OPERATION MANUAL
AURO REF / KERATOMETER URK-700
-
8/20/2019 M.I. URK700.pdf
2/25
UNICOS Co., Ltd. www.e-unicos.com
URK OPERATION MANUAL
Contents
1. Features
2. Notes forUsing the Instrument
3. Description
3.1Main Unit
3.2OperationButtons
4. Practicing throughModel Eye
5.Measurement
5.1Continuous KeratometryandRefractometry
5.2Refractometry
5.3Keratometry
5.4ContactLens BaseCurve Measurement
5.5Corneal diametermeasurement
-
8/20/2019 M.I. URK700.pdf
3/25
6. OtherMeasurementModes
6.1SETUP Mode
1. Features
1. Various Measurements Supported.
Not only the usual refractometry and keratometry, but also corneal
diameter and base curve of contact lens can be measured with this
one instrument. Thus, measurements of eye and prescriptions for
glasses and contact lenses can be made more efficiently.
2. Wide Dioptric Measurement Range
Because theURK-700 covers a wide measurement range, from
-25D to +22D, evenanexaminee with strongmyopia canbemeasured.
3.More accurate Measurement.
-
8/20/2019 M.I. URK700.pdf
4/25
The foggy method of the eye fixation target makes examinee
’s eye
comfortable and enables to getmore accurate measurementdata.
4. EasyConnection with otherEquipments
This instrument is designed to connect other equipments such as VCR
orTV monitor.
2. Notes forUsing the Instrument
1. Do not hit or drop the instrument. The instrument may be damaged
when it receives a strong impact. The impact may cause damage the
function of this instrument. Pleasedohandlewith care.
2. An exposure to the direct sunlight or too bright indoor lights may
-
8/20/2019 M.I. URK700.pdf
5/25
influence on the resultof accurate measurement.
3. To connect this with other relevant equipments, consultwith the dealer
about the wayto make it.
4. A sudden heating of the room in cold areas will cause condensation on
the protective glass in the monitor screenand onoptical parts inside the
instrument. In this case, just wait until condensation disappears before
performing measurement.
5. Keep the objective glass of the examinee side clean. If itwas stained, it
maycause onERROR or inaccurate measurements.
6. In case there is smoke, strange odor ornoise on working disconnect the
power supply and consult thedealer.
7. Don
’t use organic solution such as alcohol, thinner, benzene, etc. to
clean the surface of this instrument. Itmaydamage the instrument.
-
8/20/2019 M.I. URK700.pdf
6/25
8. When moving this URK-700, fix the stage by using clamping bolt and
stage holding knob, always keep power off, and then lift the bottom of
theunitwith both hands.
9. If you leave URK-700 without using for certain period, disconnect the
power supply and protect theunitwith dust cover.
3.Description.
3.1MainUnit.
-
8/20/2019 M.I. URK700.pdf
7/25
LCD Monitor
Functional Key
Measurement
button
Stage Clamping
Bolt
Printer
-
8/20/2019 M.I. URK700.pdf
8/25
Headrest
Measurement
Chinrest
EXT Video RS-232Port Power IN LET Fuse PowerSwitch
StageHolding
-
8/20/2019 M.I. URK700.pdf
9/25
3.2 Operation Panel
Title Technical skill
MODE To change measurementmode
CLEAR Toerase measured data on the screen
MENU To change andset up the function
PRINT To print the dataof measurement
IOL Tomeasure the patient’s eye with cataract or IOL intra-ocular lens
-
8/20/2019 M.I. URK700.pdf
10/25
4. Practicing byModel Eye
Practice measurement by using the TestModel Eye before doing the measurement of
patient’s eye.
1 Turn ON thepower
Turn ON the powerswitchonthe lower left side.
2 Setthemodel eye on
Remove the chin rest paper and align the holes on the base of the model eye with the
holes onthe chinrest. Andinsert pins.
3 Release stage lock
Fully turn the bolt in the underneath cover in counter clockwise direction. And turn the
stage holding bolt incounter clockwise directionto release thestage lock.
