MU 4082-IfC English Version 1.0.0

20
NEUROMED 4082 IFC MU_4082-IFC_English_Version:1.0.0 (032012) 1 Table of Contents 2 Presentation 2 Accessories that come with the product 3 Symbols 4 Warnings / Attention 7 Installation / Connection to the power network 8 General information 8 Interferential Current 11 Display Language 12 Controls & Control Functions 13 Operation 14 Operation Mode : Two pole Interferential 16 Operation Mode : Manual Vector Interferencial 17 Operation Mode : Automatic Vectos Interferential 19 Operation Mode : Isoplanar Interferential 21 Protocols 24 Indications & Contraindications 25 Electrode Application 26 Preventive maintenance, preservation and cleaning 27 Corrective maintenance 29 Technical Characteristics 32 Manufacturer declaration and orientation 35 Considerations 36 European Representative 37 Bibliographical reference NEUROMED 4082 IFC MU_4082-IFC_English_Version:1.0.0 (032012) 2 Presentation Carci has the honor of congratulating you in your purchase of a high-tech equipment of intrinsic safety, which we are certain will meet the most demanding quality standards. NEUROMED 4082 is an equipment for functional muscular stimulation of medium and low frequency, intended for all areas of electrotherapy, which was developed observing the safety standards NBR IEC 60601-1:1994 + amendment 1:1997, NBR IEC 60601-1-2:2006, and NBR IEC 60601-2- 10:2002, which make it a safe and highly reliable equipment. Accessories that come with the device 4 Conductive rubber (Silicone) electrodes 1 Gel Tube 2 Cables for electrode 1 Three-pole cable 1 User manual To purchase optional accessories, consult the codes on page 31, 'Accessories' item. In case of doubts, please contact: Technical Assistance Department Rua Álvares Fagundes, 359 – São Paulo – SP – Brazil -CEP [zip code] 04338-000 Phone: (0XX11) 5621 7024 / 5622 8205 email: [email protected]

description

rghnfgnfgn

Transcript of MU 4082-IfC English Version 1.0.0

  • NEUROMED 4082 IFC

    MU_4082-IFC_English_Version:1.0.0 (032012) 1

    Table of Contents

    2 Presentation 2 Accessories that come with the product 3 Symbols 4 Warnings / Attention 7 Installation / Connection to the power network 8 General information 8 Interferential Current 11 Display Language 12 Controls & Control Functions 13 Operation 14 Operation Mode : Two pole Interferential 16 Operation Mode : Manual Vector Interferencial 17 Operation Mode : Automatic Vectos Interferential 19 Operation Mode : Isoplanar Interferential 21 Protocols 24 Indications & Contraindications 25 Electrode Application 26 Preventive maintenance, preservation and cleaning 27 Corrective maintenance 29 Technical Characteristics 32 Manufacturer declaration and orientation 35 Considerations 36 European Representative 37 Bibliographical reference

    NEUROMED 4082 IFC

    MU_4082-IFC_English_Version:1.0.0 (032012) 2

    Presentation

    Carci has the honor of congratulating you in your purchase of a high-tech equipment of intrinsic safety, which we are certain will meet the most demanding quality standards.

    NEUROMED 4082 is an equipment for functional muscular stimulation of medium and low frequency, intended for all areas of electrotherapy, which was developed observing the safety standards NBR IEC 60601-1:1994 + amendment 1:1997, NBR IEC 60601-1-2:2006, and NBR IEC 60601-2-10:2002, which make it a safe and highly reliable equipment.

    Accessories that come with the device

    4 Conductive rubber (Silicone) electrodes 1 Gel Tube 2 Cables for electrode 1 Three-pole cable 1 User manual

    To purchase optional accessories, consult the codes on page 31, 'Accessories' item.

    In case of doubts, please contact: Technical Assistance Department Rua lvares Fagundes, 359 So Paulo SP Brazil -CEP [zip code] 04338-000 Phone: (0XX11) 5621 7024 / 5622 8205 email: [email protected]

  • NEUROMED 4082 IFC

    MU_4082-IFC_English_Version:1.0.0 (032012) 3

    Symbols

    Symbols on the equipment

    Means equipment off

    Means equipment on

    This symbol shows that the equipment causes physiological effects and that the user must check the instruction manual

    before using it.

    Applied part type BF

    Class II Equipment

    Manufacturer Identification

    Product Manufacturing Date

    Product Serial Number

    European Representative

    CE Mark and identification of the notifying body

    NEUROMED 4082 IFC

    MU_4082-IFC_English_Version:1.0.0 (032012) 4

    Symbols on the packaging

    Fragile. Handle with care

    Keep in a dry place

    This side up. This symbol shows the correct transport position

    Maximum piling. Shows the maximum number of boxes that can be piled. The quantity is marked in the intermediary

    square.

    Minimum and maximum temperature limits for transport and storage.

    ATTENTION / WARNINGS

    1. Read, understand and practice the operation and precaution instructions. Know the limitations and dangers associated with use of any electrical stimulation device.

    2. Do not expose the device to direct sunlight radiated from a heat radiator, excessive quantities of dust, humidity, vibrations and mechanical shocks.

    3. In the case of liquid infiltration, disconnect the device from the power network and contact the Authorized Technical Assistance.

    4. Before administering any treatment on a patient, you must become acquainted with the operational procedures for each

  • NEUROMED 4082 IFC

    MU_4082-IFC_English_Version:1.0.0 (032012) 5

    treatment modality available, as well as the indications, contraindications, warnings and precautions. Consult other resources to obtain further information on the application of electrotherapy.

    5. Operating sequence

    Connect equipment to the power network

    Turn on equipment through the

    ON/OFF switch

    Set the therapy parameters

    Connect the electrode cables

    to the patient

    Set the current intensity desired

    After the end of the time selected,

    remove the electrodes from

    the patient

    Turn on the equipment through the

    ON/OFF switch

    NEUROMED 4082 IFC

    MU_4082-IFC_English_Version:1.0.0 (032012) 6

    WARNINGS:

    1. Never turn on the ON/OFF key of Neuromed while one or more cables of the electrodes are connected to the patient and to Neuromed.

