Mass Spectrometry and Proteomics - Lecture 2 - Matthias Trost Newcastle University [email protected]

Transcript of Mass Spectrometry and Proteomics - Lecture 2 · Tandem mass spectrometry • MS/MS, MS2 or tandem...

Page 1: Mass Spectrometry and Proteomics - Lecture 2 · Tandem mass spectrometry • MS/MS, MS2 or tandem mass spectrometry • Tandem-in-space can be performed on (ion beam-transmitting)

Mass Spectrometry and Proteomics- Lecture 2 -

Matthias TrostNewcastle University

[email protected]

Page 2: Mass Spectrometry and Proteomics - Lecture 2 · Tandem mass spectrometry • MS/MS, MS2 or tandem mass spectrometry • Tandem-in-space can be performed on (ion beam-transmitting)

Previously:

40

Electrospray

MALDI

Resolution and other basics

Page 3: Mass Spectrometry and Proteomics - Lecture 2 · Tandem mass spectrometry • MS/MS, MS2 or tandem mass spectrometry • Tandem-in-space can be performed on (ion beam-transmitting)

Lecture 2

• Mass analysers• Detectors• Tandem mass spectrometry

41

Page 4: Mass Spectrometry and Proteomics - Lecture 2 · Tandem mass spectrometry • MS/MS, MS2 or tandem mass spectrometry • Tandem-in-space can be performed on (ion beam-transmitting)

42

Mass analysers

• Sector instruments (magnetic- B; electrostatic – E)• Time-of-Flight (TOF)• Quadrupol (Q)• Ion Trap (IT)• Ion Cyclotron Resonance (ICR)• Orbitrap

Page 5: Mass Spectrometry and Proteomics - Lecture 2 · Tandem mass spectrometry • MS/MS, MS2 or tandem mass spectrometry • Tandem-in-space can be performed on (ion beam-transmitting)

43

Factors affecting choice of Mass analyser

• Mass range• Resolving power• Mass accuracy and ease of calibration• Speed• Scanning or not scanning?• Ease of interfacing with ion sources• Ease of use for tandem mass spectrometry

Page 6: Mass Spectrometry and Proteomics - Lecture 2 · Tandem mass spectrometry • MS/MS, MS2 or tandem mass spectrometry • Tandem-in-space can be performed on (ion beam-transmitting)

44

Sector instruments

• Mass range up to m/z 15,000• Resolving power up to 100,000• Most suitable ionisation: EI, CI,

FAB, FD• Less suitable for ESI or MALDI• Rarely used nowadays, large,

complicated

Page 7: Mass Spectrometry and Proteomics - Lecture 2 · Tandem mass spectrometry • MS/MS, MS2 or tandem mass spectrometry • Tandem-in-space can be performed on (ion beam-transmitting)

45

Time-of-Flight (TOF)

• Requires pulsed ion packages (MALDI, pulser in ESI-QTOF)• Ions travel in a field-free region according to their kinetic energy

Ek=mv2/2; i.e. mass can be determined if speed is measured.• Theoretically unlimited mass range (realistically >250k)• Resolving power up to ~30,000• Electrostatic mirrors (Reflectrons) used to lengthen flight time and

reduce spread of ion populations higher resolution• Fast, sensitive, simple, cheap

Page 8: Mass Spectrometry and Proteomics - Lecture 2 · Tandem mass spectrometry • MS/MS, MS2 or tandem mass spectrometry • Tandem-in-space can be performed on (ion beam-transmitting)

Better resolving power: Reflectron TOF

D e te c to r

Ion source

Acceleration region Reflectron

Field-free drift region

Ion sourceReflectron

Detector Reflectron

46

Page 9: Mass Spectrometry and Proteomics - Lecture 2 · Tandem mass spectrometry • MS/MS, MS2 or tandem mass spectrometry • Tandem-in-space can be performed on (ion beam-transmitting)

47

Time-of-Flight (TOF)

©Agilent

Page 10: Mass Spectrometry and Proteomics - Lecture 2 · Tandem mass spectrometry • MS/MS, MS2 or tandem mass spectrometry • Tandem-in-space can be performed on (ion beam-transmitting)

48

Quadrupole (Q)

• Four parallel metal rods, opposite rods have DC potentials (U) on which a radio frequency is superimposed.

• Ions oscillate and at a fixed U/Rf only a certain m/z passes through the quadrupole. By continuously varying U/Rf, the Q can be used for scanning.

