Manual Servicio Pretratamiento
-
Upload
maritza-murillo -
Category
Documents
-
view
241 -
download
0
Transcript of Manual Servicio Pretratamiento
-
8/18/2019 Manual Servicio Pretratamiento
1/230
FreseniusMedicalCare
Pretreatment ComponentsBASIC / STANDARD
Service ManualEdition: 1/03.11
Part no.: 636 210 1
-
8/18/2019 Manual Servicio Pretratamiento
2/230
-
8/18/2019 Manual Servicio Pretratamiento
3/230
Fresenius Medical Care Service Manual SM-EN 1/03.11 iii
Table of contents
1 Index
2 Important information
2.1 Preface........................................................................................................................................ 2-12.1.1 Structure of the Service Manual .................................................................................................. 2-1
2.2 How to use the Service Manual ................................................................................................ 2-2
2.3 Notes........................................................................................................................................... 2-32.3.1 Significance of the warnings........................................................................................................ 2-3
2.3.2 Significance of the note ............................................................................................................... 2-32.3.3 Significance of the tip .................................................................................................................. 2-3
2.4 Brief description ........................................................................................................................ 2-42.4.1 Operator information.................................................................................................................... 2-5
2.4.2 Guarantee / warranty................................................................................................................... 2-6
2.4.3 Technical documentation............................................................................................................. 2-6
2.5 Warnings .................................................................................................................................... 2-72.5.1 Basic warnings ............................................................................................................................ 2-7
2.5.2 Electric hazards ........................................................................................................................... 2-7
2.5.3 Chemical hazards........................................................................................................................ 2-7
2.5.4 Biological hazards ....................................................................................................................... 2-8
2.6 Addresses .................................................................................................................................. 2-9
3 Component overview
3.1 Components of the water pretreatment system ..................................................................... 3-13.1.1 Water pretreatment component overview / STANDARD........................................................ 3-23.1.2 Water pretreatment component overview / BASIC................................................................. 3-3
4 Feed water inlet unit4.1 Component overview – wall-mounted prefiltration unit ........................................................ 4-14.1.1 Prefiltration unit 1‘‘, 1 1/4‘‘, 1 1/2‘‘ ................................................................................................ 4-2
4.2 Installation.................................................................................................................................. 4-34.2.1 Installation / start-up requirements .............................................................................................. 4-3
4.2.2 Data reporting.............................................................................................................................. 4-3
4.3 Specifications ............................................................................................................................ 4-44.3.1 Specifications / data sheet........................................................................................................... 4-4
4.3.2 Operating conditions.................................................................................................................... 4-5
4.3.3 Configuration options................................................................................................................... 4-5
4.4 Maintenance............................................................................................................................... 4-6
-
8/18/2019 Manual Servicio Pretratamiento
4/230
iv Fresenius Medical Care Service Manual SM-EN 1/03.11
4.5 Functional description .............................................................................................................. 4-74.5.1 Design of components................................................................................................................. 4-7
4.5.1.1 Water inlet unit ............................................................................................................. 4-7
4.5.2 Component function..................................................................................................................... 4-7
4.5.2.1 Stop valve..................................................................................................................... 4-7
4.5.2.2 Water meter - optional.................................................................................................. 4-74.5.2.3 Backwashable prefilter 105 µm .................................................................................... 4-7
4.5.2.4 Solenoid valve.............................................................................................................. 4-8
4.5.2.5 Pipe disconnector ......................................................................................................... 4-9
4.5.3 Prefiltration rack with particle filter (optional) ............................................................................. 4-10
4.5.4 Component description.............................................................................................................. 4-10
4.6 Options ..................................................................................................................................... 4-114.6.1 Prefiltration rack with particle filter (option)................................................................................ 4-11
4.6.2 Pressure reducer - optional ....................................................................................................... 4-11
4.6.3 Configuration options (feed water tank)..................................................................................... 4-11
5 Booster pump
5.1 Booster pump assembly CRN5-8 component overview ........................................................ 5-1
5.2 Installation.................................................................................................................................. 5-25.2.1 Installation / start-up requirements .............................................................................................. 5-2
5.2.2 Important start-up information...................................................................................................... 5-2
5.2.2.1 Booster pump assembly connection ............................................................................ 5-2
5.2.2.2 Connection to the domestic water supply network....................................................... 5-3
5.2.2.5 Overview of electrical connections ............................................................................... 5-9
5.2.3 Data reporting ............................................................................................................................ 5-13
5.3 Specifications .......................................................................................................................... 5-145.3.1 Specifications / data sheet......................................................................................................... 5-145.3.1.1 Performance characteristics....................................................................................... 5-14
5.3.1.2 Volumetric capacity - diagram .................................................................................... 5-15
5.3.2 Operating conditions.................................................................................................................. 5-15
5.4 Maintenance ............................................................................................................................. 5-16
5.5 Error messages........................................................................................................................ 5-175.5.1 Malfunctions, causes and correction ......................................................................................... 5-17
5.6 Functional description ............................................................................................................ 5-185.6.1 Design of components............................................................................................................... 5-18
5.6.2 Component description.............................................................................................................. 5-18
5.6.2.1 Control unit ................................................................................................................. 5-185.6.2.2 Base load and peak load pump.................................................................................. 5-19
5.6.2.3 Alternating pressure situations ................................................................................... 5-19
5.6.2.4 Run-dry protection...................................................................................................... 5-19
5.6.2.5 Tank operation ........................................................................................................... 5-19
5.6.2.6 Automatic turnoff ........................................................................................................ 5-19
5.6.3 Functional description................................................................................................................ 5-20
5.6.3.1 Setting the system to the Standby mode.................................................................... 5-20
5.6.3.2 Selector switch functions............................................................................................ 5-20
-
8/18/2019 Manual Servicio Pretratamiento
5/230
Fresenius Medical Care Service Manual SM-EN 1/03.11 v
6 Pumps
6.1 Pump CM5-5 – component overview ....................................................................................... 6-1
6.2 Installation.................................................................................................................................. 6-2
6.2.1 Installation / start-up requirements .............................................................................................. 6-26.2.2 Important start-up information ..................................................................................................... 6-2
6.2.3 Data reporting.............................................................................................................................. 6-2
6.3 Specifications ............................................................................................................................ 6-36.3.1 Specifications / data sheet (CM5-5) ............................................................................................ 6-3
6.3.2 Operating conditions.................................................................................................................... 6-3
6.4 Maintenance............................................................................................................................... 6-4
6.5 Error messages on pumps ....................................................................................................... 6-56.5.1 Visual indication on CM5-5.......................................................................................................... 6-5
6.5.1.1 Alarm messages overview ........................................................................................... 6-5
6.5.2 Pump malfunctions overview....................................................................................................... 6-6
