MATHEMATICAL MODELING OF BLOOD FLOW THROUGH AN ECCENTRIC CATHETERIZED ARTERY SIMA SARV AHRABI A dissertation submitted in partial fulfilment of the requirements for the award of the degree of Master of Science (Engineering Mathematics) Faculty of Science Universiti Teknologi Malaysia DECEMBER 2013

Transcript of MATHEMATICAL MODELING OF BLOOD FLOW THROUGH...

  • MATHEMATICAL MODELING OF BLOOD FLOW THROUGH AN

    ECCENTRIC CATHETERIZED ARTERY

    SIMA SARV AHRABI

    A dissertation submitted in partial fulfilment of the

    requirements for the award of the degree of

    Master of Science (Engineering Mathematics)

    Faculty of Science

    Universiti Teknologi Malaysia

    DECEMBER 2013

  • iii

    To my beloved Father and Mother

  • iv

    ACKNOWLEDGMENT

    Foremost, I would like to express my sincere gratitude to my supervisor

    PROF. DR. NORSARAHAIDA BINTI SAIDINA AMIN for her patience,

    motivation, enthusiasm, and immense knowledge in guiding me to complete this

    dissertation.

    Last but not least; I would like to thank my family: my parents DAVOOD

    and HAMIDE, for giving birth to me and supporting me spiritually throughout my

    life; As well as my lovely sisters Maryam and Mahshad for all their kindness towards

    me. Furthermore, my aunts Saadaat, Minoo, Nikoo and Mahboobe who did not leave

    me alone; in addition, all of my friends who really love and help me to finish this

    dissertation.

  • v

    ABSTRACT

    The mathematical model of blood flow through a catheterized stenosed artery

    is considered. A catheter is a tube, which is used in medicine for patients who are

    bedridden and whose blood pressure needs to be measured and monitored

    continuously. An example is the use of catheter during X-ray angiography or

    coronary balloon angioplasty in cardiac patients. Inserting a catheter in an artery will

    alter some characteristics of blood flow. This project investigates the effect on blood

    flow characteristics such as the velocity, the wall shear stress, the resistance

    impedance and the streamlines when a catheter is inserted into a stenosed artery.

    The catheter and the artery are assumed to be in a co-axial and eccentric position

    while blood is assumed to be Newtonian. The governing Navier-Stokes equations are

    solved analytically using perturbation method. The results show that a catheter

    placed in an eccentric position does alter the blood flow characteristics such that the

    axial velocity and the wall shear stress distribution are higher while the resistance

    impedance values are lower compared to their values in an artery where the catheter

    is concentrically placed. It is also found that under the same situation, the position of

    trapping moves closer to the wall of the stenosis while the size of the trapped bolus

    increases.

  • vi

    ABSTRAK

    Pemodelan matematik bagi aliran darah apabila kateter dimasukkan ke dalam

    arteri berstenosis dipertimbangkan. Kateter adalah suatu tiub yang digunakan dalam

    perubatan bagi pesakit yang memerlukan tekanan darah mereka diukur dan dipantau

    secara berterusan. Contohnya, kateter digunakan semasa prosidur X-ray angiografi

    atau belon angioplasti untuk pesakit jantung. Apabila kateter dimasukkan ke dalam

    arteri, ciri-ciri aliran darah akan berubah. Projek ini mengkaji kesan kateter terhadap

    ciri-ciri aliran darah seperti halaju, tekanan ricih diding, rintangan impedan dan garis

    arus apabila kateter dimasukkan pada kedudukan sepaksi eksentrik , manakala darah

    diandaikan sebagai bendalir Newtonan. Persamaan menakluk Navier-Stokes

    diselesaikan seecara analisis menggunakan kaedah usikan. Hasil kajian menunjukkan

    bahawa halaju sepaksi dan tekanan ricih dinidng adalah lebih tinggi sementara nilai

    rintangan impedan adalah lebih rendah berbanding dengan ciri-ciri aliran pada arteri

    di mana kateter diletakkan secara konsentrik. Kajian juga mendapati pada situasi

    serupa, kedudukan trapping makin hampir dengan dinding stenosis sementara saiz

    bolus trapping bertambah.