Ion motion in a Quadrupole

Quadrupole schematic

Page 11: Mass Spectrometry and Proteomics - Lecture 2 · Tandem mass spectrometry • MS/MS, MS2 or tandem mass spectrometry • Tandem-in-space can be performed on (ion beam-transmitting)

49

Quadrupole (Q)

• Advantages: small, light, portable; robust; no high voltage; high dynamic range; cheap

• Disadvantages: low resolution (~2000); mass range m/z <3,000

Page 12: Mass Spectrometry and Proteomics - Lecture 2 · Tandem mass spectrometry • MS/MS, MS2 or tandem mass spectrometry • Tandem-in-space can be performed on (ion beam-transmitting)

Miniature Mass Spectrometers

50

Page 13: Mass Spectrometry and Proteomics - Lecture 2 · Tandem mass spectrometry • MS/MS, MS2 or tandem mass spectrometry • Tandem-in-space can be performed on (ion beam-transmitting)

51

Ion Trap (IT)• Two commonly used forms exist: Linear Ion Trap (LIT – e.g. Velos,

Thermo) and 3D or Paul Ion Trap (QIT – e.g. amaZon, Bruker)• Similar to quadrupole – but in 3D.• Equations of ion motion are complex.• Space charge effects possible when many ions in trap.• Resolution 2,000-10,000 depending on scanning speed.• Mass range <2,000, can only scan down to 1/3 m/z of precursor in MS/MS loss of low molecular masses.

Linear IT

Page 14: Mass Spectrometry and Proteomics - Lecture 2 · Tandem mass spectrometry • MS/MS, MS2 or tandem mass spectrometry • Tandem-in-space can be performed on (ion beam-transmitting)

52

3D Ion Trap (QIT)

3D IT

Page 15: Mass Spectrometry and Proteomics - Lecture 2 · Tandem mass spectrometry • MS/MS, MS2 or tandem mass spectrometry • Tandem-in-space can be performed on (ion beam-transmitting)

53

Potential well and ion motion in 3D trap

Page 16: Mass Spectrometry and Proteomics - Lecture 2 · Tandem mass spectrometry • MS/MS, MS2 or tandem mass spectrometry • Tandem-in-space can be performed on (ion beam-transmitting)

54

Fourier Transform Ion Cyclotron Resonance (FT ICR)

• The m/z is determined based on the cyclotron frequency of ions in a fixed magnetic field (requiring superconducting magnet).

• Ions are trapped in a Penning Trap (ion trap in magnetic field) and excited to a cyclotron motion by an oscillating electric field.

• Very high mass accuracy (<0.5 ppm) and very high resolution (>1M)• Ions are not destroyed thus time dependent studies such as gas

reactions in the ICR cell are possible.• Very expensive in purchase and maintenance, tricky to run

Page 17: Mass Spectrometry and Proteomics - Lecture 2 · Tandem mass spectrometry • MS/MS, MS2 or tandem mass spectrometry • Tandem-in-space can be performed on (ion beam-transmitting)

FT-ICR MS – working principle

55

Cyclotron frequency of ion:

http://www.chm.bris.ac.uk/ms/theory/fticr-massspec.html

m2zBωC

Page 18: Mass Spectrometry and Proteomics - Lecture 2 · Tandem mass spectrometry • MS/MS, MS2 or tandem mass spectrometry • Tandem-in-space can be performed on (ion beam-transmitting)

56

Orbitrap

• Developed in 2000 by Alexander Makarov, commercialised in 2005 by Thermo.

• Ions are injected into the electric field between electrodes and trapped. Ions oscillate around and along the inner electrode according to the m/z. FT allowsmeasurement of mass.

• Requires ultra-high vacuum (10-10 bar)• Requires “C-trap” to inject tight ion packages.• High mass accuracy (<1 ppm), high resolution (<500k)• Sensitive and high dynamic range.• Much cheaper than FT ICR.

Page 19: Mass Spectrometry and Proteomics - Lecture 2 · Tandem mass spectrometry • MS/MS, MS2 or tandem mass spectrometry • Tandem-in-space can be performed on (ion beam-transmitting)

The Story of the Orbitrap

57The Analytical Scientist , Nov 2013

Alexander Makarov, 1990s

Alexander Makarov with Orbitrap2013

Page 20: Mass Spectrometry and Proteomics - Lecture 2 · Tandem mass spectrometry • MS/MS, MS2 or tandem mass spectrometry • Tandem-in-space can be performed on (ion beam-transmitting)

58

Ion Mobility Spectrometer

• Ions are separated by shape, not m/z.• Drift gas and electric field opposed or orthogonal.• Drift time increases with collisional cross-section.• Often combined with mass spectrometry.