4 AdjustheightofTestModel Eye.
Adjust the height of the chin rest by rotating Height Adjustment knob so that The Test
Mode Eye is aligned with theHeightAdjustmentmark on theheadrest.
5 Enter K/R or REF Mode
If neither K/R nor REF is displayed on the monitor press MODE button until either one
is displayed.
6 Adjust theposition and focus ontheModel Eye.
Looking at themonitor incline the operation joystick toward the Model Eye until a bright
dot appears near the inner alignment mark. Place the bright dot in the center of inner
alignmentmark. Focus on theModel Eye by inclining theoperation joystick forward and
backward so thatKERATO RING image is displayed clearly on the monitor, at the time
adjusted thebar wouldbeappearedabove the inner alignmentmark.
7 Measurement
Manual measurement
A. Adjust the positionand focus on the Mode Eye like the procedure 6 .
B. Press the measurement button. If measurement is failed with message like
RETRY repeattheprocedure A.andpress themeasurement buttonagain.
C. Check whether diopter value is measured or not. Diopter value is recorded in
thebottomareaof theModel Eye.
Automatic measurement
-
8/20/2019 M.I. URK700.pdf
11/25
A. To initialize programpress MENU button.
B. After user setting is displayed select AUTO START by cursor and
measurement will start automatically. Then press EXIT button, menu will be
savedandmeasurementmodeappear.
C. When the bright dot enters the inner alignment mark and Model Eye is in focus
properlymeasurementstarts automatically.
5.Measurement
5.1Continuous KeratometryandRefractometry K/R Mode
In theK/R Mode refractometry is performedafterkeratometry.
-
8/20/2019 M.I. URK700.pdf
12/25
1 Enter the K/R Mode
A. Press MODE button until “R/K” is displayedonthe topof thedisplay.
2 Adjust height of examinee’s eye
A. Have the examinee sit and place his/her chin and forehead against the chin
rest and forehead rest.
B. Adjust the height of the chin by rotating the height adjustment knob so that
the eye of the examinee is aligned with the height adjustment mark on the
headrest.
3 PerformFocusing
A. Slide the operation joystick to the left so that the right eye of the examinee is
displayed on themonitor.
B. Ask the examinee to look at the red balloon in the center of the scene eye
fixation target .
C. Looking at the monitor, check that image of the KERATO RING is not
obscured by the upper eyelid, if it is covering the ring, instruct the examinee
to keephis/her eye openedwide until measurement ends.
D. Looking at the monitor, incline the operation joystick to right or left, and turn
the operation joystick so the pupil is centeredwith the inner alignment mark
and focus on the KERATO RING, at the time adjusted the bar would be
appeared above the inner alignmentmark.
-
8/20/2019 M.I. URK700.pdf
13/25
4 Measurement
A. Press the Measurement button
B. The newestmeasuredresultwill be displayedonthe monitor.
C. According to the 6.1 SETUP MODE, you can select and change VD, CYL
points.
5 Repeated measurement
A. Measurement could be repeatedlyperformed if necessary.
B. Thenewestresultwill bedisplayedonthemonitorwhenevermeasured.
C. CAUTION: before starting the new measurement you have to press CLEAR
button to remove previous data.
6 Measure theother eye.
A. Slide the stage to the right side andmeasure the left eye.
B. After both eyes are measured, pupil distance PD will be displayed on the
monitor.
7 Print
A. Press the PRINT button.
B. The selected data inthe 6.1SETUP Mode will be printed.
5.2 Refractometry REF Mode
Only refractometrycan beperformed in theREF mode.
-
8/20/2019 M.I. URK700.pdf
14/25
1 Enter themode.
A. Press the MODE buttonuntil
“REF
”appearsonthe topofdisplay
-
8/20/2019 M.I. URK700.pdf
15/25
2 Adjust height of examinee’s eye
A. Have the examinee sit and place his/her chin and forehead against the chin
rest and forehead rest.
B. Adjust the height of the chin by rotating the height adjustment knob so that
the eye of the examinee is aligned with the height adjustment mark on the
face rest.