    2. This equipment is intended for use by health professionals only. 3. This equipment can cause radiofrequency or interrupt operation

    of nearby equipment. 4. The use of accessories not specified by the manufacturer can

    result in increase in EMISSIONS or decrease in IMMUNITY of the equipment.

    5. The use of replacement parts not specified by the manufacturer, in technical assistance, can result in increase in EMISSIONS or decrease in IMMUNITY of the equipment.

    6. We recommend not using this equipment piled or very close to other equipment.

    7. Short-distance operation (e.g. 1 m) from a shortwave or microwave therapy EQUIPMENT can produce instability in the STIMULATOR output;

    8. Application of electrodes near the thorax can increase the risk of cardiac fibrillation;

    9. Simultaneous connections of a PATIENT to a high-frequency surgical EQUIPMENT can result in burning where the STIMULATOR electrodes are applied and possible damage to the stimulator;

    10. We recommend that the current densities for any electrode exceeding 2 effective mA/cm2 may require special attention from the user;

    11. Do not place the equipment on soft base that can cover the lower vents.

    12. Keep this device out of reach of children. 13. This equipment is not suitable for use in the presence of mixture

    of anesthetics inflammable with air, oxygen or nitrous oxide.

  • NEUROMED 4082 IFC

    MU_4082-IFC_English_Version:1.0.0 (032012) 7

    Installation/Connection to the power network

    For a perfect functioning of your equipment, we suggest that the electrical installations be according to the following standards:

    Brazil :

    NBR 5410/1990 Low-voltage electrical installations; NBR13534/1995 Electrical installations in health establishments

    - Safety requirements.

    Other Countries :

    IEC 60364-7-7 (10/2002) Electrical Installations medical locations;

    Or another related standard applied in the country of use.

    Connect the power cable to the three-pole connector located in the posterior panel and connect it to the power socket.

    Your equipment uses switched power sully and operates in any voltage between 100V~ and 240V~ 50 or 60Hz.

    In case the protective fuse burns, replace it only be the value indicated: the fuses must be of 1.0A 250V~ 20AG

    Remove the cover of the fuse box using a screw driver (insert the screw driver in the groove of the fuse box, apply a little pressure and turn it counterclockwise).

    On/Off Switch

    Fuse box

    Power cable connector

    NEUROMED 4082 IFC

    MU_4082-IFC_English_Version:1.0.0 (032012) 8

    General Information

    The NEUROMED 4082 is a functional neuromuscular electrostimulator, Microcontrolled with the functions :

    o Interferential Current Two-pole, Manual Vector, Automatic and Isoplanar.

    Interferential Current

    Interferential Current is a form of medium-frequency wave, distributed through two channels (four electrodes), where the currents cross themselves in the body, in the area that needs treatment. The two currents have interference between them at the crossing point, causing intensity modulation (the current intensity increases and reduces at a regular frequency). In the interferential current, a carrier frequency of the medium frequency is used to pass the low frequency stimulation (beat) through the skin. The relatively low resistance of the skin to the carrier frequency contributes toward the patient's comfort and is oftentimes associated with this type of stimulation. Interferential Currents are all AC currents without any DC residual component.

    1 beat frequency cycle

  • NEUROMED 4082 IFC

    MU_4082-IFC_English_Version:1.0.0 (032012) 9

    Vector Interferential

    In this method of therapy, four electrodes are used, and two modulated currents are generated for each electrode pair (two channels). The frequency of a channel is set in the carrier frequency, while the other channel has a variable frequency, based on the beat frequency and frequency modulation configurations. The Interference occurs when the two currents cross in the tissue. The depth of modulation (which determines the amplitude of the stimulation current) depends on the direction of the currents, and can vary from 0 to 100%. The 100% modulation occurs only in the diagonal (and therefore, in the intersection) of both currents. It is obviously a theoretical situation, based on the presupposition that the tissue is homogeneous. In reality, the tissue is heterogeneous, such that the balance between both channels must be used to obtain the 100-% modulation depth (Fig. 1).

    Figure 1

    Isoplanar

    The isoplanar vector technique is intended to increase the area where effective stimulation occurs. Amplitude modulation occurs in the equipment and a special phase relation between both channels guarantees a 100% modulation depth between the four electrodes in all the positions. Modulation depth is 100% over the complete treatment area. The advantage of this method is that positioning of the four electrodes to effectively treat the affected tissue is less critical. The sensation of

    NEUROMED 4082 IFC

    MU_4082-IFC_English_Version:1.0.0 (032012) 10

    the isoplanar vector mode is mild and equally divided over the treatment area.

    Isoplanar Vector Model

    Two-pole Interferential.

    With the dipole vector technique, the currents of the two electrode pairs are summed vectorially in the tissue. The effect is that stimulation occurs only in direction of the resulting vector, which can be adjusted at an interval of 360. Amplitude modulation occurs in the equipment and the modulation depth is 100%. Stimulation with 100% modulation occurs only in direction of the vector. The advantage of this method is that the direction of stimulation can be adjusted electronically after positioning the electrodes. Automatic vector dipole With the automatic vector dipole technique described above, it is turned at an adjustable speed. If the increase in the current amplitude is greater than the motor threshold, the tissue will contract and relax rhythmically. The current automatic dipole vector is ideal for areas where the mechanical pressure of massage is not desirable.

