Monoplace Hyperbaric Oxygen Chamber
Manufacturing Diving & Hyperbaric Chambers
O2mono
Manufacturing Diving & Hyperbaric Chambers
-
1
-
2
O2mono CHAMBER
The Baroks O2mono Model Monoplace Hyperbaric Treatment System consists of one
compartment, is designed to safely provide hyperbaric oxygen therapy to patients at pressure
up to 2 bar( 3ATA). Monoplace chamber is designed according to European Standarts and is
tested according to EN14931 requirements.
Only quality-certified “Special Steel” with test reports is used in our Monoplace Chamber
production. In addition; before production of the chamber, ultrasonic lamination tests and after
welding process; x-ray, magnetic particle and ultrasonic tests (Non-destructive tests) have
been done by professional engineering staff. All these process is controlled and reported by
certification company. After chamber production, hydrostatic tests is done by the certification
authority surveyors
The chamber has cylindrical shape and one entrance door
-
3
We are manufacturing two type monoplace chamer : Half Acrylic (HA) and Full Acrylic (FA).
General view of two types
Half Acrylic (HA) O2mono Full Acrylic (FA) O2mono
-
4
Monoplace HBOT Chamber; are designed to treat a single person pressurized with 100% oxygen.
Figure General O2mono Body
Chamber consists of below mentioned parts
Table 1: Chamber Parts
No Part Name
1 Cylindrical Body
2 Torispherical Head
3 Door
4 Body Wall
5 Viewport
6 Nozulles
-
5
Cylindrical Shell (Body)
The chamber length is 1000 mm and inner diameter is 900 mm.
The square door size of the chamber is 1050 mm. X 1050 mm and it provides entrance and
exit easily.
Door
Door size is 1050 mm X 1050 mm.
And thickness is 40 mm. It has also
locking mechanism for safety.
Figure Door Parts
The below table shows technical specifications about monoplace chamber system.
TECHNICAL SPECIFICATIONS
Working Pressure 3 ATA (2 bar)
Hydrostatic Test Pressure 4 ATA (3 bar)
Working Temperature 0C and 45C
Chamber Capacity 1 Person
Inner Diameter 900 mm.
Chamber Length (Inside) 2,221 mm.
Total Length 2494 mm.
Material Pressure Steel
Door Number 1
Door Opening 900 mm.
Weigth ≈ 1.5 ton
Electrical Power Requirement 3 KW
Electrical Connection Voltage 220 Volt, 50 Hz
Approved By TUV
-
6
Lock Mechanism
The door of the system can be opened or closed by using Lock Arm. The system cannot be
pressurized if both lock key and door key are not closed. Camshaft provides maximum sticking
on the door Automatic Locking Cylinder locks the door when the pressure reaches 0,1 bar.
Viewport
Viewports are manufactured from PMMA:( Poly Methl Meth Acrylate ) material where
chemical and physical properties are given in ASME PVHO standard. .
-
7
CHAMBER EQUIPMENTS This chapter consists of complementary /supportive equipments inside/on the chambe
Figure Chamber Equipments
Kod Parça Adı
1 Painting
2 Breathing System
3 Safety Valves
4 Stretcher
5 Entertaintment Monitor
6 Temperature and Humidity Sensor
7 Pressure Sensor
8 Antistatic Wrıst Strap
-
8
Painting
The chambers are covered with flame retardent non-toxic epoxy paint on both exterior
and interior surfaces. The paint, due to its flame retardency property starts burning at
high temperatures, the burning only ensuses at the centre of the heat source and is
extinguished by itself when it is removed. Also the the fumation that ensues during
burning is not toxic for humans. This feature also prevents any sickness that may ensue
because of the natural fumation spread by the paint because of nominal heat.
Breathing System
Breathing system provides required breathing air/oxygen to patients and evacuates
exhaust air via exhaust system
Drager X-Plore 3300 BIBS mask is a breathing mask that is used to deliver the treatment
gas to the patient.
Avox Inlet Regulator Avox Outlet Regulator
-
9
CONTROL PANEL
Figure O2mono Control Panel
No Part Name
1 Counter
2 Door Locking Button
3 Temperature Indicator
4 Humidity Indicator
5 CO2 Indicator
6 Chamber Pressure
7 O2 Source Pressure Manometer
8 Air Source Pressure Manometer
9 Chamber Pressure
10 Flowmeter
11 O2 Analyzer
12 Patient Warning Reset Button On/Off
13 Patient Warning Alarm
14 Door Locked Light
15 Pressure Inside Light
16 Set Pressure
17 TV On/Off
18 Emergency Stop Button
19 Intercom
20 System On/Off Button
21 BIBS Mask Air or O2 Selection
22 Auxiliary Exhaust
23 Decompression
24 Compression
25 Chamber Selection Air or O2
9 7
4
3
10
12
19
5
11
2
21 8
14
22 23 24
6
1
13
15
20
18
25
17
16
-
10
O2mono Control panel consist of the following equipment’s which enables the operator take
control of the chamber safely and reliable. All main operations and functions of the chamber
can be done with this panel .The control panel of the hyperbaric chamber is located on the
chamber and provides following operations.
• Treatments control
• Maintain and adapt the interior environment of the chamber
• Communication and lighting
• Supply, evacuation and exchange of the medical working gases
• Monitoring of important information such as humidity, temperature and oxygen
concentration and patients.
-
11
ENTERTAINTMENT UNIT
During the treatment session, an LCD entertainment monitor is mounted on the chamber to
spend time during the treatment. It provides opportunity to watch video and TV to patient.
POWER SUPPLY UNIT
O2mono electric system generally;
Power Stage
Communication
Digital Indicators
Analyzers
The electrical system is supplied with 220V AC. It is also reduced 24V
DC using electronic transformers.
Communication System consists of the operator and operator warning
system.
Digital indicators show pressure, temperature, humidity and CO2 value
shows.
Oxygen and carbon dioxide analyzer is available.
The input voltage system 220 V AC ± 10%, frequency 50 Hz ± 5%. There is power supply in
order to support the system at least 30 minutes in cases where input voltage is not present,
cut or go beyond the limit value.
LCD Monitor
-
12
PATIENT TRANSFER UNIT
Stretcher is used for patient transferring and it consists of two parts as chamber internal
stretcher and transfer stretcher. Internal stretcher is transported with transfer stretcher while
the patient is transported into the chamber by pushing The stretcher is designed for working in
hyperbaric conditions as mechanically.
Figure Stretcher
-
13
-
14
-
15