Mid-InfraRed Medium Resolution Echelle Spectrometer (MIRMES)

38
Mid-InfraRed Medium Resolution Echelle Spectrometer (MIRMES) Itsuki Sakon (Univ. of Tokyo) Yuji Ikeda(Photocoding) Naofumi Fujishiro(Cybernet) Hirokazu Kataza (ISAS/JAXA) Takashi Onaka (Univ. of Tokyo), SPICA pre-project team

description

Mid-InfraRed Medium Resolution Echelle Spectrometer (MIRMES). Itsuki Sakon (Univ. of Tokyo) Yuji Ikeda(Photocoding) Naofumi Fujishiro(Cybernet) Hirokazu Kataza (ISAS/JAXA) Takashi Onaka (Univ. of Tokyo), SPICA pre-project team. Outline. - PowerPoint PPT Presentation

Transcript of Mid-InfraRed Medium Resolution Echelle Spectrometer (MIRMES)

Page 1: Mid-InfraRed Medium Resolution Echelle Spectrometer (MIRMES)

Mid-InfraRed Medium Resolution Echelle Spectrometer (MIRMES)

Itsuki Sakon (Univ. of Tokyo)

Yuji Ikeda(Photocoding)

Naofumi Fujishiro(Cybernet)

Hirokazu Kataza (ISAS/JAXA)

Takashi Onaka (Univ. of Tokyo),

SPICA pre-project team

Page 2: Mid-InfraRed Medium Resolution Echelle Spectrometer (MIRMES)

Outline

The Mid-Infrared Medium-Resolution Eschelle Spectrometer (MIRMES) is one of the focal plane instrument onboard SPICA mission in the pre-project phase. It is designed particularly for measuring the intensity and the profile of lines from ionized gas and molecules as well as the detailed spectral structure of dust band features of various compositions in the wavelengths continuously from 10 to 40 micron with moderately high spectral resolution power that is almost comparable to that of SAFARI in the far-infrared.

MIRMES consists of two channels; Arm-S covers the wavelengths from 10.3 to 19.9 micron with the resolution power of R~1200 and Arm-L covers from 19.9 to 36.0 micron with R~750. They share the same field of view by means of a beamsplitter. The FOV size for Arm-S and –L is 12” x 8”.5 for Arm-S and 14” x 12”.5 for Arm-L. The FOV is split into 5 rows by using the integral field spectroscopy (IFU) unit.

Page 3: Mid-InfraRed Medium Resolution Echelle Spectrometer (MIRMES)

Scientific Objectives/Targets& Required Specifications

Page 4: Mid-InfraRed Medium Resolution Echelle Spectrometer (MIRMES)

Exploring the process of cosmic recycling among gas, molecules and dust particlesin the context of chemical evolution history of the universe is one of the most important objective of the SPICA mission. Particularly, the following observational approaches are crucial to achieve this objective;

1) infrared spectroscopic diagnostics of the composition and the properties of dust and molecules formed in the mass loss winds from the evolved massive stars including thesupernovae (SNe), Wolf-Rayet (WR) stars, and Luminous Blue Variables (LBVs)2) infrared spectroscopic diagnostics of the composition and the properties of dust and molecules formed in the mass loss wind from low- to intermediate-mass evolved stars including post-Asymptotic Giant Branch (AGB) stars, Planetary Nebulae (PNe) and novae3) infrared spectroscopic diagnostics of the composition and the properties of molecules synthesized in the atmosphere of evolved low- to intermediate-mass stars ・ infrared spectroscopic analysis of SN remnants in the Milky Way and in nearby galaxies to understand how much fraction of newly condensed dust in the SN ejecta is destroyed and how much of them survives the shocks4) infrared spectroscopy of dense molecular clouds with embedded young stellar object to identify the infrared bands of iron sulfide grains and to systematically understand the role of cold dense molecular clouds as the site of dust synthesis and the grain growth

Scientific Targets

Page 5: Mid-InfraRed Medium Resolution Echelle Spectrometer (MIRMES)

5) Infrared spectroscopy of various ISM structures in nearby galaxies to demonstrate the cosmic recycling among ionic gas, molecules, and dust particles on a galactic scale.