6.6 Functional description .............................................................................................................. 6-76.6.1 Design of components................................................................................................................. 6-7
7 Media filters (sand filters / Birm filters)
7.1 Component overview – sand filters S140, S200, S300, S400 ................................................ 7-1
7.2 Component overview – Birm filters B140, B200, B300, B400 ................................................ 7-2
7.3 Installation of the media filters (sand / Birm).......................................................................... 7-37.3.1 Installation / start-up requirements .............................................................................................. 7-3
7.3.2 Important start-up information ..................................................................................................... 7-47.3.2.1 Start-up of the sand filters / Birm filters ........................................................................ 7-4
7.3.2.2 Valve heads and drain restrictor ................................................................................... 7-5
7.3.2.3 Valves in the control head............................................................................................ 7-9
7.3.3 Programming the control for sand filters / Birm filters................................................................ 7-11
7.3.3.1 Control / display ......................................................................................................... 7-11
7.3.3.2 Control / control elements .......................................................................................... 7-11
7.3.3.3 Control / programming................................................................................................ 7-11
7.3.4 Data reporting............................................................................................................................ 7-23
7.3.5 Use of Birm for improved water quality..................................................................................... 7-24
7.3.5.1 Removal of iron using "Birm" as filter media in the sand filter .................................... 7-24
7.3.5.2 Alarm limits for water substances............................................................................... 7-24
7.3.6 Birm filter settings ...................................................................................................................... 7-26
7.3.6.1 Weekly / semi-weekly backwash................................................................................ 7-26
7.3.7 Operating conditions.................................................................................................................. 7-28
7.4 Specifications .......................................................................................................................... 7-297.4.1 Specifications / sand filter data sheet........................................................................................ 7-29
7.4.2 Specifications / Birm filter data sheet ........................................................................................ 7-30
7.4.3 Operating conditions.................................................................................................................. 7-31
7.5 Maintenance............................................................................................................................. 7-32
7.6 Error messages ....................................................................................................................... 7-337.6.1 Sand filter / Birm filter malfunctions overview............................................................................ 7-33
7.7 Functional description ............................................................................................................ 7-34
7.7.1 Design of components............................................................................................................... 7-34
-
8/18/2019 Manual Servicio Pretratamiento
6/230
vi Fresenius Medical Care Service Manual SM-EN 1/03.11
7.7.2 Component description.............................................................................................................. 7-34
7.7.2.1 Backwash ................................................................................................................... 7-34
7.7.2.2 Flow velocities............................................................................................................ 7-35
7.7.2.3 Feed water bypass..................................................................................................... 7-35
7.7.2.4 Efficiency .................................................................................................................... 7-35
8 Particle filter (Big Blue) 20"
8.1 Component overview – particle filter ....................................................................................... 8-1
8.2 Installation.................................................................................................................................. 8-28.2.1 Installation / start-up requirements .............................................................................................. 8-2
8.2.2 Data reporting .............................................................................................................................. 8-2
8.3 Specifications ............................................................................................................................ 8-38.3.1 Specifications / data sheet........................................................................................................... 8-3
8.3.2 Operating conditions.................................................................................................................... 8-3
8.3.2.1 Particle filter selection table.......................................................................................... 8-3
8.4 Maintenance ............................................................................................................................... 8-4
8.5 Functional description .............................................................................................................. 8-58.5.1 Design of components................................................................................................................. 8-5
8.5.2 Component description................................................................................................................ 8-5
9 Softener
9.1 Component overview – softener E080(B), E150(B), E225-T, E300, E400 Magnum .............. 9-1
9.2 Modification of the softener settings in case of high brine concentrations in combinationwith Kinetico-Autorol retrofit kits............................................................................................. 9-3
9.2.1 Settings for CP 216...................................................................................................................... 9-3
9.2.2 Settings for CP 213...................................................................................................................... 9-3
9.2.3 Settings for CP 210...................................................................................................................... 9-4
9.3 Installation.................................................................................................................................. 9-59.3.1 Installation / start-up requirements .............................................................................................. 9-5
9.4 Display and operating elements............................................................................................... 9-69.4.1 Display during operation .............................................................................................................. 9-6
9.4.2 Programming the control unit....................................................................................................... 9-7
9.4.2.1 Control / display............................................................................................................ 9-7
9.4.2.2 Programming the control .............................................................................................. 9-79.4.3 Important start-up information.................................................................................................... 9-13
9.4.3.1 Softener start-up procedure ....................................................................................... 9-14
9.4.4 Start-up procedure for E080(B),/ E150(B), E225-T, E300, E400............................................... 9-16
9.4.4.3 Brine tank ................................................................................................................... 9-22
9.4.4.5 Triplex operation E225-T............................................................................................ 9-23
9.4.4.6 Alternating or parallel operation ................................................................................. 9-24
9.4.5 Data reporting ............................................................................................................................ 9-44
9.5 Specifications .......................................................................................................................... 9-459.5.1 Specifications / data sheet......................................................................................................... 9-46
9.5.2 Restrictor inserts depending on pressure range........................................................................ 9-47
9.5.3 Softener capacities .................................................................................................................... 9-48
9.5.4 Operating conditions.................................................................................................................. 9-49
-
8/18/2019 Manual Servicio Pretratamiento
7/230
Fresenius Medical Care Service Manual SM-EN 1/03.11 vii
9.5.4.1 Softener selection table.............................................................................................. 9-49
9.5.4.2 Triplex softener (E225-T) ........................................................................................... 9-50
9.5.5 Removal of iron and manganese using a softener .................................................................... 9-50
9.6 Maintenance............................................................................................................................. 9-51
9.7 Error messages ....................................................................................................................... 9-529.7.1 Softener malfunctions overview................................................................................................. 9-52
9.8 Functional description ............................................................................................................ 9-559.8.1 Design of components............................................................................................................... 9-55
9.8.2 Component description.............................................................................................................. 9-55
9.8.3 Functional description Triplex operation E225-T....................................................................... 9-56
10 Activated carbon filter
10.1 Component overview – A050, A100, A150, A300, A400 ....................................................... 10-1
10.2 Installation................................................................................................................................ 10-210.2.1 Installation / start-up requirements ............................................................................................ 10-2
10.2.2 Important start-up information ................................................................................................... 10-3
10.2.3 Start-up of the activated carbon filter ......................................................................................... 10-4
10.2.3.1 Valve heads and drain restrictor ................................................................................. 10-4
10.2.3.3 Control unit design ..................................................................................................... 10-8
10.2.4 Programming the control unit .................................................................................................. 10-10
10.2.4.1 Control / display ....................................................................................................... 10-10
10.2.4.2 Control / operating elements .................................................................................... 10-10
10.2.4.3 Control unit ............................................................................................................... 10-11
