Methods for Microbial Testing of Endoscopes
Transcript of Methods for Microbial Testing of Endoscopes
Methods for Microbial Testing of Endoscopes
• Mark Hichens Cbiol, MRSB, MSc
Determination of Total Viable Count in final rinse waters by The Method of Membrane Filtration
Applicable standards
• Washer-disinfectors Part 2: Requirements and tests for washer-disinfectors employing thermal disinfection for surgical instruments, anaesthetic equipment, bowls, dishes, receivers, utensils, glassware, etc. (ISO 15883-2:2006)
• Washer-disinfectors - Part 4: Requirements and tests for washer-disinfectors employing chemical disinfection for thermolabile endoscopes (ISO 15883-4:2008)
• Health Technical Memorandum 2030 Part 3 Washer-disinfectors Validation and verification
• HSE Standards and Recommended Practices for Endoscope Reprocessing Units. QPSD-D-005-2. V 2.0
• HSE Standards and Recommended Practices for Central Decontamination Units, Version 2.0 – 2011
Validation and process control
• Critical limits
• Reproducibility
• Repeatability
• Linearity
• Reliability
• Accuracy and sensitivity
• Sample bottles
• Media
• Filters
• Filtration apparatus
• Incubators
• Scientists
Final test Quality Control
• Internal quality control
• External quality control
Sample numberSample matrix Date Count A Count B log10A Log10B A-B A+B RSD^2 RSDR
24179 27.09.16 6 4 2 0.77815125 0.602059991 0.176091 1.3802112 0.03255476 0.18043
24220 25.05.17 4 4 2 0.602059991 0.602059991 0 1.20412 0 0.00000
24295 14.11.17 2 7 2 0.301029996 0.84509804 -0.544068 1.146128 0.45068195 0.67133
24319 30.09.16 4 4 2 0.602059991 0.602059991 0 1.20412 0 0.00000
24409 26.09.16 7 10 2 0.84509804 1 -0.154902 1.845098 0.01409629 0.11873
24498 14.08.17 2 3 2 0.301029996 0.477121255 -0.176091 0.7781513 0.1024183 0.32003
24511 20.12.16 4 6 2 0.602059991 0.77815125 -0.176091 1.3802112 0.03255476 0.18043
24608 25.11.16 5 4 2 0.698970004 0.602059991 0.09691 1.30103 0.01109667 0.10534
24622 23.05.17 6 6 2 0.77815125 0.77815125 0 1.5563025 0 0.00000
24660 03.11.17 7 5 2 0.84509804 0.698970004 0.146128 1.544068 0.01791283 0.13384
24670 23.11.17 7 5 2 0.84509804 0.698970004 0.146128 1.544068 0.01791283 0.13384
24682 01.09.17 9 4 2 0.954242509 0.602059991 0.352183 1.5563025 0.1024183 0.32003
24692 03.08.17 3 10 2 0.477121255 1 -0.522879 1.4771213 0.25061074 0.50061
24738 15.08.17 3 5 2 0.477121255 0.698970004 -0.221849 1.1760913 0.07116426 0.26677
24756 04.08.17 2 4 2 0.301029996 0.602059991 -0.30103 0.90309 0.22222222 0.47140
Uncertainty of Measurement (RSD RC)
This sheet assists in providing an estimate of the uncertainty of measurement obtained from the paired counts.
The relative s tandard deviation of reproducibi l i ty (RSD R) for each pair i s ca lculated after a common logari thms (log 10) transformation of the
count. These are combined to obtain the "average" RSD RC expressed as a log10 va lue. Log transformation provides the nummeric RSDrc
Test portions with less than two results or for which one or more of the counts are zero are excluded.
Quality Control
Process Control
Ensure the component parts of the process are working effectively.
Ensure that the component parts are working consistently. Evaluate the component parts to ensure working consistently and improve the quality of the end product.
Product control
Ensure the product meets the demands of its intended purpose
FDA Cracks Down on Duodenoscope Manufacturers - Again
• FDA in 2015 ordered manufacturers to conduct a post-market surveillance study to determine whether healthcare facilities were able to properly clean and disinfect the device –Quality Control !
• Three companies were required to conduct a study to sample and culture reprocessed duodenoscopes
• Olympus failed to commence data collection, and Pentax and Fujifilm have failed to provide sufficient data.
• FDA threatened seizure, injunction, and civil money penalties.