  • NEUROMED 4082 IFC

    MU_4082-IFC_English_Version:1.0.0 (032012) 11

    Display language

    NEUROMED 4082 allows accessing three display languages: Portuguese English Spanish

    If you wish to change the current language, follow the steps below:

    1. Connect the power cable of the NEUROMED 4082 to the power network.

    2. Turn on the device - the equipment will conduct a test of the display, presenting the Main screen.

    3. Next, press the key and keep it pressed

    4. After the equipment issues a double beep, release the Menu key

    5. The display will show the accessible languages: ENGLISH PORTUGUESE SPANISH

    6. Select the language desired through the keys

    7. Press the key

    8. The equipment will exit the language selection mode, displaying the Main menu.

    NEUROMED 4082 IFC

    MU_4082-IFC_English_Version:1.0.0 (032012) 12

    Controls

    Control functions

    Control of intensity of channels 1 and 2.

    Decrease

    Moves to next field

    Increase

    Moves to previous

    field

  • NEUROMED 4082 IFC

    MU_4082-IFC_English_Version:1.0.0 (032012) 13

    Figure : 1 Figure : 2

    Enables the funct ion chosen

    Enables the main menu screen

    Enables the NEUROMED outputs

    Disables the NEUROMED outputs

    Modes of operation

    Operation of the device:

    Press the ON/OFF at the back of the device. (Figure 1 page 7).

    ATTENTION: Never turn on the ON/OFF key of Neuromed while one or more cables of the electrodes are connected to the patient and to Neuromed.

    The display will show the Initial Screen:

    NEUROMED 4082 IFC

    MU_4082-IFC_English_Version:1.0.0 (032012) 14

    Use the keys to select the current types.

    1. Interferential : Use the and keys to enable selection of the interferential modes : Two-pole, Manual Vector, Automatic Vector or Isoplanar Vector

    1.1. Two-pole Interferential

    o Select the Two-pole Interferential mode.

  • NEUROMED 4082 IFC

    MU_4082-IFC_English_Version:1.0.0 (032012) 15

    o Press the key to enable the Two-pole Interferential mode. The Two-pole mode set screen will be displayed.

    o Through the and keys, set the parameters:

    o Set : 1 minute to 60 minutes, in steps of 1 minute.

    o

    Settings : 2 KHz, 4 KHz, 6 KHz, 8 KHz & 10 KHz.

    o

    Settings : 10 Hz to 150 Hz ( steps of 10 Hz ) o

    Settings : 10 Hz to 100 Hz ( steps of 10 Hz ) o

    Settings : Rise/fall : 0 s to 9 s (steps of 1 s) Sustaining/Rest : 1 s to 30 s (steps of 1 s)

    o After setting the parameters, connect the electrode cables in channels 1 and 2 to the electrodes in the patient. Press the

    Carrier Frequency

    Frequency Modulation

    Frequency Variation

    NEUROMED 4082 IFC

    MU_4082-IFC_English_Version:1.0.0 (032012) 16

    key , the output settings of channels 1 and 2 will be enabled.

    o Use the keys of the channels to set the intensity.

    1.2. Manual Vector Interferential

    o Select the Manual Vector Interferential mode

    o Press the key to enable the Manual Vector Interferential mode. The Manual Vector mode set screen will be displayed.

    o Through the and keys, set the parameters:

    o

    Set : 1 minute to 60 minutes, in steps of 1 minute.

  • NEUROMED 4082 IFC

    MU_4082-IFC_English_Version:1.0.0 (032012) 17

    o Settings : 2 KHz, 4 KHz, 6 KHz, 8 KHz & 10 KHz.

    o Settings : 10 Hz to 150 Hz ( steps of 10 Hz )

    o Settings : 10 Hz to 100 Hz ( steps of 10 Hz )

    o Settings : Rise/fall : 0 s to 9 s (steps of 1 s) Sustaining/Rest : 1 s to 30 s (steps of 1 s)

    o After setting the parameters, connect the electrode cables in channels 1 and 2 to the electrodes in the patient. Press the

    key , the output settings of channels 1 and 2 will be enabled.

    o Use the keys of the channels to set the intensity.

    o Use the keys to move the to the desired position.

    1.3. Automatic Vector Interferential

    o Select the Automatic Vector Interferential mode

    Carrier Frequency

    Frequency Modulation

    Frequency Variation

    NEUROMED 4082 IFC

    MU_4082-IFC_English_Version:1.0.0 (032012) 18

    o Press the key to enable the Automatic Vector Interferential mode. The Automatic Vector mode set screen will be displayed.

    o

    o Through the and keys, set the parameters:

    o

    Set : 1 minute to 60 minutes, in steps of 1 minute.

    o Settings : 2 KHz, 4 KHz, 6 KHz, 8 KHz & 10 KHz.

    o Settings : 10 Hz to 150 Hz ( steps of 10 Hz )

    o Settings : 10 Hz to 100 Hz ( steps of 10 Hz )

    o Set : 1 s to 15 s (steps of 1 s) o

    Carrier Frequency

    Frequency Modulation

    Frequency Variation

    Speed

  • NEUROMED 4082 IFC

    MU_4082-IFC_English_Version:1.0.0 (032012) 19

    Settings : Rise/fall : 0 s to 9 s (steps of 1 s) Sustaining/Rest : 1 s to 30 s (steps of 1 s)

    o After setting the parameters, connect the electrode cables in channels 1 and 2 to the electrodes in the patient. Press the

    key , the output settings of channels 1 and 2 will be enabled. The time selected will decrease.

    o Use the keys of the channels to set the intensity.

    o The will move according to the speed selected.

    1.4. Isoplanar Interferential

    o Select the Isoplanar Interferential mode

    o Press the key to enable the Isoplanar Interferential mode. The Isoplanar mode set screen will be displayed.

