FLEXIBLE AND SCALABLE The new series of Waters ® innovative preparative pump technologies provide extraordinary performance and convenience for benchtop purification of microgram to multigram sample loads. As your lab’s productivity demands increase, so do your expectations for reliability, performance, and flexibility of your solvent delivery system. The full range of Binary and Quaternary Gradient Modules have been designed to meet all of your purification needs. The modules can be easily configured to perform a variety of purification protocols covering a wide range of sample matrices including organic synthetics, natural products, peptides, and proteins. Exceptional flow rates provide you with the precision and accuracy you require, delivering the chromatographic reproducibility that you need to fully automate all your laboratory’s processes. With exceptional solvent blending, automated solvent management, and pulse-free delivery, Waters solvent delivery systems can meet even your most sensitive application challenges. Modular Pumps Increased capability and versatility for preparative chromatography Display Table of Retention Time % RSD Injection # Peak 1 Peak 2 Peak 3 Peak 4 1 1.121 1.693 2.154 3.121 2 1.124 1.697 2.159 3.122 3 1.118 1.689 2.150 3.112 4 1.122 1.694 2.156 3.122 5 1.121 1.693 2.153 3.115 6 1.122 1.696 2.158 3.122 Std Dev 0.002 0.003 0.003 0.004 % RSD 0.17 0.17 0.16 0.14 QUATERNARY GRADIENT MODULES Waters Quaternary Gradient Modules (QGM) are low-pressure mixing quaternary gradient pumps that serve as a solvent delivery device for a variety of purification systems. Available in three flow rate ranges: 0.5 to 50.0 mL/min, 0.5 to 150.0 mL/min, and 4.0 to 300.0 mL/min; the modules simplify mobile phase delivery management by accurately and precisely mixing user-programmed eluent compositions. Choose from three software options: MassLynx, ® Empower, ® or stand-alone console software, to conform to what is already being used in your laboratory. 0.08 0.06 0.04 0.02 0.00 0.00 0.50 1.00 1.50 2.00 2.50 3.00 Minutes 3.50 4.00 4.50 5.00 5.50 6.00 AU Overlay of consecutive injections performed on the 2535 QGM, at 1 mL/min demonstrates the reproducibility that ensures accurate purification of consecutive samples and time based collections. Isocratic separations with online blending of methanol and water illustrate the outstanding reproducibility of the 2535 QGM. Empower-based purification system with the 2707 Autosampler, 2535 Quaternary Gradient Module and the 2489 UV/Visible Detector.

Transcript of Modular Pumps - Waters Corporation...Waters Quaternary Gradient Modules (QGM) are low-pressure...

Page 1: Modular Pumps - Waters Corporation...Waters Quaternary Gradient Modules (QGM) are low-pressure mixing quaternary gradient pumps that serve as a solvent delivery device for a variety

FLEXIBLE AND SCALABLE

The new series of Waters® innovative preparative pump technologies provide

extraordinary performance and convenience for benchtop purification of

microgram to multigram sample loads. As your lab’s productivity demands

increase, so do your expectations for reliability, performance, and flexibility of

your solvent delivery system. The full range of Binary and Quaternary Gradient

Modules have been designed to meet all of your purification needs.

The modules can be easily configured to perform a variety of purification protocols covering a wide range of sample matrices including

organic synthetics, natural products, peptides, and proteins. Exceptional flow rates provide you with the precision and accuracy you require,

delivering the chromatographic reproducibility that you need to fully automate all your laboratory’s processes. With exceptional solvent

blending, automated solvent management, and pulse-free delivery, Waters solvent delivery systems can meet even your most sensitive

application challenges.

Modular PumpsIncreased capability and versatility for preparative chromatography

Display Table of Retention Time % RSD

Injection # Peak 1 Peak 2 Peak 3 Peak 4

1 1.121 1.693 2.154 3.1212 1.124 1.697 2.159 3.122

3 1.118 1.689 2.150 3.1124 1.122 1.694 2.156 3.1225 1.121 1.693 2.153 3.1156 1.122 1.696 2.158 3.122Std Dev 0.002 0.003 0.003 0.004% RSD 0.17 0.17 0.16 0.14

QUAT ERNARY GRADIENT MODULES

Waters Quaternary Gradient Modules (QGM) are

low-pressure mixing quaternary gradient pumps

that serve as a solvent delivery device for a

variety of purification systems. Available in

three flow rate ranges: 0.5 to 50.0 mL/min,

0.5 to 150.0 mL/min, and 4.0 to 300.0 mL/min;

the modules simplify mobile phase delivery

management by accurately and precisely

mixing user-programmed eluent compositions.

