MEIRI 2655C Version 1.8
USER MANUAL
http://www.celians-shop.com – Email : [email protected] Headquarters: Spherel Systèmes - 10bis route d’Ax – 31120 PORTET SUR GARONNE
Tel : 05 34 51 53 18 [email protected] SIRET : 791 075 401 00012
2
TABLEOFCONTENTS
1SAFETYINSTRUTIONS ...................................................................................................................................... 3
2INTRODUCTION ............................................................................................................................................... 4
3PRESENTATION ................................................................................................................................................ 4
3.1Frontpanel ............................................................................................................................................................... 4
3.2Rearpanel ................................................................................................................................................................ 5
3.3Panelinstallation ...................................................................................................................................................... 5
3.7Overview .................................................................................................................................................................. 6
4OPERATINGPRINCIPLE .................................................................................................................................... 6
5FEATURES ........................................................................................................................................................ 7
6TECHNICALFEATURES ..................................................................................................................................... 9
7CONNECTION ................................................................................................................................................. 12
7.1CAUTION ................................................................................................................................................................ 12
7.2REARCONNECTION ................................................................................................................................................ 12
7.2Internalcardlayout ................................................................................................................................................ 20
9SETUPMENU ................................................................................................................................................ 26
10EXTERNALFEATURES ................................................................................................................................... 31
10.1ANALOGICOUTPUT .................................................................................................................................. 31
10.2EXTERNALCONTROLINPUT ...................................................................................................................... 31
10.3RELAYOUTPUT ......................................................................................................................................... 31
10.4OUTPUTRS232 ......................................................................................................................................... 33
10.5EXTERNAL+5VDC .................................................................................................................................... 33
11MAINTENANCE ............................................................................................................................................ 34
12ASSISTANCE ................................................................................................................................................. 34
13CONTROLREFERENCE ................................................................................................................................. 35
http://www.celians-shop.com – Email : [email protected] Headquarters: Spherel Systèmes - 10bis route d’Ax – 31120 PORTET SUR GARONNE
Tel : 05 34 51 53 18 [email protected] SIRET : 791 075 401 00012
3
1SAFETYINSTRUTIONSPLEASEREADTHISMANUALBEFOREPROCEEDINGANYHANDLING
PRELIMINARYPRECAUTIONS
Theelementaryprecautions,safetyrulesandinstructionssetoutinthepresentdocumentmustbeobservedatalltimes inordertoensurethesafetyofpersonsandtopreventanydamagetothedeviceorto instrumentswhichareconnected to the latter. It is also essential that statutory requirements and safety requirements for the applicationconcernedshouldbeobservedduringuse.
Forreasonsofsafety,thedevicemustonlybeusedbyaqualifiedperson.Thedevicemustonlybeusedforthepurposesforwhich ithasbeendesigned,andwithin the limitsofapplication specified.Thedevicemustnotbeopened,other than in thecontext of the operations listed in the present document.No attemptmust bemade to remove internal components, or tomodifytheminanyway.Thedevice constitutesone link inan instrumentation channel. Equipment installers andoperatorsmustundertake theplanned consideration, deployment and response to safety requirements associatedwiththedevice,withtheinstrumentsconnectedtheretoandwithinstrumentationtechnology.INSTALLATIONPRECAUTIONS
Onlythevoltagerequiredbythemeasuringdeviceshouldbeused.Itmustbeconfirmedthattheinstrumentsorthepowergriddeliveringvoltagetothedeviceareconsistentwiththeratingindicatedonthelatter.
The electrical safety of this device can only be guaranteed if the latter is correctly connected to a grounding installation, inaccordance with electrical safety standards. It is essential to confirm that this fundamental safety requirement is properlyfulfilled.Forinstallationpurposes,atwo-poleswitchmustbeprovided,withaminimum3mmcontactbreakinggap.The device must be protected against any direct contact with water, and the maximum permissibleambienttemperaturesmustbeobserved.Theuseofthedevice indirectsunlight,or inveryhot,humid,dirtyordustylocations,locations.whicharesubjecttostrongvibrationsorinproximitytomagneticfieldsmayresultinincorrectoperation.Noobjectsmustbeplacedinfrontoftheventilationoutletsofthemeasuringdevice,orthecorrectventilationoftheinternalcomponentswillotherwisebeimpaired,resultinginoverheating.Ifthemeasuringdeviceshowssignsofincorrectoperation,orifanyunusualodourorsmokeisdetected,itshouldbedisconnectedfromsupplyimmediatelyandservicedbyaqualifiedMEIRIengineer.GUARANTEEMEIRI measuring devices are covered by a 1-year factory return guarantee for parts and labour, excepting anyspecificprovisions.Theguaranteewillnotbeenforceableinthefollowingspecificcases:If thedevicehasbeenoperatedatavoltageotherthanthat indicatedonthedataplateof thedevice. If theuseremploys thedevice supplied inanabnormalor impropermanner,orundertakesanymodifications thereto. If theusercausesdamagebynegligence,inadequateservicing,inexperienceortheuseofharmfulproducts.Replacementsorrepairsundertakenunderthetermsoftheguaranteeandresultingintheunavailabilityofthedeviceforaperiodoflessthansevendaysshallnotextendthetermoftheguarantee.Fortheenforcementoftheguarantee,theusermustcontacttheMEIRIdistributorwhosoldthedeviceconcerned.Repairsunderguaranteewillbeundertakeninourlaboratories,andthedevicemustbereturnedinpackagingwhichwillensurethesafetythereofduringtransport.Theuserwillassumepackagingandcarriagecostsforthereturnofthedevicetothefactory.MEIRIoritsdistributorwillassumepackagingandcarriagecostsforthereturnofthedevicetotheuserafterrepairinmainlandFranceonly.Nocompensationshallbepayableinrespectoftheunavailabilityofthedeviceassociatedwiththeconductofrepairsunderthetermsoftheguarantee.
