BC-2800 Hematology Analyzer
7.5“ color LCD , 19 parameters and 3 histograms
3-part differential of WBC
automatically cleaning sample probe
Automatic diluting, lyzing, mixing, rinsing and clog-clearing
Count speed: 30tests/hr
3 counting mode : all parameters 、 WBC/HGB 、 RBC/PLT
Automatic flushing, zapping
remain reagents alarm
expired date alarm
build-in recorder
bar code scanner and keyboard option
Up to 10000 sample results (including histograms) can be stored
BC-2800 Specification
BC-2800 Installation---Environment
Environment
The environment should be as free as possible from dust, mechanical vibrations, loud
noises and electrical interference.
Avoid proximity to brush-type motors, flickering fluorescent lights and electrical
contacts that regularly turn on and off.
Avoid placing the instrument in direct sun or in front of a source of heat.
BC-2800 Installation---Space
Space
Comfortable working height
At least 40cm on each side is the preferred access to perform service procedures.
At least 10cm behind for cabling and ventilation
There is enough space for reagents.
BC-2800 Installation---Power
Power
A properly grounded socket supplying 220 or 110 volts
The fluctuation should be in the range of ±10%.
The power frequency should be 50/60Hz.
Power consumption is less than 180VA.
Select some samples in the sample tabular review menu, and then p
ress [F5] to access “extension function” menu.
[PRINT] print the current displayed content
[HELP] help information
[MENU] exit menu
Extension function: “repetition” and “trend graph”
“trend graph” function only open to administrator
Extension function in tabular review menu
Password: 2826
Procedure: press F5 then press F1
Trend graph analysis mean: it is to monitor one patient’s illness trend.
Trend graph
Value of lyse: 0.5mL(whole blood) , 0.36mL(pre-diluted)
Blood needed:
whole blood: 13μL
pre-diluted: 20μL
Count time : less than 2 min / test
Aperture size : 80μm ( diameter ), 70mm ( depth )
Ratio of whole blood Capillary
WBC/HGB 1:308 1:400
RBC/PLT 1:44862 1:44402
Volume of volumetric testing: 300μL ×2times
BC-2800 sample and dispense
Range of linear and display
Parameter
Linear Display
WBC 0 - 99.9 ×109/L 0.0-499.9 ×109/L
RBC 0 -9.99×1012/L 0.0-9.99 ×1012/L
HGB 0 - 300g/L 0- 300g/L
MCV ------ 0-250 fL
PLT 0 - 999×109/L 0-2999 ×109/L
%100Mean
DiffCV
Parameter linear Repetition(CV%)
WBC4.0-7.5×109/L ≤ 3.0
7.6-15.0×109/L ≤ 2.0
RBC3.0-4.5×1012/L ≤ 2.0
4.5-6.5×1012/ L ≤ 1.5
HGB 110 – 180g/L ≤ 1.5
MCV 70.0 – 100.0fL ≤ 1.0
PLT100-200 ×109/L ≤ 5.0
201 - 500×109/L ≤ 4.0
Repetition
AC Input :
AC 220V±30V 50/60Hz±1Hz
Consumption :
150 VA
Fuse :
250 V 2A
Power Supply
Without any operation in 10 minutes, the system will enter the
saver program.
LCD display turns off
Power indicator is still light.
Any key exists the saver program.
Screen Saver
Power indicator
Start key
7.5” color LCD
Keypad
Record
Sample probe
Front Panel
Power switch and fuse
Equipotential terminal
Diluent coonector
Rinse connector
Lyse connector
Waste connector
Rear panel
Keyboard
Power or floppy
Transition
Printer or upgrade interface
Scanner interface
Left side
Display featureTitle Count Mode System
StatusSystem time
Window
Error Messages Reagent Status
MENU On-line Help
1 ) >99tests : remain reagents and remain waste container are enough for 99tests
2 ) 0 test : any one of the remain reagents or remain waste container is not enough for 1 test. The warning message will be show in error area. Actual tests : show the actual tests number
Remain reagents
注意: PS/2 标准键盘操作的区别
用 PS/2 标准键盘执行模式切换功能键是: [Ctrl + A]
Keypad
Menu tree
Count
Revi ew
Cal i brat i on Manual
Auto
Qual i ty control
Password
Sel f - test
Log
Status
Servi ce Mai ntenance
Confi g
Pri nt
Shutdown
Control s
X-B anal ysi s
Sett i ngs
Setup
Auto - f resh bl ood
Diluent Rinse LyseE-Z
cleanser
Normal start-up
46.5ml( clean & once background test)
16.5ml 0.5ml ---
75ml( clean & twice background test
)26ml 1ml ---
All parameters(whole blood)
28.5ml 9.5ml 0.5ml ---
All parameters(pre-diluted)
30.5ml9.5ml 0.36ml ---
Normal shut down 13.5ml 5.5 ml --- 1ml
Reagents consumption
The last shut down statusProcedure
Normal shutdown clean tubing and background testingAbnormal shutdown( Directly shutdown )
Reset motors, check software, clean tubing and background test
Package or empty tubing Prime tubing, clean tubing several times, background test.
1. Switch on BC-2800 , and the power indicator will be light. 2. The Initializing procedure will continue about 80s. 3. Self test procedure.
4 The system will enter “count” menu. If there are errors during startup, the error message will be displayed on left bottom of screen.
