Multicolor Flow Cytometry: Principles and Examples of Panel Design

30
For Research Use Only. Not for use in diagnostic or therapeutic procedures. Principles and Examples of Panel Design Presented by Christopher Alfonso, PhD, Scientist, BD Biosciences Multicolor Flow Cytometry Multicolor Flow Cytometry 23-11076-00

Transcript of Multicolor Flow Cytometry: Principles and Examples of Panel Design

Page 1: Multicolor Flow Cytometry: Principles and Examples of Panel Design

For Research Use Only. Not for use in diagnostic or therapeutic procedures.

Principles and Examples of Panel Design

Presented by Christopher Alfonso, PhD, Scientist, BD Biosciences

Multicolor Flow CytometryMulticolor Flow Cytometry

23-11076-00

Page 2: Multicolor Flow Cytometry: Principles and Examples of Panel Design

For Research Use Only. Not for use in diagnostic or therapeutic procedures.

Basic Principles

• Reagent selection starts with your instrument configuration.

• Laser and optical filter choices determine dye selection.

• Optical design impacts signal detection and selection.

• Instrument setup and QC influence performance.

• Proper fluorochrome choices lead to success.

Page 3: Multicolor Flow Cytometry: Principles and Examples of Panel Design

For Research Use Only. Not for use in diagnostic or therapeutic procedures.

Color Choices for 6, 8, 10, and More Colors

6-color 8-color 10-color AdditionalFITC or Alexa Fluor® 488

FITC or Alexa Fluor® 488

FITC or Alexa Fluor® 488

FITC or Alexa Fluor® 488

PE PE PE PE

— — PE-Texas Red® or PE-Alexa Fluor® 610

PE-Texas Red® or PE-Alexa Fluor® 610

PerCP-Cy™5.5 PerCP-Cy5.5 PerCP-Cy5.5 PerCP-Cy5.5

PE-Cy™ 7 PE-Cy7 PE-Cy7 PE-Cy7

APC or Alexa Fluor® 647

APC or Alexa Fluor® 647

APC or Alexa Fluor® 647

APC or Alexa Fluor® 647

— — Alexa Fluor® 680 or 700

Alexa Fluor® 680 or 700

APC-Cy7 APC-Cy7 APC-Cy7 APC-Cy7

— BD Horizon™ V500 or AmCyan

BD Horizon V500 or AmCyan

BD Horizon V500 or AmCyan

— BD Horizon™ V450 or Pacific Blue™

BD Horizon V450 or Pacific Blue™

BD Horizon V450 or Pacific Blue™

— — — Q-dot 655, 705…

Page 4: Multicolor Flow Cytometry: Principles and Examples of Panel Design

For Research Use Only. Not for use in diagnostic or therapeutic procedures.

Instrument Setup

• Determine PMT settings that maximize resolution sensitivity for each experiment.

• Track instrument performance to ensure consistency over time.

Page 5: Multicolor Flow Cytometry: Principles and Examples of Panel Design

For Research Use Only. Not for use in diagnostic or therapeutic procedures.

Why is PMT Optimization Important?Finding PMT settings that maximize resolution sensitivity for each experiment: move fluorescent populations out of electronic noise.

550 volts 650 volts 750 volts

CD4 dim monocytes

CD4+ lymphocytes

CD4 negative

CD4 dim monocytes

CD4+ lymphocytes

CD4 negative

CD4 dim monocytes

CD4+ lymphocytes

CD4 negative

Page 6: Multicolor Flow Cytometry: Principles and Examples of Panel Design

For Research Use Only. Not for use in diagnostic or therapeutic procedures.

