Manual Servicio Pretratamiento

download Manual Servicio Pretratamiento

of 230

Transcript of Manual Servicio Pretratamiento

  • 8/18/2019 Manual Servicio Pretratamiento

    1/230

    FreseniusMedicalCare

    Pretreatment ComponentsBASIC / STANDARD

    Service ManualEdition: 1/03.11

    Part no.: 636 210 1

  • 8/18/2019 Manual Servicio Pretratamiento

    2/230

  • 8/18/2019 Manual Servicio Pretratamiento

    3/230

    Fresenius Medical Care Service Manual SM-EN 1/03.11 iii

    Table of contents

    1 Index

    2 Important information

    2.1 Preface........................................................................................................................................ 2-12.1.1 Structure of the Service Manual .................................................................................................. 2-1

    2.2 How to use the Service Manual ................................................................................................ 2-2

    2.3 Notes........................................................................................................................................... 2-32.3.1 Significance of the warnings........................................................................................................ 2-3

    2.3.2 Significance of the note ............................................................................................................... 2-32.3.3 Significance of the tip .................................................................................................................. 2-3

    2.4 Brief description ........................................................................................................................ 2-42.4.1 Operator information.................................................................................................................... 2-5

    2.4.2 Guarantee / warranty................................................................................................................... 2-6

    2.4.3 Technical documentation............................................................................................................. 2-6

    2.5 Warnings .................................................................................................................................... 2-72.5.1 Basic warnings ............................................................................................................................ 2-7

    2.5.2 Electric hazards ........................................................................................................................... 2-7

    2.5.3 Chemical hazards........................................................................................................................ 2-7

    2.5.4 Biological hazards ....................................................................................................................... 2-8

    2.6 Addresses .................................................................................................................................. 2-9

    3 Component overview

    3.1 Components of the water pretreatment system ..................................................................... 3-13.1.1 Water pretreatment component overview / STANDARD........................................................ 3-23.1.2 Water pretreatment component overview / BASIC................................................................. 3-3

    4 Feed water inlet unit4.1 Component overview – wall-mounted prefiltration unit ........................................................ 4-14.1.1 Prefiltration unit 1‘‘, 1 1/4‘‘, 1 1/2‘‘ ................................................................................................ 4-2

    4.2 Installation.................................................................................................................................. 4-34.2.1 Installation / start-up requirements .............................................................................................. 4-3

    4.2.2 Data reporting.............................................................................................................................. 4-3

    4.3 Specifications ............................................................................................................................ 4-44.3.1 Specifications / data sheet........................................................................................................... 4-4

    4.3.2 Operating conditions.................................................................................................................... 4-5

    4.3.3 Configuration options................................................................................................................... 4-5

    4.4 Maintenance............................................................................................................................... 4-6

  • 8/18/2019 Manual Servicio Pretratamiento

    4/230

    iv Fresenius Medical Care Service Manual SM-EN 1/03.11

    4.5 Functional description .............................................................................................................. 4-74.5.1 Design of components................................................................................................................. 4-7

    4.5.1.1 Water inlet unit ............................................................................................................. 4-7

    4.5.2 Component function..................................................................................................................... 4-7

    4.5.2.1 Stop valve..................................................................................................................... 4-7

    4.5.2.2 Water meter - optional.................................................................................................. 4-74.5.2.3 Backwashable prefilter 105 µm .................................................................................... 4-7

    4.5.2.4 Solenoid valve.............................................................................................................. 4-8

    4.5.2.5 Pipe disconnector ......................................................................................................... 4-9

    4.5.3 Prefiltration rack with particle filter (optional) ............................................................................. 4-10

    4.5.4 Component description.............................................................................................................. 4-10

    4.6 Options ..................................................................................................................................... 4-114.6.1 Prefiltration rack with particle filter (option)................................................................................ 4-11

    4.6.2 Pressure reducer - optional ....................................................................................................... 4-11

    4.6.3 Configuration options (feed water tank)..................................................................................... 4-11

    5 Booster pump

    5.1 Booster pump assembly CRN5-8 component overview ........................................................ 5-1

    5.2 Installation.................................................................................................................................. 5-25.2.1 Installation / start-up requirements .............................................................................................. 5-2

    5.2.2 Important start-up information...................................................................................................... 5-2

    5.2.2.1 Booster pump assembly connection ............................................................................ 5-2

    5.2.2.2 Connection to the domestic water supply network....................................................... 5-3

    5.2.2.5 Overview of electrical connections ............................................................................... 5-9

    5.2.3 Data reporting ............................................................................................................................ 5-13

    5.3 Specifications .......................................................................................................................... 5-145.3.1 Specifications / data sheet......................................................................................................... 5-145.3.1.1 Performance characteristics....................................................................................... 5-14

    5.3.1.2 Volumetric capacity - diagram .................................................................................... 5-15

    5.3.2 Operating conditions.................................................................................................................. 5-15

    5.4 Maintenance ............................................................................................................................. 5-16

    5.5 Error messages........................................................................................................................ 5-175.5.1 Malfunctions, causes and correction ......................................................................................... 5-17

    5.6 Functional description ............................................................................................................ 5-185.6.1 Design of components............................................................................................................... 5-18

    5.6.2 Component description.............................................................................................................. 5-18

    5.6.2.1 Control unit ................................................................................................................. 5-185.6.2.2 Base load and peak load pump.................................................................................. 5-19

    5.6.2.3 Alternating pressure situations ................................................................................... 5-19

    5.6.2.4 Run-dry protection...................................................................................................... 5-19

    5.6.2.5 Tank operation ........................................................................................................... 5-19

    5.6.2.6 Automatic turnoff ........................................................................................................ 5-19

    5.6.3 Functional description................................................................................................................ 5-20

    5.6.3.1 Setting the system to the Standby mode.................................................................... 5-20

    5.6.3.2 Selector switch functions............................................................................................ 5-20

  • 8/18/2019 Manual Servicio Pretratamiento

    5/230

    Fresenius Medical Care Service Manual SM-EN 1/03.11 v

    6 Pumps

    6.1 Pump CM5-5 – component overview ....................................................................................... 6-1

    6.2 Installation.................................................................................................................................. 6-2

    6.2.1 Installation / start-up requirements .............................................................................................. 6-26.2.2 Important start-up information ..................................................................................................... 6-2

    6.2.3 Data reporting.............................................................................................................................. 6-2

    6.3 Specifications ............................................................................................................................ 6-36.3.1 Specifications / data sheet (CM5-5) ............................................................................................ 6-3

    6.3.2 Operating conditions.................................................................................................................... 6-3

    6.4 Maintenance............................................................................................................................... 6-4

    6.5 Error messages on pumps ....................................................................................................... 6-56.5.1 Visual indication on CM5-5.......................................................................................................... 6-5

    6.5.1.1 Alarm messages overview ........................................................................................... 6-5

    6.5.2 Pump malfunctions overview....................................................................................................... 6-6

    6.6 Functional description .............................................................................................................. 6-76.6.1 Design of components................................................................................................................. 6-7

    7 Media filters (sand filters / Birm filters)

    7.1 Component overview – sand filters S140, S200, S300, S400 ................................................ 7-1

    7.2 Component overview – Birm filters B140, B200, B300, B400 ................................................ 7-2

    7.3 Installation of the media filters (sand / Birm).......................................................................... 7-37.3.1 Installation / start-up requirements .............................................................................................. 7-3

    7.3.2 Important start-up information ..................................................................................................... 7-47.3.2.1 Start-up of the sand filters / Birm filters ........................................................................ 7-4

    7.3.2.2 Valve heads and drain restrictor ................................................................................... 7-5

    7.3.2.3 Valves in the control head............................................................................................ 7-9

    7.3.3 Programming the control for sand filters / Birm filters................................................................ 7-11

    7.3.3.1 Control / display ......................................................................................................... 7-11

    7.3.3.2 Control / control elements .......................................................................................... 7-11

    7.3.3.3 Control / programming................................................................................................ 7-11

    7.3.4 Data reporting............................................................................................................................ 7-23

    7.3.5 Use of Birm for improved water quality..................................................................................... 7-24

    7.3.5.1 Removal of iron using "Birm" as filter media in the sand filter .................................... 7-24

    7.3.5.2 Alarm limits for water substances............................................................................... 7-24

