Medical Product Development cycle

17
Medical Electronics ELEC5650M The Product Development Cycle in the Medical Device Industry Overview Context in which medical electronics products are used Implications for product development The Medical Devices Directive and other regulatory issues Practicalities and the product development cycle Practical examples A few comments on wider (business/marketing) issues

description

Regulation on CE Mark, ISO 13485 and other red tape stuff

Transcript of Medical Product Development cycle

Page 1: Medical Product Development cycle

Medical Electronics ELEC5650M

The Product Development Cycle in the Medical Device Industry

Overview

•  Context in which medical electronics products are used

•  Implications for product development

•  The Medical Devices Directive and other regulatory issues

•  Practicalities and the product development cycle

•  Practical examples

•  A few comments on wider (business/marketing) issues

Page 2: Medical Product Development cycle

What is a medical device?

Any instrument, apparatus, appliance, software, material or other article, whether used alone or in combination, including the software intended by its manufacturer to be used specifically for diagnostic and/or therapeutic purposes and necessary for its proper application, intended by the manufacturer to be used for human beings for the purpose of: •  Diagnosis, prevention, monitoring, treatment or alleviation of disease •  Diagnosis, monitoring, treatment, alleviation of or compensation for an injury or

handicap •  Investigation, replacement or modification of the anatomy or of a physiological

process •  Control of conception and which does not achieve its principal intended action in or on the human body by

pharmacological, immunological or metabolic means, but which may be assisted in its function by such means.

EU Medical Devices Directive (93/42/EEC)

Simplified definition

Any product used for diagnosis, treatment or prevention of an illness

But not drug-based…at least not solely •  Drug-device combinations exist and will increase in prevalence.

Includes software and software/hardware combinations.

Page 3: Medical Product Development cycle

Scope and examples

Devices used in the home •  Plasters, thermometers etc.

Devices used in hospitals and GP surgeries •  Diagnostic: ECG monitors, CT or MRI scanners,

stethoscopes, X-ray equipment etc. •  Therapeutic: surgical instruments, implants,

defibrillators etc. •  Hoists, beds, assistive aids etc.

Context is important – intended purpose.

A regulated market

A medical device cannot be ‘put on the market’ or ‘put into service’ unless it has been checked and certified that it meets the regulations for the country it is to be used in.

This applies whether the device is to be sold or used free of charge. •  Within the EU, medical devices are regulated by the Medical Devices

Directive (CE mark) •  In USA, they are regulated by FDA (Food and Drug Administration)

•  Other countries have their own systems

Page 4: Medical Product Development cycle

Principles of regulatory system

Risk management •  It must not cause any unnecessary risk to the patient, users, or anyone

else (eg. general public)

Verification •  It should ‘do what it says on the tin’ •  In other words it should perform as claimed by the manufacturer

Validation •  It must be independently evaluated by a person with appropriate clinical

qualifications

•  If necessary, clinical testing or trials should be done.

Product development

What does this mean for product development of medical devices?

Page 5: Medical Product Development cycle

The context

Who uses it?

Under what conditions and circumstances?

What could happen if something goes wrong?

Could it give the wrong result? How? What would happen?

How can it be designed, manufactured, labelled etc. to address all these points?

If something goes wrong…

Traceability is important.

If there is a problem, need to know •  Why?

•  What caused it (design error, faulty components, manufacturing error, not used correctly)?

•  What can be done to prevent it happening again (This could include withdrawal of the product certification and even prosecution)?

The regulatory bodies (MHRA, FDA etc) have extensive powers of investigation and enforcement.

Page 6: Medical Product Development cycle

Product lifecycle

The whole product lifecycle is regulated by different bodies.

Key to next slide:

MHRA – Medicines and Healthcare products Regulatory Agency (www.mhra.gov.uk) NICE – National Institute for Health and Clinical Excellence (www.nice.org.uk)

WEEE – Waste Electrical and Electronic Equipment Directive

Who regulates what?