BD™ Cytometer Setup & Tracking (CS&T) Beads

• Fully automated software and reagent system for BD digital flow cytometers

• Defines and characterizes baseline performance

• Optimizes and standardizes cytometer setup

• Tracks cytometer performance

– Benefits:

• Consistent, reproducible data

• Simplifies multicolor experiment design

• Higher quality multicolor experiment data

• Flags degrading cytometer performance

Page 7: Multicolor Flow Cytometry: Principles and Examples of Panel Design

For Research Use Only. Not for use in diagnostic or therapeutic procedures.

Important QC ParametersCS&T beads and software automatically derive and store baseline and application-specific settings.

CS&T values:– Linearity– Qr– Br– Electronic Noise (SDEN )

Page 8: Multicolor Flow Cytometry: Principles and Examples of Panel Design

For Research Use Only. Not for use in diagnostic or therapeutic procedures.

QC Parameters Defined• Linearity: Defined as proportionality of input to output (signal to #

of photons)

• Important for fluorescence compensation and quantitative measurements (eg, DNA analysis, antigen/antibody binding)

• Qr: Relative efficiency of the system to detect fluorescence

• Low Qr value = high CV = lower resolution

• High Qr value = low CV = higher resolution

• Br: A measure of true optical background in the fluorescence detector

• Unbound antibody, scatter from the flow cell and ambient light, Raman scattering, spectral overlap, and cell autofluorescence

• SDEN : The constant level background noise contributed by the electronics system

Page 9: Multicolor Flow Cytometry: Principles and Examples of Panel Design

For Research Use Only. Not for use in diagnostic or therapeutic procedures.

Application Settings Optimize Setup for Your Assay

A pre-defined application settings worksheet provides guides to setting PMT voltages based on baseline definition.

• Crosshair indicates the target value for the negative population based upon 10 x SDEN

• Adjust the negative population to fit:– Within the gray box – Center at the

crosshair

Page 10: Multicolor Flow Cytometry: Principles and Examples of Panel Design

For Research Use Only. Not for use in diagnostic or therapeutic procedures.

Rules for Proper Fluorochrome Choice

• Rule 1: Choose the brightest set of fluorochromes for your particular instrument configuration.

• Rule 2: Choose fluorochromes to minimize the potential for spectral overlap.

• Rule 3: Reserve the brightest fluorochromes for dim antibodies and vice versa.

• Rule 4: Avoid spillover from bright cell populations into detectors requiring high sensitivity for those populations.

• Rule 5: Take steps to avoid tandem dye degradation, and consider its impact upon results.

Page 11: Multicolor Flow Cytometry: Principles and Examples of Panel Design

For Research Use Only. Not for use in diagnostic or therapeutic procedures.

0

10

20

30

40

50

60

70

5.5 6.5 7.5 8.5 9.5 10.5 11.5

D

Wunstained

Wcompensated dye

1Stain Index: Metric used by Dave Parks, Stanford – Presented at ISAC 2004

Goal: Normalize the signalto the spread of backgroundwhere background may beautofluorescence, unstained cells, or compensated cellsfrom another dye dimension.

WD(SI)Index1Stain =

Relative Brightness

Where D = Difference between positive and negative peak mediansW = 2 x rSD (robust standard deviation)

Page 12: Multicolor Flow Cytometry: Principles and Examples of Panel Design

For Research Use Only. Not for use in diagnostic or therapeutic procedures.

CD4 Conjugates: Average Stain Index on BD™ LSR II and BD FACSCanto™ II for CD4 Antibody

Fluorochrome Stain IndexPE-Cy™5 353PE 302APC 278Alexa Fluor® 647 214PE-Cy7 139PerCP-Cy5.5 107BD Horizon V450 85Pacific Blue™ 80Alexa Fluor® 488 73Alexa Fluor® 700 61FITC 56APC-Cy7 37PerCP 37AmCyan 25APC-H7 24

Page 13: Multicolor Flow Cytometry: Principles and Examples of Panel Design

For Research Use Only. Not for use in diagnostic or therapeutic procedures.