    7.3.6 Birm filter settings ...................................................................................................................... 7-26

    7.3.6.1 Weekly / semi-weekly backwash................................................................................ 7-26

    7.3.7 Operating conditions.................................................................................................................. 7-28

    7.4 Specifications .......................................................................................................................... 7-297.4.1 Specifications / sand filter data sheet........................................................................................ 7-29

    7.4.2 Specifications / Birm filter data sheet ........................................................................................ 7-30

    7.4.3 Operating conditions.................................................................................................................. 7-31

    7.5 Maintenance............................................................................................................................. 7-32

    7.6 Error messages ....................................................................................................................... 7-337.6.1 Sand filter / Birm filter malfunctions overview............................................................................ 7-33

    7.7 Functional description ............................................................................................................ 7-34

    7.7.1 Design of components............................................................................................................... 7-34

  • 8/18/2019 Manual Servicio Pretratamiento

    6/230

    vi Fresenius Medical Care Service Manual SM-EN 1/03.11

    7.7.2 Component description.............................................................................................................. 7-34

    7.7.2.1 Backwash ................................................................................................................... 7-34

    7.7.2.2 Flow velocities............................................................................................................ 7-35

    7.7.2.3 Feed water bypass..................................................................................................... 7-35

    7.7.2.4 Efficiency .................................................................................................................... 7-35

    8 Particle filter (Big Blue) 20"

    8.1 Component overview – particle filter ....................................................................................... 8-1

    8.2 Installation.................................................................................................................................. 8-28.2.1 Installation / start-up requirements .............................................................................................. 8-2

    8.2.2 Data reporting .............................................................................................................................. 8-2

    8.3 Specifications ............................................................................................................................ 8-38.3.1 Specifications / data sheet........................................................................................................... 8-3

    8.3.2 Operating conditions.................................................................................................................... 8-3

    8.3.2.1 Particle filter selection table.......................................................................................... 8-3

    8.4 Maintenance ............................................................................................................................... 8-4

    8.5 Functional description .............................................................................................................. 8-58.5.1 Design of components................................................................................................................. 8-5

    8.5.2 Component description................................................................................................................ 8-5

    9 Softener  

    9.1 Component overview – softener E080(B), E150(B), E225-T, E300, E400 Magnum .............. 9-1

    9.2 Modification of the softener settings in case of high brine concentrations in combinationwith Kinetico-Autorol retrofit kits............................................................................................. 9-3

    9.2.1 Settings for CP 216...................................................................................................................... 9-3

    9.2.2 Settings for CP 213...................................................................................................................... 9-3

    9.2.3 Settings for CP 210...................................................................................................................... 9-4

    9.3 Installation.................................................................................................................................. 9-59.3.1 Installation / start-up requirements .............................................................................................. 9-5

    9.4 Display and operating elements............................................................................................... 9-69.4.1 Display during operation .............................................................................................................. 9-6

    9.4.2 Programming the control unit....................................................................................................... 9-7

    9.4.2.1 Control / display............................................................................................................ 9-7

    9.4.2.2 Programming the control .............................................................................................. 9-79.4.3 Important start-up information.................................................................................................... 9-13

    9.4.3.1 Softener start-up procedure ....................................................................................... 9-14

    9.4.4 Start-up procedure for E080(B),/ E150(B), E225-T, E300, E400............................................... 9-16

    9.4.4.3 Brine tank ................................................................................................................... 9-22

    9.4.4.5 Triplex operation E225-T............................................................................................ 9-23

    9.4.4.6 Alternating or parallel operation ................................................................................. 9-24

    9.4.5 Data reporting ............................................................................................................................ 9-44

    9.5 Specifications .......................................................................................................................... 9-459.5.1 Specifications / data sheet......................................................................................................... 9-46

    9.5.2 Restrictor inserts depending on pressure range........................................................................ 9-47

    9.5.3 Softener capacities .................................................................................................................... 9-48

    9.5.4 Operating conditions.................................................................................................................. 9-49

  • 8/18/2019 Manual Servicio Pretratamiento

    7/230

    Fresenius Medical Care Service Manual SM-EN 1/03.11 vii

    9.5.4.1 Softener selection table.............................................................................................. 9-49

    9.5.4.2 Triplex softener (E225-T) ........................................................................................... 9-50

    9.5.5 Removal of iron and manganese using a softener .................................................................... 9-50

    9.6 Maintenance............................................................................................................................. 9-51

    9.7 Error messages ....................................................................................................................... 9-529.7.1 Softener malfunctions overview................................................................................................. 9-52

    9.8 Functional description ............................................................................................................ 9-559.8.1 Design of components............................................................................................................... 9-55

    9.8.2 Component description.............................................................................................................. 9-55

    9.8.3 Functional description Triplex operation E225-T....................................................................... 9-56

    10 Activated carbon filter 

    10.1 Component overview – A050, A100, A150, A300, A400 ....................................................... 10-1

    10.2 Installation................................................................................................................................ 10-210.2.1 Installation / start-up requirements ............................................................................................ 10-2

    10.2.2 Important start-up information ................................................................................................... 10-3

    10.2.3 Start-up of the activated carbon filter ......................................................................................... 10-4

    10.2.3.1 Valve heads and drain restrictor ................................................................................. 10-4

    10.2.3.3 Control unit design ..................................................................................................... 10-8

    10.2.4 Programming the control unit .................................................................................................. 10-10

    10.2.4.1 Control / display ....................................................................................................... 10-10

    10.2.4.2 Control / operating elements .................................................................................... 10-10

    10.2.4.3 Control unit ............................................................................................................... 10-11

    10.2.5 Data reporting.......................................................................................................................... 10-21

    10.3 Specifications ........................................................................................................................ 10-2210.3.1 Specifications / data sheet....................................................................................................... 10-2210.3.2 Operating conditions................................................................................................................ 10-24

    10.3.2.1 Activated carbon filter (Single) selection table ......................................................... 10-25

    10.3.2.2 Activated carbon filter (double) selection table......................................................... 10-25

    10.4 Maintenance........................................................................................................................... 10-27

    10.5 Error messages ..................................................................................................................... 10-2810.5.1 Activated carbon filter malfunctions overview.......................................................................... 10-28

    10.6 Functional description .......................................................................................................... 10-2910.6.1 Design of components............................................................................................................. 10-29

    10.6.2 Component description............................................................................................................ 10-29

    11 USER RELEVANT – Operational data acquisition / maintenance

    11.1 Operational data acquisition / daily report............................................................................ 11-2

    11.2 System handling instructions report – water pretreatment ................................................ 11-5

    11.3 Maintenance............................................................................................................................. 11-911.3.1 Maintenance intervals for water pretreatment components....................................................... 11-9

    11.3.1.1 Feed water inlet unit ................................................................................................... 11-9

    11.3.1.2 Sand filter ................................................................................................................... 11-9

    11.3.1.3 Particle filter ............................................................................................................... 11-9

    11.3.1.4 Softener .................................................................................................................... 11-10

  • 8/18/2019 Manual Servicio Pretratamiento

    8/230

    viii Fresenius Medical Care Service Manual SM-EN 1/03.11

    11.3.1.5 Activated carbon filter ............................................................................................... 11-11

    11.3.1.6 Water leakage indicator ............................................................................................ 11-11

    12 SERVICE RELEVANT – Reports12.1 Operational qualification......................................................................................................... 12-112.1.1 Start-up report............................................................................................................................ 12-1

    12.1.2 Water pretreatment settings reports .......................................................................................... 12-6

    12.1.3 Coarse filter and booster pump assembly settings report ......................................................... 12-6

    12.1.4 Sand filter settings report........................................................................................................... 12-8

    12.1.5 Activated carbon filter settings report ...................................................................................... 12-10

    12.1.6 Softener settings report ........................................................................................................... 12-12

    12.2 Maintenance ........................................................................................................................... 12-1412.2.1 Important information on maintenance procedures ................................................................. 12-14

    12.2.2 Accessories and supplies for maintenance procedures .......................................................... 12-15

    12.2.3 Maintenance report.................................................................................................................. 12-15

    12.3 Tools and test equipment ..................................................................................................... 12-1912.3.1 Test equipment ........................................................................................................................ 12-19

    13 Definitions

    13.1 Definitions and terms .............................................................................................................. 13-1