    NEUROMED 4082 IFC

    MU_4082-IFC_English_Version:1.0.0 (032012) 20

    o Through the and keys, set the parameters:

    o Set : 1 minute to 60 minutes, in steps of 1 minute.

    o Settings : 2 KHz, 4 KHz, 6 KHz, 8 KHz & 10 KHz.

    o Settings : 10 Hz to 150 Hz ( steps of 10 Hz )

    o Settings : 10 Hz to 100 Hz ( steps of 10 Hz )

    o Set : 1 s to 15 s (steps of 1 s)

    o Settings : Rise/fall : 0 s to 9 s (steps of 1 s) Sustaining/Rest : 1 s to 30 s (steps of 1 s)

    o After setting the parameters, connect the electrode cables in channels 1 and 2 to the electrodes in the patient. Press the key

    , the output settings of channels 1 and 2 will be enabled.

    o Use the keys of the channels to set the intensity.

    o The will move according to the speed selected.

    Speed

    Carrier Frequency

    Frequency Modulation

    Frequency Variation

  • NEUROMED 4082 IFC

    MU_4082-IFC_English_Version:1.0.0 (032012) 21

    2. Protocols

    Attention

    Before applying the treatment protocols, the operator must be informed of the content of this manual as well as operation of this equipment and all aspects related to the patient's treatment.

    Liability Limitations

    Under no circumstance will Carci Ind. e Com., or its suppliers and/or dealers, be held liable for any indirect, special, incidental damage caused by use or inability to use the product, including, without being limited to damages due to loss, or every and any other damage or commercial loss, and regardless of the legal or equal theory (contract, complaint or other mode) on which the claim is based.

    Carci shall not be held liable for any consequence resulting from inaccurate information provided by its personnel, or errors included in this manual and/or other adjunct documentation (including commercial documentation).

    2.1 Select Protocol

    o In the main screen,

    NEUROMED 4082 IFC

    MU_4082-IFC_English_Version:1.0.0 (032012) 22

    o Select the option PROTOCOLS, through the keys ,

    position the cursor in the option SELECT, press the key

    o The display will show the screen:

    o Use the keys to select the desired protocol.

    o Use the keys to change the cursor to the right; and

    to select or delete the protocol.

    o Press the key to confirm.

    2.2 Save Protocol

    o Example : For the therapy

  • NEUROMED 4082 IFC

    MU_4082-IFC_English_Version:1.0.0 (032012) 23

    o Press the key o The screen will be showed

    o Use the and keys to change the cursor to select the letter to write..

    o Press the key to confirm, or to erase.

    o Press the key to save the protocol.

    2.3 Erase Protocol

    o In the Protocols screen:

    NEUROMED 4082 IFC

    MU_4082-IFC_English_Version:1.0.0 (032012) 24

    o Use the keys to select the desired protocol to erase.

    o Press the key to confirm.

    o Press the key to return to the main menu..

    Indications & Contraindications

    Indications :

    Joint Sprain Ligament lesion Muscular contusion Muscular distension Traumatic Synovitis Tendinitis Myalgias Radiculopathies ( lumbosciatalgias and cervicobrachialgias ) Post-operatory of Meniscectomy.

  • NEUROMED 4082 IFC

    MU_4082-IFC_English_Version:1.0.0 (032012) 25

    Contraindications:

    Muscular injuries, tendinitis; Dysfunctions in acute joints (arthritis, bursitis combined with heat)

    and functional lesion; Unconsolidated fibers; Forms of spasticity;

    Electrode Application

    Attention:

    The connection of accessories not specified by the manufacturer can affect the patient and correct operation of the equipment, which is not allowed.

    Before the Treatment:

    Check if the patient fits any other contraindication. Heat sensitivity test of the area to be treated. Clean the area to be treated with aqueous solution with neutral

    soap.

    Conductive Rubber (Silicone) Electrode

    The use of conductive rubber (silicone) electrode is recommended, in combination with conductor gel or wet sponge. The use of gel or duly wet sponges assures low impedance between the skin and the NEUROMED stimulator. The electrodes and sponges can easily be cleaned after treatment. To fix the electrodes with gel, or wet sponges, with the patient's skin, sticky tape of medical use or rubber band can be used.

    Self-Adhesive Electrode

    NEUROMED 4082 IFC

    MU_4082-IFC_English_Version:1.0.0 (032012) 26

    The use of self-adhesive electrodes is characterized by the ease of contact with the skin, due to their malleability. The thin layer of adherent gel allows low impedance between the skin and the Fesmed stimulator. If peradventure during the application the patient informs that the stimulus is too weak, it is a sign that the electrodes must be changed. They must be changed if the display shows the letter M on the right side of the intensity indicator, and the intensity marks zero.

    Current Density

    According to Brazilian Standard NBR IEC 60601-2-10, the maximum current density allowed is 2 mA RMS per cm2 . The NEUROMED output is limited to 50 mA RMS, with charge of 500 Ohms. To check if the current density did not exceed the maximum allowed, divide the output current in mA by the effective area of the electrode in cm2

    Electrode connection and disconnection reactions

    NEUROMED has Constant Current in the current output, which can cause unpleasant reactions if the electrodes are not correctly coupled to the skin, or if there is loss of contact with the skin. Ensure that the output is marking 0 (zero) mA when placing or removing the electrodes.

    Preventive maintenance, preservation and cleaning

    We recommend that the NEUROMED be inspected and calibrated once a year Always ensure that the device is disconnected from the power supply before cleaning. Avoid wetting the electric contacts of the wires and connectors Use a cloth wet with water and soap to clean the cabinet of your device

  • NEUROMED 4082 IFC

    MU_4082-IFC_English_Version:1.0.0 (032012) 27

    Regularly control the power supply and power cable, assuring that there are no defects. On disconnecting the power cable from the device, always remove it from the plug, thus preventing the wire from breaking Always remove the electrodes from the patient with care so as not to damage them. Regularly inspect the cables of the electrodes, checking if there is no breakage of the cable with the plugs.

    Special cares for adherent or silicone electrodes

    Cares for longer lasting electrodes:

    When using silicone electrodes, always clean the contacts of the plugs, preventing residues that render electric contact difficult.