  • vii

    TABLE OF CONTENTS

    CHAPTER TITLES PAGE

    DECLARATION ii

    DEDICATION iii

    ACKNOWLEDGEMENTS iv

    ABSTRACT v

    ABSTRAK vi

    TABLE OF CONTENTS vii

    LIST OF TABLES x

    LIST OF FIGURES xi

    LIST OF SYMBOLS xiv

    1 INTRODUCTION 1

    1.1 Research Background 1

    1.2 Problem Statement 4

    1.3 Objectives of the Study 5

    1.4 Scope of the Study 5

    1.5 Significance of the Study 5

    1.6 Outline of Dissertation

    6

  • viii

    2 LITERATURE REVIEW 7

    2.1 Introduction 7

    2.2 Blood Flow through Catheterized Artery 7

    3 PROBLEM FORMULATION 14

    3.1 Introduction 14

    3.2 Geometery of Arterial Wall 14

    3.3 Governing Equations 16

    3.4 Non-Dimensionalization of Equations 18

    4 SOLUTION PROCEDURE 25

    4.1 Introduction 25

    4.2 Solution by Perturbation Method 25

    4.3 Solution of the Axial Velocity 26

    4.4 Solution of the Stream Function 32

    4.5 Solution of the Wall Shear Stress 35

    4.6 Solution of Resisance Impedance 35

    5 RESULTS AND DISCUSSIONS 38

    5.1 Introduction 38

    5.2 Axial Velocity 38

    5.3 Wall Shear Stress 42

    5.4 Resistance Impedance 47

    5.5 Streamlines 49

  • ix

    6 CONCLUSION 56

    6.1 Summary of Work 56

    6.2 Conclusion 56

    6.3 Recommendation 57

    REFERENCES 59

  • x

    LIST OF TABLES

    TABLE NO. TITLE

    PAGE

    2.1 Summary of Literature Review 11

  • xi

    LIST OF FIGURES

    FIGURE NO. TITLE PAGE

    1.1 Arterial Catheter

    1

    1.2a A Catheter through an artery

    2

    1.2b Balloon Angioplasty

    2

    1.3 X-ray Angiography

    3

    1.4 Intravascular Ultrasound and Using Catheter

    3

    2.1 Catheterized curved artery

    8

    2.2 Catheterized artery with a clot

    9

    2.3 Catheterized artery with a balloon

    10

    3.1 Schematic diagram of an overlapping stenosed artery

    15

    3.2 Schematic diagram of a tapered overlapping stenosed

    artery

    15

    3.3 Schematic diagram of eccentric catheter through stenosed

    artery

    16

    5.1 Variation of axial velocity zv with radial distance r for

    different values of catheter radius

    39

    5.2 Variation of axial velocity zv with radial distance r for

    catheter radius 0.1 and different values of eccentricity

    parameter

    40

    5.3 Variation of axial velocity zv with radial distance r for

    different values of velocity of catheter

    41

  • xii

    5.4 Variation of axial velocity zv with radial distance r for

    different values of angle of circumferential direction

    41

    5.5 Variation of wall shear stress distributionw in the stenotic

    region for difference values of maximum height of stenosis

    42

    5.6 Variation of wall shear stress distributionw in the stenotic

    region for difference values of taper angle

    43

    5.7 Variation of wall shear stress distributionw in the stenotic

    region for difference values of catheter radius

    44

    5.8 Variation of wall shear stress distributionw in the stenotic

    region for difference values of eccentricity parameter

    45

    5.9 Variation of wall shear stress distributionw in the stenotic

    region for difference values of velocity of catheter

    46

    5.10 Variation of wall shear stress distributionw in the stenotic

    region for difference values of angle of circumferential

    direction

    46

    5.11 Variation of the resistance impedance with the maximum

    height of stenosis * for different values of catheter radius

    47

    5.12 Variation of the resistance impedance with the maximum

    height of stenosis * for different values of eccentricity

    parameter

    48

    5.13 Variation of the resistance impedance with the maximum

    height of stenosis * for different values of velocity of

    catheter

    48

    5.14 Variation of the resistance impedance with the maximum

    height of stenosis * for different values of taper angle

    49

    5.15 Plot showing streamlines for different values of the radius

    50

    5.16 Plot showing streamlines for different values of the

    eccentricity parameters

    51

  • xiii

    5.17 Plot showing streamlines for different values of the angle

    of circumferential direction

    52

    5.18 Plot showing streamlines for different values of the

    velocity of catheter

    53

    5.19 Plot showing streamlines for different values of maximum

    height of stenosis

    54

    5.20 Plot showing streamlines for different values of taper

    angle.