Stage of product lifecycle Who regulates

Product design MHRA

Approvals testing, clinical evaluation

MHRA

User protocol development MHRA, NICE, NHS

Manufacturing MHRA

Use phase MHRA, NICE

Adverse incidents MHRA

Disposal – end of life Not currently regulated specifically for medical devices – other regulations may apply, eg. WEEE

Page 7: Medical Product Development cycle

Classification

How stringent is all this accreditation process?

It depends on •  The degree of ‘invasiveness’

•  The level of risk

•  Whether there is a measuring or diagnostic function

•  Active or passive device

Each device is judged against a number of criteria to decide which class it is and the rules are applied accordingly.

Class I Class IIa Class IIb Class III Limited scrutiny, simpler processes, can self certify

High scrutiny, more clinical evaluation, trials, testing, closer monitoring.

What about medical electronic devices?

Likely to be an active device

May include software

May have a measuring or diagnostic function

…so likely to be Class IIa or IIb

Page 8: Medical Product Development cycle

Medical Devices Directive

To be CE marked (under 93/42/EEC) •  The company (or ‘organisation placing the device on the market’) must

be ISO 13485 compliant (or occasionally some other equivalent) •  There should be a ‘technical file’ for each product showing how it

complies with the Medical Device Directive, performance requirements and product standards (ISO etc).

•  Information about the product’s development, acceptance testing and information needed to manufacture, including user instructions and labelling

Essential requirements

The main technical section of the directive •  A generic checklist of potential hazards •  Every device must be address all these points

Page 9: Medical Product Development cycle

Essential requirements - electronics

Some example sections:

‘Devices incorporating electronic programmable systems must be designed to ensure repeatability, reliability and performance of these systems according to the intended use’

‘In the event of a single fault condition…appropriate means should be adopted to eliminate and reduce …risks’

‘…the software must be validated according to the state of the art taking into account the principles of the development lifecycle, risk management, validation and verification’

‘Devices intended to monitor one or more clinical parameters of a patient must be equipped with appropriate alarm systems to alert the user of situations which could lead to death or severe deterioration of the patient’s state of health’

Focus on users and ‘use case’

A medical device is certified (CE marked) for a defined ‘intended purpose’ •  The manufacturer is specifically responsible for it working if used (only)

in this way

A clinical evaluation is required to show that it works the way it is described •  At the very least this must be a ‘critical evaluation of the relevant

scientific literature’

•  Can be on similar products if you can demonstrate ‘equivalence’

Page 10: Medical Product Development cycle

The user in the design process

Iterative model (eg. Greenberg 1996) Requirements

Design

Implementation

Verification

Maintenance User evaluation

ISO 13485

Quality management system for medical device manufacturers/developers

Analogous to ISO 9000

Covers company processes for ensuring good practice

Must be applied to each product development project (also covers purchasing, production, quality assurance, continuous improvement, training etc)

States framework or policy – each company sets up own procedures and documentation structure to comply

Page 11: Medical Product Development cycle

Product specific standards

If there is one (or more than one) for your product you need to comply with it •  You can do a search for standards eg. on BSI Online

Other important standards

BS EN ISO 14971: 2001 Medical devices – application of risk management to medical devices •  Includes flowcharts for risk analysis process, medical-specific hazard checklist

BS EN ISO 60601-1-1:2006 Medical electrical equipment – General requirements for basic safety and essential performance •  More stringent specifications than for consumer products

BS EN ISO 60601-1-6: 2007 Medical electrical equipment – Usability •  Verification and validation requirements for user interfaces

BS EN ISO 60601-1-4: 1997 Medical electrical equipment – Programmable electrical medical systems •  Risk management for software and embedded systems

Page 12: Medical Product Development cycle

Risk management

Must be built into all the company’s processes

Must be specifically addressed for each product •  Risk analysis

•  Mitigation •  Regular reviews and updates of risk analysis

Comprehensive checklists in MDD Essential Requirements and in Medical devices risk management standard 14971.

