Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002...

74
Operator Manual 748223-K June 2002 http://www.processanalytic.com Model 7003M Percent Oxygen Analyzer

Transcript of Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002...

Page 1: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-KJune 2002

http://www.processanalytic.com

Model 7003MPercent Oxygen Analyzer

Page 2: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Emerson Process ManagementRosemount Analytical Inc.Process Analytic Division1201 N. Main St.Orrville, OH 44667-0901T (330) 682-9010F (330) 684-4434e-mail: [email protected]://www.processanalytic.com

ESSENTIAL INSTRUCTIONSREAD THIS PAGE BEFORE PROCEEDING!

Rosemount Analytical designs, manufactures and tests its products to meet many national andinternational standards. Because these instruments are sophisticated technical products, youMUST properly install, use, and maintain them to ensure they continue to operate within theirnormal specifications. The following instructions MUST be adhered to and integrated into yoursafety program when installing, using, and maintaining Rosemount Analytical products. Failure tofollow the proper instructions may cause any one of the following situations to occur: Loss of life;personal injury; property damage; damage to this instrument; and warranty invalidation.

• Read all instructions prior to installing, operating, and servicing the product.

• If you do not understand any of the instructions, contact your Rosemount Analytical rep-resentative for clarification.

• Follow all warnings, cautions, and instructions marked on and supplied with the product.

• Inform and educate your personnel in the proper installation, operation, and maintenance ofthe product.

• Install your equipment as specified in the Installation Instructions of the appropriate Instruc-tion Manual and per applicable local and national codes. Connect all products to the properelectrical and pressure sources.

• To ensure proper performance, use qualified personnel to install, operate, update, program,and maintain the product.

• When replacement parts are required, ensure that qualified people use replacement partsspecified by Rosemount. Unauthorized parts and procedures can affect the product’s per-formance, place the safe operation of your process at risk, and VOID YOUR WARRANTY.Look-alike substitutions may result in fire, electrical hazards, or improper operation.

• Ensure that all equipment doors are closed and protective covers are in place, except whenmaintenance is being performed by qualified persons, to prevent electrical shock and per-sonal injury.

The information contained in this document is subject to change without notice.

Ryton® is a registered trademark of Phillips Petroleum Co.

Page 3: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-K

June 2002

Rosemount Analytical Inc. A Division of Emerson Process Management Contents i

Model 7003M

TABLE OF CONTENTS

PREFACE...........................................................................................................................................P-1Definitions ...........................................................................................................................................P-1Safety Summary .................................................................................................................................P-2Documentation....................................................................................................................................P-5Compliances .......................................................................................................................................P-5

1-0 DESCRIPTION AND SPECIFICATIONS..............................................................................1-11-1 Overview................................................................................................................................1-11-2 Principle of Measurement......................................................................................................1-11-3 Sensors..................................................................................................................................1-11-4 Conditions Affecting Measurement .......................................................................................1-2

a. Gas Composition.............................................................................................................1-2b. Sample and Cal Gas Pressures......................................................................................1-2c. Environmental Factors ....................................................................................................1-2

1-5 Features.................................................................................................................................1-31-6 Specifications ........................................................................................................................1-4

a. Performance....................................................................................................................1-4b. Physical...........................................................................................................................1-4c. Electrical..........................................................................................................................1-5d. Sensors ...........................................................................................................................1-5

2-0 INSTALLATION ....................................................................................................................2-12-1 Unpacking..............................................................................................................................2-12-2 Storage ..................................................................................................................................2-12-3 Analyzer Installation ..............................................................................................................2-12-4 Sensor Installation .................................................................................................................2-1

a. Analyzer/Sensor Jumper Configuration ..........................................................................2-12-5 Electrical Connections ...........................................................................................................2-3

a. Sensor Cable ..................................................................................................................2-4b. Power Connections.........................................................................................................2-4c. Current Output Connections ...........................................................................................2-6d. Voltage Output Connections ...........................................................................................2-6

2-6 Relay Contacts for Alarms or ON/OFF-Controls ...................................................................2-8

3-0 OPERATION .........................................................................................................................3-13-1 Overview................................................................................................................................3-13-2 Keypad Operation..................................................................................................................3-1

a. Single and Double Keystroke..........................................................................................3-3b. Display Prompts ..............................................................................................................3-4c. Error Messages...............................................................................................................3-6d. Automatic Return to RUN Mode .....................................................................................3-7e. Preventing Unauthorized Access....................................................................................3-7f. Sensor Equilibrium..........................................................................................................3-7

3-3 Start-Up .................................................................................................................................3-8a. Initial Power-Up...............................................................................................................3-8b. System Restart................................................................................................................3-9c. Automatic Restart - Due to Power Failure ......................................................................3-9d. Manual Restart................................................................................................................3-9

Page 4: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-KJune 2002

ii Contents Rosemount Analytical Inc. A Division of Emerson Process Management

Model 7003M

3-4 Run Mode ..............................................................................................................................3-10a. Display Oxygen Concentration/Sensor Temperature .....................................................3-10b. Acknowledge Alarms ......................................................................................................3-10c. Access Other Modes.......................................................................................................3-10

3-5 Setup Mode ...........................................................................................................................3-11a. Security Code..................................................................................................................3-13b. Current Output Range - 0-20mA or 4-20mA................................................................3-13c. Range Endpoint Adjustments .........................................................................................3-13d. Zero Offset Adjustment ...................................................................................................3-15

3-6 Alarm Mode vs. ON/OFF-Control Mode ................................................................................3-16a. Relay Configured as Alarm .............................................................................................3-16b. Relay Configured as ON/OFF-Controller ........................................................................3-16c. Relay Contacts................................................................................................................3-18d. Fail Safe Operation .........................................................................................................3-18e. Relay Reset Upon Fatal Error.........................................................................................3-18f. Pressure Effects on Alarms ............................................................................................3-18g. Setting Alarm and ON/OFF-Controller Relays................................................................3-18h. Setting Alarm Setpoints ..................................................................................................3-18i. Setting Alarm ON and OFF Levels .................................................................................3-19

4-0 CALIBRATION AND ADJUSTMENTS.................................................................................4-14-1 Frequency of Calibration .......................................................................................................4-14-2 Calibration mode ...................................................................................................................4-1

a. Zero Calibration...............................................................................................................4-1b. Span Calibration..............................................................................................................4-1c. Pressure Compensation - Correction for Constant Pressure Differences ......................4-4

4-3 Hold Mode .............................................................................................................................4-54-4 Diagnostics Mode ..................................................................................................................4-8

a. Raw Sensor Voltage .......................................................................................................4-8b. Test Current Output ........................................................................................................4-8c. Linearizing Current Output..............................................................................................4-8d. Test Alarm Relays...........................................................................................................4-10

4-5 Adjustments And Settings .....................................................................................................4-10

5-0 THEORY................................................................................................................................5-15-1 Principles Of Operation .........................................................................................................5-1

a. Electrochemical Theory ..................................................................................................5-15-2 Effects of Sample Conditions ................................................................................................5-1

a. Barometric Pressure .......................................................................................................5-1b. Humidity ..........................................................................................................................5-1c. Sample Temperature ......................................................................................................5-1d. Interfering Gases.............................................................................................................5-2

5-3 Circuit Descriptions................................................................................................................5-2a. Signal Board....................................................................................................................5-2b. Power Supply Board .......................................................................................................5-2c. Microprocessor Board.....................................................................................................5-2d. Alarm Circuits..................................................................................................................5-2

6-0 ROUTINE SERVICING AND TROUBLESHOOTING...........................................................6-16-1 Routine Servicing ..................................................................................................................6-16-2 Troubleshooting.....................................................................................................................6-1

a. Symptoms .......................................................................................................................6-1

Page 5: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-K

June 2002

Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii

Model 7003M

7-0 REPLACEMENT PARTS......................................................................................................7-17-1 Circuit Board Replacement Policy.........................................................................................7-17-2 Matrix .....................................................................................................................................7-17-3 Replacement Parts - Model 7003M.......................................................................................7-27-4 Replacement Parts - Sensors................................................................................................7-3

a. Rechargeable Sensors ...................................................................................................7-3b. Disposable Sensors ........................................................................................................7-4

8-0 RETURN OF MATERIAL......................................................................................................8-18-1 Return Of Material .................................................................................................................8-18-2 Customer Service ..................................................................................................................8-18-3 Training..................................................................................................................................8-1

LIST OF ILLUSTRATIONS

Figure 2-1. Sensor Ordering Matrix.......................................................................................... 2-2Figure 2-2. Model 7003M Rear Panel Connections................................................................. 2-3Figure 2-3. Power Jumpers and Fuses .................................................................................... 2-5Figure 2-4. Current Output Connections .................................................................................. 2-7Figure 3-1. Model 7003M Front Panel Controls and Indicators ............................................... 3-1Figure 3-2. Model 7003M Operation Flow Diagram................................................................. 3-2Figure 3-3. SETUP Mode Flowchart ...................................................................................... 3-12Figure 3-4. Security Access Routine Flowchart ..................................................................... 3-14Figure 3-5. ALARM Mode Flowchart...................................................................................... 3-17Figure 3-6. Action of Alarm Relay ON and OFF Setpoints .................................................... 3-19Figure 4-1. CALIBRATION Mode and PRESSURE COMPENSATION Flowchart.................. 4-3Figure 4-2. HOLD and DIAGNOSTICS Modes Flowchart ....................................................... 4-7

