KENYA STANDARD KS ISO 80601-2-74:2017 ICS 11.040.10 Second Edition © KEBS 2020 Medical electrical equipment Part 2-74: Particular requirements for basic safety and essential performance of respiratory humidifying equipment APPROVED 2020-04-08 FREE LICENSE BY KENYA BUREAU OF STANDARDS (KEBS) TO INTERESTED PARTIES, STAKEHOLDERS MANUFACTURERS AND PUBLIC, FOR A LIMITED ACCESS DURING COVID-19 EMERGENCY, ON 2020-04-15 - FOR PERSONAL USE ONLY. COPYING, INTRANET AND INTERNET PROHIBITED

Transcript of Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee...

Page 1: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

KENYA STANDARD KS ISO 80601-2-74:2017

ICS 11.040.10

Second Edition

© KEBS 2020

Medical electrical equipment

Part 2-74: Particular requirements for

basic safety and essential performance of

respiratory humidifying equipment

APPROVED 2020-04-08

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 2: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

KS ISO 80601-2-74:2017

ii © KEBS 2020– All rights reserved

TECHNICAL COMMITTEE REPRESENTATION

The following organizations were represented on the Technical Committee:

Alpha Medical Manufacturers Ltd.

Association of Medical Engineers of Kenya (AMEK)

Christian Health Association of Kenya (CHAK)

Diverse Management Consultants Ltd.

Equra Health Kenya

Getrudes Children Hospital

Hydromatics Ltd.

Jad’s Diagnostics EA Ltd.

Kenya Accreditation Service (KENAS)

Kenya Association of Physicians

Kenya Medical Supplies Agency (KEMSA)

Kenya National Chambers of Commerce and Industry

Medtronic

Ministry of Health

MP Shah Hospital

Roche Diagnostics

Stratus Medical Imaging Solutions Ltd.

Sub-county Health Services, Kamukunji

Kenya Bureau of Standards — Secretariat

REVISION OF KENYA STANDARDS

In order to keep abreast of progress in industry, Kenya Standards shall be regularly reviewed. Suggestions

for improvements to published standards, addressed to the Managing Director, Kenya Bureau of Standards,

are welcome.

© Kenya Bureau of Standards, 2020

Copyright. Users are reminded that by virtue of Section 25 of the Copyright Act, Cap. 130 of 2001 of the Laws of Kenya, copyright

subsists in all Kenya Standards and except as provided under Section 25 of this Act, no Kenya Standard produced by Kenya Bureau of

Standards may be reproduced, stored in a retrieval system in any form or transmitted by any means without prior permission in writing

from the Managing Director.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 3: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

KENYA STANDARD KS ISO 80601-2-74:2017

ICS 11.040.10

Second Edition

© KEBS 2020– All rights reserved iii

Medical electrical equipment

Part 2-74: Particular requirements for basic safety and

essential performance of respiratory humidifying

equipment

Kenya Bureau of Standards, Popo Road, Off Mombasa Road, P.O. Box 54974 - 00200, Nairobi, Kenya

+254 020 6948000, + 254 722202137, + 254 734600471

[email protected]

@KEBS_ke

kenya bureau of standards (kebs)

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 4: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

KS ISO 80601-2-74:2017

iv © KEBS 2020 – All rights reserved

NATIONAL FOREWORD

This Kenya Standard was prepared by the Medical Devices, Instruments and Hospital Equipment Technical

Committee under the guidance of the Standards Projects Committee, and it is in accordance with the

procedures of the Kenya Bureau of Standards

This standard is identical with and has been reproduced from ISO 80601-2-74, Medical electrical equipment

— Part 2-74: Particular requirements for basic safety and essential performance of respiratory humidifying

equipment, published by the International Organization for Standardization (ISO). The National Standards

Council has endorsed the adoption of the 2017 edition of this standard as a Kenya Standard.

For the purposes of this standard, the ISO text should be modified as follows:

a) Terminology

The words ‘this Kenya Standard’ should replace the words ‘this ISO Standard’ whenever they appear.

b) References

The references to ISO should be replaced by references to the appropriate Kenya Standards, where they

have been declared.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 5: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

© ISO 2017

Medical electrical equipment —Part 2-74: Particular requirements for basic safety and essential performance of respiratory humidifying equipmentAppareils électromédicaux —Partie 2-74: Exigences particulières pour la sécurité de base et les performances essentielles des équipements d’humidification respiratoire

INTERNATIONAL STANDARD

ISO80601-2-74

First edition2017-05

Reference numberISO 80601-2-74:2017(E)

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 6: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

ii © ISO 2017 – All rights reserved

COPYRIGHT PROTECTED DOCUMENT

© ISO 2017, Published in SwitzerlandAll rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized otherwise in any form or by any means, electronic or mechanical, including photocopying, or posting on the internet or an intranet, without prior written permission. Permission can be requested from either ISO at the address below or ISO’s member body in the country of the requester.

ISO copyright officeCh. de Blandonnet 8 • CP 401CH-1214 Vernier, Geneva, SwitzerlandTel. +41 22 749 01 11Fax +41 22 749 09 [email protected]

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 7: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved iii

Contents Page

Foreword ............................................................................................................................................................. v 

Introduction ..................................................................................................................................................... vii 

201.1 Scope, object and related standards ................................................................................... 1 201.2 Normative references ............................................................................................................... 3 201.3 Terms and definitions .............................................................................................................. 5 201.4 General requirements .............................................................................................................. 7 201.5 General requirements for testing of ME EQUIPMENT ..................................................... 10 201.6 Classification of ME EQUIPMENT and ME SYSTEMS .............................................................. 11 201.7 ME EQUIPMENT identification, marking and documents .............................................. 11 201.8 Protection against electrical HAZARDS form ME EQUIPMENT ........................................ 19 201.9 Protection against mechanical hazards of ME EQUIPMENT and ME SYSTEMS ........... 19 201.10 Protection against unwanted and excessive radiation HAZARDS .......................... 21 201.11 Protection against excessive temperatures and other HAZARDS .......................... 21 201.12 Accuracy of controls and instruments and protection against

hazardous outputs .................................................................................................................... 24 201.13 HAZARDOUS SITUATIONS and fault conditions for ME EQUIPMENT .............................. 27 201.14 Programmable electrical medical systems (pems) .................................................. 28 201.15 Construction of ME EQUIPMENT ........................................................................................... 28 201.16 ME SYSTEMS ............................................................................................................................... 29 201.17 Electromagnetic compatibility of ME EQUIPMENT and ME SYSTEMS .......................... 29 201.101 BREATHING SYSTEM connectors and ports ..................................................................... 29 201.102 Requirements for the BREATHING SYSTEM and ACCESSORIES ..................................... 32 201.103 LIQUID CONTAINER .................................................................................................................. 33 201.104 FUNCTIONAL CONNECTION ...................................................................................................... 34 202 Electromagnetic disturbances — Requirements and tests ......................................... 34 206 Usability ......................................................................................................................................... 35 208 General requirements, tests and guidance for alarm systems in medical

electrical equipment and medical electrical systems .................................................. 36 211 Requirements for medical electrical equipment and medical electrical

systems used in the home healthcare environment ..................................................... 36 Annex C (informative) Guide to marking and labelling requirements for

ME EQUIPMENT and ME SYSTEMS ................................................................................................. 38 

Annex D (informative) Symbols on Marking .................................................................................... 44 

Annex AA (informative) Particular guidance and rationale ...................................................... 45 

Annex BB (normative) * Determination of the accuracy of the displayed MEASURED GAS TEMPERATURE ......................................................................................................................... 61 

Annex CC (normative) * Determination of the HUMIDIFICATION OUTPUT .................................... 63 

Annex DD (normative) * Specific enthalpy calculations .............................................................. 67 

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 8: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

iv ©ISO/IEC2016–Allrightsreserved

Annex EE (normative) Removable temperature sensors and mating ports ......................... 69 

Annex FF (normative) * Standard temperature sensor ............................................................... 73 

Annex GG (informative) Saturation vapour pressure ................................................................... 76 

Annex HH (informative) Reference to the essential principles of safety and performance of medical devices in accordance with ISO 16142-1:2016[7] .......... 77 

Annex II (informative) Terminology — Alphabetized index of defined terms .................... 81 

Bibliography ................................................................................................................................................... 85 

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 9: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved v

Foreword

ISO (the InternationalOrganization for Standardization) is aworldwide federationofnationalstandards bodies (ISO member bodies). The work of preparing International Standards isnormally carried out through ISO technical committees. Each member body interested in asubjectforwhichatechnicalcommitteehasbeenestablishedhastherighttoberepresentedonthat committee. International organizations, governmental and non‐governmental, in liaisonwith ISO, also take part in the work. ISO collaborates closely with the InternationalElectrotechnicalCommission(IEC)onallmattersofelectrotechnicalstandardization.

Theproceduresusedtodevelopthisdocumentandthoseintendedforitsfurthermaintenanceare described in the ISO/IEC Directives, Part1. In particular the different approval criterianeededforthedifferenttypesofISOdocumentsshouldbenoted.Thisdocumentwasdraftedinaccordance with the editorial rules of the ISO/IECDirectives, Part2 (seewww.iso.org/directives).

Attention is drawn to the possibility that some of the elements of this documentmay be thesubjectofpatentrights.ISOshallnotbeheldresponsibleforidentifyinganyorallsuchpatentrights.Detailsofanypatentrightsidentifiedduringthedevelopmentofthedocumentwillbeinthe Introduction and/or on the ISO list of patent declarations received (seewww.iso.org/patents).

Any tradenameused in this document is information given for the convenience of users anddoesnotconstituteanendorsement.

ForanexplanationonthemeaningofISOspecifictermsandexpressionsrelatedtoconformityassessment, as well as information about ISO's adherence to the World Trade Organization(WTO) principles in the Technical Barriers to Trade (TBT) see the following URL:www.iso.org/iso/foreword.html.

The committee responsible for this document is ISO/TC121, Anaesthetic and respiratory equipment,SubcommitteeSC3,Lung ventilators and related equipmentandIEC/TC62,Electrical equipment in medical practice,SubcommitteeSC62D,Electromedical equipment.

This firsteditionof ISO80601‐2‐74cancelsandreplaces the thirdeditionof ISO8185:2007[1],which has been technically revised. It also incorporates the third edition of IEC60601‐1,includingamendment1,thefourtheditionofIEC60601‐1‐2,thethirdeditionofIEC60601‐1‐6,includingamendment1, thesecondeditionof IEC60601‐1‐8, includingamendment1,andthesecondeditionofIEC60601‐1‐11.

Themostsignificantchangesarethefollowingmodifications:

— extendingthescopetoincludetheHUMIDIFIERanditsACCESSORIES,wherethecharacteristicsofthoseACCESSORIEScanaffecttheBASICSAFETYandESSENTIALPERFORMANCEoftheHUMIDIFIER,andthusnotonlytheHUMIDIFIERitself;

— identificationofESSENTIALPERFORMANCEforaHUMIDIFIERanditsACCESSORIES;

— modification of the humidification test PROCEDURE and the disclosure of humidificationperformance;

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 10: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

vi ©ISO2017–Allrightsreserved

— morefullydimensioningtheremovabletemperaturesensorportandsensor;

— removalofrequirementsforso‐called“bubble”HUMIDIFIERSasaseparatedocumentisbeingpreparedforthem[8];

andthefollowingadditions:

— requirementsformechanicalstrength(viaIEC60601‐1‐11);

— newsymbols;

— requirementsforaHUMIDIFIERasacomponentofanMESYSTEM;

— requirementsforENCLOSUREintegrity(wateringressviaIEC60601‐1‐11);

— requirementsforcleaninganddisinfectionPROCEDURES(viaIEC60601‐1‐11);

— requirementsforBIOCOMPATIBILITY;

— requirementsforUSABILITY.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 11: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved vii

Introduction

ThisdocumentspecifiesrequirementsforrespiratoryhumidifyingequipmentintendedforuseonPATIENTSinHOMEHEALTHCAREENVIRONMENTandinhealthcarefacilities.HUMIDIFIERSareusedtoraise thewater content of gasesdelivered to PATIENTS.Gases available formedical usedonotcontain sufficient moisture and can damage or irritate the respiratory tract or desiccatesecretions of PATIENTSwhose upper airways have been bypassed. Inadequate humidity at thePATIENT‐CONNECTION PORT can cause drying of the upper airway, or desiccation of tracheo‐bronchialsecretions inthetrachealortracheostomytube,whichcancausenarrowingorevenobstructionoftheairway[19][20].HeatisemployedtoincreasethewateroutputoftheHUMIDIFIER.

In addition, many HUMIDIFIERS utilize heated BREATHING TUBES in order to increase operatingefficiency and reduce water loss (condensate) as well as heat loss in the BREATHING TUBE.Ventilator and anaesthesia BREATHING TUBES in common use might not withstand the heatgeneratedbyHUMIDIFIERSandBREATHINGTUBEheatingmechanisms.

Many HUMIDIFIER MANUFACTURERS use off‐the‐shelf electrical connectors for their electricallyheatedBREATHINGTUBES.However,sincedifferentMANUFACTURERShaveusedthesameelectricalconnectorfordifferentpoweroutputs,electricallyheatedBREATHINGTUBEScanbephysically,butnotelectrically,interchangeable.UseofimproperelectricallyheatedBREATHINGTUBEShascausedoverheating,circuitmelting,PATIENTandOPERATORburnsandfires.Itwasnotfoundpracticaltospecify the interface requirements for electrical connectors to ensure compatibility betweenHUMIDIFIERSandBREATHINGTUBESproducedbydifferentMANUFACTURERS.

Since the safe use of a HUMIDIFIER depends on the interaction of the HUMIDIFIERwith itsmanyACCESSORIES, this document sets total system performance requirements up to the PATIENT‐CONNECTION PORT. These requirements are applicable to ACCESSORIES such as BREATHING TUBES(both heated and non‐heated), temperature sensors and equipment intended to control theenvironmentwithintheseBREATHINGTUBES.

Humidification can also be used by respiratory support MEEQUIPMENT to increase PATIENTcomfortandcompliancewiththetherapy.Examplesareobstructivesleepapnoeaandnasalhighflowtherapyequipment.TheHUMIDIFICATIONOUTPUTrequirementsofsuchMEEQUIPMENTis lessdemandingasthePATIENT’Supperairwayisnotbypassed.

HUMIDIFIERSarecommonlyusedwithairandair‐oxygenmixturesandanyHUMIDIFIERshouldbeabletooperatewiththesegases.Careshouldbetakenifusingothergasmixessuchashelium‐oxygenmixtures,asthedifferentphysicalandthermalpropertiesofthesegasesmaydisturbtheoperationoftheHUMIDIFIER.

Inthisdocument,thefollowingprinttypesareused:

—Requirementsanddefinitions:romantype;

—Test specifications: italic type;

—Informativematerialappearingoutsideoftables,suchasnotes,examplesandreferences:insmallertype.Normativetextoftablesisalsoinasmallertype;

—Terms defined in CLAUSE3 of the general standard, in this document or as noted: smallcapitals;

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 12: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

viii ©ISO2017–Allrightsreserved

Inreferringtothestructureofthisdocument,theterm

—“clause”meansoneofthefivenumbereddivisionswithinthetableofcontents,inclusiveofallsubdivisions(e.g.Clause201includessubclauses201.7,201.8,etc.);

—“subclause”meansanumberedsubdivisionof a clause (e.g.201.7,201.8and201.9areallsubclausesofClause201).

Referencestoclauseswithinthisdocumentareprecededbytheterm“Clause”followedbytheclausenumber.Referencestosubclauseswithinthisdocumentarebynumberonly.

Inthisdocument, theconjunctive“or” isusedasan“inclusiveor”soastatement is true ifanycombinationoftheconditionsistrue.

TheverbalformsusedinthisdocumentconformtousagedescribedinAnnexHoftheISO/IECDirectives,Part2.Forthepurposesofthisdocument,theauxiliaryverb:

— “shall”meansthatcompliancewitharequirementoratestismandatoryforcompliancewiththisdocument;

— “should”means that compliancewith a requirement or a test is recommended but is notmandatoryforcompliancewiththisdocument;

— “may” is used todescribeapermissibleway to achieve compliancewith a requirementortest.

Anasterisk (*)as the first characterof a titleorat thebeginningof aparagraphor table titleindicatesthatthereisguidanceorrationalerelatedtothatiteminAnnexAA.

TheattentionofMemberBodiesandNationalCommitteesisdrawntothefactthatequipmentmanufacturersandtestingorganizationsmayneedatransitionalperiodfollowingpublicationofanew,amendedorrevisedISOorIECpublicationinwhichtomakeproductsinaccordancewiththe new requirements and to equip themselves for conducting new or revised tests. It is therecommendation of the committees that the content of this publication be adopted forimplementationnationallynotearlierthan3years fromthedateofpublicationforequipmentnewlydesignedandnotearlierthan5yearsfromthedateofpublicationforequipmentalreadyinproduction.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 13: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

INTERNATIONAL STANDARD ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved 1

Medical electrical equipment —

Part 2-74: Particular requirements for basic safety and essential performance of respiratory humidifying equipment

201.1 Scope, object and related standards

IEC60601‐1:2005+AMD1:2012,Clause1applies,exceptasfollows.

201.1.1 * Scope

Replacement:

This document applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of a HUMIDIFIER, alsohereafterreferredtoasMEEQUIPMENT,incombinationwithitsACCESSORIES,thecombinationalsohereafterreferredtoasMESYSTEM.

This document is also applicable to those ACCESSORIES intended by their MANUFACTURER to beconnected to a HUMIDIFIER where the characteristics of those ACCESSORIES can affect the BASICSAFETYorESSENTIALPERFORMANCEoftheHUMIDIFIER.

EXAMPLE1 Heated BREATHING TUBES (heated‐wire BREATHING TUBES) or MEEQUIPMENT intended tocontroltheseheatedBREATHINGTUBES(heatedBREATHINGTUBEcontrollers).

NOTE1 Heated BREATHING TUBES and their controllers are MEEQUIPMENT and are subject to therequirementsofIEC60601‐1.

NOTE2 ISO5367specifiesothersafetyandperformancerequirementsforBREATHINGTUBES.

Thisdocument includesrequirements for thedifferentmedicalusesofhumidification,suchasinvasive ventilation, non‐invasive ventilation, nasal high‐flow therapy, and obstructive sleepapnoeatherapy,aswellashumidificationtherapyfortracheostomyPATIENTS.

NOTE3 AHUMIDIFIERcanbeintegratedintootherequipment.Whenthisisthecase,therequirementsoftheotherequipmentalsoapplytotheHUMIDIFIER.

EXAMPLE2 Heated HUMIDIFIER incorporated into a critical care ventilator where ISO80601‐2‐12[12]alsoapplies.

EXAMPLE3 HeatedHUMIDIFIERincorporatedintoahomecareventilatorfordependentPATIENTSwhereISO80601‐2‐72[14]alsoapplies.

EXAMPLE4 HeatedHUMIDIFIERincorporatedintosleepapnoeatherapyequipmentwhereISO80601‐2‐70[13]alsoapplies.

This document also includes requirements for an ACTIVEHME (HEAT AND MOISTURE EXCHANGER),MEEQUIPMENTwhichactively addsheat andmoisture to increase thehumidity level of thegasdeliveredfromtheHMEtothePATIENT.ThisdocumentisnotapplicabletoapassiveHME,whichreturnsaportionoftheexpiredmoistureandheatofthePATIENTtotherespiratorytractduringinspirationwithoutaddingheatormoisture.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 14: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

2 ©ISO2017–Allrightsreserved

NOTE4 ISO9360‐1[5]andISO9360‐2[6]specifythesafetyandperformancerequirementsforapassiveHME.

If a clause or subclause is specifically intended to be applicable to MEEQUIPMENT only, or toMESYSTEMSonly, the title and contentof that clauseor subclausewill say so. If that isnot thecase,theclauseorsubclauseappliesbothtoMEEQUIPMENTandtoMESYSTEMS,asrelevant.

HAZARDS inherent in the intendedphysiological functionofMEEQUIPMENTorMESYSTEMSwithinthescopeofthisdocumentarenotcoveredbyspecificrequirementsinthisdocumentexceptinIEC60601‐1:2005+AMD1:2012,7.2.13and8.4.1.

NOTE5 AdditionalinformationcanbefoundinIEC60601‐1:2005+AMD1:2012,4.2.

This document does not specify the requirements for cold pass‐over or cold bubble‐throughhumidificationdevices,therequirementsforwhicharegiveninISO20789:—.[8]

Thisdocument isnotapplicable toequipmentcommonlyreferred toas “roomhumidifiers”orhumidifiers used in heating, ventilation and air conditioning systems, or HUMIDIFIERSincorporatedintoinfantincubators.

ThisdocumentisnotapplicabletonebulizersusedforthedeliveryofdrugstoPATIENTS.

NOTE6 ISO27427[10]specifiesthesafetyandperformancerequirementsfornebulizers.

ThisdocumentisaparticularstandardintheIEC60601‐1andtheISO/IEC80601series.

201.1.2 Object

Replacement:

The object of this document is to establish particular BASIC SAFETY and ESSENTIAL PERFORMANCErequirementsforaHUMIDIFIER,asdefinedin201.3.209,anditsACCESSORIES.

NOTE ACCESSORIESareincludedbecausethecombinationoftheHUMIDIFIERandtheACCESSORIESneedsto be adequately safe. ACCESSORIES can have a significant impact on the BASIC SAFETY or ESSENTIALPERFORMANCEofaHUMIDIFIER.

201.1.3 Collateral standards

Addition:

Thisdocumentrefers to thoseapplicablecollateralstandards thatare listed inClause2of thegeneralstandard1andClause201.2ofthisdocument.

IEC60601‐1‐2, IEC60601‐1‐6, IEC60601‐1‐8 and IEC60601‐1‐11 apply as modified inClauses202, 206, 208 and 211, respectively. IEC60601‐1‐3[15] does not apply. All otherpublishedcollateralstandardsintheIEC60601‐1seriesapplyaspublished.

201.1.4 Particular standards

Replacement:

In the IEC60601 series, particular standards may modify, replace or delete requirementscontained in the general standard and collateral standards as appropriate for the particular

1ThegeneralstandardisIEC60601‐1:2005+AMD1:2012.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 15: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved 3

MEEQUIPMENTunderconsideration,andmayaddotherBASICSAFETYandESSENTIALPERFORMANCErequirements.

Arequirementofaparticularstandardtakespriorityoverthegeneralstandard.

For brevity, IEC60601‐1 is referred to in this document as the general standard. Collateralstandardsarereferredtobytheirdocumentnumber.

Thenumberingof clausesandsubclausesof thisdocumentcorresponds to thatof thegeneralstandardwiththeprefix“201”(e.g.201.1inthisdocumentaddressesthecontentofClause1ofthegeneralstandard)orapplicablecollateralstandardwiththeprefix“20x”,wherexisthefinaldigit(s)ofthecollateralstandarddocumentnumber(e.g.202.4inthisdocumentaddressesthecontentofClause4oftheIEC60601‐1‐2collateralstandard,208.6inthisdocumentaddressesthecontentofClause6oftheIEC60601‐1‐8collateralstandard,etc.).Thechangestothetextofthegeneralstandardarespecifiedbytheuseofthefollowingwords:

“Replacement” means that the clause or subclause of the general standard or applicablecollateralstandardisreplacedcompletelybythetextofthisdocument.

“Addition”meansthatthetextofthisdocumentisadditionaltotherequirementsofthegeneralstandardorapplicablecollateralstandard.

“Amendment”meansthattheclauseorsubclauseofthegeneralstandardorapplicablecollateralstandardisamendedasindicatedbythetextofthisdocument.

Subclauses,figuresortableswhichareadditionaltothoseofthegeneralstandardarenumberedstarting from 201.101. However, due to the fact that definitions in the general standard arenumbered3.1 through3.139,additionaldefinitions in thisdocumentarenumberedbeginningfrom201.3.201.AdditionalannexesareletteredAA,BB,etc.,andadditionalitemsaa),bb),etc.

Subclauses,figuresortableswhichareadditionaltothoseofacollateralstandardarenumberedstartingfrom20x,where“x”isthenumberofthecollateralstandard,e.g.202forIEC60601‐1‐2,203forIEC6060‐1‐3,etc.

The term “this document” is used tomake reference to the general standard, any applicablecollateralstandardsandthisparticulardocumenttakentogether.

Wherethereisnocorrespondingclauseorsubclauseinthisdocument,theclauseorsubclauseofthegeneralstandardorapplicablecollateralstandard,althoughpossiblynotrelevant,applieswithoutmodification;where it is intended thatanypartof thegeneral standardorapplicablecollateralstandard,althoughpossiblyrelevant,isnottobeapplied,astatementtothateffectisgiveninthisdocument.

201.2 Normative references

The following documents are referred to in the text in such a way that some or all of theircontentconstitutesrequirementsofthisdocument.Fordatedreferences,onlytheeditioncitedapplies. For undated references, the latest edition of the referenced document (including anyamendments)applies.

IEC60601‐1:2005+AMD1:2012,Clause2applies,exceptasfollows.

