  • 8/20/2019 M.I. URK700.pdf

    1/25

    OPERATION MANUAL

    AURO REF / KERATOMETER URK-700

  • 8/20/2019 M.I. URK700.pdf

    2/25

    UNICOS Co., Ltd. www.e-unicos.com

    URK OPERATION MANUAL

    Contents

    1. Features

    2. Notes forUsing the Instrument

    3. Description

    3.1Main Unit

    3.2OperationButtons

    4. Practicing throughModel Eye

    5.Measurement

    5.1Continuous KeratometryandRefractometry

    5.2Refractometry

    5.3Keratometry

    5.4ContactLens BaseCurve Measurement

    5.5Corneal diametermeasurement

  • 8/20/2019 M.I. URK700.pdf

    3/25

    6. OtherMeasurementModes

    6.1SETUP Mode

    1. Features

    1. Various Measurements Supported.

    Not only the usual refractometry and keratometry, but also corneal

    diameter and base curve of contact lens can be measured with this

    one instrument. Thus, measurements of eye and prescriptions for

    glasses and contact lenses can be made more efficiently.

    2. Wide Dioptric Measurement Range

    Because theURK-700 covers a wide measurement range, from

    -25D to +22D, evenanexaminee with strongmyopia canbemeasured.

    3.More accurate Measurement.

  • 8/20/2019 M.I. URK700.pdf

    4/25

    The foggy method of the eye fixation target makes examinee

    ’s eye

    comfortable and enables to getmore accurate measurementdata.

    4. EasyConnection with otherEquipments

    This instrument is designed to connect other equipments such as VCR

    orTV monitor.

    2. Notes forUsing the Instrument

    1. Do not hit or drop the instrument. The instrument may be damaged

    when it receives a strong impact. The impact may cause damage the

    function of this instrument. Pleasedohandlewith care.

    2. An exposure to the direct sunlight or too bright indoor lights may

  • 8/20/2019 M.I. URK700.pdf

    5/25

    influence on the resultof accurate measurement.

    3. To connect this with other relevant equipments, consultwith the dealer

    about the wayto make it.

    4. A sudden heating of the room in cold areas will cause condensation on

    the protective glass in the monitor screenand onoptical parts inside the

    instrument. In this case, just wait until condensation disappears before

    performing measurement.

    5. Keep the objective glass of the examinee side clean. If itwas stained, it

    maycause onERROR or inaccurate measurements.

    6. In case there is smoke, strange odor ornoise on working disconnect the

    power supply and consult thedealer.

    7. Don

    ’t use organic solution such as alcohol, thinner, benzene, etc. to

    clean the surface of this instrument. Itmaydamage the instrument.

  • 8/20/2019 M.I. URK700.pdf

    6/25

    8. When moving this URK-700, fix the stage by using clamping bolt and

    stage holding knob, always keep power off, and then lift the bottom of

    theunitwith both hands.

    9. If you leave URK-700 without using for certain period, disconnect the

    power supply and protect theunitwith dust cover.

    3.Description.

    3.1MainUnit.

  • 8/20/2019 M.I. URK700.pdf

    7/25

    LCD Monitor

    Functional Key

    Measurement

    button

    Stage Clamping

    Bolt

    Printer

  • 8/20/2019 M.I. URK700.pdf

    8/25

    Headrest

    Measurement

    Chinrest

    EXT Video RS-232Port Power IN LET Fuse PowerSwitch

    StageHolding

  • 8/20/2019 M.I. URK700.pdf

    9/25

    3.2 Operation Panel

    Title Technical skill

    MODE To change measurementmode

    CLEAR Toerase measured data on the screen

    MENU To change andset up the function

    PRINT To print the dataof measurement

    IOL Tomeasure the patient’s eye with cataract or IOL intra-ocular lens

  • 8/20/2019 M.I. URK700.pdf

    10/25

    4. Practicing byModel Eye

    Practice measurement by using the TestModel Eye before doing the measurement of

    patient’s eye.

     1 Turn ON thepower

    Turn ON the powerswitchonthe lower left side.

     2 Setthemodel eye on

    Remove the chin rest paper and align the holes on the base of the model eye with the

    holes onthe chinrest. Andinsert pins.

     3 Release stage lock

    Fully turn the bolt in the underneath cover in counter clockwise direction. And turn the

    stage holding bolt incounter clockwise directionto release thestage lock.