Page 6: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-KJune 2002

iv Contents Rosemount Analytical Inc. A Division of Emerson Process Management

Model 7003M

LIST OF TABLES

Table 2-1. Sensor Cables ....................................................................................................... 2-4Table 2-2. Analyzer Cable Connections ................................................................................. 2-4Table 3-1. Key Functions ........................................................................................................ 3-3Table 3-2. Display Indicators .................................................................................................. 3-3Table 3-3. Security Access Prompts and Responses............................................................. 3-4Table 3-4. Alarm Settings Prompts and Responses............................................................... 3-4Table 3-5. Calibration Prompts and Responses ..................................................................... 3-4Table 3-6. System Setup Prompts and Responses................................................................ 3-4Table 3-7. System Hold and Diagnostics Codes .................................................................... 3-5Table 3-8. Data Displayed in DIAGNOSTICS Mode............................................................... 3-5Table 3-9. Miscellaneous Prompts and Responses ............................................................... 3-5Table 3-10. Power ON, Reset Diagnostics Errors (Fatal Errors ............................................... 3-6Table 3-11. RUN Mode Errors (Non-Fatal Errors ..................................................................... 3-6Table 3-12. Signal Board Errors ............................................................................................... 3-6Table 3-13. Microprocessor Errors That Reset The System .................................................... 3-6Table 3-14. Default Settings at Power-Up ................................................................................ 3-8Table 3-15. RUN Mode Key Functions ................................................................................... 3-10Table 3-16. SETUP Mode Key Functions .............................................................................. 3-11Table 3-17. ALARM Mode Key Functions.............................................................................. 3-16Table 3-18. Alarm Relay Contact Positions/Terminal Locations............................................. 3-18Table 4-1. CALIBRATION Mode Key Functions........................................................................1Table 4-2. Composition of Dry Atmospheric Air...................................................................... 4-2Table 4-3. Composition of Humid Atmospheric Air................................................................. 4-2Table 4-4. Barometric Pressure Effect on Apparent Oxygen Concentration ..............................4Table 4-5. PRESSURE COMPENSATION Mode Key Functions..............................................4Table 4-6. HOLD Mode Key Functions ......................................................................................5Table 4-7. Caret Readings...................................................................................................... 4-6Table 4-8. DIAGNOSTICS Mode Key Functions ................................................................... 4-8

LIST OF DRAWINGS

652542 Outline and Mounting – Panel Mount652543 Outline and Mounting – Wall Mount652544 Outline and Mounting – Pipe Mount652682 Pictorial Wiring Diagram652683 Interconnect Diagram652699 Installation Drawing

Page 7: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-K

June 2002

Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-1

Model 7003M

PREFACE

The purpose of this manual is to provide information concerning the components,functions, installation and maintenance of the Model 7003M Percent Oxygen Analyzer

Some sections may describe equipment not used in your configuration. The user shouldbecome thoroughly familiar with the operation of this module before operating it. Readthis instruction manual completely.

DEFINITIONS

The following definitions apply to DANGERS, WARNINGS, CAUTIONS and NOTES found throughoutthis publication.

DANGER .

Highlights the presence of a hazard which will cause severe personal injury, death, or substantialproperty damage if the warning is ignored.

WARNING .

Highlights an operation or maintenance procedure, practice, condition, statement, etc. If notstrictly observed, could result in injury, death, or long-term health hazards of personnel.

CAUTION.

Highlights an operation or maintenance procedure, practice, condition, statement, etc. If notstrictly observed, could result in damage to or destruction of equipment, or loss of effectiveness.

NOTEHighlights an essential operating procedure,condition or statement.

Page 8: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-KJune 2002

P-2 Preface Rosemount Analytical Inc. A Division of Emerson Process Management

Model 7003M

SAFETY SUMMARY

If this equipment is used in a manner not specified in these instructions, protective systems may beimpaired.

AUTHORIZED PERSONNEL

To avoid explosion, loss of life, personal injury and damage to this equipment and on-site property,all personnel authorized to install, operate and service the this equipment should be thoroughlyfamiliar with and strictly follow the instructions in this manual. SAVE THESE INSTRUCTIONS.

DANGER.

ELECTRICAL SHOCK HAZARDDo not operate without doors and covers secure. Servicing requires access to live parts which cancause death or serious injury. Refer servicing to qualified personnel.

For safety and proper performance this instrument must be connected to a properly groundedthree-wire source of power.

Relay contacts wired to separate power sources must be disconnected before servicing.

This instrument was shipped from the factory set up to operate on either 115 VAC 50/60 Hz, or 230VAC 50/60 Hz, as specified by sales order. Verification that this voltage is proper for the installationis the responsibility of the purchaser. See Section 2-5b on page 2-4.

WARNING .

OXYGEN LEVEL WILL NOT BE MONITOREDThe instrument will not return automatically to the RUN mode from HOLD or DIAGNOSTICS modes.Operator must manually return instrument to RUN mode to continue monitoring oxygen.

WARNING.

ERRONEOUS OXYGEN READINGSIf the pressure compensation function is not applied correctly, erroneous readings can be dis-played. The display using span gas during calibration should be the same as the reading using thesame gas in RUN mode if the pressure values are correctly entered.

Page 9: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-K

June 2002

Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-3

Model 7003M

WARNING.

PARTS INTEGRITYTampering or unauthorized substitution of components may adversely affect safety of this product.Use only factory documented components for repair

WARNING.

SENSOR NOT INTRINSICALLY SAFEThe oxygen sensors used with this analyzer are not intrinsically safe. Use of the sensors in or nearflammable liquids, gases or vapors or in Class I, Division 1, Hazardous Locations should be care-fully evaluated by qualified personnel at the site and is entirely the responsibility of the user.

The sensor is approved as non-incendive for use in Class I, Division 2 locations.

CAUTION .

HIGH PRESSURE GAS CYLINDERSThis module requires periodic use of pressurized gas. See General Precautions for Handling andStoring High Pressure Gas Cylinders, page P-4

Page 10: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-KJune 2002

P-4 Preface Rosemount Analytical Inc. A Division of Emerson Process Management

Model 7003M

GENERAL PRECAUTIONS FOR HANDLING AND STORING HIGHPRESSURE GAS CYLINDERS

Edited from selected paragraphs of the Compressed Gas Association's "Handbook of CompressedGases" published in 1981

Compressed Gas Association1235 Jefferson Davis HighwayArlington, Virginia 22202

Used by Permission

1. Never drop cylinders or permit them to strike each other violently.

2. Cylinders may be stored in the open, but in such cases, should be protected against extremes of weatherand, to prevent rusting, from the dampness of the ground. Cylinders should be stored in the shade when lo-cated in areas where extreme temperatures are prevalent.

3. The valve protection cap should be left on each cylinder until it has been secured against a wall or bench, orplaced in a cylinder stand, and is ready to be used.

4. Avoid dragging, rolling, or sliding cylinders, even for a short distance; they should be moved by using a suit-able hand-truck.

5. Never tamper with safety devices in valves or cylinders.

6. Do not store full and empty cylinders together. Serious suckback can occur when an empty cylinder is at-tached to a pressurized system.

7. No part of cylinder should be subjected to a temperature higher than 125°F (52°C). A flame should never bepermitted to come in contact with any part of a compressed gas cylinder.

8. Do not place cylinders where they may become part of an electric circuit. When electric arc welding, precau-tions must be taken to prevent striking an arc against the cylinder.

Page 11: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-K

June 2002

Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-5

Model 7003M

DOCUMENTATION

The following Model 7003M instruction materials are available. Contact Customer Service Center or thelocal representative to order.

748223 Operator Manual (this document)

COMPLIANCES

The Model 7003M Percent Oxygen Analyzer and Models 4000 and 5000 series sensors are suitable foruse in Class I, Groups A, B, C, and D, Division 2 locations per North American electrical codes. FactoryMutual Research (FM) approved under J.I. 1T5AO.AX. Canadian Standards Association (CSA) certifiedunder file LR 93812.

APPROVED

FM ®

Page 12: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-KJune 2002

P-6 Preface Rosemount Analytical Inc. A Division of Emerson Process Management

Model 7003M

Page 13: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-K

June 2002

Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-1

Model 7003M

SECTION 1DESCRIPTION AND SPECIFICATIONS

1-1 OVERVIEW

The Model 7003M Percent Oxygen Analyzercontinuously monitors the concentration ofoxygen in a gaseous sample. It consists of asensor and the analyzer. The sensor ishoused in a chamber assembly and isconnected to the analyzer by a shielded multi-conductor cable.

The Model 7003M Percent Oxygen Analyzerprovides direct digital readout, isolated current

output, automatic temperature compensationcircuitry, and dual relay contact closures foralarms or ON/OFF control devices asstandard features. The alarm ranges can beselected independently from the output range,eliminating the need to readjust alarmsetpoints if the range is changed.

The analyzer has six modes of operation:

MODE FUNCTIONRUN (see Section 3-4 on page 3-10) Normal monitoringSETUP (see Section 3-5 on page 3-11) Changing parametersALARM/CONTROL (see Section 3-6 on page 3-16) Set alarms or ON/OFF controllersCALIBRATION (see Section 4-2 on page 4-1) Gas calibration/pressure inputsHOLD (see Section 4-3 on page 4-5) Retain last output valueDIAGNOSTICS (see Section 4-4 on page 4-8) Electronics checks

The Model 7003M Percent Oxygen Analyzeris housed in a NEMA 4X (CSA enclosure 4),corrosion resistant housing suitable for panel,pipe or wall mounting.

The analyzer has a microprocessor controlledkeyboard for operation. Signals from theremote oxygen and temperature sensors areprocessed to compute real time oxygenconcentration corrected for temperature,constant pressure differences and residualsensor current. The oxygen concentration iscompared with user-set alarm parameters,and also generates an isolated 4 to 20 or 0 to20mA signal. Data is stored in a memorycircuit that retains the data in case power isremoved.

The analyzer is shipped from the factoryconfigured for operation on either 115 VAC or230 VAC, 50/60 Hz as specified by thecustomer when ordered.