Replacement:

IEC60601‐1‐2:2014, Medical electrical equipment — Part 1-2: General requirements for basic safety and essential performance — Collateral Standard: Electromagnetic disturbances — Requirements and tests

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 16: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

4 ©ISO2017–Allrightsreserved

IEC60601‐1‐6:2010+AMD1:2013,Medical electrical equipment — Part 1-6: General requirements for basic safety and essential performance — Collateral Standard: Usability

IEC60601‐1‐8:2006+AMD1:2012,Medical electrical equipment — Part 1-8: General requirements for basic safety and essential performance — Collateral standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems

Addition:

ISO3744:2010,Acoustics — Determination of sound power levels and sound energy levels of noise sources using sound pressure — Engineering methods for an essentially free field over a reflecting plane

ISO5356‐1:2015,Anaesthetic and respiratory equipment — Conical connectors — Part 1: Cones and sockets

ISO5367:2014,Anaesthetic and respiratory equipment — Breathing sets and connectors

ISO7396‐1:2016,Medical gas pipeline systems — Part 1: Pipeline systems for compressed medical gases and vacuum

ISO14937:2009, Sterilization of health care products — General requirements for characterization of a sterilizing agent and the development, validation and routine control of a sterilization process for medical devices

ISO15223‐1:2016,Medical devices — Symbols to be used with medical device labels, labelling and information to be supplied — Part 1: General requirements

ISO17664:2004, Sterilization of medical devices — Information to be provided by the manufacturer for the processing of resterilizable medical devices

ISO18562‐1:—2, Biocompatibility evaluation of breathing gas pathways in healthcare applications — Part 1: Evaluation and testing within a risk management process

ISO80369‐1:2010, Small-bore connectors for liquids and gases in healthcare applications — Part 1: General requirements

IEC60601‐1:2005+AMD1:2012,Medical electrical equipment — Part 1: General requirements for basic safety and essential performance

IEC60601‐1‐11:2015,Medical electrical equipment — Part 1-11: General requirements for basic safety and essential performance — Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment

IEC60601‐2‐19:2009,Medical electrical equipment — Part 2-19: Particular requirements for the basic safety and essential performance of infant incubators

2Tobepublished.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 17: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved 5

IEC61672‐1:2013,Electroacoustics — Sound level meters — Part 1: Specifications

IEC62304:2006+AMD1:2015,Medical device software — Software life cycle processes

EN15986:2011,Symbol for use in the labelling of medical devices — Requirements for labelling of medical devices containing phthalates

201.3 Terms and definitions

For the purposes of this document, the terms and definitions given in ISO4135:2001[1],ISO7396‐1:2016, ISO9360‐1:2000[5], ISO23328‐2:2002[9], IEC60601‐1:2005+AMD1:2012,IEC60601‐1‐2:2014, IEC60601‐1‐8:2006+AMD1:2012, IEC60601‐1‐11:2015,IEC62366‐1:2015[18]andthefollowingapply.

ISO and IEC maintain terminological databases for use in standardization at the followingaddresses:

— IECElectropedia:availableathttp://www.electropedia.org/

— ISOOnlinebrowsingplatform:availableathttp://www.iso.org/obp

NOTE AnalphabetizedindexofdefinedtermsisfoundbeginninginAnnexDD.

IEC60601‐1:2005+AMD1:2012,Clause3applies,exceptasfollows:

Additions:

201.3.201ACTIVE HMEACTIVE HEAT AND MOISTURE EXCHANGERHUMIDIFIERwherewater,water vapourorheat is actively added to theHEAT ANDMOISTURE EXCHANGER(HME)toincreasethehumiditylevelofthegasflowingfromtheHMEtothePATIENT

201.3.202AIRWAY PRESSUREPawpressureatthepatient‐connectionport

201.3.203BODY TEMPERATURE PRESSURE, SATURATEDBTPSambientatmosphericpressureandaRELATIVEHUMIDITYof100%atatemperatureof37°C

Note1toentry:RespiratoryphysiologylungvolumesandflowsarestandardizedtoBTPS.

201.3.204DELIVERED GAS TEMPERATUREtemperatureofthegas,oraerosol,orboth,atthePATIENT‐CONNECTIONPORT

201.3.205FLOW-DIRECTION-SENSITIVE COMPONENTcomponentorACCESSORYthroughwhichgasflowhastobeinonedirectiononlyforproperfunctioningorPATIENTsafety

[SOURCE:ISO4135:2001,3.1.7[1],modified— added“orACCESSORY” andreplaced“must”with“hasto”.]

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 18: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

6 ©ISO2017–Allrightsreserved

201.3.206HEATED BREATHING TUBE CONTROLLERMEEQUIPMENTwhichcontrolsthetemperatureortheheatingofaBREATHINGTUBE

Note1toentry:AHEATEDBREATHINGTUBECONTROLLERcanbeeitherstand‐aloneorpartoftheHUMIDIFIER.

201.3.207HUMIDIFICATION CHAMBERpartoftheHUMIDIFIERinwhichvaporizationornebulizationtakesplace

201.3.208HUMIDIFICATION OUTPUTtotalmassofwatervapourperunitvolumeofgasatthePATIENT‐CONNECTIONPORTunderBTPSconditions

201.3.209HUMIDIFIERMEEQUIPMENTthataddswaterintheformofdropletsorvapour,orboth,totheinspiredgas

EXAMPLEVaporizingandultrasonicequipmentandanACTIVEHMEaretypesofHUMIDIFIERS.

201.3.210LIQUID CONTAINERpartoftheHUMIDIFIERwhichholdstheliquid

Note1toentry:TheLIQUIDCONTAINERcanbeaccessibletothebreathinggas.

Note2toentry:TheLIQUIDCONTAINERcanalsobepartoftheHUMIDIFICATIONCHAMBER.

Note3toentry:TheLIQUIDCONTAINERcanbedetachableforfilling.

201.3.211LIQUID RESERVOIRpartoftheHUMIDIFIERwhichreplenishestheLIQUIDCONTAINER

201.3.212MAXIMUM LIMITED PRESSUREPLIMmaxhighestAIRWAYPRESSUREduringNORMALUSEorunderSINGLEFAULTCONDITION

201.3.213MAXIMUM OPERATING PRESSUREmaximumpressureintheHUMIDIFICATIONCHAMBERduringNORMALUSE

201.3.214MEASURED GAS TEMPERATUREtemperatureofthegas,oraerosol,orboth,thattheMESYSTEMismeasuringand,ifapplicable,displaying

201.3.215MONITORING EQUIPMENTMEEQUIPMENTorpartthatcontinuouslyorcontinuallymeasuresandindicatesthevalueofavariabletotheOPERATOR

[SOURCE:ISO80601‐2‐12:2011,201.3.217[12]]

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 19: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved 7

201.3.216PATIENT-CONNECTION PORTportatthePATIENT‐endoftheBREATHINGTUBESintendedforconnectiontoanairwaydevice

EXAMPLEA tracheal tube, tracheostomy tube, facemask and supralaryngeal airway are all airwaydevices.

201.3.217PROTECTION DEVICEpartorfunctionofMEEQUIPMENTthat,withoutinterventionbytheOPERATOR,protectsthePATIENTfromhazardousoutputduetoincorrectdeliveryofenergyorsubstances

[SOURCE:ISO60601‐2‐12:2011,201.3.220[12]]

201.3.218RELATIVE HUMIDITYwater vapour pressure, expressed as a percentage of the saturation vapour pressure, at a particulartemperature

201.3.219SET TEMPERATUREtemperatureatwhichtheHUMIDIFIERattemptstomaintainMEASUREDGASTEMPERATURE

Note1toentry:TheSETTEMPERATUREmaybeOPERATOR‐adjustable.

201.3.220STANDARD TEMPERATURE AND PRESSURE, DRYSTPDpressureof101,325kPaatanoperatingtemperatureof20°C

201.3.221VALIDATIONconfirmation,throughtheprovisionofOBJECTIVEEVIDENCE,thattherequirementsforaspecificINTENDEDUSEorapplicationhavebeenfulfilled

Note1toentry: The OBJECTIVE EVIDENCE needed for a VALIDATION is the result of a test or other form ofdeterminationsuchasperformingalternativecalculationsorreviewingdocuments.

Note2toentry:Theterm“VALIDATED”isusedtodesignatethecorrespondingstatus.

Note3toentry:TheuseconditionsforVALIDATIONcanberealorsimulated.

[SOURCE:ISO9000:2015,3.8.13[4]]

201.4 General requirements

IEC60601‐1:2005+AMD1:2012,Clause4applies,exceptasfollows.

201.4.3 ESSENTIAL PERFORMANCE

IEC60601‐1:2005+AMD1:2012,4.3applies,exceptasfollows.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 20: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

8 ©ISO2017–Allrightsreserved

Additional subclause:

201.4.3.101 * Additional requirements for ESSENTIAL PERFORMANCE

Additional ESSENTIAL PERFORMANCE requirements are found in the subclauses listed inTable201.101.

Table 201.101 — Distributed ESSENTIAL PERFORMANCE requirements

Requirement Subclause

Forcategory1HUMIDIFIERS,deliveryofHUMIDIFICATIONOUTPUTorgenerationofaTECHNICALALARMCONDITION

201.12.1.101a

For HUMIDIFIERS equipped with SET TEMPERATURE MONITORINGEQUIPMENT,deliveryofHUMIDIFICATIONOUTPUTorgenerationofaTECHNICALorPHYSIOLOGICALALARMCONDITION

201.12.1.101a

For all other category 2 HUMIDIFIERS for the purposes of thisdocument, category 2 HUMIDIFIERS are considered to not haveESSENTIAL PERFORMANCE. Notwithstanding this fact, when thisdocumentreferstoESSENTIALPERFORMANCEasacceptancecriteria,thedeliveryofHUMIDIFICATIONOUTPUTisevaluated.a

aSubclause 202.8.1.101 indicates methods of evaluating delivery of HUMIDIFICATION OUTPUT asacceptancecriteriafollowingspecifictestsrequiredbythisdocument.

201.4.6 * ME EQUIPMENT or ME SYSTEM parts that contact the PATIENT

Amendment (add at end of 4.6 prior to the compliance check):

aa)TheGASPATHWAYSshallbesubject totherequirements forAPPLIEDPARTSaccordingtothissubclause.

bb)TheHUMIDIFIERoritspartsorACCESSORIESthatcancomeintocontactwiththePATIENTshallbesubjecttotherequirementsforAPPLIEDPARTSaccordingtothissubclause.

NOTE Forthepurposesofthisdocument,GASPATHWAYSareconsideredtobeanypartorsurfacethatcanbeexposedtothePATIENTrespiratorygas.

Additional subclauses:

201.4.11.101 * Additional requirements for pressurized gas input

201.4.11.101.1 Overpressure requirement

a) IftheHUMIDIFIERisintendedtobeconnectedtoaMEDICALGASPIPELINESYSTEMcomplyingwithISO7396‐1:2016,thenit:

1) shalloperateandmeettherequirementsofthisdocumentthroughoutitsRATEDrangeofinputpressure;

2) shallnotcauseanunacceptableRISKundertheSINGLEFAULTCONDITIONof1000kPa.

NOTE1 An internal pressure regulator can be required to accommodate the SINGLE FAULTCONDITIONofmaximuminputpressure,aswellastheRATEDrangeofinputpressure.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 21: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved 9

NOTE2 UndertheSINGLEFAULTCONDITIONofoverpressure,itisdesirableforgastocontinuetoflowtotheBREATHINGSYSTEM.Underthiscondition,theflowratefromtheHUMIDIFIERislikelytobeoutsideofitsspecification.

b) If theHUMIDIFIERhasamaximumRATEDinputpressureinexcessof600kPa,theHUMIDIFIERshallnotcauseanunacceptableRISKundertheSINGLEFAULTCONDITIONoftwicethemaximumRATEDinputpressure.

Check compliance by functional testing in NORMAL USE and under NORMAL CONDITION with the most adverse operating settings, by functional testing in SINGLE FAULT CONDITION and inspection of the RISK MANAGEMENT FILE.

201.4.11.101.2 Compatibility requirement

IftheHUMIDIFIERisintendedtobedirectlyconnectedtoaMEDICALGASPIPELINESYSTEMcomplyingwithISO7396‐1:2016then:

a) theRATEDrangeofinputpressureshallcovertherangespecifiedinISO7396‐1:2016;

b) underNORMALCONDITION,

1) themaximum10saverageinputflowraterequiredbytheHUMIDIFIERforeachgasshallnotexceed60l/minatapressureof280kPa,measuredatthegasinputport,and

2) thetransientinputflowrateshallnotexceed200l/minaveragedfor3s.

or:

3) theACCOMPANYINGDOCUMENTSshalldisclose:

i) themaximum10saverageinputflowraterequiredbytheHUMIDIFIERforeachgasatapressureof280kPa,measuredatthegasinputport;

ii) the maximum transient input flowrate averaged for 3s required by theHUMIDIFIER for each gas at a pressure of 280kPa, measured at the gas inputport;

iii) awarningtotheeffectthatthisHUMIDIFIERisahighflowdeviceandshouldonlybe connected to a pipeline installation designed using a diversity factor thatallowsfortheindicatedhighflowataspecifiednumberofterminaloutlets,inordertoavoidexceedingthepipelinedesignflow,therebyminimizingtheRISKthattheHUMIDIFIERinterfereswiththeoperationofadjacentequipment.

Check compliance by functional testing in NORMAL USE and under NORMAL CONDITION with the most adverse operating settings and by inspection of the ACCOMPANYING DOCUMENTS.

EXAMPLE Highest driving gas consumption, highest gas delivery and, if provided, the highest RATED gas consumption at any gas power supply output.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 22: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

10 ©ISO2017–Allrightsreserved

201.4.101 Additional general requirements

HUMIDIFIERSarefrequentlyusedincombinationwithotherrespiratoryMEEQUIPMENTormedicaldevices. The BASIC SAFETY and ESSENTIAL PERFORMANCE of both the HUMIDIFIER and the otherrespiratoryMEEQUIPMENTormedicaldeviceareinterdependent.

a) Where a HUMIDIFIER is intended to be used in combination with other respiratoryMEEQUIPMENTormedicaldevicesasindicatedinitsinstructionsforuse,itshallbeevaluatedincombinationwiththeotherrespiratoryMEEQUIPMENTormedicaldeviceswhenapplyingtherequirementsofthisdocument.

b) As appropriate, the requirements of the particular standards of the other respiratoryMEEQUIPMENTormedicaldevicesindicatedintheinstructionsforuseoftheHUMIDIFIERshallalso apply to the combination of the HUMIDIFIER and other respiratory MEEQUIPMENT ormedicaldevices.

201.5 General requirements for testing of ME EQUIPMENT

IEC60601‐1:2005+AMD1:2012,Clause5applies,exceptasfollows.

201.5.4 Other conditions

Amendment (add to the list):

aa)Unlessotherwisespecified,theLIQUIDCONTAINERandLIQUIDRESERVOIR,ifprovided,shallbefilled tomaximumcapacity, as indicated in the instructions foruse, at thebeginningof a testwithdistilledwaterattheambienttesttemperature.

bb)Forthepurposeofcheckingcompliancewithrequirementsofthisdocument,theDELIVEREDGASTEMPERATUREshallbesensedintheBREATHINGTUBEnotmorethan50mmfromthePATIENT‐CONNECTIONPORT(seeAnnexBB).

Additional subclauses:

201.5.101 Additional requirements for general requirements for testing of ME EQUIPMENT

201.5.101.1 Humidifier test conditions

a) Fortesting,theHUMIDIFIER:

1) shall be connected to gas supplies as specified for NORMAL USE, except that industrialgradeoxygen;and

2) airmaybesubstitutedfortheequivalentmedicalgas,asappropriate,unlessotherwisestated.

b) Whenusingsubstitutegases,careshouldbetakentoensurethatthetestgasesareoil‐freeandappropriatelydry.

c) Themoisturecontentofallgassuppliesshallbelessthan1mg/l.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 23: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved 11

201.5.101.2 * Gas flowrate and leakage specifications

Inthisdocument,requirementsfortheflowrate,volumeandleakageareexpressedatSTANDARDTEMPERATURE AND PRESSURE, DRY (STPD), except for those associatedwith the BREATHING SYSTEM,whichareexpressedatBODYTEMPERATUREANDPRESSURE,SATURATED(BTPS).

Correct all test measurements to STPD or BTPS, as appropriate.

201.5.101.3 * HUMIDIFIER testing errors

a) For the purposes of this document, declared tolerances shall be adjusted by themeasurementuncertainty.

b) TheMANUFACTURERshalldisclosethemeasurementuncertaintyofeachdisclosedtoleranceinthetechnicaldescription.

Check compliance by inspection of the instructions for use and the technical description.

201.6 Classification of ME EQUIPMENT and ME SYSTEMS

IEC60601‐1:2005+AMD1:2012,Clause6applies,exceptasfollows.

Additional subclause:

201.6.101 HUMIDIFIER classification

201.6.101.1 Category 1

AHUMIDIFIER intended foruse inPATIENTSwhoseupperairwayshavebeenbypassed (invasiveventilation),shallbeclassifiedcategory1.

NOTE Acategory1HUMIDIFIERcanalsobesuitableforuseinPATIENTSwhoseupperairwayshavenotbeenbypassed.

201.6.101.2 Category 2

A HUMIDIFIER not intended for use in PATIENTSwhose upper airways have been bypassed (i.e.intended fornon‐invasiveventilation,nasalhigh flow therapy, sleepapnoeaCPAP treatment),shallbeclassifiedcategory2.

201.6.101.3 Classification

a) AHUMIDIFIERshallbeclassifiedeitherascategory1orcategory2.

b) A particular HUMIDIFIER may be classified as category1 over a certain specified range offlowratesandtemperatures,butalsoascategory2overawiderspecifiedrangeofflowratesandtemperatures.

201.7 ME EQUIPMENT identification, marking and documents

IEC60601‐1:2005+AMD1:2012,Clause7applies,exceptasfollows:

Additional subclause:

201.7.2.4.101 Additional requirements for ACCESSORIES

a) ACCESSORIESsuppliedseparatelyshall

1) fulfiltherequirementsof201.7.2.101,201.7.2.13.101and201.7.2.17.101,and

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 24: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

12 ©ISO2017–Allrightsreserved

2) bemarkedwithanindicationofanylimitationsoradverseeffectsoftheACCESSORYontheBASICSAFETYorESSENTIALPERFORMANCEoftheHUMIDIFIER,ifapplicable.

b) If marking the ACCESSORY is not practicable, this information may be placed in theinstructionsforuse.

NOTE The MANUFACTURER of the ACCESSORY can be the HUMIDIFIER MANUFACTURER or another entity(“third‐partymanufacturer”, healthcareproviderordurablemedical equipmentprovider) andall theseentitiesareexpectedtoensurecompliancewiththisrequirement.Additionalrequirementsarefoundin201.102.

Check compliance by inspection and inspection of the RISK MANAGEMENT FILE for any limitations or adverse effects of the ACCESSORY.

201.7.2.5 ME EQUIPMENT intended to receive power from other equipment

Amendment (add before the last paragraph):

NOTE For a heated BREATHING TUBE, the connector to the HUMIDIFIER or HEATED BREATHING TUBECONTROLLERisaconnectiontothesupplythatmightneedthismarking.

201.7.2.8.2 * Other power sources

Amendment (add at the end of the subclause):

NOTE TheconnectorontheHUMIDIFIERorHEATEDBREATHINGTUBECONTROLLERforaheatedBREATHINGTUBEisaconnectorthatmightneedthismarking.

Additional subclauses:

201.7.2.13.101 Additional requirements for physiological effects

a) Anynaturalrubberlatex‐containingcomponentsintheGASPATHWAYSorACCESSORIESshallbemarkedascontaininglatex.

b) SuchmarkingshallbeCLEARLYLEGIBLE.

c) Symbol5.4.5fromISO15223‐1:2016(Table201.D.1.101,symbol4)maybeused.

d) Theinstructionsforuseshalldiscloseanynaturalrubberlatex‐containingcomponents.

Check compliance by inspection.

201.7.2.17.101 Additional requirements for protective packaging

a) ThemarkingonpackagesshallbeCLEARLYLEGIBLEandshallinclude

1) adescriptionofthecontents,

2) anidentificationreferencetothebatch,typeorserialnumberorsymbols5.1.5,5.1.6or5.1.7fromISO15223‐1:2016(Table201.D.1.101,symbol1,symbol2orsymbol3),and

3) forpackagescontainingnaturalrubber latex, theword“LATEX”,orsymbol5.4.5 fromISO15223‐1:2016(Table201.D.1.101,symbol4).

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 25: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved 13

b) ForaspecificMODELORTYPEREFERENCE,theindicationofsingleuseshallbeconsistentfortheMODELORTYPEREFERENCE.

Check compliance by inspection.

201.7.2.101 Additional requirements for marking on the outside of ME EQUIPMENT or ME EQUIPMENT parts

a) ThemarkingofMEEQUIPMENT,partsorACCESSORIESshallbeCLEARLYLEGIBLEandshallincludethefollowing:

1) themaximumandminimumliquidlevels,ifthesearenecessarytomaintainBASICSAFETYorESSENTIALPERFORMANCEoftheHUMIDIFIER.

b) If applicable, marking of OPERATOR‐accessible MEEQUIPMENT, parts or ACCESSORIES shall beCLEARLYLEGIBLEandshallincludethefollowing:

2) thespecialstorage,handlingoroperatinginstructions;

3) theparticularwarningsand/orprecautionsrelevanttotheimmediateoperationoftheHUMIDIFIER;

4) anarrowindicatingthedirectionof the flowforFLOW‐DIRECTION‐SENSITIVECOMPONENTSthatareOPERATOR‐removablewithouttheuseofaTOOL;

5) ifapressure‐reliefPROTECTIONDEVICE isprovided, thepressureatwhich itopens.Thismarkingshallbeonornearthepressure‐reliefPROTECTIONDEVICE.

Check compliance by inspection.

201.7.4.3 * Units of measurement

IEC60601‐1:2005+AMD1:2012,7.4.3applies,exceptasfollows:

Amendment (add to the bottom as a new row in Table 1):

aa)Allgasvolume,flowrateandleakagespecifications

100) shallbeexpressedatstpd(standardtemperatureandpressure,dry),

101) exceptthoseassociatedwiththeBREATHINGSYSTEMwhichshallbeexpressedatBTPS(BODYTEMPERATUREANDPRESSURE,SATURATED).

201.7.9.1 Additional general requirements

IEC60601‐1:2005+AMD1:2012,7.9.1applies,exceptasfollows:

Amendment (replace the first dash with):

— Nameortradenameandaddressof:

— themanufacturer;

— where the MANUFACTURER does not have an address within the locale, an authorizedrepresentativewithinthelocale,

towhichtheRESPONSIBLEORGANIZATIONcanrefer;

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 26: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

14 ©ISO2017–Allrightsreserved

201.7.9.2 Instructions for use

IEC60601‐1:2005+AMD1:2012,7.9.2applies,exceptasfollows:

Additional subclauses:

201.7.9.2.1.101 Additional general requirements

a) ForaHUMIDIFIERintendedforuseintheHOMEHEALTHCAREENVIRONMENT,separateinstructionsforuseshallbeprovidedfor:

1) theLAYOPERATOR;

2) thesupervisingclinicianorthehealthcareprofessionalOPERATOR.

b) The MANUFACTURER may choose in which instructions for use to place the informationrequired by this document unless otherwise indicated in this document based on RISKMANAGEMENTandUSABILITYconsiderations.

c) ThesupervisingclinicianorthehealthcareprofessionalOPERATORinstructionsforuseshallincludetheinformationcontainedintheLAYOPERATORinstructionsforuse.

Check compliance by inspection of the instructions for use, the RISK MANAGEMENT FILE and USABILITY ENGINEERING FILE.

201.7.9.2.1.102 Additional general requirements

Theinstructionsforuseshallinclude:

a) iftheHUMIDIFIER,itspartsorACCESSORIESareintendedforsingleuse,informationonknowncharacteristicsandtechnicalfactorsknowntotheMANUFACTURERthatcouldposeaRISKiftheHUMIDIFIER,itspartsorACCESSORIESwerereused;

b) iftheHUMIDIFIER,itspartsorACCESSORIESareintendedforsingleuse,thatitissingleuseandinformationregardingtheintendeddurationofuse;

c) a statement on the quality and purity of thewater to be used in the HUMIDIFIER, and thataddingothersubstancescanhaveadverseeffects.

NOTE A nebulizer, located between the ventilator and the gas inlet port of the HUMIDIFICATIONCHAMBERisasourceofsuchsubstances.

Check compliance by inspection.

201.7.9.2.2.101 * Additional requirements for warnings and safety notices

Theinstructionsforuseshallinclude:

a) *a warning statement to the effect that “WARNING: Do not add any attachments oraccessoriestothehumidifierthatarenotlistedintheinstructionforuseofthehumidifieroraccessoryorthehumidifiermightnotfunctioncorrectlyaffectingthequalityofthetherapyorinjuringthepatient.”

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 27: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved 15

b) awarningstatementtotheeffect that“WARNING:Donotusethehumidifieratanaltitudeabove [insert maximum RATED altitude] or outside a temperature of [insert RATEDtemperature range].Using thehumidifieroutsideof this temperature rangeor above thisaltitudecanaffectthequalityofthetherapyorinjurethepatient.”

c) awarningstatementtotheeffectthat“WARNING:Topreventdisconnectionofthetubingortubingsystemduringuse,especiallyduringambulatoryuse,onlytubesincompliancewithISO5367orISO80601‐2‐74shouldbeused”.

d) if applicable, awarning statement to the effect that “WARNING: Covering breathing tubeswithablanketorheating them inan incubatororwithanoverheadheater canaffect thequalityofthetherapyorinjurethepatient.”

Check compliance by inspection of the instructions for use.

201.7.9.2.6 Installation

Amendment (add at the end of the subclause):

The instructions for use shall give recommended mounting methods and other relevantinformationforinstallationoftheHUMIDIFIER.

Additional subclauses:

201.7.9.2.8.101 * Additional requirements for start-up PROCEDURE

NOTE For the purposes of this document, a start‐up PROCEDURE is a pre‐use test that is used todeterminewhethertheHUMIDIFIERisreadyforuse.

a) IftheHUMIDIFIERisequippedwithanALARMSYSTEM,thentheinstructionsforusefortheLAYOPERATORshalldiscloseamethodbywhichtheALARMSIGNALScanbefunctionallytestedtodetermineiftheyareoperatingcorrectly.

b) Portions of this test method may be performed automatically by the HUMIDIFIER or mayrequireOPERATORaction.

EXAMPLE1 Combinationofthepower‐onself‐testroutinesandOPERATORactionsthatfunctionallychecktheALARMSIGNALS.

c) The specificationsof any required ACCESSORIES or test equipment needed toperform thesetestsshallbedisclosedintheinstructionsforuse.

EXAMPLE2 AtesttemperatureprobethatactivatestheALARMCONDITION.

Check compliance by inspection of the instructions for use.

