miRSIGN Colon Cancer Test (RUO) Expression Analysis Instruction manual v1.1 #80102 August 2016

Transcript of MiRSIGN™ Colon Cancer Test (RUO) - Exiqon

Page 1: MiRSIGN™ Colon Cancer Test (RUO) - Exiqon

miRSIGN Colon Cancer Test (RUO)

Expression Analysis

Instruction manual v1.1 #80102 August 2016

Page 2: MiRSIGN™ Colon Cancer Test (RUO) - Exiqon

2

miRSIGN Colon Cancer Test (RUO) · Instruction Manual

Table of contents

Product Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3I . Reagent kits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5II . Primer Sets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Additional required materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Recommended accompanying products . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

Product description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Control Assays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

Before starting the Experiment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

Protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15

Tips to protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22

Troubleshooting guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30

FAQs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31

Related products . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33

Page 3: MiRSIGN™ Colon Cancer Test (RUO) - Exiqon

3

miRSIGN Colon Cancer Test (RUO) · Instruction Manual

Product summary

Intended use miRSIGN Colon Cancer Test (RUO) is a Real Time Polymerase Chain Reaction (RT-PCR) assay intended for the detection of cancer cells in formalin-fixed, paraffin-embedded tissue specimens . The test is for Research Use Only (RUO) and not for use in diagnostic procedures .

Summary and ExplanationThe miRSIGN Colon Cancer Test (RUO) may assists in determining whether colon cancer cells are present or absent in the formalin fixed paraffin (FFPE) embedded colon tissue specimens . The assay detects the expression level of four microRNAs (hsa-miR-21-5p in combination with three other human microRNAs) which provides a numerical combined expression level that can be compared to a normal reference range and used to categorize the tissue specimen . The assay has shown very high specificity in distinguishing colon cancer specimens versus normal cancer specimens (Figure 1) . Internal validation studies have proven the assay to be reproducible with a high level of accuracy and to be very robust allowing for more stable RNA testing from FFPE samples which are typically challenging .

The miRSIGN Colon Cancer Test (RUO) may potentially aid where histological tumor assessment is undecided or difficult to perform due to limited number of cancer cells or suboptimal pre-analytical processing . A scoring algorithm may be used to categorize the sample as tumor or normal based on a deltaCp value; the Cp of microRNAs hsa-miR-21-5p and hsa-miR-34a-5p subtracted from the sum of Cp values of hsa-miR-126-3p and hsa-miR-143-3p . Although the signature is indicative of malignancy this test is designed for Research Use Only and is not to be used in a clinical setting .

Normal Cancer samples

∆Cq

(miR

-21

and

miR

-34a

vs

miR

-126

and

miR

-143

)

TPR97.3%

SPC97.3%

FPR2.7%

ACC97.3%

PPV97.3%

NPV97.3%

FDR2.7%

6

4

2

0

-2

-4

-6

-8

Actual

CancerCancer

Normal

3736 1

361

37

37

37

Prediction

Normal

Figure 1. miRSIGN Colon Cancer Assay (RUO) Application curve. The dCq values were used as a test score and the cutoff determined on the scatter plot shown . For this microRNA score, the Cq of microRNAs hsa-miR-21-5p and hsa-miR-34a-5p were subtracted from the sum of Cq values of hsa-miR-126-3p and miR-143-3p) . The individual samples are presented as dark blue dots (normal) and light blue dots (cancer samples) .

Page 4: MiRSIGN™ Colon Cancer Test (RUO) - Exiqon

4

miRSIGN Colon Cancer Test (RUO) · Instruction Manual

Principle of ProcedureThe miRSIGN Colon Cancer Test (RUO) is based on the miRCURY LNA™ Universal RT microRNA PCR system and contains LNA™ primer sets and accessory reagents required to perform the RT-PCR procedure for analyzing the expression levels of of a four-microRNA signature in formalin-fixed, paraffin-embedded colon carcinoma tissue .

The miRCURY LNA™ Universal RT microRNA PCR system is a microRNA-specific, LNA™-based system designed for sensitive and accurate detection of microRNA by quantitative real-time PCR using SYBR® Green . The method is based on universal reverse transcription (RT) followed by real-time PCR amplification with two microRNA specific LNA™ enhanced primers .

The miRSIGN Colon Cancer Test (RUO) comprise individual primer sets specific for 4 human microRNAs (hsa-miR-21-5p and three other microRNAs) accompanied by two reagent kits:

• Universal cDNA synthesis kit II (2 kits)• ExiLENT SYBR® Green master mix kit (1 kit)• Individual LNA™ primer sets for hsa-miR-21-5p, hsa-miR-34a-5p, hsa-miR-126-3p, hsa-miR-143-3p

References• Schetter et al . MicroRNA Expression Profiles Associated with Prognosis and Therapeutic Outcome in Colon Adenocarcinoma . JAMA . 2008;299:425-436 .• Slaby et al . Altered Expression of miR-21 . miR-31 . miR-143 and miR-145 is related to clinicopathologic features of colorectal cancer . Oncology . 2007;72:397-402 .• Calin G .A . and Croce C .M . MicroRNA signatures in human cancers . Nature . 2006;6:857-866 .• Kloosterman et al . In situ detection of miRNAs in animal embryos using LNA™ - modified oligonucleotide probes . Nature . 2006; 3:1:27-29 . • Peterson et al . Locked Nucleic Acid (LNATM) Recognition of RNA: NMR Solution Structures of LNATM - RNA hybrids . J Am Chem Soc . 2002;124:21:5974-5982

Page 5: MiRSIGN™ Colon Cancer Test (RUO) - Exiqon

5

miRSIGN Colon Cancer Test (RUO) · Instruction Manual

I . Reagent kits

The miRSIGN Colon Cancer Test (RUO) is a complete kit containing all reagents required for analyzing the expression levels of a cancer related four-microRNA signature in FFPE colon cancer specimens . The kit comprise the following products:

Universal cDNA synthesis kit II, 8-64 rxns (product # 203301)This kit contains all reagents required for first-strand cDNA synthesis for 8-64 reactions1) .

ExiLENT SYBR® Green master mix, 2.5 ml (product # 203403)This kit contains all reagents required for the PCR amplification step . In addition, a positive control assay, the UniSp6 RNA Spike-in control primer set, is provided with this kit for amplification of the synthetic UniSp6 RNA spike-in provided in the Universal cDNA synthesis kit II . The reagents are provided in amounts sufficient for 500 reactions of 10 µL .