    13.2 Abbreviations........................................................................................................................... 13-1

    13.3 Symbols.................................................................................................................................... 13-1

    14 Appendix

  • 8/18/2019 Manual Servicio Pretratamiento

    9/230

    Chapter 1: Index

    Fresenius Medical Care Service Manual SM-EN 1/03.11 1-1

    1 Index

    A Abbreviations 13-1 Activated carbon filter  10-1

     Activated carbon filter

    malfunctions overview 10-28

     Activated carbon filter settings

    report 12-10

     Addresses 2-9

     Alternating or parallel operation 9-24

     Appendix 11-1, 12-1

    B

    Basic warnings 2-7

    Biological hazards 2-8

    Booster pump 5-1

    Brief description 2-4

    C

    Chemical hazards 2-7

    Coarse filter and booster pumpassembly settings report 12-6

    D

    Definitions and terms 13-1

    Disclaimer of liability 2-6

    Duties of the responsible

    organization 2-5

    E

    Electric hazards 2-7Error messages on pumps 6-5

    F

    Feed water bypass 7-35

    Feed water inlet unit 4-1

    G

    Guarantee / warranty 2-6

    IImportance of the instructions 2-2Important information 2-1

    Improved water quality 7-24

    M

    Maintenance intervals 11-9

    Maintenance report 12-15

    Manual regeneration 9-14

    Manually starting the regeneration 

    9-21Media filters (sand filters / Birm

    filters) 7-1

    N

    Note, significance 2-3

    Notes 2-3

    O

    Operational data acquisition / daily

    report 11-2Operational qualification 12-1

    Operator information 2-5

    Operator responsibility 2-6

    P

    Particle filter (Big Blue) 20" 8-1

    Preface 2-1

    Prefiltration rack with particle filter  4-10

    Prefiltration unit 4-2

    Pressure reducer  4-11

    Pump malfunctions overview 6-6

    Pumps 6-1

    S

    Sand filter malfunctions overview 7-33

    Sand filter selection table 7-25,7-31

    Sand filter settings report 12-8

    Selection table activated carbon

    filter (double) 10-25

    Selection table activated carbon

    filter (Single) 10-25

    Softener  9-1

    Softener capacities 9-48

    Softener malfunctions overview 9-52

    Softener selection table 9-49

    Softener settings in case of high

    brine concentration 9-3

    Softener settings report 12-12Start-up report 12-1

    Symbols 13-1

    System handling instructions

    report – Water pre-treatment 11-5

    T

    Target group 2-5

    Technical documentation 2-6

    Tip, significance 2-3

    W

    Warnings 2-7

    Warnings, significance 2-3

    Water pretreatment / BASIC 3-3

    Water pretreatment / STANDARD 3-2

    Water pretreatment component

    overview 3-2, 3-3

    Water pretreatment settings

    reports 12-6

  • 8/18/2019 Manual Servicio Pretratamiento

    10/230

    Chapter 1: Index

    1-2 Fresenius Medical Care Service Manual SM-EN 1/03.11

  • 8/18/2019 Manual Servicio Pretratamiento

    11/230

    Chapter 2: Important information

    Fresenius Medical Care Service Manual SM-EN 1/03.11 2-1

    2 Important information

    2.1 Preface

    2.1.1 Structure of the Service Manual

    This Service Manual contains a number of individual components or

    overall concepts that are not products of Fresenius Medical Care

    Deutschland GmbH.

    Due to the standardization of documents, the structure for standardized

    Service Manuals was applied to list the contents for service technicians

    in their usual workflow.

    Each component is therefore described in this Service Manual as an

    individual device using the general structure.

    When applying the master structure, empty chapters were omitted if the

    chapter contents were not applicable.

  • 8/18/2019 Manual Servicio Pretratamiento

    12/230

    Chapter 2: Important information

    2-2 Fresenius Medical Care Service Manual SM-EN 1/03.11

    2.2 How to use the Service Manual

    Identification The document can be identified by the following information on the titlepage and on the labels, if any:

     – Edition of the document – Part number of the technical document

    Page identification The page identification 1-3, for example, refers to Chapter 1, page 3.

    Editorial information The editorial information 1/01.05, for example, refers to: 1st edition,January 2005.

    Organization of thechapters

    To facilitate the use of documents from Fresenius Medical Care, the

    organization of the chapters has been standardized in all manuals.

    There may therefore be chapters within this document without any

    content. Chapters without content are identified.

    Illustrations The illustrations used in the documents (e.g. screens, photos, etc.) maydiffer from the original if this does not have any influence on the

    function.

    Importance of theinstructions

    This document is part of the accompanying documents and is an

    essential part of the device. It includes information necessary for use,

    programming and start-up of the device.

    This document must be studied carefully before attempting to operate

    the device.

    Before the responsible organization may begin operating the device,

    the individual responsible for the operation must have been instructed

    by the manufacturer on how to use the device and must be thoroughlyfamiliar with the contents of the document.

    The device may only be operated by individuals certificated to have

    been instructed on the proper operation and handling of the system.

    Changes Changes to the technical document will be released as new editions orsupplements. In general, this manual is subject to change without

    notice.

    Reproduction Reproduction, even in parts, only with written approval.

  • 8/18/2019 Manual Servicio Pretratamiento

    13/230

    Chapter 2: Important information

    Fresenius Medical Care Service Manual SM-EN 1/03.11 2-3

    2.3 Notes

    2.3.1 Significance of the warnings

    2.3.2 Significance of the note

    2.3.3 Significance of the tip

    Warning 

     Advises the operator that failure to observe this information can result

    in personal injury.

    Warning 

     Advises the operator that failure to observe this information can cause

    damage to the equipment.

    Note 

    Information alerting the operator to the fact that failure to follow the

    steps as specified may result in the specific function not being executed

    correctly, not being executed at all, or not producing the desired effect.

    Tip 

    Information providing useful tips for easy handling.

  • 8/18/2019 Manual Servicio Pretratamiento

    14/230

    Chapter 2: Important information

    2-4 Fresenius Medical Care Service Manual SM-EN 1/03.11

    2.4 Brief description

    Design BASIC /STANDARD

    Components The water pretreatment may consist of the following individualcomponents

     – Feed water inlet unit (single components A-F)

     – Booster pump

     – Media filter (sand filter / Birm filter)

     – Particle filter (Big Blue)

     – Softener 

     – Activated carbon filter 

    If necessary, additional modules can be connected upstream ordownstream for improved quality.

    The water pretreatment components must be installed in a dry, dust-

    free room. The responsible organization must ensure that the design

    specifications of the pretreatment components comply with the

    requirements of the other components used. This applies especially to

    the following system components such as reverse osmosis system,

    permeate ring main and dialysis systems.

     As the complete system consists of separate, individual systems, the

    responsible organization is responsible for the complete system.

  • 8/18/2019 Manual Servicio Pretratamiento

    15/230

    Chapter 2: Important information

    Fresenius Medical Care Service Manual SM-EN 1/03.11 2-5

    2.4.1 Operator information

    Target group The water pretreatment may only be installed, operated and used byindividuals with the appropriate training, knowledge and experience.

     Assembly, extensions, adjustments, modifications or repairs may onlybe carried out by the manufacturer or persons authorized by him.

    Duties of the responsibleorganization

    The responsible organization assumes the following responsibilities:

     – Compliance with the national or local installation, operation, use and

    maintenance regulations

     – Compliance with the accident prevention regulations

     – Correct and safe condition of the water pretreatment.

     – Permanent availability of accompanying documents.

    The water pretreatment consists of components designed to improve

    the permeate conductivity and is extended with additional componentsto a complete water treatment system for use in a dialysis unit by the

    responsible organization.

    The water pretreatment must be installed in a dry room not reserved for

    medical use. The staff call should also be connected. The responsible

    organization must ensure that the technical adjustments of water

    pretreatment match the requirements of the other components and that

    they make up the complete system.

    This applies especially to the following system components such as:

     – Rooms with outlets and drains for the installation of the water

    pretreatment

     – Leakage protection systems

     – Sampling site in the soft water supply line

     – Sampling sites in the ring main system

     – Alert systems for malfunctions (alarms)

     – Operational organization

    The components of the water pretreatment must be easily accessible

    from all sides. The responsible organization must ensure that its

    operators have been trained.