    After using the silicone electrodes, wash them with running water and dry them, keeping in a fresh and dry place.

    After using the adherent electrodes, carefully remove them from the electrode cable, wet them with water and keep in their envelope, in a fresh and dry place.

    Corrective maintenance Important notes:

    Corrective maintenance of this equipment, in addition to the prescriptions above, must be conducted by the CARCI authorized technical assistance only. Electric diagrams, circuits and parts lists are not supplied to end consumers. If needed, contact the Carci Authorized Technical Assistance.

    Problems Solutions

    NEUROMED 4082 IFC

    MU_4082-IFC_English_Version:1.0.0 (032012) 28

    Equipment does not come on Check if the power cable is connected to the power network Check if the fuse(s) is(are) not burnt

    Patient reports that the sensation of the stimulus is greater in one electrode than in the other

    Change the electrodes (silicone or adherent) because they have become worn out

    The intensity is increased to the maximum and the patient reports little stimulation

    Check the patient's sensitivity Change the electrodes (silicone or adherent)

    On increasing the intensity, after 10 mA, the intensity is zeroed, and the letter m is displayed on the right side of the intensity

    Press the Start key and start the treatment again. Check continuity of the electrode cable Check conductivity of the electrodes

    On increasing the intensity, it is zeroed, and the letter M is displayed on the right side of the intensity. This means that there was excess current

    Press the Start key and start the treatment again. Check conductivity of the electrodes. Check if there is short circuit in the electrode cable plug Contact technical assistance,

    Device comes on normally but the patient reports non-stimulation

    Check the coupling and continuity of the electrode cables with the patient Check conductivity of the electrodes

  • NEUROMED 4082 IFC

    MU_4082-IFC_English_Version:1.0.0 (032012) 29

    Technical characteristics

    Origin : CARCI Brazil Models : NEUROMED 4082 Function and Application : Equipment for Electrotherapy Power supply : 100230 V~ automatic ( 10%)

    : 50 / 60 Hz Maximum Consumption : 100 VA ANVISA [National Health Surveillance Agency] Registration : 10314290035

    -------------------------------------------------------------------------------------------------

    Classification according to the standards NBR IEC 60601.1 and NBR IEC 60601.2.10:

    - Type of protection against electric shock:............................... class II - Level of protection against electric shock of the part applied: type BF - Level of protection against harmful water penetration: IPX0 - Disinfection methods : see preventive maintenance - Level of application safety in the presence of an anesthetic mixture inflammable with air, oxygen or nitrous oxide: not suitable - Mode of operation: continuous --------------------------------------------------------------------------------------------------

    Fuse : 1,0 A 250 V 20 AG Dimensions : 31 x 6 x 18 (LxHxD) (cm) Weight : 2.6 kg Excess current protection Continuous component of current equal to Zero

    Neuromed Output Parameters: Output channels : 4 channels Output characteristics : Constant Current Current Amplitude Range : Depends on the waveform Current Amplitude Resolution : 1 mA Maximum Amplitude ( Ipp) : 140 mA (charge of 500 Ohms) Timer : 0 - 60 minutes

    Interferential

    NEUROMED 4082 IFC

    MU_4082-IFC_English_Version:1.0.0 (032012) 30

    Automatic Vector Interferential Frequency : 2, 4, 6, 8 and 10 kHz Frequency Modulation (AMF) : 10 to 150 Hz (steps of 10 Hz) Frequency Spectrum : 10 to 100 Hz (steps of 10 Hz) Ramp rise/descent time: 0 to 9 (steps of 1s) Ramp sustaining time: 1 to 30 s (steps of 1s) Ramp rest time : 1 to 30 s (steps of 1s) Vector Speed : 1 to 9 S Amplitude: 0 - 140 mA

    Isoplanar Interferential Frequency : 2, 4, 6, 8 and 10 kHz Frequency Modulation (AMF) : 10 to 150 Hz (steps of 10 Hz) Frequency Spectrum : 10 to 100 Hz (steps of 10 Hz) Ramp rise/descent time : 0 to 9 (steps of 1s) Ramp sustaining time : 1 to 30 s (steps of 1s) Ramp rest time : 1 to 30 s (steps of 1s) Vector Speed : 1 to 9 S Amplitude : 0 - 140 mA

    Two-pole Interferential Frequency : 2, 4, 6, 8 and 10 kHz Frequency Modulation (AMF) : 10 to 150 Hz (steps of 10 Hz) Frequency Spectrum : 10 to 100 Hz (steps of 10 Hz) Ramp rise/descent time : 0 to 9 (steps of 1s) Ramp sustaining time : 1 to 30 s (steps of 1s) Ramp rest time : 1 to 30 s (steps of 1s) Amplitude : 0 - 140 mA

    Exponential Monophase Phase Duration : 0.05 to 2000 uS ( steps of 10 uS ) Phase Interval : 1 100 ms ( steps of 1 mS ) Polarity : Positive, Negative or Automatic. Amplitude : 0 to 40 mA

    Trbert Ultra Exciting Phase Duration : 2 mS Phase Interval : 5 mS Polarity : Positive, Negative or Automatic. Amplitude : 0 to 40 mA

  • NEUROMED 4082 IFC

    MU_4082-IFC_English_Version:1.0.0 (032012) 31

    Microcurrents Frequency : 1 Hz to 200 Hz ( steps of 1 Hz ) Polarity : Positive, Negative or Automatic. Amplitude : 0 to 40 mA

    Equipment working conditions: Room temperature : +10C to +40C Relative humidity : 0% to 80% Atmospheric pressure : 700 hPa to 1060hPa

    Environmental conditions for transport and storage: Room temperature : -10C to 60oC Relative humidity : 20% to 90% Atmospheric pressure : 500hPa to 1060hPa

    Optional Accessories:

    Conductive rubber (Silicone) electrodes Code 04033 Electrode cables

    Yellow Code 21305G Green Code 21306G Blue Code 20307G Purple Code 21308G

    Three-pole cable Code 14747G User manual Code 10395G

    Note: In case of accessory replacement, only use those indicated by CARCI. The use of accessories not specified by CARCI can compromise equipment safety. CARCI will not be held liable for such procedure.