201.7.9.2.9.101 Additional requirements for operating instructions

201.7.9.2.9.101.1 LAY OPERATOR operating instructions

TheinstructionsforusefortheLAYOPERATORshallinclude:

a) the conditions under which the HUMIDIFIER maintains the accuracy of controlled anddisplayedvariablesasdisclosedintheinstructionsforuse;

EXAMPLE1 Acceptablerangeofwaterlevel.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 28: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

16 ©ISO2017–Allrightsreserved

EXAMPLE2 Intervalofcalibrationofasensor.

b) anexplanationofthemeaningoftheIPclassificationmarkedontheMEEQUIPMENT;

c) themaximumvolumeofwater,expressedinml,availableforvaporizationcontainedintheLIQUIDCONTAINERand,ifprovided,intheLIQUIDRESERVOIR;

d) anindicationoftheexpecteddurationofoperationbetweenrefills,underspecifiedoperatingconditions;

Check compliance by inspection of the instructions for use.

201.7.9.2.9.101.2 * Supervising clinician operating instructions

The instructions for use intended for the supervising clinician or the healthcare professionalOPERATORshallinclude:

a) theMAXIMUMLIMITEDPRESSUREoftheHUMIDIFIERandACCESSORIES;

b) theMAXIMUMOPERATINGPRESSURE;

c) theRATEDrangeofenvironmentaloperatingconditions(temperatureandaltitude)ofNORMALUSE;

d) themaximumDELIVEREDGASTEMPERATURE,iftheHUMIDIFIERisnotprovidedwithameansofcontinuouslyindicatingtheMEASUREDGASTEMPERATURE;

e) *the location in the HUMIDIFIER or ACCESSORIES to which the displayed MEASURED GASTEMPERATUREisreferenced;

f) thegasleakageoftheHUMIDIFIERorindividualcomponents,asappropriate,atthemaximumRATEDpressure.Thegas leakageshouldbedetermined inaccordancewith ISO5367oranequivalentmethod.ThegasleakageforanACTIVEHMEshouldbedeterminedinaccordancewithISO9360‐1[5]orISO9360‐2[6];

g) *unlesstheHUMIDIFIERisintegratedintootherequipment,

1) theRATEDrangeofthefollowingcharacteristicsoftheassembledOPERATOR‐detachableparts,overwhichtheaccuraciesofsetandmonitoredhumidificationaremaintained:

i) flowrate;

ii) GASPATHWAYresistance;

iii) GASPATHWAYcompliance.

2) Thesespecificationsmaybepresentedinranges.

3) The accuracies of set andmonitored valuesmay be presented as a function of thesecharacteristics.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 29: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved 17

4) Since these values can be affected by the depletion of the liquid, the minimum andmaximumvaluesshallbedisclosed.

5) Compliance and resistance can be nonlinear. These characteristics might need to bespecifiedoverarange(e.g.at15l/min,30l/min,60l/min,maximumflowrateandthemaximumRATEDpressure).

6) TheresistanceandcomplianceshouldbedeterminedinaccordancewithISO5367oranequivalentmethod.

7) Theresistanceandcompliance foranACTIVEHMEshouldbedetermined inaccordancewithISO9360‐1[5]orISO9360‐2[6].

h) *unlesstheHUMIDIFIERisintegratedintootherequipment,thepressuredrop,asafunctionofflowrate,acrosstheHUMIDIFIERandACCESSORIESorindividualcomponents.

8) ThepressuredropshouldbedeterminedinaccordancewithISO5367oranequivalentmethod.

9) The pressure drop for an ACTIVE HME should be determined in accordance withISO9360‐1[5]orISO9360‐2[6].

i) the known adverse effects on the performance of the HUMIDIFIER when exposed to, forexample, electrocautery, electrosurgery, defibrillation, X‐ray (gamma radiation), infraredradiation,conductedtransientmagneticfieldsincludingmagneticresonanceimaging(MRI),andradiofrequencyinterference.

Ifapplicable,instructionsforuseshalldisclose

j) theessentialtechnicalcharacteristicsofeachrecommendedBREATHINGSYSTEMFILTER;

EXAMPLE Deadspaceandresistance.

k) *foraHUMIDIFIERthatentrainsairforthepurposeofdilutingoxygen:

10)a statement to the effect that the oxygen concentration can be affected by a partialobstructiondownstreamoftheHUMIDIFIER,e.g.whenusingACCESSORYequipment;

11)arecommendationthattheoxygenconcentrationbemeasuredatthepointofdeliverytothePATIENT.

Check compliance by inspection of the instructions for use.

201.7.9.2.12 Cleaning, disinfection and sterilization

IEC60601‐1:2005+AMD1:2012,7.9.2.12applies,exceptasfollows.

Amendment (add after NORMAL USE):

andSINGLEFAULTCONDITION

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 30: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

18 ©ISO2017–Allrightsreserved

Amendment (replace the second bullet):

aa)List the applicable parameters such as temperature, pressure, humidity, time limits andnumberofcyclesthatsuchMEEQUIPMENT,partsorACCESSORIEScantolerate.

bb)Alternatively,anothermethodtodeterminereducedperformanceandtheendofusefullifemaybeprovided.

Amendment (add after bulleted list):

cc)The instructions for use shall identify the portions of the GAS PATHWAYS through theHUMIDIFIERthatcanbecomecontaminatedwithbodyfluidsorexpiredgasesduringbothNORMALCONDITIONandSINGLEFAULTCONDITION.

Additional subclauses:

201.7.9.2.13.101 Additional requirements for maintenance

Theinstructionsforuseshalldisclose

a) adescriptionofperiodicvisualsafetyinspectionsthatshouldbeperformedbytheOPERATOR,and

b) if applicable, the INTERNAL ELECTRICAL POWER SOURCE care and maintenance PROCEDURES,includinginstructionsforrechargingorreplacement.

Check compliance by inspection of the instructions for use.

201.7.9.2.14.101 Additional requirements for ACCESSORIES, supplementary equipment, used material

TheinstructionsforuseofaHUMIDIFIERshallidentify

a) atleastonesetofACCESSORIES,and

b) ifapplicable,theMEEQUIPMENTnecessaryfortheHUMIDIFIER’SINTENDEDUSE.

Ifapplicable,theinstructionsforuseshalldisclose

c) anyrestrictionsonthepositioningofcomponentswithintheBREATHINGSYSTEM,and

EXAMPLE WheresuchcomponentsareFLOW‐DIRECTION‐SENSITIVECOMPONENTS.

d) any adverse effect of any recommendedACCESSORYon the ESSENTIAL PERFORMANCEor BASICSAFETYoftheHUMIDIFIERorequipmenttowhichitisconnected.

Check compliance by inspection of the instructions for use and inspection of the RISK MANAGEMENT FILE for any adverse effect of any recommended ACCESSORY.

201.7.9.3.1.101 Additional general requirements

Thetechnicaldescriptionshalldisclose

a) theinterdependenceofcontrolfunctions,and

b) astatementtotheeffectthattheresponsibleorganizationshouldensurethecompatibilityofthehumidifierandallofthepartsandaccessoriesusedtoconnecttothepatientorotherequipmentbeforeuse.

Check compliance by inspection of the technical description.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 31: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved 19

201.7.9.3.101 Additional requirements for the technical description

a) Thetechnicaldescriptionshalldiscloseadescriptionofamethodforcheckingthefunctionof the ALARM SYSTEM for each of the ALARM CONDITIONS specified in this document, if notperformedautomaticallyduringstart‐up.

b) Thetechnicaldescriptionshalldisclosewhichchecksareperformedautomatically.

Check compliance by inspection of the technical description.

201.8 Protection against electrical HAZARDS form ME EQUIPMENT

IEC60601‐1:2005+AMD1:2012,Clause8applies,exceptasfollows.

Additional subclause:

201.8.3.101 Additional requirements for classification of APPLIED PARTS

TheAPPLIEDPARTSofaHUMIDIFIERanditsACCESSORIESshallbeF‐TYPEAPPLIEDPARTS.

Check compliance by inspection.

201.8.7.4.7 Measurement of the PATIENT LEAKAGE CURRENT

Replacement:

e) Assemble the HUMIDIFIER to the BREATHING TUBE and other necessary ACCESSORIES. Wrap the metal foil around the PATIENT-CONNECTION PORT as mentioned under 8.7.4.6. The metal foil is considered as the only PATIENT CONNECTION for the APPLIED PART concerned.

201.9 Protection against mechanical hazards of ME EQUIPMENT and ME SYSTEMS

IEC60601‐1:2005+AMD1:2012,Clause9applies,exceptasfollows.

Additional subclauses:

201.9.4.3.101 * Additional requirements for instability from unwanted lateral movement

a) ATRANSIT‐OPERABLEHUMIDIFIERintendedforuseineithertheHOMEHEALTHCAREENVIRONMENTorEMERGENCYMEDICALSERVICESENVIRONMENTshallincludeameansbywhichtheHUMIDIFIERcan be easily attachedwithout the use of a TOOL to prevent unwantedmovement duringtransportwhileinuse.

b) Themeans shall hold the HUMIDIFIER to withstand accelerations or decelerations of 1,0glongitudinal(forward,backward)and1,0gtransverse(left,right)foratleast5seach.

c) Nomoreliquidthanisspecifiedin201.13.1.101shallexittheHUMIDIFICATIONCHAMBERoutletfromtheseaccelerationsordecelerations.

EXAMPLE Means to be physically restrained during transport in a personal vehicle, in anambulanceoronawheelchair.

Check compliance by functional testing and the testing of 201.13.1.101.

201.9.6.2.1.101 Additional requirements for audible acoustic energy

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 32: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

20 ©ISO2017–Allrightsreserved

The A‐weighted sound pressure level emitted by the HUMIDIFIER shall be less than 50dB asdeterminedbythetestmethodofthisdocument.

Check compliance with the following test.

a) Place the HUMIDIFIER on the sound-reflecting plane, fill the HUMIDIFICATION CHAMBER to the least favourable level and attach the least favourable set of ACCESSORIES from those indicated in the instructions for use.

— Connect a ventilator or other appropriate flow source to the input of the HUMIDIFIER.

— Acoustically isolate the test lung, if used, and flow source by a suitable means so that any noise caused by the test lung and flow source does not interfere with the sound measurement of the HUMIDIFIER.

b) If the flow source is a ventilator,

— set the ventilator to volume control mode that generates ventilation as indicated in Table 201.102,

— configure the test lung with the compliance and resistance components whose values are indicated in Table 201.102, and

— connect the PATIENT-CONNECTION PORT to the test lung.

c) Otherwise, configure the flow source to worst case flow.

d) Using a microphone of the sound level meter complying with the requirements of class 1 specified in IEC 61672-1:2013 with frequency-weighting characteristic A and the time-weighting characteristic F as specified in ISO 3744:2010, measure the sound pressure levels in accordance with ISO 3744:2010, 8.2.1, at 10 positions in a hemisphere with a radius of 1 m from the geometric centre of the HUMIDIFIER in a free field over a reflecting plane as specified in ISO 3744:2010, 7.2.3 and 8.1.1. ISO 3744:2010, Annex F, shall not be utilized.

e) Confirm that the A-weighted background level of extraneous noise is at least 6 dB below that measured during the test.

f) Calculate the A-weighted sound pressure level averaged over the measurement surface according to ISO 3744:2010, 8.2.4.

g) Confirm that the measured sound pressure level is less than 50 dB.

201.9.6.2.1.102 Additional requirements for audible acoustic energy for use with an incubator

AHUMIDIFIERwiththeBREATHINGTUBEandothernecessaryACCESSORIESintendedforusewithanincubator shall comply with the sound pressure level requirements of IEC60601‐2‐19:2009,201.9.6.2.1.101.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 33: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved 21

Check compliance by application of the tests of IEC 60601-2-19:2009, 201.9.6.2.1.101.

Table 201.102 — Test conditions for acoustic tests

Adjustable parameter

Test condition

For a HUMIDIFIER intended to provide an inspired volume

Vinsp≥300ml 300ml≥Vinsp≥50ml Vinsp≤50ml

Inspiredvolume,Vinspa 500ml 150ml 30ml

Ventilatoryfrequency,f 10min−1 20min−1 30min−1

I:Eratio 1/2 1/2 1/2

PEEPb 5hPa 5hPa 5hPa

Resistance,Rc[21][22][23] 5hPa(l/s)−1±10% 20hPa(l/s)−1±10% 50hPa(l/s)−1±10%

IsothermalCompliance,Cc 50mlhPa−1±5% 20mlhPa−1±5% 1mlhPa−1±5%

aVinspismeasuredbymeansofapressuresensoronthetestlung,whereVT=C×Pmax.bIfthePEEPisadjustable.cTheaccuracyforCandRappliesovertherangesofthemeasuredparameters.

201.10 Protection against unwanted and excessive radiation HAZARDS

IEC60601‐1:2005+AMD1:2012,Clause10applies.

201.11 Protection against excessive temperatures and other HAZARDS

IEC60601‐1:2005+AMD1:2012,Clause11applies,exceptasfollows.

201.11.1.2.2 * APPLIED PARTS not intended to supply heat to a PATIENT

Amendment:

Notwithstanding the requirements of IEC60601‐1:2005+AMD1:2012, 11.1.2.2, the allowablemaximumtemperatureof theACCESSIBLEPARTsurfacesofBREATHINGTUBESwithin25cmof thePATIENT‐CONNECTIONPORTshallnotexceed44°C.

201.11.6.2 * Overflow in ME EQUIPMENT

Replacement:

a) LiquidoverflowingfromtheLIQUIDCONTAINERorLIQUIDRESERVOIRshall

1) notwetanyMEANSOFPROTECTIONthatisliabletobeadverselyaffectedbyliquid,nor

2) resultinthelossofBASICSAFETYorESSENTIALPERFORMANCE.

b) No HAZARDOUS SITUATION (as specified in IEC60601‐1:2005+AMD1:2012,13.1 or201.13.1.101)orunacceptableRISKduetooverflowshallbedeveloped:

1) iftheLIQUIDCONTAINERorLIQUIDRESERVOIRisfilledtoitsmaximumcapacity;

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 34: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

22 ©ISO2017–Allrightsreserved

2) foraPORTABLEHUMIDIFIER(e.g.tabletop),ifitistiltedthroughanangleof10°fromanyposition of NORMAL USE when operated under NORMAL CONDITION at the maximumflowrateofNORMALUSE;

3) foraMOBILEHUMIDIFIER(e.g.pole‐mounted),ifitis

i) tiltedthroughanangleof20°fromanypositionofNORMALUSE,and

ii) moved over a threshold as described in IEC60601‐1:2005+AMD1:2012, 9.4.2.4.3,and

4) foranACTIVEHME,intheleastfavourableorientation;

whenoperatedunderNORMALCONDITIONatthemaximumflowrateofNORMALUSE.

Check compliance by the following:

c) Fill the LIQUID CONTAINER and LIQUID RESERVOIR to the indicated maximum level. Operate the HUMIDIFIER at its maximum RATED flowrate.

d) PORTABLE ME EQUIPMENT is subsequently tilted through an angle of 10° in the least favourable direction(s) (if necessary with refilling) starting from the position of NORMAL USE.

e) MOBILE ME EQUIPMENT is subsequently tilted through an angle of 20° in the least favourable direction(s) (if necessary with refilling) starting from the position of NORMAL USE and is moved over a threshold as described inIEC 60601-1:2005+AMD1:2012, 9.4.2.4.3.

f) Return the HUMIDIFIER to normal orientation and subsequently refill the LIQUID CONTAINER to the maximum level then add a further quantity equal to 15 % of the capacity of the LIQUID CONTAINER, poured in steadily over a period of 1 min.

g) After these PROCEDURES, the HUMIDIFIER is to pass the appropriate dielectric strength and LEAKAGE CURRENT tests and is to show no signs of wetting of uninsulated electrical parts or electrical insulation of parts that could result in the loss of BASIC SAFETY or ESSENTIAL PERFORMANCE in NORMAL CONDITION or in combination with a SINGLE FAULT CONDITION (based on a visual inspection).

h) In addition, confirm that there is no more liquid than is specified in 201.13.1.101 exits the HUMIDIFICATION CHAMBER outlet.

201.11.6.6 * Cleaning and disinfection of ME EQUIPMENT or ME SYSTEM

Amendment (add additional requirement as new first paragraph):

aa)GAS PATHWAYS through the HUMIDIFIER and its ACCESSORIES that can become contaminatedwithbodyfluidsorexpiredgasesduringNORMALCONDITIONorSINGLEFAULTCONDITIONthatarenotsingleuse

1) shall be designed to allow for cleaning and disinfection or cleaning and sterilization(additional requirements are found in IEC60601‐1:2005+AMD1:2012, 11.6.7 andIEC60601‐1‐11:2015,Clause8).

2) Dismantlingmaybeused.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 35: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved 23

Amendment (add additional requirement and replace the compliance test):

bb)HUMIDIFIER ENCLOSURES shall bedesigned to allow for surface cleaning anddisinfection toreduce to acceptable levels the RISK of cross infection of the OPERATOR, other persons or nextPATIENT.

cc)InstructionsforprocessingandreprocessingtheHUMIDIFIERanditsACCESSORIESshall

102) complywithISO17664:2004andISO14937:2009,and

103) bedisclosedintheinstructionsforuse.

NOTE1 ISO14159providesguidanceforthedesignofENCLOSURES.

Check compliance by inspection of the RISK MANAGEMENT FILE. When compliance with this document could be affected by the cleaning or the disinfecting of the HUMIDIFIER or its parts or ACCESSORIES, clean and disinfect them 10 times in accordance with the methods indicated in the instruction for use, including any cooling or drying period. After these PROCEDURES, confirm that BASIC SAFETY and ESSENTIAL PERFORMANCE are maintained. Confirm that the MANUFACTURER has evaluated the effects of multiple PROCESS cycles and the effectiveness of those cycles.

NOTE2 Additionalinformationregardingtheorderoftestisfoundin211.10.1.1.

201.11.6.7 Sterilization of ME EQUIPMENT or ME SYSTEM

Amendment (add note before compliance test):

NOTE Additionalrequirementsare foundinIEC60601–1:2005+AMD1:2012,11.6.6andIEC60601–1‐11:2015,Clause8.

201.11.7 BIOCOMPATIBILITY of ME EQUIPMENT and ME SYSTEMS

Amendment (add after existing text prior to the compliance statement):

aa)TheMANUFACTURERofaHUMIDIFIER,BREATHINGSYSTEM,itspartsandACCESSORIESshalladdressintheRISKMANAGEMENTPROCESStheRISKSassociatedwiththeleachingor leakingofsubstancesintotheGASPATHWAY.

bb)TheGASPATHWAYSshallbeevaluatedforBIOCOMPATIBILITYaccordingtoISO18562‐1:—.

cc)Specialattentionshallbegiventosubstanceswhicharecarcinogenic,mutagenicortoxictoreproduction.

dd)The ACCESSIBLE PARTS and GAS PATHWAYS of a HUMIDIFIER, BREATHING SYSTEM, its parts orACCESSORIES that containphthalateswhichare classifiedas carcinogenic,mutagenicor toxic toreproductionshallbemarkedascontainingphthalatesonthedeviceitselforonthepackagingthatitcontainsphthalates.

ee)ThesymbolofEN15986:2011(Table201.D.2.101,symbol5)maybeused.

ff)If the INTENDED USE of a HUMIDIFIER, BREATHING SYSTEM, its parts or ACCESSORIES includetreatmentofchildrenortreatmentofpregnantornursingwomen,aspecificjustificationfortheuseofthesephthalatesshallbeincludedintheRISKMANAGEMENTFILE.

gg)The instructions for use of a HUMIDIFIER, BREATHING SYSTEM, its parts or ACCESSORIES thatcontainsuchphthalatesshallcontain

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 36: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

24 ©ISO2017–Allrightsreserved

100)informationonRESIDUALRISKSforthesePATIENTgroups,and

101)ifapplicable,onappropriateprecautionarymeasures.

201.12 Accuracy of controls and instruments and protection against hazardous outputs

IEC60601‐1:2005+AMD1:2012,Clause12applies,exceptasfollows.

201.12.1 Accuracy of controls and instruments

Amendment (add after existing sentence):

aa)The HUMIDIFIERmay providemeans to reduce the visibility of its controls and indicatorseitherautomaticallyorbytheOPERATORaction.

bb)If provided, the HUMIDIFIER shall automatically resume normal visibility during an ALARMCONDITION.

Check compliance by functional testing and application of the tests of IEC 60601-1:2005+AMD1:2012, 7.1.2.

Additional subclauses:

201.12.1.101 * HUMIDIFICATION OUTPUT

a) Overtherangeof flowrates,settings,ambienttemperature,andgas inlet temperatureandhumidityofNORMALUSE,theHUMIDIFICATIONOUTPUTatthePATIENT‐CONNECTIONPORTshallnotbelessthan:

1) 33mg/lforacategory1HUMIDIFIERoperatinginamodeintendedforaPATIENTwhoseupperairwayshavebeenbypassed(invasiveventilation).

b) Overtherangeofflowrates,ambienttemperature,andgasinlettemperatureandhumidityofNORMALUSE,theHUMIDIFIERshallbecapableofdeliveringaHUMIDIFICATIONOUTPUTatthePATIENT‐CONNECTIONPORTofatleast:

2) 12mg/lforacategory1HUMIDIFIERoperatinginaamodeintendedforaPATIENTwhoseupperairwayshavenotbeenbypassed(non‐invasiveventilation);

3) 12mg/lforacategory2HUMIDIFIER.

c) TheHUMIDIFICATIONOUTPUTshalleitherbe

4) determined for each BREATHING SYSTEM configuration indicated in the instructions foruse,or

5) determined for the worst‐case BREATHING SYSTEM configurations indicated in theinstructionsforuse.

NOTE The worst‐case BREATHING SYSTEM configuration can be different for differentflowratesandHUMIDIFICATIONOUTPUT.

d) Ifworst‐caseBREATHINGSYSTEMconfigurationsareused,therationalefortheirselectionshallbedocumentedintheRISKMANAGEMENTFILE.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 37: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved 25

e) TheHUMIDIFICATIONOUTPUT(inmg/l)overtheRATEDrangeofgasflowratesandsettingsshallbedisclosedintheinstructionsforuse.

Check compliance by inspection of the instructions for use and RISK MANAGEMENT FILE for the rationale, if applicable, and with the tests of Annex C.

201.12.1.102 * SET TEMPERATURE accuracy and MONITORING EQUIPMENT

a) AHUMIDIFIERshallbeequippedwithanALARMSYSTEMthatincludes:

1) atleastaMEDIUMPRIORITYALARMCONDITIONforacategory1HUMIDIFIER;or

2) ifequippedwithSETTEMPERATUREMONITORINGEQUIPMENT,atleastaLOWPRIORITYALARMCONDITIONforacategory2HUMIDIFIER;

toindicatethattheMEASUREDGASTEMPERATURE,whenaveragedovera5minperiod,differsbymorethan±2°CfromtheSETTEMPERATUREduringNORMALUSE.

b) These ALARM CONDITIONS need not be activated during the start‐up period or during thetransitiontoanewstateofthermalequilibriumfollowingachangeingasflowrateorchangeinSETTEMPERATURE.

NOTE The requirements for thermal overshoot of 201.12.4.101 applyduring theseperiodsoftransition.

c) The maximum start‐up period in NORMAL USE, the warm‐up time for the MEASURED GASTEMPERATUREtoreachtheSETTEMPERATUREfromastartingtemperatureof(23±2)°C,shallbedisclosedintheinstructionsforuse.

Check compliance by functional testing.

201.12.1.103 * MEASURED GAS TEMPERATURE MONITORING EQUIPMENT

a) TheHUMIDIFIERmaybeequippedwithMEASUREDGASTEMPERATUREMONITORINGEQUIPMENTthatdisplaysthetemperature.

b) Ifequipped,theMEASUREDGASTEMPERATUREMONITORINGEQUIPMENTshall

1) haveaRATEDrangeofatleast25°Cto45°C,and

2) beaccurateto±2°CovertheRATEDrange.

c) TheaccuracyoftheMEASUREDGASTEMPERATUREMONITORINGEQUIPMENTshallbedisclosedintheinstructionsforuse.

Check compliance by inspection of the instructions for use, functional testing and with the tests of Annex BB.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 38: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

26 ©ISO2017–Allrightsreserved

201.12.4 Protection against hazardous output

Additional subclause:

201.12.4.101 * Thermal overshoot

In NORMAL USE and SINGLE FAULT CONDITIONS and over the RATED flowrate range and at themaximumRATEDoperatingtemperature,theDELIVEREDGASTEMPERATUREoftheHUMIDIFIERshallnotexceedanenergyequivalentto43°Cand100%RELATIVEHUMIDITY(aspecificenthalpynottoexceed197kJ/m3dryair)whenaveragedover120s.

Table201.103containsexamplesofcombinationsoftemperatureandRELATIVEHUMIDITYforairwithsuchaspecificenthalpy.

NOTE HUMIDIFIERSarecommonlyusedwithairandoxygenmixtures.Thethermalovershootlimitof197kJ/m3ofdrygaswhenaveragedover120salsoappliestoothergasmixtures,suchashelium‐oxygenmixtures,where the instructions foruse includesusewithothergasmixtures.Thetemperature limit islikelydifferentforothergasmixturesandwillneedtobecalculated.

Check compliance with the following test.

a) Assemble the HUMIDIFIER, BREATHING TUBES and ACCESSORIES as indicated in the instructions for use.

b) Fill the HUMIDIFICATION CHAMBER to its maximum level prior to each test. Operate the HUMIDIFIER at its minimum RATED flowrate and wait 30 min.

c) Quickly adjust the flowrate to the maximum RATED flowrate and with a sampling period no greater than 2 s, measure the DELIVERED GAS TEMPERATURE for the next 240 s.