http://www.celians-shop.com – Email : [email protected] Headquarters: Spherel Systèmes - 10bis route d’Ax – 31120 PORTET SUR GARONNE
Tel : 05 34 51 53 18 [email protected] SIRET : 791 075 401 00012
4
2INTRODUCTIONThank you for choosing our conditionning electronics from the MEIRI range.Read this document carefully and keep this manual in a safe place for futurereference. Our warranty will only apply if the products are installed and usedaccordingtotheinstructionsgiven.
3PRESENTATION
TheME2655Cisadigitalpaneldisplayorpanelmount.It ismainlyintendedforthemeasurement of signals and the conditioning of sensors in the context ofprocess controls. It can provide an analog output that can be used by anacquisitionsystemoraPLC,transmitinRS232andhasdigitalinputsandoutputs(TOR).TheME2655C isavailable inmultiplerangesadaptedtothedifferentelectricaland physical quantities (force, torque, weighing, pressure, displacement,inclination ...). The sensor's sensitivity and measuring range information aresufficientforquickcommissioning. 3.1Frontpanel
1.2linesof16charactersdisplay2. Voyantsd’étatdesseuils«Mini/OK/Maxi»3.«ResetandEscape»Key4. «Menu»Key5. PrintKeyandUPScroll6. ZeroKey/UPScroll
1
2 3
4
5
6
http://www.celians-shop.com – Email : [email protected] Headquarters: Spherel Systèmes - 10bis route d’Ax – 31120 PORTET SUR GARONNE
Tel : 05 34 51 53 18 [email protected] SIRET : 791 075 401 00012
5
3.2Rearpanel
3.3Panelinstallation
1. RS232Connection2. Straingaugeinput(A)3. Fuseholder4. Mainsupply85à264Vac–50/60Hz5. Outputconnection(contacts,print,tareandzero)6. Electronictorquesensorandquadratureencoderinput7. Mainson/offswitch
6
1 2 3 4
7 5
http://www.celians-shop.com – Email : [email protected] Headquarters: Spherel Systèmes - 10bis route d’Ax – 31120 PORTET SUR GARONNE
Tel : 05 34 51 53 18 [email protected] SIRET : 791 075 401 00012
6
3.7Overview
4OPERATINGPRINCIPLETheME2655Cconditionsthesensorsbyprovidingahighstabilitypowersupply.Then, itamplifiesthesignalscomingfromtheseandconvertsthemintodigital.The internalprocessor calculatesand scales in real time theexact valueof themeasuredphysicalparameteranddisplaysitonthedisplay.Dependingonthemeasurementmodeselected,thereal-timevalueorthepeakvalueobtained. In thismeasurementmode, theelectronicsacquire thedataathigh speed and realize a sliding average allowing to "capture" a fugitive signalwithgreatprecision.Thedevicetransmits itsdisplayviaRS232link inrealtimeoroperatoraction. Italsocontrolstworelayswhencrossingthresholdsconfiguredbytheoperator.AnanalogdigitalconverterallowstheME2655Ctoprovidetwoanalogoutputs±10Vand4-20mA,availableontherearpanel.Theentireconfigurationiscarriedoutinthedisplaybyscrollingthemenusoftheprogram.