6) Infrared spectroscopic diagnostics of ISM properties of remote galaxies, which provide us unique physical parameter space in terms of metallicity and morphology.

All these observational approaches require spectroscopic abilities covering thoroughly from 10m to 40m with moderately high (R>1000) spectral resolution power in order to measure the intensity and profiles of ionic lines, molecular lines and various dust band features in this wavelength regime. In addition, observations of time-varying objects such as supernovae, LBVs, WRs and novae are indispensable to achieve our scientific purpose, simultaneous operation of shorter wavelength module covering from 10-20m and longer wavelength module covering from 20-40m is indispensable. Moreover, accuracy in the absolute flux level and, especially, their relative consistency between the shorter wavelength module covering from 10-20m and longer wavelength module covering from 20-40m is indispensable to obtain the accurate ionic line ratios and profiles of dust band features distributed widely in 10-40m. The SPICA/MIRMES is the instrument that is designed to fulfill those requests given above.

Scientific Targets

Page 6: Mid-InfraRed Medium Resolution Echelle Spectrometer (MIRMES)

Scientific Targets

Key Scientific Targets1. Molecules and dust formation in the ejecta of Core-collapse Supernovae,

Luminous Blue Variables (LBVs) and WR stars

2. Destruction and survival of dust in Supernova remnants (SNRs)

in the Milky Way, Magellanic Clouds, and in Nearby galaxies

3. Molecular Chemistry in the atmosphere of AGB stars and PNe

4. Dust formation in recurrent Novae and Type Ia supernovae

5. Molecules and dust formation in dense clouds with embedded YSOs

6. Cosmic recycling taken place within nearby galaxies

7. Mid-infrared Spectroscopic diagnostics of ISM condition in remote galaxies

8. Mid-infrared spectroscopy of emission lines associated with warm (100-1000K) gas in proto-planetary disks

9. Distribution and physical state of solid materials in proto-planetary disks and dust disks in the main-sequence stars.

Page 7: Mid-InfraRed Medium Resolution Echelle Spectrometer (MIRMES)

Consistency with MRD

1. Multi-epoch MIR spectroscopy of Core-collapse Supernova, LBVs and WR stars

consistent with “Life Cycle of Dust; Objective #1” in MRD.

2. Demonstrating Material Circulation in Supernova remnant

consistent with “Life Cycle of Dust; Objective #3” in MRD.

3. Understanding the Chemistry in the atmosphere of AGB stars and PNe

consistent with “Life Cycle of Dust; Objective #2” in MRD.

4. Measuring the Dust Yields by Nova, Type Ia supernova

consistent with “Life Cycle of Dust; Objective #2” in MRD.

5. Understanding the Dust formation in Dense Interstellar Clouds

consistent with “Life Cycle of Dust; Objective #4” in MRD.

6. Understanding the cosmic recycling taken place within nearby galaxies

consistent with “Life Cycle of Dust; Objective #5” in MRD.

7. MIR Spectroscopic diagnostics of ISM conditions in remote galaxies

consistent with “Extragalactic Science; Objective #3 & #4” in MRD.

8. Mid-infrared spectroscopy of lines emitted from warm gas in protoplanetary disks

consistent with “Planetary System; Objective #2” in MRD.

1. Understanding the role of solid state materials for planet formation.

consistent with “Planetary System; Objective #6” in MRD.

Page 8: Mid-InfraRed Medium Resolution Echelle Spectrometer (MIRMES)

Specification of MIRMES

ARM-S   ARM-L

array format Si:As (2k x 2k) 25 m/pix

Si:Sb (1k x 1k) 18 m/pix

Wavelength coverage 10.0m-19.9m 19.9m-35.0m

Spectral resolution (R=/) ~1400 ~750

pixel scale 0.409 (“/pix) 0.413 (“/pix)

Slit width 1”.7 (4.2pixel) 2”.5 (6.1pixel)

FOV size 12” x 8”.5 14” x 12”.5

Arm-S Echelle order min (m) max (m)