Table 201.103 — Examples of permissible combinations of temperature and RELATIVE HUMIDITY in air

Temperature °C

RELATIVE HUMIDITY%

43 100

44 95

45 90

48 76

50 71

d) Calculate the specific enthalpy using the method of Annex DD.

e) Confirm that the specific enthalpy averaged over 120 s does not exceed 197 kJ/m3 during the 240 s measurement period.

f) Repeat b) to e), starting at the maximum RATED flowrate and quickly adjusting to the minimum RATED flowrate.

g) Operate the HUMIDIFIER at its minimum RATED flowrate and maximum RATED SET TEMPERATURE and wait 30 min.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 39: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved 27

h) Turn off the gas flow for 3 min.

i) Return the gas flow to the minimum RATED flowrate and with a sampling period no greater than 2 s, measure the DELIVERED GAS TEMPERATURE and AIRWAY PRESSURE for the next 240 s.

j) Calculate the specific enthalpy using the method of Annex DD.

k) Confirm that the specific enthalpy averaged over 120 s does not exceed 197 kJ/m3 during the 240 s measurement period.

l) Repeat g) to k), starting at the average RATED flowrate.

m) Repeat g) to k), starting at the maximum RATED flowrate.

n) Operate the HUMIDIFIER with no flow and wait 30 min.

o) Quickly adjust the flowrate to the maximum RATED flowrate and with a sampling period no greater than 2 s, measure the DELIVERED GAS TEMPERATURE and AIRWAY PRESSURE for the next 240 s.

NOTE Equipment modifications can be necessary for a HUMIDIFIER that is activated by its integrated flow generator.

p) Calculate the specific enthalpy using the method of Annex DD.

q) Confirm that the specific enthalpy does not exceed 197 kJ/m3 when averaged over 120 s during the 240 s measurement period.

201.13 HAZARDOUS SITUATIONS and fault conditions for ME EQUIPMENT

IEC60601‐1:2005+AMD1:2012,Clause13applies,exceptasfollows.

Additional subclauses:

201.13.1.101 * Additional specific HAZARDOUS SITUATIONS

InNORMAL CONDITIONand SINGLE FAULT CONDITION, aHUMIDIFIER shall be so constructed that thefollowingHAZARDOUSSITUATIONSshallnotoccur:

a) thevolumeofliquidexitingtheHUMIDIFICATIONCHAMBERoutletshallnotexceed:

1) 1,0mlin1minor2,0mlin1hwhenintendedforusewithPATIENTSweighinglessthan5kg;

2) 5mlin1minor20mlin1hforallotherPATIENTS.

Check compliance with functional testing.

201.13.2.101 Additional specific SINGLE FAULT CONDITIONS

AHUMIDIFIERshallbesoconstructedthatthefollowingSINGLEFAULTCONDITIONSshallnotcauseanunacceptableRISK:

a) operationoftheHUMIDIFIERwithoutanyliquid;

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 40: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

28 ©ISO2017–Allrightsreserved

b) if theHUMIDIFIER includesasensororsensors thatareresponsible for theconditionof thegasdeliveredtothePATIENT,anyfailureofasensororthesensingsystem.

EXAMPLE1 Sensorsingleopen‐circuit.

EXAMPLE2 Sensorsingleshort‐circuit.

EXAMPLE3 SensordisconnectedfromtheHUMIDIFIERcontrolsystem.

EXAMPLE4 SensordisconnectedfromBREATHINGTUBEorHUMIDIFIER.

Check compliance by functional testing and inspection of RISK MANAGEMENT FILE.

201.13.102 * Independence of HUMIDIFICATION control function and related RISK CONTROL measures

a) A SINGLE FAULT CONDITION shall not cause the HUMIDIFIER‐control function and thecorrespondingPROTECTIONDEVICEtofailsimultaneously.

b) ASINGLEFAULTCONDITIONshallnotcause

1) theHUMIDIFIER‐controlfunctionandthecorrespondingMONITORINGEQUIPMENT,or

2) theHUMIDIFIER‐controlfunctionandthecorrespondingALARMSYSTEM

tofailinsuchawaythatthelossoftheHUMIDIFIER‐controlfunctionisnotdetected.

Check compliance by inspection and functional testing.

201.14 Programmable electrical medical systems (pems)

IEC60601‐1:2005+AMD1:2012,Clause14applies,exceptasfollows.

201.14.1 General

Amendment (extend the last paragraph prior to the compliance check with):

aa)ThehumidityandtemperaturecontrolPESSoftheHUMIDIFIERPEMSwithoutanindependenthardwareRISKCONTROLmeasureshallbeconsideredas

100) for a category 1 HUMIDIFIER, software safety ClassC as defined inIEC62304:2006+AMD1:2015.

101) for a category 2 HUMIDIFIER, at least software safety ClassB as defined inIEC62304:2006+AMD1:2015.

bb)ThesoftwaresafetyclassforaCategory2HUMIDIFIERshallnotbereducedfromClassBtoClassAwithanindependenthardwareRISKCONTROLmeasure.

201.15 Construction of ME EQUIPMENT

IEC60601‐1:2005+AMD1:2012,Clause15applies,exceptasfollows.

Additional subclause:

201.15.101 Mode of operation

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 41: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved 29

AHUMIDIFIERshallbesuitableforCONTINUOUSOPERATION.

Check compliance by inspection.

201.16 ME SYSTEMS

IEC60601‐1:2005+AMD1:2012,Clause16applies,exceptasfollows.

Additional subclause:

201.16.1.101 Additional general requirements for ME SYSTEMS

ACCESSORIES connected to the HUMIDIFIER shall be considered to form an MESYSTEM with theHUMIDIFIER.

Check compliance by application of the relevant tests of IEC 60601-1:2005+AMD1:2012.

201.17 Electromagnetic compatibility of ME EQUIPMENT and ME SYSTEMS

IEC60601‐1:2005+AMD1:2012,Clause17applies.

Additional clauses:

201.101 BREATHING SYSTEM connectors and ports

201.101.1 * General

a) IfaHUMIDIFIERisintendedtobeplacedinaBREATHINGSYSTEM,anyconicalconnectorshall

1) complywithISO5356‐1:2015,or

2) not engage with those connectors or with connectors complying withISO80369‐1:2010.

b) Anon‐conicalconnectorshall

1) not engage with a conical connector complying with ISO5356‐1:2015, unless theycomply with the engagement, disengagement and leakage requirements of thatstandard.

Check compliance by application of the tests of ISO 5356-1:2015 and functional testing.

201.101.2 PATIENT-CONNECTION PORT

Ifequipped,thePATIENT‐CONNECTIONPORTshallbeoneofthefollowing:

a) afemale15mmconicalconnectorcomplyingwithISO5356‐1:2015;

b) acoaxial15mm/22mmconicalconnectorcomplyingwithISO5356‐1:2015.

Check compliance by application of the tests of ISO 5356-1:2015.

201.101.3 FLOW-DIRECTION-SENSITIVE components

Any OPERATOR‐detachable FLOW‐DIRECTION‐SENSITIVE COMPONENT of the HUMIDIFIER shall be sodesigned that it cannot be fitted in such a way that it presents an unacceptable RISK to thePATIENT.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 42: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

30 ©ISO2017–Allrightsreserved

Check compliance by inspection of OPERATOR-detachable FLOW-DIRECTION-SENSITIVE COMPONENTS and inspection of the RISK MANAGEMENT FILE.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 43: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved 31

201.101.4 * ACCESSORY port

Ifprovided, eachACCESSORYportof theHUMIDIFIER, BREATHING SYSTEM, itspartsandACCESSORIESshall

a) complywithISO80369‐1:2010,and

b) beprovidedwithameanstosecuretheACCESSORYinpositionandameanstosecureclosureafterremovaloftheACCESSORY.

NOTE1 ItisexpectedthattheRESP‐125(R1)connectorofISO80369‐2[11]willmeetthiscriterion.

NOTE2 Thisportisgenerallyusedformeasuringpressure,samplingofgasesorforintroductionoftherapeuticaerosols.

Check compliance by inspection and application of the tests of ISO 80369-1:2010.

201.101.5 Monitoring probe port

Ifaportisprovidedforintroductionofamonitoringprobe,it:

a) shallnotbecompatiblewithconnectorsspecifiedinISO5356‐1:2015;

b) shall be provided with a means to secure the probe in position and a means to secureclosureafterremovaloftheprobe.

Check compliance by inspection and application of the tests of ISO 5356-1:2015.

201.101.6 Oxygen inlet port

a) AnoxygeninletconnectoroftheHUMIDIFIER,BREATHINGSYSTEM,itspartsandACCESSORIESthatisOPERATOR‐accessiblewithouttheuseofaTOOLshallcomplywithISO80369‐1:2010.

b) A HUMIDIFIER with this inlet connector shall maintain BASIC SAFETY and ESSENTIALPERFORMANCEwithoxygensupplysystemsupto600kPa,inNORMALCONDITION.

NOTE It is expected that the RESP‐6000 (R2) connector of ISO80369‐2[11] will meet thiscriterion.

Check compliance by functional testing and application of the tests of ISO 80369-1:2010.

201.101.7 Other connectors

If the HUMIDIFIER, BREATHING SYSTEM, its parts or ACCESSORIES is fitted with any other type ofconnector,theseconnectors:

a) shallmatewithBREATHINGTUBESthatcomplywithISO5367:2014;

b) shallnotacceptorpermitconnectionwitheither

1) the15mmor22mmconicalconnectorscomplyingwithISO5356‐1:2015,or

2) theconnectorscomplyingwithISO80369‐1:2010.

Check compliance by inspection and application of the tests of ISO 5367:2014.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 44: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

32 ©ISO2017–Allrightsreserved

201.101.8 Removable temperature sensors and ports

201.101.8.1 Security

WhenthesensorsormatingportsareengagedinNORMALUSE,theconnectionshallnotbecomedisconnectedundertheconditionsof

a) noflow,or

b) maximumRATEDflowrate.

Check compliance by functional testing under the conditions of no flow or maximum RATED flowrate.

201.101.8.2 * Leakage

Theleakagefromanengagedsensorormatingportshallnotexceed5ml/minatapressureof60cmH2O.

Check compliance by functional testing.

201.101.8.3 Construction

Removablesensorsandportsshall

a) meetthedimensionalrequirementsofAnnexEE,or

b) besufficientlydifferentthattheycannotbeinterchangedwiththosethatdo.

Check compliance by inspection and functional testing or by application of the tests of Annex EE.

201.101.9 Other orifices

IftheHUMIDIFIER,BREATHINGSYSTEM,itspartsandACCESSORIESincorporateanindependentfillingorACCESSORYorifice(e.g.anairentrainmentoraheaterorifice),thatorificeshallnotacceptanyof

a) theconnectorsspecifiedinISO5356‐1:2015,or

b) theconnectorscomplyingwithISO80369‐1:2010.

Check compliance by inspection.

201.102 Requirements for the BREATHING SYSTEM and ACCESSORIES

201.102.1 * General

All BREATHING SYSTEMS, their parts and ACCESSORIES shall complywith the requirements of thisdocument, whether they are produced by the MANUFACTURER of the HUMIDIFIER or by anotherentity(“third‐partymanufacturer”orhealthcareprovider).

Check compliance by the tests of this document.

201.102.2 Labelling

a) TheMODELORTYPEREFERENCEofatleastonecompatibleHUMIDIFIERshallbedisclosedintheACCOMPANYING DOCUMENT, or packaging label provided with each BREATHING SYSTEM orACCESSORY,compliantwith201.102.1.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 45: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved 33

b) Statements shall be included in the ACCOMPANYING DOCUMENT or packaging label of eachBREATHINGSYSTEM,partorACCESSORYtotheeffectthat:

1) breathing systems, their parts and accessories are validated for use with specifichumidifiers;

2) incompatiblepartscanresultindegradedperformancewhichcanaffectsafety;

3) theresponsibleorganizationisaccountableforthecompatibilityofthehumidifierandall of the parts and accessories used to connect to the patient and other equipmentbeforeuse.

Check compliance by inspection of the ACCOMPANYING DOCUMENT.

201.102.3 * BREATHING TUBES

201.102.3.1 Non-heated BREATHING TUBES

BREATHINGTUBES,other thanheatedBREATHINGTUBES, intended foruse in theBREATHINGSYSTEMshallcomplywithISO5367:2014atthemaximumHUMIDIFICATIONOUTPUToftheHUMIDIFIER.

Check compliance by application of the tests of ISO 5367:2014 while connected to the specified HUMIDIFIER operated at its maximum RATED output.

201.102.3.2 Heated BREATHING TUBES

HeatedBREATHINGTUBESintendedforuseintheBREATHINGSYSTEMshallnotcollapseonbending,occludeorotherwisecause lossofBASIC SAFETYorESSENTIALPERFORMANCEwhen theBREATHINGTUBESaresubjecttothemaximumRATEDoutputpowerofthespecifiedHEATEDBREATHINGTUBECONTROLLER,includingunderconditionsofnoflow.

Check compliance by application of the tests of ISO 5367:2014, Annex E and Annex G, while connecting the HEATED BREATHING TUBES to the specified HEATED BREATHING TUBE CONTROLLER operated at its maximum RATED output at both the maximum flowrate and no flow conditions.

201.103 LIQUID CONTAINER

201.103.1 Liquid level

For category1HUMIDIFIERS,means shall beprovided topermit theOPERATOR todetermine theliquidlevel:

a) intheLIQUIDCONTAINER;

b) ifprovided,theLIQUIDRESERVOIR;

withoutdismantlingtheHUMIDIFIER.

Check compliance by inspection.

201.103.2 Filling cap

Reusablefillingcaps,ifprovided,shallbetetheredtopartoftheHUMIDIFIER.

Check compliance by inspection.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 46: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

34 ©ISO2017–Allrightsreserved

201.104 FUNCTIONAL CONNECTION

201.104.1 * General

BASICSAFETYandESSENTIALPERFORMANCEoftheHUMIDIFIERshallbemaintainedif

a) connections to the FUNCTIONAL CONNECTION of a HUMIDIFIER, including the heated BREATHINGTUBEcontroller,aredisrupted,or

b) theequipmentconnectedtothosepartsfails.

Check compliance by functional testing.

201.104.2 * Connection to an electronic health record

A Category 1 HUMIDIFIER should be equippedwith a FUNCTIONAL CONNECTION that permits datatransmissionfromtheHUMIDIFIERto,forexample,anelectronichealthrecord.

201.104.3 * Connection to a DISTRIBUTED ALARM SYSTEM

A Category 1 HUMIDIFIER should be equipped with a FUNCTIONAL CONNECTION that permitsconnectiontoaDISTRIBUTEDALARMSYSTEM.

201.104.4 Connection for remote control

A HUMIDIFIER may be equipped with a FUNCTIONAL CONNECTION for external control of theHUMIDIFIER.

202 Electromagnetic disturbances — Requirements and tests

IEC60601‐1‐2:2014appliesexceptasfollows.

Addition:

202.4.3.1 Configurations

IEC60601‐1‐2:2014appliesexceptasfollows.

Amendment (add after the last dash of 4.3.1):

aa)attachmentofthebreathingtubestothehumidifierorheatedbreathingtubecontroller;

bb)if applicable, attachment of ACCESSORIES as necessary to achieve the BASIC SAFETY andESSENTIALPERFORMANCEoftheHUMIDIFIER.

202.5.2.2.1 Requirements applicable to all ME EQUIPMENT and ME SYSTEMS

Amendment [add note to list element b)]:

NOTE Therequirementsofthisdocumentarenotconsidereddeviationsorallowances.

Addition:

202.8.1.101 * Additional general requirements

a) Thefollowingdegradations,ifaffectingBASICSAFETY,shallnotbeallowed:

1) componentfailures;

2) changesinprogrammableparametersorsettings;

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 47: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved 35

3) resettodefaultsettings;

4) changeofoperatingmode;

5) thespecificenthalpyattheHUMIDIFIERoutputaveragedover120sexceeding197kJ/m3.

NOTE1 TableAA.1 indicates the conversion from the values of dew point to the units of mg/l(referenceBTPS) forusewhenusingadewpointhygrometer tomonitor theHUMIDIFICATIONOUTPUTwhendeterminingthespecificenthalpyattheHUMIDIFIERoutput.

NOTE2 MANUFACTURERS are encouraged to considermonitoring a control signal to theheater todetermineifcontrolledheatingisoccurringinlieuofmonitoringspecificenthalpy.

b) TheHUMIDIFIERmayexhibittemporarydegradationofperformance(e.g.deviationfromtheperformance indicated in the instructions for use during IMMUNITY testing) that does notaffectBASICSAFETYorESSENTIALPERFORMANCE.

206 Usability

IEC60601‐1‐6:2010+AMD1:2013appliesexceptasfollows.

ForaHUMIDIFIER,thefollowingshallbeconsideredPRIMARYOPERATINGFUNCTIONS:

a) observingmonitoredhumidificationparameters;

b) fillingtheLIQUIDCONTAINERand,ifprovided,theLIQUIDRESERVOIR;

c) observingthewaterlevelintheLIQUIDCONTAINERand,ifprovided,intheLIQUIDRESERVOIR;

d) configuringtheACCESSORIESincludingconnectionofthedetachablepartstotheHUMIDIFIER;

EXAMPLE1 Heatedbreathingtubecontroller,watertrap,breathingtubes,breathingsystemfilter,monitoringequipment.

e) connectingthePATIENT‐CONNECTIONPORTtothePATIENT‐interface;

f) disconnectingthePATIENT‐CONNECTIONPORTfromthePATIENT‐interface;

g) reprocessingtheACCESSORIES;

h) startingtheHUMIDIFIERfrompoweroff;

i) turningofftheHUMIDIFIER;

j) performingabasicpre‐usefunctionalcheckoftheHUMIDIFIERincludingtheALARMSIGNALS.

Thefollowingfunctions,ifavailable,shallalsobeconsideredPRIMARYOPERATINGFUNCTIONS:

k) connectingtheHUMIDIFERtotheflowsource(e.g.ventilator);

l) disconnectingtheHUMIDIFERfromtheflowsource(e.g.ventilator);

m)settingtheOPERATOR‐adjustablecontrols,including:

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 48: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

36 ©ISO2017–Allrightsreserved

1) switchingbetweendifferenthumidificationmodes;

2) settinghumidificationcontrolparameters;

3) settingALARMLIMITS;

4) inactivatingALARMSIGNALS;

n) startinghumidificationfromstandby;

o) activatingstandby.

ThefollowingactionsassociatedwithhumidificationalsoshallbeconsideredPRIMARYOPERATINGFUNCTIONS:

NOTE For the purposes of this document, the following functions are considered PRIMARY OPERATINGFUNCTIONSeventhoughtheyarenotperformedontheHUMIDIFIER’SOPERATOR‐EQUIPMENTINTERFACE.

p) addingmedicationtothegasflowingintothePATIENT;

EXAMPLE2 InjectingfluidsintotheancillaryportconnectionoftheBREATHINGSYSTEM.

q) for a TRANSIT‐OPERABLE equipment, positioning the PATIENT and the equipment on awheelchairortrolley.

206.5 Replacement of requirements given in IEC 62366[17]

Amendment (add at the end of the last paragraph of Clause 5):

ThesummaryoftheapplicationspecificationshallincludeHUMIDIFIERclassificationasspecifiedin201.6.101.

NOTE The “application specification” of IEC62366[17] is called the USE SPECIFICATION in IEC62366–1:2015[18].

208 General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems

IEC60601‐1‐8:2006+AMD1:2012appliesexceptasfollows.

Additional subclauses:

208.6.8.4.101 * Additional requirements for termination of ALARM SIGNAL inactivation

For category1 HUMIDIFIERS, the duration of AUDIO PAUSED or ALARM PAUSED for the ALARMCONDITIONSrequiredbythisdocumentshallnotexceed120swithoutOPERATORintervention.

NOTE ThispermitsanOPERATORtodeliberatelyextendthedurationofAUDIOPAUSEDbydirectaction.

Check compliance by functional testing.

211 Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment

IEC60601‐1‐11:2015appliesexceptasfollows.

211.10.1.1 General requirements for mechanical strength

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 49: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved 37

Amendment (add the before first paragraph):

a) The tests of IEC60601‐1‐11:2015, Clause10 and of IEC60601‐1:2005+AMD1:2012, 15.3shallbeperformedonthesamesampleoftheHUMIDIFIERafterthecleaninganddisinfectionPROCEDURES of 201.11.6.6 of this document have been performed unless there are nocleaninganddisinfectionPROCEDURESspecifiedintheinstructionsforuse.

b) IfmorethanonePROCEDUREisspecifiedintheinstructionsforuse,eachPROCEDUREshallbesotested.

c) AseparatesampleoftheHUMIDIFIERmaybeusedforeachspecifiedPROCEDURE.

IEC60601‐1:2005+AMD1:2012,annexesapply,exceptasfollows.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 50: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

38 ©ISO2017–Allrightsreserved

Annex C (informative)

Guide to marking and labelling requirements for ME EQUIPMENT and

ME SYSTEMS

IEC60601‐1:2005+AMD1:2012,AnnexCofthegeneralstandardapplies,exceptasfollows.

201.C.1 Marking on the outside of ME EQUIPMENT, ME SYSTEMS or their parts

Addition:

201.C.1.101 Marking on the outside of a HUMIDIFIER or its parts

Additional requirements formarking on the outside of a HUMIDIFIER or its parts are found inTable201.C.101.

Table 201.C.101 — Marking on the outside of a HUMIDIFIER or its parts

Description of marking Subclause

Anyparticularstorageand/orhandlinginstructions,ifapplicable 201.7.2.101b)

Anyparticularwarningsand/orprecautionsrelevanttotheimmediateoperationoftheHUMIDIFIER,ifapplicable

201.7.2.101c)

Arrow indicating the direction of the flow for FLOW‐DIRECTION‐SENSITIVE COMPONENTS, ifapplicable

201.7.2.101d)

Containingnaturalrubberlatex,ifapplicable 201.7.2.13.101

ForACCESSORIES suppliedseparately, indicationofany limitationsoradverseeffectsofthe ACCESSORY on the BASIC SAFETY or ESSENTIAL PERFORMANCE of the HUMIDIFIER, ifapplicable

201.7.2.4.101

For ACCESSORIES supplied separately, the requirements of 201.7.2.101, 201.7.2.13.101and201.7.2.17.101

201.7.2.4.101

ForeachHUMIDIFIER,partandACCESSORY,containsphthalates,ifapplicable 201.11.7

Forpackaging,containingnaturalrubberlatex,ifapplicable 201.7.2.17.101c)

Forpackaging,descriptionofthecontents 201.7.2.17.101a)

Forpackaging,identificationreferencetothebatch,typeorserialnumber 201.7.2.17.101b)

Maximumandminimumliquidlevels 201.7.2.101a)

Pressureatwhichthepressure‐reliefPROTECTIONDEVICEopens,ifprovided 201.7.2.101e)

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 51: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved 39

201.C.4 ACCOMPANYING DOCUMENTS, general

Addition:

201.C.4.101 ACCOMPANYING DOCUMENTS, general, of a HUMIDIFIER

Additional requirements for ACCOMPANYING DOCUMENTS, general, of a HUMIDIFIER are found inTable201.C.102.