Page 21: Mass Spectrometry and Proteomics - Lecture 2 · Tandem mass spectrometry • MS/MS, MS2 or tandem mass spectrometry • Tandem-in-space can be performed on (ion beam-transmitting)

Ion mobility spectrum

59

Kloniecki et al, 2011

Oligomers of Alzheimer’s disease Aβ1-40 form several different structural states.

Page 22: Mass Spectrometry and Proteomics - Lecture 2 · Tandem mass spectrometry • MS/MS, MS2 or tandem mass spectrometry • Tandem-in-space can be performed on (ion beam-transmitting)

60

Detectors• Commonly used are Secondary Electron multipliers (SEM),

Channel electron multipliers (CEM) and Microchannel Plates (MCP).

• These work by the principle that energetic particles hitting a semiconductor or metal surface lead to emission of secondary electrons. These can than be used to further amplify the signal in a cascade.

• Gain of 106-108.• Detectors “age” and require regular replacement.

Page 23: Mass Spectrometry and Proteomics - Lecture 2 · Tandem mass spectrometry • MS/MS, MS2 or tandem mass spectrometry • Tandem-in-space can be performed on (ion beam-transmitting)

Detectors: MCP

61

Microchannel Plates (MCP)

http://encyclopedia2.thefreedictionary.com/

Note: ion counting detectors also give signals of impact of energetic neutrals, electrons and photons. Therefore, care has to be taken, not to allow other particles than the mass-analysed ions to hit the detector.

Page 24: Mass Spectrometry and Proteomics - Lecture 2 · Tandem mass spectrometry • MS/MS, MS2 or tandem mass spectrometry • Tandem-in-space can be performed on (ion beam-transmitting)

62

Tandem mass spectrometry

• MS/MS, MS2 or tandem mass spectrometry• Tandem-in-space can be performed on (ion beam-

transmitting) Hybrid instruments with 2 distinct mass analysers.

• Tandem-in-time in Ion traps: ion selection, activation and product ion analysis in the very same place.

Page 25: Mass Spectrometry and Proteomics - Lecture 2 · Tandem mass spectrometry • MS/MS, MS2 or tandem mass spectrometry • Tandem-in-space can be performed on (ion beam-transmitting)

63

Product Ion Scan1. Precursor or “Parent”

Ions are selected and isolated

2. Fragmentation by CID

3. Product or “Daughter” Ions are characterised by Q3

4. Most common MS/MS experiment –used for protein ID

Page 26: Mass Spectrometry and Proteomics - Lecture 2 · Tandem mass spectrometry • MS/MS, MS2 or tandem mass spectrometry • Tandem-in-space can be performed on (ion beam-transmitting)

64

Precursor Ion Scan1. Product Ion is

selected and Q3 is parked

2. Q1 scans normally3. Only precursors

which fragment to produce a selected product ion are detected.

4. Often used to target peptides with specific modifications resulting in “reporter ions”.

Page 27: Mass Spectrometry and Proteomics - Lecture 2 · Tandem mass spectrometry • MS/MS, MS2 or tandem mass spectrometry • Tandem-in-space can be performed on (ion beam-transmitting)

65

Neutral Loss Scan1. The mass of a functional

group whose loss is to be detected is selected

2. Both Q1 and Q3 scan simultaneously, offset by the selected “neutral loss” mass

3. Fragmentation by CID4. Product ions are

detected only when specified loss occurs in Q2, indicating the presence of the moiety of interest.

5. Often used for peptide modifications such as phosphorylation.

Page 28: Mass Spectrometry and Proteomics - Lecture 2 · Tandem mass spectrometry • MS/MS, MS2 or tandem mass spectrometry • Tandem-in-space can be performed on (ion beam-transmitting)

66

Selected Reaction Monitoring (SRM)

1. Precursor and product ion mass need to be known.

2. Q1 is fixed on specified mass.

3. Fragmentation by CID.4. Q3 is fixed on specified

fragment of mass selected in Q1.

5. Very commonly used for accurate peptide/small molecule quantitation.