1-2 PRINCIPLE OF MEASUREMENT

Measurement is based on the electro-chemical reaction between oxygen and thesensor cathode which produces a low levelelectrical current. The sensor signal isproportional to the partial pressure of oxygenpresent at the membrane. This signal and athermistor signal are received by the analyzerand processed by the electronics to output acalibrated oxygen signal that is compensatedfor temperature.

1-3 SENSORS

Rosemount Analytical offers rechargeable anddisposable oxygen sensors which can beused with the Model 7003M. These sensorsare supplied in kits: Submersion, in-line flow,fast response, and a cell separation kit.Sensors are available constructed ofpolypropylene or Ryton. See Section 7-4REPLACEMENT PARTS - SENSORS onpage 7-3, for additional information.

Page 14: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-KJune 2002

1-2 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management

Model 7003M

1-4 CONDITIONS AFFECTING MEASUREMENT

Oxygen measurement is affected by thefollowing:

• Composition of the sampled gas

• Sample and calibration gas pressures

• Environment of sensor, cable andanalyzer

Each must be considered when calibratingand operating the analyzer to obtain accuratemeasurements.

a. Gas Composition

Any substance in the sample thatsignificantly interferes or "poisons" thesensor oxygen reading should beremoved by a chemical and/ormechanical sample handling system (seeSection 5-2 on page 5-1). If ambient air isnot used as the span gas, thecomposition of calibration gases shouldalso be representative of the sample.

The sample and calibration gases mustmeet certain specifications (seeSpecifications in Preface) to ensureaccurate measurement. Physicalparameters that affect measurementinclude:

• Pressure

• Humidity

• Temperature

• Gas composition

Also refer to Section 5-2 on page 5-1 foradditional information about how gasconditions affect the oxygenmeasurement.

b. Sample and Cal Gas Pressures

The sample and calibration gases may beat different pressures as long as thepressure of each remains constant. Anyvariation in the pressure of either gas willresult in an incorrect output reading of thesame proportion.

Two methods of sample calibration arerecommended:

1. Calibrate at the sample pressure, or

2. Calibrated at a different pressure, andenter the two pressure in the analyzer(Section 4-2c on page 4-4)

Refer to Section 5-2 on page 5-1 for amore detailed discussion of therelationship between gas pressures andapparent oxygen concentration.

c. Environmental Factors

If radio frequency interference (RFI)normally occurs in the operatingenvironment, the sensor cable should beelectrically shielded during installation.

Page 15: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-K

June 2002

Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-3

Model 7003M

1-5 FEATURES

Suppressed RangesWhen setting the Model 7003M to asuppressed output range (15 to 25%, etc.),note that the absolute accuracy, drift, andnoise of the oxygen measurement will remainthe same for all ranges (see Specifications inPreface).

ALARM ModeALARM Mode is not self-resetting. Operatormust press ACK to bring relays back to theirno-alarm state. Used when immediateoperator action is required.

NOTE:Operator Can Bypass Alarm Relays

When relays are set to ALARM Mode (Sec-tion 3-6a on page 3-16) and an alarm con-dition has occurred, pressing the ACK keywill turn off the relay switch, whether theoxygen level is out of range or not. If youdo not want an operator to be able to by-pass the relay switches in this way, set theinstrument to ON/OFF -CONTROL Mode.

ON/OFF-Controller ModeON/OFF-CONTROLLER Mode is self-resetting. Pressing ACK has no effect onrelay coils. Used when automated action isrequired.

Page 16: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-KJune 2002

1-4 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management

Model 7003M

1-6 SPECIFICATIONS

a. Performance

Accuracy........................................ ±0.25% oxygenOperating Range ........................... 0 to 25% oxygen (endpoints set by user)Linearity ......................................... ±0.25% oxygen

(with constant sample temperature after sensor zero offset correc-tion)

Zero Stability.................................. ±0.25% oxygen/week at 25°CSpan Stability................................. ±0.25% oxygen/24 hours at 25°CNoise ............................................. ±0.25% oxygenResponse Time ............................. 90% of reading in 20 seconds at 25°CTemperature Effects

32 to 110°F (0 to 44°C) ......... ±6% reading60 to 90°F (15 to 32°C) ......... ±3% readingother 30°F (16°C) ranges ...... +4% reading

b. Physical

MountingStandard ................................ Panel MountOptional ................................. Wall Mount, Pipe Mount

Dimensions.................................... 5.7 x 5.7 x 7.6 inches (14 x 14 x 19 cm) HxWxDWeight ........................................... Approximately 3.0 pounds (1.36 kg)Enclosure....................................... ABS Plastic, Black, NEMA Type 4X, IP65Ambient Humidity .......................... 95% relative humidity, non-condensing (maximum)Ambient Temperature.................... 14 to 122°F (-10 to 50°C)Sensor Cable................................. 2 twisted pairs, specified length up to 1000 feet (305 m) maximumPower Cable .................................. 3 conductor, Type SJT, 18 AWG minimumConnection Conduit

Power..................................... 1/2 inchSensor ................................... 3/4 inch

Page 17: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-K

June 2002

Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-5

Model 7003M

c. Electrical

Power Requirements ..................... 115/230 VAC ±10%, 50/60 Hz ±3 HzDisplay........................................... 3-1/2 digit liquid crystal (LCD)Signal Output................................. Isolated 4 to 20 or 0 to 20 mA (selectable)

600 ohms maximum loadAlarm Relay Contacts.................... Two Form C Relays

SPDT, resistive load, independently adjustableDeadband adjustable over full rangeAlarm Relay Contacts RatingsGeneral Purpose (Ordinary) Locations:115 VAC, 3.0 A (resistive), 230 VAC, 1.5 A resistiveClass I, Division 2 Locations:CSA: 120 VAC, 3.0 A (non-inductive)240 VAC, 1.5 A (non-inductive).FM: 28 VDC (resistive) (Ci = 0, Li = 0)Groups A&B: 150 mAGroup C: 400 mAGroup D: 540 mA

Recorder Output ............................ Isolated 4 to 20 mA

d. Sensors

Types............................................. Rechargeable and DisposableProcess Connections .................... Refer to flow chamber (if supplied)Sample Pressure ........................... 0 to 50 psig (0 to 345 kPa)Sample Temperature..................... 32 to 110°F (0 to 44°C)Ambient Temperature.................... 32 to 110°F (0 to 44°C)Ambient Humidity .......................... 95% relative humidity, non-condensing (maximum)

Also refer to instructions supplied with sensor.

Page 18: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-KJune 2002

1-6 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management

Model 7003M

Page 19: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-K

June 2002

Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-1

Model 7003M

SECTION 2INSTALLATION

2-1 UNPACKING

Carefully examine the shipping carton andcontents for signs of damage. Immediatelynotify the shipping carrier if the carton orcontents is damaged. Retain the carton andpacking material until all componentsassociated with the Model 7003M PercentOxygen Analyzer are operational.

2-2 STORAGE

The sensors and analyzer unit may beshipped together or separately. The sensorsare shipped assembly and charged, ready foruse.

If rechargeable sensors are stored longer thansix months, they may need to be recharged orrejuvenated if performance is marginal.Disposable sensors are not rechargeable.

Refer to documentation supplied with sensorsfor more information.

2-3 ANALYZER INSTALLATION

The Model 7003M Percent Oxygen Analyzeris designed to meet NEMA Type 4X (CSAenclosure 4) requirements, provided that thecutouts in the bottom of the enclosure arefitted with approved conduit fittings or areproperly sealed by the user with Hoffmanplugs (PN A-S050) or equivalent.

The Model 7003M is supplied with the PanelMounting Kit PN 652527 as standard.Optional Wall Mount Kit PN 652539 and PipeMount Kit PN 652529 are available. Refer tothe Outline and Mounting drawings providedin the rear of this manual.

2-4 SENSOR INSTALLATION

NOTE:Refer to Figure 2-1 Sensor Ordering Matrixon page 2-2, and verify that the sensor tobe used corresponds with the application.If an incorrect sensor is used, calibrationmay be impossible or incorrect. If a differ-ent type of sensor is to be used, the rockerswitch (SW1) on the Signal Board mayhave to be changed (see Section 2-4abelow).

Install the sensor in an area wheretemperature is relatively constant, taking careto avoid changes in exposure to sunlight ordue to sudden drafts. A room-temperatureenvironment is preferable. The flowchamber/sensor kit contains detailedmounting and sample flow information. Adddistilled or deionized water to the sensor tothe maximum level indication on the sensorreservoir. Let sensor stand for approximately15 minutes and check for leaks around thebase of the reservoir, and at the seams andcorners. If a leak is found, contact the factorybefore proceeding. Drain the sensor.

a. Analyzer/Sensor Jumper Configuration

The Model 7003M is configured to matchthe type of sensor ordered with theanalyzer. If the Signal Board or Sensor isrepaired or replaced, verify rocker switchSW1 on the Signal Board is properly set.

Rechargeable Sensor: Closepositions 1, 4, and 5

Disposable Sensor: Close positions2, 4, and 6

Page 20: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-KJune 2002

2-2 Installation Rosemount Analytical Inc. A Division of Emerson Process Management

Model 7003M

Figure 2-1. Sensor Ordering Matrix

4 0 0 0 3

6

4000 Rechargeable Sensor5000 Disposable Sensor

1 Polypropylene2 Ryton (rechargeable only)9 Special (consult factory)

1 Submersion2 In-Line Flow3 Fast Response (rechargeable only)9 Special (consult factory)

Position →

Positions 1 through 4 = Sensor TypePosition 5 = Material of Construction (Sensor and Flow Assembly)Position 6 = Flow Assembly Configuration

Example 400013 is a Rechargeable sensor, constructed of polypropylene, mounted in aFast-Response Flow Chamber Assembly.