SECTEUR
OUTPUTS:-RS232-Contactssec-ANALOG
CAPTEUR
A M PL I
EXCITATION
SUPPLY
PROCESSORADC
DISPLAY
DAC
RELAISEXCITATION
RS232
EXTERNALCONTROL
http://www.celians-shop.com – Email : [email protected] Headquarters: Spherel Systèmes - 10bis route d’Ax – 31120 PORTET SUR GARONNE
Tel : 05 34 51 53 18 [email protected] SIRET : 791 075 401 00012
7
5FEATURESGENERALFEATURESME2655C
§ Displayofthemeasurementinrealtime§ Storingthemaximumorminimumvalue(fastpeakmeasurement)§ Automaticormanualreset§ Visuallevelindicator(bargraph)§ Invertingthepolarityoftheinputsignal§ Voltageoutput±10V§ Powerourput4–20mA§ Output2relaycontactsdry§ RS232output§ ExternalresettablecommandRESET,PRINT,TARE§ Scrolling Menu configuration en 4 languages (English, German, French,
Italian)FEATURESME2655CAJ
§ Conditionoramplifierstraingaugesensor§ 4-wiressensor
§ Bridgepowersupply10V § Adjustablesensitivityfrom0.313mV/Vto9.999mV/V
FEATURESME2655CLVDT698
§ ConditionoramplifierforLVDTSensor§ A/Bdemodulation
§ 3-wireshalf-bridgecompatible§ Adjustablesensitivityfrom31.3mV/Vto999.9mV/V
§ FrequencyandexcitationvoltageondemandCTIONSME520LVD
http://www.celians-shop.com – Email : [email protected] Headquarters: Spherel Systèmes - 10bis route d’Ax – 31120 PORTET SUR GARONNE
Tel : 05 34 51 53 18 [email protected] SIRET : 791 075 401 00012
8
T698FEATURESME2655CLVDT598
§ ConditionoramplifierforLVDT5or6-wiressensors§ (A-B)/(A+B)Demodulations
§ Adjustablesensitivityfrom31.3mV/Vto999.9mV/V § Frequencyandexcitationvoltageondemand
E52T698FEATURESME520U
§ Voltageamplifier§ Compatiblesensorelectroniintegrated§ Sensorpowersupply+15Vdc1W§ Adjustablesensitivityfrom0.313to9.999V
FFEATURESME520DPONCTIONSME520FU-DSR
§ Conditionoramplifierpotentiometricsensor§ Sensorpowersupply4.1V§ Adjustablesensitivityfrom31.3mV/Vto999.9mV/V
http://www.celians-shop.com – Email : [email protected] Headquarters: Spherel Systèmes - 10bis route d’Ax – 31120 PORTET SUR GARONNE
Tel : 05 34 51 53 18 [email protected] SIRET : 791 075 401 00012
9
6TECHNICALFEATURES
GENERALFEATURESME2655CPOWERSUPPLYPowersupplyvoltage 95-264VAC/50-60HzPower Typical 15WSupplufilter Schaffner Fuse 5*20mm;temporisé 1ADIMENSIONSize H*W*D(socketsandbases
included86*144*157mm
Weight 1.2kgSetting Tableorpanel CONDITIONSOFUSETemperaturecontrol 0à+50°CStoragetemperature -10à+70°CProtectionrating Ontable
Assembledonpanelwithwaterproofingjoint(notincluded)
IP50IP55
DIGITALPARTMicrocontroller RISCClock 20MHzAnalogconverter/digital 16BitsSamplingfrequency maximum 100kHzDigitalconverter/analog 12BitsResolutionfro±E.M. ±32768Display 2lines16characters LCDOUTPUTRS232Direction Transmission Speed Adjustable(réglageBauds) 1200,2400,9600,19200Bitsnumber Setting 7,8(1bitdeSTOP)Parity Variable PairouimpairFormat +___1.0000__mm___> Responsetime Betweenthecontroland1stfront 50msmaxTansmissionlength Accordingconfiguredspeed 170,80,20,11msMaximalvelocityoftheexternalcontrol
Accordingconfiguredspeed 5,9,20,26Hz
Continousexternal Externcontrolmaintained 20Hz
http://www.celians-shop.com – Email : [email protected] Headquarters: Spherel Systèmes - 10bis route d’Ax – 31120 PORTET SUR GARONNE
Tel : 05 34 51 53 18 [email protected] SIRET : 791 075 401 00012
10
conrolspeed ANALOGOUTPUTSVoltageoutput Configurable(For±PE) ±5Vou10VPoweroutput Configurable(For±PE) 8-16mAou4-20mABandwidth at-3db 500HzRisetime 2msLinearity Typical 0.05%RELAYOUTPUTSContact Break/Working 0.5A/220VacmaxiHysteresis Withmaximumnumberofdecimal
Withotherdisplaymodes±10points±1point
EXTERNALINPUTCONTROLRAZ Bypotential-freecontactorTTL Assetclosedorat0Print Bypotential-freecontactorTTL Assetclosedorat0Tara Bypotential-freecontactorTTL Assetclosedorat0
FEATURESME2655CAJSensorinput Completebridgeofconventionalor
semiconductorgauges4Wires
Inputimpedance Typical 15MΩBridgeimpedance 350Ωà10kΩBridgeexcitation Voltagemaxi30mA 10VDCTRMC Typical 110dBSensitivityinput Adjustable 0.313à9.999mV/VNosensibilitysettings 0.001mV/VLinearconditioner Typical 0.01%
FEATURESME520LVDT698Inputsensor LVDTCompletbridgeor
semibridge3ou4Wires
