in LA-ICP-MS analysis through selection of a large

area of analysis.

Acknowledgments This research was supported by American Councils for International Education through an NEH International Collaborative Research

Fellowship to David Peterson (2009-2011), as well as funding from the National Science Foundation (BCS 0821783 and OPP 0722771)

and the Idaho State University Office of Research, Faculty Research Committee, and Humanities and Social Sciences Research Com-

mittee. Dr. Laure Dussubieux (Field Museum) graciously provided copper and bronze standards for the LA-ICP-MS analysis. Thanks to

Dr. Philip Kohl and Dr. Ruben Badalyan for use of materials from their excavations at Horom.

Figure 9. EDS map of large scale microstructure variations displaying dendritic inclusions characteristic of cast bronze (left) and of a corroded surface of bronze sample where oxidation has obscured the microstructure (right).

Materials and Methods

Sample Preparation

Portions of 74 copper and bronze artifacts were embedded in epoxy, then sanded and pol-

ished to ensure that they were free of residue and could be oriented perpendicular to the la-

ser beam. Before LA-ICP-MS analysis, imaging and analysis was conducted by Scanning Elec-

tron Microscopy with Energy Dispersive Spectroscopy (SEM-EDS) using a FEI Quanta 200F

field emission gun SEM with a Bruker Quantax 200 SDD EDS system (Figure 7 and 9).

LA-ICP-MS Analysis

LA-ICP-MS was performed on 66 copper-based samples. The remaining eight samples pre-

pared for this project are lead-based and are not considered here. Samples were ablated us-

ing a New Wave UP-213 (213 nm wavelength) laser ablation system in imaged aperture

mode. Ablated particulate material was analyzed by a Thermo X Series 2 Inductively Coupled Plasma-Mass Spectrometer. Abla-

tion of each sample was performed using an 80 μm beam, traveling at 50 μm/s over three separate 500 x 500 μm raster

patterns (Figure 5).

Experiment Standardization and Calibration

A total of 10 analytes were considered: 59Co, 62Ni, 75As, 109Ag, 117Sn, 121Sb, 125Te, 197Au, 204Pb, 209Bi. In addition 27Al,

55Mn, 57Fe, 67Zn, 82Se, 139La, 140Ce, 153Eu, 172Yb were collected but subsequently removed from the analysis due to poor

detection or non-detection (Table 1). 65Cu was collected, but due to our internal standardization method we did not use the LA-

ICP-MS data. A combination of copper and bronze standards (B10, B12, 51.13-4, 71.32-4 and SRM-494) were analyzed to effec-

tively bracket the unknown samples for calibration. The basic approach to data calibration is a slight modification of the ap-

proach described previously by Neff and Dudgeon (2006). ICP-MS data calibration was performed by fitting standardized con-

centrations (in ratio to copper) in the standards to standardized counts (ratios of raw counts to raw copper counts in the stand-

ard). We adopted a procedure to measure the copper concentration at each ablation site using SEM-EDS for incorporation into

the LA-ICP-MS data calculations. Replicate EDS measures of copper were averaged and the

mean value was used in the internal standard correction/calibration of the unknowns.

Optical Emission Spectrometry (OES)

The samples were taken by Dr. Aram Gevorkyan for OES analysis in the early 1990s. He ana-

lyzed 47 samples from the Horom site (six from LBA II graves, 41 from Iron I burials) at the

Institute for Archaeology and Ethnography in Yerevan, Armenia. The results are unpublished

and had to be transcribed from photographs of handwritten records for this study (Figure 6).

In these analyses, copper concentrations were not quantified and values for other elements

were not normalized. Copper was instead identified as the “base” material. Maximum non-

normalized detection limits were 0.003-0.001 percent.

Microstructural Variation

Elements commonly found as trace elements, impurities and alloy constituents in copper

and bronze may segregate as cast metal cools, as in the case of lead (Pb), or may form differ-

ent phases with copper, as arsenic (As) and tin (Sn) will do (Figure 7 and 9). This structural

variation can be seen under high magnification. SEM-EDS adds the capacity for chemical

analysis of the phases that are present, and can be used to determine whether a sample ar-

ea targeted for laser ablation is characteristic of the overall structure. SEM-EDS maps

demonstrate the potential for identifying regions of sample heterogeneity (Figure 7 and 9)

within the raster patterns (Figure 5) created by LA-ICP-MS. Concerns about microstructural

variation for LA-ICP-MS can be mitigated if identified beforehand by SEM-EDS.

Figure 5. SEM image of laser ablation raster pattern.

Figure 6. Original handwritten record of OES analysis.

Figure 7. EDS map of small scale microstructure

with tin.

Table 1 LA-ICP-MS experimental detection limits.

Background Images The background images of the ceremonial standard from Lchashen (left), the Karashamb goblet (center) and the representation of the solar system from the Sevan Basin (right) are from Les Dossiers d’Archeolo-

gie 321 (22, 23, 43).

27Al 55Mn 57Fe 59Co 62Ni 67Zn 75As 82Se 109Ag 117Sn 121Sb 139La 140Ce 153Eu 172Yb 197Au 204Pb 209Bi

Detection Limit (ppm) 1338.02 133.08 732.98 12.93 350.39 1166.77 1312.58 2.61 47.38 5671.30 136.24 5.71 0.08 0.00 0.00 1.44 1458.83 22.29

Figure 2. Location of the Horom fortress and necropolis within Armenia.