    Operators of the water pretreatment, the reverse osmosis system and

    the dialysis systems must have received instructions on how to operatethe system.

    The responsible organization should inform the local water supplier of

    the dialysis operation and insist on early provision of data regarding

    water composition, availability etc.. This measure does not relieve the

    responsible organization of its obligation to regularly check the feed

    water composition.

    Samples for microbial testing must therefore be collected from the

    system and from the individual parts of the system in accordance with

    the applicable regulations.

  • 8/18/2019 Manual Servicio Pretratamiento

    16/230

    Chapter 2: Important information

    2-6 Fresenius Medical Care Service Manual SM-EN 1/03.11

     As the water supply system consists of a series of smaller systems, the

    responsible organization is responsible for the complete water supply

    system.

    Operator responsibility When entering parameters, the following must be observed:

     – The parameters entered must be verified by the operator, i.e. the

    operator must check that the values entered are correct. If the

    verification reveals a deviation between the desired parameters and

    the parameters displayed, the setting must be corrected before

    activating the function.

     – The actual values displayed must be compared with the desired

    values specified.

    Disclaimer of liability Should the responsible organization wish to use other consumablesand accessories than those listed in the accompanying documents, the

    responsibility to ensure the correct function of the device lies exclusively

    with the responsible organization. The applicable legal regulations mustbe complied with (e.g. in Germany the Medical Device Directive, MDD

    and the MPBetreibV = German regulation for the operation of medical

    products).

    The manufacturer does not assume any responsibility or liability for

    personal injury or other damage and excludes any warranty for damage

    to the components of the water pretreatment resulting from the use of

    non-approved or unsuitable consumables or accessories.

    2.4.2 Guarantee / warranty

    Guarantee The extent of the guarantee is stipulated in the respective purchaseorders.

    Warranty The purchaser's warranty rights are governed by the applicable legalregulations.

    Proper function of the water pretreatment components is guaranteed for

    the period specified in the sale contract.

     Any use of the device which is not in accordance with its intended use

    will void any liability and warranty.

    2.4.3 Technical documentation

    On request, descriptions and other documents are made available by

    the manufacturer. These are intended to support trained personnel of

    the responsible organization in servicing and repairing the device.

  • 8/18/2019 Manual Servicio Pretratamiento

    17/230

    Chapter 2: Important information

    Fresenius Medical Care Service Manual SM-EN 1/03.11 2-7

    2.5 Warnings

    2.5.1 Basic warnings

    2.5.2 Electric hazards

    ,

    2.5.3 Chemical hazards

    Warning 

    The design of the water pretreatment components must comply with

    the requirements of the Fresenius Installation Guideline (IGL)

    Warning 

    Danger to life caused by electrical voltage.

    Touching live parts will cause an electric shock.

     – Before opening electrical devices, disconnect the power plug.

     Actuating the ON/OFF switch stops operation of electrical devices,

    but does not disconnect the system from the supply voltage!

    Warning 

    The national standards and regulations must be observed when

    connecting the system to the power supply system.

    Warning 

    Use of protective equipments

    Rubber gloves (acrylonnitril latex, cotton-lined) should be worn to avoid

    contact of chemicals or disinfectant with the skin.

    When working with acidic substances:

    Wear goggles! Observe the safety precautions of the concentratedsubstance used!

    In the event of contact with acid:

    Eye: Immediately flush with gently flowing water for 15 minutes.

    Skin: Use soap under flowing water for neutralization.

    Ingestion: Do not induce vomiting, but have the person affected drink

    plenty of still water. Seek medical advice.

  • 8/18/2019 Manual Servicio Pretratamiento

    18/230

    Chapter 2: Important information

    2-8 Fresenius Medical Care Service Manual SM-EN 1/03.11

    2.5.4 Biological hazards

    Warning 

    The system may only be disinfected after previous consultation with the

    manufacturer of the system or by personnel authorized by the

    manufacturer.

    Warning 

    Risk of infection

    Observe the applicable local laws and regulations concerning handling

    of potentially infectious material.

  • 8/18/2019 Manual Servicio Pretratamiento

    19/230

    Chapter 2: Important information

    Fresenius Medical Care Service Manual SM-EN 1/03.11 2-9

    2.6 Addresses

    Please address any inquiries to:

    Manufacturer  Fresenius Medical Care AG & Co. KGaAD-61346 Bad Homburg

    Germany

    Phone: + 49 (0)6172 609-0

    www.fmc-ag.com

    Internationalservice

    Fresenius Medical Care

    Deutschland GmbH

    Service Support International

    Hafenstraße 9

    D-97424 Schweinfurt

    Germany

    Phone: +49 (0)9721 678-333 (Hotline)Fax: +49 (0)9721 678-130

    Localservice

  • 8/18/2019 Manual Servicio Pretratamiento

    20/230

    Chapter 2: Important information

    2-10 Fresenius Medical Care Service Manual SM-EN 1/03.11

  • 8/18/2019 Manual Servicio Pretratamiento

    21/230

    Chapter 3: Component overview

    Fresenius Medical Care Service Manual SM-EN 1/03.11 3-1

    3 Component overview

    3.1 Components of the water pretreatment system

    The following pages show the two overviews of the BASIC and

    STANDARD water pretreatment system.

  • 8/18/2019 Manual Servicio Pretratamiento

    22/230

    Chapter 3: Component overview

    3-2 Fresenius Medical Care Service Manual SM-EN 1/03.11

    3.1.1 Water pretreatment component overview / STANDARD

       F  e  e   d  w  a   t  e  r   i  n   l  e   t  u  n   i   t

       P  r  e   f   i   l   t  r  a   t   i  o  n  –   R  a  c   k  w   i   t   h  p  a  r   t   i  c   l  e   f   i   l   t  e  r   (  o  p   t   i  o  n  a   l   )

       B   i  r  m    f

       i   l   t  e  r

       B  o  o  s   t  e  r  p  u  m  p

       P  a  r   t   i  c

       l  e   f   i   l   t  e  r   B   i  g   B   l  u  e ,   5  µ  m    (

       2   0   “   )

       S  o   f   t  e  n  e  r

       A

      c   t   i  v  a   t  e   d  c  a  r   b  o  n   f   i   l   t  e  r

       A   S   t  o  p  v  a   l  v  e

       B   W  a   t  e  r  m  e   t  e  r   (  o  p   t   i  o  n  a   l   )

       C   B  a  c   k  w  a  s   h  a   b   l  e  p  r  e   f   i   l   t  e  r   1   0   5  µ  m

       D   S  o   l  e  n  o   i   d  v  a   l  v  e

       E   P   i  p  e   d   i  s  c  o  n  n  e  c   t  o  r

       F   P  r  e  s  s  u  r  e  r  e   d  u  c  e  r

       B   1   4   0

       B   2   0   0

       B   3   0   0

       B   4   0   0

       E   0   8   0

       E   1   5   0

       E   2   2   5  -   T

       E   3   0   0

       E   4   0   0

       A

       0   5   0

       A

       1   0   0

       A

       1   5   0

       A

       3   0   0

       A

       4   0   0

       S  o   f   t  w  a   t  e  r  m  o  n   i   t  o  r   i  n  g  u  n   i   t   /   A  q  u  a   S   O   F   T

       P  a  r   t   i  c   l  e   f   i   l   t  e

      r   B   i  g   B   l  u  e ,   5  µ  m    (

       2   0   “   )

       F  e  e   d  w  a   t  e  r   t  a  n   k

       T  u   b   i  n  g

       M  e   d   i  a   f   i   l   t  e  r

       S  a  n   d   f   i   l   t  e  r

       S   1   4   0

       S   2   0   0

       S   3   0   0

       S   4   0   0

  • 8/18/2019 Manual Servicio Pretratamiento

    23/230

    Chapter 3: Component overview

    Fresenius Medical Care Service Manual SM-EN 1/03.11 3-3

    3.1.2 Water pretreatment component overview / BASIC

       F  e  e   d  w  a   t  e  r   i  n   l  e   t  u  n   i   t

       P  r  e   f   i   l   t  r  a   t   i  o  n  –   R  a  c   k  w   i   t   h  p  a  r   t   i  c   l  e   f   i   l   t  e  r   (  o  p   t   i  o  n  a   l   )