4 15.53 19.97 5 12.71 15.53 6 10.75 12.71 7 (9.98) 10.75

※ min and max are defined as the wavelength at which the grating efficiency drops to 40% of the peak

Arm-L Echelle order min (m) max (m)

5 29.4 35.0 6 24.4 29.4 7 21.2 24.4 8 (19.9) 21.2

Page 9: Mid-InfraRed Medium Resolution Echelle Spectrometer (MIRMES)

Specification of Instrument

Echelle Formats on detector arrays of Arm-S and Arm-L

Page 10: Mid-InfraRed Medium Resolution Echelle Spectrometer (MIRMES)

Concept Study Current Status

Page 11: Mid-InfraRed Medium Resolution Echelle Spectrometer (MIRMES)

Optics & Optical Elements

Fore-Optics for MIRMES Image Slicer (with 5 slitlets)

side view

top view

Page 12: Mid-InfraRed Medium Resolution Echelle Spectrometer (MIRMES)

Optics & Optical Elements

Spectrograph for MIRMES/Arm-S

Si:As detector Array (2k x 2k ; 25um)

Pseudo Slit (60”.8)

30mmCollimator mirrors

Cameramirrors

Page 13: Mid-InfraRed Medium Resolution Echelle Spectrometer (MIRMES)

Optics & Optical Elements

Pseudo Slit (70”)

30mm

Collimator mirrors

Cameramirrors

Spectrograph for MIRMES/Arm-L

Si:Sb detector Array (1k x 1k ; 18um)

Page 14: Mid-InfraRed Medium Resolution Echelle Spectrometer (MIRMES)

Detectors

MIRMES/Arm-S (TBD; same as that used for MIRACLE)Si:As 2kx2k (Raytheon)

Pixel pitch; 25um/pix

Dark current; 0.1e/sec

Full well; 1.0x106 (electron/pix)

Thermal output; 1mW

Quantum Efficiency; N/A

MIRMES/Arm-L (TBD; same as that used for MIRACLE)Si:Sb 1kx1k (DRS)

Pixel pitch; 18um/pix

Dark current; 1e/sec

Full well; 1.0x106 (electron/pix)

Thermal output; 1mW

Quantum Efficiency; N/A

Page 15: Mid-InfraRed Medium Resolution Echelle Spectrometer (MIRMES)

Volume & Structure

Total Volume MIRMES/Arm-S; 300 x 300 x 150MIRMES/Arm-L; 200 x 250 x 150 (in units of mm)

Page 16: Mid-InfraRed Medium Resolution Echelle Spectrometer (MIRMES)

Thermal Design

-No driving modules in MIRMES

-Calibration lamp & shutter for dark measurements are shared with MIRACLE

Detectors (parasitic) (active)

Si:As 2K x 2K array x 1 (5K)

4.5K J-T stage  ・・・ 0 mW 1 mW (TBD)

Si:Sb 1K x 1K array x 1 (3K)

1.7K J-T stage  ・・・ 0 mW 1 mW (TBD)

Wire (20K) 12K 2ST stage  ・・・     1 mW 50 mW (TBD)

Annealing function

for each detectors ・・・ 0 mW N/A mW (TBD)

Page 17: Mid-InfraRed Medium Resolution Echelle Spectrometer (MIRMES)

Expected Performance

Assumptions;-Surface Brightness of the Background

dominated by zodiacal emission

(cf. Reach et al. (2003));

Low-background case; Zodiacal emission modeled by blackbody of Tdust= 274.0K normalized at the 25m flux of 15.5MJy/sr

High-background case; Zodiacal emission modeled by blackbody of Tdust= 268.5K normalized at the 25m flux of 79.42MJy/sr

-Spatial scale of 1 pixel in the sky;

0”.409 for Arm-S, 0”.413 for Arm-L

-Effective area of the primary mirror;

(3.0/2)2 x 0.8 [m2]

-Slit Width;

1”.7 (4.2pix) for Arm-S, 2”.5 (6.1pix) for Arm-L

-Effective Image Size;