10.2.5 Data reporting.......................................................................................................................... 10-21
10.3 Specifications ........................................................................................................................ 10-2210.3.1 Specifications / data sheet....................................................................................................... 10-2210.3.2 Operating conditions................................................................................................................ 10-24
10.3.2.1 Activated carbon filter (Single) selection table ......................................................... 10-25
10.3.2.2 Activated carbon filter (double) selection table......................................................... 10-25
10.4 Maintenance........................................................................................................................... 10-27
10.5 Error messages ..................................................................................................................... 10-2810.5.1 Activated carbon filter malfunctions overview.......................................................................... 10-28
10.6 Functional description .......................................................................................................... 10-2910.6.1 Design of components............................................................................................................. 10-29
10.6.2 Component description............................................................................................................ 10-29
11 USER RELEVANT – Operational data acquisition / maintenance
11.1 Operational data acquisition / daily report............................................................................ 11-2
11.2 System handling instructions report – water pretreatment ................................................ 11-5
11.3 Maintenance............................................................................................................................. 11-911.3.1 Maintenance intervals for water pretreatment components....................................................... 11-9
11.3.1.1 Feed water inlet unit ................................................................................................... 11-9
11.3.1.2 Sand filter ................................................................................................................... 11-9
11.3.1.3 Particle filter ............................................................................................................... 11-9
11.3.1.4 Softener .................................................................................................................... 11-10
-
8/18/2019 Manual Servicio Pretratamiento
8/230
viii Fresenius Medical Care Service Manual SM-EN 1/03.11
11.3.1.5 Activated carbon filter ............................................................................................... 11-11
11.3.1.6 Water leakage indicator ............................................................................................ 11-11
12 SERVICE RELEVANT – Reports12.1 Operational qualification......................................................................................................... 12-112.1.1 Start-up report............................................................................................................................ 12-1
12.1.2 Water pretreatment settings reports .......................................................................................... 12-6
12.1.3 Coarse filter and booster pump assembly settings report ......................................................... 12-6
12.1.4 Sand filter settings report........................................................................................................... 12-8
12.1.5 Activated carbon filter settings report ...................................................................................... 12-10
12.1.6 Softener settings report ........................................................................................................... 12-12
12.2 Maintenance ........................................................................................................................... 12-1412.2.1 Important information on maintenance procedures ................................................................. 12-14
12.2.2 Accessories and supplies for maintenance procedures .......................................................... 12-15
12.2.3 Maintenance report.................................................................................................................. 12-15
12.3 Tools and test equipment ..................................................................................................... 12-1912.3.1 Test equipment ........................................................................................................................ 12-19
13 Definitions
13.1 Definitions and terms .............................................................................................................. 13-1
13.2 Abbreviations........................................................................................................................... 13-1
13.3 Symbols.................................................................................................................................... 13-1
14 Appendix
-
8/18/2019 Manual Servicio Pretratamiento
9/230
Chapter 1: Index
Fresenius Medical Care Service Manual SM-EN 1/03.11 1-1
1 Index
A Abbreviations 13-1 Activated carbon filter 10-1
Activated carbon filter
malfunctions overview 10-28
Activated carbon filter settings
report 12-10
Addresses 2-9
Alternating or parallel operation 9-24
Appendix 11-1, 12-1
B
Basic warnings 2-7
Biological hazards 2-8
Booster pump 5-1
Brief description 2-4
C
Chemical hazards 2-7
Coarse filter and booster pumpassembly settings report 12-6
D
Definitions and terms 13-1
Disclaimer of liability 2-6
Duties of the responsible
organization 2-5
E
Electric hazards 2-7Error messages on pumps 6-5
F
Feed water bypass 7-35
Feed water inlet unit 4-1
G
Guarantee / warranty 2-6
IImportance of the instructions 2-2Important information 2-1
Improved water quality 7-24
M
Maintenance intervals 11-9
Maintenance report 12-15
Manual regeneration 9-14
Manually starting the regeneration
9-21Media filters (sand filters / Birm
filters) 7-1
N
Note, significance 2-3
Notes 2-3
O
Operational data acquisition / daily
report 11-2Operational qualification 12-1
Operator information 2-5
Operator responsibility 2-6
P
Particle filter (Big Blue) 20" 8-1
Preface 2-1
Prefiltration rack with particle filter 4-10
Prefiltration unit 4-2
Pressure reducer 4-11
Pump malfunctions overview 6-6
Pumps 6-1
S
Sand filter malfunctions overview 7-33
Sand filter selection table 7-25,7-31
Sand filter settings report 12-8
Selection table activated carbon
filter (double) 10-25
Selection table activated carbon
filter (Single) 10-25
Softener 9-1
Softener capacities 9-48
Softener malfunctions overview 9-52
Softener selection table 9-49
Softener settings in case of high
brine concentration 9-3
Softener settings report 12-12Start-up report 12-1
Symbols 13-1
System handling instructions
report – Water pre-treatment 11-5
T
Target group 2-5
Technical documentation 2-6
Tip, significance 2-3
W
Warnings 2-7
Warnings, significance 2-3
Water pretreatment / BASIC 3-3
Water pretreatment / STANDARD 3-2
Water pretreatment component
overview 3-2, 3-3
Water pretreatment settings
reports 12-6
-
8/18/2019 Manual Servicio Pretratamiento
10/230
Chapter 1: Index
1-2 Fresenius Medical Care Service Manual SM-EN 1/03.11
-
8/18/2019 Manual Servicio Pretratamiento
11/230
Chapter 2: Important information
Fresenius Medical Care Service Manual SM-EN 1/03.11 2-1
2 Important information
2.1 Preface
2.1.1 Structure of the Service Manual
This Service Manual contains a number of individual components or
overall concepts that are not products of Fresenius Medical Care
Deutschland GmbH.
Due to the standardization of documents, the structure for standardized
Service Manuals was applied to list the contents for service technicians
in their usual workflow.
Each component is therefore described in this Service Manual as an
individual device using the general structure.
When applying the master structure, empty chapters were omitted if the
chapter contents were not applicable.
-
8/18/2019 Manual Servicio Pretratamiento
12/230
Chapter 2: Important information
2-2 Fresenius Medical Care Service Manual SM-EN 1/03.11
2.2 How to use the Service Manual
Identification The document can be identified by the following information on the titlepage and on the labels, if any:
– Edition of the document – Part number of the technical document
Page identification The page identification 1-3, for example, refers to Chapter 1, page 3.
Editorial information The editorial information 1/01.05, for example, refers to: 1st edition,January 2005.
Organization of thechapters
To facilitate the use of documents from Fresenius Medical Care, the
organization of the chapters has been standardized in all manuals.
There may therefore be chapters within this document without any
content. Chapters without content are identified.
Illustrations The illustrations used in the documents (e.g. screens, photos, etc.) maydiffer from the original if this does not have any influence on the
function.
Importance of theinstructions
This document is part of the accompanying documents and is an
essential part of the device. It includes information necessary for use,
programming and start-up of the device.
This document must be studied carefully before attempting to operate
the device.
Before the responsible organization may begin operating the device,
the individual responsible for the operation must have been instructed
by the manufacturer on how to use the device and must be thoroughlyfamiliar with the contents of the document.
The device may only be operated by individuals certificated to have
been instructed on the proper operation and handling of the system.
Changes Changes to the technical document will be released as new editions orsupplements. In general, this manual is subject to change without
notice.
Reproduction Reproduction, even in parts, only with written approval.
-
8/18/2019 Manual Servicio Pretratamiento
13/230
Chapter 2: Important information
Fresenius Medical Care Service Manual SM-EN 1/03.11 2-3
2.3 Notes
2.3.1 Significance of the warnings
2.3.2 Significance of the note
2.3.3 Significance of the tip
Warning
Advises the operator that failure to observe this information can result
in personal injury.
Warning
Advises the operator that failure to observe this information can cause
damage to the equipment.
Note
Information alerting the operator to the fact that failure to follow the
steps as specified may result in the specific function not being executed
correctly, not being executed at all, or not producing the desired effect.
Tip
Information providing useful tips for easy handling.
-
8/18/2019 Manual Servicio Pretratamiento
14/230
Chapter 2: Important information
2-4 Fresenius Medical Care Service Manual SM-EN 1/03.11
2.4 Brief description
Design BASIC /STANDARD
Components The water pretreatment may consist of the following individualcomponents
– Feed water inlet unit (single components A-F)
– Booster pump
– Media filter (sand filter / Birm filter)
– Particle filter (Big Blue)
– Softener
– Activated carbon filter
If necessary, additional modules can be connected upstream ordownstream for improved quality.
The water pretreatment components must be installed in a dry, dust-
free room. The responsible organization must ensure that the design
specifications of the pretreatment components comply with the
requirements of the other components used. This applies especially to
the following system components such as reverse osmosis system,
permeate ring main and dialysis systems.
As the complete system consists of separate, individual systems, the
responsible organization is responsible for the complete system.
-
8/18/2019 Manual Servicio Pretratamiento
15/230
Chapter 2: Important information
Fresenius Medical Care Service Manual SM-EN 1/03.11 2-5
2.4.1 Operator information
Target group The water pretreatment may only be installed, operated and used byindividuals with the appropriate training, knowledge and experience.
Assembly, extensions, adjustments, modifications or repairs may onlybe carried out by the manufacturer or persons authorized by him.
Duties of the responsibleorganization
The responsible organization assumes the following responsibilities:
– Compliance with the national or local installation, operation, use and
maintenance regulations
– Compliance with the accident prevention regulations
– Correct and safe condition of the water pretreatment.