Demodulation A/BExcitationfrecuencysensor Standard(othersondemand) 5kHz±10%Excitationvoltagesensor Standard(othersondemand) 3VRMS±10%Inputimpedance Amplifierinput 200kΩSensitivityinput Adjustable 31.3à999.9mV/VNosensibilitysettings 0.001mV/VLinearconditioner Typical 0.05%
http://www.celians-shop.com – Email : [email protected] Headquarters: Spherel Systèmes - 10bis route d’Ax – 31120 PORTET SUR GARONNE
Tel : 05 34 51 53 18 [email protected] SIRET : 791 075 401 00012
11
FEATURESME520LVDT598Inputsensor LVDT 5ou6filsDemodulation (A-B)/(A+B)Excitationfrecuencysensor Standard(othersondemand) 5kHz±10%Excitationvoltagesensor Standard(othersondemand) 3VRMS±10%Inputimpedance Amplifierinput 200kΩSensitivityinput Adjustable 31.3à999.9mV/VNosensibilitysettings 0.001mV/VLinearconditioner Typical 0.05%
FEATURESME520UInput Voltagereferencedtoground Inputimpendance 17kΩExcitationvoltagesensor 15VDCSensitivityinput Adjustable 0.313à9.999VLinearconditioner Typical 0.01%
FEATURESME520DPInput Voltagereferencedtoground Inputimpedance >200MΩImpedanceSensor 500Ωà10kΩExcitationvoltagesensor 4.1VDCSensitivityinput Adjustable 31.3à999.9mV/VLinearconfitioner Typical 0.01%
http://www.celians-shop.com – Email : [email protected] Headquarters: Spherel Systèmes - 10bis route d’Ax – 31120 PORTET SUR GARONNE
Tel : 05 34 51 53 18 [email protected] SIRET : 791 075 401 00012
12
7CONNECTION
7.1CAUTION Before connecting or separating the measuring device to other electroniccomponents, turn off the measuring device. Disconnect all cables beforedismantling or moving the unit. Do not handle the adjustment devices toosuddenly.7.2REARCONNECTIONNOTE:Thecableshieldmustbeconnectedtotheconnectorhousings
RS232CONNECTOR
PlugSub-Dfemale9contactsPIN2 TXPIN3 RXPIN5 GNDI/OPLUGPlugSub-Dmale15contactsPIN1 Lowthreshold-ContactN/CPIN2 Lowthreshold-ContactN/OPIN3 Highthreshold-Contact
communPIN4 GNDPIN5 OutputtensionPIN6 GNDPIN7 PRINTPIN8 NCPIN9 Lawthreshold-Contact
communPIN10 Highthreshold-ContactN/OPIN11 Highthreshold-ContactN/CPIN12 +5Vdc10mAmaxiPIN13 Voltageoutput
http://www.celians-shop.com – Email : [email protected] Headquarters: Spherel Systèmes - 10bis route d’Ax – 31120 PORTET SUR GARONNE
Tel : 05 34 51 53 18 [email protected] SIRET : 791 075 401 00012
13
SensorconnectorE520AJM
2655CAJPlugDIN7contactsPIN6 (+)ExcitationPIN1 (-)SignalPIN4 (-)ExcitationPIN2 (+)SignalPIN5 (+)ExcitationPIN3 (-)ExcitationPIN7 GND 4-WIRESSTRAINGAUDESENSORSINPUTSON7-PINCONNECTORM
SENSOR 2655CAJ
(+)Signal PIN2
(+)Excitation PIN6
(-)Signal PIN1
Cableshielding PIN7
(-)Excitation PIN4
E520AJM
PIN14 RESETPIN15 TARE
http://www.celians-shop.com – Email : [email protected] Headquarters: Spherel Systèmes - 10bis route d’Ax – 31120 PORTET SUR GARONNE
Tel : 05 34 51 53 18 [email protected] SIRET : 791 075 401 00012
14
2655CUPlugDIN7contactsPIN6 Excitation+PIN1 NCPIN4 Excitation-(GND)PIN2 Signal+PIN5 NCPIN3 NCPIN7 GND INTEGRATEDELECTRICSENSORSCONNECTIOR
SENSOR 2655CU
(+)Signal PIN2
(+)Excitation PIN6
Cableshielding PIN7
(-)Excitation PIN4
ONCTION
+VCC
GND
OUT
http://www.celians-shop.com – Email : [email protected] Headquarters: Spherel Systèmes - 10bis route d’Ax – 31120 PORTET SUR GARONNE
Tel : 05 34 51 53 18 [email protected] SIRET : 791 075 401 00012
15
2655CDPPlugDIN7contactsPIN6 Excitation+PIN1 NCPIN4 Excitation-(GND)PIN2 Signal+PIN5 NCPIN3 NCPIN7 GND 3-WIRESPOTENTIOMETICCONNECTOR
SENSOR 2655CDP
(+)Signal PIN2
(+)Excitation PIN6
Cableshielding PIN7
(-)Excitation PIN4
http://www.celians-shop.com – Email : [email protected] Headquarters: Spherel Systèmes - 10bis route d’Ax – 31120 PORTET SUR GARONNE
Tel : 05 34 51 53 18 [email protected] SIRET : 791 075 401 00012
16
ONCTION 2655CLVDT598PlugDIN7contactsPIN6 (+)ExcitationPIN1 (-)SignalPIN4 (-)ExcitationPIN2 (+)SignalPIN5 (+)ExcitationPIN3 (-)ExcitationPIN7 GND 5-WIRESINDUCTIVESENSORCONNECTORM
SENSOR 2655CLVDT598
(+)Signal PIN2
(+)Excitation PIN6
(-)Signal PIN1
CableShielding PIN7
(-)Excitation PIN4
GND PIN7
E520AJM
http://www.celians-shop.com – Email : [email protected] Headquarters: Spherel Systèmes - 10bis route d’Ax – 31120 PORTET SUR GARONNE
Tel : 05 34 51 53 18 [email protected] SIRET : 791 075 401 00012
17