     4 AdjustheightofTestModel Eye.

    Adjust the height of the chin rest by rotating Height Adjustment knob so that The Test

    Mode Eye is aligned with theHeightAdjustmentmark on theheadrest.

     5 Enter K/R or REF Mode

    If neither K/R nor REF is displayed on the monitor press MODE button until either one

    is displayed.

     6 Adjust theposition and focus ontheModel Eye.

    Looking at themonitor incline the operation joystick toward the Model Eye until a bright

    dot appears near the inner alignment mark. Place the bright dot in the center of inner

    alignmentmark. Focus on theModel Eye by inclining theoperation joystick forward and

    backward so thatKERATO RING image is displayed clearly on the monitor, at the time

    adjusted thebar wouldbeappearedabove the inner alignmentmark.

     7 Measurement

      Manual measurement

    A. Adjust the positionand focus on the Mode Eye like the procedure 6 .

    B. Press the measurement button. If measurement is failed with message like

    RETRY repeattheprocedure A.andpress themeasurement buttonagain.

    C. Check whether diopter value is measured or not. Diopter value is recorded in

    thebottomareaof theModel Eye.

      Automatic measurement

  • 8/20/2019 M.I. URK700.pdf

    11/25

    A. To initialize programpress MENU button.

    B. After user setting is displayed select AUTO START by cursor and

    measurement will start automatically. Then press EXIT button, menu will be

    savedandmeasurementmodeappear.

    C. When the bright dot enters the inner alignment mark and Model Eye is in focus

    properlymeasurementstarts automatically.

    5.Measurement

    5.1Continuous KeratometryandRefractometry K/R Mode

    In theK/R Mode refractometry is performedafterkeratometry.

  • 8/20/2019 M.I. URK700.pdf

    12/25

     1 Enter the K/R Mode

    A. Press MODE button until “R/K” is displayedonthe topof thedisplay.

     2 Adjust height of examinee’s eye

    A. Have the examinee sit and place his/her chin and forehead against the chin

    rest and forehead rest.

    B. Adjust the height of the chin by rotating the height adjustment knob so that

    the eye of the examinee is aligned with the height adjustment mark on the

    headrest.

     3 PerformFocusing

    A. Slide the operation joystick to the left so that the right eye of the examinee is

    displayed on themonitor.

    B. Ask the examinee to look at the red balloon in the center of the scene eye

    fixation target .

    C. Looking at the monitor, check that image of the KERATO RING is not

    obscured by the upper eyelid, if it is covering the ring, instruct the examinee

    to keephis/her eye openedwide until measurement ends.

    D. Looking at the monitor, incline the operation joystick to right or left, and turn

    the operation joystick so the pupil is centeredwith the inner alignment mark

    and focus on the KERATO RING, at the time adjusted the bar would be

    appeared above the inner alignmentmark.

  • 8/20/2019 M.I. URK700.pdf

    13/25

     4 Measurement

    A. Press the Measurement button

    B. The newestmeasuredresultwill be displayedonthe monitor.

    C. According to the 6.1 SETUP MODE, you can select and change VD, CYL

    points.

     5 Repeated measurement

    A. Measurement could be repeatedlyperformed if necessary.

    B. Thenewestresultwill bedisplayedonthemonitorwhenevermeasured.

    C. CAUTION: before starting the new measurement you have to press CLEAR

    button to remove previous data.

     6 Measure theother eye.

    A. Slide the stage to the right side andmeasure the left eye.

    B. After both eyes are measured, pupil distance PD will be displayed on the

    monitor.

     7 Print

    A. Press the PRINT button.

    B. The selected data inthe 6.1SETUP Mode will be printed.

    5.2 Refractometry REF Mode

    Only refractometrycan beperformed in theREF mode.

  • 8/20/2019 M.I. URK700.pdf

    14/25

     1 Enter themode.

    A. Press the MODE buttonuntil

    “REF

    ”appearsonthe topofdisplay

  • 8/20/2019 M.I. URK700.pdf

    15/25

     2 Adjust height of examinee’s eye

    A. Have the examinee sit and place his/her chin and forehead against the chin

    rest and forehead rest.

    B. Adjust the height of the chin by rotating the height adjustment knob so that

    the eye of the examinee is aligned with the height adjustment mark on the

    face rest.