– Permanent availability of accompanying documents.
The water pretreatment consists of components designed to improve
the permeate conductivity and is extended with additional componentsto a complete water treatment system for use in a dialysis unit by the
responsible organization.
The water pretreatment must be installed in a dry room not reserved for
medical use. The staff call should also be connected. The responsible
organization must ensure that the technical adjustments of water
pretreatment match the requirements of the other components and that
they make up the complete system.
This applies especially to the following system components such as:
– Rooms with outlets and drains for the installation of the water
pretreatment
– Leakage protection systems
– Sampling site in the soft water supply line
– Sampling sites in the ring main system
– Alert systems for malfunctions (alarms)
– Operational organization
The components of the water pretreatment must be easily accessible
from all sides. The responsible organization must ensure that its
operators have been trained.
Operators of the water pretreatment, the reverse osmosis system and
the dialysis systems must have received instructions on how to operatethe system.
The responsible organization should inform the local water supplier of
the dialysis operation and insist on early provision of data regarding
water composition, availability etc.. This measure does not relieve the
responsible organization of its obligation to regularly check the feed
water composition.
Samples for microbial testing must therefore be collected from the
system and from the individual parts of the system in accordance with
the applicable regulations.
-
8/18/2019 Manual Servicio Pretratamiento
16/230
Chapter 2: Important information
2-6 Fresenius Medical Care Service Manual SM-EN 1/03.11
As the water supply system consists of a series of smaller systems, the
responsible organization is responsible for the complete water supply
system.
Operator responsibility When entering parameters, the following must be observed:
– The parameters entered must be verified by the operator, i.e. the
operator must check that the values entered are correct. If the
verification reveals a deviation between the desired parameters and
the parameters displayed, the setting must be corrected before
activating the function.
– The actual values displayed must be compared with the desired
values specified.
Disclaimer of liability Should the responsible organization wish to use other consumablesand accessories than those listed in the accompanying documents, the
responsibility to ensure the correct function of the device lies exclusively
with the responsible organization. The applicable legal regulations mustbe complied with (e.g. in Germany the Medical Device Directive, MDD
and the MPBetreibV = German regulation for the operation of medical
products).
The manufacturer does not assume any responsibility or liability for
personal injury or other damage and excludes any warranty for damage
to the components of the water pretreatment resulting from the use of
non-approved or unsuitable consumables or accessories.
2.4.2 Guarantee / warranty
Guarantee The extent of the guarantee is stipulated in the respective purchaseorders.
Warranty The purchaser's warranty rights are governed by the applicable legalregulations.
Proper function of the water pretreatment components is guaranteed for
the period specified in the sale contract.
Any use of the device which is not in accordance with its intended use
will void any liability and warranty.
2.4.3 Technical documentation
On request, descriptions and other documents are made available by
the manufacturer. These are intended to support trained personnel of
the responsible organization in servicing and repairing the device.
-
8/18/2019 Manual Servicio Pretratamiento
17/230
Chapter 2: Important information
Fresenius Medical Care Service Manual SM-EN 1/03.11 2-7
2.5 Warnings
2.5.1 Basic warnings
2.5.2 Electric hazards
,
2.5.3 Chemical hazards
Warning
The design of the water pretreatment components must comply with
the requirements of the Fresenius Installation Guideline (IGL)
Warning
Danger to life caused by electrical voltage.
Touching live parts will cause an electric shock.
– Before opening electrical devices, disconnect the power plug.
Actuating the ON/OFF switch stops operation of electrical devices,
but does not disconnect the system from the supply voltage!
Warning
The national standards and regulations must be observed when
connecting the system to the power supply system.
Warning
Use of protective equipments
Rubber gloves (acrylonnitril latex, cotton-lined) should be worn to avoid
contact of chemicals or disinfectant with the skin.
When working with acidic substances:
Wear goggles! Observe the safety precautions of the concentratedsubstance used!
In the event of contact with acid:
Eye: Immediately flush with gently flowing water for 15 minutes.
Skin: Use soap under flowing water for neutralization.
Ingestion: Do not induce vomiting, but have the person affected drink
plenty of still water. Seek medical advice.
-
8/18/2019 Manual Servicio Pretratamiento
18/230
Chapter 2: Important information
2-8 Fresenius Medical Care Service Manual SM-EN 1/03.11
2.5.4 Biological hazards
Warning
The system may only be disinfected after previous consultation with the
manufacturer of the system or by personnel authorized by the
manufacturer.
Warning
Risk of infection
Observe the applicable local laws and regulations concerning handling
of potentially infectious material.
-
8/18/2019 Manual Servicio Pretratamiento
19/230
Chapter 2: Important information
Fresenius Medical Care Service Manual SM-EN 1/03.11 2-9
2.6 Addresses
Please address any inquiries to:
Manufacturer Fresenius Medical Care AG & Co. KGaAD-61346 Bad Homburg
Germany
Phone: + 49 (0)6172 609-0
www.fmc-ag.com
Internationalservice
Fresenius Medical Care
Deutschland GmbH
Service Support International
Hafenstraße 9
D-97424 Schweinfurt
Germany
Phone: +49 (0)9721 678-333 (Hotline)Fax: +49 (0)9721 678-130
Localservice
-
8/18/2019 Manual Servicio Pretratamiento
20/230
Chapter 2: Important information
2-10 Fresenius Medical Care Service Manual SM-EN 1/03.11
-
8/18/2019 Manual Servicio Pretratamiento
21/230
Chapter 3: Component overview
Fresenius Medical Care Service Manual SM-EN 1/03.11 3-1
3 Component overview
3.1 Components of the water pretreatment system
The following pages show the two overviews of the BASIC and
STANDARD water pretreatment system.
-
8/18/2019 Manual Servicio Pretratamiento
22/230
Chapter 3: Component overview
3-2 Fresenius Medical Care Service Manual SM-EN 1/03.11
3.1.1 Water pretreatment component overview / STANDARD
F e e d w a t e r i n l e t u n i t
P r e f i l t r a t i o n – R a c k w i t h p a r t i c l e f i l t e r ( o p t i o n a l )
B i r m f
i l t e r
B o o s t e r p u m p
P a r t i c
l e f i l t e r B i g B l u e , 5 µ m (
2 0 “ )
S o f t e n e r
A
c t i v a t e d c a r b o n f i l t e r
A S t o p v a l v e
B W a t e r m e t e r ( o p t i o n a l )
C B a c k w a s h a b l e p r e f i l t e r 1 0 5 µ m
D S o l e n o i d v a l v e
E P i p e d i s c o n n e c t o r
F P r e s s u r e r e d u c e r
B 1 4 0
B 2 0 0
B 3 0 0
B 4 0 0
E 0 8 0
E 1 5 0
E 2 2 5 - T
E 3 0 0
E 4 0 0
A
0 5 0
A
1 0 0
A
1 5 0
A
3 0 0
A
4 0 0
S o f t w a t e r m o n i t o r i n g u n i t / A q u a S O F T
P a r t i c l e f i l t e
r B i g B l u e , 5 µ m (
2 0 “ )
F e e d w a t e r t a n k
T u b i n g
M e d i a f i l t e r
S a n d f i l t e r
S 1 4 0
S 2 0 0
S 3 0 0
S 4 0 0
-
8/18/2019 Manual Servicio Pretratamiento
23/230
Chapter 3: Component overview
Fresenius Medical Care Service Manual SM-EN 1/03.11 3-3
3.1.2 Water pretreatment component overview / BASIC
F e e d w a t e r i n l e t u n i t
P r e f i l t r a t i o n – R a c k w i t h p a r t i c l e f i l t e r ( o p t i o n a l )
B o o s t e r p u m p
S o f t e n e r
A
c t i v a t e d c a r b o n f i l t e r
A S t o p v a l v e
B W a t e r m e t e r ( o p t i o n a l )
C B a c k w a s h a b l e p r e f i l t e r 1 0 5 µ m
D S o l e n o i d v a l v e
E P i p e d i s c o n n e c t o r
F P r e s s u r e r e d u c e r
E 0 8 0
E 1 5 0
E 2 2 5 - T
E 3 0 0
E 4 0 0
A
0 5 0
A
1 0 0
A
1 5 0
A
3 0 0
A
4 0 0
S o f t w a t e r m o n i t o r i n g u n i t / A q u a S O F T
P a r t i c l e f i l t e
r B i g B l u e , 5 µ m (
2 0 “ )
F e e
d w a t e r t a n k
T u b i n g
B i r m f
i l t e r
B 1 4 0
B 2 0 0
B 3 0 0
B 4 0 0
M e d i a f i l t e r
S a n d f i l t e r
S 1 4 0
S 2 0 0
S 3 0 0
S 4 0 0
-
8/18/2019 Manual Servicio Pretratamiento
24/230
Chapter 3: Component overview
3-4 Fresenius Medical Care Service Manual SM-EN 1/03.11
-
8/18/2019 Manual Servicio Pretratamiento
25/230
Chapter 4: Feed water inlet unit
Fresenius Medical Care Service Manual SM-EN 1/03.11 4-1
4 Feed water inlet unit
4.1 Component overview – wall-mounted prefiltration unit
Part number: Complete assembly – description
F00002261 BASIC Feed water inlet unit 1 1/4" (Inline operation)
F00001403 BASIC Feed water inlet unit 1" (Tank operation)
POS Description
A Stop valve
B Backwashable prefilter (105 µm)
C Solenoid valve
Note:
An additional leakage detector (6328611), which switches the solenoid valve, must be implemented.