http://www.celians-shop.com – Email : [email protected] Headquarters: Spherel Systèmes - 10bis route d’Ax – 31120 PORTET SUR GARONNE
Tel : 05 34 51 53 18 [email protected] SIRET : 791 075 401 00012
18
2655CLVDT698PlugDIN7contactsPIN6 (+)ExcitationPIN1 (-)SignalPIN4 (-)ExcitationPIN2 (+)SignalPIN5 (+)ExcitationPIN3 (-)ExcitationPIN7 GND 4-WIRESCOMPLETBRIDGESENSORCONNECTORM
SENSOR 2655CLVDT698
(+)Signal PIN2
(+)Excitation PIN6
(-)Signal PIN1
Cableshielding PIN7
(-)Excitation PIN4
PIN7
E520AJM
http://www.celians-shop.com – Email : [email protected] Headquarters: Spherel Systèmes - 10bis route d’Ax – 31120 PORTET SUR GARONNE
Tel : 05 34 51 53 18 [email protected] SIRET : 791 075 401 00012
19
3-WIRESHALF-BRIDGEINDUCTIVESENSORCONNECTOR
SENSOR 2655CLVDT698
(+)Signal PIN2
(+)Excitation PIN6
Cableshielding PIN7
(-)Excitation PIN4
E520AJM 3-WIRESINDUCTIVECOMPLETBRIDGESENSORCONNECTORM
SENSOR 2655CLVDT698
(+)Signal PIN2
(+)Excitation PIN6
PIN1
Cableshielding PIN7
(-)Excitation PIN4
E520AJM
http://www.celians-shop.com – Email : [email protected] Headquarters: Spherel Systèmes - 10bis route d’Ax – 31120 PORTET SUR GARONNE
Tel : 05 34 51 53 18 [email protected] SIRET : 791 075 401 00012
20
7.2Internalcardlayout For the2655C LVDT6982Cmodelonly, it ispossible to configure several parameters onthepackagingcard:LayoutLVDTou1/2bridge:VoieA Pontcomplet 1/2pont VoieB Pontcomplet 1/2pontS4àS10 Switchontheleft Switchontheright S14àS20 Switchontheleft
Switchontheright
Configurationdel'impédancedecharge:A 10KΩ 2KΩ 100KΩ B 10KΩ 2KΩ 100KΩ
S3
Jumperontheright
Jumperontheleft
withoutjumper
S13
Jumperontheright
Jumperontheleft
withoutjumper
Excitationfrequencylayout:
Fréquence S1 S2 configuration13kHz
Withoutjumper Ontheright
10kHz
Withoutjumper Ontheleft
5kHz
Jumprontheleft Without
2.5kHz
Jumperontheright Without
S1 S2
S1 S2
S1 S2
S1 S2
http://www.celians-shop.com – Email : [email protected] Headquarters: Spherel Systèmes - 10bis route d’Ax – 31120 PORTET SUR GARONNE
Tel : 05 34 51 53 18 [email protected] SIRET : 791 075 401 00012
21
8QUICKLYSETUP
Settingexampleof2655CAJLocate the measuring range and sensitivity on the sensor certificate or label.Connectittothecabinetandturniton.In our example, a force sensorwith a capacity of ± 50daN and a sensitivity of2.000mV / V is connected to the 2655C. The display is configured to view themeasurementinrealtime.
Action/description DISPLAYToenterthesetupmenus,pressthekey:LeavethefunctiononinputAandmovetothenextmenuwiththekey:
SelectthedaNunitwiththekey:Switchtothenextmenuwiththekey:
Hold thebutton to set themeasuringrangeto50.00daN.Switchtothenextmenuwiththekey
Holdthekeytosetthesensitivityto2.000mV/V.
SENSITIVITY A 2.000 mV/V
FUNCTION A
Zéro
FULL SCLAE ±50.00 daN
Zéro
MEASURING UNIT daN
Zéro
http://www.celians-shop.com – Email : [email protected] Headquarters: Spherel Systèmes - 10bis route d’Ax – 31120 PORTET SUR GARONNE
Tel : 05 34 51 53 18 [email protected] SIRET : 791 075 401 00012
22
Pressthekeyrepeatedlytogotothe"measuringmode"menu.Checkthatitisinfollowermode
PressthekeyrepeatedlytogotothebackupmenuPressthekeythesensorparametersaresaved,thedevicewillrestartautomatically.
Themeasuringmodeisnowdisplayed:Thresholdstatusindicator.Real-timedisplayofmeasuredvalueWithunitmeasurementRecalltheinputdisplayedBargraph,evolveswiththemeasurement.Thekeyallowsyoutoperformarelativezero(tara)atanytime.The key allows the measurementinformation to be sent instantaneously via theRS232output.
BACK UP PRESS ZERO
Zéro
+ 0.138 daN = A ..
Zéro
TYPE OF MEASURE FOLLOWER
http://www.celians-shop.com – Email : [email protected] Headquarters: Spherel Systèmes - 10bis route d’Ax – 31120 PORTET SUR GARONNE
Tel : 05 34 51 53 18 [email protected] SIRET : 791 075 401 00012
23
Settingexample2655CLVDT698Locatethemeasuringrangeandsensitivityonthesensorcertificateorlabel.IfitiswritteninmV/V/mm,convertittomV/Vbymultiplyingitbythefullscaleofthesensor.Connectittothecabinetandturniton.In our example, a displacement sensor with a capacitance of ± 5mm and asensitivityof200.0mV/Visconnectedtothe2655C.Thedisplayisconfiguredtoretain themaximum size of a samplemeasured by the sensor, with amanualresetofthispeakvalue.