1 2 3 4 5

1

Page 21: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-K

June 2002

Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-3

Model 7003M

FUSE RATING:115V: 1/4A 250V230V: 1/8A 250V

1 +12V2 -12V1 ANODE2 TEMP SENSE3 NC4 NC5 EPOL6 GND7 Tx8 Rx9 RET1 COM2 NC3 NO4 COM5 NC6 NO7 ISO GND8 ISO CUR-9 ISO CUR+

ALARM CONTACT RATING:3A 120 VAC1.5A 240 VAC3A 30 VDC

AL 1

AL 2

AC POWER

HOT NEUT GND L1 L2

2-5 ELECTRICAL CONNECTIONS

Observe all precautions stated in this manualand in the drawings provided at the back ofthis manual before installation.

All electrical connections to the analyzer aremade on the Power Supply Board connectionpanel (see Figure 2-2 below and drawing652683), which is accessed by removing therear cover.

WARNING

ELECTRICAL SHOCK HAZARDDo not operate without rear cover secured.Servicing requires access to live partswhich can cause death or serious injury.Refer servicing to qualified personnel.

For safety and proper performance this in-strument must be connected to a properlygrounded three-wire source of power.

Electrical installation must be made in ac-cordance with the National Electrical Code(ANSI/NFPA-70) and/or any applicable na-tional or local codes.

WARNING

ENCLOSURE INTEGRITYUnused cable conduit entries must be se-curely sealed by non-flammable closuresto provide enclosure integrity in compli-ance with personnel safety and environ-mental protection requirements. Forwatertight installation conforming to therequirements of NEMA Type 4X, useHoffman plugs (PN A-S050) or equivalentto seal any unused openings.

Provide three separate 1/2 inch conduits asfollows:

• Power cable

• Sensor cable

• Alarm cables

Figure 2-2. Model 7003M Rear Panel Connections

Page 22: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-KJune 2002

2-4 Installation Rosemount Analytical Inc. A Division of Emerson Process Management

Model 7003M

a. Sensor Cable

The sensor cable used with the Model7003M may be of any length up to amaximum of 1000 feet (305 m). (SeeTable 2-1 below.)

The signal cable should be routed througha conduit (customer-supplied) inpermanent installations or in cases wherethe sensor is located more than a few feetfrom the control unit. (See Table 2-2below.)

NOTE:In humid environments, temperaturechanges can cause moisture to con-dense in the conduit, resulting in sig-nal noise and corrosion. To minimizethe effect of condensation, put desic-cant (available locally) inside the con-duit before sealing it.

CABLE PARTNUMBER

LENGTH

191748 10 Feet (3.1 m)193265-SPECL

Any specified lengthup to 1000 Feet (305m)

Table 2-1. Sensor Cables

If conduit used:1/2 inch NPT female connection onbottom of control unit. (Conduit cus-tomer-supplied.

If no conduit used:Use cable connector assembly (PN856831) to provide watertight seal. As-sembly includes: Aluminum inner seal-ing grommet for connector compressionnut.

Table 2-2. Analyzer CableConnections

Connect Sensor Cable1. Ensure that power is disconnected

and remove rear cover.

2. Install the conduit or cable clamp inthe left opening in the bottom rear ofthe case.

3. Install the sensor cable in theinstrument through the cable clampand tighten the nut on the plug tocreate a seal around the cable.

4. Refer to Figure 2-2 on page 2-3.Using a flat-blade screwdriver,connect the wires of the sensor cableto the TB2-1, TB2-2, TB2-5 and TB2-6 on the rear panel of the analyzer.

5. Replace the rear cover if no otherconnections are to be made. Do notovertighten the cover retaining screwsto avoid stripping the threads in theplastic case.

b. Power Connections

Electrical power is supplied to theanalyzer via a customer-supplied threeconductor cable, type SJT, minimum 18AWG. Refer to the installation drawing inthe rear of this manual.

Jumpers and FusesTo verify the jumper location for thespecified power (115 or 230 VAC) or if theanalyzer is to be connected to a differentpower source, refer to Figure 2-3 on page2-5and set jumpers and fuse as follows:

WARNING

ELECTRICAL SHOCK HAZARDDisconnect power to analyzer before con-tinuing.

Page 23: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-K

June 2002

Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-5

Model 7003M

F1 E3 E4 E2 E1

1/4 AMP 115 VAC

F1 E3 E4 E2 E1

1/8 AMP 230 VAC

Power Supply Board

115 VAC Jumpers: E1 - E2 E3 - E4 Fuse: 1/4 Amp

115 VAC Jumpers: E2 - E4 Fuse: 1/8 Amp

1. Remove the rear cover from theanalyzer.

2. Remove Power Supply Board.

3. Verify and/or set Jumpers E1, E2, E3,and E4 per Figure 2-3 below.

4. Verify that fuse (F1) is correct forpower.

5. Re-install Power Supply Board.

6. Replace the rear cover if no otherconnections are to be made. Do notovertighten the cover retaining screwsto avoid stripping the threads in theplastic case.

Conduit ConnectionsRoute the power cable through conduitand into the central opening in the bottomof the case. The inlet hole is sized forstandard 1/2 inch conduit fittings.

WARNING

ENCLOSURE NOT GROUNDEDThe non-metallic enclosure does not pro-vide grounding between conduit connec-tions. Use grounding-type bushing andjumper wires.

Connect Power CableConnect power wiring to the compressiontype connectors on the rear panel (Figure2-2 on page 2-3) as follows:

1. Loosen the connector screw.

2. Strip insulation back approximately1/4 inch (6.35 mm), twist strandedwires together. DO NOT tin braidwire or the connection can workloose.

3. Insert wire into connector about 1/4inch (6.25 mm), ensuring that wiresdo not touch any other wire orconnector.

4. Tighten screw.

Figure 2-3. Power Jumpers and Fuses

Page 24: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-KJune 2002

2-6 Installation Rosemount Analytical Inc. A Division of Emerson Process Management

Model 7003M

CAUTION

TOTAL OUTPUT RESISTANCETotal resistance of all output devices andassociated interconnection cable must notexceed 600 ohms.

c. Current Output Connections

Refer to Figure 2-4 on page 2-7. Isolatedcurrent output (0 to 20 or 4 to 20 mA) isstandard on the Model 7003M. Toconvert from current output to voltageoutput, see Section 2-5d below.

1. Disconnect power to the analyzer.Remove rear cover.

2. Connect the shielded recorder cableleads to the rear panel of the analyzer(Figure 2-2 on page 2-3):

TB3-7 has no connectionTB3-8 negative (-)TB3-9 positive (+)

3. Connect the other end of the outputcable to terminals of the outputdevice. Ensure that polarity iscorrect. Connect the shield to asuitable ground point.

NOTE:Current output terminals are notgrounded. Ground the current loop atsome point within the circuit to mini-mize noise or other undesirable inter-actions.

4. Replace the rear cover. Do notovertighten cover retaining screws, toavoid stripping the threads in theplastic case.

5. To connect recorder(s), see Figure2-4 on page 2-7.

d. Voltage Output Connections

Refer to Figure 2-4 on page 2-7. Toconvert the voltage output to currentoutput see Section 0 above.

To monitor the signal output with avoltage recorder or other voltage device,the standard current output must beconverted into a voltage by connecting aresistor in parallel with the output deviceas follows:

1. Determine the resistor value (R)required:

R1 ohms = (fullscale voltage,volts)/(0.02 amps)

2. Disconnect power; remove the rearcover.

3. Connect resistor (1/4 watt or greater)of value R between TB3-8 and TB3-9(Figure 2-4 on page 2-7).

4. Connect leads of a shielded recordercable to the rear panel of the analyzer(Figure 2-2 on page 2-3):

TB3-7 has no connectionTB3-8 negative (-)TB3-9 positive (+)

5. Connect other end of output cable toterminals of output device. Ensurethat polarity is correct.

6. Replace the rear cover. Do notovertighten cover retaining screws, toavoid stripping the threads in theplastic case.

7. For the low end point to be 0 volts, setthe current output range to 0 to 20mAwhen setting up the instrument(Section 3-5b on page 3-13).

Page 25: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-K

June 2002

Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-7

Model 7003M

Figure 2-4. Current Output Connections

789

RECORDER

S - +

NEG (-)POS (+)

TB3

789

RECORDERA

S - +

NEG (-)POS (+)

RECORDERB

S - +

ONE RECORDER

MULTIPLERECORDERS

Page 26: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-KJune 2002

2-8 Installation Rosemount Analytical Inc. A Division of Emerson Process Management

Model 7003M

2-6 RELAY CONTACTS FOR ALARMS ORON/OFF-CONTROLS

NOTE:Whenever "Alarm" is used, it refers toeither alarm or ON/OFF Control, de-pending on how the relays are con-nected and configured by thecustomer.

RatingsTwo sets of alarm relay contacts areprovided for actuation of customer-supplied alarm and/or ON/OFF-Controlfunctions. (see Section 1-6 specificationson page 1-4).

Power FailureRemoval of AC power from the analyzer,as in a power failure, removes power fromboth relay coils. The Model 7003Mshould operate on a different AC powersource than the Alarm relay contacts.

Radio Frequency Interference (RFI)If alarm contacts are connected to anydevice that causes radio frequencyinterference (RFI), an arc suppressor (PN858728) should be used to minimize RFI.

Connecting Devices to Relay Contacts

WARNING

RELAY CONTACTSRelay contacts wired to separate powersource must be disconnected beforeservicing.