D Pipe disconnector (Inline operation)
B C DA BA C
TANK OPERATIONINLINE OPERATION
-
8/18/2019 Manual Servicio Pretratamiento
26/230
Chapter 4: Feed water inlet unit
4-2 Fresenius Medical Care Service Manual SM-EN 1/03.11
4.1.1 Prefiltration unit 1‘‘, 1 1/4‘‘, 1 1/2‘‘
Part number: Complete assembly – description
F00001116 Prefiltration unit complete 1"
F00001117 Prefiltration unit complete 1 1/4“
F00001118 Prefiltration unit complete 1 1/2“
POS Description
A Stop valve
B Backwashable prefilter
C Solenoid valve (24 V)
D Pipe disconnector
E Particle filter (Big Blue) (5 µm); 20“
F Drain kit
G Duplex softener (option)
H Electronics unit
I Leakage sensor
F
B
E
G
A
C D
H
I
-
8/18/2019 Manual Servicio Pretratamiento
27/230
Chapter 4: Feed water inlet unit
Fresenius Medical Care Service Manual SM-EN 1/03.11 4-3
4.2 Installation
4.2.1 Installation / start-up requirements
General information
4.2.2 Data reporting
Note
Start-up of the system must be performed by a trained technician. The
wiring must be adapted to fit the respective field of application.
Note
Prior to its start-up, check the system for damage caused during
transport. If there are signs of damage to the electric components, donot turn the system on.
Note
Reports for the items of the operational qualification listed beloware available as master copies in chapter 12:
– Start-up
– Setting
– Maintenance
-
8/18/2019 Manual Servicio Pretratamiento
28/230
Chapter 4: Feed water inlet unit
4-4 Fresenius Medical Care Service Manual SM-EN 1/03.11
4.3 Specifications
4.3.1 Specifications / data sheet
Rated output of RO system Tank operation Inline operation Rack operation
Feed water path
Type, size
1 “ 1 1/4 “ 1 1/2 “
Rated flow 1.5m³/h
25 L/min
2.9 m³/h
48 L/min
5.8 m³/h
97 L/min
Peak performance during softener regenera-
tion
2.4 m³/h
40 L/min
4 m³/h
67 L/min
7.8 m³/h
130 L/min
Weight
Weight without rack 12kg empty;
~17 kg; operation
17kg empty;
~21 kg; operation
22kg empty
~27 kg; operation
Operating weight approx. Weight
~42 kg
Weight
~46 kg
Weight
~52 kg
Operating data
Kv 2.2 m³/h 2.5 m³/h 6.0 m³/h
Maximum pressure 8.0 bar 8.0 bar 6.8 bar
Ambient temperature 1–45 °C 1–45 °C 1–45 °C
Water temperature 1-36 °C 1-36 °C 1-36 °C
Voltage supply 230 V 50 Hz 230 V 50 Hz 230 V 50 Hz
Floor space
Prefiltration rack
(HxWxD in mm)
1700 x 1200 x 800 1700 x 1400 x 800 1700 x 1500 x 800
Connections
Inlet and outlet 1 1/4 “
External thread
1 1/4 “
External thread
1 1/4 “
External thread
-
8/18/2019 Manual Servicio Pretratamiento
29/230
Chapter 4: Feed water inlet unit
Fresenius Medical Care Service Manual SM-EN 1/03.11 4-5
4.3.2 Operating conditions
For information regarding the operating conditions refer to the latest
edition of the Installation Guideline (IGL).
4.3.3 Configuration options
Possible configurations:
– Configuration with AquaSOFT (see chapter 10)
– Automatic backwash function
– Timer
-
8/18/2019 Manual Servicio Pretratamiento
30/230
Chapter 4: Feed water inlet unit
4-6 Fresenius Medical Care Service Manual SM-EN 1/03.11
4.4 Maintenance
Note
The information given below is summarized for all waterpretreatment components in Chapter 14 (Appendix).:
– Important information on maintenance procedures
– Accessories and supplies
– Test report - maintenance
– Overview / maintenance intervals for water pretreatment
components
– Explanations for the test report
-
8/18/2019 Manual Servicio Pretratamiento
31/230
Chapter 4: Feed water inlet unit
Fresenius Medical Care Service Manual SM-EN 1/03.11 4-7
4.5 Functional description
4.5.1 Design of components
4.5.1.1 Water inlet unit
Standard configuration
The standard configuration of a feed water inlet path consists of the stop
valve, a backwashable prefilter, a solenoid valve, a pipe disconnectorand a drain connector.
4.5.2 Component function
4.5.2.1 Stop valve
The function of the stop valve is to stop the water supply for servicing to
perform maintenance procedures. The stop valve is provided with a
sampling port for the collection of feed water samples.
4.5.2.2 Water meter - optional
An optional water meter can be installed to document the water
consumption.
4.5.2.3 Backwashable prefilter 105 µm
The backwashable prefilter is installed in the feed water path upstream
of the solenoid valve to prevent contamination and therefore an early
failure of the solenoid valve.
Water inlet unit
Components:
A) Stop valve
B) Water meter (optional)
C) Backwashable prefilter (105 µm)
D) Solenoid valve
E) Pipe disconnector
F) Pressure reducer (optional)
F) Drain
-
8/18/2019 Manual Servicio Pretratamiento
32/230
Chapter 4: Feed water inlet unit
4-8 Fresenius Medical Care Service Manual SM-EN 1/03.11
In its standard version, the backwashable prefilter has 105 µm filterinserts. As an option this prefilter can also be designed with a 50 µm filter insert.
As the first filter in the complete water pretreatment system, this insert
is usually exposed to a high concentration of particles. Regual
backwashing of the filter during non-dialysis hours must therefore be
ensured to prevent early pressure drops. This can either be done
manually (by turning the rotary pin 180° clockwise) or with the automatic
backwashing feature.