     3 PerformFocusing

    A. Slide the operation joystick to the left so that the right eye of the examinee is

    displayed on themonitor.

    B. Ask the examinee to look at the red balloon in the center of the scene eye

    fixation target .

    C. Looking at the monitor, check that image of the KERATO RING is not

    obscured by the upper eyelid, if it is covering the ring, instruct the examinee

    to keephis/her eye openedwidely untilmeasurement ends.

    D. Looking at the monitor, incline the operation joystick to right or left, and turn

    the operation joystick so the pupil is centeredwith the inner alignment mark

    and focus on the KERATO RING, at the time adjusted the bar would be

    appeared above the inner alignmentmark.

     4 Measurement

    A. Press the Measurement button

    B. The newestmeasuredresultwill be displayedonthe monitor.

    C. According to the 6.1 SETUP MODE, you can select and change VD, CYL

    points.

     5 Repeated measurement

    A. Measurement could be repeatedlyperformed if necessary.

    B. Thenewestresultwill bedisplayedonthemonitorwhenevermeasured.

    C. CAUTION: before starting the new measurement you have to press CLEAR

    button to remove previous data.

     6 Measure theother eye.

    A. Slide the stage to the right side andmeasure the left eye.

    B. After both eyes are measured, pupil distance PD will be displayed on the

    monitor.

     7 Print

    A. Press the PRINT button.

    B. The selected data inthe 6.1SETUP Mode will be printed.

  • 8/20/2019 M.I. URK700.pdf

    16/25

    5.3Keratometry KER Mode

    Onlythe radius ofcurvature of the corneacan be measured inKER Mode.

  • 8/20/2019 M.I. URK700.pdf

    17/25

     1 Enter the KER Mode.

    A. Press the MODE button until “KER” is displayed inthe topof the display

     2 Adjust height of examinee’s eye

    A. Have the examinee sit and place his/her chin and forehead against the chin rest

    andforehead rest.

    B. Adjust the height of the chinby rotating the height adjustmentknob so thatthe eye

    of theexaminee is aligned with theheightadjustmentmark onthe face rest.

     3 PerformFocusing

    A. Slide the operation joystick to the left so that the right eye of the examinee is

    displayed on themonitor.

    B. Ask the examinee to look at the redballoon in the center of the scene eye fixation

    target .

    C. Looking at the monitor, check that image of the KERATO RING is not obscured by

    theupper eyelid, if it is covering the ring, instruct theexaminee to keep his/hereye

    opened widelyuntil measurementends.

    D. Looking at the monitor, incline the operation joystick to right or left, and turn the

    operation joystick so the pupil is concentric with the inner alignment mark and

    focus on the KERATO RING, at the time adjusted the bar would be appeared

    above the inner alignmentmark.

     4 Measurement

  • 8/20/2019 M.I. URK700.pdf

    18/25

    A. Press the Measurement button

    B. The newestmeasuredresultwill be displayedonthe monitor.

    C. According to the 6.1SETUP Mode you can select and change VD, CYL, KERATO

    points.

     5 Repeated measurement

    A. Measurement could be repeatedlyperformed if necessary.

    B. The newestresultwill bedisplayedonthemonitorwhenevermeasured.

    C. CAUTION: before starting the newmeasurement youhave to press CLEAR button

    to remove previous data.

     6 Measure theother eye.

    A. Slide the stage to the right side andmeasure the left eye.

    B. Afterboth eyes are measured, pupil distance PD will be displayedonthe monitor.

     7 Print

    A. Press the PRINT button.

    The selecteddata in the 6.1SETUP Mode will be printed.

    5.4Contact Lens BaseCurve Measurement CLBC mode

    Base curve ofhard contact lens canbe measured inthe CLBC mode.

  • 8/20/2019 M.I. URK700.pdf

    19/25

     1 Enter theMode.

    A. Press the MODE button until “CLBC MODE” is displayed inthe topof the display

     2 AttachContactLens.

    A. Put water in the concave section of contact lens holder at backside of the TEST

    Model Eye and place the contact lens on the holder with the concave surface

    facingupward. Take care that thecontact lens notto drop.

     3 Attachthe testModel Eye.

    A. Remove the chin rest paper. Fix the Model Eye attached with contact lens using

    fixationpin.