    55

  • xiv

    LIST OF SYMBOLS

    d - Location of the stenosis

    eR - Reynolds number

    ( , )R z t - Radius of the tapered arterial region

    0R - Radius of the normal artery without stenosis

    p - Fluid pressure

    t - Time

    - Density of fluid

    - Angular frequency

    - Dynamic viscosity

    - Angle of tapering

    - Shear stress

    - Resistance impedance

    - Radius of catheter

    - Eccentricity parameter

    - Stream function

  • CHAPTER 1

    INTRODUCTION

    1.1 Research Background

    A catheter is a thin, flexible, hollow and lengthy plastic tube that is generally

    inserted into an artery (large blood vessel) placed in the wrist; but can also be

    inserted into the elbow, groin, foot or the inside of the arm. However, in theory,

    every artery could be used for injection of catheter but in practical situation, arteries

    of the mentioned organs are commonly placed for the injection. (Figure 1.1)

    Figure 1.1 Arterial Catheter, [wikipedia.org/wiki/File:Arterial_kateter,(n.d.)]

    http://en.wikipedia.org/wiki/File:Arterial_kateter_(Seldinger).jpg

  • 2

    Catheters have an important role in medicine for heart problems. For

    instance, they are usually used to diagnose and treat heart conditions. During

    catheterization, small tubes (catheters) are inserted into the circulatory system under

    x-ray guidance in order to obtain information about blood flow and pressures within

    the heart and to determine if there are obstructions within the blood vessels feeding

    the heart muscle (coronary arteries). Usage of the catheter is well known in medicine

    of various arterial diseases as ―coronary balloon angioplasty‖ (Figure 1.2a,b), ―X-ray

    angiography‖(Figure 1.3),―and ―intravascular ultrasound‖ (Figure 1.4).

    Figure 1.2a A Catheter through an artery, [ arterydisease.org/(n.d.)]

    Figure 1.2b Balloon Angioplasty, [ arterydisease.org/(n.d.)]

    http://virchicago.com/peripheral-artery-disease/http://virchicago.com/peripheral-artery-disease/

  • 3

    Figure 1.3 X-ray Angiography, [ arterydisease.org/(n.d.)]

    Figure 1.4 Intravascular Ultrasound and Using Catheter, [Tekes et al.,(2009)]

    http://virchicago.com/peripheral-artery-disease/

  • 4

    In addition, a catheter is used for the measurement of various physiological

    flow characteristics such as arterial blood pressure or pressure gradient and flow

    velocity or flow rate. Examples include patients in the intensive care unit (ICU)

    requiring isotropic support or patients with severe cardiovascular disease undergoing

    surgery.

    Furthermore, catheter can be used for periodic arterial blood gas analysis in

    patients with respiratory failure, or severe acid/base disturbance. When a patient has

    a lung problem that is so severe it requires checking the levels of oxygen or carbon

    dioxide of the blood more than 3 to 4 times a day regularly, the arterial catheter is

    used to draw blood without having to repeatedly stick a needle into the patient body.

    1.2 Problem Statement

    When a catheter is inserted into a blood vessel, blood clots can be formed on

    the tips of arterial catheters, the clots can block blood flow. Furthermore,

    bleeding can occur at the time of inserting the catheter. Patient becomes

    uncomfortable resulting from the injection. However, more important, the injection

    of a catheter alters the flow field and disturbs the hemodynamic conditions in the

    artery. In practical situation, the injection of a catheter cannot be exactly concentric

    with the artery. It is in fact placed in an eccentric position. Hence, this study

    investigates what happen to the blood flow characteristics when catheter is inserted

    eccentrically through an artery.

  • 5

    1.3 Objectives of the Study

    The main objective of this research is to determine the effects of eccentric

    catheterization on blood flow characteristics in a tapered and stenosed artery.

    Specific objectives are to calculate the axial velocity, stream function and trapping,

    resistance impedance and wall shear stress.

    1.4 Scope of the Study

    This study is based on the work of Mekheimer and Kot (2012). The blood is

    considered Newtonian fluid, incompressible, steady and laminar; In addition,

    catheter and artery supposed to be two tubes, which are co-axial together, and inner

    tube is catheter that assumed to be in eccentric position with the outer tube (tapered

    artery).

    1.5 Significance of the Study

    A motivation of the present analysis is the hope that such a problem will be

    applicable in many clinical applications. It is hoped that with this investigation, the

    injection can be carried out more proficiently. In addition, the results are used to

    obtain the estimates of increased or decreased of blood flow characteristics across an

    artery stenosis during catheterization.

  • 6

    1.6 Outline of Dissertation

    In Chapter 1, the nature and use of catheter the problem statement and

    objectives are presented. In Chapter 2, overview of previous studies on catheterized

    artery and blood flow changes are discussed. In Chapter 3, problem is formulated

    mathematically. In Chapter 4, method to solve the problem is expressed and the

    equations of some characteristics of blood flow are obtained. In Chapter 5, the results

    are displayed graphically. In Chapter 6, conclusions and comments about the results

    and recommendation for future studies are interpreted.