Choose from three software options: MassLynx,®

Empower,® or stand-alone console software,

to conform to what is already being used in

your laboratory.

0.08

0.06

0.04

0.02

0.00

0.00 0.50 1.00 1.50 2.00 2.50 3.00Minutes

3.50 4.00 4.50 5.00 5.50 6.00

AU

Overlay of consecutive injections performed on the 2535 QGM, at 1 mL/min demonstrates the reproducibility that ensures accurate purification of consecutive samples and time based collections.

Isocratic separations with online blending of methanol and water illustrate the outstanding reproducibility of the 2535 QGM.

Empower-based purification system with the 2707 Autosampler, 2535 Quaternary Gradient Module and the 2489 UV/Visible Detector.

Page 2: Modular Pumps - Waters Corporation...Waters Quaternary Gradient Modules (QGM) are low-pressure mixing quaternary gradient pumps that serve as a solvent delivery device for a variety

4

BINARY GRADIENT MODULES

Waters Binary Gradient Modules (BGM) incorporate extremely efficient preparative flow rates with greatly reduced system volumes,

capable of providing equivalent performance in both analytical and preparative modes.

Time

Time

Time

Time

-0.00 0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40 2.60 2.80 3.00 3.20 3.40 3.60 3.80 4.00 4.20 4.40 4.60 4.80 5.00 5.20 5.40 5.60 5.80 6.00 6.20 6.40 6.60 6.80 7.00 7.20 7.40 7.60 7.80 8.00

AU

0.0

1.0

2.0

3.0

4.0

5.0

-0.00 0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40 2.60 2.80 3.00 3.20 3.40 3.60 3.80 4.00 4.20 4.40 4.60 4.80 5.00 5.20 5.40 5.60 5.80 6.00 6.20 6.40 6.60 6.80 7.00 7.20 7.40 7.60 7.80 8.00

AU

0.0

2.0

4.0

6.0

8.0

1.0e+1

-0.00 0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40 2.60 2.80 3.00 3.20 3.40 3.60 3.80 4.00 4.20 4.40 4.60 4.80 5.00 5.20 5.40 5.60 5.80 6.00 6.20 6.40 6.60 6.80 7.00 7.20 7.40 7.60 7.80 8.00

AU

0.0

2.0

4.0

6.0

8.0

1.0e+1

-0.00 0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40 2.60 2.80 3.00 3.20 3.40 3.60 3.80 4.00 4.20 4.40 4.60 4.80 5.00 5.20 5.40 5.60 5.80 6.00 6.20 6.40 6.60 6.80 7.00 7.20 7.40 7.60 7.80 8.00

AU

0.0

1.0

2.0

3.0

4.0

5.0 4.6 mm column at 1.5 mL/min

10 mm column at 7.1 mL/min

19 mm column at 25.6 mL/min

30 mm column at 63.8 mL/min

0.43

0.28

0.11

0.73 2.221.58 3.34

4.264.83 6.07

3.84

3.82

0.523.242.09

4.244.46 6.124.84 5.68

3.85

0.48

3.292.130.65

4.24

4.43 6.104.83 5.62

3.870.43

0.28

0.11

0.73 2.221.58 3.34

4.264.83 6.07

* For more information, refer to Waters application note Analytical HPLC to Preparative HPLC: Scale-up Techniques Using a Natural Product Extract (720003120EN).

2545 BGM

The 2545 BGM is a high-pressure mixing binary gradient pump

that serves as the primary solvent delivery device for the

Waters AutoPurification™ System. The 2545 BGM provides

consistently smooth, pulse-free solvent flow for both analytical

and preparative scale flow rates under MassLynx Software

control with the FractionLynx™ Application Manager.

■■ Built on Waters proven high-pressure gradient

mixing technology

■■ Maintained pumping performance and flow rate range

■■ 0.5 to 150.0 mL/min

■■ Self priming from solvent container on floor

■■ Optional leak sensor

The broad flow rate range available with the 2545 BGM enables method development optimization on an analytical scale followed by preparative scale purifications, with no system modifications.* Scaled flow rates and injection volumes were calculated using the Waters OBD™ Calculator.

1525EF HPLC Pump

The 1525EF HPLC Pump is a high-pressure mixing binary

gradient pump that is utilized in many semi-preparative

configurations. Its exceptional flow rate provides you with

precision and accuracy, delivering the chromatographic

reproducibility your lab requires.