Spectra Predict Spillover• Minimize the potential for spectral overlap

– Spillover estimates available with the BD Fluorescence Spectrum Viewer

Page 14: Multicolor Flow Cytometry: Principles and Examples of Panel Design

For Research Use Only. Not for use in diagnostic or therapeutic procedures.

Determine Antigen/Fluorochrome Combos

• Match brightest fluors with dimmest antigens– Antigen density– Intracellular antigens are usually dimmer and/or less discrete populations

than surface antigens • Review antibody/fluorochrome combinations in Technical Data Sheet

(TDS)– Visually compare all antigens conjugated to the same fluorochromes

CD5 CD8

Page 15: Multicolor Flow Cytometry: Principles and Examples of Panel Design

For Research Use Only. Not for use in diagnostic or therapeutic procedures.

Example“Bright” antibodies go on “dim” fluorochromes

Example: CD8 “bright” Pacific Blue™ (SI = 80) CD7 “less bright” PE (SI = 302)

CD8 = 90K molecules/cellCD7 = 20K molecules/cell

Page 16: Multicolor Flow Cytometry: Principles and Examples of Panel Design

For Research Use Only. Not for use in diagnostic or therapeutic procedures.

Tandem Dyes

0 hours

2 hours

22.5 hours

PE

CD8 CD3PE-Cy5PE-Cy7 Time Sample

Left in Light

Page 17: Multicolor Flow Cytometry: Principles and Examples of Panel Design

For Research Use Only. Not for use in diagnostic or therapeutic procedures.

False Positives Due to Tandem Degradation

A.

False positives inAPC channel reducedin absence of APC-Cy7

False positivesin PE channelremain

CD8 APC-Cy7+ cells CD4 PE-Cy7+ cells

B.

With CD8 APC-Cy7 and CD4 PE-Cy7

Without CD8 APC-Cy7

Page 18: Multicolor Flow Cytometry: Principles and Examples of Panel Design

For Research Use Only. Not for use in diagnostic or therapeutic procedures.

New Tandems Are More StableAPC-H7 to replace APC-Cy7:

Comparison of Sample Stability(in BD™ stabilizing fixative at RT)

0

50

100

150

200

250

0 1 2 4 6 8 24 48Hours of light exposure

% S

pillo

ver

CD4 APC-H7

CD8 APC-H7

CD4 APC-Cy7

CD8 APC-Cy7

Page 19: Multicolor Flow Cytometry: Principles and Examples of Panel Design

For Research Use Only. Not for use in diagnostic or therapeutic procedures.

Introducing BD Horizon V500• Novel organic dye for the violet laser

– Reduced spillover into the FITC channel when compared to AmCyan– Brighter than Pacific Orange™– Maximizes choice and flexibility in multicolor panel design

• Excitation max of 415 nm, emission max of 500 nm

Pacific Orange™

Horizon V500

AmCyan

Page 20: Multicolor Flow Cytometry: Principles and Examples of Panel Design

For Research Use Only. Not for use in diagnostic or therapeutic procedures.

Horizon V500 Features

% Spillover into FITC

V500 AmCyan

Hu CD4 3 26

Hu CD19 5 30

Hu CD45 3 28

Freshly isolated lymphocytes stained with human CD4 (RPA-T4) conjugated to Pacific Orange™, Horizon V500, or AmCyan, run on a BD FACSCanto II (using a 510/50 nm, 502 LP filter). All conjugates run at 0.25 µg/test.

Freshly isolated lymphocytes stained with CD4 (RPA-T4) conjugated to V500 or AmCyan, CD19 (SJ25C1) V500 or AmCyan, or CD45 (HI30) V500 or AmCyan. Run on a BD LSR II, spillover detected in the FITC channel.

Improved Brightness over Pacific Orange™ Reduced Spillover into the FITC Channel

Page 21: Multicolor Flow Cytometry: Principles and Examples of Panel Design

For Research Use Only. Not for use in diagnostic or therapeutic procedures.