Action/description DISPLAYToenterthesetupmenus,pressthekey:LeavethefunctiononinputAandmovetothenextmenuwiththekey:
Selectthemmunitwiththekey:Switchtothenextmenuwiththekey:
Hold thebutton to set themeasuringrangeto5.000mm.Switchtothenextmenuwiththekey
Holdthekeytosetthesensitivityto200.0mV/V.
SENSITIVITY A 200.0 mV/V
FUNCTION A
MEASURINF UNIT mm
Zéro
FULL SCALE ±5.000 mm
Zéro
Zéro
http://www.celians-shop.com – Email : [email protected] Headquarters: Spherel Systèmes - 10bis route d’Ax – 31120 PORTET SUR GARONNE
Tel : 05 34 51 53 18 [email protected] SIRET : 791 075 401 00012
24
Pressthekeyrepaetedlytoreachthemenu«modeofmeasure»Configure thePeak+measurementmodewith thekeySwitchtothenextmenuwiththekey
Check that the RAZ of the measurement isconfiguredtobemanually.Switchtothenextmenuwiththekey
Check that the trip time is not less than 0.003seconds.Switchtothenextmenuwiththekey
Checkthatthestartingthresholdis+0.010mm.Switchtothenextmenuwiththekey
MODE OF MEASURE PEAK +
Zéro
R.A.Z. MEASURE MANUAL
T. TRIGGER 0.003 seconds
TRIGGER TIME +0.010 mm
http://www.celians-shop.com – Email : [email protected] Headquarters: Spherel Systèmes - 10bis route d’Ax – 31120 PORTET SUR GARONNE
Tel : 05 34 51 53 18 [email protected] SIRET : 791 075 401 00012
25
PressthekeyrepeatedlytogotothebackupmenuPressthekeythesensorparametersaresaved,thedevicewillrestartautomatically..
Themeasuringmodeisnowdisplayed:Thresholdstatusindicator.Real-timedisplayofmeasuredvalueWithunitmeasurementRecalltheinputdisplayedBargraph,evolveswiththemeasurement.Thekeyisusedtoresetthepeakvalue.Ittransmits the measurement informationsimultaneouslyviatheRS232output.Thekeyallowsyoutoperformarelativezero(tara)atanytimeThe key allows the measurementinformation to be sent instantaneously via theRS232output.
BACK UP PRESS ZERO
Zéro
+ 4.138 mm = A ..
Zéro
RST
http://www.celians-shop.com – Email : [email protected] Headquarters: Spherel Systèmes - 10bis route d’Ax – 31120 PORTET SUR GARONNE
Tel : 05 34 51 53 18 [email protected] SIRET : 791 075 401 00012
26
9SETUPMENUTheprogramisexecutedbypressingthe"Menu"keys:Theparameterscanbechangedbypressingthe"Param."key:
DISPLAY Fonction/Description MeansthattheboxusesinputAforthesensor.If
inputBisavailable(optional),theusercanchoosethis input alone or in mathematical combinationwithinputA.Selection:A,B,A+B,A-B,(A+B)/2,(A-B)/2
Allowstheusertoselecttheunitofmeasurementtobedisplayed.Note: no conversion operation is applied to themeasurement.Unitsavailable:Nm,mKg,mdaN,lbf.in,ft.lbf,N,daN,KN,gramme,Kgf,tonne,pound,mm,cm,m,inch,feet,bar,psi,Pa,mmHg,mmH2O,g,mm/s,m/s,C,F,V,mV,A,mA,aucuneunité
Adjusts the measuring range of the sensor (seesensorordatasheet).Adjustablefrom0.1000to999.9Note:ThisparameteriscommontoinputsAandB.
FUNCTION A
MEASURE UNIT daN
FULL SCALE ±5.000 daN
Zéro Print
http://www.celians-shop.com – Email : [email protected] Headquarters: Spherel Systèmes - 10bis route d’Ax – 31120 PORTET SUR GARONNE
Tel : 05 34 51 53 18 [email protected] SIRET : 791 075 401 00012
27
Thisalertmessageappearsifthemeasuringrangehas changed, especially if the decimal point hasmoved. In this case, the following menu goesdirectlytothe"HighthresholdandLowthreshold"to check the value of the threshold setpoints, asthese have also been modified. Also check thethresholdvalueifthePEAKmodeisactive.
Sensor sensitivity for input A, mV / V or Vdepending on the function of the enclosure (AJ,LVDT, U ...). Values can range from 0.314 to9.999mV/VorVorfrom31.3to999.9mV/VfortheLVDT.
This menu allows to modify the sign of themeasurement according to the direction ofvariationofthesensorA.Twopossibleselections:Normalorinverse
REAL:DisplaystherealvalueofthesensorATARA:displaystherelatuvevalueofthesensorA
Thetareissettozerobypressingthekey
Accesstithesensorcalibrationmenubypressingthekey:The calibrationmenu recalls themeasured range(E :)and thesensitivity (S :) configuredpreviouslyand displays the measurement in real time. Thesettingkeysareusedtochangethesensitivity.Thisfunction is used when calibrating a cabinet andsensorassembly.
!! WARNING!! CHANGED THRESHOLDS !