■■ 225 µL pump heads deliver up to 22.5 mL/min

■■ Optional FlexInject Manual Dual Injector Module for both

analytical and semi-preparative injections

■■ For semi-preparative applications: 10 to 30 mm I.D.

column separations

■■ Maximum operating pressure: 6000 psi up to 22.5 mL/min

Page 3: Modular Pumps - Waters Corporation...Waters Quaternary Gradient Modules (QGM) are low-pressure mixing quaternary gradient pumps that serve as a solvent delivery device for a variety

1

2535 QGM ■■ Flow rate range from 0.5 to 50.0 mL/min, with separate

flow paths for analytical and preparative capability

■■ Programmable four-solvent selection

■■ Self priming from solvent container on floor

■■ Inlet manifold valve for large volume sample loading

■■ Maximum operating pressure: 6000 psi

2545 QGM■■ Flow rate range from 0.5 to 150.0 mL/min

■■ Programmable four-solvent selection

■■ Self priming from solvent container on floor

■■ Inlet manifold valve for large volume sample loading

■■ Maximum operating pressure: 6000 psi up to

100 mL/min, and 5000 psi at 150 mL/min

2555 QGM ■■ Flow rate range from 4 to 300 mL/min

■■ Programmable four-solvent selection

■■ Self priming from solvent container on floor

■■ Inlet manifold valve for large volume sample loading

■■ Maximum operating pressure: 3000 psi up to

200 mL/min, up to 2500 psi at 300 mL/min

AU

2.06

3.52

Time

This preparative separation of a pharmaceutical sample demonstrates the ability of the 2555 QGM to quickly and accurately resolve a large volume sample load.

Total load 500 mg Column: 50 x 50 mm SunFire™ C18, 10 µm

Conditions:Gradient from 10% A to 25% B over seven minutesSolvent A: 0.1% Formic acid in waterSolvent B: Acetonitrile

Flow rate: 250 mL/minSample: 0.2 g/mL Acetylsalicylic acid with ~0.02g/mL salicylic acidInjection vol: 2.5 mL (20 mL Loop)

The 2545 QGM with a manual inlet valve for loading large

volume samples.

Page 4: Modular Pumps - Waters Corporation...Waters Quaternary Gradient Modules (QGM) are low-pressure mixing quaternary gradient pumps that serve as a solvent delivery device for a variety

Versatile injection capability

Solvent delivery systems can provide access to both analytical and

preparative scale injectors, with the Waters FlexInject Manual Dual

Injector Module mounted on the side of the unit, or the standalone

prep chromatography rack which can be located beside the pump.

To maximize the performance of the combination pump and dual

mode injector for both analytical and preparative flow rates,

the optical flow cells of the Waters 2489 TUV Detector and the

2998 PDA Detector are available in different formats. A dual

inlet, analytical and preparative flow cell, and a semi-prep flow

cell are available for both detectors. A variable pathlength flow

cell is available for the 2489 TUV Detector for high flow, large

concentration sample purifications.

For sample volumes that are too large for conventional injectors, the

QGMs include a manual, large volume inlet manifold. In addition,

automated, repetitive injections of sample solutions may be

introduced to the column through one of the four solvent lines, a

function that can be programmed through the instrument method.

*For more information, refer to Waters application note: A Modular Preparative HPLC System for the Isolation of Puerarin from Kudzu Root Extracts (720003797EN).

The separation of kudzu root extract using the 2707 Sample Manager (A). Very large injection volumes can be achieved by delivering sample through

a solvent line from the 2545 QGM (B) defined by the instrument method.*

Minutes

2.00

1.80

1.60

1.40

1.20

1.00

0.80

0.60

0.40

0.20

0.00

-0.20

0.00 0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00 5.50 6.00 6.50 7.00 7.50 8.00

AU

500 µL injection

Minutes

AU

2.00

1.80

1.60

1.40

1.20

1.00

0.80

0.60

0.40

0.20

0.00

-0.200.00 0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00 5.50 6.00 6.50 7.00 7.50 8.00

2 mL of sample loaded

A

B

AUTOMAT ED SOLV ENT MANAGEMENT

Auto•Blend™ Technology enables you to generate

gradients automatically without having to manually

alter the buffer concentration, giving you the freedom

of using up to four solvents simultaneously. Auto•Blend

programming technology improves uptime, since no

premixing of mobile phases is required. You can build

both column cleanup and system flushing into your

methods. In addition to method development, Auto•Blend

simplifies maintenance by offering straightforward eluent

management, increasing the flexibility and efficiency of

your laboratory.