1”.7 (4.2pix) for Arm-S, 2”.5 (6.1pix) for Arm-L

-Optical System Efficiency (including filters, mirror transmittance, beamsplitter, etc.); 0.3 for Arm-S, 0.3 for Arm-L-Dark Current; 0.1[e/sec] for Si:As 2Kx2K array (Arm-S), 1.0[e/sec] for Si:Sb 1Kx1K array(Arm-L)-Readout Noise; 10[e] for Arm-S, 20[e] for Arm-L-Maximum lamp time per exposure; 600 [sec]

-Minimum lamp time per exposure; 2 [sec]

-Full well per pixel;

1.0x106[e/pix] for Arm-S

1.0x106[e/pix] for Arm-L

-Linearity warranty;

0.5 x full well

-Maximum encircled energy fraction

in a central pixel for a point source;

0.12 for Arm-S, 0.06 for Arm-L

Page 18: Mid-InfraRed Medium Resolution Echelle Spectrometer (MIRMES)

Expected Performance

SPICA/MIRMES Sensitivity for point sources

Page 19: Mid-InfraRed Medium Resolution Echelle Spectrometer (MIRMES)

Expected Performance

SPICA/MIRMES Sensitivity for diffuse sources

Page 20: Mid-InfraRed Medium Resolution Echelle Spectrometer (MIRMES)

Expected Performance

SPICA/MIRMES Saturation for point sources

Page 21: Mid-InfraRed Medium Resolution Echelle Spectrometer (MIRMES)

Resource Requirements

Page 22: Mid-InfraRed Medium Resolution Echelle Spectrometer (MIRMES)

Field-of-View Requirement

FOV size for Arm-S; 12”x8.5”

for Arm-L; 14”x12”.5

- Arm-S and Arm–L share the same field of view by means of a beam splitter installed in the fore-optics.

- Each FOV is divided into 5 slitlets with image slicer.

- The size of each slitlet;

12” x 1”.7 for Arm-S

14” x 2”.5 for Arm-L

12

”.58”.

5

12”.0

14”.0FOV of Arm-L

FOV of Arm-S

Page 23: Mid-InfraRed Medium Resolution Echelle Spectrometer (MIRMES)

Thermal & Cryogenic Requirement

MIRMES  Requirement

Remarksunit parasitic Active Total

1.7K J-T stage       parasitic : instrument offactive : ON, incl. parasitic temperature of stage 1.7 K

temperature range requirement K 3KTemp. at detectorSi:Sb 1kX1K x 1

Low temp to reduce dark

temperature stability requirement mK 100mK

Average lift mW 0.0 1.0 1.0

Peak lift mW 0.0 1.0 1.0

4.5K J-T stage        

temperature of stage 4.5 K  

temperature range requirement K 5.0K Temp. at detector

temperature stability requirement mK 100mK  

Average lift mW 0.0 1.0 1.0 Si:As 2kX2K x 1Low temp to reduce dark Peak lift mW 0.0 1.0 1.0

12K 2ST stage        

Heat sink for wirePre-Amp at low temp.

stage

temperature of stage 12K

temperature range requirement K 20K

temperature stability requirement K 1K

Average lift mW 1.0 50.0  

Peak lift mW 1.0 50.0  

Page 24: Mid-InfraRed Medium Resolution Echelle Spectrometer (MIRMES)

Pointing / Attitude control Requirement

Requested pointing/attitude control accuracy; 0”.425

The widths of slitlet of Arms-S and –L are 1.7” and 2.5”, respectively.

The pointing accuracy corresponding to the ¼ of the width of slitlet

is requested; 1”.7 x 0.25 ~ 0”.425 for Arm-S

2”.5 x 0.25 ~ 0”.625 for Arm-L

Page 25: Mid-InfraRed Medium Resolution Echelle Spectrometer (MIRMES)

Structural Requirement

TBD

Page 26: Mid-InfraRed Medium Resolution Echelle Spectrometer (MIRMES)

Data Generation Rate & Data Handling Requirement

Simultaneous readout of data of Arms-S and -L

1 pixel = 16bit(=2Byte), 2K x 2K pixels = 8.4MB (Arm-S), 1K x 1K pixels = 2.1MB (Arm-L)