     4 PerformFocusing

    A. Looking at the monitor, incline the operation joystick to right or left, and turn the

    operation joystick so thepupil is centeredwith the inner alignment red mark.

    B. Moving operation joystick straight and back fix up KERATO RING to be clear at

    the time adjusted thebar wouldbeappearedabove the inner alignmentmark.

     5 Measurement

    A. Press the Measurement button

  • 8/20/2019 M.I. URK700.pdf

    20/25

    B. The newestmeasuredresultwill be displayedonthe monitor.

    C. According to the 6.1SETUP Mode you can select and change VD, CYL, KERATO

    points.

     6 Repeated measurement

    A. Measurement could be repeatedlyperformed if necessary.

    B. Thenewestresultwill bedisplayedonthemonitorwhenevermeasured.

    C. CAUTION: before starting the newmeasurement youhave to press CLEAR button

    to remove previous data.

     7 Print

    A. Press the PRINT button.

    The selecteddata in the 6.1SETUP Mode will be printed.

    5.5 Corneal diameterMeasurement SIZE Mode

    Diameterofcornea can bemeasured intheSIZE mode.

     1 Enter the Mode

    A. Press MODE button until “SIZE MODE” is displayed inthe topof the display.

     2 Adjust height of examinee’s eye

    A. Have the examinee sit and place his/her chin and forehead against the chin

    rest and forehead rest.

    B. Adjust the height of the chin by rotating the height adjustment knob so that

    the eye of the examinee is aligned with the height adjustment mark on the

    headrest.

     3 PerformFocusing

    A. Slide the operation joystick to the left so that the right eye of the examinee is

    displayed on themonitor.

    B. Ask the examinee to look at the red balloon in the center of the scene eye

    fixation target .

    C. Looking at the monitor, check that image of the KERATO RING is not

    obscured by the upper eyelid, if it is covering the ring, instruct the examinee

    to keephis/her eye openedwidely untilmeasurement ends.

    D. Looking at the monitor, incline the operation joystick to right or left, and turn

    the operation joystick so the pupil is centeredwith the inner alignment mark

    and focus on the KERATO RING.

     4 Measurement

    A. Press the Measurement button

  • 8/20/2019 M.I. URK700.pdf

    21/25

    B. To move the left/right bar bypressing the

    /

     

    button.

    C. The newestmeasuredresultwill be displayedonthe monitor.

    D. According to the 6.1 SETUP Mode you can select and change VD, CYL,

    KERATO points.

     5 Repeated measurement

    A. Measurement could be repeatedlyperformed if necessary.

    B. Thenewestresultwill be displayedonthemonitorwhenevermeasured.

    C. CAUTION: before starting the new measurement you have to press CLEAR

    button to remove previous data.

     6 Measure theother eye.

    A. Slide the stage to the right side andmeasure the left eye.

     7 Print

    A. Press the PRINT button.

    B. The selected data inthe 6.1SETUP Mode will be printed.

    6. OtherMeasurementMode

    6.1SETUP MODE

  • 8/20/2019 M.I. URK700.pdf

    22/25

    Change all the settings aboutmeasurements, printouts.

  • 8/20/2019 M.I. URK700.pdf

    23/25

     1 STEP

    To change measurementdata units.

    Select0.25/0.12 pressingbutton

    /

     

    2 VD

    To selectVertexDistance VD

    Select0/10/12/13.5/15 pressingbutton→/ 

    3 CYL

    To selectCylinder valueand form

    Select -/+/+- pressingbutton

    /

     

    4 START

    To selectManual orAuto measurement

    SelectManual/Auto pressingbutton

    /

     

    .

     5 REPEAT

    To select repeatedmeasurements.

    Select1/3/5/7 pressingbutton →/ .

     6 TARGET

    Target LED brightness control “1” is the brightest

    Select1/2/3/4/5/6/7 pressingbutton

    /

     

    .

     7 MONITOR

    Monitorbrightness control “1” is thebrightest

    Select1/2/3/4 pressingbutton

    /

     

  • 8/20/2019 M.I. URK700.pdf

    24/25

  • 8/20/2019 M.I. URK700.pdf

    25/25

     17 CODE

    To initialize programontheoccasion ofoccurring errors among useddata.

    After pressing EDIT “5” Reload the factory data? will appear. If you answer “yes”

    programcanbe initialized.