    Terminologies and Definitions:

    For purposes of the standard NBR IEC 60601-1:1994 + amendment 1997, the following conditions apply, used in the Manufacturer Declarations & Orientations :

    - Meaning given to the verbal forms should and may

    NEUROMED 4082 IFC

    MU_4082-IFC_English_Version:1.0.0 (032012) 32

    should means that compliance with the requirement or test in question is mandatory in order to comply with the Standard.

    would means that compliance with the requirement or test in question is strongly recommended, but not mandatory in order to comply with the Standard.

    may means that compliance with the requirement or test in question is only a particular way of obtaining compliance with the Standard.

    MANUFACTURER DECLARATION AND ORIENTATION - ELECTROMAGNETIC EMISSIONS

    NEUROMED is intended to be used in the electromagnetic environment described below. The buyer and operator of NEUROMED would assure that it is being used in such

    environment. Emission tests Compliance Electromagnetic environment - orientation

    RF Emission CISPR 11 Group 1

    NEUROMED uses RF energy for its internal operation only. Thus, its RF emission is very low and not likely to cause any interference in another nearby electronic equipment.

    RF Emission CISPR 11 Class B NEUROMED is intended to be used in all

    establishments, including homes and those connected directly to the public power network that supplies power to constructions with domestic purposes.

    Harmonics emission IEC 61000-3-2 Class A

    Voltage fluctuation / Flicker emission IEC 61000-3-3

    Compliant

    MANUFACTURER DECLARATION AND ORIENTATION - ELECTROMAGNETIC IMMUNITY

    NEUROMED is intended to be used in the electromagnetic environment specified below. The buyer and operator of NEUROMED would assure that it is being used in such

    environment.

    Immunity tests Test level of IEC 60601 Level of

    compliance Electromagnetic

    environment - orientation

  • NEUROMED 4082 IFC

    MU_4082-IFC_English_Version:1.0.0 (032012) 33

    Electrostatic discharge IEC 61000-4-2

    6 kV contact 8 kV air

    6 kV contact 8 kV air

    The floor must be made of wood, concrete or ceramic. If the floor is covered with synthetic material, the relative humidity of the air must be at least 30%.

    Immunity tests Test level of IEC 60601 Level of

    compliance Electromagnetic

    environment - orientation

    Fast transients / Bursts IEC 61000-4-4

    2 kV power supply line 1 kV signal input and output line

    2 kV power supply line

    Not applicable

    The quality of the power network would be that of a typical hospital or commercial environment.

    Surge IEC 61000-4-5

    1 kV differential mode 2 kV ordinary mode

    1 kV differential mode 2 kV ordinary mode

    Voltage dips, short interruptions and voltage variations in the power supply. IEC 61000-4-11

    95% dip in Ut) For 0.5 cycle

    40% Ut (60% dip in Ut) For 5 cycles

    70% Ut (30% dip in Ut) For 25 cycles

    95% dip in Ut) For 5 s

    95% dip in Ut) For 0.5 cycle

    40% Ut (60% dip in Ut) For 5 cycles

    70% Ut (30% dip in Ut) For 25 cycles

    95% dip in Ut) For 5 s

    Magnetic fields of the network frequencies (50/60 Hz) IEC 61000-4-8

    3 A/m 3 A/m

    The magnetic fields of the network frequencies would be characteristic levels of a typical commercial or hospital environment.

    NEUROMED 4082 IFC

    MU_4082-IFC_English_Version:1.0.0 (032012) 34

    Note: Ut is the voltage of the AC network before applying the test level.

    MANUFACTURER DECLARATION AND ORIENTATION - ELECTROMAGNETIC IMMUNITY

    NEUROMED is intended to be used in the electromagnetic environment specified below. The buyer and operator of NEUROMED would assure that it is being used in such

    environment. Immunity

    tests Test level of IEC 60601

    Level of compliance

    Electromagnetic environment - orientation

    Conducted RF IEC 61000-4-6

    Radiated RF IEC 61000-4-3

    3 Vrms 150 kHz to 80 MHz

    3 V/m 80 MHz to 2.5 GHz

    3 V

    3 V/m

    Portable and mobile RF communication equipment would not be used closer to any part of the NEUROMED, including cables, other than the separation distance recommended, calculated from the equation applicable for the transmitter frequency. Recommended separation distance

    Pd .17.1=

    Pd .17.1= 80 MHz to 800 MHz

    Pd .3.2= 800 MHz to 2.5 GHz

    Where P is the maximum output power of the transmitter in watts (W), according to the transmitter manufacturer, and d is the recommended separation distance in meters (m). The field generated by fixed RF transmitters, as determined by an electromagnetic field study in the sitea, would be lower than the level of compliance in each frequency range. b There may be interference in the equipment surroundings with the following symbol:

    NOTE 1: in the range of 80 MHz and 800 MHz, the highest frequency of the range is applied. NOTE 2: this procedure cannot be applied in all situations. The electromagnetic propagation is affected by absorption and reflection of structures, objects and people. a. The intensity of the fields generated by fixed transmitters, such as cell sites for

    telephones (mobile/wireless) and land mobile radios, amateur radios, AM, FM and TV radio broadcasting stations cannot be theoretically prognosed with precision. To

  • NEUROMED 4082 IFC

    MU_4082-IFC_English_Version:1.0.0 (032012) 35

    evaluate the electromagnetic environment due to the fixed RF transmitters, a study of the electromagnetic field in the site would be considered. If the field intensity measured in the site where NEUROMED is used exceeds the level of compliance above, the NEUROMED would be observed to verify if it is operating normally. If abnormal performance is observed, additional measures may be required,, such as reorientation or reallocation of the NEUROMED;

    b. Above the frequency scale of 150 kHz to 80 MHz, the field intensity should be less than 3 V/m.