(1) The case of longest ramp time (texp=600 sec)

If we try to downlink the data sampled before and after the reset,

the data generation rate becomes; 10.5(MB) x 2 / 600(sec) = 35KBps

(2) The case of shortest ramp time (texp=2.0)

If we try to downlink the data sampled before and after the reset,

the data generation rate becomes; 10.5(MB) x 2 / 2(sec) = 10.5MBps

If we calculate the differential between data sampled before and after the reset, integrate 4 exposures, and downlink only the result, the data generation rate becomes;

18bit /16bit x (8.4MB +2.1MB) / 2(sec) / 4 = 1.48MBps (preferred)

 Onboard computer that can handle the image operation is requested

Page 27: Mid-InfraRed Medium Resolution Echelle Spectrometer (MIRMES)

Warm Electronics

TBD(common among MIRACLE, MIRMES and MIRHES)

FunctionComponent

#. Pcs.

Power (W)

Power Dissipation

Power (W)

Power Dissipation

Focal-Plane Array 2 9 9 9 9

observing standby

Page 28: Mid-InfraRed Medium Resolution Echelle Spectrometer (MIRMES)

Operation & Observing Mode

Operation; TBD

power data generation rate

(W) (MB/s)

Parasitic 0 0

Active

-standby TBD < 0.01

-observing

single texp=2 (sec) TBD 1.48 ※

single texp=20/60/120/600 (sec) TBD 1.05/0.35/0.18/0.04

step texp=20/60/120/600 (sec) TBD 1.05/0.35/0.18/0.04

-calibration

dark texp=2/20/60/120/600 TBD 1.48/1.05/0.35/0.18/0.04

cal. lamp on texp=2(sec) TBD 1.48※

※4 exposure cycles are 1 unit.

Single mode Parameters; Texp, ncycle

- dithering mode required Step Mapping mode Parameters;

Texp, dstep , nstep, mstep, ncycle

step

off

set

dste

p

/・・・

12

”.58”.

5

12”.0

14”.0FOV of Arm-L

FOV of Arm-S

・・・

nstep

mstep12”.0

Page 29: Mid-InfraRed Medium Resolution Echelle Spectrometer (MIRMES)

Development and Test Plan

Page 30: Mid-InfraRed Medium Resolution Echelle Spectrometer (MIRMES)

Key Technical Issues & TRL

SPICA/MIRMES-optics design(I) ; almost completed.

Volume reduction is under consideration

-structure design(I); starting analyses with SHI (~ 30 Apr. 2010)

-optics design(II); Not yet (scheduled from 1st May 2010 ~)

-structure design(II); Not yet (scheduled from 1st May 2010 ~)

-beam splitter / filter; starting analyses with JDS Uniphase (Feb. 2010 ~)

-slice mirror; starting analyses of manufacturing the slice mirror

- detector/ electric design; common with MIRACLE

Page 31: Mid-InfraRed Medium Resolution Echelle Spectrometer (MIRMES)

Development Plan

Page 32: Mid-InfraRed Medium Resolution Echelle Spectrometer (MIRMES)

Test & Verification PlanFMPM

MIRMES Prototype Model(PM) test & Verification (2013/6—2014/6)Room temperature optical source test (optical alignment check)Cryogenic temperature infrared source test (infrared alignment check, detector electric circuit test)Room temperature vibration testCryogenic temperature vibration test

MIRMES Flight Model(FM) test & Verification (2015/1—2015/9)Room temperature optical source test (optical alignment check)Cryogenic temperature infrared source test (infrared alignment check, detector electric circuit test)Room temperature vibration testCryogenic temperature vibration testCalibration (wavelength, flux)

MIRMES FM combination test with MIRACLE (2015/9—2016/3)Room temperature optical source test (optical alignment check)Cryogenic temperature infrared source test (infrared alignment check, detector circuit test)Room temperature vibration testCryogenic temperature vibration test

Page 33: Mid-InfraRed Medium Resolution Echelle Spectrometer (MIRMES)