Connect alarm or ON/OFF-Controldevices to the relay contacts as follows:

1. Disconnect power; remove rear coverof analyzer.

2. Connect leads from the externalalarm or ON/OFF Control system(see Figure 2-2 on page 2-3).

TB3-1 Alarm 1 ComTB3-2 Alarm 1 NCTB3-3 Alarm 1 NOTB3-4 Alarm 2 ComTB3-5 Alarm 2 NCTB3-6 Alarm 2 NO

3. Connect other end of output cable toterminals of alarm device.

4. Replace the rear cover. Do notovertighten cover retaining screws, toavoid stripping the threads in theplastic case.

NOTE:For installation in a Class I, Division 2location, refer to the Specifications inthe Preface section, and the approvallabel on the analyzer for Factory Mu-tual limitations on contact ratings.

Page 27: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-K

June 2002

Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-1

Model 7003M

Rosemount Analytical

MODEL 7003M

DISPLAY ACK ALARM1

ALARM2

CAL ENTER

PERCENT OXYGENANALYZER

ALARM1

ALARM2

%O2

°C

SECTION 3OPERATION

3-1 OVERVIEW

It is recommended that the operator read theentire manual through before first operatingthe instrument.

The operator can generally operate the Model7003M using the flow chart in Figure 3-2 onpage 3-2 after the display symbols and modesof operation are understood.

3-2 KEYPAD OPERATION

During each particular keystroke sequence,the analyzer will recognize only certain keys.If an unrecognized key is pressed, theinstrument will either not respond or willdisplay "SrY" ("Sorry") and return to RUNMode.

Generally, to escape from any routine, pressENTER repeatedly until the instrument returnsto the RUN Mode and the oxygenconcentration is displayed.

Changing Parameters With The KeypadAll operation parameters are set with thekeypad. When the instrument is waiting foroperator input, the display will flash a digit. Atthis point, if the number itself can be changed,the operator can generally increase thenumber (UP arrow) or decrease it (DOWNarrow) at this time. An exception is when theoperator is restricted to pre-selected choices(e.g., output current ranges, see Section 3-5bon page 3-13).

When changing a value, an arrow key can bepressed for a moment, and the display willchange by one digit. If it is held down, thedisplay will change, slowly at first, thenrapidly.

If the display has a digit flashing and will notchange, check the range of that parameter.For example, an alarm setpoint cannot be sethigher than 25%.

When the number desired is displayed, pressENTER. The Model 7003M will accept thatnumber and continue with the next step of thatroutine.

Figure 3-1. Model 7003M Front Panel Controls and Indicators

Page 28: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-KJune 2002

3-2 Operation Rosemount Analytical Inc. A Division of Emerson Process Management

Model 7003M

RUN MODE

DISPLAY ACK Reveal Security Code

DISPLAY ACK Acknowledge AlarmDisplay Oxygen Temperature

ALARM MODEALARM1*

SETUP MODECAL + ENTER

CALIB. MODECAL

HOLD MODE

+

ENTER SECURITYCODE

ENTER SECURITYCODE

ENTER SECURITYCODE

ENTER SECURITYCODE

Set NewSecurityCode

Set SpanConcentration

ENTER ACK ENTER

ENTER

ACK

ENTERSet Concentrationfor MinimumOutput Current(0 or 4mA)

ENTERSet Concentrationfor MaximumOutput Current(20mA)

Select0-20 or 4-20mA

ENTER

FLOWZERO GAS

Set ResidualCurrent Offset

ENTER

Return toRUN Mode

ALARM 1*

ENTER

Change "OFF"Setpoint

ENTER

Return toRUN Mode

Change "ON"Setpoint

DisableAlarm

ToggleON/OFF

PRESSURECOMPENSATION

ROUTINE

CAL

ENTER

Set Span GasPressure

ENTER

Return toRUN Mode

Set SamplePressure

Return toRUN Mode

DIAGNOSTICSMODE

DISPLAY + ACK

Return toRUN Mode

DISPLAY

ALARM 1*

ToggleON/OFF

ChangeCurrentOutput

ENTER

Return toRUN Mode

Select DisplayedParameter

*Press ALARM 2 to set Alarm 2.

Figure 3-2. Model 7003M Operation Flow Diagram

Page 29: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-K

June 2002

Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts 3-3

Model 7003M

a. Single and Double Keystroke

Most procedures involve a sequence ofsingle keystrokes by which the operator

presses one key at a time. Some of themore frequently used single-keycommands are listed in Table 3-1 below.

KEY FUNCTION HOW TO USE ITACK Acknowledges an alarm condition Press once to silence alarms

DISPLAY Switch between %O2 and °C Press to display °C; press again to display O2.Output signals are not affected.

ARROW keys Increase/Decrease displayed value Press once to change value by one digit. Hold keydown to change value quicker.

CAL Access CALIBRATION mode Press to calibrate. (If security is enabled, promptsfor ID code first)

ENTER End command sequence Press to enter value or to end a command se-quence

ALARM1 Access ALARM 1 mode Press to check, set or disable Alarm 1 or ON/OFF-Control relays

ALARM2 Access ALARM 2 mode Press to check, set or disable Alarm 2 or ON/OFF-Control relays

CAL + ENTER Access SETUP mode From RUN mode, press CAL and ENTER keyssimultaneously.

UP + DOWN ARROW Access HOLD mode. From RUN mode, press UP ARROW and DOWNARROW keys simultaneously.

DISPLAY + ACK Access DIAGNOSTICS mode From HOLD mode, press DISPLAY and ACK keyssimultaneously.

DISPLAY + ACK Access security ID code From RUN mode, press DISPLAY and ACK keyssimultaneously.

Table 3-1. Key Functions

INDICATOR CARET/STATUS DESCRIPTION%O2 Caret ON Oxygen percentage displayed°C Caret ON Sensor temperature displayed, °C

ALARM1,2 Caret OFF Normal conditionALARM1,2 Caret blinking Alarm condition

- Caret ON Displays current output in mA(DIAGNOSTICS mode only)

Table 3-2. Display Indicators

Page 30: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-KJune 2002

3-4 Operation Rosemount Analytical Inc. A Division of Emerson Process Management

Model 7003M

b. Display Prompts

The Model 7003M prompts the operatorfor inputs and also indicates the function

in progress. For example, when enteringthe calibration mode, the prompt "CAL" isdisplayed for one second after the CALkey is pressed. Display acronyms arelisted in Table 3-3 through Table 3-9.

CODE DEFINITIONid enter security code promptno invalid security code entered - access denied

YES access granted or security code entered is correct

Table 3-3. Security Access Prompts and Responses

CODE DEFINITIONA-1 setpoints for Alarm 1 - ALARM functionA-2 setpoints for Alarm 2 - ALARM functionC-1 setpoints for Alarm 1 - ON/OFF - Controller functionC-2 setpoints for Alarm 2 - ON/OFF - Controller functionOn value in which to initiate an Alarm Action (On-setpoint)

OFF value in which to terminate an Alarm Action (Off-setpoint)dis Alarm is now disabled

Table 3-4. Alarm Settings Prompts and Responses

CODE DEFINITIONCAL initiate a span calibrationPPr enter process pressure valueCPr enter calibration pressure valueGAS Initiate a zero offset calibration (gas)

Table 3-5. Calibration Prompts and Responses

CODE DEFINITIONSet enter system setup mode4-0 define the current output as 4-20 or 0-20 mACLo % concentration that 0/4 mA represents (calibration low)CHi % concentration that 20 mA represents (calibration high)

Table 3-6. System Setup Prompts and Responses

Page 31: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-K

June 2002

Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-5

Model 7003M

CODE DEFINITIONA-1 toggle ON/OFF check of relay coil for Alarm 1A-2 toggle ON/OFF check of relay coil for Alarm 2D-2 performing non-intrusive test Internal RAM (16#00#..16#FF#)D-3 performing non-intrusive test on Internal RAM (16#4000#..16#5FFF)D-4 performing non-intrusive test on EEPROM (16#B600#..16#B7FF#)D-5 performing non-intrusive test on PROM (16#C000#..16#FFFF#)F-2 failed D-2 testF-3 failed D-3 testF-4 failed D-4 testF-5 failed D-5 testhld entering system Hold

OFF alarm relay is de-energized (normally-closed)On alarm relay is energized (normally-open)tSt entering on-line test diagnostics mode

Table 3-7. System Hold and Diagnostics Codes

CODE DEFINITIONcaret lit displayed

O2 sensor membrane voltage from ADCTemp sensor thermistor voltage from DC

minus sign milliamps sent to current output

Table 3-8. Data Displayed in DIAGNOSTICS Mode

CODE DEFINITIONSry invalid keypad function (Sorry)run current function completed - entering RUN Mode

Table 3-9. Miscellaneous Prompts and Responses

Page 32: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-KJune 2002

3-6 Operation Rosemount Analytical Inc. A Division of Emerson Process Management

Model 7003M

c. Error Messages

If an error condition occurs, E-# (non-fatalerror) or F-# (fatal error) will flash on the

display. Table 3-10 through Table 3-13list error codes and corrective actions.Pressing "ACK" or "ENTER" willacknowledge any alarm and the errormessage will disappear.

CODE DEFINITIONF-1 Invalid PROM (16#C000#..16#FFFF#) checksum detectedF-2 Internal RAM (16#00#..16#FF#) failed testingF-3 External RAM (16#4000#..16#5FFF) failed testingF-4 EEPROM (16#B600#.. 16#B7FF#) bulk erase failureF-5 Signal Board not connected

F-6 thru F-9 EEPROM write failure

Table 3-10. Power ON, Reset Diagnostics Errors (Fatal Errors

CODE DEFINITIONE-1 5 volts power supply above 5.5 VE-2 5 volts power supply below 4.5 VE-3 12 volts power supply above 13.2 VE-4 12 volts power supply below 10.8 VE-5 sensor output too low for calibrationE-6 sensor missingE-8 sample temperature is below 1°CE-9 sample temperature is above 44°C

Table 3-11. RUN Mode Errors (Non-Fatal Errors

CODE DEFINITIONHI sample temperature is above 50°CLO sample temperature is below 0°CAhi excessive sensor signal

Table 3-12. Signal Board Errors

CODE DEFINITIONCOP Computer Operating Properly error (error detected)

iot Invalid Op code Trap (internal computer command is invalid)E-0 power interruption occurred

Table 3-13. Microprocessor Errors That Reset The System

Page 33: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-K

June 2002

Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-7

Model 7003M

d. Automatic Return to RUN Mode

If no keys are pressed for five minutes,the instrument will automatically return tothe RUN Mode, except when in ZeroOffset, HOLD, or DIAGNOSTICS Modes.