       B  o  o  s   t  e  r  p  u  m  p

       S  o   f   t  e  n  e  r

       A

      c   t   i  v  a   t  e   d  c  a  r   b  o  n   f   i   l   t  e  r

       A   S   t  o  p  v  a   l  v  e

       B   W  a   t  e  r  m  e   t  e  r   (  o  p   t   i  o  n  a   l   )

       C   B  a  c   k  w  a  s   h  a   b   l  e  p  r  e   f   i   l   t  e  r   1   0   5  µ  m

       D   S  o   l  e  n  o   i   d  v  a   l  v  e

       E   P   i  p  e   d   i  s  c  o  n  n  e  c   t  o  r

       F   P  r  e  s  s  u  r  e  r  e   d  u  c  e  r

       E   0   8   0

       E   1   5   0

       E   2   2   5  -   T

       E   3   0   0

       E   4   0   0

       A

       0   5   0

       A

       1   0   0

       A

       1   5   0

       A

       3   0   0

       A

       4   0   0

       S  o   f   t  w  a   t  e  r  m  o  n   i   t  o  r   i  n  g  u  n   i   t   /   A  q  u  a   S   O   F   T

       P  a  r   t   i  c   l  e   f   i   l   t  e

      r   B   i  g   B   l  u  e ,   5  µ  m    (

       2   0   “   )

       F  e  e

       d  w  a   t  e  r   t  a  n   k

       T  u   b   i  n  g

       B   i  r  m    f

       i   l   t  e  r

       B   1   4   0

       B   2   0   0

       B   3   0   0

       B   4   0   0

       M  e   d   i  a   f   i   l   t  e  r

       S  a  n   d   f   i   l   t  e  r

       S   1   4   0

       S   2   0   0

       S   3   0   0

       S   4   0   0

  • 8/18/2019 Manual Servicio Pretratamiento

    24/230

    Chapter 3: Component overview

    3-4 Fresenius Medical Care Service Manual SM-EN 1/03.11

  • 8/18/2019 Manual Servicio Pretratamiento

    25/230

    Chapter 4: Feed water inlet unit

    Fresenius Medical Care Service Manual SM-EN 1/03.11 4-1

    4 Feed water inlet unit

    4.1 Component overview – wall-mounted prefiltration unit

    Part number: Complete assembly – description

    F00002261 BASIC Feed water inlet unit 1 1/4" (Inline operation)

    F00001403 BASIC Feed water inlet unit 1" (Tank operation)

    POS Description

    A Stop valve

    B Backwashable prefilter (105 µm)

    C Solenoid valve

    Note:

     An additional leakage detector (6328611), which switches the solenoid valve, must be implemented.

    D Pipe disconnector (Inline operation)

    B C DA BA C

    TANK OPERATIONINLINE OPERATION

  • 8/18/2019 Manual Servicio Pretratamiento

    26/230

    Chapter 4: Feed water inlet unit

    4-2 Fresenius Medical Care Service Manual SM-EN 1/03.11

    4.1.1 Prefiltration unit 1‘‘, 1 1/4‘‘, 1 1/2‘‘

    Part number: Complete assembly – description

    F00001116 Prefiltration unit complete 1"

    F00001117 Prefiltration unit complete 1 1/4“

    F00001118 Prefiltration unit complete 1 1/2“

    POS Description

    A Stop valve

    B Backwashable prefilter 

    C Solenoid valve (24 V)

    D Pipe disconnector 

    E Particle filter (Big Blue) (5 µm); 20“

    F Drain kit

    G Duplex softener (option)

    H Electronics unit

    I Leakage sensor 

    F

    B

    E

    G

    A

    C D

    H

    I

  • 8/18/2019 Manual Servicio Pretratamiento

    27/230

    Chapter 4: Feed water inlet unit

    Fresenius Medical Care Service Manual SM-EN 1/03.11 4-3

    4.2 Installation

    4.2.1 Installation / start-up requirements

    General information

    4.2.2 Data reporting

    Note 

    Start-up of the system must be performed by a trained technician. The

    wiring must be adapted to fit the respective field of application.

    Note 

    Prior to its start-up, check the system for damage caused during

    transport. If there are signs of damage to the electric components, donot turn the system on.

    Note 

    Reports for the items of the operational qualification listed beloware available as master copies in chapter 12:

     – Start-up

     – Setting

     – Maintenance

  • 8/18/2019 Manual Servicio Pretratamiento

    28/230

    Chapter 4: Feed water inlet unit

    4-4 Fresenius Medical Care Service Manual SM-EN 1/03.11

    4.3 Specifications

    4.3.1 Specifications / data sheet

    Rated output of RO system Tank operation Inline operation Rack operation

    Feed water path

    Type, size

    1 “ 1 1/4 “ 1 1/2 “

    Rated flow 1.5m³/h

    25 L/min

    2.9 m³/h

    48 L/min

    5.8 m³/h

    97 L/min

    Peak performance during softener regenera-

    tion

    2.4 m³/h

    40 L/min

    4 m³/h

    67 L/min

    7.8 m³/h

    130 L/min

    Weight

    Weight without rack 12kg empty;

    ~17 kg; operation

    17kg empty;

    ~21 kg; operation

    22kg empty

    ~27 kg; operation

    Operating weight approx. Weight

    ~42 kg

    Weight

    ~46 kg

    Weight

    ~52 kg

    Operating data

    Kv 2.2 m³/h 2.5 m³/h 6.0 m³/h

    Maximum pressure 8.0 bar 8.0 bar 6.8 bar  

     Ambient temperature 1–45 °C 1–45 °C 1–45 °C

    Water temperature 1-36 °C 1-36 °C 1-36 °C

    Voltage supply 230 V 50 Hz 230 V 50 Hz 230 V 50 Hz

    Floor space

    Prefiltration rack

    (HxWxD in mm)

    1700 x 1200 x 800 1700 x 1400 x 800 1700 x 1500 x 800

    Connections

    Inlet and outlet 1 1/4 “

    External thread

    1 1/4 “

    External thread

    1 1/4 “

    External thread

  • 8/18/2019 Manual Servicio Pretratamiento

    29/230

    Chapter 4: Feed water inlet unit

    Fresenius Medical Care Service Manual SM-EN 1/03.11 4-5

    4.3.2 Operating conditions

    For information regarding the operating conditions refer to the latest

    edition of the Installation Guideline (IGL).

    4.3.3 Configuration options

    Possible configurations:

     – Configuration with AquaSOFT (see chapter 10)

     – Automatic backwash function

     – Timer 

  • 8/18/2019 Manual Servicio Pretratamiento

    30/230

    Chapter 4: Feed water inlet unit

    4-6 Fresenius Medical Care Service Manual SM-EN 1/03.11

    4.4 Maintenance

    Note 

    The information given below is summarized for all waterpretreatment components in Chapter 14 (Appendix).:

     – Important information on maintenance procedures

     – Accessories and supplies

     – Test report - maintenance

     – Overview / maintenance intervals for water pretreatment

    components

     – Explanations for the test report

  • 8/18/2019 Manual Servicio Pretratamiento

    31/230

    Chapter 4: Feed water inlet unit

    Fresenius Medical Care Service Manual SM-EN 1/03.11 4-7

    4.5 Functional description

    4.5.1 Design of components

    4.5.1.1 Water inlet unit

    Standard configuration

    The standard configuration of a feed water inlet path consists of the stop

    valve, a backwashable prefilter, a solenoid valve, a pipe disconnectorand a drain connector.

    4.5.2 Component function

    4.5.2.1 Stop valve

    The function of the stop valve is to stop the water supply for servicing to

    perform maintenance procedures. The stop valve is provided with a

    sampling port for the collection of feed water samples.

    4.5.2.2 Water meter - optional

     An optional water meter can be installed to document the water

    consumption.

    4.5.2.3 Backwashable prefilter 105 µm

    The backwashable prefilter is installed in the feed water path upstream

    of the solenoid valve to prevent contamination and therefore an early

    failure of the solenoid valve.

    Water inlet unit

    Components:

     A) Stop valve

    B) Water meter (optional)

    C) Backwashable prefilter (105 µm)

    D) Solenoid valve

    E) Pipe disconnector 

    F) Pressure reducer (optional)

    F) Drain

  • 8/18/2019 Manual Servicio Pretratamiento

    32/230

    Chapter 4: Feed water inlet unit

    4-8 Fresenius Medical Care Service Manual SM-EN 1/03.11

    In its standard version, the backwashable prefilter has 105 µm filterinserts. As an option this prefilter can also be designed with a 50 µm filter insert.