    Separation distances recommended between portable and mobile RF communication equipment and NEUROMED

    NEUROMED is intended for use in an electromagnetic environment in which RF disorders are controlled. The buyer or operator of the NEUROMED may help prevent electromagnetic interference by keeping a minimum distance between portable and mobile RF communication equipment (transmitters) and NEUROMED as recommended below, according to the maximum output power of the communication equipment.

    Declared maximum output power of the

    transmitter (W)

    Separation distance according to the transmitter frequency 150 kHz to 80

    MHz

    Pd 17.1=

    80 MHz to 800 MHz

    Pd 17.1=

    800 MHz to 2.5 GHz

    Pd 3.2= 0,01 11.70 cm 11.7 0cm 23.00 cm 0,1 37.00 cm 37.00 cm 72.70 cm 1 1.17 m 1.17 m 2.30 m

    10 3.70 m 3.70 m 7.27 m 100 11.70 m 11.70 m 23.00 m

    For transmitters with declared maximum output power not listed above, the recommended separation distance (d in meters) may be determined using the equation applicable to the transmitter frequency; where P is the maximum output power of the transmitter in watts (W) according to its manufacturer. NOTE 1: at 80 MHz and 800 MHz, the separation distance for the highest frequency is applied. NOTE 2: this procedure may be applied in all situations. The electromagnetic propagation is affected by absorption and reflection of structures, objects and people.

    Final considerations

    It has been Carci's policy to continuously improve the quality of its products. Carci reserves the right to make changes in the design and

    NEUROMED 4082 IFC

    MU_4082-IFC_English_Version:1.0.0 (032012) 36

    specifications, as well as to add and improve its products, without being obliged to install them in already manufactured products.

    The text, illustrations and specifications contained in this manual are based on information available during the printing. All rights reserved.

    This manual cannot be reproduced in whole or in part without the written consent of Carci.

    Manufacturer Data CARCI Indstria e Comrcio de Aparelhos Cirrgicos e Ortopdicos Ltda. CGC [Corporate Taxpayer's Roll]: 61.461.034/0001-78 I.E. [State Tax ID] : 110.182.450.113 Rua lvares Fagundes, 359 So Paulo SP Brazil CEP [zip code] 04338-000 Phone: (11) 3346 2100 Fax.: (11) 3270 8027 email: [email protected] website : www.carci.com.br

    Carci Registration at the Ministry of Health : MS-1.03.142-9

    For Technical Assistance : Technical Assistance Department Rua lvares Fagundes, 359 So Paulo SP Brazil CEP [zip code] 04338-000 Phone: (0XX11) 5621 7024 email: [email protected]

    Technical Responsible: Orlando Orlandi Melo de Carvalho - CREA 5061377287/D

    European Representative

    OBELIS SA 34, Av. de Tervuren, bte 44 B 1040 Brussels , BELGIUM Phone: (32) 2.732.59.54

  • NEUROMED 4082 IFC

    MU_4082-IFC_English_Version:1.0.0 (032012) 37

    Fax: (32) 2.732.60.03

    Email: [email protected]

    WARRANTY CERTIFICATE IN ANNEX

    Bibliographical references

    1. Frampton, VM (1994) Transcutaneous electrical nerve stimulation and Chronic Pain, in Wells, PE, Frampton, VM, Bowsher, D (eds) Pain Management by Physiotherapy. Butterworth Heinemann, London.

    2. Johnson, MI, Ashton, CH, Thompson, JW (1991a) An indepth study of long term users of transcutaneous electrical nerve stimulation (TENS). Implications for clinical use of TENS Pain 44: 221- 229.

    3. Johnson, MI, Ashton, CH, Thompson, JW (1991b) The consistency of pulse frequencies and pulse patterns of transcutaneous electrical nerve stimulation (TENS) used by chronic pain patients. Pain 44: 231- 234.

    4. Tulgar, M, McGlone, F, Bowsher, D, Miles, JB (1991) Comparative effectiveness of different stimulation modes in relieving pain. Part I A Pilot Study. Pain 47: 151 155.

    5. Alberts, P.D., Bakker, M., Spierversterking door Middel van Middenfrequente Wisselstromen. Nederlands Tijdschrift foor Fysiotherapie 10:318 et seq. 1977.

    6. Bernards, J.A., Bouman, L.N., Fysiologie van de Mens. Bohn, Scheltema en Holkema, Utrecht, 5e herziene druk, 1988.

    7. Bernard, Pierre D., La therapie Diadynamique Paris, Editions "PHYSIO", 1962.

    8. Besson, J.M., Chaouch, A., Peripheral and Spinal Mechanisms of Nociception. Physiological Reviews 67, 1. January 1987.

    9. Carley, P., Wainapel, S., Electrotherapy for Acceleration of Wound Healing; Low Intensity Direct Current. Arch. Phys. Med. Rehab., 66, pp 443-446. 1985.

    NEUROMED 4082 IFC

    MU_4082-IFC_English_Version:1.0.0 (032012) 38

    10. Hogenkamp, M., Mittelmeijer, E., Smits, I., Stralen, C. Van, Interferential Therapy. B.V. Enraf-Nonius Delft, Holland, May 1990.

    11. Hoogland, R., Strengthening and Stretching of Muscles using Electrical Current. B.V. Enraf-Nonius Delft, Holland, December 1988.

    12. Howson, D.C., Peripheral Nerve Excitability, Implications for Transcutaneous Electrical Nerve Stimulation. Physical Therapy, Vol. 58, 12, December 1978.

    13. Janda, V., Muskelfunktionsdiagnostik, Muskeltest Untersuchung Verkrzter Muskeln, Untersuchung der Hypermobilitt. Verlag Acco, Leuven, Belgi, 1979.