CAUTION

OXYGEN WILL NOT BE MONITOREDThe instrument will remain in HOLD or DI-AGNOSTICS Modes until the operator re-turns it to the RUN Mode or otheroperation.

e. Preventing Unauthorized Access

The Model 7003M is equipped with asecurity feature to prevent unauthorizedtampering with the instrument settings.The instrument is shipped from the factorywith the security feature disabled. It canbe enabled or disabled at any time.

Security access is not required to:

• Display sensor data• Acknowledge alarms• Reveal security code

Security access is required to:

• Calibrate with span gas• Access options and diagnostics• Change operation parameters• Enable or disable alarms and

ON/OFF-Control• Disable or change the security code

After the security code is entered, theoperator can change any instrumentsetting and can go back and forth to theRUN Mode without having to enter thecode again.

If security is enabled, and the instrumentis left in the RUN Mode for five minutes,security will automatically be re-activated.Within this five minute period, anyone canaccess the functions without re-enteringthe security code.

f. Sensor Equilibrium

When a new sensor is installed, power upthe instrument and let the sensor stabilizefor about 5 hours before calibrating it.

If the oxygen sensor is calibrated at arelatively high oxygen concentration, suchas with ambient air (typically 20.95%) andthen is exposed to a low oxygen level, alonger equilibrium time period may berequired for the sensor to equilibrate.This time period depends on themonitoring concentration level and howlong the sensor was operated at thehigher level.

Page 34: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-KJune 2002

3-8 Operation Rosemount Analytical Inc. A Division of Emerson Process Management

Model 7003M

3-3 START-UP

Before proceeding, the operator should befamiliar with the keypad and indicators(Section 3-2 on page 3-1).

Operational checks and adjustments are notnormally required, provided that the sensor,inter-connection cable and analyzer havebeen properly installed. The analyzercontinually performs self-checking routinesduring operation.

Calibration with ambient air is recommendedafter start-up and at regular intervals (seeSection 4-2 on page 4-1)

The following sections provide instructions forinitial start-up and subsequent standardizationof the analyzer.

a. Initial Power-Up

1. Ensure that the proper sensor andsensor cable are connected to theanalyzer.

2. Apply power to the analyzer.

The following internal tests areperformed:

RAM Read/write test

ROM "Checksum" testEEPROM Read/write test

3. If an error code flashes, refer to theerror code lists in Table 3-10 throughTable 3-13. Press ACK to continuewith the startup procedure.

After internal tests are passed, thesoftware version is displayed for afew seconds, and the analyzer goesdirectly into RUN mode. "RUN" isbriefly displayed, then currentoxygen level.

Error codes E-0 (power failure) andE-5 (not calibrated) typically occurupon power-up.

4. Monitor the oxygen output or displayuntil the level stabilizes (typically 5hours). The alarms and currentoutput levels can be set immediately ifdesired.

5. Once the sensor has stabilized, theanalyzer is operational and is readyfor subsequent operations. Refer toTable 3-14 below for the defaultsettings present upon power-up.

Mode: AlarmState: DisabledAlarm 1Setpoints: On: 0.0 Off: 0.1%Mode: AlarmState: DisabledAlarm 2Setpoints: On: 0.0 Off: 0.1%0% of scale: 0% OxygenRange 100% of scale: 25% Oxygen

Current Output Range 4-20 mASecurity Disabled

Calibration: 76.0 cm HgPressure Process: 76.0 cm HgResidual Zero Offset 0Calibration Status Not Calibrated

Table 3-14. Default Settings at Power-Up

Page 35: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-K

June 2002

Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-9

Model 7003M

b. System Restart

The Model 7003M will attempt to restartitself in the event of a power interruption,power surge or other outside interferenceto microprocessor operation (Section 3-3cbelow).

The analyzer can also be manuallyrestarted if the RAM or EEPROMbecomes corrupted (Section 3-3d below).

c. Automatic Restart - Due to PowerFailure

If a power failure or other event occursthat affects operation of themicroprocessor the system will attempt tore-start itself. If it is successful, theinstrument will continue to operatenormally, except that the appropriate errorcode will be displayed until the ACKbutton in pressed. See Section 5-3c onpage 5-2 for circuit description.

d. Manual Restart

1. Turn power to analyzer OFF.

2. Wait 30 seconds, and turn the powerto the analyzer ON.

3. While the software revision is beingdisplayed and the self-tests are beingperformed, press DISPLAY and ACKsimultaneously.

"dYn" is displayed.

4. Press ENTER to clear the RAM andEEPROM.

The display will go blank, then continuewith the SETUP routine.

Continue with Section 3-5a on page 3-13,step 2.

Page 36: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-KJune 2002

3-10 Operation Rosemount Analytical Inc. A Division of Emerson Process Management

Model 7003M

3-4 RUN MODE

Refer to Figure 3-2 on page 3-2 and Table3-15 below. The RUN mode is the normalmode of operation, from which the operatorcan:

• Display oxygen concentration and sensortemperature

• Acknowledge alarms

• Display security code

• Access other modes

KEY DISPLAY WHAT HAPPENSDISPLAY Number change Toggles between O2 reading and temperatureACK Clears message Acknowledges AlarmsALARM 1 "A-1" then ON To Alarm 1 SETUP modeALARM 2 "A-2" then ON To Alarm 2 SETUP modeUP ARROW "SrY" "run" Stays in RUN modeDOWN ARROW "SrY" "run" Stays in RUN modeCAL "CAL" then ON To CALIBRATION modeENTER "SrY" "run" Stays in RUN modeCAL + ENTER "SEt" then ON Moves to SETUP modeUP + DOWN "hLd" then ON Moves to HOLD modeDISPLAY + ACK "123" Shows the security code

Table 3-15. RUN Mode Key Functions

a. Display Oxygen Concentration/SensorTemperature

Press DISPLAY once to display sensortemperature. Press DISPLAY again todisplay oxygen concentration.

b. Acknowledge Alarms

If an alarm setpoint has been reached, "A-1" or "A-2" will flash with a concentrationvalue. Acknowledging the alarm will stopthe flashing display and turn off the relayswitch.

To Acknowledge Alarm - Press ACKkey. If relays are configured as ON/OFFcontrollers, pressing the ACK key has noeffect.

c. Access Other Modes

The following lists which key(s) to press toaccess various modes:

SETUP = CAL+ENTERALARM = ALARM1 (or ALARM2)CALIBRATION = CALHOLD = UP and DOWN ARROWSDIAGNOSTICS = DISPLAY+ACK (from HOLD)

Page 37: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-K

June 2002

Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-11

Model 7003M

3-5 SETUP MODE

Refer to Figure 3-3 on page 3-12 and Table3-16 below. From the SET-UP mode theoperator can:

• Change security code

• Set output current range (0-20mA or 4-20mA)

• Set output concentration range endpoints

• Adjust zero offset

KEY DISPLAY WHAT HAPPENSDISPLAY Nothing NothingACK "<4>-0" / "4-<0> Selects 4-20 or 0-20 mA operationALARM 1 Nothing NothingALARM 2 Nothing NothingUP ARROW 10.8 →10.9 Changes the number displayedDOWN ARROW 10.9 →10.8 Changes the number displayedCAL Nothing NothingENTER various to "run" Moves to the next step or goes to RUN mode

Table 3-16. SETUP Mode Key Functions

Page 38: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-KJune 2002

3-12 Operation Rosemount Analytical Inc. A Division of Emerson Process Management

Model 7003M

ALARM1

ALARM2

%O2

°C

SET-UP Mode

+

SECURITY ACCESS ROUTINE(see Figure 3-4)

ALARM1

ALARM2

%O2

°C

ALARM1

ALARM2

%O2

°C

ALARM1

ALARM2

%O2

°C

FlashingMay not be displayedeach time.

ALARM1

ALARM2

%O2

°CFlashing

Prompts for Security Code.If left at 000, code will be disabled.Use Arrow keys to set Security Code.

ALARM1

ALARM2

%O2

°C

Select either 4-20mA or0-20mA Output

ACK

Prompts for low end output.(i.e., 0% O2 would be the low rangerepresenting O2 level of 0 or 4mA.

ENTER

CAL ENTER

ENTER

ALARM1

ALARM2

%O2

°CFlashing

ALARM1

ALARM2

%O2

°CENTER

Flash for 2 seconds(Hi range output selection)

ENTER

ALARM1

ALARM2

%O2

°C

Set Span output level = 20mA

FlashingMay not be displayedeach time.

Flashing

ALARM1

ALARM2

%O2

°C

ENTERSet Span output level = 20mA

ALARM1

ALARM2

%O2

°CFlash for 2 seconds Prompts for zero offset

ALARM1

ALARM2

%O2

°C

ALARM1

ALARM2

%O2

°C

Figure 3-3. SETUP Mode Flowchart

Page 39: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-K

June 2002

Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-13

Model 7003M

a. Security Code

Refer to Figure 3-4 on page 3-14.

Reveal Current Security CodeFrom RUN mode, hold down DISPLAYand ACK keys simultaneously.