Contents

5 x Reaction buffer2) 128 µL, 5 x concentrated

Enzyme mix 64 µL, 10 x concentrated

Nuclease free water 1 ml

UniSp6, RNA Spike-in template3) 12 fmol, dried down

Contents

ExiLENT SYBR® Green master mix, 2x concentrated 2 x 1 .25 ml, 2x concentrated

Nuclease free water 2 x 1 .25 ml

UniSp6 RNA Spike-in control primer set v2 4), lyophilized 200 rxn

1) Number of reactions is based on a standard reaction volume of 10 µL to 80 µL . Reaction volume depends on the application and number of assays to profile . Please consult Figure 4 for details .

2) Includes universal reverse transcription primer .3) Used exclusively for the UniSp6 RNA Spike-in control primer set included in the ExiLENT SYBR® Green master mix kit . 4) Used exclusively for detection of the UniSp6 RNA spike-in provided with the Universal cDNA synthesis kit II . Assay is redesigned

for improved performance, compared to the assay in product # 203551, 203403 and 203421 . Target sequence is unchanged .

Page 6: MiRSIGN™ Colon Cancer Test (RUO) - Exiqon

6

miRSIGN Colon Cancer Test (RUO) · Instruction Manual

II . Primer Sets

MicroRNA LNA™ PCR primer set (product # 204230, # 204486)LNA™ PCR primer sets are designed for optimal performance with the Universal cDNA Synthesis Kit II and the ExiLENT SYBR® Green master mix, kit II . The performance of LNA™ primer sets will be affected if used in combination with less than optimal reagents . The primer sets are supplied in sufficient amounts for 200 reactions of 10 µL .

Reference gene primer set (product # 204227 and product #205992)The Reference gene primer sets are designed for use with the microRNA primers sets above as reference genes for normalization . The performance of LNA™ primer sets will be affected if used in combination with less than optimal reagents . The primer mix is supplied in sufficient amount for 200 reactions of 10 µL .

Contents

LNA™ PCR Primer mix (dried down)

Contents

LNA™ PCR Primer mix (dried down)

Page 7: MiRSIGN™ Colon Cancer Test (RUO) - Exiqon

7

miRSIGN Colon Cancer Test (RUO) · Instruction Manual

Storage

LNA™ primer sets and Reference gene primer setsThe primer sets are shipped dried down at room temperature . The primers can be stored between +4°C and -20°C . Under these conditions, all components are stable for at least 12 months . After resuspension, it is recommended to store LNA™ primer sets and Reference gene primer sets in aliquots at -20°C to avoid repeated freeze-thaw cycles .

Universal cDNA synthesis kit II and ExiLENT SYBR® Green master mixThese kits are shipped on dry ice in polystyrene containers and should be stored at -15°C to -25°C . Do not store in a frost-free freezer . Under these conditions, all components are stable until the expiry date on the package or vial . It is recommended that the RNA spike-in be stored in aliquots at -20°C after re-suspension to avoid repeated freeze-thaw cycles .

Additional required materials

Reagents not supplied• ROX or other passive reference dye (required on some PCR cyclers)

Materials and Equipment not supplied• Nuclease-free PCR tubes or plates for use with individual assays • Nuclease-free, aerosol barrier pipette tips• Nuclease-free, low nucleic acid binding microcentrifuge tubes (e .g . Eppendorf DNA LoBind tubes product and original NUNC vials) • Sealing foils for PCR plates• Micro-centrifuge and plate centrifuge• Heating block, thermal cycler or other incubators• Real-time PCR instrument

Optional but recommended product• miRCURY LNA™ Universal RT microRNA PCR, RNA Spike-in kit (product# 203203)

Page 8: MiRSIGN™ Colon Cancer Test (RUO) - Exiqon

8

miRSIGN Colon Cancer Test (RUO) · Instruction Manual

Recommended accompanying products

Exiqon recommends the Exiqon GenEx qPCR software for comprehensive and convenient data analysis . GenEx includes a wizard for import of Exiqon miRCURY LNA™ Universal RT microRNA PCR data and offers advanced methods to analyze real-time qPCR data in a few simple steps . The software includes tools for selection and validation of reference genes, data pre-processing and comprehensive statistical analyses . For more information and to download a free trial, please go to www .exiqon .com/qpcr-software . The following Exiqon GenEx products are available:

Exiqon GenEx6 Industrial - Exiqon version of GenEx, qPCR analysis software, industrial license

Exiqon GenEx6 Academic - Exiqon version of GenEx, qPCR analysis software, academic license

Exiqon recommends the miRCURY™ RNA Isolation kits for purification of total RNA or small RNA fraction . RNA purified using the miRCURY™ RNA Isolation kits is fully compatible with the miRSIGN Colon Cancer Test (RUO) and the miRCURY LNA™ Universal RT microRNA PCR System .

miRCURY™ RNA Isolation Kit - FFPESpecifically designed for purification of total RNA from FFPE tissue samples .

Page 9: MiRSIGN™ Colon Cancer Test (RUO) - Exiqon

9

miRSIGN Colon Cancer Test (RUO) · Instruction Manual

Product description

A unique system for microRNA profilingmiRCURY LNA™ Universal RT microRNA PCR offers the best available combination of performance and ease-of-use on the microRNA real-time PCR market because it unites two important features (Figure 2):

1 . Universal RT – One first-strand cDNA synthesis reaction provides template for all microRNA real-time PCR assays . This saves precious sample, reduces technical variation, requires use of less reagents, and saves time in the laboratory .

2 . LNA™ PCR amplification – Both PCR amplification primers (forward and reverse) are microRNA specific and optimized with LNA™ . The result is: 1) exceptional sensitivity as well as extremely low background enabling accurate quantification of very low microRNA levels and 2) highly specific assays that allow discrimination between closely related microRNA sequences .

miRCURY LNA™ Universal RT microRNA PCR offers solutions both for high-throughput microRNA expression profiling and for quantification of individual microRNAs .

Step 1: First-strand synthesis (RT)

Step 2: Real-time PCR amplification

Mature microRNAAAAAAAAAAAAAAAAAAAAA

AAAAAAAAAAAAAAAAAAAA

A)

B)

A)

B)

3’ degenerate anchor

miR-specific forward primer

miR-specific reverse primer

TTTTTTTTTTTTTTT

TTTTTTTTTTTTTTT

5’ universal tag

LNA LNA LNA

LNA LNA LNA

Figure 2. Schematic outline of the miRCURY LNA™ Universal RT microRNA PCR System. A poly-A tail is added to the mature microRNA template (step 1A) . cDNA is synthesized using a poly-T primer with a 3’ degenerate anchor and a 5’ universal tag (step 1B) . The cDNA template is then amplified using microRNA-specific and LNA™-enhanced forward and reverse primers (step 2A) . SYBR® Green is used for detection (step 2B) .

