    Panoramic X-Ray Detector XXH-3005

    USER MANUAL

    HUATEC GROUP

  • 8/11/2019 Manual of X Ray Flaw Detector XXH3005

    2/14

    HUATEC GROUP

    - 2 -

    Catalog

    1. Summary 3

    2.Technical targets 4

    3.Configuration 4

    3.1 X-ray generator 4

    3.2 control unit 5

    4.Installation 7

    5.Using 8

    5.1Start 8

    5.1.1 System checking itself 8

    5.2 Exposure operation 9

    5.2.1 Exposure parameter setting 9

    5.2.2 Start exposure 10

    5.2.3 Running period 10

    5.3 Alarm protection 12

    5.4 Turn off the unit 13

    6.Notes 13

  • 8/11/2019 Manual of X Ray Flaw Detector XXH3005

    3/14

    HUATEC GROUP

    - 3 -

    1. Summary

    Welcome for using Panoramic X-ray detector XXH-3005, this type of unit is

    the most ideal equipment for Non-destructive testing in the fields such as

    petroleum, aviation, mechanical manufacturing, boiler, pressure vessels,

    ship building, national defence industries and so on.We manufacture X-Ray

    flaw detector according to the international standard of ISO3999.

    The control unit of Panoramic X-ray detector XXH-3005 combines the

    advantages of controllers both in china and foreign countries, it applies high

    ability micro-processor and effective professional integrated circuit

    components, with the advanced controlling technology, It not only can adaptto different instability environment, but also have the high reliability.

    Panoramic X-ray detector XXH-3005 has new characters as following:

    Checking the software and hardware itself after start the machine

    Excellent anti-jamming performance, which can adapt for field operation

    without power supply, because of the motor supports

    Can adjust exposure parameters with keys ( easy operation.)

    Has the function of Delay time opening HV. Convenient for operator keep

    away from dangerous field

    Can adjust high voltage slowly for protecting X-ray generator

    Automatically controlling according to 1:1 working and rest operating way,

    the longest continuous exposure period is 5 minutes.

    Including complete protecting system: over KV, over KV, low mA, over

    mA, over temperature and optical-sound alarming

    Convenient for repairing and maintenance

    Much more Substance for structure designing, much more convenient for

    local operation

    Time can be changed for0.3 according to the power supply net to make

    the exposure more precise

  • 8/11/2019 Manual of X Ray Flaw Detector XXH3005

    4/14

    HUATEC GROUP

    - 4 -

    We have introduced the detail operation information of Panoramic X-ray

    detector XXH-3005 in the manual, For better understanding various

    functions of the system, please read it carefully before using.

    2. Technical targets

    Panoramic X-ray detector (with glass tube)

    3.Configuration

    This detector is made up of X-ray generator, control unit, power supply

    cable, connection cable and accessories.

    (note): users must not open the X-ray generator and control unit

    3.1 X-ray generator

    A set of X-ray generator combines X-ray tube, HT transformer (include X-

    ray tube filament winding) and insulation gas (SF6) into a aluminum barrel.

    One side of X-ray generator installed fan and radiator which is for tubes

    cooling. for high voltage , SF6 gas is good insulation medium.

    Components for X-ray generator including:

    A. X-ray tube

    B. HV transformer (include filament winding)

    C. Temperature relay

    Output

    voltage

    (KV)

    Input

    frequency

    (KW)

    Radiation target

    Max penetration

    for A3 steel

    Weight(kg) Dimension(mm)

    flat cone flat cone generator controller generator controller

    XXH-

    3005

    170-300 3.0

    25+

    5

    30+5 47 40 42 11 335335835 350290160

  • 8/11/2019 Manual of X Ray Flaw Detector XXH3005

    5/14

    HUATEC GROUP

    - 5 -

    D. Gas pressure

    E. cable Connect socket

    F. alarming lamp socket

    G. X-ray tube cooling fan

    In this system the X-ray tube applies with complete electrical defeat

    protection, also with the anode of X-ray tube grounded connection to stable

    the single pulse voltage, there sets a temperature relay inside of X-ray

    generator, once it appears over temperature, the high voltage will be cut off

    from the control unit automatically to make sure the safety of the whole

    system. When the alarming lamp lights, it stands for that X-ray detector is

    producing X-ray.

    The end collar on two side of X ray generator can be laid vertically or flat, italso can be used for moving when is working in the high altitude.

    3.2 control unit

    The main function of controller is to control X ray generator working on a

    regulated voltage value automatically to make sure produces stable X ray,

    and also controlling the exposure time automatically.

    The operate panel of controller can refer to diagram 1.