SENSITIVITY A 2.000 mV/V
SENSOR DIRECTION INORMAL
REAL A: - 0.736 TARA A: + 0.001
CALIBRATION A PRESS ZERO
E : 5.000 S : 2.000 - 0.001 daN
Zéro
Zéro
http://www.celians-shop.com – Email : [email protected] Headquarters: Spherel Systèmes - 10bis route d’Ax – 31120 PORTET SUR GARONNE
Tel : 05 34 51 53 18 [email protected] SIRET : 791 075 401 00012
28
Allowstoselect themeasurementmodethatwillbeusedtodisplaythesensorvalue.Threepossibleselections:FOLLOWER:DisplaysthevalueinrealtimePEAK+:Displaysthemaximumvalue
PEAK-:Displaysthemaximumvalue
MENUISONLYDISPLAYEDIFTHEMEASURINGPEAKMODEISSETUP:
Resetting the peak value can be configured intwodifferentways:MANUAL:TheresetisdonebypressingtheRSTkey.ThenewPEAKvalueor0.000(ifthestartingthreshold has not been crossed) will bedisplayed.AUTOMATIC: the displayed peak value will beupdatedwhenthestartingthresholdiscrossed.
Allows you to configure the minimum timebeforethepeakvalueistakenintoaccount.Thistiming starts as soon as the starting thresholdhasbeencrossed.Configurablefrom0.001to1.000seconds.An acceptable setting is approximately 0.007seconds.
Setsthetriggerthresholdforthemeasurement.Storing thepeak valuewill only startwhen thislevelisexceededandafterthetriptime.An acceptable setting is 20% of themeasurement.
MODE OF MEASURE FOLLOWER
R.A.Z. MEASURE MANUAL
T. TRIGGER TIME 0.010 seconds
START MEASURE +00.20 daN
http://www.celians-shop.com – Email : [email protected] Headquarters: Spherel Systèmes - 10bis route d’Ax – 31120 PORTET SUR GARONNE
Tel : 05 34 51 53 18 [email protected] SIRET : 791 075 401 00012
29
The2655Chastwoalarmsthatcontrolthe"<","="and">"LEDsonthefrontpanel,aswellastworelays(seesection9.3).These menus allow you to set the high and lowsetpointsofthealarms.Thesevaluescanbeupto±2timesthemeasuringrange.The thresholds can be positive or negative. If apositive threshold and a negative threshold arerequired,thehighthresholdmustbeconfiguredinpositiveandthethresholdlowinnegative.Ifthetwothresholdsareofthesamepolarity,thehighthresholdmustbesethigherinabsolutevaluethanthelowthreshold.
Inmeasurementmode the user can apply a zerorelativetothedisplayviathekeyThismenuallowsyoutoactivateorblocktheuseofthisbutton.
Increases or decreases the number of digits afterthedecimalpoint.Thissettingdependsontheconfiguredmeasuringrange,withinthelimitoffivesignificantdigits.
Choiceofdisplaylanguage.Possible selections: German, English, French,Italian
MEASURE HIGH +5.010 daN
MEASURE LOW -5.010 daN
PRESS ZERO APPROVE
DECIMAL NUMBERS - 0.001 daN
LANGUAGE ENGLISH
Zéro
http://www.celians-shop.com – Email : [email protected] Headquarters: Spherel Systèmes - 10bis route d’Ax – 31120 PORTET SUR GARONNE
Tel : 05 34 51 53 18 [email protected] SIRET : 791 075 401 00012
30
AllowsadjustmentoftheparametersoftheRS232link.Innumberofdatabits:7or8WithorWithoutparitySpeed:from1200to19200
Themeasuringrangesoftheanalogoutputvoltageandcurrentcanbeassignedacoefficientx1:±5V&8-16mAx2:±10V&4-20mA
AdjuststhesensitivityofthebargraphdeviationConfigurablefrom01to10.
BackupoftheentireconfigurationAfterpressingthekeytheparametersaresavedandthedeviceisrestartedautomatically.
Allows you to restore the default settings in allmenus of the device. Caution, all settingspreviously saved by the userwill be permanentlylost
CALABRITION RS232 9600, SANS, 8
OUTPUT U et I x 1
ZOOM BARGRAPH 02
BACK UP PRESS ZERO
FACTORY SETTINGS PRESS ZERO
Zéro
http://www.celians-shop.com – Email : [email protected] Headquarters: Spherel Systèmes - 10bis route d’Ax – 31120 PORTET SUR GARONNE
Tel : 05 34 51 53 18 [email protected] SIRET : 791 075 401 00012
31
10EXTERNALFEATURES 10.1ANALOGICOUTPUT Areconversionofthedisplayisavailableviatheanalogoutputs.The voltageoutput and the currentoutput are available simultaneously. Theseanalog outputs are active inmeasuringmode. They are automatically disabledwhenaccessingtheprogrammingmenus.Depending on the setting (see chapter of the configuration menus), differentoutputsareavailable:±10V,±5Vand4-20mAfortheconfiguredsensorrange.Theoutputsof the twoanalogoutputsareconnected to theGND,seechapter7.2forconnectionsandchapter6forconnections. 10.2EXTERNALCONTROLINPUT Threeinputsareavailabletotheusertocontrolthefollowingfunctions:RESET,PRINTandTARE.Theseinputsperformthesamefunctionaspressingthebuttonsonthefrontpanel.INPUTMIRRORSCHEME INPUTSCHARACTERITICS
- Activelow- TTLcompatible/relaycontact- Maximuminputcurrent4mA
Only one entry can be active at a time. If a series of functions are beingprocessed,theresponsetimeof30msmaybeexceeded.IfthePRINTcommandisheld, thecommand is repeatedautomatically.Thecontrolsare in relation totheGND,seechapter7.2forconnection.