Page 10: MiRSIGN™ Colon Cancer Test (RUO) - Exiqon

10

miRSIGN Colon Cancer Test (RUO) · Instruction Manual

Control Assays

There are different types of control assays available in the miRSIGN Colon Cancer Test (RUO):• Reference assays• RNA spike-in assay

These control assays are available in the miRSIGN miR-21 Oncogene Assay kit . One RNA spike-in template is provided with the Universal cDNA Synthesis Kit II, while the assay that will detect this RNA spike-in is available in the ExiLENT SYBR® Green master mix . Additionally 4 RNA spike-in templates are available as a spike-in kit . The assays for detection of these 4 templates as well as endogenous reference assays are available as individual primer sets .

Reference assaysThe selection of reference genes to be used for normalization was based on a determination of microRNAs that contributed the most to precision in distributing samples into cancer and non-cancer categories . The miRSIGN signature test runs multiple microRNAs and controls simultaneously thus reducing noise and increasing the sensitivity of the assay .

See Tip 11

Page 11: MiRSIGN™ Colon Cancer Test (RUO) - Exiqon

11

miRSIGN Colon Cancer Test (RUO) · Instruction Manual

RNA spike-ins (synthetic control templates)The primary purpose of the RNA spike-in and the matching primer pairs for detection is to provide controls for the quality of the PCR . RNA isolations may vary in yield, purity and integrity . Some sample types may contain compounds that inhibit the cDNA synthesis or the PCR even though the RNA has been purified using the best standard procedures . This may result in different efficiencies of the reverse transcription or PCR between compared samples . One way to control for differences in efficiencies at each experimental level (cDNA synthesis and PCR) is by adding a known RNA spike-in to the sample prior to cDNA synthesis . After conducting the PCR but before initiating the data analysis, wells detecting RNA spike-in are compared and outlier samples may be identified and considered for exclusion in the further data analysis .

We have designed a flight of RNA spike-ins for this purpose . The UniSp6 RNA Spike-in template is provided in the miRSIGN Colon Cancer Test (RUO) . Additionally four RNA spike-in templates can be obtained with the separate RNA Spike-in kit . Here, a vial of three RNA spike-in templates, UniSp2, UniSp4 and UniSp5, mixed at different concentrations can be used during RNA isolation . The cel-miR-39-3p RNA template provided in a separate vial in the RNA Spike-in kit can be mixed with the UniSp6 template from the Universal cDNA synthesis kit II to obtain two different template concentrations . This combination can be added during the cDNA synthesis . Five wells in pre-defined PCR panel plates contain the matching primer sets . The selection of RNA Spike-in control primer set in the Pick-&-Mix PCR Panel can be customized to the specific need . A UniSp6 control primer set is also provided with the ExiLENT SYBR® Green master mix kit, which is to be used with our non-plate based PCR primer set products .The RNA spike-ins are shipped dried down and must be re-suspended before use .

If the RNA Spike-in kit with multiple RNA spike-ins is used, follow the protocol accompanying that product .

Use of UniSp6:1 . Re-suspend the UniSp6 RNA spike-in by adding 80 μL nuclease free water to the tube .2 . Mix by vortexing and spin down . Leave for 20-30 min . on ice to fully dissolve RNA spike-in . Mix by vortexing and spin down . Store in aliquots at -20°C3 . Prior to the RT reaction, add 1 μL synthetic spike-in (108 copies/μL) per 20 ng sample RNA .

Page 12: MiRSIGN™ Colon Cancer Test (RUO) - Exiqon

12

miRSIGN Colon Cancer Test (RUO) · Instruction Manual

Experimental designBefore starting the experiment, it is essential to consider the experimental setup and consider the number of replicates needed for obtaining significant results – replicates being technical as well as biological . The number of biological replicates required varies from experiment to experiment depending on the variation within and between the groups . We recommend that a No Template Control (NTC) is included in the study every time a new experiment is set up, to set the background level . The most optimal NTC is a mock up sample preparation including only carrier RNA as sample . The NTC should be run on all assays included in the study . We define the background of an assay as 5 Cq values below the NTC level .

Figure 3. Experimental design

UniSp2

Sample n

Sample n+1 Sample preparation evaluation

Sold as separate Spike-in kit (203203)

cDNA synthesis evaluation

Comes with miRSIGN Colon Cancer Test (RUO)

Included in all panel products

Inter plate Calibration

NTC

UniSp4 UniSp5 cel-miR-39-3p UniSp6 UniSp3

Page 13: MiRSIGN™ Colon Cancer Test (RUO) - Exiqon

13

miRSIGN Colon Cancer Test (RUO) · Instruction Manual

Before starting the Experiment

Before setting up a real-time PCR experiment, there are a number of practical experimental design parameters that should be considered:

RNA input - The miRSIGN Colon Cancer Test (RUO) protocol is optimized for use of 20 ng total RNA per 20µL cDNA synthesis reaction . Prior to conducting a larger microRNA profiling study, it is recommended to optimize the amount of input RNA to the RT reaction in order to avoid conducting a larger study where inhibition occurs sporadically throughout the data set .

Information on how to extract and handle RNA can be found in the tips section . In short, total RNA should be prepared using a method that preserves small RNA species . DNase treatment may be necessary . When using commercially available kits, please ensure that the total RNA preparation is guaranteed to contain microRNAs .

Excess volumes required for pipetting – Liquid handling with pipettes or pipetting robots require excess volumes of reagents due to loss during pipetting . The loss depends on the available pipetting system but losses in the range from 10% -25% are not uncommon . All protocols in the current instruction manual reflect the required reaction volumes and pipetting volumes should be adjusted according to accommodate the pipetting loss of the available pipetting system .

ROX – ROX is a passive reference dye used by some PCR cycles to obtain a robust read over the entire array of wells in a 96- or 384-well PCR plate . The requirement for ROX is instrument dependent and we recommend to follow the instrument manufactures guidelines on this (see also Tip 8) .

RNA work requires specific handling and precautions to prevent RNase contamination of the reagents and degradation of the RNA sample . Find information on how to handle RNA in the tips section starting on page 52 . The tips section also provides simple guidelines for good laboratory practice to ensure optimal performance of PCR experiments .