    14. Johnson, M.A., a.o., Data on Distribution of Fibre Types in thirty-six Human Muscles. An Autopsy Study. Journal of Neurological Science, 18, pp. 111-129.

    15. Kaada, B., Vasodilation induced by Transcutaneous Nerve Stimulation in Peripheral Ischemia. European Heart Journal, 3, pp 303-314, 1982.

    16. Khan, J., Use of Iontophoresis in Peyronies Disease. Physical Therapy, 7, July 1982.

    17. Kloth, L.C., Feedar, J.A., Acceleration of Wound Healing with High Voltage, Monophasic, Pulsed Current. Physical Therapy, 68, pp 503-508, 1988.

    18. Kots, Y.M., Lectures and Laboratory Periods. Symposium on Electrostimulation of Skeletal Muscles. Concordia University, Montreal/Quebec, Canada, December 6th-15th, 1977.

    19. Kovanen, V., Suominen, H., Heillinen, E., Collagen of Slow Twitch and Fast Twitch Muscle Fibres in Different Types of Rat Skeletal Muscle. Eur. J. Appl. Physiol., 52, pp 235-242, 1984.

    20. Kuo, K.H.M., Clamann, H., Coactivation of Synergistic Muscles of Different Fibre types in Fast and Slow Contractions. American Journal of Physical Medicine, vol.60, 5, pp 219-238, 1981.

    21. Kousemaeker, E.A.M., Aarts, N.J.M., Duy, A. Van der, De Behandeling van Slecht Genezende Wonden met Behulp Xanthinolnicotinaat. Nederlands Tijdschrift voor Fysiotherapie, 6, pp 174-178, 1974.

    22. Levine, J.D., Codere, T.J., Basbaum, A.I., Proceedings of the Vth World Congress on Pain. Elsevier Science Publishers B.V. (Biomedical Division), 1988.

    23. Lullies, H., Trincker, D., Taschenbuch der Physiologie II, Gustv Fischer Verlag, Stuttgart, 2e druk, 1973.

  • NEUROMED 4082 IFC

    MU_4082-IFC_English_Version:1.0.0 (032012) 39

    24. Lullies, H., Elektrophysiologische voraussetzungen der Elektrodiagnostik and Elektrotherapie. Elektromedizin. Band 6, 2, 1961.

    25. Lundeberg, T., Kjartansson, J., Samuelsson, J., Effect of Electrical Nerve Stimulation on Healing of Ischaemic Skin Flaps. The lancet, pp 712-714, September 1988.

    26. Melzack, R., Wall, P.D., Pain Mechanisms: A New Theory. Science 150, pp 971-979, November 1965.

    27. Niele, R., Treatment Protocol for the Endomed CV 405. B.V. Enraf-Nonius Delft, Holland, November 1988.

    28. Ottoson, D., Lundeberg, T., Pain Treatment: A Practical Manual by Transcutaneous Electrical Nerve Stimulation. Springer Verlag. Berlin, 1988.

    29. Pouliart, C., De Elektrische Weerstand van de Huid. Belgisch Tijdschrift voor Reumatologie en Fysische Geneeskunde, Vol 29, Fasc. 2, pp 64-79, 1974.

    30. Pronk, N., Het Twee (Vier) Cellenbad by de Behandeling van Hyperhydrosis Palmoplantaris. Nederlands Tijdschrift voor Fysiotherapie, Vol 96, 6, pp 120-121, 1986.

    31. Sato, A., Schmidt, R.F., Somatosympathetic Reflexes: Afferent Fibres, Central Pathways, Discharge Characteristics. Physiological Reviews, Vol, 53, 4, pp 916-947, October 1973.

    32. Sjlund, B.H., Eriksson, M.B.E., Endorphins and Analgesia Produced by Peripheral Conditioning Stimulation. Advances in Pain Research and Therapy, vol 3, 1979.

    33. Steuernagel, O., Skripten zur Elektrotherapie, Band II. Heidelberger Reprographie A. Grosch KG, Eppelheim b. Heidelberg, 1978.

    34. Stralen, C.J.C. van, Transcutane elektrische prikkeling bij Claudicatio Intermittens. Nederlands Tijdschrift voor Fysiotherapie, Vol. 94, nr. 11, blz. 226, 1984.

    35. Wheeler, P.C., Wolcott, L.E., Morris, J.L, Spangler, M.R., Neural Considerations in the Healing of Ulcerated Tissue by Clinical Electrotherapeutic Application of Weak Direct Current. Findings and Theory. Neuro-Electric Reasearch, 1970.

    36. Wolcot, L.E., Gault, W.R., Gatens, P., Acceleration of Healing Rate in Ischaemic Skin Ulcers with Low Intensity Direct Current (LIDC). Lecture on June 14th, Las Vegas, 1972.

    37. Wolf, S.L., Perspectives on Central Nervous System Responsiveness to Transcutaneous Electrical Nerve Stimulation. Physical Therapy, Vol, 58, 12, December 1978.

    NEUROMED 4082 IFC

    MU_4082-IFC_English_Version:1.0.0 (032012) 40

    38. Yaksh, T.L. Substance-P Release from Knee Joint Afferent Terminals: Modulation by Opiods. Brain Research, 458, 319-324, 1988.

    39. Zimmermann, M., Pain Mechanisms and Mediators in Osteoarthritis. Seminars in Arthritis and Rheumatism, vol 18m 4, suppl. 2, pp 22-29, 1989.

    40. Zutphen, H.C.F. van, Stralen, C.J.C. van, Pls, P., Bernards, J.A., Kolle, L.F.J.Th.M., Rens, P.P.Th.G. van, Nederlands Leerboek der Fysische Therape in Engere Zin, Wetenschappelijke Uitgeverij bunge, 3e druk, 1986.