Available Products

Visit bdbiosciences.com/colors for more information.

Page 22: Multicolor Flow Cytometry: Principles and Examples of Panel Design

For Research Use Only. Not for use in diagnostic or therapeutic procedures.

Panel - Naïve and Memory Human T Cells (Cell Surface Staining)

• CD57 FITC• CD11a PE• CD28 PE-Cy5• CD27 PE-Cy7• CD8 PerCP-Cy5.5

• CCR7 A647• CD45RA APC-H7

• CD3 V450• CD4 V500

Blue (488 nm)

Red (633 nm)

Violet (405 nm)

Page 23: Multicolor Flow Cytometry: Principles and Examples of Panel Design

For Research Use Only. Not for use in diagnostic or therapeutic procedures.

CD4+ T-Cell Subsets

Page 24: Multicolor Flow Cytometry: Principles and Examples of Panel Design

For Research Use Only. Not for use in diagnostic or therapeutic procedures.

CD8+ T-Cell Subsets

Page 25: Multicolor Flow Cytometry: Principles and Examples of Panel Design

For Research Use Only. Not for use in diagnostic or therapeutic procedures.

Panel - Naïve and Memory Human T cells (Cell Surface and Intracellular Staining)

• IFN-γ FITC• CD11a PE• CD28 PE-Cy5• CD27 PE-Cy7• CD8 PerCP-Cy5.5

• CCR7 A647• CD45RA APC-H7

• CD3 V450• CD4 V500

Blue (488 nm)

Red (633 nm)

Violet (405 nm)

Page 26: Multicolor Flow Cytometry: Principles and Examples of Panel Design

For Research Use Only. Not for use in diagnostic or therapeutic procedures.

Functional Correlation with Cell Surface Markers in CD4 and CD8 Subsets after Stimulation

2h stim: CD4 cells

5h stim: CD4 cells

5h stim: CD8 cells

Page 27: Multicolor Flow Cytometry: Principles and Examples of Panel Design

For Research Use Only. Not for use in diagnostic or therapeutic procedures.

Considerations with Cell Stimulation In Vitro

Down-regulation of CD4: Stimulation followed by fixation down-modulates CD3 & CD42h stim

Partial solution to down-regulation of CD4: Stain before stimulation

5h stim

5h stim

Page 28: Multicolor Flow Cytometry: Principles and Examples of Panel Design

For Research Use Only. Not for use in diagnostic or therapeutic procedures.

Tools and Resources Available at bdbiosciences.com/colors

Fluorescence Spectrum Viewer Absorption and Emission Spectra

• Posters– Fluorochrome Specifications Chart and the newly updated

Fluorochrome Reference Guide • Application Note

– Selecting Reagents for Multicolor Flow Cytometry• eLearning Courses

– Introduction to Flow course available online

Page 29: Multicolor Flow Cytometry: Principles and Examples of Panel Design

For Research Use Only. Not for use in diagnostic or therapeutic procedures.

Acknowledgments

• Jocelyn Cheung• Joe Trotter• Mark Edinger• Ravi Hingorani• Cynthia Lane• Dave Ernst• Bob Balderas• Immunocytometry Research group at BD

Pharmingen• BD Pharmingen Flow-Core Facility

Page 30: Multicolor Flow Cytometry: Principles and Examples of Panel Design

For Research Use Only. Not for use in diagnostic or therapeutic procedures.

If you have further questions, ask us!

We are fully committed to helping you achieve the most from multicolor flow cytometry.

Contact your US Reagent Sales Rep, e-mail [email protected],

or call 877.232.8995.Please visit our BD Colors page at:

bdbiosciences.com/colors.

Alexa Fluor®, Texas Red®, Pacific Blue™, and Pacific Orange™ are registered trademarks or trademarks of Molecular Probes, Inc.

Cy™ is a trademark of Amersham Biosciences Corp.

Class I (1) laser product.