Important note

See Tip 8

Page 14: MiRSIGN™ Colon Cancer Test (RUO) - Exiqon

14

miRSIGN Colon Cancer Test (RUO) · Instruction Manual

ABI instruments – the default settings on ABI real-time PCR cyclers are not suitable for running miRCURY LNA™ Universal RT microRNA PCR . Settings need to be changed from automatic to manual background and threshold settings to obtain valid PCR data (see also Tip 10) . Furthermore, if the dataset is to be analyzed using the GenEx software, it is important that the experiment is set up as an AQ experiment, not RQ . To ensure correct settings, download the instrument settings file at www .exiqon .com/sds .

See Tip 10

Page 15: MiRSIGN™ Colon Cancer Test (RUO) - Exiqon

15

miRSIGN Colon Cancer Test (RUO) · Instruction Manual

Protocol

This protocol is used for conducting the first-strand cDNA synthesis and real-time PCR, using the individual assays provided in the miRSIGN Colon Cancer Test (RUO) .

Before using the LNA™ PCR primer mix, the Reference gene primer mix or the UniSp6 RNA spike-in control primer mix for the first time, the primers must be re-suspended:• Re-suspend the primer mix by adding 220 μL nuclease free water to the tube . Mix by

vortexing and spin down . Leave on ice for 20-30 minutes .

Additional required materials:• 96- or 384-well plate real-time PCR cycler• Thermocycler for first-strand cDNA synthesis• 96/384-well plates or tube strips compatible with available real-time PCR cycler• Micro centrifuge• Swing bucket centrifuge for 96-/384-well plates

Checklist:• Have you considered excess volumes required for using liquid handling robotics – see page 18• Did you consider how to use the RNA spike-ins – please see page 15• ROX: The ExiLENT SYBR® Green master mix, does not include the ROX passive reference

dye . Please follow instrument manufactures recommendations• ABI instruments: The use of manual background and threshold settings is necessary for

obtaining correct PCR data . Make sure to have the optimal settings by downloading the instrument settings file at www .exiqon .com/sds . Furthermore, if the data is to be analyzed using GenEx, the experiment must be set up as an AQ experiment, not RQ

• Have you optimized the input amount to the RT reaction in order to avoid inhibition?

Page 16: MiRSIGN™ Colon Cancer Test (RUO) - Exiqon

16

miRSIGN Colon Cancer Test (RUO) · Instruction Manual

Phase I: Prepare RNA sample(miRCURY FFPE kit)

Phase II: cDNA Synthesis:One universal cDNA synthesis

Phase III: real-time PCR amplificationOne for each primerset- Mix cDNA with PCR Master mix- Add cDNA:PCR Master mix to each individual primer set- Perform replicates (optional)

Phase IV: Data analysisSee data analysis guide online- Export data for further analysis- Data pre-processing, normalization and statistical analysis

4

3

2

1

0

-1Normal Tumor Tumor

stroma

Tumor Total

• miR-21

• let-7a

Rel

ativ

e ex

pres

sion

(log

2)

miR-21-5p miR-126-3p miR-34a-5p miR-143-3p Spike-inUniSp6

Workflow for miRSIGN Colon Cancer Test (RUO)

Page 17: MiRSIGN™ Colon Cancer Test (RUO) - Exiqon

17

miRSIGN Colon Cancer Test (RUO) · Instruction Manual

ProtocolThe miRSIGN Colon Cancer Test (RUO) protocol is a two-part protocol consisting of:1 . First-strand cDNA synthesis (Step 1-5)2 . Real-time PCR amplification (Step 6-11)

Important: Keep reagents and reactions on ice (or at 4˚C) at all times . First strand synthesis

Step 1 Dilute template RNA

Adjust each of the template RNA samples to a concentration of 5 ng/μL using nuclease free water .

Step 2 Prepare reagents

Gently thaw the 5x Reaction buffer and nuclease-free water, and immediately place on ice . Mix by vortexing . Re-suspend the RNA spike-ins according to the appropriate RNA Spike-ins protocol (see page 15), leave on ice for 15-20 minutes . Immediately before use, remove the Enzyme mix from the freezer, mix by flicking the tubes and place on ice . Spin down all reagents .

Page 18: MiRSIGN™ Colon Cancer Test (RUO) - Exiqon

18

miRSIGN Colon Cancer Test (RUO) · Instruction Manual

Step 4Mix and spin reagents

Mix the reaction by very gentle vortexing or pipetting to ensure that all reagents are thoroughly mixed . After mixing, spin down .

Step 5Incubate and heat inactivate1)

• Incubate for 60 min at 42˚C .• Heat-inactivate the reverse transcriptase

for 5 min at 95˚C . • Immediately cool to 4°C .• Store at 4°C or freeze .

Step 3 Combine reagents according to Table 2

Note: remember to calculate necessary excess volume for pipetting and robotic dead volume .

If performing first-strand cDNA synthesis on multiple RNA samples, it is recommended to prepare an RT working solution of the 5x Reaction buffer, water, Enzyme mix and RNA spike ins (in the proportion indicated in the first four lines of Table 2) .The following procedure is recommended:1 . Prepare the required amount of RT working solution and place it on ice .2 . Dispense RT working solution into nuclease free tubes .3 . Dispense template RNA in each tube .

Table 2 – Reverse transcription reaction setup

Reagent Volume (µL), RT reaction

5x Reaction buffer 2

Nuclease-free water 4 .5

Enzyme mix 1

Synthetic RNA spike ins, optional replace with H2O if omitted

0 .5

Template total RNA (5 ng/µL) 2

Total volume 10

1) The protocol can be interrupted at this stage . The undiluted cDNA may be kept at -20˚C for up to 5 weeks (optional store at 4˚C for up to 4 days) . It is recommended that synthesized cDNA is stored in “low-nucleic acid binding” tubes or plates .

Page 19: MiRSIGN™ Colon Cancer Test (RUO) - Exiqon

19

miRSIGN Colon Cancer Test (RUO) · Instruction Manual

qPCR protocol

Step 6Prepare reagents for real-time PCR

Place cDNA (from Step 5), nuclease free water and PCR Master mix on ice and thaw for 15-20 min . Protect the PCR Master mix vials from light . Immediately before use, mix the PCR Master mix by pipetting up and down . The rest of the reagents are mixed by vortexing and spun down .

Step 7Dilute cDNA template 80x in nuclease free water2)

Immediately before use, dilute only the amount of cDNA template needed for the planned real-time PCR reactions 80x in nuclease free water (e .g . add 395 μL nuclease free water to each 5 μL of reaction) . It is important that “low-nucleic acid binding” tubes or plates are used . It is not recommended to store the 1:80 dilution of cDNA .