3 PerformFocusing
A. Slide the operation joystick to the left so that the right eye of the examinee is
displayed on themonitor.
B. Ask the examinee to look at the red balloon in the center of the scene eye
fixation target .
C. Looking at the monitor, check that image of the KERATO RING is not
obscured by the upper eyelid, if it is covering the ring, instruct the examinee
to keephis/her eye openedwidely untilmeasurement ends.
D. Looking at the monitor, incline the operation joystick to right or left, and turn
the operation joystick so the pupil is centeredwith the inner alignment mark
and focus on the KERATO RING, at the time adjusted the bar would be
appeared above the inner alignmentmark.
4 Measurement
A. Press the Measurement button
B. The newestmeasuredresultwill be displayedonthe monitor.
C. According to the 6.1 SETUP MODE, you can select and change VD, CYL
points.
5 Repeated measurement
A. Measurement could be repeatedlyperformed if necessary.
B. Thenewestresultwill bedisplayedonthemonitorwhenevermeasured.
C. CAUTION: before starting the new measurement you have to press CLEAR
button to remove previous data.
6 Measure theother eye.
A. Slide the stage to the right side andmeasure the left eye.
B. After both eyes are measured, pupil distance PD will be displayed on the
monitor.
7 Print
A. Press the PRINT button.
B. The selected data inthe 6.1SETUP Mode will be printed.
-
8/20/2019 M.I. URK700.pdf
16/25
5.3Keratometry KER Mode
Onlythe radius ofcurvature of the corneacan be measured inKER Mode.
-
8/20/2019 M.I. URK700.pdf
17/25
1 Enter the KER Mode.
A. Press the MODE button until “KER” is displayed inthe topof the display
2 Adjust height of examinee’s eye
A. Have the examinee sit and place his/her chin and forehead against the chin rest
andforehead rest.
B. Adjust the height of the chinby rotating the height adjustmentknob so thatthe eye
of theexaminee is aligned with theheightadjustmentmark onthe face rest.
3 PerformFocusing
A. Slide the operation joystick to the left so that the right eye of the examinee is
displayed on themonitor.
B. Ask the examinee to look at the redballoon in the center of the scene eye fixation
target .
C. Looking at the monitor, check that image of the KERATO RING is not obscured by
theupper eyelid, if it is covering the ring, instruct theexaminee to keep his/hereye
opened widelyuntil measurementends.
D. Looking at the monitor, incline the operation joystick to right or left, and turn the
operation joystick so the pupil is concentric with the inner alignment mark and
focus on the KERATO RING, at the time adjusted the bar would be appeared
above the inner alignmentmark.
4 Measurement
-
8/20/2019 M.I. URK700.pdf
18/25
A. Press the Measurement button
B. The newestmeasuredresultwill be displayedonthe monitor.
C. According to the 6.1SETUP Mode you can select and change VD, CYL, KERATO
points.
5 Repeated measurement
A. Measurement could be repeatedlyperformed if necessary.
B. The newestresultwill bedisplayedonthemonitorwhenevermeasured.
C. CAUTION: before starting the newmeasurement youhave to press CLEAR button
to remove previous data.
6 Measure theother eye.
A. Slide the stage to the right side andmeasure the left eye.
B. Afterboth eyes are measured, pupil distance PD will be displayedonthe monitor.
7 Print
A. Press the PRINT button.
The selecteddata in the 6.1SETUP Mode will be printed.
5.4Contact Lens BaseCurve Measurement CLBC mode
Base curve ofhard contact lens canbe measured inthe CLBC mode.
-
8/20/2019 M.I. URK700.pdf
19/25
1 Enter theMode.
A. Press the MODE button until “CLBC MODE” is displayed inthe topof the display
2 AttachContactLens.
A. Put water in the concave section of contact lens holder at backside of the TEST
Model Eye and place the contact lens on the holder with the concave surface
facingupward. Take care that thecontact lens notto drop.
3 Attachthe testModel Eye.
A. Remove the chin rest paper. Fix the Model Eye attached with contact lens using
fixationpin.