     As the first filter in the complete water pretreatment system, this insert

    is usually exposed to a high concentration of particles. Regual

    backwashing of the filter during non-dialysis hours must therefore be

    ensured to prevent early pressure drops. This can either be done

    manually (by turning the rotary pin 180° clockwise) or with the automatic

    backwashing feature.

    The automatic backwashing feature is available as an option. It should

    always be used in combination with a timer to be able to define a fixed

    backwash time.

    4.5.2.4 Solenoid valve

    Function The feed water inlet solenoid valve is used to disconnect the watersupply in case of a leakage. In case of rack installation it is activated bya single-station leakage detection or in case of wall-mounted installation

    by a leakage detector.

    Usage as tank fill valve In combination with a feed water tank, the feed water inlet solenoidvalve is also used as a tank fill valve. In this case, the solenoid valve is

    controlled by a float switch which is supplied with the feed water tank.

    Manual opening If necessary, the solenoid valve can be opened manually.

    In case of a power failure push the lever in and turn it by 90°. In this po-sition it is automatically locked and the valve is permanently open.

    Water supply is therefore still ensured even without power supply.

    If the solenoid valve is manually bypassed, the room

    must never be left unattended.Otherwise the valve must be left in its original position.

    Normal condition

    (original position)

    Valve open

    (manually bypassed)

    Note 

    Without power supply, leakage monitoring is not active.

     After dialysis hours, the valve should be closed again by unlocking to

    stop the water supply in case of leakage.

  • 8/18/2019 Manual Servicio Pretratamiento

    33/230

    Chapter 4: Feed water inlet unit

    Fresenius Medical Care Service Manual SM-EN 1/03.11 4-9

    4.5.2.5 Pipe disconnector 

    If the inlet pressure drops, the pipe disconnector ensures a

    disconnection from the system (DIN EN 1717) as requested by the

    Drinking Water Decree if there is a risk of back-contamination.

  • 8/18/2019 Manual Servicio Pretratamiento

    34/230

    Chapter 4: Feed water inlet unit

    4-10 Fresenius Medical Care Service Manual SM-EN 1/03.11

    4.5.3 Prefiltration rack with particle filter (optional)

    4.5.4 Component description

    Stop valve (see chapter 4.5.2.1 , page 4-7)

    Backwashable prefilter  (see chapter 4.5.2.3 , page 4-7)

    Solenoid valve (see chapter 4.5.2.4 , page 4-8)

    Pipe disconnector  (see chapter 4.5.2.5 , page 4-9)

    Particle filter (Big Blue)5 µm

    see chapter 8 (particle filter)

    Drain kit The drain kit is used to direct the draining water to the drain in thebuilding.

    Duplex softener (option) see chapter 9 (Softener)

    Electronics unit The electronics unit supplies the connected components with electricalvoltage.

    Leakage sensor  The leakage sensor detects (inadvertent) water leakage. In this case,the connected valve is closed. An audible alarm and a red LED indicator

    signal a leakage alarm.

    Prefiltration rack with particle filter (optional)

    Components:

     – Stop valve

     – Backwashable prefilter 

     – Solenoid valve (24 V)

     – Pipe disconnector 

     – Particle filter (Big Blue) (5 µm); 20“

     – Drain kit

     – Duplex softener (option)

     – Electronics unit

     – Leakage sensor 

  • 8/18/2019 Manual Servicio Pretratamiento

    35/230

    Chapter 4: Feed water inlet unit

    Fresenius Medical Care Service Manual SM-EN 1/03.11 4-11

    4.6 Options

    4.6.1 Prefiltration rack with particle filter (option)

    4.6.2 Pressure reducer - optional

     An optional pressure reducer can be installed in the feed water path. Its

    function is to avoid pressure peaks and thus to limit the system load. A

    pressure reducer is recommended if the static pressure at the user

    points exceeds 5 bar .

    The installation of a pressure reducer is recommended in case of

    variations in pressure caused be pressure increases / domestic water

    supply system. As an option, a domestic water station can be used

    instead of a backwashable prefilter. It includes the backwashable

    prefilter and a pressure reducer unit.

    4.6.3 Configuration options (feed water tank)

    If a feed water tank is installed, it will have the requested system

    disconnection. In this case the installation of a pipe disconnector is not

    required.

  • 8/18/2019 Manual Servicio Pretratamiento

    36/230

    Chapter 4: Feed water inlet unit

    4-12 Fresenius Medical Care Service Manual SM-EN 1/03.11

  • 8/18/2019 Manual Servicio Pretratamiento

    37/230

    Chapter 5: Booster pump

    Fresenius Medical Care Service Manual SM-EN 1/03.11 5-1

    5 Booster pump

    5.1 Booster pump assembly CRN5-8 component overview

    Part number: Complete assembly – description

    F00001935 Booster pump assembly (vertical pumps CRN5-8)

    POS Description

    A Connection to water supply(Possible on both sides. The connection which is not used must be provided with a cap nut)

    B1 Pump 1 (vertical pump CRN5-8)

    B2 Pump 2 (vertical pump CRN5-8)

    C Control unit

    D Pressurizer vessel

    W2

    A

    B1

    D

    C

  • 8/18/2019 Manual Servicio Pretratamiento

    38/230

    Chapter 5: Booster pump

    5-2 Fresenius Medical Care Service Manual SM-EN 1/03.11

    5.2 Installation

    5.2.1 Installation / start-up requirements

    General information

    5.2.2 Important start-up information

    5.2.2.1 Booster pump assembly connection

    The suction line can be connected either on the right or on the outer

    left of the booster pump assembly. In case of tank operation it is

    recommended to connect two suction tubes in parallel to avoid

    pressure drops in the suction line.

    The pressure port is located in the center and can be connected

    using a tubing or a pipe for a fixed connection. A high-pressure

    tubing is included in the accessories package of the system.

     After connection to the water supply, both pumps must be ventilated

    by opening the vent screws.

    Note 

    Start-up of the system must be performed by a trained technician. The

    wiring must be adapted to fit the respective field of application.

    Note 

    Prior to its start-up, check the system for damage caused during

    transport. If there are signs of damage to the electric components, donot turn the system on.

    Booster pump assembly (CRN5-8)

    The pumps must be ventilated prior to start-up of the switching unit.

  • 8/18/2019 Manual Servicio Pretratamiento

    39/230

    Chapter 5: Booster pump

    Fresenius Medical Care Service Manual SM-EN 1/03.11 5-3

    5.2.2.2 Connection to the domestic water supply network

    Positioning the pressure switch

    The pressure switch installed at the water inlet ensures protection

    against running dry. Upon delivery, the pressure switch is set to 1 bar.

    It is recommended to retain this setting. If the pressure in the water inlet

    drops below 1 bar, the booster pump assembly will be stopped. If this

    dynamic minimum pressure is not reached, it is recommended to use a

    feed water tank.

    Pressure switch installation position

    Note 

    If a pipe separator or a pressure reducer is installed on the outlet side

    of the booster pump, the pressure sensor must always be installed after

    these components. For the installation, remove the screw-type adapter

    in the outlet block and integrate it after the pipe disconnector.

    State upon delivery: Dismounting the adapter 

    State upon delivery: When the system is supplied, the pressuresensor is installed with an adapter at the pump outlet, this adapter

    must be removed and must be installed after the pressure reducer

    and the pipe separator.

    The pressure sensor is installed on the outlet side after the pressure

    reducer and the pipe separator.

    Pressure switch installation position and connection (run-dry protection)

    Pressure switch installation position (run-dry protection)

  • 8/18/2019 Manual Servicio Pretratamiento

    40/230

    Chapter 5: Booster pump

    5-4 Fresenius Medical Care Service Manual SM-EN 1/03.11

    Booster pump assembly with pump switching

    Pressure switch installation position and connection

    Pressure switch installation position (after pipe separator and

    pressure reducer)

    Pressure switch installation position and connection (run-dry protection)

    Electrical connection of the pressure switch for the run-dry protection

    Pressure switch installation position and connection

    Electrical connection of the pressure sensor 

  • 8/18/2019 Manual Servicio Pretratamiento

    41/230

    Chapter 5: Booster pump

    Fresenius Medical Care Service Manual SM-EN 1/03.11 5-5

    Verification / setting of the pressure switch

    Connection of the pressure switch

    Pressure switch connection

    Connection 1: white

    Connection 2: brown

    Pressure sensor connection

    Pressure sensor connection Connection 1: brown

    Connection 2: white

    Reference point on the scale

    Locate the reference point.