Recommendation: Include a passive reference dye in the cDNA dilution if advised by instrument manufacturer . Please note that the PCR Master mix does not include ROX . The amount of ROX required is instrument dependent and it is important to refer to the manufacturer’s recommendations when deciding how much ROX to use, see Tip 8 .

2) Adjust volumes to accommodate your in-house liquid handling system volume loss when pipetting

See Tip 8

Page 20: MiRSIGN™ Colon Cancer Test (RUO) - Exiqon

20

miRSIGN Colon Cancer Test (RUO) · Instruction Manual

Step 9Mix and spin reagents

Mix the reaction by gentle pipetting to ensure that all reagents are mixed thoroughly . After mixing cap tubes or strips or seal the plate with optical sealing as recommended by the manufacturer . Spin down in a centrifuge (1500g for 1 minute) . The experiment can be paused at this point . Store the reactions protected from light at 4°C for up to 24 hours .

Step 8Combine PCR Master mix, PCR primer mix and cDNA according to Table 3 .

Mix thoroughly

Note: remember to calculate necessary excess volume for pipetting and robotic dead volume .

When multiple real-time PCR reactions are performed with the same microRNA primer set, it is recommended to prepare a primer master mix working-solution of the PCR primers and the PCR Master mix (in the proportion indicated in Table 3) . The following procedure is recommended:1 . Prepare the required amount of primer:master mix working-solution

(see Table 3) and place it on ice . It is recommended to include excess of all reagents in the master mix to compensate for pipetting excess material .

2 . Place the relevant volume of primer:master mix working-solution in PCR tubes/wells (see Table 3) and spin tubes/plate briefly in a centrifuge (1500g for 1 minute), to remove air bubbles .

3 . Add cDNA template to each tube/well .

Table 3 – Real-time PCR reaction, pr. 10 µL reaction3)

Reagent Volume (µL), 96/384-well plate, tubes or strips

PCR Master mix 5

PCR primer mix4) 1

Diluted cDNA template 4

Total volume 10

3) If using a 96-well cycler with a minimum recommended volume of 20 µL (as some ABI instruments), then use 10 µL reaction volume and set the instrument settings at 20 µL .

4) The PCR primer mix must be dissolved prior to real-time PCR set-up, see page 20 .

Page 21: MiRSIGN™ Colon Cancer Test (RUO) - Exiqon

21

miRSIGN Colon Cancer Test (RUO) · Instruction Manual

5) Five additional amplification cycles are required when using the LC480 instrument to allow collection of assay data with Cp-values up to 40 .

6) The ramp-rate of cooling from 95˚C to 60˚C should be set to 1 .6°C/s . This is equivalent to 100% under standard cycling conditions on the ABI 7500, 7900 and Viia7 instruments . If the ramp rate of cooling is too rapid, performance may be compromised .

7) Melting curve analysis of the PCR product(s) is recommended to verify specificity and identity of the amplification reaction . Melting curve analysis is an analysis step built into the software of instruments . Please follow the instructions provided by the supplier . Note: The Tm of a PCR product depends on buffer composition, salt concentration and the PCR instrument .

Step 10Real-time PCR amplification

Perform real-time PCR amplification followed by melting curve analysis according to Table 4 .

Table 4 – Real-time PCR cycle conditions

Process step Settings, LC480 instrument5)

Settings, other instruments3)

Polymerase Activation/Denaturation 95˚C, 10 min 95˚C, 10 min

Amplification 45 amplification cycles at 95˚C, 10 s60˚C, 1 min, ramp-rate 1 .6˚C/s6)

Optical read

40 amplification cycles at95˚C, 10 s60˚C, 1 min, ramp-rate 1 .6˚C/s6)

Optical read

Melting curve analysis7) Yes Yes

Step 11Analyze data

Perform initial data analysis using the software supplied with the real-time PCR instrument to obtain raw Cq values (Cp or Ct, depending on PCR instrument) . If you are using an ABI instrument, please note that it is not recommended to use auto Ct settings . Furthermore, if the data is to be analyzed using Exiqon GenEx, the experiment must be set up as an AQ experiment, not RQ . For a guide on how to set manual baseline and threshold, refer to Tip 10, page 56 in the tips section . If you are using a Roche LC480 instrument, we recommend analysis using the 2nd derivative method .

A scoring algorithm can calculated as follows: (miR-126-3p + miR-143-3p) -(miR-21-5p + miR-34a-5p)

ABI instrument user?Apply manual baseline and threshold settings! Go to www.exiqon.com/sds to download settings file

Page 22: MiRSIGN™ Colon Cancer Test (RUO) - Exiqon

22

miRSIGN Colon Cancer Test (RUO) · Instruction Manual

Tips to protocol

Purification and preparation of total RNA that includes small RNAs (<200 nt) from a biological sample is the first critical step for a successful expression profiling analysis of microRNAs . Therefore, the method used for RNA sample preparation is critical to the success of the experiment .

The following points should be considered before starting the experiment: • We recommend using the miRCURY™ RNA Isolation kit for isolation of RNA that contains

the small RNA fraction . If not using one of the miRCURY™ RNA Isolation kits, it is important that the method used to isolate RNA from a sample should give a quantitative recovery of small RNAs and should not result in a substantial loss of the small RNA fraction . This also applies when using commercially available kits . Make sure the total RNA preparation is guaranteed to contain microRNA .

• If commercially available purified RNA is used, it is important to make sure that the RNA is guaranteed to contain small RNAs and is representative of the microRNAs in the species and/or tissue from which it was isolated .

• The comparison of samples prepared using different RNA isolation methods is not recommended . However, if this is necessary, it is recommended to include the measure of a reference small RNA which has a consistent and unvaried expression level in order to allow for comparison of microRNA levels in the different sample preparations .

• The isolation of RNA and the reaction steps preceding real-time PCR should be performed in rooms separate from where real-time PCR experiments will take place in order to avoid contamination with amplicon .

• It is recommended that the integrity of isolated RNA be assessed before proceeding with quantitative real-time PCR experiments . This may be performed either on the Agilent 2100 Bioanalyzer or by denaturing gel electrophoresis .

• If necessary, treat RNA preparations with DNases to remove contaminating DNA that may interfere with the real-time PCR results .

• Purified total RNA should be dissolved in nuclease-free water at a stock concentration of at least 1 μg/μL . See recommendations for storage conditions in Tip 3 .