The automatic backwashing feature is available as an option. It should
always be used in combination with a timer to be able to define a fixed
backwash time.
4.5.2.4 Solenoid valve
Function The feed water inlet solenoid valve is used to disconnect the watersupply in case of a leakage. In case of rack installation it is activated bya single-station leakage detection or in case of wall-mounted installation
by a leakage detector.
Usage as tank fill valve In combination with a feed water tank, the feed water inlet solenoidvalve is also used as a tank fill valve. In this case, the solenoid valve is
controlled by a float switch which is supplied with the feed water tank.
Manual opening If necessary, the solenoid valve can be opened manually.
In case of a power failure push the lever in and turn it by 90°. In this po-sition it is automatically locked and the valve is permanently open.
Water supply is therefore still ensured even without power supply.
If the solenoid valve is manually bypassed, the room
must never be left unattended.Otherwise the valve must be left in its original position.
Normal condition
(original position)
Valve open
(manually bypassed)
Note
Without power supply, leakage monitoring is not active.
After dialysis hours, the valve should be closed again by unlocking to
stop the water supply in case of leakage.
-
8/18/2019 Manual Servicio Pretratamiento
33/230
Chapter 4: Feed water inlet unit
Fresenius Medical Care Service Manual SM-EN 1/03.11 4-9
4.5.2.5 Pipe disconnector
If the inlet pressure drops, the pipe disconnector ensures a
disconnection from the system (DIN EN 1717) as requested by the
Drinking Water Decree if there is a risk of back-contamination.
-
8/18/2019 Manual Servicio Pretratamiento
34/230
Chapter 4: Feed water inlet unit
4-10 Fresenius Medical Care Service Manual SM-EN 1/03.11
4.5.3 Prefiltration rack with particle filter (optional)
4.5.4 Component description
Stop valve (see chapter 4.5.2.1 , page 4-7)
Backwashable prefilter (see chapter 4.5.2.3 , page 4-7)
Solenoid valve (see chapter 4.5.2.4 , page 4-8)
Pipe disconnector (see chapter 4.5.2.5 , page 4-9)
Particle filter (Big Blue)5 µm
see chapter 8 (particle filter)
Drain kit The drain kit is used to direct the draining water to the drain in thebuilding.
Duplex softener (option) see chapter 9 (Softener)
Electronics unit The electronics unit supplies the connected components with electricalvoltage.
Leakage sensor The leakage sensor detects (inadvertent) water leakage. In this case,the connected valve is closed. An audible alarm and a red LED indicator
signal a leakage alarm.
Prefiltration rack with particle filter (optional)
Components:
– Stop valve
– Backwashable prefilter
– Solenoid valve (24 V)
– Pipe disconnector
– Particle filter (Big Blue) (5 µm); 20“
– Drain kit
– Duplex softener (option)
– Electronics unit
– Leakage sensor
-
8/18/2019 Manual Servicio Pretratamiento
35/230
Chapter 4: Feed water inlet unit
Fresenius Medical Care Service Manual SM-EN 1/03.11 4-11
4.6 Options
4.6.1 Prefiltration rack with particle filter (option)
4.6.2 Pressure reducer - optional
An optional pressure reducer can be installed in the feed water path. Its
function is to avoid pressure peaks and thus to limit the system load. A
pressure reducer is recommended if the static pressure at the user
points exceeds 5 bar .
The installation of a pressure reducer is recommended in case of
variations in pressure caused be pressure increases / domestic water
supply system. As an option, a domestic water station can be used
instead of a backwashable prefilter. It includes the backwashable
prefilter and a pressure reducer unit.
4.6.3 Configuration options (feed water tank)
If a feed water tank is installed, it will have the requested system
disconnection. In this case the installation of a pipe disconnector is not
required.
-
8/18/2019 Manual Servicio Pretratamiento
36/230
Chapter 4: Feed water inlet unit
4-12 Fresenius Medical Care Service Manual SM-EN 1/03.11
-
8/18/2019 Manual Servicio Pretratamiento
37/230
Chapter 5: Booster pump
Fresenius Medical Care Service Manual SM-EN 1/03.11 5-1
5 Booster pump
5.1 Booster pump assembly CRN5-8 component overview
Part number: Complete assembly – description
F00001935 Booster pump assembly (vertical pumps CRN5-8)
POS Description
A Connection to water supply(Possible on both sides. The connection which is not used must be provided with a cap nut)
B1 Pump 1 (vertical pump CRN5-8)
B2 Pump 2 (vertical pump CRN5-8)
C Control unit
D Pressurizer vessel
W2
A
B1
D
C
-
8/18/2019 Manual Servicio Pretratamiento
38/230
Chapter 5: Booster pump
5-2 Fresenius Medical Care Service Manual SM-EN 1/03.11
5.2 Installation
5.2.1 Installation / start-up requirements
General information
5.2.2 Important start-up information
5.2.2.1 Booster pump assembly connection
The suction line can be connected either on the right or on the outer
left of the booster pump assembly. In case of tank operation it is
recommended to connect two suction tubes in parallel to avoid
pressure drops in the suction line.
The pressure port is located in the center and can be connected
using a tubing or a pipe for a fixed connection. A high-pressure
tubing is included in the accessories package of the system.
After connection to the water supply, both pumps must be ventilated
by opening the vent screws.
Note
Start-up of the system must be performed by a trained technician. The
wiring must be adapted to fit the respective field of application.
Note
Prior to its start-up, check the system for damage caused during
transport. If there are signs of damage to the electric components, donot turn the system on.
Booster pump assembly (CRN5-8)
The pumps must be ventilated prior to start-up of the switching unit.
-
8/18/2019 Manual Servicio Pretratamiento
39/230
Chapter 5: Booster pump
Fresenius Medical Care Service Manual SM-EN 1/03.11 5-3
5.2.2.2 Connection to the domestic water supply network
Positioning the pressure switch
The pressure switch installed at the water inlet ensures protection
against running dry. Upon delivery, the pressure switch is set to 1 bar.
It is recommended to retain this setting. If the pressure in the water inlet
drops below 1 bar, the booster pump assembly will be stopped. If this
dynamic minimum pressure is not reached, it is recommended to use a
feed water tank.
Pressure switch installation position
Note
If a pipe separator or a pressure reducer is installed on the outlet side
of the booster pump, the pressure sensor must always be installed after
these components. For the installation, remove the screw-type adapter
in the outlet block and integrate it after the pipe disconnector.
State upon delivery: Dismounting the adapter
State upon delivery: When the system is supplied, the pressuresensor is installed with an adapter at the pump outlet, this adapter
must be removed and must be installed after the pressure reducer
and the pipe separator.
The pressure sensor is installed on the outlet side after the pressure
reducer and the pipe separator.
Pressure switch installation position and connection (run-dry protection)
Pressure switch installation position (run-dry protection)
-
8/18/2019 Manual Servicio Pretratamiento
40/230
Chapter 5: Booster pump
5-4 Fresenius Medical Care Service Manual SM-EN 1/03.11
Booster pump assembly with pump switching
Pressure switch installation position and connection
Pressure switch installation position (after pipe separator and
pressure reducer)
Pressure switch installation position and connection (run-dry protection)
Electrical connection of the pressure switch for the run-dry protection
Pressure switch installation position and connection
Electrical connection of the pressure sensor
-
8/18/2019 Manual Servicio Pretratamiento
41/230
Chapter 5: Booster pump
Fresenius Medical Care Service Manual SM-EN 1/03.11 5-5
Verification / setting of the pressure switch
Connection of the pressure switch
Pressure switch connection
Connection 1: white
Connection 2: brown
Pressure sensor connection
Pressure sensor connection Connection 1: brown
Connection 2: white
Reference point on the scale
Locate the reference point.