References Chernykh, E. N. (1992) Ancient Metallurgy in the USSR. Cambridge, England: Cambridge University Press. Djindjian, F. (2007) Une histoire de l’Areménie, des origins aux débuts de IV° siècle. Les Dossiers d’Archeologie 321: 6-19. Neff, H. and J. V. Dudgeon (2006) LA-ICP-MS Analysis of Ceramics and Ceramic Raw Materials from the Gila River Indian Community, Arizona. Research Report: Institute for Integrated Research in Ma-

terials, Environments, and Society, California State University, Long Beach. Pollard, M., C. Batt, B. Stearn, and S. Young (2007) Analytical Chemistry in Archaeology. New York: Cambridge University Press.

Figure 1. The Early Iron Age fortress at Horom (Djindjian 2007: 14).

Figure 4. Ornament from the sampled objects.

Figure 3. Bronze point from the sampled objects.

Discussion The results obtained suggest a process of alloying that was shaped by technical and aesthetic preferences based on the materiality of copper

and bronze. Arsenic improves the fluidity of bronzes and reduces gas emitted by molten copper, allowing for molding of ornate designs. In

concentrations of up to about 5.0 wt %, arsenic improves the malleability of copper, but higher concentrations result in a more brittle mate-

rial. Most of the implements analyzed in this study are cast points (Figure 3) and the use of copper-arsenic alloys enhanced the casting of

these objects. While lead also enhances malleability and casting, it segregates in molten copper and through practical experience, the met-

alworkers that made the Horom assemblages probably knew that high concentrations could result in weaker products. The tendency for lead

to vary with arsenic and tin in the ternary plot (Figure 8) suggests recycling has taken place and was driving down levels of alloy constituents

in the local metal pool. The association of tin bronze with ornaments could indicate that the alloy was valued for its aesthetic properties.

However, high tin concentrations also increase fluidity of copper when molten, and its malleability when solid. While they share some tech-

nical properties, the selection of high tin for ornaments and high arsenic for implements appears to have been a matter of aesthetic prefer-

ence based on visible differences (i.e., tin gives copper a golden hue while arsenic lightens and gives it a silvery tint).

Objectives

Dr. David Peterson, Dr. John Dudgeon and students at Idaho State University conducted Laser Ablation Inductively Coupled

Plasma Mass Spectrometry (LA-ICP-MS) on 74 metal artifacts from burials dating to the Late Bronze Age II (LBA II, 13th-12th

century BCE) and Early Iron Age (Iron I, 11th-9th century BCE) at the Horom necropolis in northern Armenia. This study had

three goals:

1. To explore the effectiveness of LA-ICP-MS for analyzing systematic differences in the composition of copper and bronze

artifacts both within and between assemblages, in order to characterize metal technologies and to aid in the discrimi-

nation of copper sources,

2. To compare LA-ICP-MS results with earlier Optical Emission Spectrometry (OES) results for 47 of the same samples,

3. To improve the efficiency and economy of archaeometallurgical research on materials from Armenia by utilizing facilities

at Idaho State University in the Center for Archaeology, Materials and Applied Spectroscopy laboratory (CAMAS).

The analyses were performed with certified physical standards (4 bronze, 1 copper). Hence our play on words in the title, in

stating that we’ve added ‘new standards’ to analysis of ancient Eurasian metalwork.

Background Samples from prehistoric copper and bronze artifacts from Armenia were collected in 2009 and 2010 by the South Caucasus

Archaeometallurgy Project, a collaboration between researchers from the National Academy of Sciences of the Republic of

Armenia and Idaho State University. All of the sampled artifacts date to the LBA II and Iron I periods. Archaeometallurgy has

been an important part of Armenian archaeology since the 1960s, but has suffered a sharp decrease in funding since the late

1980s. This research previously had centered on OES conducted by Dr. Aram Gevorkyan. Since 2000, Dr. Khachatur Me-

liksetian has conducted XRF, INAA and MC-ICP-MS analysis of metal artifacts from Armenia at the

Curt Engelhorn Center for Archaeometry (Mannheim).

Armenia is in the South Caucasus Mountains region, where copper metalwork appeared in the

6th millennium BC. Copper ore deposits were exploited there by the early 4th or perhaps 5th mil-

lennium BC. By the Early Bronze Age (ca. 3750/3500–2500 BC) evidence is found for regular pro-

duction of copper and arsenic bronze. From the Early Bronze Age onward, early metal producers

in the Caucasus were part of geographically broad networks in which metal, artifacts and tech-

nical knowledge circulated (Chernykh 1992). At first, much of this occurred in the context of small

villages, while the Middle Bronze Age saw an increase in the consumption of metalwork in lavish

mound burials like those at Karashamb (background, center). By the Late Bronze II and Iron I peri-

ods, the consumption of copper and bronze increased even further in necropoli such as Artik, Lori

Berd, and Horom. While in many cases iron came to replace copper and bronze for implements, it

was still used in ornaments (Figure 4) and symbolic items such as representations of the solar sys-

tem and ceremonial standards (background, right and left respectively). Iron did not replace

bronze, but contributed to reorientations in technology, while new alloys like brass and antimony

bronze were developed as a complement to arsenic bronze and tin bronze, which had been used

since the EBA.

Increasing Tin (Sn)

Increasing Arsenic (As)

Incr

easi

ng L

ead

(Pb)