Tip 1. RNA extraction and template preparation

Page 23: MiRSIGN™ Colon Cancer Test (RUO) - Exiqon

23

miRSIGN Colon Cancer Test (RUO) · Instruction Manual

Plasma and serum are essentially cell free liquid samples . This means that only circulating RNA is extracted from these sample types, resulting in low total RNA concentrations, even if the microRNA fraction is readily detectable . The result of this is that measuring correct RNA concentrations is difficult, and that there is a high risk of increased loss during extraction . For this reason, we recommend using RNA amounts based on starting volume rather than RNA quantity . Comprehensive guidelines for microRNA profiling in blood serum/plasma can be downloaded at www .exiqon .com/serum-plasma-guidelines .

The following precautions should be taken to prevent RNase contaminationand degradation of the RNA sample and reagents:• Always wear disposable gloves .• Use nuclease-free, low nucleic acid binding plasticware and filter barrier pipette tips .• Keep tubes capped when possible . Always spin tubes before opening .• For long-time storage, RNA may be stored at -80°C . Avoid repeated freezing and thawing

cycles .

To reduce the risk of contaminating PCRs with “old” PCR amplicons, and consequently obtain false results: • Always wear a clean lab coat . Use separate lab coats for RNA sample preparation, cDNA

synthesis and when setting up PCR reactions or handling PCR products .• Change gloves often, especially whenever you suspect they may have been contaminated . • Establish and maintain designated areas for PCR set-up, PCR amplification, and gel

electrophoresis of PCR products .• Never bring amplified PCR products into the PCR set-up area .• Spin down all reaction and sample tubes before opening . Open and close all reagent and

sample tubes carefully, trying not to splash or spray PCR samples .• Keep reactions and components capped whenever possible .• Use filter barrier pipette tips to avoid aerosol-mediated contamination of your pipetting

device .• Clean laboratory benches and equipment regularly .

Tip 2. Guidelines for serum, plasma and biofluid samples

Tip 3. General guidelines for handling and storage of RNA

Tip 4. Good PCR laboratory practice

Page 24: MiRSIGN™ Colon Cancer Test (RUO) - Exiqon

24

miRSIGN Colon Cancer Test (RUO) · Instruction Manual

In order to ensure optimal performance of the miRCURY LNA™ Universal RT microRNA PCR system it is important that the reagents and reactions are kept on ice (or at 4°C) as much as possible during the protocol (apart from reaction steps specifically involving raised temperatures) .

Store cDNA samples in nuclease-free low nucleic acid-binding micro centrifuge tubes, e .g . Eppendorf DNA LoBind tubes .

The following controls are recommended and should be included in the experimental set-up:• Reverse transcription/no enzyme controls, i .e . first-strand cDNA synthesis reactions

performed without the Enzyme mix . If a PCR product is amplified from this control reaction it indicates genomic DNA or PCR product contamination of the template RNA .

• Non-template controls in the real-time PCR amplification, i .e . real-time PCR reactions performed without any cDNA template . This control will reveal PCR product contamination of the reaction .

• Blank purification or carrier only, i .e . when purifying RNA in the presence of carrier RNA (such as for serum and plasma samples) . This control will reveal any non-specific signals originating from the carrier RNA alone .

Many real-time PCR instruments will only produce reliable results when a passive reference dye such as ROX is added to the PCR reaction . The reference dye is used to normalize signals from individual PCR wells in order to enable comparison of real-time PCR amplification signals across an entire PCR-plate .It is recommended to determine whether your real-time PCR instrument has this type of requirement . The amount of ROX to include in the PCR reaction depends on the requirements of the real-time PCR instrument and must be adjusted accordingly . Please follow the supplier’s instructions for preparation and concentrations of ROX solutions . Typically, real-time PCR instruments that allow excitation at individual wavelengths for individual dyes (most filter wheel based instruments) require less ROX than instruments that use a single excitation wavelength for all fluorophores (most laser based instruments use excitation at 488 nm) .

Tip 5. Keep reagents and reactions cool at all times

Tip 6. First strand cDNA synthesis

Tip 7. Recommended controls

Tip 8. Passive reference dye (ROX)

Page 25: MiRSIGN™ Colon Cancer Test (RUO) - Exiqon

25

miRSIGN Colon Cancer Test (RUO) · Instruction Manual

It is important to note that excessive amounts of ROX may inhibit the PCR reaction . It may be recommended to perform a titration of ROX amounts in order to determine the optimal concentration for a particular instrument-system combination . It is possible to use the spike-in template (provided in the Universal cDNA synthesis kit II) and the Control primers (provided in the ExiLENT SYBR® Green master mix kit) to perform such titrations .

Recommended ROX concentrations are found here:

When setting up large scale experiments with several PCR plates involved, individual run differences may be observed . One way of avoiding these having an effect on the results is setting up the experiment in such a way that all samples for one gene are run on the same plate . However, this is not always feasible . When Cq values for one gene have to be compared across plates, it is recommended to employ an inter-plate calibrator .

An inter-plate calibrator is a template-assay combination which is the same in all plates, and always located in the same well across different plates . This can then be used to calibrate all plates to give the same Cq value for the calibrator, thereby reducing run-to-run variance . Please note that all panels contain inter-plate calibrators .

Tip 9. Inter-plate calibration

Instrument ROX concentration

ABI 7900HT 300-500 nM

ABI 7900HT FAST 300-500 nM

ABI Viia7 30-50 nM

ABI Viia7 FAST 30-50 nM

ABI 7500 FAST 30-50 nM

ABI StepOnePlus 300-500 nM

ABI 7300 300-500 nM

ABI 7500 30-50 nM

ABI 7700 300-500 nM

ABI 7000 300-500 nM

Page 26: MiRSIGN™ Colon Cancer Test (RUO) - Exiqon

26

miRSIGN Colon Cancer Test (RUO) · Instruction Manual

On cyclers using baseline and threshold values for Cq (Ct) calculations, such as ABI 7900HT, it is important that the proper settings are used . Use of the automatic function of the software for these settings does not seem to produce optimal results for SYBR® Green based assays . Often the baseline is set erroneously on non-detected assays, and this in turn gives false positives, therefore do not use automatic settings . Another issue to consider when using automatic settings is that the settings may differ between plates resulting in data that cannot be compared directly . Inter-plate calibration may not fully resolve this issue, since each assay has a separately calculated baseline and threshold . Instead, both threshold and baseline should be set manually, applying the same settings for all assays on the plate .

The following principles should be applied to manual baseline and threshold settings:

Baseline:The baseline should be calculated in the cycle interval before the amplification takes off (see Figure 3) .

Threshold: The threshold should then be set with the Y-axis in log scale where all assays are in the log linear phase, and the threshold above background for all assays (see Figure 3) . Note: The optimal threshold value may vary between individual machines and experiments .