Table 201.C.102 — ACCOMPANYING DOCUMENTS, general, of a HUMIDIFIER

Description of requirement Subclause

For each BREATHING SYSTEM and ACCESSORY, theMODEL OR TYPE REFERENCE of at leastonecompatibleHUMIDIFIER

201.102.2

For each BREATHING SYSTEM, part and ACCESSORY, a statement to the effect thatbreathing systems, their parts andaccessories are validated for usewith specifichumidifiers

201.102.2a)

For each BREATHING SYSTEM, part and ACCESSORY, a statement to the effect thatincompatiblepartscanresultindegradedperformance

201.102.2b)

For each BREATHING SYSTEM, part and ACCESSORY, a statement to the effect that theresponsibleorganizationisaccountableforthecompatibilityofthehumidifierandallofthepartsandaccessoriesusedtoconnecttothepatientbeforeuse

201.102.2c)

HUMIDIFIERisahighflowdevicewarning,ifapplicable 201.4.11.101.23)iii)

Maximumtime‐weightedaverageinputflowrateforeachgas,ifapplicable 201.4.11.101.23)i)

Maximumtransientinputflowrateforeachgas,ifapplicable 201.4.11.101.23)ii)

NameortradenameandaddressoftheMANUFACTURERandwheretheMANUFACTURERdoesnothaveanaddresswithinthelocaleanauthorizedrepresentative

201.7.9.1

Units of measure for volumes, flows and leakages expressed as STPD or BTPS, asappropriate

201.7.4.3

201.C.5 ACCOMPANYING DOCUMENTS, instructions for use

Addition:

201.C.5.101 ACCOMPANYING DOCUMENTS, instructions for use of a HUMIDIFIER

Additional requirements for ACCOMPANYINGDOCUMENTS, instructions foruseof a HUMIDIFIER arefoundinTable201.C.103.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 52: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

40 ©ISO2017–Allrightsreserved

Table 201.C.103 — ACCOMPANYING DOCUMENTS, instructions for use of a HUMIDIFIER

Description of requirement Subclause

AccuracyoftheMEASUREDGASTEMPERATUREMONITORINGEQUIPMENT 201.12.1.103

Any adverse effect of any recommendedACCESSORYon the BASIC SAFETY or ESSENTIALPERFORMANCEoftheHUMIDIFIER,ifapplicable

201.7.9.2.14.101c)

Anynaturalrubberlatex‐containingcomponents,ifapplicable 201.7.2.13.101

Description of the INTERNAL ELECTRICAL POWER SOURCE care and maintenancePROCEDURES,includinginstructionsforrechargingorreplacement,ifapplicable

201.7.9.2.13b)

DescriptionoftheperiodicvisualsafetyinspectionsthatshouldbeperformedbytheOPERATOR

201.7.9.2.13a)

DisclosedapplicationspecificationincludestheHUMIDIFIERclassificationasspecifiedin201.6.101

206.5

Disclosure of any restrictions on the placing of components within the BREATHINGSYSTEM,ifapplicable

201.7.9.2.14.101b)

For a HUMIDIFIER, its parts or ACCESSORIES intended for single‐use, information onknowncharacteristics and technical factors known to theMANUFACTURER that couldposeaRISKiftheHUMIDIFIER,itspartsorACCESSORIESwouldbereused

201.7.9.2.1.102a)

ForaHUMIDIFIER,itspartsorACCESSORIESintendedforsingle‐use,thatitissingleuseandtheintendeddurationofuse

201.7.9.2.1.102b)

ForACCESSORIESsuppliedseparatelywheremarkingtheACCESSORYisnotpracticable,therequirementsof201.7.2.4.101,201.7.2.13.101and201.7.2.17.101,ifnotmarked

201.7.2.4.101

ForeachHUMIDIFIER,partandACCESSORY,informationonRESIDUALRISKSforchildrenortreatment of pregnant or nursing women and, if applicable, on appropriateprecautionarymeasuresfordevicesthatcontainphthalates

201.11.7

For the LAY OPERATOR instructions, an explanation of the meaning of the IPclassification

201.7.9.2.9.101.1b)

FortheLAYOPERATORinstructions,anindicationoftheexpecteddurationofoperationbetweenrefills

201.7.9.2.9.101.1d)

FortheLAYOPERATOR instructions,conditionsunderwhichtheHUMIDIFIERmaintainstheaccuracyofcontrolledanddisplayedvariables

201.7.9.2.9.101.1a)

For the LAYOPERATOR instructions,methodbywhichallof theALARM SIGNALS canbefunctionallytestedtodetermineif theyareoperatingcorrectly, ifequippedwithanALARMSYSTEM

201.7.9.2.8.101

For the LAYOPERATOR instructions, themaximumvolumeofwater, expressed inml,available forvaporizationcontained in the LIQUID CONTAINERand, ifprovided, in theLIQUIDRESERVOIR

201.7.9.2.9.101.1c)

FortheLAYOPERATORinstructions,thespecificationsofanyACCESSORIESorequipmentrequiredtoperformthetestsdescribedin201.7.9.2.8.101

201.7.9.2.8.101

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 53: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved 41

Table 201.C.103 (continued)

Description of requirement Subclause

ForthesupervisingclinicianorthehealthcareprofessionalOPERATORinstructionsandfor a HUMIDIFIER that entrains air for the purpose of diluting oxygen, arecommendationthattheoxygenconcentrationbemeasuredatthepointofdeliverytothePATIENT

201.7.9.2.9.101.2k)

ForthesupervisingclinicianorthehealthcareprofessionalOPERATORinstructionsandforaHUMIDIFIER thatentrainsair for thepurposeofdilutingoxygen,astatement tothe effect that the oxygen concentration can be affected by a partial obstructiondownstreamoftheHUMIDIFIER

201.7.9.2.9.101.2k)

ForthesupervisingclinicianorthehealthcareprofessionalOPERATORinstructions,theessential technical characteristics of each recommendedBREATHING SYSTEM FILTER, ifapplicable

201.7.9.2.9.101.2j)

ForthesupervisingclinicianorthehealthcareprofessionalOPERATORinstructions,thegas leakage of the HUMIDIFIER or individual components, as appropriate, at themaximumRATEDpressure

201.7.9.2.9.101.2f)

ForthesupervisingclinicianorthehealthcareprofessionalOPERATORinstructionsforuse,theinformationcontainedininstructionsforuseforLAYOPERATOR

201.7.9.2.1.101

ForthesupervisingclinicianorthehealthcareprofessionalOPERATORinstructions,theknown adverse effects on the performance of the HUMIDIFIERwhen exposed to, forexample, electrocautery, electrosurgery, defibrillation, X‐ray (gamma radiation),infraredradiation,conductedtransientmagneticfieldsincludingmagneticresonanceimaging(MRI),andradiofrequencyinterference

201.7.9.2.9.101.2i)

ForthesupervisingclinicianorthehealthcareprofessionalOPERATORinstructions,thelocation in the HUMIDIFIER or ACCESSORIES to which the displayed MEASURED GASTEMPERATUREisreferenced

201.7.9.2.9.101.2e)

ForthesupervisingclinicianorthehealthcareprofessionalOPERATORinstructions,themaximumDELIVEREDGASTEMPERATURE,iftheHUMIDIFIERisnotprovidedwithameansofcontinuouslyindicatingtheMEASUREDGASTEMPERATURE

201.7.9.2.9.101.2d)

ForthesupervisingclinicianorthehealthcareprofessionalOPERATORinstructions,theMAXIMUMLIMITEDPRESSUREoftheHUMIDIFIERandACCESSORIES

201.7.9.2.9.101.2a)

ForthesupervisingclinicianorthehealthcareprofessionalOPERATORinstructions,theMAXIMUMOPERATINGPRESSURE

201.7.9.2.9.101.2b)

ForthesupervisingclinicianorthehealthcareprofessionalOPERATORinstructions,theRATED range of environmental operating conditions (temperature and altitude) ofNORMALUSE

201.7.9.2.9.101.2c)

ForthesupervisingclinicianorthehealthcareprofessionalOPERATORinstructions,theminimum,maximum and RATED range of flowrate, GAS PATHWAY resistance and GASPATHWAY compliance of the assembled OPERATOR‐detachable parts, over which theaccuraciesofsetandmonitoredhumidificationaremaintained,unlesstheHUMIDIFIERisintegratedintootherequipment

201.7.9.2.9.101.2g)

For the supervising clinician or the healthcare professional OPERATOR instructions,pressure drop, as a function of flowrate, across the HUMIDIFIER and ACCESSORIES orindividualcomponents,unlesstheHUMIDIFIERisintegratedintootherequipment

201.7.9.2.9.101.2h)

HUMIDIFICATIONOUTPUT(inmg/l)overtheRATEDrangeofgasflowratesandsettings 201.12.1.101

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 54: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

42 ©ISO2017–Allrightsreserved

Table 201.C.103 (continued)

Description of requirement Subclause

List of the applicable parameters such as temperature, pressure, humidity, timelimitsandnumberofcyclesthatsuchMEEQUIPMENT,partsorACCESSORIEScantolerate

201.7.9.2.12

Maximumstart‐upperiodinNORMALUSE 201.12.1.102

ProcessingorreprocessingPROCESSinstructionsfortheHUMIDIFIERanditsACCESSORIES 201.11.6.6

Qualityandpurityof thewater tobeused in theHUMIDIFIER, and thataddingothersubstancescanhaveadverseeffects

201.7.9.2.1.102c)

Recommendedmountingmethodsandotherrelevantinstallationinformation 201.7.9.2.6

SeparateinstructionsforuseforLAYOPERATOR 201.7.9.2.1.101

SeparateinstructionsforuseforsupervisingclinicianorthehealthcareprofessionalOPERATOR

201.7.9.2.1.101

SetofACCESSORIESand, ifapplicable, theMEEQUIPMENTnecessary fortheHUMIDIFIER’SINTENDEDUSE

201.7.9.2.14.101a)

Warning statement to the effect that covering breathing tubes with a blanket orheatingtheminanincubatororwithanoverheadheatercanaffectthequalityofthetherapyorinjurethepatient,ifapplicable

201.7.9.2.2.101d)

Warningstatementtotheeffectthatdonotaddanyattachmentsoraccessoriestothehumidifierthatarenotlistedintheinstructionforuseofthehumidifieroraccessoryorthehumidifiermightnotfunctioncorrectlyaffectingthequalityofthetherapyorinjuringthepatient

201.7.9.2.2.101a)

Warningstatementtotheeffect thatdonotusethehumidifieratanaltitudeabove[insert maximum RATED altitude] or outside a temperature of [insert RATEDtemperaturerange].Usingthehumidifieroutsideofthistemperaturerangeorabovethisaltitudecanaffectthehumidifierperformancewhichconsequentlycanresultinpatientdeath

201.7.9.2.2.101b)

Warningstatementtotheeffectthattopreventdisconnectionofthetubingortubingsystemduringuse,especiallyduringambulatoryuse,onlytubesincompliancewithISO5367orISO80601‐2‐74shouldbeused

201.7.9.2.2.101c)

WhichportionsoftheGASPATHWAYSthroughtheHUMIDIFIERcanbecomecontaminatedwith body fluids or expired gases during both NORMAL CONDITION and SINGLE FAULTCONDITION

201.7.9.2.12

201.C.6 ACCOMPANYING DOCUMENTS, technical description

Addition:

201.C.6.101 ACCOMPANYING DOCUMENTS, technical description of a HUMIDIFIER

AdditionalrequirementsforACCOMPANYINGDOCUMENTS,technicaldescriptionofaHUMIDIFIERarefoundinTable201.C.104.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 55: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved 43

Table 201.C.104 — ACCOMPANYING DOCUMENTS, technical description of a HUMIDIFIER

Description of requirement Subclause

Descriptionof amethod for checking the function of ALARM SYSTEM for eachof theALARMCONDITIONsofthisdocument,ifnotperformedautomaticallyatstart‐up

201.7.9.3.101

Disclosureoftheinterdependenceofcontrolfunctions 201.7.9.3.1.101a)

Disclosureofthemeasurementuncertaintyforeachdisclosedtolerance 201.5.101.3

ListingofwhichALARMCONDITIONsthatarecheckedautomaticallyatstart‐up 201.7.9.3.101

Statement to the effect that the responsible organization should ensure thecompatibility of thehumidifier andall of theparts andaccessories intended tobeusedtoconnecttothepatientpriortouse

201.7.9.3.1.101b)

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 56: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

44 ©ISO2017–Allrightsreserved

Annex D(informative)

Symbols on Marking

IEC60601‐1:2005+AMD1:2012,AnnexDapplies,exceptasfollows.

Addition:Table 201.D.2.101 — Additional symbols on marking

No Symbol Reference Title and description

1

IEC80878:2015[16]

ISO7000‐2492[3]

Symbol5.1.5ISO15223‐1:2016

Batchcode

ToidentifytheMANUFACTURER'Sbatchorlotcode,forexample,onamedicaldeviceorthecorrespondingpackaging.Thecodeshallbeplacedadjacenttothesymbol.

2

IEC80878:2015[16]

ISO7000‐2493[3]

Symbol5.1.6ISO15223‐1:2016

Cataloguenumber

ToidentifytheMANUFACTURER’Scataloguenumber,forexample,onamedicaldeviceorthecorrespondingpackaging.Thecataloguenumbershallbeplacedadjacenttothesymbol.

3

IEC80878:2015[16]

ISO7000‐2498[3]

Symbol5.1.7ISO15223‐1:2016

Serialnumber

ToidentifytheMANUFACTURER’Sserialnumber,forexample,onamedicaldeviceoritspackaging.Theserialnumbershallbeplacedadjacenttothesymbol.

4

IEC80878:2015[16]

ISO7000‐2725[3]

Symbol5.4.5ISO15223‐1:2016

Containsorpresenceof(naturalrubberlatex)

Onmedicaldevices:toindicatethattheequipmentcontainstheidentifiedproductorsubstance.

NOTEReplace“XXX”withthesymbolorotheridentificationofthesubstancethatiscontainedorpresent,whereLATEXisusedfornaturalrubberlatex.

5

IEC80878:2015[16]

ISO7000‐2725[3]

EN15986:2011

Containsorpresenceof[XXX]

Onmedicaldevices:toindicatethattheequipmentcontainstheidentifiedproductorsubstance.

NOTEReplace“XXX”withthesymbolorotheridentificationofthesubstancethatiscontainedorpresent,wherePHTisusedforphthalate.

Additional Annexes:

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 57: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved 45

Annex AA(informative)

Particular guidance and rationale

AA.1 General guidance

Thisannexprovidesrationalefortheimportantrequirementsofthisdocumentandisintendedforthosewhoarefamiliarwiththesubjectofthisdocumentbutwhohavenotparticipatedinitsdevelopment.Anunderstandingof the reasons for themain requirements is considered tobeessentialforitsproperapplication.Furthermore,asclinicalpracticeandtechnologychange,itisbelievedthatrationaleforthepresentrequirementswillfacilitateanyrevisionofthisdocumentnecessitatedbythosedevelopments.

AA.2 Rationale for particular clauses and subclauses

Thefollowingarerationalesforspecificclausesandsubclauseinthisdocument,withclauseandsubclausenumbersparalleltothoseinthebodyofthedocument.

Theclausesandsubclausesinthisannexhavebeensonumberedtocorrespondtotheclausesand subclauses in this document to which they refer. The numbering is, therefore, notconsecutive.

Subclause 201.1.1 — Scope

An ACTIVE HME is also electrically powered MEEQUIPMENT that contains heater elements tovaporize liquidwater that is added into the breathing gas. Thiswater vapour augments thatdelivered to the respiratory tract of the PATIENT from the HME. Most requirements of thisdocumentthereforeapply.Thoserequirementsthatdonotapplyareclearlyindicated.

SomeHUMIDIFIERShavebuilt‐inflowsourcesthatgeneratetheflowofbreathinggas,forexample,sleepapnoeatherapyequipment.ForsuchMEEQUIPMENT,mostrequirementsof thisdocumentapply.Thoserequirementsthatdonotapply,suchaspressuredrop,areclearlyindicated.

Subclause 201.4.3.101 — Additional requirements for ESSENTIAL PERFORMANCE

The ESSENTIAL PERFORMANCE of a HUMIDIFIER fundamentally is maintaining its HUMIDIFICATIONOUTPUT or informing the OPERATOR that this performance is not being attained. However,accurateandtimelymeasurementsofHUMIDIFICATIONOUTPUTareproblematicatbest.

Given the nature of the physics of humidification, short transient changes in HUMIDIFICATIONOUTPUTarenotclinicallysignificant.Therefore,changes inandnottheabsoluteHUMIDIFICATIONOUTPUTaveragedovertimeareusedasacceptancecriteriafollowingthespecifictestsrequiredbythisdocumenttodemonstratethemaintenanceofESSENTIALPERFORMANCE.

TheMANUFACTURERshouldconsider the timelinessof thenotificationof theOPERATORprovidedbytheALARMCONDITIONSorindicationofabnormaloperationwhenHUMIDIFICATIONOUTPUTisnotmaintained.

Subclause 201.4.6 — ME EQUIPMENT or ME SYSTEM parts that contact the PATIENT

SincemuchoftheBREATHINGSYSTEMislikelytobedrapedoveroraroundthePATIENT,itislikelytocomeintodirectcontactwiththePATIENTduringNORMALUSE.Additionally,theGASPATHWAYSconductfluidsintooroutofthePATIENT.Assuch,theGASPATHWAYSoftheBREATHINGSYSTEMand

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 58: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

46 ©ISO2017–Allrightsreserved

the HUMIDIFIER need to be investigated regarding biocompatibility and compatibility withsubstancesthatmightpassintothePATIENTviatheGASPATHWAYS.AlsoofconcernareelectricalHAZARDSshouldanycircuitrybeincorporatedintotheBREATHINGSYSTEM(particularlyforheatedBREATHING SYSTEMS). By ensuring that those items are subject to the requirements for APPLIEDPARTS,theseissuesareaddressedbytherequirementsalreadyinthegeneralstandard.

Subclause 201.11.101 — Additional requirements for pressurized gas input

AHUMIDIFIER designed to be connected to a pressurized gas supply is required to continue tooperatereliablythroughoutitsRATEDrangeofsupplypressures,andthesepressurescanonlybemaintainediftheHUMIDIFIERinNORMALCONDITIONdoesnotattempttodrawmoreflowfromthegassourcethanthegassource isdesignedtosupply. It isalsoexpectedthattheseHUMIDIFIERSshouldbedesigned topreventanunacceptableRISKunderpossible SINGLEFAULTCONDITIONSofthepressurizedgassupply.

Pressurizedmedicalgassupplies, includingMEDICALGASPIPELINESYSTEMSandcylinderpressureregulators conforming to current relevant standards, supply gas‐specific terminal outlets at apressure that iswithin an internationally agreedpressure rangeof 280kPa to600kPaunderNORMALCONDITION.ItisexpectedthatHUMIDIFIERSshouldoperatetotheirdeclaredspecificationatanysupplypressurewithinthisrange.

In the case of a pressure regulator failure, the gas supply pressure could rise to thepressureregulator’ssupplypressure,whichcanbecylinder(tank)pressure.Tosafeguardagainstthisorsimilareventualities,gas‐specificmedicalgassupplysystemsarerequiredtobeprovidedwithameanstolimittheiroutputpressuretonotmorethan1000kPa.Allgas‐poweredMEEQUIPMENTshouldbedesignedsoasnottopresentanunacceptableRISK if itssupplypressurerisesuptothisvalue.

HUMIDIFIERSwithmaximumRATEDinputpressuresexceeding600kPaarerequiredtofulfiltheseconditionsatuptotwicetheirmaximumRATEDinputpressure.

To ensure that theminimum pressure of 280kPa can bemaintained in practice, MEDICAL GASPIPELINESYSTEMSsupplyingcompressedmedicalgasesthroughgas‐specific terminaloutletsaredesigned so that they can maintain this pressure at the input of gas‐powered devices whilesupplying steady‐state flowrates up to 60l/min at a single outlet connected directly to thepipeline. Account is taken of the pressure drop in the pipeline supplying the outlet and thepressuredrop,at60l/min,acrosstheterminalunitandthehoseassemblyconnectingthedevicetothepipeline.

TheMEDICALGASPIPELINESYSTEMisalsorequiredtobecapableofsupplyingsufficientgasthatthisflowcanbedrawnfromapredeterminednumberofadjacentterminalunitssimultaneously.TheactualnumberwillhavebeendeterminedduringthedesignandinstallationoftheMEDICALGASPIPELINESYSTEMby theapplicationofa “diversity factor”, a factoragreedbetween thesupplierandRESPONSIBLEORGANIZATIONtobeappropriateforeachsectionoftheinstallationaccordingtothedesignatedpurposeofeachareasupplied.Recommendeddiversityfactorsareformulatedtoensure that the MEDICAL GAS PIPELINE SYSTEM is capable of supplying an average flowrate of60l/mintotherequiredproportionofterminaloutlets.However,iftheflowratedemandfrommany adjacent MEEQUIPMENT exceeds 60l/min, there is an increased possibility that theHUMIDIFIER inputpressure could fall below280kPa,mainly becauseof the increasedpressuredrop across the terminal unit and input hose assembly (also because of the flow‐dropcharacteristicinthecaseofpressureregulatorssupplyingasingleterminaloutlet).

Inadditiontosteady‐stateflowratesof60l/min,theswitchingoftheinternalpneumaticsystemand theoperationof a PATIENTdemand systemcan result inMEEQUIPMENT requiring transient

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 59: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved 47

input flowrates far in excess of 60l/min. Because of the compressibility of gas at pipelinepressuresandthediameterofpipingthatisemployedinordertominimizethepressuredrop,such transientdemands can generallybe accommodated from the gaswithin thepipesof theMEDICALGASPIPELINESYSTEM.Therecanbetemporarypressuredropsoftheinputpressureattheinletof theMEEQUIPMENT, tobelow280kPa,due to transient flowrates in excessof200l/min(over3s) butmost of thesedropswill bewithin the supplyhose assemblies specifiedby theMANUFACTURER. MANUFACTURERS need to evaluate their own designs to establish whether anyconsequent transientpressuredrop affects theperformanceof theirMEEQUIPMENTwhenusedwithrecommendedsupplyhoseconfigurationsandwhenconnectedtoalternativegas‐specificterminaloutlets,suchasthosefittedtocylinderpressureregulatorsconformingtoISO10524‐1.

MEEQUIPMENT that can draw greater average or transient flows during INTENDED USE arepermitted,buttheirACCOMPANYINGDOCUMENTSarerequiredtodisclosethoseflowratesandwarnoftheneedforadifferentdiversityfactor.

The average flowrate of 60l/min is greater than the test flowrate used during thecommissioningofMEDICALGASPIPELINESYSTEMS.Initself,thisshouldbeofnoconcernbecausethespecific conditions specified for the test do not allow a direct comparison between the twovalues.Thecommitteeresponsibleforpipelinestandards,ISO/TC121/SC6,inconsultationwithISO/TC121/SC1andISO/TC121/SC3,agreedtothe60l/minaverageflowratevalue,andalsothe200l/minforupto3s transient flowrates,duringthepreparationof thecurrentstandardforMEDICALGASPIPELINESYSTEMSandwereawareof theneedtosatisfythatspecificationwhenfinalizingtheMEDICALGASPIPELINESYSTEMtestrequirements.

MANUFACTURERS should be aware that othermedical gas supply system standards permit thefitting of gas‐specific terminal outlets to supply systems such as pendant supply units. Suchsubsystemsrestricttheflowthatcanbedrawnfromtheirterminaloutlets.

Subclause 201.5.101.2 — Gas flowrate and leakage specifications

Quantitiesofgasarefrequentlyexpressedasthevolumethatthegasoccupiesatstandardizedconditions. Generally, one atmosphere (101,325kPa) is used as standard pressure. However,several standard temperatures are used. Whereas 0°C is used as standard temperature inphysics,either20°Cor21,1°C(70°F)isoftenusedinengineering.Inventilation,thegasinthelungshasatemperatureidenticaltobodytemperature(~37°C)irrespectiveofthetemperatureofthegasdeliveredtothePATIENT.Thevolumeofagivenamountofdrygasincreasesbyabout13,6%from0°Cto37°Corby5,8%from20°Cto37°C.

Gas delivery systems supplying pressurized gas tomedical equipment, including HUMIDIFIERS,followengineeringconventionsandspecifygasquantitiesandflowratesatSTPDconditions.Thispracticeisfollowedinthisdocumentforallrequirementsconcerninggasinput.

However,HUMIDIFIERScomplyingwiththisdocumentarelikelytobeusedwithventilatorsthatinflatethePATIENT’Slungsrelativetoalocalatmosphericpressurebetween70kPaand110kPa.In addition, the gas in the lungs is always saturated with water vapour regardless of thehumidityofthegasdeliveredtothePATIENT’Sairway.Withastandardtemperatureof0°C,1lofgasreferencedtoSTPD(STANDARDTEMPERATUREPRESSUREDRY)canexpandthelungsby1,8latapressure of 70kPa. In order to have the values comparable among different HUMIDIFIERS, it isessentialthattheinformationforallHUMIDIFIERSisreferencedtothesamestandardconditions.Becauseitisthevolumeofgasandnotthenumberofmoleculesthatexpandsthelungs,BTPSistheappropriatesetofreferenceconditionstouse.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 60: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

48 ©ISO2017–Allrightsreserved

Subclause 201.5.101.3 — HUMIDIFIER testing errors

When testing HUMIDIFIER performance, several of the test parameters cannot be measuredwithoutasignificantdegreeofmeasurementuncertaintyduetolimitationsoftheaccuracythatcanbe achieved, particularlywhenmeasuring volumesby the integrationof rapidly changingflows.

Because of the relative significance of these uncertainties, it is important thatMANUFACTURERSallowforthemwhendeclaringparameteraccuracy.

Similarly,itisimportantforthird‐partytesterstorecognizethesignificanceoftheuncertaintyintheirownmeasurementswhentestingtothisdocument.

Inpractice,thismeansthat,forexample,ifaMANUFACTURERdeterminesthataparameterhasatoleranceof±7%butthatthemeasurementuncertaintyis±3%,thenaparametertoleranceof±10%isdeclared. Ifa third‐party testersubsequentlyobtainsanerrorof themeasuredvaluefor that parameter of ±15%, with a measurement uncertainty of ±5%, then the third‐partytesterhastoaccepttheMANUFACTURER’Sclaim.

Furthermore, theMANUFACTURER is required todisclose themeasurementuncertainty foreachdeclared value in order to provide both information to the RESPONSIBLE ORGANIZATION andguidance fora third‐party testeras to theneededmeasurementaccuracywhen testing to thisdocument.

Subclause 201.7.2.8.2 — Other power sources

TheHEATEDBREATHINGTUBECONTROLLERMANUFACTURERneedstomarkthemaximumamountofpowerthatcouldbedeliveredtotheBREATHINGTUBE.ThisisnecessarytopermitMANUFACTURERSof BREATHING TUBES to be able to test completely their BREATHING TUBES in order to meet therequirementsof201.102.SincemanyoftheBREATHINGTUBEMANUFACTURERS“reverse‐engineer”the BREATHING TUBE, the worst‐case, maximum power output of the HEATED BREATHING TUBECONTROLLERneedstobeknown.

Subclause 201.7.4.3 — Units of measurement

Additionalinformationisfoundinrationalefor201.5.101.2.

Subclause 201.7.9.2.2.101 — Additional requirements for warnings and safety notices

a)

TheOPERATORshouldbeawarethatonlythepartsorACCESSORIESlistedintheinstructionforusehave been VALIDATED by the MANUFACTURER. The use of non‐VALIDATED parts can represent anunacceptableRISK.

Forexample,

— apowersupplyunitotherthantheonerecommendedbytheMANUFACTURERcanbedesignedand manufactured with inferior quality (poor reliability), can affect the electromagneticcompatibilityoftheHUMIDIFIER,etc.;

— theconnectionofpartstotheBREATHINGSYSTEMthatarenotlistedintheinstructionforusecan increase the inspiratory or expiratory pathway resistance of the BREATHING SYSTEM orcan increase the unintentional leakage of the BREATHING SYSTEM to a level that affects theBASICSAFETYandESSENTIALPERFORMANCE.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 61: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved 49

Subclause 201.7.9.2.8.101 — Additional requirements for start-up PROCEDURE

Insomedesigns,adequatecheckingoftheALARMSYSTEMcanbeperformedwithacombinationofOPERATORactionandthepower‐onself‐testroutinesthatVERIFYtheintegrityofthesoftwareandthe integrityof thecomputercontrolling theHUMIDIFIER,aswellas themeasuringsensorsandtheALARMSIGNALgeneration.

Subclause 201.7.9.2.9.101.2 — Supervising clinician operating instructions

e)

HUMIDIFIERScanincorporatetemperaturesensorsthatmeasureanddisplaythetemperatureofthegasatvariouslocations.Manydifferentdesignapproachesexist.ThetemperatureofthegasatthePATIENT‐CONNECTIONPORTisnotalwaysthemostclinicallyusefultemperaturetodisplay,forexample.

EXAMPLE1 HUMIDIFIERwithaheatedwireBREATHINGTUBE.