  • 8/18/2019 Manual Servicio Pretratamiento

    42/230

    Chapter 5: Booster pump

    5-6 Fresenius Medical Care Service Manual SM-EN 1/03.11

    Rotate the selector fully counter-clockwise.

    The selector must first be rotated fully counter-clockwise.

    Adjust the selector to the reference point.

    Rotate the selector approx. ¼ of a turn clockwise until the figure "1"

    is above the reference point.

    Adjust the selector to the reference point.

     Adjust the scale reference point.

    The pressure switch is now set to 1 bar. This means the operatingpressure of the pump is 1 bar.

    Due to the pressure switch's hysteresis, the switch-off pressure is

    slightly below 1 bar (approx. 0.5 bar).

  • 8/18/2019 Manual Servicio Pretratamiento

    43/230

    Chapter 5: Booster pump

    Fresenius Medical Care Service Manual SM-EN 1/03.11 5-7

    5.2.2.3 Connection to a feed water tank

    Note 

    In case of tank operation, the tank must be provided with a float switchwhich will open a contact if the tank is empty. This contact is connected

    in the pump inlet instead of the pressure switch.

    Tank connection

    The inlet of the booster pump assembly is directly connected to the

    outlet of the tank using the pump connection set no. 6319211.

    The suction line can be connected either on the right or on the outer

    left of the booster pump assembly.

    In case of tank operation it is recommended to connect two suction

    tubes in parallel to avoid pressure drops in the suction line. The pressure port is located in the center and can be connected

    using a tubing or a pipe for a fixed connection. A high-pressure tubing

    with a length of x cm is included in the accessories package of the

    system.

    The pressure sensor is connected directly to the pump outlet.

    The inlet pressure switch on the water inlet side of the pump is not

    used during tank operation. In its place the float switch (see

    illustration) must be wired in. This must be done during installation.

    Pressure switch installation position and connection (run-dry protection)

     A float switch which is installed in the feed water tank ensures run-

    dry protection.

    The brown and white conductor must be connected.

  • 8/18/2019 Manual Servicio Pretratamiento

    44/230

    Chapter 5: Booster pump

    5-8 Fresenius Medical Care Service Manual SM-EN 1/03.11

    Pressure switch installation position and connection

    Installation position of the pressure sensor on the pump outlet (state

    upon delivery)

    Electrical connection of the pressure sensor in the switching cabinet.

  • 8/18/2019 Manual Servicio Pretratamiento

    45/230

    Chapter 5: Booster pump

    Fresenius Medical Care Service Manual SM-EN 1/03.11 5-9

    5.2.2.4 Power supply

     A 400 V; 3 N ~; 50 Hz; 16 A three-phase current connection is required

    to supply the pump switching unit with voltage.

    Connection of a leakage detector 

    The floating contact of a leakage detection system can be connected to

    the booster pressure assembly. The contact must be closed if there is

    no leakage and must open if a leakage occurs.

    It is recommended to use the insert of the leakage detecting system. It

    can be connected directly to the pump control (see illustration )

    5.2.2.5 Overview of electrical connections

    Connection option for a leakage detection system (jumper must be removed)

    Remove the jumper and connect the line to the floating contacts of

    the leakage detector.

    Note 

    When the system is supplied, positions 1 to 6 are wired.

    Positions 7 and 8 must be adjusted by the technician depending on

    the respective situation (feed water tank or direct operation).

  • 8/18/2019 Manual Servicio Pretratamiento

    46/230

    Chapter 5: Booster pump

    5-10 Fresenius Medical Care Service Manual SM-EN 1/03.11

    POS Connections

    1 Pump 1

    2 Pump 2

    3 Power cable

    4 Temperature switch 1

    5 Temperature switch 2

    6 Pressure sensor 

    7 Connection for run-dry protection (pressure switch or float switch)

    8 Connection option for the floating contact (e.g. leakage detector)

    1 2 3 4 5 6

    7

    8

  • 8/18/2019 Manual Servicio Pretratamiento

    47/230

    Chapter 5: Booster pump

    Fresenius Medical Care Service Manual SM-EN 1/03.11 5-11

    5.2.2.6 Settings on the booster pump assembly (CRN5-8)

    Should the booster pump assembly not operate as desired when it is

    supplied, the setting of the pressure level and of the delay times can be

    checked and changed.

    Pressure levels

    Pressuretype

    Description

    Pmin Operating pressure for base load and peak load pump (0-10 bar):

    Pmin is set to the pressure at which the pump is to start.

    Pmax1 Cutoff pressure level for the peak load pump (difference to Pmin), (0-4 bar):

    If this pressure is achieved at parallel operation, the peak load pump is turned off. Pmax1 should

    be set to a value between Pmin and Pmax2.

    Pmax2 Cutoff pressure level for the base load pump (difference to Pmax1), (0-2 bar):

    If this pressure is reached during operation of one pump, this pump is switched off. Pmax2

     

    should be set to approx. 0.5 bar below the pump's nominal pressure.

    Caution:If a pressure reducer was installed downstream of the booster pump assembly, the pressure

    reducer should be set to 6 bar minimum as the pressure on the pressure sensor can otherwise

    never reach the cutoff pressure.

    Note 

    The settings of max1 and Pmax

    2 are difference values which add to the

    lower pressure set.

  • 8/18/2019 Manual Servicio Pretratamiento

    48/230

    Chapter 5: Booster pump

    5-12 Fresenius Medical Care Service Manual SM-EN 1/03.11

    Calculation example

    Setting Pmin: 3 bar 

    Setting Pmax1: 2 bar 

    Setting Pmax2: 1 bar 

    This means Operating pressure level (Pmin) = 3 bar 

    Cutoff pressure level for peak load pump (Pmax1) = 5 bar 

    Cutoff pressure level for base load pump (Pmax 

    2) = 6 bar 

    Delay times

    T  = Over-run (0–2 min): 

    Minimum operation time of the pump to prevent permanent start and

    stop of the pump. Default setting: 2 minutes. When set to "0 minutes",the minimum operation time is approx. 30 seconds.

    T  = run-dry delay (0–2 min):

    time after which die pumps are turned off in case the pump runs dry.

    Default setting: 0 minutes.

    When set to "0 minutes", the delay time is approx. 10 seconds.

    Motor protection

    Potentiometers P1 and P2 are set to the nominal current of the motor.

    Motor protection setting (nominal current of

    pump 1 and pump 2)

    Setting Pmax2: Cutoff pressure level for base

    load pump (difference to Pmax1)

    Setting Pmax1: Cutoff pressure for peak loadpump (difference to Pmin)

    Setting Pmin: Operating pressure for baseload and peak load pump

    Minimum operation time setting

    Water alarm delay time setting

  • 8/18/2019 Manual Servicio Pretratamiento

    49/230

    Chapter 5: Booster pump

    Fresenius Medical Care Service Manual SM-EN 1/03.11 5-13

    5.2.3 Data reporting

    Note 

    Documentation of the settings in the Settings report

     As part of the Operational Qualification, the settings must be

    documented in the Settings reports upon start-up.