  • REFERENCES

    Anderson, J. D. (1995). Computational fluid dynamics (Vol. 206): McGraw-Hill

    New York.

    Biswas, D., & Chakraborty, U. S. (2010). Steady Flow of Blood through a

    Catheterized Tapered Artery with Stenosis: A Theoretical Model. Assam

    University Journal of Science and Technology, 4(2), 7-16.

    Daripa, P., & Dash, R. K. (2002). A numerical study of pulsatile blood flow in an

    eccentric catheterized artery using a fast algorithm. Journal of engineering

    mathematics, 42(1), 1-22.

    Dash, R., Jayaraman, G., & Mehta, K. (1996). Estimation of increased flow

    resistance in a narrow catheterized artery—a theoretical model. Journal of

    Biomechanics, 29(7), 917-930.

    Dash, R., Jayaraman, G., & Mehta, K. (1999). Flow in a catheterized curved artery

    with stenosis. Journal of Biomechanics, 32(1), 49-61.

    Doffin, J., & Chagneau, F. (1981). Oscillating flow between a clot model and a

    stenosis. Journal of Biomechanics, 14(3), 143-148.

    Hoffmann, K. A., & Chiang, S. T. (2000). {Computational fluid dynamics, Vol. 1}.

    Wichita, KS: Engineering Education System.

    Ismail, Z., Abdullah, I., Mustapha, N., & Amin, N. (2008). A power-law model of

    blood flow through a tapered overlapping stenosed artery. Applied

    Mathematics and Computation, 195(2), 669-680.

    Jayaraman, G., & Sarkar, A. (2005). Nonlinear analysis of arterial blood flow—

    steady streaming effect. Nonlinear Analysis: Theory, Methods &

    Applications, 63(5), 880-890.

    Jayaraman, G., & Tewari, K. (1995). Flow in catheterised curved artery. Medical and

    Biological Engineering and Computing, 33(5), 720-724.

    Kanai, H., Iizuka, M., & Sakamoto, K. (1970). One of the problems in the

    measurement of blood pressure by catheter-insertion: Wave reflection at the

    tip of the catheter. Medical and biological engineering, 8(5), 483-496. doi:

    10.1007/BF02477185

    Kundu, P. K., & Cohen, I. M. (1990). Fluid mechanics: Academic Press.

  • 60

    Labadin, J., & Walton, A. (2006). Modeling of Axial Flow between eccentric

    cylinders. Paper presented at the Proceedings of the 2 nd IMT-GT Regional

    Conference on mathematics, Statistics and Applications Universiti sains

    Malaysia, Penang, June.

    Mandal, P. K. (2005). An unsteady analysis of non-Newtonian blood flow through

    tapered arteries with a stenosis. International Journal of Non-Linear

    Mechanics, 40(1), 151-164.

    Mekheimer, K. S., & El Kot, M. (2012). Mathematical modeling of axial flow

    between two eccentric cylinders: Application on the injection of eccentric

    catheter through stenotic arteries. International Journal of Non-Linear

    Mechanics, 47(8), 927-937.

    Mekheimer, K. S., & Kot, M. E. (2010). Suspension model for blood flow through

    arterial catheterization. Chemical Engineering Communications, 197(9),

    1195-1214.

    Sankar, D. (2009). A two-fluid model for pulsatile flow in catheterized blood vessels.

    International Journal of Non-Linear Mechanics, 44(4), 337-351.

    Sankar, D., & Hemalatha, K. (2007). A non-Newtonian fluid flow model for blood

    flow through a catheterized artery—steady flow. Applied mathematical

    modelling, 31(9), 1847-1864.

    Sankar, D., & Lee, U. (2008). Two-fluid non-linear model for flow in catheterized

    blood vessels. International Journal of Non-Linear Mechanics, 43(7), 622-

    631.

    Sankar, D., & Lee, U. (2010). Pulsatile flow of two-fluid nonlinear models for blood

    flow through catheterized arteries: a comparative study. Mathematical

    Problems in Engineering, 2010.

    Tekes, C., Karaman, M., & Degertekin, F. L. (2009). Co-array optimization of

    CMUT arrays for Forward-Looking IVUS. Paper presented at the Ultrasonics

    Symposium (IUS), 2009 IEEE International.

    Walton, A. G. (2003). The nonlinear instability of thread-annular flow at high

    Reynolds number. Journal of Fluid Mechanics, 477, 227-257.

    Young, D. (1968). Effect of a time-dependent stenosis on flow through a tube.

    Journal of Engineering for Industry, 90, 248.