Water is heated in a HUMIDIFICATION CHAMBER until the MEASURED GAS TEMPERATURE in theHUMIDIFICATIONCHAMBERreaches37°C.ThegasisthenheatedintheBREATHINGTUBEto40°Catthe PATIENT‐CONNECTION PORT to prevent condensation. The RELATIVE HUMIDITY exiting theHUMIDIFICATION CHAMBER is approximately 100% and approximately 85% at the PATIENT‐CONNECTION PORT. In addition, because the total heat content (or energy content) of the gas isprimarily due to the water vapour content, the total heat content of the gas at the PATIENT‐CONNECTIONPORTisonlyslightlyhigher(duetotheincreasedtemperature)thanthatofthegasexitingtheHUMIDIFICATIONCHAMBER.ThismeansthatoncethegasleavesthePATIENT‐CONNECTIONPORT,rapidcoolingwilloccuruntilthegasreturnsto100%RELATIVEHUMIDITYandthereforehascooled back to the HUMIDIFICATION CHAMBER temperature (the saturated gas temperature).Following this rapid cooling, the gas will more slowly equilibrate to the temperature of thePATIENT.

Inexample1,theMEASUREDGASTEMPERATUREintheHUMIDIFICATIONCHAMBERisthebestindicationof the humidity being delivered to the PATIENT because it represents the saturated gastemperature. TheMEASURED GAS TEMPERATURE at the PATIENT‐CONNECTION PORT represents a gastemperature that has a RELATIVE HUMIDITY of less than 100% and is therefore not a goodindicationofthehumiditythePATIENTreceives.

EXAMPLE2 HUMIDIFIERwithnon‐heatedBREATHINGTUBE.

WaterisheatedinaHUMIDIFICATIONCHAMBERandtheresultantvapouriscarriedbygasflowingthroughtheHUMIDIFIERtoanunheatedBREATHINGTUBE.AsthegastravelsthroughtheBREATHINGTUBEitcoolsuntilitreaches37°CatthePATIENT‐CONNECTIONPORT.TheRELATIVEHUMIDITYatthePATIENT‐CONNECTIONPORTisapproximately100%.Toachievethistemperatureandhumidity,theMEASUREDGASTEMPERATUREintheHUMIDIFICATIONCHAMBERcanbeashighas55°Cbutneedonlyhave42%RELATIVEHUMIDITY.

Inexample2,theMEASUREDGASTEMPERATUREatthePATIENT‐CONNECTIONPORTistheappropriateindicationofthehumiditybeingdeliveredtothePATIENT.

Therefore,inordertoencompassdifferenttechnologies,itisnotappropriateforthisdocumentto require a temperature display to show only the temperature delivered at the PATIENT‐CONNECTIONPORToratanyspecificpointintheHUMIDIFIERandACCESSORIES.TheMANUFACTURERisrequiredtostateclearlyintheinstructionsforusethesitetowhichthedisplayedMEASUREDGASTEMPERATUREisreferenced.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 62: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

50 ©ISO2017–Allrightsreserved

In both examples, BASIC SAFETY and ESSENTIAL PERFORMANCE is maintained as both HUMIDIFIERSmeet the requirements of this document (e.g. 201.12.1.101, HUMIDIFICATION OUTPUT at thePATIENT‐CONNECTION PORT, and 201.12.4, protection against hazardous output) despite the factthattheyaredisplayingtemperaturesfromdifferentlocations.

g)

ResistancetoflowanywhereintheGASPATHWAYcanincreasetheworkofbreathing.Itcanalsointerfere with the effectiveness of intermittent mandatory ventilation (IMV) or triggeringmechanismsinlungventilators.

h)

The internal compliance of the BREATHING SYSTEM, which includes the HUMIDIFIER, needs to beknown in order to accurately determine the tidal volume settings of volume‐controlledventilators.

k)

TheamountofairentrainedbyaHUMIDIFIER(e.g.byaVenturimechanism)isafunctionofgasvelocity.Changesingasvelocity(e.g.becauseofapartialobstructionoftheventilationcircuit)directlyaffectstheoxygenconcentration.

Subclause 201.9.4.3.101 — Additional requirements for instability from unwanted lateral movement

TRANSIT‐OPERABLE MEEQUIPMENT needs to be capable of being attached to wheelchairs andparticularlyautomobileswhenthePATIENTisusingtheMEEQUIPMENTwhiletravelling.AsuddenstopinanautomobilecancauseMEEQUIPMENTtobecomeahazardousflyingobject.Thismeansof attachment should not involve the use of a TOOL as the MEEQUIPMENT needs to be easilyattachedanddetached.

Subclause 201.11.1.2.2 — APPLIED PARTS not intended to supply heat to a PATIENT

TheobjectiveofthisrequirementistoprotectthePATIENTfromskinburnsduetocontactwiththeexternalsurfaceoftheBREATHINGTUBE.Seerationaleto201.12.4.101forselecting44°C.

Subclause 201.11.6.2 — Overflow in ME EQUIPMENT

HUMIDIFIERSareoftenmountedonpolesinNORMALUSE.However,theseHUMIDIFIERSareoftennotmounted exactly horizontally. The committee felt that a 20° tilt (beyond that of NORMAL USE)couldbeconstruedasreasonablyforeseeable,andthereforerequiredthattheHUMIDIFIERshouldoperatenormally,whichincludesnotspillinganyliquid,beyondthatspecifiedin201.13.1.101whenoperatedunderNORMALCONDITIONSatthisposition.PermanentlymountedMEEQUIPMENTisunlikelytobesubjecttosuchtilting,norwouldMEEQUIPMENTthatwasintendedtobeoperatedwhile placed on a table or floor, such as sleep apnoea therapy equipment, and so 10° wasregardedasasufficienttestangle forthese.A15%overfill isalsoreasonably foreseeableandthe HUMIDIFIER should operate normally, which includes not spilling any liquid, beyond thatspecifiedin201.13.1.101whenoperatedunderNORMALCONDITIONSwiththisamountofoverfill.

An ACTIVE HME that is located proximal to the PATIENT can be in any orientation. The leastfavourableorientationneedstobedeterminedfortheoverflowtest.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 63: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved 51

Subclause 201.11.6.6 — Cleaning and disinfection of ME EQUIPMENT or ME SYSTEM

TheessentialprinciplesofISO16142‐1[7]requirethatmedicaldevicesarenottobeoperatedorusediftheirconditioncouldcompromisethehealthandsafetyofthePATIENTonwhomtheyarebeingusedortheemployeesorthirdpartiesinteractingwiththem.

Thismeans thatHUMIDIFIERS, theirACCESSORIESandparts cannotbeused if there is apotentialRISK of the PATIENT, OPERATOR or other person being infected as a result of contact with theHUMIDIFIER,ACCESSORYorpart.

Therefore,non‐singleuseHUMIDIFIERS,theirACCESSORIESandpartsrequireanappropriatelevelofdisinfection,dependingontheiruse,butrarelyneedtobesterile.

Recommendations for hygienic reprocessing of HUMIDIFIERS, their ACCESSORIES and parts arebasedonthegeneralhygienerequirementsforthereprocessingofmedicaldevicesandneedtotake into consideration the special requirements and needs of PATIENT care in the clinicalenvironment[7].Therequirementsforhygienicreprocessingofthisdocumentareintendedto

— make the RESPONSIBLE ORGANIZATION for reprocessing the HUMIDIFIER aware of how toimplementthesetasksinaresponsiblemannerthroughappropriatedelegation,and

— helpall parties involved in the reprocessingof HUMIDIFIERS, theirACCESSORIES andparts tocomplywiththeMANUFACTURER’Sinstructions.

The cleaning and disinfection PROCEDURES of the MANUFACTURER are also intended to providepracticalsupporttoallthoseinvolvedinPATIENTcareintheclinicalenvironmentwithregardtoimplementingthehygienemeasuresrequiredforthePATIENT’Ssafety.

It shouldbenoted that HUMIDIFIERS, as all othermedicaldevices thathavebeen contaminatedwith human pathogenic microorganisms, are a potential source of infection for humans. AnyHUMIDIFIER that has already been used on another PATIENT is potentially contaminated withcontagious pathogenic microorganisms until proven otherwise. Appropriate handling andreprocessingPROCEDURESareessentialtoprotectthenextpersonhandlingthedeviceorthenextPATIENTonwhom thedevice is used.HenceHUMIDIFIERS, their re‐usableACCESSORIES andpartsthat have been used are required to undergo a reprocessing PROCESS, following theMANUFACTURER’Sinstructions,priortoreusebyanotherPATIENT.

ThefollowingbasicconsiderationsneedtobeaddressedbytheMANUFACTURERwhenspecifyingthereprocessinginstructionsofaHUMIDIFIER,itsACCESSORIESorparts:

a) protectingthePATIENT,theOPERATORandtheRESPONSIBLEORGANIZATION(includingpersonnelinvolvedinperformingthereprocessingPROCESS);

b) the limits of the PROCEDURES used for reprocessing (such as the number of reprocessingcycles);

c) thenecessity toguarantee theprovenstandardizedPROCEDURES ina consistentlyhighandverifiablequality,basedonanestablishedqualitymanagementsystem.

TherecommendedreprocessingPROCESSshouldbedeterminedby:

— thepotentialdegreeandtypeofcontaminationoftheHUMIDIFIER,ACCESSORIESorparts;

— theRISKofinfectinganotherPATIENTresultingfromtheirreuseandthetypeofapplicationoftheHUMIDIFIER.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 64: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

52 ©ISO2017–Allrightsreserved

Specialconsiderationof thepossibleRISKassociatedwiththecontaminationofgas‐conductingcomponents due to the PATIENT’S re‐breathing under SINGLE FAULT CONDITION should beconsidered.

Onthebasisoftheabove,aVERIFIEDandVALIDATEDdocumentedreprocessingPROCEDUREneedstobespecified insuchdetail so that theoutcome is reproducible.AnacceptableRESIDUALRISKfromtheHAZARDofinfectionforthenextPATIENTcanbeassumedif:

a) the documented reprocessing PROCEDURE’S effectiveness has been VERIFIED throughappropriatescientificmethodsbytheMANUFACTURER;

b) the reliability of the documented reprocessing PROCEDURES has been VERIFIED in practicethroughappropriatequalityassurancemeasuresbytheRESPONSIBLEORGANIZATIONcarryingoutthereprocessingPROCEDURES.

WhenselectingandevaluatingthereprocessingPROCEDURES,theMANUFACTURERshouldconsider:

— theamountandtypeofpathogenicmicroorganismsexpectedtocontaminatetheHUMIDIFIER,ACCESSORIESorparts;

— theRISK for thepathogenicmicroorganisms tobe transmitted to the PATIENT,OPERATORorotherpersons;

— themicroorganism'sresistancetotherecommendedreprocessingPROCEDURES.

The RISKS posed by a reprocessed HUMIDIFIER, ACCESSORIES or parts are determined by thefollowingfactors:

a) undesiredeffects,whichcanresultfrom:

— theprevioususe;

— thepreviousreprocessingPROCESSES;

— transportationandstorage;

b) theRISKSfromsubsequentuses,suchasthefollowing:

— residuesfromtheprevioususe(suchassecretions,otherbodyfluids,anddrugs);

— residues from the previous reprocessing PROCESSES (such as cleaning agents,disinfectantsandothersubstances,includingtheirreactionproducts);

— changesofphysical,chemicalorfunctionalpropertiesofthedevice;

— changes in the condition of the material (such as accelerated wear and tear,embrittlementandchangedsurfaceconditions,connectorsandadhesivejoints);

c) theRISKoftransmissionofanypathogenicmicroorganisms.

When considering the suitability of the reprocessing PROCESS and the feasibility of thereprocessingPROCESSfortheHUMIDIFIER,ACCESSORIESorparts,theMANUFACTURERshouldconsiderthefollowingpoints:

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 65: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved 53

— theRISKSinvolvedinthereprocessingPROCESS;

— thecosteffectivenessofthereprocessingPROCESS;

— thepracticabilityofthereprocessingPROCESS;

— the availability of the cleaning equipment and the cleaning agents specified in thereprocessingPROCESS;

— theefficiencyofthereprocessingPROCESS;

— thereproducibilityofthereprocessingPROCESS;

— qualitymanagementrequirementsofthereprocessingPROCESS;

— the environmental impact of the reprocessing PROCESS and the disposal of the HUMIDIFIER,ACCESSORIESorparts.

The MANUFACTURER should VERIFY all cleaning agents and reprocessing PROCEDURES used withregardtotheirsuitabilityandrepeatabilitywiththeHUMIDIFIER,ACCESSORIESorparts,dependingonthetypeofuse.

The RESPONSIBLE ORGANIZATION should VERIFY that manual cleaning and disinfection of theHUMIDIFIER, ACCESSORIES or parts are always carried out in accordance with the PROCEDURESspecifiedintheACCOMPANYINGDOCUMENT.

TheMANUFACTURERshouldspecifyVALIDATEDautomatedcleaninganddisinfectionPROCEDURES.Iftheyarenot followed, theeffectivenessof thecleaninganddisinfectioncannotbeguaranteed.Such parameters could include the volume of water used, water pressure, temperature, pH,dosageofcleaningagentsanddisinfectantsandresidencetime.

ToensurethereproducibilityofautomatedreprocessingPROCEDURES,testsshouldbecarriedoutonaregularbasis.

TheMANUFACTURERshouldensure that thespecifieddisinfectionPROCEDURESareVERIFIED tobebactericidal,fungicidalandvirucidalsothatthecleanedanddisinfectedHUMIDIFIER,ACCESSORIESor parts do not pose an unacceptable RISK of infection by reproductive pathogenicmicroorganismswhenanyoftheseelements,collectivelyorindividuallycomesincontactwiththenextPATIENT,OPERATORorotherperson.

Effectivedisinfectionrequiresthattheinstructionsforthedisinfectant,especiallywithregardtoconcentrationandresidencetime,arefollowed.

Following any reprocessing PROCEDURE, safety and functional testing of the HUMIDIFIER andACCESSORIES (as specified by the MANUFACTURER’S instructions) needs to be carried out. Ifnecessary, safety‐relevant functional testing can be carried out directly before use of theHUMIDIFIER.

TheextentandtypeofthetestsdependsontheHUMIDIFIER,ACCESSORYorpartandtheseneedtobedefinedintheACCOMPANYINGDOCUMENT.

Subclause 201.12.1.101 — HUMIDIFICATION OUTPUT

HUMIDIFIERS can be used with PATIENTS whose upper airways have been bypassed by atracheostomy or tracheal tube (invasive ventilation). The upper airway provides the major

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 66: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

54 ©ISO2017–Allrightsreserved

portionoftheheatandmoisturesuppliedtothealveoli.Whentheupperairwaysarebypassed,the HUMIDIFIER needs to supply this missing heat and moisture. The humidity in the tracheaduringnormalrespirationcanrangefrom36mg/lto40mg/l.ThisfigureisforahealthyPATIENTand still requires healthy functioning airways to further condition the gas to reach BTPS —addingmoistureandraisingthetemperaturetoachieve37°C/44mg/lforalveolarconditions.

Physiological humidity levels (37°C/44mg/l) prevent depletion of moisture from themucociliary transport system and maintain normal mucus clearance. When the airway isexposed to low levelsofhumidity, theaqueous layerdecreases, themucus layer thickens,andcilia beat slows[24]. This reduces the airway defence mechanism and increases the RISK ofrespiratoryinfection[25].

IfthePATIENThascompromisedairwaysasistypicalofalong‐termintubatedPATIENT,thentheyshould be supplied with breathing gasses as close to 37°C/44mg/l (alveolar conditions) aspossibleinordertominimizetheextraloadplacedontheremainingairways.

Since theoptimalmoisture levelbelow the carina is44mg/l (100%RHat37°C), amoisturecontentofmorethan33mg/l,andupto44mg/latthePATIENT‐CONNECTIONPORT,isrequiredtopreventthedryingoutofsecretionsintheartificialairway.

Theminimum temperature of gas able to hold 33mg/l of water at BTPS conditions is 32,2°C(refer toTableAA.1).However, this document refers to the conditionsof gasdelivered to thePATIENT‐CONNECTION PORT. There will be some temperature drop across the tracheal tubeconnectorandtracheal tube,whichareexposedto theambientair.Typically, the temperaturewilldropafewdegreesCelsiusalongthelengthofatrachealtubeconnector.

The requirement is for a minimum of 32,2°C delivered to the lower trachea, so the gas cancontain33mg/lofwateratBTPSconditions.Therefore,forinvasiveventilation,thetemperatureat the PATIENT‐CONNECTION PORT should be at least 34°C to allow for a minimum of a 2°Ctemperaturedropalonganycathetermount.

A tracheal tube does not increase the temperature and humidity as efficiently as the upperrespiratory tract.Asa tracheal tube limits thewarmingandhumidificationof gas as itpassesintothePATIENTairway,thetemperatureandhumidityrequirementsofthegasarehigheratthePATIENT‐CONNECTIONPORTtoaccommodateforthelackofgasconditioning.ThemeasurementofgastemperatureandhumidityasitenterstheintubatedPATIENTandtravelsdownthetrachealtube has shown that if the gas is not warmed and humidified to BTPS prior to entering thePATIENT, then furtherwarming andhumidification of the gas is requiredby the lower airway.However, the lower airway in healthy non‐intubated person contributes little to gasconditioning[25].Thus, instrivingforphysiologicalhumidity levelsatthecarina(37°C/44mg/l)in PATIENTS whose upper airways have been bypassed (invasive ventilation), the HUMIDIFIERneeds to be capable of producing a HUMIDIFICATION OUTPUT of at least 33mg/l at the PATIENT‐CONNECTIONPORT.

HUMIDIFIERS canalsobeused in PATIENTSwhoseupperairwayshavenotbeenbypassed (non‐invasiveventilation,nasalhighflowtherapy,sleepapnoeacontinuouspositiveairwaypressureor CPAP treatment). Adequacy of ventilation is dependent on providing sufficient airflow tomaintainapressuregradient fromthe interface to thePATIENTairways.Thenasalmucosahasconsiderablecapacitytoheatandhumidifyinspiredair,butthiscapacitycanbeoverwhelmedathigh flowrates during ventilation. These conditions provoke mucosal hypertonicity, nasalcongestionandincreasednasalresistance.

Duringnormalbreathinginspiredairiswarmedandhumidifiedintheupperairwaysto31°C,30,8mg/l BTPS, by the time it reaches in the pharynx. By heating and humidifying the non‐

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 67: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved 55

invasive gas flow to the same levelwhich occurs naturally in the nasopharynx, the effects ofdryingcanbeprevented,increasingPATIENTcomfortandtolerancetotherapy[26][27][28].

Nasal congestion, dry nose and dry throat are common in PATIENTS requiring non‐invasiveventilationandsleepapnoeaCPAPtreatment.Highpressuresandflowsprovokechangesinthenasal cell structure, mucosal hypertonicity, nasal congestion, nasal discharge, and increasednasal resistance[28][29][30]. This airwaydrying effect is amplifiedwithoral breathing,mask leakandunidirectionalflowassociatedwithnasalmasksandmouthleak[32][33].

Increased nasal resistance is attenuatedwith humidity levels of 21mg/l andminimizedwith>30mg/l.Comfortdatasuggeststhathumiditylevelsabove15mg/lincreasesPATIENTtoleranceand30mg/lreducesthefeelingoforaldryness.Unheatedpass‐overhumidificationcanachievehumidification levels of 7mg/l to 12mg/l. However, cold‐pass‐over does not achieve theimprovement in respiration water loss or the increase in PATIENT compliance that can beachieved using heated humidification[25][26][29][30][31][32][33][34][35][36][37][38][39]. Thus, 12mg/l shouldbeseenasaminimumlevelofhumidificationforPATIENTSwhoseupperairwayshavenotbeenbypassed.

AlloftheabovediscussionassumesaPATIENTatstandardbodytemperatureof37°C.Forcasesofintentionalhypothermiaorhyperthermia,thelimitsshouldbeadjustedaccordingly.

ForanACTIVEHME,theHUMIDIFICATIONOUTPUTcannotbedeterminedwiththetestsdescribedinthis document. The moisture loss test specified in ISO9360‐1[5] and ISO9360‐2[6] is moreappropriateandcanprovideanindicationofperformanceforanACTIVEHME.SinceanACTIVEHMEaddswater,thetestsofISO9360‐1[5]andISO9360‐2[6]mightnotindicateamoistureloss,butanaddition of moisture. The moisture loss value and HUMIDIFICATION OUTPUT are not directlycomparable.

Subclause 201.12.1.102 — SET TEMPERATURE accuracy and MONITORING EQUIPMENT

ThedisplayedMEASUREDGASTEMPERATUREneedstobeasaccurateaspracticable.Thecommitteeconsideredthataninaccuracyoflessthan2°CofthedisplayedMEASUREDGASTEMPERATUREdoesnot compromise the clinical condition or the safety of the PATIENT. It is important that theOPERATOR ismadepromptly awarewhen theMEASURED GAS TEMPERATUREhas exceeded the SETTEMPERATUREbymorethananacceptableamount.ThecommitteeagreedthatwhatconstitutedanacceptableamountcouldbelefttotheMANUFACTURER.

Theover‐temperatureALARMLIMITissetat43°C.Additionalinformationisfoundintherationalefor201.13.1.101.Becauseof theabilityof theairways tosupplyorabsorbheatandmoisture,thisovershootisrequiredtonotexceedanenergyequivalenttoairat43°Cand100%RELATIVEHUMIDITY (a specific enthalpy 197kJ/m3 of dry air), averaged over 120s. If the delivered gasenergy exceeds a specific enthalpy of 197kJ/m3 of dry air) when averaged over 120s underNORMALCONDITIONorSINGLEFAULTCONDITION,theHUMIDIFIERisrequiredhaveameanstodetectanextremeover‐energyALARMCONDITION.

Subclause 201.12.1.103 — MEASURED GAS TEMPERATURE MONITORING EQUIPMENT

ForaHUMIDIFIER thatdisplays theMEASUREDGASTEMPERATURE, the committee concluded thatarangeof25°Cto45°CwastheminimumthatanOPERATORrequiredtooperatetheHUMIDIFIER.ItshouldbeverycleartotheOPERATORifthedisplayedMEASUREDGASTEMPERATUREishigherthan45°Corlowerthan25°C.

AHUMIDIFIERcontroller,byitsnature,continuouslyadjuststhecomponentsoftheHUMIDIFICATIONSYSTEMthataffect thetemperatureof thehumidifiedgasdeliveredtothePATIENT. It isnormal,therefore,fortheMEASUREDGASTEMPERATUREtocycleabouttheSETTEMPERATURE,butthisisnot

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 68: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

56 ©ISO2017–Allrightsreserved

considered to be clinically significant providing the temperature is within ±2°C of the SETTEMPERATUREwhenaveragedover5min.

Subclause 201.12.4.101 — Thermal overshoot

The human airway has a very significant ability to absorb or deliver heat and moisture.Reference the commonpracticeof sitting in a saunawithoutHARM to the respiratory track[40].Fullysaturatedgasat45°Ccanbe inspired forupto1hwithoutdamagingthemucosaof therespiratory tract[41]. Amore recent study reported tolerance of inspired gas temperatures of46,9°Cto49,3°C,100%RH(256kJ/m3)for49min[42].

Taking intoaccount theenthalpyof inspiredgas thathasbeenshown tobe toleratedwithoutcausing thermal injury to the human airways and the very short exposure times of thermalovershoot from a heated HUMIDIFIER in clinical practice, the delivered gas energy limit of197kJ/m3ofdrygaswhenaveragedover120scanbeused.

Whenconsideringgasmixturesotherthanoxygen/air,thefollowingshouldbeobserved.Giventhatmostoftheenergyiscontainedinthewatervapour,theequivalentofairat43°C,100%RHisthemaximumenthalpythatshouldbeallowed.Thishasaspecificvolumeof0,9786m3/kgofdryairandanenthalpycontentof197kJ/m3ofdryair.AssumingthevolumebreathedbythePATIENTisthesame,whatevergasmixtureisused,thenthesafeenthalpylimit is197kJ/m3ofdrygas.ThisenthalpyperunitvolumegivesamorerelevantmeasureoftheenergydeliveredtothePATIENT.

SustainedDELIVEREDGASTEMPERATURESabove41°CrepresentapotentialthermalHAZARDtothePATIENT.AlthoughitisrarelyneededforPATIENTcare,asustainedDELIVEREDGASTEMPERATUREof41°C at any level of saturation is not a thermal HAZARD to the PATIENT. DELIVERED GASTEMPERATURESabove41°C,dependingonthecombinationofgastemperature,levelofsaturationandPATIENTexposuretime,canbehazardous.

Studies to measure the relative importance of exposure time and temperature in causingcutaneousburnsdeterminedthatsurfacetemperaturesofatleast44°Cand6hexposurewererequired to cause irreversible damage to epidermal cells[43]. This is confirmed by studiesconductedby theUSNavyMedicalResearchandDevelopmentCommand[41],whichconcludedthatfullysaturatedgasat45°Ccanbeinspiredforupto1hwithoutdamagingthemucosaoftherespiratorytract.

Gas at body temperature and fully saturated (37°C and 100%RH)will not transfer thermalenergy to or from the PATIENT with a normal body temperature of 37°C. Dry gas at bodytemperature(37°Cand0%RH)willdrawheatawaythroughevaporation.Gasat41°Candfullysaturatedhasthecapacitytodeliverlessthan181,3kJ/m3ofdrygasbreathedbythePATIENT.ToprotectthePATIENTfromthermalinjury,heatingoftheHUMIDIFIERisinterruptediftheDELIVEREDGASTEMPERATUREexceeds43°C.Athermalovershoot,nottoexceedenergyequaltoairat43°Cand100%RHaveragedover120s(197kJ/m3ofdrygas),isinconsequentialtothePATIENTandispermittedtosimplifyconstructionoftheHUMIDIFIER.

It is therefore important that the OPERATOR be provided with a continuous display of theMEASURED GAS TEMPERATURE, and that the HUMIDIFIER automatically interrupts heating andactivates theover‐temperatureALARMCONDITIONwhen theMEASUREDGAS TEMPERATURE exceeds43°C.

The testing conditions required in 201.12.4.101 are important to ensure that large shifts inflowrates do not result in excessive overshoots of high‐energy gases being delivered to thePATIENT.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 69: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved 57

Subclause 201.13.1.101 — Additional specific HAZARDOUS SITUATIONS

Excessive liquid output could cause PATIENT injury and an accumulation of water in theBREATHINGTUBE.