Tip 10. Guidelines for real-time PCR data collection using ABI instruments

If ROX passive reference dye has not been used in the PCR reactions, make sure the SDS software is set-up without reference dye correction .

Important note

Page 27: MiRSIGN™ Colon Cancer Test (RUO) - Exiqon

27

miRSIGN Colon Cancer Test (RUO) · Instruction Manual

Threshold above background

Amplification take-off

Baseline interval

Figure 5.

Page 28: MiRSIGN™ Colon Cancer Test (RUO) - Exiqon

28

miRSIGN Colon Cancer Test (RUO) · Instruction Manual

The purpose of normalization is to remove technical and biological variation between samples that is not related to the biological changes under investigation . Proper normalization is critical for the correct analysis and interpretation of results from real-time PCR experiments . The most commonly used option for normalization is to use stably expressed reference genes .

In general it is recommended to test several endogenous control candidates (reference genes) before setting up the actual microRNA expression analysis . These candidates should be chosen among genes that can be expected to be stably expressed over the whole range of samples being investigated . They can be stably expressed small non-coding RNA, or stably expressed microRNAs, and chosen based on literature or pre-existing data (e .g . microarray analysis or qPCR panel screening) .

Exiqon offers primer sets for a number of different small RNAs which tend to be stably expressed, and are therefore often good candidates for reference genes . It’s important to keep in mind that in spite of being small non-coding RNAs, most of these are significantly larger than microRNA and therefore may have different extraction efficiency and stability . U6 is one such reference gene which is often used . However, U6 is significantly larger than microRNAs and has a different sub-cellular distribution . The existence of several different isoforms also makes it a suboptimal reference gene . 5S ribosomal RNA is another popular option, but this RNA has a much higher expression level than most microRNAs, and is often found as a PCR contaminant .

If working with blood serum or plasma, please note that only circulating RNA are present . In this case the small non-coding RNAs (5S, U6, SNORs etc) are not good candidates for reference genes, since they are most probably not present in the sample .

Tip 11. Quick guide to normalization of microRNA qPCR experiments

Page 29: MiRSIGN™ Colon Cancer Test (RUO) - Exiqon

29

miRSIGN Colon Cancer Test (RUO) · Instruction Manual

Using stably expressed microRNAs as reference genes offers several advantages such as equal size, extraction efficiency and stability, as well as having expression levels within a similar range of the target microRNAs . Several candidates can be found in the literature, including miR-191, miR-103, let-7a, and miR-16 . Microarray or qPCR panel screening data may also be used when choosing candidate reference genes .

All reference gene candidates should be empirically validated for each study . A number of different software packages exist for evaluating the optimal nature and number of endogenous controls, and for applying multiple endogenous controls for normalizing target expression . One such option is the GenEx software from MulitD, sold through Exiqon with a special application for Exiqon PCR panels . GenEx incorporates both GeNorm and Normfinder for finding the optimal reference genes, and is easy and intuitive to use for the actual normalization .

An alternative option for normalization of data from panels (profiling a high number of microRNAs) is to normalize against the global mean; that is, the average of all expressed microRNAs This can be a good option in samples with a high call-rate (expressed microRNAs), but should be used with caution in samples with low call-rates . It is also not a good option in samples for which the general microRNA expression level is changed .

For further information on normalization and references, we recommend our data-analysis guide available online as well as the guide to microRNA normalization from www .gene-quantification .de .

Page 30: MiRSIGN™ Colon Cancer Test (RUO) - Exiqon

30

miRSIGN Colon Cancer Test (RUO) · Instruction Manual

Troubleshooting guide

Problem Suggestion

PCR signal in samples amplified from first-strand synthesis reactions performed without reverse transcriptase

This typically indicates contamination of the template RNA with genomic DNA . Perform DNase treatment of the RNA sample . If this does not solve the problem RNA samples or other reagent may be contaminated with PCR products .

PCR signal in no-template PCR reaction

This typically indicates contamination of the cDNA template or PCR reagents with amplified PCR product . Exposing the reactions to elevated temperatures (i .e . room temperature) during any part of the protocol increases the risk of background signals . It is important that the reagents and assembled reactions are kept cool (on ice or 4°C) at all times (see Tip 4, p . 53 for details) .

Generated signals are weak

• On some real-time PCR cyclers, gain-settings are adjustable . Make sure the gain settings of your real-time PCR cycler have been set to accommodate the signals generated from the specific assay .

• RNA samples may contain PCR inhibitors . Further purification or an alternative RNA extraction method may be necessary . Check positive controls .

No fluorescent signal is detected during the PCR

Confirm that you have a PCR product by running an aliquot of your PCR reaction on an agarose gel .

No fluorescent signal detected during the PCR, but a PCR amplicon can be detected by agarose gel electrophoresis

• Check that the filter in the real-time PCR cycler was set to either SYBR® Green or FAM/FITC .

• Check that the optical read is at the correct step of the real-time PCR cycles .• Adjust the baseline in the real-time PCR cycler software .

Page 31: MiRSIGN™ Colon Cancer Test (RUO) - Exiqon

31

miRSIGN Colon Cancer Test (RUO) · Instruction Manual

FAQs

What kind of real-time PCR instruments are miRSIGN Colon Cancer Test (RUO) and the miRCURY LNA™ Universal RT microRNA PCR compatible with?miRSIGN Colon Cancer Test (RUO) and the miRCURY LNA™ Universal RT microRNA PCR system in general are compatible with all instruments capable of reading green fluorophores such as fluorescein/FITC/FAM and SYBR® Green . The system has been tested and found to work on real-time PCR instruments from several leading suppliers of this type of instrument .

What kind of settings should I use on my real-time PCR instrument?If your real-time PCR instrument supports fluorophores such as fluorescein/FITC/FAM or SYBR® Green your instrument must be set to detect these fluorophores .

Is miRSIGN Colon Cancer Test (RUO) compatible with other SYBR® Green master mixes?We do not recommend using other SYBR® Green master mixes for real-time PCR analysis with the LNA™ primer sets . The primer sets have been optimized and validated using the miRCURY LNA™ ExiLENT SYBR® Green master mix and the performance of the primer sets will be compromised by using a different master mix (which may contain different salt and/or enzyme concentrations) .

My RNA is already enriched for microRNA, how much should I use in the real-time PCR experiments?miRCURY LNA™ Universal RT microRNA PCR system including the miRSIGN miR-21 Oncogene Assay is developed for use on total RNA and we do not recommend enriching for small RNAs . Samples of enriched microRNAs are difficult to quantitate accurately making it very tricky to ensure the same amount of sample is added to each reaction . If necessary, a total RNA equivalent should be used for the enriched sample, e .g . use a proportional amount of enriched sample resulting from 20 ng of total RNA . It may be necessary to try a couple of different amounts of enriched sample to ensure that the results fall within the linear range of the assay .