Change Current Security Code1. From RUN mode, press CAL and

Enter simultaneously.

2. SET, id are briefly displayed. Then"000" is displayed with the right digitflashing.

NOTEIf RUN and CAL are not press exactlyat the same time, "SRY", or "RUN" willappear, and the analyzer will return toRUN mode.

3. If the security code has beendisabled, skip to step 4. Use theARROW keys to enter the currentsecurity code, then press ENTER.

4. If the correct code is entered, YES,SET, id will be displayed briefly, thecode will flash. If the incorrect code isentered the analyzer will return toRUN mode.

5. Use the ARROW keys to set the newsecurity code (or "000" to disable),press enter.

6. Continue with the next set-up routine,or hold down ENTER until "run"appears (the instrument is back inRUN mode).

b. Current Output Range - 0-20mA or 4-20mA

Following the last procedure, "4-0" isdisplayed with either digit flashing. "4"indicated 4-20mA (default), and "0"indicates 0-20mA.

1. Press ACK until the desired range isflashing, press ENTER.

2. Continue with the next SET-UProutine, or hold ENTER key to returninstrument to RUN mode.

c. Range Endpoint Adjustments

The oxygen values corresponding tominimum and maximum output currentmust be set. Default is 0% for 0 or 4mAand 25% for 20mA. Endpoints must bewithin the range of the instrument andmust be at least 1% oxygen apart.

Set Oxygen Low Concentration End-pointFollowing the last procedure, "CLo" isdisplayed briefly, followed by the oxygenconcentration last stored to correspond tominimum output current.

NOTEIf "CLo" is set to a value greater than"CHi", "CHi" will be automatically resetto a value 1% greater than "CLo".

Use the ARROW keys to set the lowoxygen concentration (0 to 24%), pressENTER.

"CHi" is displayed briefly, followed by thelatest oxygen concentration storedcorresponding to maximum outputcurrent.

Set Oxygen High Concentration End-point"CHi" cannot be set to a value equal to orless than "CLo".

Use the ARROW keys to set the highoxygen concentration (1 to 25%), pressENTER.

Continue with next SET-UP routine orhold ENTER to return to RUN mode.

Page 40: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-KJune 2002

3-14 Operation Rosemount Analytical Inc. A Division of Emerson Process Management

Model 7003M

Figure 3-4. Security Access Routine Flowchart

RUN

SET-UP Mode

CAL ENTER+

YES

ALARM CAL

HOLD Mode

+

NO

YES

OPERATOR GIVENACCESS TO SE-LECTED MODE

ENTER ID

ENTER

NO

ACCESS DENIED

Returns toRUN mode

WAS SECURITY IDENTERED <5 MIN.

AGO

IDCORRECT

?

Page 41: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-K

June 2002

Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-15

Model 7003M

d. Zero Offset Adjustment

Following the last procedure, the currentzero offset value will be flashing.

NOTEWhen flowing zero gas over the sen-sor, ensure that there are no leaks andthe gas actually has zero oxygen con-centration. Otherwise, significant er-rors will result. Zero offset adjustmentis generally not required. Errors aremore often due to incorrect zero gasoxygen concentration than residualcurrent of the oxygen sensor.

The zero offset adjustment allows theoperator to compensate for any residualsensor current. Adjustment range is 0-5%equivalent oxygen.

Sensor Current = Current fromOxygen + Residual Current

NOTEAdjusting the zero offset affects theoutput of the Model 7003M. Therefore,if the zero offset is changed, the sys-tem must be re-calibrated with spangas.

Zero Offset Is Not To Be ChangedPress ENTER to return analyzer to RUNmode, and skip the remainder of thisprocedure.

Zero Offset Is To Be Removed1. Press the UP ARROW until the

maximum value is displayed.

2. Press ENTER and re-calibrate withspan gas per Section 4-2b on page 4-1.

Zero Offset To Be Adjusted1. Expose the sensor to 100% oxygen-

free nitrogen or other inert gas andallow the sensor signal to stabilize (2-5 hours depending on sampleconditions).

2. Use the DOWN ARROW to bring thedisplayed value to zero (for maximumoffset).

3. Press ENTER to return to RUN mode.

4. Re-calibrate with span gas perSection 4-2b on page 4-1.

Page 42: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-KJune 2002

3-16 Operation Rosemount Analytical Inc. A Division of Emerson Process Management

Model 7003M

3-6 ALARM MODE VS. ON/OFF-CONTROLMODE

Refer to Figure 3-5 on page 3-17 and Table3-17 below. From the ALARM or ON/OFF-Control Mode, the operator can:

• Set alarm ON and OFF setpoints.

• Select alarm or ON/OFF-Control functionfor relays

• Disable alarm(s)

KEY DISPLAY WHAT HAPPENSDISPLAY Nothing NothingACK "dIS" Disables alarmsALARM 1 "C-1" / "A-1" Toggles between alarm and controller modeALARM 2 "C-2" / "A-2" Toggles between alarm and controller modeUP ARROW 10.8 →10.9 Changes alarm setpointDOWN ARROW 10.9 →10.8 Changes alarm setpointCAL Nothing NothingENTER "OFF" / "run" Moves to the "OFF" setpoint step or goes back to RUN mode

Table 3-17. ALARM Mode Key Functions

a. Relay Configured as Alarm

Relays are configured as alarms (ALARMMode) when immediate operator action isrequired.

When alarms are selected, "A-1" and "A-2" are displayed in the ALARM mode set-up procedure.

When an alarm condition occurs, thecaret blinks, the relay activates, and thedisplay flashes. Press the ACK key toacknowledge the alarm, de-energize therelay coil and stop the alarm actions.

b. Relay Configured as ON/OFF-Controller

Relays are configured as ON/OFF-Controllers (ON/OFF-Controller mode)when a non-critical, non-hazardousautomated action is desired, or when it isimportant that the relay output not bebypassed by the operator.

When ON/OFF-Controllers are selected,"C-1" and/or "C-2" is displayed in theALARM mode set-up procedure.

When an ON/OFF-Controller alarmcondition occurs, Alarm 1 or 2 caret blinksand the relay coil is energized. However,the display does not flash, and the ACKkey is inactive. Alarm conditions areautomatically reset when the sensorindicates that an alarm condition nolonger exists.

Page 43: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-K

June 2002

Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-17

Model 7003M

Figure 3-5. ALARM Mode Flowchart

ALARM/CONTROLLER MODE

SECURITY ACCESS ROUTINE(see Figure 3-4)

ALARM1

ALARM2

%O2

°C

ALARM1

ALARM2

%O2

°CPrompts "ON" Setpoint

ALARM1

ALARM2

%O2

°CDisplays "ON" Setpoint

Note: Alarms are disabled when instrument is received.Alarm 2 uses identical sequence.

To Disable Alarm ACK

Adjust Alarm OFF Setpoint

ENTER

ALARM1

ALARM

1

Indicates Alarmis disabled

ALARM1

ALARM2

%O2

°C

ENTER

Flashes AlarmConfiguration

Flashing

Flashes ControlConfiguration

ALARM1

ALARM2

%O2

°C

ALARM1

ALARM2

%O2

°C

ALARM1

ALARM2

%O2

°C

OR

ALARM1

ALARM2

%O2

°C ALARM1

ALARM2

%O2

°CPrompts for OFF Setpoint

Adjust Alarm ON SetpointSTANDARD ALARM Change Relay

configuration

ALARM1

ALARM2

%O2

°C

ALARM1

ALARM2

%O2

°C

1

1

New values/functions will not be stored until instrument returns to RUN mode with relay activated.1

Page 44: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-KJune 2002

3-18 Operation Rosemount Analytical Inc. A Division of Emerson Process Management

Model 7003M

c. Relay Contacts

Relay contacts are listed in Table 3-18 below:

TERMINAL LOCATIONSRELAY POS ALARM1 ALARM2NO (Normally open) TB3-3 TB3-6NC (Normally closed) TB3-2 TB3-5COM (Common) TB3-1 TB3-4

Relay positions are with power removed.

Table 3-18. Alarm Relay Contact Positions/Terminal Locations

d. Fail Safe Operation

The relays are setup to operated in "FailSafe" configuration. In this configurationthe relay coils are powered when the unitis NOT in an alarm condition. When analarm condition occurs, power is removedfrom the relay coils.

If a power failure occurs, power will beremoved from the relays and an alarmcondition will be indicated to the deviceconnected to the relays.

e. Relay Reset Upon Fatal Error

If a fatal error is detected, the unit willattempt to reset itself which will makeboth relays go into alarm mode. If themicroprocessor is able to continue itsoperation, the relays will then be reset totheir normal operating position.

f. Pressure Effects on Alarms

If the sample pressure changes, thepartial pressure of oxygen, and thus thereading, will change and the alarm relayswill not actuate at the same oxygen levelsas they were set when the interment waslast calibrated. See Section 5-2a on page5-1 for a more detailed description ofpressure effects.

g. Setting Alarm and ON/OFF-ControllerRelays

Wherever "Alarm" is referred to in thissection, the term "ON/OFF-Controller" canbe substituted, unless specifically calledout. Also, any reference to A1 or Alarm 1also applies to Alarm 2, and to ON/OFF-Controllers 1 and 2.

h. Setting Alarm Setpoints

This procedure can be used to:

• Verify alarm relay setpoints

• Change setpoints

• Enable alarms

• Disable alarms

Perform the following procedure for eachAlarm setpoint, as required. Figure 3-6illustrates how relays are energized andde-energized in High and Low Alarmconfigurations.