Patient involvement

If the product is to be used by/on a patient to assess the clinical benefits, this is a clinical trial. It is usually best for device to be CE marked first UNLESS you need to do the trials to validate it (in which case permission from MHRA for a ‘clinical investigation’ is required) Trials require ethical approvals http://www.nres.npsa.nhs.uk/ Alternatively, patients may be asked to comment on a design in a focus group, or through a questionnaire etc. Nurses often have significant insights into patient behaviour. In the case of certain patient groups (eg. children, patients with dementia) this may be more reliable. It is usually possible to observe clinics to get information about patient behaviour.

Page 13: Medical Product Development cycle

Manufacturing stage

Compliance requirements continue •  Quality assurance – ensure products conform with own specifications

•  Assessment of supplied components

•  Test/inspection of own products

•  Monitoring of customer feedback (post market surveillance) – root cause analysis, corrective action, preventative action

•  Periodic design review, updating

•  Respond to external changes, eg. obsolete components, legislative changes

•  Repeat risk analysis

•  Decide action, implement, monitor

•  Document everything

Practicalities

Manufacturers are audited and monitored by ‘notified bodies’ who are regulated by MHRA

The notified bodies are private sector bodies who charge a fee to the manufacturer •  Initial accreditation to ISO 13485 •  Annual audit to maintain certification

•  Additional charges to CE mark each product

Page 14: Medical Product Development cycle

Exercise

Consider a medical electronic device •  What is its intended purpose? By what kind of user and in what context? •  What are the possible hazards in using it?

•  Include general hazards and risks of it giving ‘wrong’ results

•  How do these affect the product design? •  What are the design issues?

•  How can some of the risks be managed? •  How can it be made easier to use?

Some more general issues

Product design is not all about regulations, even for medical devices…

Page 15: Medical Product Development cycle

Product development in context

To develop a good product, it is necessary to understand the use model

This is particularly true in medical device development •  Even more complicated for future e-health medical devices

And…we have not yet talked about marketing, business, purchasing, financial issues.

A major target for the NHS is ‘efficiency’ – which tends to mean ‘cost reduction’

Multiple stakeholders

Depending on the product, a number of people may be involved: •  Consultant clinician •  Senior nurse •  Allied health professionals – physiotherapists, occupational therapists etc. •  Patient (consumer/customer) •  Patient’s family/carers •  NHS procurement staff •  NHS managers and accountants •  Other service providers (public or private sector)

•  Call centre •  Personal trainers/gyms •  Social services

Page 16: Medical Product Development cycle

Influencing opinion leaders

Risk averse market

But opinion leaders easy to identify •  Professional structures (BMA, RCN, other professional institutions)

•  Conferences and journals

Clinical evaluation should be done by someone with good reputation

Work with NHS procurement agency PASA http://www.pasa.nhs.uk/PASAweb

NHS Technology Adoption Centre

The NHS Technology Adoption Centre, based in Manchester, was launched in September 2007 with the following key aims: •  To increase the uptake of new technology in all areas of the NHS. •  To work with partners to identify excellent technologies which will

improve healthcare in the NHS.

•  To promote greater cooperation between all organisations involved in the development and use of healthcare technologies in the NHS.

http://www.technologyadoptioncentre.nhs.uk/home.html

Page 17: Medical Product Development cycle

Criteria for adoption

A set of criteria have been established by the Technology Adoption Centre, against which all the technologies are evaluated. •  Has to involve step change innovation.

•  Has to have an adoption issue. The technology must have strong independently reviewed evidence (possibly NICE, HTA, CEP) of improved patient outcomes and / or systems efficiencies but still minimal uptake in the NHS.

•  Supports NHS national policy priorities. •  Should be in a clinical area of major focus for the NHS.

The private sector in UK

Conventional private healthcare sector (BUPA, Nuffield) offer same services as NHS •  Major innovation will not be adopted •  May be more receptive to step change innovation

Growth in consumer healthcare may change this •  Sales direct to customer – self care, self diagnosis, self monitoring

•  More freedom within allied professions – dentists, opticians, audiologists?

•  Products are there, service sector still immature