Health Service Executive Standards and Recommended Practices for Commissioning, Validation and Testing in Endoscope Decontamination Units
The guidance is close to publication
It is comprehensive
Introduces new testing procedures, acceptable parameters
and recommends frequency of testing
Endoscope cleaning and disinfection process (Major components).
Cleaning Efficacy by residual soil detection Weekly Testing
Contamination of endoscopes will be proteinaceous in nature.
Cleaning efficacy tests undertaken using an artificial test soil and surrogate device accordance with EN ISO 15883 Part 5.
Qualification:- The chamber walls and load carrier should be free of residual soil
Process Challenge Device (PCD) tests (Hexalumen, Tosi, Flexicheck)
Qualification:- Process challenge devices test should be in accordance with the limits described by the manufacturer.
Frequency - weekly.
Residual Protein Test
• Determine the effectiveness of the cleaning process by determining residual protein
• Residual contamination can be made clearly visibly by swabbing the surfaces or lumens of the endoscope using residual protein tests capable of detecting >1µg of protein.
• Swabbing methods based on Bradford assay or Coomassie blue reagents are widely available
• If appropriately validated/quality controlled.
Residual Protein -frequency of testing
Ensuring testing of every product family in every chamber every quarter and each scope annually
DESCRIPTION
A
B
C
D
E
F
G
Endoscopes and
Product Families
Processed
Total Number of
Product
Families
Processed
Basins
(positions)
Required
Number of
Test for Machine Performance
per quarter
(A x B)
Required
Number of
Test for Machine Performance
per year
(C X 4)
Total number of
scopes
available to
the user
Required
number of
additional
tests to
establish
scope integrity.
E-D when E
is more than D
Total number of
tests per
year
D+F
Bronchoscope and
Cystoscopes, both
belonging to product family number 3
1 2 2 tests 8 5 0 8
Gastroscope product
family number 1
ERCP product family
number 2
Cystoscope product family number 3
3 8 24 tests 96 120 24 120
Disinfection efficacy
• Annual EWD Disinfection Efficacy testing
• Disinfection efficacy should be verified using the test method described in EN15883 (2009) part 4 section 4.4.2.4.
• Endoscope Lumen Sampling
Endoscope Lumen Sampling
• The Endoscope is brushed through with a sterile disposable long-stemmed cleaning brush after processing
• The brush end is aseptically transferred to 10 mL of sterile peptone water containing a neutraliser
• The eluate is filtered through a 47 mm diameter 0.45 m filter and then the filter placed onto a R2A agar plate and incubated at 30±2°C for up to five days
• The number of colony forming units (cfu) are counted
Guidance Levels
A result of less than 10 cfu per brush sample
should be obtained if the Endoscope
has been satisfactorily disinfected
Alternative technique - Lumen flushing
• Passing sterile water/neutralising buffer down the Endoscope lumens- suction, air/water and biopsy lumen
• Flush ~10 mL of water down the lumen and collect into a sterile sample bottle at the distal end.
• Laboratory analysis, examine for total viable count on R2A TSA or YEA media with incubation at 30±2°C for up to five days.
• The results should be below 10 cfu/100 mL.
• Good aseptic technique are essential to obtain good results
Endoscopes and Product Families Processed
Total Number of Product Families Processed
Scopes available to the user in each product family
Sampling Strategy
Cystoscopes, belonging to product family number 3
1 1 Cystoscope The single endoscope will be sampled every
quarter.
Bronchoscopes and Cystoscopes, both belonging to product family number 3
1 1 Bronchoscopes
1 Cystoscope
The two endoscopes will be sampled
alternatively every quarter as they are part of
the same product family. Therefore, each scope
will be sampled twice in the year.
ERCPs product family number 2 Gastroscope / Colonoscopes both belonging to product family number 1
2 1 ERCP endoscope
1 Gastroscope
1 Colonoscope
The ERCP endoscope will be sampled every
quarter.
The gastroscope and colonoscope will be
sampled alternatively every quarter as they are
part of the same product family.
ERCPs and EUS both belonging to product family number 2 Gastroscope product family number 1
2 1 ERCP endoscope
3 EUS endoscopes
1 Gastroscope
The ERCP and EUS endoscopes will be sampled
once each over the 12 month period as they are
part of the same product family and testing
should be rotated amongst them each quarter.
The gastroscope will be sampled every quarter
as it is the only member of that product family.