Development Cost

Development Cost (x 1,000 JPY)Optical design(I) 1,000 (Arm-S) + 1,000 (Arm-L) [Photocoding, Cybernet]Structure design(I) 10,000 (Arm-S+Arm-L) [SHI]Optical design(II) 1,000 (Arm-S) + 1,000 (Arm-L) [Photocoding. Cybernet]Structure design(II) 10,000 (Arm-S+Arm-L) [SHI]

Detector, Detector Circuits; common development system with MIRACLE   primary investment 200,000 Si:As 2Kx2K 100,000 x 2 (FM + PM) [Laytheon] Si:Sb 1Kx1K 100,000 x 2 (FM + PM) [DRS]

Beamsplitter, filter; 1 sample test 2,000 x 5 (TBD; negotiation with JDS Uniphase) Slice mirror; 2,000 x 2 (FM + PM)Mirror; common development system with MIRACLE

PM x 2 + FM; 1,000,000 (TBD; negotiation with SHI) Test & Verification; 100,000

Page 34: Mid-InfraRed Medium Resolution Echelle Spectrometer (MIRMES)

Observing Program

Page 35: Mid-InfraRed Medium Resolution Echelle Spectrometer (MIRMES)

Observation Plan to perform Science Targets

1. Multi-epoch MIR spectroscopy of Core-collapse Supernova, LBVs and WR stars

see “Life Cycle of Dust; Objective #1” of MRD in details.

2. Demonstrating Material Circulation in Supernova remnant

see “Life Cycle of Dust; Objective #3” of MRD in details.

3. Understanding the Chemistry in the atmosphere of AGB stars and PNe

see “Life Cycle of Dust; Objective #2” of MRD in details.

4. Measuring the Dust Yields by Nova, Type Ia supernova

see “Life Cycle of Dust; Objective #2” of MRD in details.

5. Understanding the Dust formation in Dense Interstellar Clouds

see “Life Cycle of Dust; Objective #4” of MRD in details.

6. Understanding the cosmic recycling taken place within nearby galaxies

see “Life Cycle of Dust; Objective #5” of MRD in details.

7. MIR Spectroscopic diagnostics of ISM conditions in remote galaxies

see “Extragalactic Science; Objective #3 & #4” of MRD in details.

8. Mid-infrared spectroscopy of lines emitted from warm gas in protoplanetary disks

see “Planetary System; Objective #2” of MRD in details.

1. Understanding the role of solid state materials for planet formation.

see “Planetary System; Objective #6” of MRD in details.

Page 36: Mid-InfraRed Medium Resolution Echelle Spectrometer (MIRMES)

Outline of Ground Data Processing

All the procedures used for the data processing of MIRMES are generally same as those of Subaru/COMICS.

The data reduction pipeline and contribution softwares are prepared during the performance verification phase.

The calibration datasets (monitoring standard star, flat fielding, ) are taken regularly (say, once a month) to check the stability of the on-orbit performance.

Page 37: Mid-InfraRed Medium Resolution Echelle Spectrometer (MIRMES)

Organization & Structure for Development

※Detector, electric circuit, mirrors; common with MIRACLE team

General: I. Sakon (Univ. of Tokyo), H.Kataza (ISAS/JAXA),

T. Onaka (Univ. of Tokyo) [adviser]

Optical Design : Y. Ikeda (Photocoding) ・ N.Fujishiro (CYBERNET.Co.)

Structural Design : Sumitomo Heavy Industry (SHI)

Detector : T. Wada, H.Kataza (ISAS/JAXA) ・ technical staff 1

Electric Circuit : technical staff 2

Test and Verification of MIRMES (including development of beam splitter, filters, and slice mirrors, experiments and performance test) :

I. Sakon (Univ. of Tokyo), 1-2 graduate school students, technical staff 3

Sciene Discussion : I.Sakon, T.Onaka (Univ. of Tokyo), Y.Okada (Univ. of Cologne), H.Kaneda (Nagoya University), T. Nozawa (IPMU), T. Kozasa (Hokkaido Univ.), M. Matsuura (University College London) [TBD]    

Page 38: Mid-InfraRed Medium Resolution Echelle Spectrometer (MIRMES)

Summary