5%

90%

5%

0%

A

B

C

D

Water

Acetonitrile

100 mMAmmonium

Formate

100 mMAmmoniumBicarbonate

Initial Conditions Final Conditions

Initial Final

90%

5%

5%

0%

5 mM

buffer

5 mM

buffer

Page 5: Modular Pumps - Waters Corporation...Waters Quaternary Gradient Modules (QGM) are low-pressure mixing quaternary gradient pumps that serve as a solvent delivery device for a variety

SOF T WARE CONT ROL

Integrated system monitoring and diagnostic software provided

with the instrument enables operators to monitor several key

parameters including solvent usage, system pressure, and

gradient composition. Interactive control capabilities provide

unique diagnostic tools with visual indicators allowing for easy

troubleshooting and visualization of both the pump and the

UV detectors’ current status.

Console control of the solvent delivery system and the detector for a simplified solution with user diagnostics and independent instrument control.

Channel A and organic composition change are monitored, and the information collected, which can then be reported in a simple JPEG or BMP format.

With several data channels to choose from, users can easily

configure the system to monitor the data channels they require;

such as gradient composition, flow rate, and detector absorbance.

Fully integrated control of Waters Quaternary Gradient Modules

(QGMs) with autosamplers, detectors, and fraction collectors

is available through Empower Software, our industry-leading

chromatography data software (CDS) platform, or MassLynx

Software.

The Waters FlexInject Manual Dual Injector Module is available with any of the Waters preparative pumps, and

provides easy access to both analytical and preparative scale injections.

For CDS independent control of both the QGM and the 2489 TUV

Detector, the instrument console provides complete access to

method creation and a sample list format for system use with

software independent injectors and collectors.

The instrument console provides an interactive diagnostic tool for easy visualization of the overall pump status.

Page 6: Modular Pumps - Waters Corporation...Waters Quaternary Gradient Modules (QGM) are low-pressure mixing quaternary gradient pumps that serve as a solvent delivery device for a variety

Waters Corporation 34 Maple Street Milford, MA 01757 U.S.A. T: 1 508 478 2000 F: 1 508 872 1990 www.waters.com

Waters, T he Science of What’s Possible, MassLynx, and Empower are registered trademarks of Waters Corporation. FractionLynx, SunFire, OBD, Auto•Blend, and AutoPurification are trademarks of Waters Corporation. All other trademarks are the property of their respective owners.

©2011 Waters Corporation. Printed in the U.S.A. July 2016 720003897EN LM-SIG

ORDERING INFORMATION

Quaternary Gradient Modules2535 QGM – Flow rate range 0.5–50.0 mL/min 1760017392545 QGM – Flow rate range 0.5–150.0 mL/min 1760017362555 QGM – Flow rate range 4.0–300.0 mL/min 176001741

Prep Chromatography Rack 205000619

Binary Gradient Modules1525 Binary Gradient Pump 0–10 mL/min 1860015251525 EF Upgrade Kit, for flow rate to 22.5 mL/min 2050003242545 BGM – Flow rate range 0.5–150.0 mL/min 186025450

BUILD A P REPARAT IV E CHROMATOGRAPHY SYST EM TO MATCH YOUR NEEDS, WORKLOAD, AND BUDGET

Choose from a variety of high performance components to configure a preparative chromatography system that’s right for your lab.

Whether your application requires low or high pressure mixing, low or high flow rates, manual or automated injection and collection,

you can choose the exact level of functionality and capacity that your application requires. And since Waters’ purification systems are

upgradable, you don’t have to worry about your investment as your workload increases.

Sample Load Column I.D. Fluid Handling Unit Max Flow Rate

µg – mg 3.9 – 7.8 mm 1525 Binary HPLC Pump* 10 mL/min

µg –10s mg 3.9 – 19 mm 1525 Binary HPLC Pump* and EF Kit 22.5 mL/min

mg –10s mg 3.9 – 50 mm 2535 Quaternary Gradient Module 50 mL/min

m – g 4.6 – 50 mm 2545 Binary Gradient Module* 150 mL/min

m – g 4.6 – 50 mm 2545 Quaternary Gradient Module 150 mL/min

mg – 10s g 7.8 – 75 mm 2555 Quaternary Gradient Module 300 mL/min

* High pressure mixing

Choose a solvent delivery system based on the amount of material you will be loading on the column and the size of the column.

2545 BGM.