4 PerformFocusing
A. Looking at the monitor, incline the operation joystick to right or left, and turn the
operation joystick so thepupil is centeredwith the inner alignment red mark.
B. Moving operation joystick straight and back fix up KERATO RING to be clear at
the time adjusted thebar wouldbeappearedabove the inner alignmentmark.
5 Measurement
A. Press the Measurement button
-
8/20/2019 M.I. URK700.pdf
20/25
B. The newestmeasuredresultwill be displayedonthe monitor.
C. According to the 6.1SETUP Mode you can select and change VD, CYL, KERATO
points.
6 Repeated measurement
A. Measurement could be repeatedlyperformed if necessary.
B. Thenewestresultwill bedisplayedonthemonitorwhenevermeasured.
C. CAUTION: before starting the newmeasurement youhave to press CLEAR button
to remove previous data.
7 Print
A. Press the PRINT button.
The selecteddata in the 6.1SETUP Mode will be printed.
5.5 Corneal diameterMeasurement SIZE Mode
Diameterofcornea can bemeasured intheSIZE mode.
1 Enter the Mode
A. Press MODE button until “SIZE MODE” is displayed inthe topof the display.
2 Adjust height of examinee’s eye
A. Have the examinee sit and place his/her chin and forehead against the chin
rest and forehead rest.
B. Adjust the height of the chin by rotating the height adjustment knob so that
the eye of the examinee is aligned with the height adjustment mark on the
headrest.
3 PerformFocusing
A. Slide the operation joystick to the left so that the right eye of the examinee is
displayed on themonitor.
B. Ask the examinee to look at the red balloon in the center of the scene eye
fixation target .
C. Looking at the monitor, check that image of the KERATO RING is not
obscured by the upper eyelid, if it is covering the ring, instruct the examinee
to keephis/her eye openedwidely untilmeasurement ends.
D. Looking at the monitor, incline the operation joystick to right or left, and turn
the operation joystick so the pupil is centeredwith the inner alignment mark
and focus on the KERATO RING.
4 Measurement
A. Press the Measurement button
-
8/20/2019 M.I. URK700.pdf
21/25
B. To move the left/right bar bypressing the
→
/
button.
C. The newestmeasuredresultwill be displayedonthe monitor.
D. According to the 6.1 SETUP Mode you can select and change VD, CYL,
KERATO points.
5 Repeated measurement
A. Measurement could be repeatedlyperformed if necessary.
B. Thenewestresultwill be displayedonthemonitorwhenevermeasured.
C. CAUTION: before starting the new measurement you have to press CLEAR
button to remove previous data.
6 Measure theother eye.
A. Slide the stage to the right side andmeasure the left eye.
7 Print
A. Press the PRINT button.
B. The selected data inthe 6.1SETUP Mode will be printed.
6. OtherMeasurementMode
6.1SETUP MODE
-
8/20/2019 M.I. URK700.pdf
22/25
Change all the settings aboutmeasurements, printouts.
-
8/20/2019 M.I. URK700.pdf
23/25
1 STEP
To change measurementdata units.
Select0.25/0.12 pressingbutton
→
/
2 VD
To selectVertexDistance VD
Select0/10/12/13.5/15 pressingbutton→/
3 CYL
To selectCylinder valueand form
Select -/+/+- pressingbutton
→
/
4 START
To selectManual orAuto measurement
SelectManual/Auto pressingbutton
→
/
.
5 REPEAT
To select repeatedmeasurements.
Select1/3/5/7 pressingbutton →/ .
6 TARGET
Target LED brightness control “1” is the brightest
Select1/2/3/4/5/6/7 pressingbutton
→
/
.
7 MONITOR
Monitorbrightness control “1” is thebrightest
Select1/2/3/4 pressingbutton
→
/
-
8/20/2019 M.I. URK700.pdf
24/25
-
8/20/2019 M.I. URK700.pdf
25/25
17 CODE
To initialize programontheoccasion ofoccurring errors among useddata.
After pressing EDIT “5” Reload the factory data? will appear. If you answer “yes”
programcanbe initialized.