Quality of water
• Critical to the outcome of the decontamination process.
• Microbiologically
• Efficacy of cleaning agents and disinfectants
Supplied water quality Determinant and Unit Maximum Permitted Values
Final Rinse Other Stages
Appearance Clear, colourless
Degree of acidity (pH) 5.5 to 8.0
Conductivity at 25º (uS/cm) 30
Total dissolved solids (mg/100mL) 4
Total hardness, CaCO3 (mg/L) 50 200
Chloride, CI (mg/L) 10 120 10 120
Heavy metals, determined as Lead, Pb (mg/L)
10
Iron, Fe (mg/L) 2
Phosphate, P2O5 (mg/L) 0.2
Silicate, SiO2 (mg/L) 0.2
Total viable count (TVC) at 30ºC <10
Bacterial endotoxins (EU/mL) 0.25
Frequency – annually.
Final Rinse Water
• Microbiological quality of the water needs to be assessed regularly in accordance with EN ISO 15883 Part 1:2014 Annex D
• Total viable counts
• pH
• Conductivity
• Hardness
• Weekly
Aerobic colony count
(cfu) in 100mL
Interpretation Action and advice Colour grade
<1 cfu Satisfactory Use as normal Green
1 – 9 cfu on a regular
basis
Acceptable –indicates
that bacterial
numbers are under a
reasonable level of
control
Use as normal Green
10 – 100cfu Risk Assess
Request identification
of predominant
organisms and
resample
Carry out formal risk
assessment, based
upon clinical needs in
consultation with the
consultant
Microbiologist the IPC
team and the AE(D)
Orange
>100cfu Risk Assess
As above. Request
identification of
predominant
organisms and
resample
EWD should be taken
out of service until
water quality has
improved / issue
resolved.
RED
Pseudomonas aeruginosa
• Gram negative opportunistic pathogen. Ubiquitous in the environment
• Able to form biofilms and exhibit resistance to disinfectants
• In endoscopy is the most commonly reported micro-organism responsible for transmission of infection
Environmental mycobacteria (rapid growing)
• Opportunistic pathogens found widely in the environment. Commonly found in water distribution systems
• Exhibit resistance to high level disinfectants due to presence of hydrophobic lipid rich cell wall
• Infection of the patient may lead to miss-diagnosis of tuberculosis and implementation of un-necessary antibiotic therapy
Storage Cabinets for Heat Labile Endoscopes (CESC’s)
ISO EN 16442 Controlled environment storage cabinet for disinfected thermolabile endoscopes
The controlled environment provided by any CESC should ensure that there is no deterioration of the microbiological quality of the Endoscope
Routine Testing of CESC’s
Surface contamination
Indication of overall cleanliness
Indication of cleaning efficacy
Aero-contamination
Indication of the circulating air, and air that may be entering the cabinet
Annual re-validation
Placing of a sterile surrogate device in the cabinet for the validated time period confirmed at initial validation
Description of Test
Frequency
Result Required
Comments
Determine the contamination levels on the inside surfaces of the cabinet
Quarterly
The contamination levels identified shall be less than 25 cfu/25 cm² EN 16442 – Section 6.5
Reference should be ne made to EN 16442 – section 6.5
NB: Pathogenic organisms to be identified.
Check that the cabinet is capable of maintaining the quality of the Endoscopes
Annual The acceptable result for the test is
<10 cfu/lumen. Sterile surrogate device to be fitted to the Endoscope drying cabinet and left for a period of no less than what was established during commissioning.
Evaluate airborne microbial contamination.
Annual Reference should be made to EN
16442 Annex C
Environmental monitoring
• Alert level – CFU Levels that when exceeded, signal a possible deviation from that normally seen.
• Action Level – CFU levels that significantly exceed those normally seen.
Vacuum storage systems
• Validation of storage time
• Periodic validation of such systems is required to verify performance
• As specified by the manufacturer and/or by risk assessment
• Endoscope Lumen Sampling
• At the end of the defined storage period.
• There should be no degradation of the microbiological quality.
• Less than 10 cfu per sample
Final release testing
• Residual protein tests on reprocessed endoscopes as a measure of cleaning efficacy
• Endoscope Lumen Sampling as a measure of disinfection and storage efficacy
There is no such thing as chance or accident; the words merely signify our ignorance of some real and immediate cause. (Adam Clarke 1760-1832)