    Components on control panel including:

    a) preparing lamp: when the lamp lights, it means machine can do

    exposure this time

    b) voltage indicator: when the lamp lights, X-ray generator is exposure

    now

    c) current indicator: when the lamp lights, means X ray generator works

    well

    d) 2 digital LED displayer: used for displaying exposure time or running

    status

    e) time adjusting: used for setting the exposure time

    f) voltage setting: used for setting exposure voltage value for X ray

    generator

  • 8/11/2019 Manual of X Ray Flaw Detector XXH3005

    6/14

    HUATEC GROUP

    - 6 -

    g) X ray on: used for X ray generator exposure starting

    h) X ray off: used for stop X ray generator exposure or delay KV start

    i) exposure curve: adjust proper exposure parameters according to the

    inspected workpiece.

    voltage indicator current indicator 2 digital LED displayer

    preparing

    lamp

    time adjusting time adjusting

    voltage setting voltage setting

    exposure curve X ray on X ray off

    Pic. 1General drawing of front panel of control unit

    Here has a socket box in the right side of controller (refer to diagram 2)

    Power cable connector Ground wire plug 2A fuse

    Connection

    Power cable Connector

    switch cable connector

    Pic. 2Side panel of control unit

  • 8/11/2019 Manual of X Ray Flaw Detector XXH3005

    7/14

    HUATEC GROUP

    - 7 -

    4.Installation

    1. After open the package box, check if there has any damages or miss

    assembling according to packing list.

    2. Confirm if the standard of controller is same as X ray generator, or it

    will lead for the damage of X ray generator

    3. Cut off the controllers power supply, connect power supply cable

    well, at the same time, the power supply should satisfied with the

    requirements in Technical targets

    4. Connect one side of ground wire on the controller, the other side on

    the ground

    5. Connect the control unit to X ray generator with connected cable

    6. Connect the alarming lamp to X ray generator

    7. It should have X ray protect system when is using. If using in the wild

    field, we can use one 2mm lead board for protection, or we can make sure

    nobody in the circle of 20 meters semi-diameter which of X ray generator

    as focal

    8. After finish the above steps ,we should also check if the gas pressure

    of X ray generator satisfied with the requirements in Technical targets, if

    not, please mustnt open X rayon, avoid for the damage of X ray generator

  • 8/11/2019 Manual of X Ray Flaw Detector XXH3005

    8/14

    HUATEC GROUP

    - 8 -

    5. Using

    5.1 system starting

    After finish the installation, please confirm if the cables on controller is

    connected correctly, and switch the side power supply on, at this time, the

    2 digital LED displayer on operate panel lights.

    5.1.1system self-checking

    The controller will do complete tests itself after opening. At the same time

    every section of 2 digital LED displayer continuously rotate lights (if anysection not lights may have some damages)

    When system self-checking, the following status stands for normal:

    Preparing lamp, voltage indicator, current indicator lamp extinguish

    X ray on X ray off keys have no use

    following status stands for abnormal:

    2 digital LED display some codes

    Buzzer sounds continuously

    5.1.1.1 If there has any failure, corresponding codes will shows on LED

    displayer

    F1: tests 89C51 RAM failure

    F2: tests 89C51 EPROM failure

    F3: tests X ray on key failure

    F4: tests X ray off key failure

    When find out the failure, users should write down the codes carefully, turn

    off the power supply and contact with the manufacturer for repairing.

    5.1.1.2 If the system works normally, it will display information as following

    after about 30 seconds:

    lights preparing lamp

  • 8/11/2019 Manual of X Ray Flaw Detector XXH3005

    9/14

    HUATEC GROUP

    - 9 -

    buzzer sounds for once or twice

    LED displayer shows xx (x indicates for the value between 0 and 5.5),

    at this time, operator can set detector for exposure

    time can be changed for 0.3 according to net power supply fluctuation.

    5.2 exposure operation

    After finish the above processes, system has prepared well for exposure,

    at this time, preparing lamp lights, system goes into exposure parameter

    setting period.

    5.2.1 exposure parameters setting

    Before exposure, we should set the exposure parameters, such as the

    voltage value or exposure time.

    Please set the exposure parameter according to the following steps:

    5.2.1.1 confirm the material and thickness of inspected workpiece

    5.2.1.2 according to the density & thickness of inspected workpiece

    calculate the thickness of A3 steel

    5.2.1.3 check the A3 steel thickness from exposure curve table to confirm

    the voltage value and exposure time

    (notes): because of the differences of X ray generator, for a separate

    equipment, the parameter on the curve table is not sure the best values for

    exposure, operators should applies the standard modules to make their

    best exposure curve according to original equipment curve table

    (1) modification of voltage value

    revolve the KV setting keys clockwise or anticlockwise to corresponding

    the indicating line with the exposure voltage value. If revolve clockwise, the

    KV value will be increased, or it will be decreased.

  • 8/11/2019 Manual of X Ray Flaw Detector XXH3005

    10/14

    HUATEC GROUP

    - 10 -

    The setted KV value should be corresponding to the used equipment, the

    range of output KV value of different type controller & X ray generator as

    following:

    1005 type: 20-100kv

    1605 type: 40-160kv

    2005 type: 100-200kv

    2505 type: 150-250kv

    3005 type: 170-300kv

    (2) modification of exposure time setting value

    exposure time can be setted by time choosing hand rotary and 2digital

    LED displayer.

    rotary the time choosing key clockwise or anticlockwise, at the sametime2 digital LED displaying value increase or decrease correspondingly

    left side of 2digital LED stands for minutes, right side for one tenth

    of minute, every value is multiple times of 6 seconds. For example, 4.5

    stands for that the setted exposure time is 4 minutes and 30 seconds.