Page 32: MiRSIGN™ Colon Cancer Test (RUO) - Exiqon

32

miRSIGN Colon Cancer Test (RUO) · Instruction Manual

What is the recommended experimental set up for Universal RT microRNA real-time PCR?It is generally accepted that the reverse transcription (RT) reaction gives rise to more variation than the PCR reaction . It is therefore advisable to perform replicate RT reactions, ideally 3 separate reactions with 1-2 PCR reactions for each RT . It is further recommended to always include at least three biological replicates (separate RNA extractions) of each sample type in order to allow statistical analysis of the results . If small changes in microRNA expression are expected, it may be necessary to include more replicates to ensure a signifi cant result . In general it is recommended that replicates should be included at any stage during sample procurement, processing, RNA isolation, etc . that could give rise to variation between samples .

A tech note on guidelines for setting up microRNA qPCR experiments can be downloaded at www .exiqon .com/miRNA-qPCR-guidelines

Additional information is available at www .exiqon .com/mirna-pcr

Page 33: MiRSIGN™ Colon Cancer Test (RUO) - Exiqon

33

miRSIGN Colon Cancer Test (RUO) · Instruction Manual

Related products

Exiqon offers a broad variety of tools enabling new discoveries concerning the expression, function and spatial distribution of microRNAs:

miRCURY LNA™ microRNA Detection ProbesFor in situ hybridization and northern blotting of all annotated microRNAs .

miRCURY LNA™ microRNA ISH Optimization kit (FFPE)Complete kit with control probes and hybridization buffer for easy set up of microRNA in situ hybridization .

miRCURY LNA™ microRNA Array, microarray kitPre-printed miRCURY LNA™ microRNA Array microarray slides, available for hsa, mmu & rno and other species . Kit includes hybridization and wash buffers as well as synthetic spike-in microRNAs .

miRCURY LNA™ microRNA Power labeling kitFor fluorescent labeling of microRNAs from total RNA samples ready for array hybridization .

miRCURY LNA™ microRNA Array, ready-to-spot probe setReady-to-spot oligos for direct printing of arrays, or coupling in bead-based applications .

miRCURY LNA™ microRNA Inhibitors, Power Inhibitors, Family Inhibitors and MimicsUnravel the function of microRNAs by microRNA inhibition and overexpression, respectively . Sophisticated LNA™ design ensures potent inhibition and overexpression, respectively, of all microRNAs regardless of their GC content . Chemically modified, highly stable Power Inhibitors for unrivalled potency . Available for in vitro and in vivo studies .

miRCURY LNA™ microRNA Inhibitor LibraryFor genome-wide high throughput screening of microRNA function .

miRCURY LNA™ Target Site BlockersFor inhibition of microRNA binding to a specific mRNA target .

Page 34: MiRSIGN™ Colon Cancer Test (RUO) - Exiqon

34

miRSIGN Colon Cancer Test (RUO) · Instruction Manual

Notice to purchaserExiqon, LNA and miRCURY are registered trademarks of Exiqon A/S, Vedbaek, Denmark . SYBR Green is a licensed trademark of Invitrogen . All other trademarks are the property of their respective owners .

Locked-nucleic Acids (LNAs™) are protected by US Pat No . 6,268,490, US Pat No . 6,770,748, US Pat No . 6,639,059, US Pat No . 6,734,291 and other applications and patents owned or licensed by Exiqon A/S . Products are provided to buyers for research use only . The products in their original or any modified form may be used only for the buyer’s internal research purposes and not for commercial, diagnostic, therapeutic, or other use, including contract research . The buyer may not provide products to third parties in their original or any modified form . The purchase of products does not include or carry an implied right or license for the buyer to use such products in their original or any modified form in the provision of services to third parties, and a license must be obtained directly from Exiqon A/S for such uses .

Use of this product is covered by one or more of the following US patents and corresponding patent claims outside the US: 5,994,056 and 6,171,785 . The purchase of this product includes a limited, non transferable immunity from suit under the foregoing patent claims for using only this amount of product solely in Contract Research, including reporting results of purchaser’s activities for a fee or other commercial consideration, and also for the purchaser’s own internal research . No right under any other patent claim is conveyed expressly, by implication, or by estoppel . Further information on purchasing licenses may be obtained by contacting the Director of Licensing, Applied Biosystems, 850Lincoln Centre Drive, Foster City, California 94404, USA .

Furthermore, this product is provided under an agreement between Molecular Probes, Inc ., a wholly owned subsidiary of Invitrogen Corporation, and EXIQON and the manufacture, use, sale or import of this product is subject to one or more U .S . Patents and corresponding international equivalents . The purchase of this product conveys to the buyer the non-transferable right to use the purchased amount of the product and components of the product in research conducted by the buyer, where such research does not include testing, analysis or screening services for any third party in return for compensation on a per test basis . The buyer cannot sell or otherwise transfer (a) this product (b) its components or (c) materials made using this product or its components to a third party or otherwise use this product or its components or materials made using this product or its components for Commercial Purposes . Commercial Purposes means any activity by a party for consideration and may include, but is not limited to: (1) use of the product or its components in manufacturing; (2) use of the product or its components to provide a service, information, or data; (3) use of the product or its components for therapeutic, diagnostic or prophylactic purposes; or (4) resale of the product or its components, whether or not such product or its components are resold for use in research . For information on purchasing a license to this product for purposes other than research, contact Molecular Probes, Inc ., Business Development, 29851 Willow Creek Road, Eugene, OR 97402 . Tel: (541) 465-8300, Fax: (541) 335-0354 .

Further, the purchase of this product includes a limited, non-transferable license under specific claims of U .S . Patent Nos . 6,174,670 and 6,569,627, owned by the University of Utah Research Foundation and licensed to Roche Diagnostics GmbH and Idaho Technology, Inc ., to use only the enclosed amount of product according to the specified protocols . No right is conveyed, expressly, by implication, or by estoppel, to use any instrument or system under any claim of U .S . Patent Nos . 6,174,670 and 6,569,627, other than for the amount of product contained herein .

For life science research use only . Not for use in diagnostic procedures .

Page 35: MiRSIGN™ Colon Cancer Test (RUO) - Exiqon

35

miRSIGN Colon Cancer Test (RUO) · Instruction Manual