-
8/18/2019 Manual Servicio Pretratamiento
42/230
Chapter 5: Booster pump
5-6 Fresenius Medical Care Service Manual SM-EN 1/03.11
Rotate the selector fully counter-clockwise.
The selector must first be rotated fully counter-clockwise.
Adjust the selector to the reference point.
Rotate the selector approx. ¼ of a turn clockwise until the figure "1"
is above the reference point.
Adjust the selector to the reference point.
Adjust the scale reference point.
The pressure switch is now set to 1 bar. This means the operatingpressure of the pump is 1 bar.
Due to the pressure switch's hysteresis, the switch-off pressure is
slightly below 1 bar (approx. 0.5 bar).
-
8/18/2019 Manual Servicio Pretratamiento
43/230
Chapter 5: Booster pump
Fresenius Medical Care Service Manual SM-EN 1/03.11 5-7
5.2.2.3 Connection to a feed water tank
Note
In case of tank operation, the tank must be provided with a float switchwhich will open a contact if the tank is empty. This contact is connected
in the pump inlet instead of the pressure switch.
Tank connection
The inlet of the booster pump assembly is directly connected to the
outlet of the tank using the pump connection set no. 6319211.
The suction line can be connected either on the right or on the outer
left of the booster pump assembly.
In case of tank operation it is recommended to connect two suction
tubes in parallel to avoid pressure drops in the suction line. The pressure port is located in the center and can be connected
using a tubing or a pipe for a fixed connection. A high-pressure tubing
with a length of x cm is included in the accessories package of the
system.
The pressure sensor is connected directly to the pump outlet.
The inlet pressure switch on the water inlet side of the pump is not
used during tank operation. In its place the float switch (see
illustration) must be wired in. This must be done during installation.
Pressure switch installation position and connection (run-dry protection)
A float switch which is installed in the feed water tank ensures run-
dry protection.
The brown and white conductor must be connected.
-
8/18/2019 Manual Servicio Pretratamiento
44/230
Chapter 5: Booster pump
5-8 Fresenius Medical Care Service Manual SM-EN 1/03.11
Pressure switch installation position and connection
Installation position of the pressure sensor on the pump outlet (state
upon delivery)
Electrical connection of the pressure sensor in the switching cabinet.
-
8/18/2019 Manual Servicio Pretratamiento
45/230
Chapter 5: Booster pump
Fresenius Medical Care Service Manual SM-EN 1/03.11 5-9
5.2.2.4 Power supply
A 400 V; 3 N ~; 50 Hz; 16 A three-phase current connection is required
to supply the pump switching unit with voltage.
Connection of a leakage detector
The floating contact of a leakage detection system can be connected to
the booster pressure assembly. The contact must be closed if there is
no leakage and must open if a leakage occurs.
It is recommended to use the insert of the leakage detecting system. It
can be connected directly to the pump control (see illustration )
5.2.2.5 Overview of electrical connections
Connection option for a leakage detection system (jumper must be removed)
Remove the jumper and connect the line to the floating contacts of
the leakage detector.
Note
When the system is supplied, positions 1 to 6 are wired.
Positions 7 and 8 must be adjusted by the technician depending on
the respective situation (feed water tank or direct operation).
-
8/18/2019 Manual Servicio Pretratamiento
46/230
Chapter 5: Booster pump
5-10 Fresenius Medical Care Service Manual SM-EN 1/03.11
POS Connections
1 Pump 1
2 Pump 2
3 Power cable
4 Temperature switch 1
5 Temperature switch 2
6 Pressure sensor
7 Connection for run-dry protection (pressure switch or float switch)
8 Connection option for the floating contact (e.g. leakage detector)
1 2 3 4 5 6
7
8
-
8/18/2019 Manual Servicio Pretratamiento
47/230
Chapter 5: Booster pump
Fresenius Medical Care Service Manual SM-EN 1/03.11 5-11
5.2.2.6 Settings on the booster pump assembly (CRN5-8)
Should the booster pump assembly not operate as desired when it is
supplied, the setting of the pressure level and of the delay times can be
checked and changed.
Pressure levels
Pressuretype
Description
Pmin Operating pressure for base load and peak load pump (0-10 bar):
Pmin is set to the pressure at which the pump is to start.
Pmax1 Cutoff pressure level for the peak load pump (difference to Pmin), (0-4 bar):
If this pressure is achieved at parallel operation, the peak load pump is turned off. Pmax1 should
be set to a value between Pmin and Pmax2.
Pmax2 Cutoff pressure level for the base load pump (difference to Pmax1), (0-2 bar):
If this pressure is reached during operation of one pump, this pump is switched off. Pmax2
should be set to approx. 0.5 bar below the pump's nominal pressure.
Caution:If a pressure reducer was installed downstream of the booster pump assembly, the pressure
reducer should be set to 6 bar minimum as the pressure on the pressure sensor can otherwise
never reach the cutoff pressure.
Note
The settings of max1 and Pmax
2 are difference values which add to the
lower pressure set.
-
8/18/2019 Manual Servicio Pretratamiento
48/230
Chapter 5: Booster pump
5-12 Fresenius Medical Care Service Manual SM-EN 1/03.11
Calculation example
Setting Pmin: 3 bar
Setting Pmax1: 2 bar
Setting Pmax2: 1 bar
This means Operating pressure level (Pmin) = 3 bar
Cutoff pressure level for peak load pump (Pmax1) = 5 bar
Cutoff pressure level for base load pump (Pmax
2) = 6 bar
Delay times
T = Over-run (0–2 min):
Minimum operation time of the pump to prevent permanent start and
stop of the pump. Default setting: 2 minutes. When set to "0 minutes",the minimum operation time is approx. 30 seconds.
T = run-dry delay (0–2 min):
time after which die pumps are turned off in case the pump runs dry.
Default setting: 0 minutes.
When set to "0 minutes", the delay time is approx. 10 seconds.
Motor protection
Potentiometers P1 and P2 are set to the nominal current of the motor.
Motor protection setting (nominal current of
pump 1 and pump 2)
Setting Pmax2: Cutoff pressure level for base
load pump (difference to Pmax1)
Setting Pmax1: Cutoff pressure for peak loadpump (difference to Pmin)
Setting Pmin: Operating pressure for baseload and peak load pump
Minimum operation time setting
Water alarm delay time setting
-
8/18/2019 Manual Servicio Pretratamiento
49/230
Chapter 5: Booster pump
Fresenius Medical Care Service Manual SM-EN 1/03.11 5-13
5.2.3 Data reporting
Note
Documentation of the settings in the Settings report
As part of the Operational Qualification, the settings must be
documented in the Settings reports upon start-up.