    5.2.2 start exposure

    After finish setting voltage value and exposure time, press X ray on key to

    start exposure

    5.2.3 running period

    Exposure period has three processes: delay starting, exposure period and

    rest period.

    5.2.3.1 delay starting

    After press X ray on key, convenient for operator go away from scene to

    avoid radiation of X ray, before X ray on, system will delay stating for 1

    minute automatically.

  • 8/11/2019 Manual of X Ray Flaw Detector XXH3005

    11/14

    HUATEC GROUP

    - 11 -

    During the delayed period, AC and rest time timing take turns on

    displayer, the value will be decreased by 6 seconds a time, at the same

    time, you can also heard the alarm of didi. Sounds. Before it

    decreased until 0. users can stop the system by pressing X ray off, then

    the system will back to exposure setting interface; you can also press X

    ray on to finish the delay stating period.

    5.2.3.2 exposure period

    When the delaying time decreased until 0 or operator press X ray on to

    finish the delaying period, the whole system will goes into exposure period

    automatically.

    At first, the system will slowly increase the tube voltage and current to the

    regulate value in the advanced control way. After preparing lamp on

    control panel extinguished, KV lamp lights, it stands for that X ray

    generator has been connected, after a while, mA lamp lights, it means

    the system is on normal exposure status.

    On displayer it shows the rest exposure time, when the rest exposure time

    decreased until 0, the whole system will be cut off and have a rest

    automatically.

    If press X ray off during exposure period, the whole system will stop the

    operation and cut off the high voltage of X ray generator into rest status,

    after the rest time equals to the tubes working time, it will back to exposure

    parameter setting status.

    When there has any failure during the exposure period, controller will cut

    off the high voltage of X ray generator immediately. At the same time

    2digital LED displays the failure codes, buzzer rings continuously until

    press X ray off

  • 8/11/2019 Manual of X Ray Flaw Detector XXH3005

    12/14

    HUATEC GROUP

    - 12 -

    5.2.3.3 rest period

    After finish the exposure period ( or stop by pressing X ray off ), system

    will be cut off automatically and goes into rest period, the rest time should

    equals to the tubes working time. The rest period is forced. During this

    period, any key has no use, the whole lamp all extinguishes.

    On displayer it shows the rest relax time, when the time decreases until 0.

    the preparing lamp lights, buzzer rings a hinting sound, displayer re-

    shows the exposure time, then the system re-back to exposure parameter

    setting status.

    5.3 alarming protection

    After the X ray generator connected with high voltage. The whole system

    real time monitors KV, mA, tube head temperature values of X ray

    generator all the time. If there has any abnormal phenomenon such as

    over KV, over mA, low mA or over temperature of the tube head, the

    controller will cut off the high voltage of X ray generator , at the same time,

    2digital LED displays failure codes, buzzer rings continuously, different

    codes has different meaning:

    UU: controller controls over KV failure

    UA: system controls over mA failure

    LA: system controls low mA failure

    PT: X ray generator over temperature failure

    AA: controlling over mA of main loop failure

    If press X ray off under the alarming status, the system will stop currently

    operation, after into rest period, the rest time should equals to working

    time, then back to exposure parameters setting status.

  • 8/11/2019 Manual of X Ray Flaw Detector XXH3005

    13/14

    HUATEC GROUP

    - 13 -

    (notes): when it appears failure during exposure, users should check if

    here has any incorrectness of cable connection, if power supply voltage

    works normally or contact with supplier for repairing.

    5.4 turn off the unit

    To make sure the system work stable, users just can cut off the controlling

    power supply under the parameter setting status, it mustnt stop X ray unit

    when X ray generator connecting with high voltage or during the rest

    period.

    Please reuse the unit after turn it off at least 10 seconds, or it will destroy

    the inside components of the unit.

    6. notices

    1. make sure the standard of control unit conform to the standard of

    tube head

    2. make sure the cables connects well, the cable sockets clean

    3. make sure the controller connected ground well

    4. check the pressure of X ray generator with pressure table, it mustnt

    be used below 0.35 Mpa

    5. when operates access to tube head (for example, stick the film and

    so on) should turn off the high voltage first.

    6. make sure not turn off the machine, when tube head connected with

    high voltage (during the exposure operation, include stop period) or during

    the rest time of tube head

    7. its better to aging for one time before using everyday, age the unit to

  • 8/11/2019 Manual of X Ray Flaw Detector XXH3005

    14/14

    HUATEC GROUP

    - 14 -

    a voltage value, which you want to use during the exposure time, it will

    benefit for the normal working of X ray tube and longer its using life.

    8. when adding the length of connected cable, the sectional area of

    every pin lead should fit for standards.

    9. if not using the unit, it should be better to put it in a dry environmental

    and also get it away from heat source, or harm gas to avoid the sunshine.