Subclause 201.13.102 — Independence of HUMIDIFICATION control function and related RISK CONTROL measures

Thisrequirementpreventstheuseofamonitoringdevicetocontrolanactuatorthatwouldleadtoanundetectedmalfunctionoftheactuatorincaseofmonitoringfailure.

Subclause 201.101.1 — General

Non‐standardBREATHINGSYSTEMconnectorscanrepresentanunacceptableRISKasattemptsaremadetofitastandardBREATHINGSYSTEMtoaventilatororHUMIDIFIERinanemergencysituation.Non‐standard BREATHING SYSTEM connectors can cause leaks if used with similar but notcompatibleconnectors.

Subclause 201.101.4 — ACCESSORY port

TheuseofLuer taperorLuer‐lockconnectorscomplyingwith ISO594‐1or ISO594‐2arenotpermitted for use in a BREATHING SYSTEM as there are several case reports of accidentalconnectionwithintravenousfluidsandparenteralandenteralfeedingsolutionscausingseriousmorbidityandmortalityduetoaspirationoftheseforeignsubstancesintothelungs.

Subclause 201.101.8.2 — Leakage

60cmH2O is thepressure currentlyproposed for testing leaks in ISO5367.5ml/min leakagefrom the engaged sensor or mating port is 10% of the total allowable leakage, for a wholeBREATHINGSYSTEM,inISO5367.

Subclause 201.102.1 — General

As safeusedependson the interactionof theHUMIDIFIERwithACCESSORIES, thisdocument setstotal‐systemperformancerequirementsreferenced to thePATIENT‐CONNECTIONPORT.Therefore,total system performance requirements are applicable to both MANUFACTURERS of HUMIDIFIERSandBREATHINGTUBESintendedforusewithaHUMIDIFIER(bothHEATEDBREATHINGTUBESandnon‐heated BREATHING TUBES). The HUMIDIFIER with ACCESSORIES should have a means of reducingcondensateintheBREATHINGTUBES.

EXAMPLE HeatingtheBREATHINGTUBEortheplacementofwatertraps.

It is the responsibility of theMANUFACTURER of a BREATHING SYSTEM, its parts or ACCESSORIES toVERIFYthattheirproductcomplieswiththerequirementsofthisdocument.

BREATHINGTUBESuptothePATIENT‐CONNECTIONPORTformpartofthetotalsystemperformancerequirements.MANUFACTURERSofBREATHINGTUBESneedtoensurethattotal‐systemperformancerequirements aremet by testing the BREATHING TUBES (both heated and non‐heated)with therecommendedHUMIDIFIER.

Subclause 201.102.3 — BREATHING TUBES

BREATHINGTUBEShavebeenreported tocollapseonbending,occludeandperforatedue to theheat generated by HUMIDIFIERS and supplemental electrical heating. It is believed that aBREATHINGTUBEwhichistestedtomeettherequirementsofthisdocument,andwhichdoesnotkink, occlude and perforate during these tests, can provide BASIC SAFETY and ESSENTIALPERFORMANCEinclinicaluse.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 70: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

58 ©ISO2017–Allrightsreserved

Subclause 201.104.1 — General

The connections between a HUMIDIFIER and the source of breathing gas, and the HEATEDBREATHINGTUBECONTROLLERandaheaterwithinaBREATHINGTUBE,areFUNCTIONALCONNECTIONS.IntheeventofdisruptionoftheseFUNCTIONALCONNECTIONS,orfailureoftheconnectedequipment,theHUMIDIFIERisrequiredtomaintainESSENTIALPERFORMANCE,asindicatedinTable201.101.Fora CATEGORY 1 HUMIDIFIER, loss of HUMIDIFICATION OUTPUT is acceptable provided an ALARMCONDITION is created. For a CATEGORY 2 HUMIDIFIER, the ALARM CONDITION is not required. In allcases,BASICSAFETYisrequiredtobemaintained.

Subclause 201.104.2 — Connection to electronic health record

ElectronicdocumentationofPATIENTcareinterventionsisrapidlybecomingthestandardofcare.Theprimarymotivationsare to improve thequalityof care for an individual PATIENT throughaccurate and complete documentation, and to improve the completeness and accuracy ofaggregated data to facilitate continuous quality improvement. Providing remote supervisorycapabilityisrapidlybecomingthestandardofcareintheHOMEHEALTHCAREENVIRONMENT.

Subclause 201.104.3 — Connection to DISTRIBUTED ALARM SYSTEM

PATIENTS are not always located near enough to the OPERATOR to ensure that ALARM SIGNALScomingfromthePATIENT'Sroomcanbeheard.ItisreasonablyforeseeablethatsomeroomsofaPATIENT’Shome,limitedcarefacilityorhealthcarefacilityareoutofearshotofotherrooms.Asaresult,itisrecommendedforaHUMIDIFIERbeequippedwithameanstoconnecttoaDISTRIBUTEDALARMSYSTEMthatcanprovideadditionalALARMSIGNALpresentationpoints.ADISTRIBUTEDALARMSYSTEMfacilitatesdeliveryofALARMSIGNALStootherroomswheretheOPERATORmightbelocated,therebypermittingatimelyresponseandinterventiontosupportPATIENTcare.

Subclause 202.8.1.101 — Additional general requirements

It isnot the intentof the committee to require that the IMMUNITY testsbeperformedmultipletimes at several HUMIDIFICATION OUTPUTS, but that the MANUFACTURER should determine whichHUMIDIFICATION OUTPUT represents the worst case for a given IMMUNITY test and use thoseconditions.

Commercially available hygrometers can be used to monitor changes in the HUMIDIFICATIONOUTPUT as acceptance criteria following the specific tests required by this document todemonstratethemaintenanceofESSENTIALPERFORMANCE.

Subclause 208.6.8.4.101 — Additional requirements for termination of ALARM SIGNAL inactivation

Permitting very long pauses of ALARM SIGNALS can be hazardous for the PATIENT since theOPERATOR will not be notified of the existence of an ALARM CONDITION. However, PATIENTmanagement often requires PROCEDURES that can be disrupted by auditory ALARM SIGNALS.Therefore,extendingAUDIOPAUSEDbyOPERATORactionisusefultopreventtheHUMIDIFIERfromdisturbingtheOPERATORorothersinthevicinity.

HUMIDIFIERS shouldbeequippedwithan AUDIO PAUSED capability thatpermits the OPERATOR topausetheALARMSIGNALSpriortothecreationofanALARMCONDITION.SuchacapabilitypermitstheOPERATORtominimizenuisanceauditoryALARMSIGNALS.

Annex BB — Determination of the accuracy of the displayed MEASURED GAS TEMPERATURE

It is difficult to measure temperature at precisely the same location as the HUMIDIFIER’Stemperature sensor without modifying the gas flowrate, and hence, the thermal transfercharacteristicsfromthegastotheHUMIDIFIER’Stemperaturesensor.Theobjectofthetest isto

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 71: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved 59

measuretemperatureoneithersideoftheHUMIDIFIER’Stemperaturesensorandinterpolatethetemperaturetothesiteundertest.Temperaturedropinthecircuitcanbenonlinear.Hence,theobjective is toplace the standard temperature sensorsas closeaspossible to theHUMIDIFIER’Stemperaturesensor,butwithminimaldisruptionofgasflowpatterns.

Annex CC — Determination of the HUMIDIFICATION OUTPUT

TheHUMIDIFICATIONOUTPUTisdefinedasthemassofwatervapourperunitvolumeofgas(mg/l)at BTPS reference conditions, which is physiologically more appropriate than other referenceconditions.

Commercially available hygrometers do not have the necessary speed of response to provideconsistent and correct results when operated in the non‐isothermal environment of theBREATHINGSYSTEM.Therefore,theuseofsuchinstrumentsshouldberestrictedtoconstant‐flowhumidity measurements. However, the use of such instruments is suitable to VERIFY themaintenance of ESSENTIAL PERFORMANCE during all of the tests of this document except thedetermination of the HUMIDIFICATION OUTPUT required in 201.12.1.101 for ventilated gas flowconditions,whicharenotconstantandforwhichthetimeresponseoftheinstrumentistooslowtorespondtotherapidchangesintheventilatedflowwaveform.

The use of a dew point hygrometer requires a conversion from the specified units of mg/l(referenceBTPS)toequivalentvaluesfordewpoint.SeeTableAA.1.

Table AA.1 — Equivalent dew point for minimum required HUMIDIFICATION OUTPUT

HUMIDIFIER classification

Absolute humidity under BTPS conditions

mg/l

Equivalent dew point

°C

Category1 33 32,2

Category2 12 15,9

Annex DD — Specific enthalpy calculations

Formula (DD.2)

Thetotalenthalpycalculationcomesfromalogicalunderstandingofthefundamentalpropertiesof ideal gases. Amagat’s Law tells us that the total volume is the sum of the partial volumes.Therefore,thetotalenthalpyisthesumoftheenthalpiescalculatedforeachgascomponent(i.e.the sum of the dry gas enthalpy and the water vapour enthalpy). It is observed thatFormula(DD.2) has the same form as that of ISO8185[2], but with the result multiplied bydensity to give the desired volumetric reference units. This formula can also be found instandardreferencebooks[44].

Annex FF — Standard temperature sensor

The standard temperature sensor includes an additional copper thermal mass to effect anaveraging of temperature across the circuit,minimize effects of condensation forming on thesensor, reduce effects of precise positioning of the sensor, increase thermal transfer to thesensor,minimizethestemeffectandensureastabletemperaturemeasurement.

Themeasuringofanytemperatureinanon‐isothermalsituationisdifficult.Itisnotamatterofjust using a thermometer. Energy input, energy output, thermal conductivities all have to betakenintoaccount.Good,calibratedtemperaturemeasuringinstruments,whenusedtomeasureflowinggastemperatureinanarrowtube,canhavereadingsthatdifferbyupto10°Cfromeach

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 72: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

60 ©ISO2017–Allrightsreserved

other.Themethodprescribedinthisdocumentisanattempttolimitthevariability,anddefineastandardtemperature‐measuringmethod,toallowcomparisonandrepeatability.

Thermistors are chosen as the sensing element as they are small, readily available, have lowdrift,haveasignificantoutputsignaloverthetemperaturesofinterestandareeasilyinterfacedtosimplecircuits.

The particular values and dimensions of the thermistor are specified to match commonlyavailable commercial devices and are to standardize the thermal loss and stem effect of theconstructedsensor.

Whenplacingthethermistorsensorinameasuringcircuit,caremustbetakentoensurethatthecurrentuseddoesnotcausesignificantself‐heatingofthethermistor.

Agood41/2digit(orbetter)multimetercanbeusedtomeasuretheresistancedirectlybutcareneedstobetakentobesurethatthecurrentsthemetergeneratestomakesuchmeasurementsdonotcausesignificantself‐heating.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 73: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved 61

Annex BB (normative)

* Determination of the accuracy of the displayed MEASURED GAS

TEMPERATURE

BB.1 Test preparation

ConfirmtheaccuracyofthedisplayedMEASUREDGASTEMPERATUREbyintroducingtwostandardtemperature sensors, as defined in AnnexFF, into the HUMIDIFIER and ACCESSORIES that areconfiguredinaccordancewiththeACCOMPANYINGDOCUMENTS.TheconfigurationofthesensorsisshowninFigureBB.1.

If necessary, add extension tubing so that ambient drafts and temperatures do not undulyinfluencethesensors.Suchtubingshouldbeof thesamediameterastheBREATHINGTUBESandlongenoughso thatall sensorsare locatedatadistance fromambientdraftsequal toat least10timestheBREATHINGTUBEdiameter.

ThedistancesfromthenormallocationoftheHUMIDIFIER'Stemperaturesensortothelocationofthestandard temperaturesensors (distance l inFigureBB.1)shallbeequalandshallmeasurefrom20mmto30mm.

Key

1 normallocation

2 seal

3 standardtemperaturesensor

Figure BB.1 — Configuration for displayed temperature accuracy test

BB.2 Test PROCEDURE

Carry out testing as follows.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 74: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

62 ©ISO2017–Allrightsreserved

a) Sample the temperature at least every 2 s.

b) Operate the HUMIDIFIER over the flowrate range indicated in the ACCOMPANYING DOCUMENTS.

c) Set the minimum SET TEMPERATURE and confirm that the MEASURED GAS TEMPERATURE is within ±2 °C of the arithmetic mean of the two standard temperature sensors in a steady-state condition.

d) Change the SET TEMPERATURE from the minimum to the maximum setting.

e) Confirm that the DISPLAYED GAS TEMPERATURE is within ±2 °C of the arithmetic mean of the two standard temperature sensors in a steady-state condition for the maximum SET TEMPERATURE.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 75: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved 63

Annex CC(normative)

* Determination of the HUMIDIFICATION OUTPUT

CC.1 Principle

The HUMIDIFICATION OUTPUT is measured using the gravimetric method, which is a simpletechnique.

This testmethodusesdry input air, so anymoisture in theoutput airhasbeen addedby theHUMIDIFIER.Thevolumeofoutputair isnormalized to thevolume itwouldoccupyunderBTPSconditions.

CC.2 Test conditions

There are several practical considerations required to perform this test method in order tocalculate the HUMIDIFICATION OUTPUT. As several different definitions of “standard conditions”exist,itisimportanttoknowthereferenceconditionsfortheflowmetercalibration.

TableCC.1providesthemostcommonlyencounteredreferenceconditionsandacorrespondingcorrectionfactortoscalethevolumetricflowratereadingreferencedtostandardconditionstoavolumetric flowrate referenced to BTPS conditions as required to perform the gravimetriccalculation.

Table CC.1 — Common reference conditions and their correction factors

Publishing entity

Standard conditions standard l/min

Scaling factor to BTPSl/min

NIST,ISO10780 0°C,101,325kPa 1,210

EPA,NIST 20°C,101,325kPa 1,127

AMCA 21,1°C,101,325kPa 1,123

NIST(USNationalInstituteofStandardsandTechnology)

EPA(USEnvironmentalProtectionAgency)

AMCA(AirMovementandControlAssociationInternational)

For amore precise conversion to BTPS, if independent accuratemeans ofmeasuring ambientpressureandtemperatureareavailable,theconversionfromVATPtoVBTPSiscalculatedusingthefollowingFormula(CC.1)basedontheIdealGasLawassumingthattheinputgasisdry:

body amb

BTPS ATPamb amb v body

T PV V

T P P T

(CC.1)

where

Pamb istheambientpressure;

Pv(Tbody) isthesaturationvapourpressureatTbody (inK)ascalculatedusingFormula (CC.2);

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 76: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

64 ©ISO2017–Allrightsreserved

VATP isthevolumeatATP(ambienttemperature,pressure)conditions;

Tbody is310°K;

Tamb isthetemperatureofthedeliveredgas.

Pv(Tbody)=

3h h

h

3 142,3130,590 51 8,2lg 2,480 4 10

10T T

T

(CC.2)

NOTE WhenTambisequaltoTbody,Pv(Tbody)equals6,270987kPa.

CC.3 Apparatus

The following items shall be utilizedwithmeasurement apparatus and for sufficient durationsuchthatatotalmeasurementaccuracyof±1mg/lisachieved:

a) theHUMIDIFIERundertestwithitsACCESSORIESasindicatedintheinstructionsforuse;

b) aweightscale;

c) atimer;

d) aflowmeter;

NOTE1 Most readily available commercial air flowmeters are calibrated for dry gas and it isthereforedesirabletomeasurethegasflowonthedryside(i.e.theinputtotheHUMIDIFIER).

e) astandardtemperaturesensorasdefinedinAnnexFF.

f) agassource

NOTE2 Additionalinformationisprovidedin201.5.

CC.4 PROCEDURE

ThegravimetricmethodHUMIDIFICATIONOUTPUTisdeterminedasfollows.

a) Configure the HUMIDIFIER with its ACCESSORIES as indicated in the instructions for use.

b) Install the standard temperature sensor at a site representing the DELIVERED GAS TEMPERATURE. Call this temperature T2 (°C).

c) If necessary, add extension tubing such that ambient drafts and temperatures do not unduly influence sensors. Ensure that the extension tubing is of equal diameter to the BREATHING TUBE and of length such that all sensors are located at a distance from ambient drafts equal to at least 10 times the BREATHING TUBE diameter.

d) Confirm that the ambient temperature of the HUMIDIFIER and ACCESSORIES are within the range of 17 °C to 23 °C. Measure the input gas temperature and call this temperature T1 (°C). Confirm that the inlet gas temperature is within the RATED ambient temperature range of the HUMIDIFIER.

e) Arrange relative elevations of the HUMIDIFIER, BREATHING TUBE and HUMIDIFICATION CHAMBER, as applicable, such that:

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 77: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved 65

— Condensation that does not represent humidification reaching the PATIENT (e.g. condensation in the inspiratory limb) does not leave the humidification system and is included in m1 (the mass of the HUMIDIFIER), its contents and the recommended BREATHING TUBE as defined in h).

— Condensation that represents humidification reaching the PATIENT (e.g. condensation occurring in the instrumental dead space) leaves the HUMIDIFIER and is not included in m1 [see h)].

f) Connect the HUMIDIFIER to the gas source.

g) Set the gas flowrate with the flowmeter to the maximum RATED flowrate ±5 % and turn the HUMIDIFIER on at its maximum output setting. Confirm that at least 2 h have elapsed to ensure that the error due to the warm-up time of the HUMIDIFIER is minimized.

h) Turn off the HUMIDIFIER. Disconnect all ACCESSORIES, including the gas source, electrical connections and any extension tubing to remove any extraneous influences on the mass measurement. Weigh only the HUMIDIFIER, its contents and the recommended BREATHING TUBE; record this mass as m0. This is the initial mass of the HUMIDIFIER.

i) Reconnect all ACCESSORIES.

j) Turn the HUMIDIFIER on and start the timer.

k) Record the start time as t0 and maintain OPERATOR control settings throughout the test.

l) Stop the test when the measurement of the following quantities maintains a total measurement accuracy of ±1 mg/l:

— the HUMIDIFIER has used a sufficient quantity of the usable capacity of the LIQUID CONTAINER;

— the test is of sufficient duration.

m) Record the time as t1 and record the duration of the test (t1 - t0).

n) Note the total elapsed time and convert to suitable time units with respect to the flowmeter reading.

NOTE Special attention is drawn to the objective of a total measurement error of less than 1 mg/l. Measurement of time, temperature, especially flowrate and mass used, should be sufficiently accurate relative to the value of the quantity to maintain this objective. In practice, the mass and estimated output of the HUMIDIFIER gives a guide as to the minimum duration of the test to maintain overall accuracy. An error analysis of the measurement apparatus and estimated results is strongly recommended as a guide.

o) Weigh the HUMIDIFIER at the end, and record the difference with respect to the starting mass.

p) Calculate the volume of gas humidified, VBTPS, (referenced to BTPS) during the test using Formula (CC.3):

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 78: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

66 ©ISO2017–Allrightsreserved

VBTPS=qSLPM×SF×TDUR (CC.3)

where

qSLPM istheflowrateofgasreferencedtostandardconditions;

SF isthescalingfactortoBTPS (fromTable CC.1);

TDUR isthedurationofthetest.

q) Calculate HUMIDIFICATION OUTPUT, Hout, (mg/l) using Formula (CC.4).

outBTPS

mH

V

(CC.4)

where

Δm isthemasschangeoftheHUMIDIFIER (inmg).

r) Confirm that HUMIDIFICATION OUTPUT exceeds the value indicated in the instructions for use.

s) Repeat e) to r)with the ambient temperature at the HUMIDIFIER’S minimum RATED operating temperature ±1 °C.

t) Repeate) to r)with the ambient temperature at the HUMIDIFIER’S maximum RATED operating temperature ±1 °C.

u) Repeatd)tot)at the HUMIDIFIER’S minimum RATED flowrate.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 79: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved 67

Annex DD(normative)

* Specific enthalpy calculations

DD.1 Calculation of specific enthalpy

Thespecificenthalpy,ΔHs,ofamoistgas is theenergyreleasedbythegas,perunitvolumeofdrygas,whencooledorheatedataconstantpressureto0°C,inclusiveofanycontributionfromwatervapourandcondensation,butexclusiveofanycontributionfromwaterinitiallyinaliquidform(suchaswaterdroplets).Thevolumereferredto istheinitialvolumeofthedrygas,andthepreferredunitsofmeasurementforΔHsarekJm‐3.

TableDD.1listssymbolsandvaluesforpropertiesusedinthecalculationofspecificenthalpy.

Table DD.1 — Symbols and values for the calculation of specific enthalpy

Symbol Value Description Reference

Mwv 18,0153gmol−1 Molecularweightofwatervapour [45]

Mair 28,9645gmol−1 Molecularweightofdryair [45]

cp,air 1,0051kJkg−1K−1 Constantpressurespecificheatcapacityofdryair [46][47]

cp,wv 1,8662kJkg−1K−1 Constantpressurespecificheatcapacityofwatervapour

[46][47]

Lwv 2501kJkg−1 Latentheatofvaporizationofwaterat0°C [48]

R 8,31446JK−1mol−1 Universalgasconstant —

Formulaeforthesaturationvapourpressure,Psat,forwateraregiveninAnnexGG.Thespecifichumidityofagas,hs(Td,P),(massofwatervapourperunitmassofdryair), iscalculatedusingFormula(DD.1):

sat dwvs d

air sat d

,P TM

h T PM P P T

(DD.1)

DD.2 Specific enthalpy measurement

Tocalculatethespecificenthalpyofagas,thetemperature,T in°C,pressure,P inPa,anddewpoint temperature,Td in °C, needs to be known. The specific enthalpy limit in this documentreferstothespecificenthalpyofthedeliveredgas.Thus,thePROCEDUREoutlinedbelowisusedtomeasure the specific enthalpy. Deviations may be made from this PROCEDURE if justifiedappropriately.

a) With a sampling period of no longer than 2 s, measure:

1) the DELIVERED GAS TEMPERATURE, T (i.e. at the PATIENT-CONNECTION PORT);

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 80: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

68 ©ISO2017–Allrightsreserved

2) the temperature of the gas at the HUMIDIFICATION CHAMBER outlet, Th, and take Td as the minimum of T and Th;

3) the pressure of the delivered gas, or an appropriately assumed value (such as 101 325 Pa at sea level).

b) Calculate the specific enthalpy using Formula (DD.2) and Formula (DD.3) below.

c) Filter the specific enthalpy using a moving average with a 120 s window period.

d) Confirm that the filtered specific enthalpy does not exceed 197 kJ/m3 at any time.

For20°C≤Td≤60°C and20°C≤T≤80°C, the specific enthalpy ofmoist air, ΔHs, is givenbyFormula(DD.2):

ΔHs=ρair[cp,airT+hs(Td,P)(Lwv+cp,wvT)] (*DD.2)

where

air sat dair 273,15

M P P T

R T

(DD.3)

Formula(DD.3)dividesthecomputationintotwoparts:thecontributiontotheenthalpyduetothedryair,andduetothewatervapour.Theenthalpyisfirstcalculatedperunitmassofdryair,andthenscaledbythedensityofair(ρair)tobeperunitvolumeofdryair.Thecontributionduetothedryairissimplycp,airT.Thecontributionduetothewatervapourisfoundbymultiplyingthe specific humidity (kg ofwater vapourper kg of dry air) by the specific enthalpyper unitmassofwatervapour,whichisapproximatedbyLwv+cp,wvT.

Ifthedeliveredgasisnotmoistair(e.g.humidifiedoxygen),appropriatevaluesforMairandcp,airaresubstituted(e.g.MO2andcp,O2).

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 81: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved 69

Annex EE(normative)

Removable temperature sensors and mating ports

EE.1 General

InteroperabletemperaturesensorsandtheirmatingportsareneededtoensurethatBREATHINGSYSTEMSandHUMIDIFERSarecompatible.ForPATIENTsafety,thetemperaturesensorneedstobesecurelyretainedinacompatiblematingportonceinserted.

EE.2 Dimensional requirements for removable temperature sensors

a) Removabletemperaturesensorsshall

1) meetthedimensionalrequirementsofFigureEE.1,or

2) besufficientlydifferentthattheycannotbeinterchangedwiththosethatdomeettheserequirements.

b) Temperaturesensormatingportsshall

3) meetthedimensionalrequirementsofFigureEE.2,or

4) shallbesufficientlydifferentthattheycannotbeinterchangedwiththosethatdomeettheserequirements.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 82: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

70 ©ISO2017–Allrightsreserved

Dimensionsinmillimetres,unlessotherwiseindicated

NOTE TableEE.1containsthedimensionsforthisfigure.

Figure EE.1 — Removable temperature sensor port

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 83: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved 71

Table EE.1 — Removable temperature sensor port dimensions

Dimensionsinmillimetres,unlessotherwiseindicated

Removable temperature sensor port

Reference Designation Dimension

Minimum Nominal Maximum

(A)Conical taper of 1:40 included angle(degrees,reference) (1,43)

ØDInsidediameterattheopeningofthefemaletaper 7,83 7,87 7,91

E Depthoffemaletaper 5,00 10,00 12,00

ØGInside diameter of the smaller end of thefemale taperat4,3mmfromtheopeningofthefemaletaper

7,72 7,76 7,80

ØW

Diameter of the smallest cylinder thatencompasses the outside surfaces of theexternal featuresof the temperature sensorport

9,07 9,37 9,67

R Radius or chamfer at the entrance of thefemaletaper 0,00 0,25 0,50

NOTE TableEE.2containsthedimensionsforthisfigure.

Figure EE.2 — Removable temperature sensor dimensions

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 84: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

72 ©ISO2017–Allrightsreserved

Table EE.2 — Removable temperature sensor dimensions

Dimensionsinmillimetres,unlessotherwiseindicated

Removable temperature sensor

Reference Designation Dimension

Minimum Nominal Maximum

(a) Conicaltaperof1:40includedangle(degrees,reference) (1,43°)

Ød Diameteratlargeendofthemaletaper 7,89 7,93 7,97

e Depthofmaletaper 10,50 — —

Øg Diameteratsmallendofthemaletaper 7,56 7,60 7,64

r Radiusorchamferatthesmallendofthemaletaper

0,00 0,25 0,50

EE.3 Test method for security of engagement of removable temperature sensors to mating ports

Ifataperedconicalconnectorisusedfortheremovabletemperaturesensor,thenitshallcomplywiththefollowingseparationforcerequirement.