    Note 

    Reports for the items of the operational qualification listed beloware available as master copies in chapter 12:

     – Start-up

     – Setting

     – Maintenance

  • 8/18/2019 Manual Servicio Pretratamiento

    50/230

    Chapter 5: Booster pump

    5-14 Fresenius Medical Care Service Manual SM-EN 1/03.11

    5.3 Specifications

    5.3.1 Specifications / data sheet

    5.3.1.1 Performance characteristics

    Parameter Function / specifications

    Usage Pressure increase

    Component Booster pump assembly

    Type CRN5-8

    Connection 1 1/4", external thread

    Sealing material Shaft seal carbon silicide, o-ring EPDM

    Installation position Vertical

    Operating current 2.5 kW / kVA

    Power supply 380–400 V; 3N~ 50 Hz; 16A

    Operating pressure, adjustable 1–6 bar (preset to x bar)

    Switch-off pressure 1–6 bar (preset to x bar)

    Pressure, minimum 2.2 bar (in case of tank operation)

    Pressure, maximum 6.5 bar 

    Water temperature, minimum 1 °C

    Water temperature, maximum 40 °C

    Flow rate 5.7 m³/h at 3.8 bar (at a nominal pumping head of 37.8m)

    Ambient temperature, max. 40°C

    Parameter Function / specifications

    Electrical connection 380–400 V 3N~ 50 Hz

    Power consumption 2.5 kW / kVA

    Operating pressure  Adjustable 1–6 bar (preset to x bar)

    Cutoff pressure  Adjustable 1–6 bar (preset to x bar)

    Dimensions (width/ height/depth) Width / height / depth 101 cm / 92 cm / 45 cm

  • 8/18/2019 Manual Servicio Pretratamiento

    51/230

    Chapter 5: Booster pump

    Fresenius Medical Care Service Manual SM-EN 1/03.11 5-15

    5.3.1.2 Volumetric capacity - diagram

    5.3.2 Operating conditions

    For information regarding the operating conditions refer to the latest

    edition of the Installation Guideline (IGL).

    (one pump active) (two pumps active)

    Delivery volume

       P  u  m  p   i  n  g   h  e  a   d

  • 8/18/2019 Manual Servicio Pretratamiento

    52/230

    Chapter 5: Booster pump

    5-16 Fresenius Medical Care Service Manual SM-EN 1/03.11

    5.4 Maintenance

    Note 

    The information given below is summarized for all waterpretreatment components in Chapter 14 (Appendix):

     – Important information on maintenance procedures

     – Accessories and supplies

     – Test report – maintenance

     – Overview / maintenance intervals for water pretreatment

    components

     – Explanations for the test report

  • 8/18/2019 Manual Servicio Pretratamiento

    53/230

    Chapter 5: Booster pump

    Fresenius Medical Care Service Manual SM-EN 1/03.11 5-17

    5.5 Error messages

    5.5.1 Malfunctions, causes and correction

    Setting the system to the Standby mode

    Components:

     A) Operating mode selector switch

    B) Operation / malfunction indicator (pump 1 and pump 2)

    C) Run-dry / leakage indicator 

    D) Main power switch

    Selection Description

    Green LED flashing. The motor protection has operated. No automatic confirmation aftercorrection of the problem.

    Possible causes:

     – Motor current too high.

     – The motor temperature is too high, possible defective pressure

    sensor causing permanent operation and overheating of the pump.

    Confirmation:

    Set the operating mode selector switch to "OFF" and then back to

    "ON" again. If necessary, allow the motor to cool down first.

    Red LED illuminated System was turned off due to lack of water or leakage. Automaticconfirmation after correction of the problem.

    B

    D

    C

    A

  • 8/18/2019 Manual Servicio Pretratamiento

    54/230

    Chapter 5: Booster pump

    5-18 Fresenius Medical Care Service Manual SM-EN 1/03.11

    5.6 Functional description

    5.6.1 Design of components

    5.6.2 Component description

    The booster pump assembly (CRN5-8) is an assembly which is ready

    for connection in accordance with the requirements for the minimum

    flow pressure with normal centrifugal suction pumps.

    The booster pump assembly in conjunction with the two vertical pumpsprovides increased operational safety of a water treatment system from

    Fresenius Medical Care.

    The booster pump assembly is used for the alternating control of two

    booster pumps. Automatic switching ensures uniform distribution of

    stress on both connected pumps and prevents damage to the pump

    caused by prolonged idle times when it is not in use. To increase the

    efficiency, the second pump will be turned on additionally in peak

    situations.

    The booster pump assembly is intended for the pressure increase in the

    feed water section of the water pretreatment system. It can be used

    both for increasing the pressure downstream of a feed water tank andfor direct connection to the feed water line.

    5.6.2.1 Control unit

    The electronic control of the booster pump assembly controls and

    monitors the two connected pumps.

    Booster pump assembly (CRN5-8)

    Components:

     A) Feed water connection

    B1) Pump 1 (vertical pump CRN5-8)

    B2) Pump 2 (vertical pump CRN5-8)

    C) Control unit

    D) Pressurizer vessel

    B2

    D

    C

    A

    B1

  • 8/18/2019 Manual Servicio Pretratamiento

    55/230

    Chapter 5: Booster pump

    Fresenius Medical Care Service Manual SM-EN 1/03.11 5-19

    5.6.2.2 Base load and peak load pump

    The control differentiates between a base load and peak load pump. If

    water requirements are low, only the base load pump will deliver, if

    water requirements rise the peak load pump will also be switched on.

    This concept has the advantage that the pumps are always operated intheir optimum performance range. This contributes to a high efficiency

    of the complete system. The base load and the peak load pump will be

    activated alternating after each restart. This ensures uniform wear of

    both pumps.

    5.6.2.3 Alternating pressure situations

    If the system is not in operation and the pressure in the system drops

    below the start-up pressure level (Pmin), the base load pump will start.

    If the pressure drops again to the start-up pressure (Pmin) caused by

    an increasing water demand, the peak load pump will also be started.If, with decreasing water demand, the pressure in the system rises

    above the pressure level (Pmax 1), the peak load pump will be stopped.

    If water consumption stops, the pressure will rise above the pressure

    level (Pmax 2) and the basic load pump will be stopped.

    5.6.2.4 Run-dry protection

    The adjustable pressure switch at the booster pump assembly inlet

    protects the system against running dry and ensures a minimum

    pressure of 1 bar at the booster pump inlet.

    If the pressure at the inlet drops below 1 bar, first the peak load pumpwill stop. If the pressure at the inlet remains below 1 bar  after this pump

    stop, the base load pump will also be stopped.

    5.6.2.5 Tank operation

    When operated with a tank, a float switch will be connected instead of

    the pressure to ensure protection against running dry.

    5.6.2.6 Automatic turnoff 

    The water temperature is also monitored; if the temperature exceeds

    60 °C, e.g. by a defect of the control unit and a constantly running pump

    without water consumption, the respective pump will be stopped.

  • 8/18/2019 Manual Servicio Pretratamiento

    56/230

    Chapter 5: Booster pump

    5-20 Fresenius Medical Care Service Manual SM-EN 1/03.11

    5.6.3 Functional description

    5.6.3.1 Setting the system to the Standby mode

    5.6.3.2 Selector switch functions

    Setting the system to the Standby mode

    Components:

     A) Operating mode selector switch

    B) Operation / malfunction indicator (pump 1 and pump 2)

    C) Run-dry/leakage indicator 

    D) Main power switch

     After turning on the main power switch and after selection of

    automatic 1 + 2 (position as in illustration), the operating modeswitch is used to set the system to the Standby mode.

    Functions of the selector switch

    Components:

     A) Test mode pump 2

    B) Off 

     A) Test mode pump 1

    D) Automatic mode pump 1

    E) Automatic mode pumps 1 + 2

    D) Automatic mode pump 2

    Selection Description

    Automatic mode pump 1 Only pump 1 runs in the automatic mode (e.g. if pump 2 is defective).

    Automatic mode pumps 1+2 Both pumps run in the automatic mode. Depending on the consumptionvolume the pumps are started and stopped (default setting). The green

    LED of the active pump is constantly illuminated to indicate operation.

    Automatic mode pump 2 Only pump 2 runs in the automatic mode (e.g. if pump 1 is defective).

    B

    D

    C

    A

    B

    ED

    A

    F

    C

  • 8/18/2019 Manual Servicio Pretratamiento

    57/230

    Chapter 5: Booster pump

    Fresenius Medical Care Service Manual SM-EN 1/03.11 5-21

    Test mode pump 1 Pump 1 runs irrespective of the prevailing pressures and safetyfunctions. The test mode will be automatically stopped after approx. 1.5

    minutes.

    Test mode pump 2 Pump 2 runs irrespective of the prevailing pressures and safetyfunctions. The test mode will be automatically stopped after approx. 1.5

    minutes.

    Off  Both pumps are off.

    Selection Description

  • 8/18/2019 Manual Servicio Pretratamiento

    58/230

    Chapter 5: Booster pump

    5-22 Fresenius Medical Care Service Manual SM-EN 1/03.11

  • 8/18/2019 Manual Servicio Pretratamiento