Note
Reports for the items of the operational qualification listed beloware available as master copies in chapter 12:
– Start-up
– Setting
– Maintenance
-
8/18/2019 Manual Servicio Pretratamiento
50/230
Chapter 5: Booster pump
5-14 Fresenius Medical Care Service Manual SM-EN 1/03.11
5.3 Specifications
5.3.1 Specifications / data sheet
5.3.1.1 Performance characteristics
Parameter Function / specifications
Usage Pressure increase
Component Booster pump assembly
Type CRN5-8
Connection 1 1/4", external thread
Sealing material Shaft seal carbon silicide, o-ring EPDM
Installation position Vertical
Operating current 2.5 kW / kVA
Power supply 380–400 V; 3N~ 50 Hz; 16A
Operating pressure, adjustable 1–6 bar (preset to x bar)
Switch-off pressure 1–6 bar (preset to x bar)
Pressure, minimum 2.2 bar (in case of tank operation)
Pressure, maximum 6.5 bar
Water temperature, minimum 1 °C
Water temperature, maximum 40 °C
Flow rate 5.7 m³/h at 3.8 bar (at a nominal pumping head of 37.8m)
Ambient temperature, max. 40°C
Parameter Function / specifications
Electrical connection 380–400 V 3N~ 50 Hz
Power consumption 2.5 kW / kVA
Operating pressure Adjustable 1–6 bar (preset to x bar)
Cutoff pressure Adjustable 1–6 bar (preset to x bar)
Dimensions (width/ height/depth) Width / height / depth 101 cm / 92 cm / 45 cm
-
8/18/2019 Manual Servicio Pretratamiento
51/230
Chapter 5: Booster pump
Fresenius Medical Care Service Manual SM-EN 1/03.11 5-15
5.3.1.2 Volumetric capacity - diagram
5.3.2 Operating conditions
For information regarding the operating conditions refer to the latest
edition of the Installation Guideline (IGL).
(one pump active) (two pumps active)
Delivery volume
P u m p i n g h e a d
-
8/18/2019 Manual Servicio Pretratamiento
52/230
Chapter 5: Booster pump
5-16 Fresenius Medical Care Service Manual SM-EN 1/03.11
5.4 Maintenance
Note
The information given below is summarized for all waterpretreatment components in Chapter 14 (Appendix):
– Important information on maintenance procedures
– Accessories and supplies
– Test report – maintenance
– Overview / maintenance intervals for water pretreatment
components
– Explanations for the test report
-
8/18/2019 Manual Servicio Pretratamiento
53/230
Chapter 5: Booster pump
Fresenius Medical Care Service Manual SM-EN 1/03.11 5-17
5.5 Error messages
5.5.1 Malfunctions, causes and correction
Setting the system to the Standby mode
Components:
A) Operating mode selector switch
B) Operation / malfunction indicator (pump 1 and pump 2)
C) Run-dry / leakage indicator
D) Main power switch
Selection Description
Green LED flashing. The motor protection has operated. No automatic confirmation aftercorrection of the problem.
Possible causes:
– Motor current too high.
– The motor temperature is too high, possible defective pressure
sensor causing permanent operation and overheating of the pump.
Confirmation:
Set the operating mode selector switch to "OFF" and then back to
"ON" again. If necessary, allow the motor to cool down first.
Red LED illuminated System was turned off due to lack of water or leakage. Automaticconfirmation after correction of the problem.
B
D
C
A
-
8/18/2019 Manual Servicio Pretratamiento
54/230
Chapter 5: Booster pump
5-18 Fresenius Medical Care Service Manual SM-EN 1/03.11
5.6 Functional description
5.6.1 Design of components
5.6.2 Component description
The booster pump assembly (CRN5-8) is an assembly which is ready
for connection in accordance with the requirements for the minimum
flow pressure with normal centrifugal suction pumps.
The booster pump assembly in conjunction with the two vertical pumpsprovides increased operational safety of a water treatment system from
Fresenius Medical Care.
The booster pump assembly is used for the alternating control of two
booster pumps. Automatic switching ensures uniform distribution of
stress on both connected pumps and prevents damage to the pump
caused by prolonged idle times when it is not in use. To increase the
efficiency, the second pump will be turned on additionally in peak
situations.
The booster pump assembly is intended for the pressure increase in the
feed water section of the water pretreatment system. It can be used
both for increasing the pressure downstream of a feed water tank andfor direct connection to the feed water line.
5.6.2.1 Control unit
The electronic control of the booster pump assembly controls and
monitors the two connected pumps.
Booster pump assembly (CRN5-8)
Components:
A) Feed water connection
B1) Pump 1 (vertical pump CRN5-8)
B2) Pump 2 (vertical pump CRN5-8)
C) Control unit
D) Pressurizer vessel
B2
D
C
A
B1
-
8/18/2019 Manual Servicio Pretratamiento
55/230
Chapter 5: Booster pump
Fresenius Medical Care Service Manual SM-EN 1/03.11 5-19
5.6.2.2 Base load and peak load pump
The control differentiates between a base load and peak load pump. If
water requirements are low, only the base load pump will deliver, if
water requirements rise the peak load pump will also be switched on.
This concept has the advantage that the pumps are always operated intheir optimum performance range. This contributes to a high efficiency
of the complete system. The base load and the peak load pump will be
activated alternating after each restart. This ensures uniform wear of
both pumps.
5.6.2.3 Alternating pressure situations
If the system is not in operation and the pressure in the system drops
below the start-up pressure level (Pmin), the base load pump will start.
If the pressure drops again to the start-up pressure (Pmin) caused by
an increasing water demand, the peak load pump will also be started.If, with decreasing water demand, the pressure in the system rises
above the pressure level (Pmax 1), the peak load pump will be stopped.
If water consumption stops, the pressure will rise above the pressure
level (Pmax 2) and the basic load pump will be stopped.
5.6.2.4 Run-dry protection
The adjustable pressure switch at the booster pump assembly inlet
protects the system against running dry and ensures a minimum
pressure of 1 bar at the booster pump inlet.
If the pressure at the inlet drops below 1 bar, first the peak load pumpwill stop. If the pressure at the inlet remains below 1 bar after this pump
stop, the base load pump will also be stopped.
5.6.2.5 Tank operation
When operated with a tank, a float switch will be connected instead of
the pressure to ensure protection against running dry.
5.6.2.6 Automatic turnoff
The water temperature is also monitored; if the temperature exceeds
60 °C, e.g. by a defect of the control unit and a constantly running pump
without water consumption, the respective pump will be stopped.
-
8/18/2019 Manual Servicio Pretratamiento
56/230
Chapter 5: Booster pump
5-20 Fresenius Medical Care Service Manual SM-EN 1/03.11
5.6.3 Functional description
5.6.3.1 Setting the system to the Standby mode
5.6.3.2 Selector switch functions
Setting the system to the Standby mode
Components:
A) Operating mode selector switch
B) Operation / malfunction indicator (pump 1 and pump 2)
C) Run-dry/leakage indicator
D) Main power switch
After turning on the main power switch and after selection of
automatic 1 + 2 (position as in illustration), the operating modeswitch is used to set the system to the Standby mode.
Functions of the selector switch
Components:
A) Test mode pump 2
B) Off
A) Test mode pump 1
D) Automatic mode pump 1
E) Automatic mode pumps 1 + 2
D) Automatic mode pump 2
Selection Description
Automatic mode pump 1 Only pump 1 runs in the automatic mode (e.g. if pump 2 is defective).
Automatic mode pumps 1+2 Both pumps run in the automatic mode. Depending on the consumptionvolume the pumps are started and stopped (default setting). The green
LED of the active pump is constantly illuminated to indicate operation.
Automatic mode pump 2 Only pump 2 runs in the automatic mode (e.g. if pump 1 is defective).
B
D
C
A
B
ED
A
F
C
-
8/18/2019 Manual Servicio Pretratamiento
57/230
Chapter 5: Booster pump
Fresenius Medical Care Service Manual SM-EN 1/03.11 5-21
Test mode pump 1 Pump 1 runs irrespective of the prevailing pressures and safetyfunctions. The test mode will be automatically stopped after approx. 1.5
minutes.
Test mode pump 2 Pump 2 runs irrespective of the prevailing pressures and safetyfunctions. The test mode will be automatically stopped after approx. 1.5
minutes.
Off Both pumps are off.
Selection Description
-
8/18/2019 Manual Servicio Pretratamiento
58/230
Chapter 5: Booster pump
5-22 Fresenius Medical Care Service Manual SM-EN 1/03.11
-
8/18/2019 Manual Servicio Pretratamiento