Power on and shut down
Background test
Parameter Acceptable limit
WBC 0.3 109 / L
RBC 0.03 1012/ L
HGB 1 g / L
HCT 0.5 %
PLT 10 109 / L
1. Only ID mode
2. All information
3. Batch information mode
Sample information input
1. Press [MODE] to select the prediluted mode (any mode preceded by PB). 2. Press [DILUENT] to enter the Add Diluent screen3. Prepare a clean sample cup under the sample probe and press [START] to dispense 1.6ml diluent to the cup.
4. Take out the sample cup after the sample probe rose up and press [ENTER] to exit the Add Diluent screen. 5. Collect 20μL capillary blood specimen and immediately add it to the sample cup and shake to mix them.
Preparation of pre-diluted blood
[F1] is to select the setting items.
[SETUP]-[Settings]
Reagent Valid volume
Diluent ( 0-30.0 ) L
Rinse ( 0-30.0 ) L
Lyse ( 0-999 ) mL
Waste ( 0-99.9 ) L
[SETUP]-[Settings]--Reagents Remain
[SETUP]-[Settings]—Print & Communication
[SETUP]-[Settings]—Print & Communication
Printing Device
Format1 Format2 Format3 Format4
Recorder Parameters + histograms
Parameters Parameters+ histograms
Parameters
Printer Whole page with histograms
Whole page without
histograms
Whole page with histograms
Whole page without
histograms
[SETUP]-[Settings]—Date & time
[Setup]-[Settings]---[Gain]
[Setup]-[Settings]---[Gain]--WBC
• too small WBC gain that should be increased
• too large WBC gain that should be decreased.
[Setup]-[Settings]---[Gain]--RBC
• When the difference between the actual MCV analysis result and the expected result exceeds 6%, need to change the RBC gain.
• For example, assuming the expected MCV result is 90.0fL, while the actual analysis result is 82.0fL, the gain factor as below:
098.10.82
0.90=
ActualMCV
VExpectedMC
[Setup]-[Settings]---[Gain]--HGB
• adjust the HGB gain to change the HGB blank voltage, which usually should be set between 4.3 ~ 4.5V.
[Setup]-[Settings]--count
[Setup]-[Settings]--Reference RangePatient Group Sex Age
General Not specified Not specified
Not specified > 13 years
Male adults Man > 13 years
Female adults Woman > 13 years
Pediatric patients
Not specified>= 1 month and <= 13
years
Neonates Not specified <1 month
[Setup]-[Settings]--other
[Service]-[Maintenance]--Prime
[Service]-[Maintenance]—Zap & Flush
[Service]-[Maintenance]—Probe cleanser clean
• If your analyzer is to run 24 hours a day, you should perform this procedure every 3 days.
• If you follow the shutdown procedure to turn off your analyzer everyday, you should perform this procedure every week.
• To make sure this analyzer functions normally, every time the accumulated analyzed samples reach 200, a dialog box will pop up to remind the user to perform the Probe cleanser cleaning
procedure.
“Lyse test” is to check the injected volume for lyse is correct or not. When shall we use this function?When the WBC result or its histogram is abnormal, we should run this function.
Screw off these two screws
[Service]-[Maintenance]—Lyse test
Remove the Count bath shield cover
The injected lyse must over the second mark
2ml lyse will be dispensed to the count bath.
[Service]-[Maintenance]—Lyse test
[Service]-[Maintenance]—Drain
• Drain bath
Empty the count bath, and prime it again
• Drain Tubing
Empty all tubing, then turn off the equipment
[Service]-[Maintenance]—Clean
• Clean bathUse diluent to clean count bath
• Clean wipe blockAfter being used for a long time, the bottom of the probe wipe
block may be contaminated by blood.
When the accumulated analyzed samples reach 4,000, a dialog box will
pop up to remind the user to clean the wipe block
[Service]-[Maintenance]—Prepare to ship
• Empty the tubing
• Wash the tubing
• Empty the tubing
• Turn off the equipment
[Service]-[System test]
The available self-test items are divided into four groups, the
Tubing, Machine, Valve and Circuit.
[Service]-[self-test]
[Service]-[self-test]—[Test tubing]
Count Time WBC( S) & Count Time RBC( S) It measures the duration of a WBC or RBC countAperture(v) It measures the voltage (v) over the aperture. Vacuum It checks whether the vacuum system functions normally. Pressure It checks whether the system flushes the aperture at a normal press
ure. Filter It checks whether the filter functions normally.
[Service]-[self-test]—[Machine]
[Service]-[self-test]—[Valve]
It is to monitor that WBC, RBC and PLT signals can be properly converted
into digital signals or not.
[Service]-[self-test]—[Circuit]
The recorded events are divided into four groups, All, Para set, Error and Other
(including setting discriminators, system self-test and updating system software)
This analyzer can save maximum 100 log files and once the maximum number has been reached, the newest log will automatically cover the oldest one.
[Service]-[Log]
Every screen displays 13 items
It only shows BC-2800 current configuration, no change.
[Service]-[Configuration]
The printing tasks are queued in this screen
[Service]-[Print management]
[Calibration]---[Manual]
Password: 2826
Valid range:75%---125%
[Calibration]---[Auto]
Password: 2826
Valid range:75%---125%
[Calibration]---[Fresh blood]
Password: 2826
Valid range:75%---125%