Carry out security of engagement testing as follows.

a) Condition removable temperature sensors and mating ports at (41 ± 2) °C and (95 ± 5) % RH for at least 1 h.

b) Engage the removable temperature sensor with the mating port in accordance with the ACCOMPANYING DOCUMENTS.

c) Condition the engaged components, without activation of any disengagement mechanism, for at least 1 h at the conditions specified in a).

d) Apply an axial separation force of (25 ± 2,5) N for 10 s at a rate not exceeding 20 N/s.

e) Confirm that sensor remains engaged.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 85: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved 73

Annex FF(normative)

* Standard temperature sensor

FF.1 Test preparation

a) The standard temperature sensor shall meet the requirements indicated below. See alsoFigureFF.1.

b) Aproperly assembled standard temperature sensor shall have the followingperformancecharacteristics:

1) timeconstant>0,5sand<1,0sforastepchangeof22°Cto37°Cinwaterataflowrateof1m/s,and

NOTE1 Thisplacesboundsonthethermalconductivityoftheconstructedsensor.

2) influencedbychangesinambienttemperatureof<0,01°Cper1°C.

NOTE2 Thisplacesaboundonthestemeffectoftheconstructedsensor.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 86: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

74 ©ISO2017–Allrightsreserved

a) Sheath for standard temperature sensor

b) Standard temperature sensor construction

Key

1 electricalinsulationonthermistorleads

2 thermallyconductivepasteorepoxy

Figure FF.1 — Standard temperature sensor

Anexampleoftheconstructionofsuchasensorisasfollows.

— A 6mm length of 3,18mm outside diameter copper tube. The inside diameter should bechosen to create sufficient clearance on the thermistor bead. Any other material withthermalconductivity>386W/(m·K)withthedimensionsgiveninFigureFF.1isacceptable.Athermistorwithanaccuracyof±0,2°Cfrom25°Cto45°C(wheninisothermalstirredairorwater).Thesupplierstypicallyhavethermistorsof2252Ohmresistanceat25°C.Thisisasuitablevaluefortheaccuracythetestrequires.

— Theelectricalleadsofthethermistorshouldbe#32gaugetinnedcopperwire,aminimumof60mmlong(anyothermaterialwithconductivity<180W/(mK)isacceptable.

— The electrical leads of the thermistor shouldbe sleeved to achieve electrical and thermalinsulation. Plastic tubing achieves both aims, but any electrically insulatingmaterialwiththermalconductivity<0,02W/(m·K)isacceptable.

— Thethermistorshouldbecentralizedinthecoppertube,andthevoidsfilledbyathermallyconductive but electrically insulating compound (such as thermal heatsink paste orthermallyconductivelow‐stressepoxy)havingathermalconductivity>0,183W/(m·K).

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 87: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved 75

NOTE1 The copper sleeve is most easily manufactured from standard copper tube, diameter3,18mm (1/8in). This gives a standard surface area and surface thermal conductivity to theconstructedsensor.

NOTE2 Thetemperaturebeadspecifiedisreadilyavailablefrommanysources.3

NOTE3 Whenplacingthethermistorsensorinameasuringcircuit,careneedsbetakentoensurethatthemeasuringcurrentuseddoesnotcausesignificantself‐heatingofthethermistor.

3Complianttemperaturebeads(thermistors)arecommerciallyavailablefromsourcesincluding:Honeywell,Yellow

Springs Instruments (YSI), Fenwal and Omega. This information is given for the convenience of users of thisdocumentanddoesnotconstituteanendorsementbyISOorIECoftheseproducts.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 88: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

76 ©ISO2017–Allrightsreserved

Annex GG (informative)

Saturation vapour pressure

GG.1 Test preparation

ThesaturationvapourpressureandenhancementfactorforwaterareusedinthecalculationofspecificenthalpyinAnnexDD.Thesaturationvapourpressure,Psat,andenhancementfactor,f,forwateraregivenbyFormula(GG.1)andFormula(GG.2)respectively[49].

18,678

234,5 257,154sat 611,21

T T

TP T e

(GG.1)

f(T,P)=1,00072+(P)[3,2·10−8+(5,9·10−12)(T2)] (GG.2)

where

T isthetemperatureofthegasin°C;

P istheatmosphericpressureinPa.

Touse theenhancement factor,multiplyFormula(GG.1)byFormula(GG.2) forallenthalpyorhumiditycalculationsusingPsat(seeAnnexCCandAnnexDD).

The coefficients for Formula(GG.1) (saturation vapour pressure) and Formula(GG.2)(enhancementfactorforwater)weretakenfromReference[50].

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 89: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved 77

Annex HH (informative)

Reference to the essential principles of safety and performance of

medical devices in accordance with ISO 16142-1:2016[7]

ThisdocumenthasbeenpreparedtosupporttheessentialprinciplesofsafetyandperformanceofaCLINICALTHERMOMETERasamedicaldeviceaccordingtoISO16142‐1:2016[7].Thisdocumentisintendedtobeacceptableforconformityassessmentpurposes.

Compliancewith this document provides onemeans of demonstrating conformancewith thespecificessentialprinciplesofISO16142‐1:2016[7].Othermeansarepossible.TableHH.1mapstheclausesandsubclausesofthisdocumentwiththeessentialprinciplesofISO16142‐1:2016.

Table HH.1 — Correspondence between the essential principles and this document

Essential principle ofISO 16142-1:2016[7]

Corresponding clause(s)/subclause(s) of this

document

Qualifying remarks/Notes

1 all Thepartrelatingtomanufacturingisnotaddressed.

a) 206

b) 206

2 Thepartrelatingtomanufacturingisnotaddressed.

a) all

b) 201.4 Thepartrelatingtomanufacturingisnotaddressed.

c) 201.1.3,201.7

d) 201.7

3 all Thepartrelatingtomanufacturingisnotaddressed.

4 all

5 201.1.3,201.4,201.15

6 201.4

8.1 —

a) 201.11

b) 201.4,201.11

c) 201.9,201.15

8.2 201.11.7

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 90: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

78 ©ISO2017–Allrightsreserved

Table HH.1 (continued)

Essential principle ofISO 16142-1:2016[7]

Corresponding clause(s)/subclause(s) of this

document

Qualifying remarks/Notes

8.3 201.11

8.4 201.11.7

8.5 201.11.6.2 Thepartrelatingtoingressisnotaddressed.

9.1 201.11.6.6

a) 201.11.6.6

b) 201.11.6.6

c) 201.11.6.6

12.1 201.7,201.14,201.16,201.101,201.102,201.104

12.2 —

a) 206

b) 206

c) 201.5,202,211

d) 201.11

e) 201.14,201.16

f) 201.11

g) 202

12.4 201.11

12.5 201.7,201.8

12.6 201.1.3

13.1 201.12.1

13.2 201.7

13.3 201.7,201.12.1,206,208

13.4 201.7

15.1 201.14

15.2 201.14

16.1 201.4,201.13

16.5 202

16.6 202

16.7 201.8

17.1 201.9,211

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 91: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved 79

Table HH.1 (continued)

Essential principle ofISO 16142-1:2016[7]

Corresponding clause(s)/subclause(s) of this

document

Qualifying remarks/Notes

17.2 201.9

17.3 201.9

17.4 201.8,201.15,201.101

17.5 201.15,201.101,201.102,201.103

17.6 201.11

18.1 201.12

18.2 201.12,201.13.1.101,201.13.102

19.1 201.7

19.2 201.7,206,211

20.1 211

20.2 206

20.3 201.7.9.2.8.101

21.1 201.7

21.2 201.7

21.3 201.7

21.4 201.7.2.3,201.7.2.13.101,201.7.2.17.101,201.8,201.9,

201.11.7

21.5 —

a) 201.7 Thepartrelatingtoauthorizedrepresentativeisnotaddressed.

b) 201.7,201.7.2.17.101a)

f) 201.7

i) 201.7.2.101b),211

j) 201.7.2.101b),211

k) 201.7.2.101c)

l) 201.7

m) 201.7

21.6 201.7

21.7 —

a) 201.7.9.1

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 92: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

80 ©ISO2017–Allrightsreserved

Table HH.1 (continued)

Essential principle ofISO 16142-1:2016[7]

Corresponding clause(s)/subclause(s) of this

document

Qualifying remarks/Notes

b) 201.7

d) 201.7.2.17.101,201.7.9.2.1.102b)

g) 201.7

h) 201.7

i) 201.7.9.2.2

k) 201.7.9.2.9.101.2g),201.7.9.2.9.101.2h),201.7.9.2.9.101.2j),

201.7.9.2.14.101,201.102

l) 201.7,201.7.9.2.13.101

n) 201.7.9.2.2

p) 201.7.9.2.12,201.11.6.6,211

q) 201.7

21.8 201.7

21.9 211

a) 211

b) 211

c) 211

d) 211

e) 211

f) 211

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 93: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved 81

Annex II (informative)

Terminology — Alphabetized index of defined terms

NOTE TheISOOnlineBrowsingPlatform(OBP)4andtheIECElectropedia5provideaccesstomanyofthesetermsanddefinitions.

Term Source

ACCESSIBLEPART IEC60601–1:2005+AMD1:2012,3.2

ACCESSORY IEC60601‐1:2005+AMD1:2012,3.3

ACCOMPANYINGDOCUMENT IEC60601‐1:2005+AMD1:2012,3.4

ACTIVEHEATANDMOISTUREEXCHANGER 201.3.201

ACTIVEHME 201.3.201

AIRWAYPRESSURE(Paw) 201.3.202

ALARMCONDITION IEC60601–1:2005+AMD1:2012,3.141

ALARMLIMIT IEC60601–1‐8:2006,3.3

ALARMPAUSED IEC60601–1‐8:2005,3.5

ALARMSIGNAL IEC60601–1‐8:2005,3.9

ALARMSYSTEM IEC60601–1‐8:2006,3.11

APPLIEDPART IEC60601–1:2005,3.8

AUDIOPAUSED IEC60601–1‐8:2006,3.13

BASICSAFETY IEC60601–1:2005,3.10

BODYTEMPERATUREPRESSURE,SATURATED 201.3.203

BREATHINGSYSTEM ISO4135:2001,4.1.1[1]

BREATHINGSYSTEMFILTER ISO23328‐2:2002,3.1[9]

BREATHINGTUBE ISO4135:2001,4.1.2[1]

BTPS 201.3.203

CLEARLYLEGIBLE IEC60601–1:2005+AMD1:2012,3.15

aTobepublished.

4Availableathttps://www.iso.org/obp/ui/#home.

5Availableathttp://www.electropedia.org.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 94: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

82 ©ISO2017–Allrightsreserved

Term Source

CONTINUOUSOPERATION IEC60601–1:2005,3.18

DELIVEREDGASTEMPERATURE 201.3.204

DISTRIBUTEDALARMSYSTEM IEC60601–1‐8:2006,3.17

ENCLOSURE IEC60601–1:2005,3.26

ESSENTIALPERFORMANCE IEC60601–1:2005+AMD1:2012,3.27

F‐TYPEAPPLIEDPART IEC60601–1:2005,3.29

FUNCTIONALCONNECTION IEC60601–1:2005,3.33

FLOW‐DIRECTION‐SENSITIVECOMPONENT 201.3.205

GASPATHWAY ISO18562‐1:2016a,3.5

HARM IEC60601–1:2005,3.38

HAZARD IEC60601–1:2005+AMD1:2012,3.39

HAZARDOUSSITUATION IEC60601–1:2005+AMD1:2012,3.40

HEATANDMOISTUREEXCHANGER ISO9360‐1:2000,3.1[5]

HEATEDBREATHINGTUBECONTROLLER 201.3.206

HME ISO9360‐1:2000,3.1[5]

HOMEHEALTHCAREENVIRONMENT IEC60601–1‐11:2015,3.1

HUMIDIFICATIONCHAMBER 201.3.207

HUMIDIFICATIONOUTPUT 201.3.208

HUMIDIFIER 201.3.209

IMMUNITY IEC60601–1‐2:2014,3.8

INTENDEDUSE IEC60601–1:2005+AMD1:2012,3.44

INTERNALELECTRICALPOWERSOURCE IEC60601–1:2005,3.45

LAY IEC60601–1‐11:2015,3.2

LEAKAGECURRENT IEC60601–1:2005,3.47

LIQUIDCONTAINER 201.3.210

LIQUIDRESERVOIR 201.3.211

MANUFACTURER IEC60601–1:2005+AMD1:2012,3.55

MAXIMUMLIMITEDPRESSURE(PLIMmax) 201.3.212

MAXIMUMOPERATINGPRESSURE 201.3.213

MEEQUIPMENT IEC60601–1:2005,3.63

MESYSTEM IEC60601–1:2005,3.64

aTobepublished.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 95: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved 83

Term Source

MEDICALELECTRICALEQUIPMENT IEC60601–1:2005,3.63

MEDICALELECTRICALSYSTEM IEC60601–1:2005,3.64

MEANSOFPROTECTION(MOP) IEC60601–1:2005,3.60

MEASUREDGASTEMPERATURE 201.3.214

MEDICALGASPIPELINESYSTEM ISO7396‐1:2016,3.29

MEDIUMPRIORITY IEC60601–1‐8:2006,3.28

MOBILE IEC60601–1:2005+AMD1:2012,3.65

MODELORTYPEREFERENCE IEC60601–1:2005,3.66

MONITORINGEQUIPMENT 201.3.215

NOMINAL(VALUE) IEC60601–1:2005,3.69

NORMALCONDITION IEC60601–1:2005,3.70

NORMALUSE IEC60601–1:2005+AMD1:2012,3.71

OPERATOR IEC60601–1:2005,3.73

PATIENT IEC60601–1:2005+AMD1:2012,3.76

PATIENTCONNECTION IEC60601–1:2005,3.78

PATIENT‐CONNECTIONPORT 201.3.216

PATIENTLEAKAGECURRENT IEC60601–1:2005,3.80

PORTABLE IEC60601–1:2005+AMD1:2012,3.85

PEMS IEC60601–1:2005,3.90

PESS IEC60601–1:2005,3.90

PRIMARYOPERATINGFUNCTION IEC62366–1:2015,3.13[18]

PROCEDURE IEC60601–1:2005+AMD1:2012,3.88

PROCESS IEC60601–1:2005+AMD1:2012,3.89

PROGRAMMABLEELECTRICALMEDICALSYSTEM IEC60601–1:2005,3.90

PROGRAMMABLEELECTRONICSUBSYSTEM IEC60601–1:2005,3.90

PROTECTIONDEVICE 201.3.217

RATED(VALUE) IEC60601–1:2005,3.97

RELATIVEHUMIDITY 201.3.218

RESIDUALRISK IEC60601–1:2005+AMD1:2012,3.100

RESPONSIBLEORGANIZATION IEC60601–1:2005,3.101

RISK IEC60601–1:2005+AMD1:2012,3.102

aTobepublished.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 96: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

84 ©ISO2017–Allrightsreserved

Term Source

RISKCONTROL IEC60601–1:2005+AMD1:2012,3.105

RISKMANAGEMENT IEC60601–1:2005+AMD1:2012,3.107

RISKMANAGEMENTFILE IEC60601–1:2005+AMD1:2012,3.108

SETTEMPERATURE 201.3.219

SINGLEFAULTCONDITION IEC60601–1:2005+AMD1:2012,3.116

STANDARDTEMPERATUREANDPRESSURE,DRY 201.3.220

STPD 201.3.220

TOOL IEC60601–1:2005,3.127

TRANSIT‐OPERABLE IEC60601–1‐11:2015,3.4

USABILITY IEC62366–1:2015,3.16[18]

USABILITYENGINEERINGFILE IEC62366–1:2015,3.18[18]

USESPECIFICATION IEC62366–1:2015,3.23[18]

VALIDATION(VALIDATED) 201.3.221

VERIFICATION(VERIFIED) IEC60601–1:2005+AMD1:2012,3.138

aTobepublished.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 97: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved 85

Bibliography

[1] ISO4135:2001,Anaesthetic and respiratory equipment — Vocabulary

[2] ISO8185:2007,Respiratory tract humidifiers for medical use — Particular requirements for respiratory humidification systems

[3] ISO7000:2014,Graphical symbols for use on equipment — Registered symbols

[4] ISO9000:2015,Quality management systems — Fundamentals and vocabulary

[5] ISO9360‐1:2000,Anaesthetic and respiratory equipment — Heat and moisture exchangers (HMEs) for humidifying respired gases in humans — Part 1: HMEs for use with minimum tidal volumes of 250 ml

[6] ISO9360‐2:2001,Anaesthetic and respiratory equipment — Heat and moisture exchangers (HMEs) for humidifying respired gases in humans — Part 2: HMEs for use with tracheostomized patients having minimum tidal volumes of 250 ml

[7] ISO16142‐1:2016, Medical devices — Recognized essential principles of safety and performance of medical devices — Part 1: General essential principles and additional specific essential principles for all non-IVD medical devices and guidance on the selection of standards

[8] ISO20789:—6,Anaesthetic and respiratory equipment – Passive humidifiers

[9] ISO23328‐2:2002,Breathing system filters for anaesthetic and respiratory use — Part 2: Non-filtration aspects

[10] ISO27427:2013, Anaesthetic and respiratory equipment — Nebulizing systems and components

[11] ISO80369‐27, Small-bore connectors for liquids and gases in healthcare applications — Part 2: Connectors for breathing systems and driving gases applications

[12] ISO80601‐2‐12:2011,Medical electrical equipment — Part 2-12: Particular requirements for basic safety and essential performance of critical care ventilators

[13] ISO80601‐2‐70:2015,Medical electrical equipment — Part 2-70: Particular requirements for basic safety and essential performance of sleep apnoea breathing therapy equipment

6Tobepublished.

7Tobepublished.

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 98: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

86 ©ISO2017–Allrightsreserved

[14] ISO80601‐2‐72:2015,Medical electrical equipment — Part 2-72: Particular requirements for basic safety and essential performance of home healthcare environment ventilators for ventilator-dependent patients

[15] IEC60601‐1‐3,Medical electrical equipment – Part 1-3: General requirements for basic safety and essential performance – Collateral Standard: Radiation protection in diagnostic X-ray equipment

[16] IEC/TR60878:2015,Graphical symbols for electrical equipment in medical practice

[17] IEC62366:2014+AMD1:2015,Medical devices — Application of usability engineering to medical devices

[18] IEC62366‐1:2015, Medical devices — Part 1: Application of usability engineering to medical devices

[19] DOYLEA., JOSHIM., FRANKP., CRAVENT.,MOONDIP., YOUNGP.A change inhumidificationsystemcaneliminateendotrachealtubeocclusion.J. Crit. Care.2011,(Mar)p.23

[20] MIYAOH.,HIROKAWAT.,MIYASAKAK.,KAWAZOET.Relativehumidity,notabsolutehumidity,isofgreatimportancewhenusingahumidifierwithaheatingwire.Crit. Care Med.1996,20(5)pp.674–679

[21] BURTONG.W.,FOXD.E.R.Anairwayresistance foruse inanartificial lung.Br. J. Anaesth.1972,44pp.1253–1262

[22] MECKLENBURGHJ.S.Constructionoflinearresistanceunitsforamodellung.Med and Biol Elect and Comp.1988,26pp.552–554

[23] MUSHINW.W.,RENDELL‐BAKERL.,THOMPSONP.W.,MAPLESONW.W.Automatic Ventilation of the Lungs.ThirdEdition,1980,pp.182–3.

[24] WILLIAMSR.,RANKINN.,SMITHT.,GALLERD.,SEAKINSP.Relationshipbetweenthehumidityand temperatureof inspiredgasand the functionof theairwaymucosa.Crit. Care Med.1996,24(11)pp.1920–1929

[25] KONRAD F., SCHREIBER T., BRECHT‐KRAUSD.Mucociliary Transport in ICU patients.Chest.1994,105(1)pp.237–241

[26] MASSIEC.,HARTR.,PERALEZK.,RICHARDSG.Effectsofhumidificationonnasalsymptomsandcomplianceinsleepapneapatientsusingcontinuouspositiveairwaypressure.Chest.•••,116pp.403–408

[27] PRIMIANO F., SAIDE G., MONTAGUE F., KUSE K., GREEN C., HOROWITZ J. Water vapour andemperaturedynamicsintheupperairwaysofnormalandCFsubjects.Eur. Respir. J.1968,1pp.401–414

[28] WIESTG.,LEHNERTG.,BRUCKW.,MEYERM.,HAHNE.,FICKERJ.Aheatedhumidifierreducesupperairwaydrynessduringcontinuouspositiveairwaypressuretherapy.Respir. Med.1999,93pp.21–26

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 99: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

©ISO2017–Allrightsreserved 87

[29] KOUTSOURELAKISI.,VAGIAKISE.,PERRAKIE.,KARATZAM.,MAGKOUC.,KOPAKAM.etal.Nasalinflammation in sleep apnoea patients using CPAP and effect of heated humidification.Eur. Respir. J.2011,37pp.587–594

[30] CONSTANTINDISJ.,KNOBBERD.,STEINHARTH.,KUHNJ.,IROH.Fine‐structuralinvestigationsofthe effect odnCPAP‐mask applicationon thensalmucosa.Acta Otolaryngol. 2000,120pp.432–437

[31] RICHARDSG.,CISTULLIP.,UNGARR.,BERTHON‐JONESM., SULLIVANC.Mouth leakwithnasalcontinuouspositiveairwaypressureincreasesnasalairwayresistance.Am. J. Respir. Crit. Care Med.1996,154pp.182–186

[32] MARTINDEARAUJOM.,VIERAS.,VASQUEZE.,FLEURYB.Heatedhumidificationorfacemaskto prevent upper airway dryness during continuous positive airway pressure therapy.Chest.2000,117pp.142–147

[33] OTO J., NAKATAKI E., OKUDA N., ONODERA M., IMANAKA H., NISHIMURA M. Hygrometricproperties of inspired gas and oral dryness in patients with acute respiratory failureduringnoninvasiveventilation.Respir. Care.2014,39(45)pp.39–45

[34] LELLOUCHE F.,MAGGIOREA., LYAZIDI A., DEYEN., TAILLE S., BROCHARD L.Water content ofdeliveredgasesduringnon‐invasive ventilation inhealthy subjects. Intensive Care Med.2009,37(3)pp.939–945

[35] ORECR.etal.Mechanismsofbacterialmovementinventilatorcircuits.Anaesth. Intensive Care.1997,25(5)p.568

[36] RAKOTONANHARY D., PELLETIER‐FLEURY N., GAGNADOUX F., FLEURY B. Predictive factors fortheneedforadditionalhumidificationduringnasalcontinuouspositiveairwaypressuretherapy.Chest.2001,119pp.460–465

[37] RANDERRATHW.,MEIERJ.,GENGERH.,DOMANSKIU.,RUHLEK.‐H.Efficiencyofcoldpassoverand heated humidification under continuous positive airway pressure. Eur. Respir. J.2002,20pp.183–286

[38] RYANS.,RANKINN.,MEYERE.WilliamsR.EnergyBalanceintheintubatedhumanairwayisanindicatorofoptimalgasconditioning.Crit. Care Med.2002,30(2)pp.355–361

[39] TUGGEY J.,DELMASTROM.,ELLIOTTM.Theeffectofmouth leakandhumidificationduringnasalnon‐invasiveventilation.Respir. Med.2007

[40] KILLICK E.Physiological response to breathing hot air. Dept of Physiology, University ofLeeds,1931

[41] ReportoftheNavalMedicalResearchandDevelopmentCommand,PhysiologicalDesignGoalsforThermalProtectionforDivers,Conferencereportof5September1980

[42] GALLAGHER S., VERCRUYSSENM., DENO N. Hot air breathing: effects of elevated wet bulbtemperatures on tissue temperatures of themouth.Am. Ind. Hyg. Assoc. J. 1985,46(6)pp.332–335

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 100: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

88 ©ISO2017–Allrightsreserved

[43] MORITZA.R.,HENRIQUESF.C.JR.StudiesofthermalinjuryI.Therelativeimportanceoftimeand surface temperature in the causation of cutaneous burns. Am. J. Pathol. 1974, 23pp.695–720

[44] CENGELY.,BOLESM.Thermodynamics–AnEngineeringApproach(2nded),McGraw‐HillInc,1994,equation13‐7,p.693

[45] US Standard Atmosphere.USgovernmentprintingoffice,1976

[46] NISTChemistryWebBook.(2011)http://webbook.nist.gov/chemistry/,Retrieved1May2013,NationalInstituteofStandardsandTechnology

[47] CHASEM.W.JR.NIST‐JANAFThermochemicalTables,FourthEdition,(1998)J.Phys.Chem.Ref.Data,Monograph9,1‐1951

[48] HandbookofChemistryandPhysics.73rdEdition,(1992)Ed.DavidR.Lide.CRCPress:BocaRatonChapter6,p.11

[49] BUCKA.L.NewEquationsforComputingVaporPressureandEnhancementFactor.J. Appl. Meteorol.1981,20pp.1527–1531

[50] BuckResearchInstrumentsLLC,Model CR-1A operating manual,2012

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 101: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D

Page 102: Medical electrical equipment Part 2-74: Particular ......equipment in medical practice, Subcommittee SC 62D, Electromedical equipment. This first edition of ISO 80601‐2‐74 cancels

ISO 80601-2-74:2017(E)

© ISO 2017 – All rights reserved

ICS 11.040.10Price based on 88 pages

FR

EE

LIC

EN

SE

BY

KE

NY

A B

UR

EA

U O

F S

TA

ND

AR

DS

(K

EB

S)

TO

INT

ER

ES

TE

D P

AR

TIE

S, S

TA

KE

HO

LDE

RS

MA

NU

FA

CT

UR

ER

S A

ND

PU

BLI

C, F

OR

A L

IMIT

ED

AC

CE

SS

DU

RIN

G C

OV

ID-1

9 E

ME

RG

EN

CY

,O

N 2

020-

04-1

5 -

FO

R P

ER

SO

NA

L U

SE

ON

LY. C

OP

YIN

G, I

NT

RA

NE

T A

ND

INT

ER

NE

T P

RO

HIB

ITE

D