http://www.celians-shop.com – Email : [email protected] Headquarters: Spherel Systèmes - 10bis route d’Ax – 31120 PORTET SUR GARONNE
Tel : 05 34 51 53 18 [email protected] SIRET : 791 075 401 00012
32
10.3RELAYOUTPUT The 2655C incorporates two configurable alarms: high threshold and lowthreshold. In addition to the visual indicators, alarms are available on the rearpanelviatworelays(seechapter7.2forconnection).The relays are in positive safety mode, ie the contacts are in alarm if theenclosure is switched off. The relays are also in alarm when accessing theconfigurationmenus.Innormaloperation,aslongasthedisplayisbetweentheconfiguredthresholdvalues,therelaycontactsare"normallyopen"(N/O).If the display exceeds the low threshold or high threshold value, thecorrespondingrelayisinalarmanditscontactsare"normallyclosed"(N/C).
LOWTHRESHOLDRELAYMIRRORSCHEME
OK ALARM
HIGHTHRESHOLDRELAYMIRRORSCHEME
OK ALARM
http://www.celians-shop.com – Email : [email protected] Headquarters: Spherel Systèmes - 10bis route d’Ax – 31120 PORTET SUR GARONNE
Tel : 05 34 51 53 18 [email protected] SIRET : 791 075 401 00012
33
10.4OUTPUTRS232 The 2655C can transmit themeasurement display via its RS232 port. The portconfigurationmustmatch the device towhich it is connected. For connection,seechapter7.2andchapter6forspecifications.RS232transmissionisviathePRINTandRESETfunctions.OutputFormat(20ASCIIcharacters0-19):0 1à10 11 12à
1617 18 19
Sign Measure Space Unit Threshold Endofline NewlineLegendSign +or-Measure XXXX.XXXXXX Thenon-significantzerosarereplacedbyspaces.
Unit Measureunit:mm,daN,mil,etc(charactersnote16=space)Threshold Alarmstatusrepresentedbythesymbols"<","=",or">".Thesymbol
"!"Meansafalsemeasurementorasaturatedsystem.
Examples:-0.2556mm<
+21.596Nm= +915.96Nm>10.5EXTERNAL+5VDCIn pin 12 of the I / O connector (see chapter 7.2), a voltage of 5 volts DC,maximum10mA,isavailabletotheuserforaninterfacecircuit,forexample.
http://www.celians-shop.com – Email : [email protected] Headquarters: Spherel Systèmes - 10bis route d’Ax – 31120 PORTET SUR GARONNE
Tel : 05 34 51 53 18 [email protected] SIRET : 791 075 401 00012
34
11MAINTENANCE
Beforecleaningorservicing,disconnecttheappliancebydisconnectingtheplugorturningoffthepowerswitch.Whenthemeterbecomesdirty,cleanitwithaclean,drycloth.Donotuseliquidcleaningagentssuchasbenzineorthinner,orevenflammableproducts.Donotusepaintthinners,solvents,cleaningagentsorchemical-impregnated cleaning pads.MEIRI is not liable for damage caused byimproperuseofthemeasuringdeviceorbymodificationsmadebytheuser,norcanhecoverthelostordestroyeddata.
12ASSISTANCE Before returning your ME520 to the after-sales service, make sure that theME520iscorrectlywiredbothtothesensorandtotheinstallationtowhichitisconnected.Also,makesurethattheapplianceisingoodworkingorder.Thetablebelowprovidessometroubleshootinghelp:
WARNINGSIGN POSSIBLESOLUTIONDisplayflashes -Thesensor input is"intheair".Checkthecorrect
connectionofthesensorandthecorrectwiringonthe2655Cconnector.- The sensor input is full. Check that the sensorsensitivity is correct. Also make sure that it isfunctioningproperly.
Themeasureseemstosaturateorisfalse.
- Theremay be a gain error in your configuration,check the correct configuration of the measuringrangeandsensitivityofthesensor.
The measurement fluctuates or isnotstable.
- Use shielded cables to connect the sensors,connecttheshieldtopin7oftheconnectoroftheconnectoroftheenclosure.)- Beware of earth loops. Connect it to themeasuring ground at only one point of yourmeasuringchain.
http://www.celians-shop.com – Email : [email protected] Headquarters: Spherel Systèmes - 10bis route d’Ax – 31120 PORTET SUR GARONNE
Tel : 05 34 51 53 18 [email protected] SIRET : 791 075 401 00012
35
13CONTROLREFERENCEMCREATYOUROWNREFERENCEModel TypeAJ,U,DP,LVDT598ouLVDT698
ME2655C -
http://www.celians-shop.com – Email : [email protected] Headquarters: Spherel Systèmes - 10bis route d’Ax – 31120 PORTET SUR GARONNE
Tel : 05 34 51 53 18 [email protected] SIRET : 791 075 401 00012
36
ENDOFDOCUMENT