Page 45: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-K

June 2002

Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-19

Model 7003M

ACTIVATES DE-ACTIVATES

OFF SETPOINT

SETPOINTON

OFF8.24%

8.10%

TIME

ALARM ON

DEADBAND

LOW Alarmset for 8.10% O2

ACTIVATES DE-ACTIVATES

SETPOINT

OFF SETPOINT

ON

OFF

8.24%

8.10%

TIME

ALARM ON

HIGH Alarmset for 8.24% O2

i. Setting Alarm ON and OFF Levels

Refer to flowchart Figure 3-5 on page 3-17.

From RUN mode, press ALARM 1.

A-1, YES, A-1 (or C-1), ON flash briefly;the alarm ON setpoint is displayed withthe right digit blinking.

A-1 indicates relay 1 acts as an ALARM;C-1 indicates relay one acts as anON/OFF-Control.

To change the relay action fromALARM to ON/OFF Control function (orvise versa) press ALARM 1. C-1 or A-1will flash indicating a new mode.

To disable ALARM 1 - Press ACK."DIS" will flash briefly and return to RUNmode.

To set the ON setpoint for ALARM 1 -Use the ARROW keys, press ENTER.OFF is flashed briefly, then the alarm OFFsetpoint is displayed with the right digitflashing.

To set the OFF setpoint for ALARM 1 -Use the ARROW keys, press ENTER.The instrument displays "run" and returnsto RUN mode.

Figure 3-6. Action of Alarm Relay ON and OFF Setpoints

Page 46: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-KJune 2002

3-20 Operation Rosemount Analytical Inc. A Division of Emerson Process Management

Model 7003M

Page 47: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-K

June 2002

Rosemount Analytical Inc. A Division of Emerson Process Management Calibration and Adjustments 4-1

SECTION 4 CALIBRATION AND ADJUSTMENTS

4-1 FREQUENCY OF CALIBRATION

When a new sensor is installed, adjust thezero offset current (Section 3-5d on page 3-15) and calibrate the instrument with spangas (4-2b on page 4-1. Span theinstrument weekly at first and note thestability. Use this information to determinethe regular calibration frequency requiredby this particular application.

4-2 CALIBRATION MODE

Refer to Figure 4-1 on page 4-3and Table4-1 below. From the CALIBRATION Mode,the operator can:

• Calibrate with span gas

• Enter sample and span gas pressures

KEY DISPLAY WHAT HAPPENSDISPLAY Nothing NothingACK Nothing NothingALARM 1 Nothing NothingALARM 2 Nothing NothingUP ARROW 10.8 →10.9 Changes calibration factorDOWN ARROW 10.9 →10.8 Changes calibration factorCAL "PPr" Moves to pressure compensation modeENTER "run" Drops back to RUN mode

Table 4-1. CALIBRATION Mode Key Functions

a. Zero Calibration

Each oxygen sensor produces aconstant residual electrical current inthe absence of oxygen due tocharacteristics of the sensor. Typicallythis current is less that 0.1% apparentoxygen. It is different for each sensor,but will remain constant for the life ofthat sensor. The instrument can bezeroed by adjusting the residual currentper Section 3-5d on page 3-15 Afterthis procedure is performed once, thezero should not need adjusting againduring the life of the sensor.

b. Span Calibration

1. Expose the sensor to ambient air1

and let the oxygen reading andsensor temperature stabilize.

1 We recommend using ambient air in calibration, re-

gardless of the range. Only when monitoring at 0 to

2. Press CAL to enter theCALIBRATION mode. Enter thesecurity code if required.

3. CAL flashes and oxygenconcentration is displayed with theright digit flashing. Output is heldat last reading until returned toRUN mode.

4. Use the ARROW keys to set theinstrument to the concentration ofthe calibration gas. See Table 4-2on page 4-2 and Table 4-3 on page4-2 if using ambient air as spangas.

1% range, can the accuracy of the instrument beslightly improved by using a gas of lower concentra-tion (e.g., 1%) for calibration. If bottled air is used,verify the oxygen concentration against National Bu-reau of Standards certified calibration gas. The trueoxygen concentration in cylinders is often differentfrom the concentration indicated on the cylinder label.

Page 48: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-KJune 2002

4-2 Calibration and Adjustments Rosemount Analytical Inc. A Division of Emerson Process Management

Model 7003M

5. Press ENTER to return to the RUNmode.

6. If the ratio between theconcentration value entered andthe sensor voltage is excessive, anerror message will be flashed.

7. If the sample gas and thecalibration gas are each at constantpressure, continue with Section 4-2c on page 4-4 and enter thepressure of each.

Oxygen 20.946% ±0.002Nitrogen 78.084% ±0.004Argon 0.934% ±0.004CO2 0.033% ±0.001Water Vapor 0.000%Other Gases Balance

760 mm Hg; 0% Relative Humidity1

Table 4-2. Composition of Dry Atmospheric Air

PERCENT COMPOSITION ( .002%) A@GAS 15°C 20°C 25°C 30°COxygen 20.641% 20.510% 20.338% 20.115%Water Vapor 1.683 2.307 3.126 4.187Other Gases Balance Balance Balance Balance

760 mm Hg; 100% Relative Humidity2

Table 4-3. Composition of Humid Atmospheric Air

1 CRC Handbook of Chemistry and Physics, R. Weast, ed., Chemical Rubber Co., Boca Raton, FL: CRC Press, 1987, p. F-

148.2 Computed from data in CRC Handbook of Chemistry and Physics, p. D-190.

Page 49: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-K

June 2002

Rosemount Analytical Inc. A Division of Emerson Process Management Calibration and Adjustments 4-3

Model 7003M

Figure 4-1. CALIBRATION Mode and PRESSURE COMPENSATION Flowchart

CALIBRATION MODE

CAL

ALARM1

ALARM2

%O2

°C

SECURITY ACCESS ROUTINE(see Figure 3-4)

ALARM1

ALARM2

%O2

°C

Flash for 2 seconds(Process Pressure)

ENTER

ENTER

PRESSURE COMPENSATION SEQUENCE(i.e., CAL at atmospheric pressure, put sensorin different process pressure.)

ALARM1

ALARM2

%O2

°CFlash for 2 seconds(Calibration Pressure)

Input Process Pressure in cm Hg

ALARM1

ALARM2

%O2

°C

ALARM1

ALARM2

%O2

°C

ALARM1

ALARM2

%O2

°C

ALARM1

ALARM2

%O2

°C

CURRENT %O2READING

ALARM1

ALARM2

%O2

°C

ENTER

STANDARD CALIBRATIONPROCEDURE

Input Calibration Pressure in cm Hg

Example ifusing in Air

ALARM1

ALARM2

%O2

°C

ENTER

ALARM1

ALARM2

%O2

°C

Page 50: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-KJune 2002

4-4 Calibration and Adjustments Rosemount Analytical Inc. A Division of Emerson Process Management

Model 7003M

BAROMETRICPRESSURE

CM HG (IN. HG)

APPARENT O2 PARTIALPRESSURE

CM HG (DRY AIR)O2 CONCENTRATION WITH

NO PRESSURE CORRECTION1

73 (28.7) 15.3 20.24%74 (29.1) 15.5 20.5275 (29.5) 15.7 20.8076 (29.9) 15.8 20.9577 (30.3) 16.1 21.3678 (30.7) 16.3 21.63

1 Computed using 20.946% oxygen at 760 m.

Table 4-4. Barometric Pressure Effect on Apparent Oxygen Concentration

KEY DISPLAY WHAT HAPPENSDISPLAY Nothing NothingACK Nothing NothingALARM 1 Nothing NothingALARM 2 Nothing NothingUP ARROW 10.8 →10.9 Changes pressure valueDOWN ARROW 10.9 →10.8 Changes pressure valueCAL Nothing NothingENTER "CPr" / "run" Moves to calibration pressure input or goes back to RUN mode

Table 4-5. PRESSURE COMPENSATION Mode Key Functions

c. Pressure Compensation - Correctionfor Constant Pressure Differences

Refer to Figure 4-1 on page 4-3 andTable 4-5 above. If the sample andcalibration gases are each at constantpressure, the following procedure shouldbe used to correct for the pressuredifference between them. However, theinstrument does not monitor pressure andwill not correct the output if either sampleor calibration gas pressure changes.(See Section 5-2 on page 5-1).

1. From RUN mode, press CAL.

2. CAL, YES, CAL flash briefly, then theoxygen concentration is displayedwith the right digit flashing.

3. Press CAL again.

4. PPr flashes, then previous samplepressure setting is displayed with rightdigit flashing (76.0 cm Hg is default).

5. Use ARROW keys to enter samplepressure.

6. Press ENTER to accept the value.

7. CPr flashes, then previous span gaspressure setting is displayed with rightdigit flashing (76.0 cm Hg is default).

8. Use the ARROW keys to enterpressure of the calibration gas to beused. If the sample and calibrationgas will always be at the samepressure, set PPr and CPr at thesame value.

9. Press ENTER to accept the value.Instrument returns to RUN mode.

Page 51: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-K

June 2002

Rosemount Analytical Inc. A Division of Emerson Process Management Calibration and Adjustments 4-5

Model 7003M

4-3 HOLD MODE

Refer to Figure 4-2 on page 4-7 and Table 4-6below. In the HOLD mode:

• All instrument functions are locked

• All existing values are retained

• The instrument stops responding tosensor signals

• The last measured value is displayed

• The last output value remains at theisolated current output connections

• Alarms and ON/OFF-Controllers are heldin their last state

• No keys are recognized except entry tothe diagnostic routing.

KEY DISPLAY WHAT HAPPENSDISPLAY Nothing NothingACK Nothing NothingALARM 1 Nothing NothingALARM 2 Nothing NothingUP ARROW Nothing NothingDOWN ARROW Nothing NothingCAL Nothing NothingENTER "run" Drops back to RUN modeDISPLAY + ACK "d-4" "tSt" Moves to DIAGNOSTIC mode

Table 4-6. HOLD Mode Key Functions

If the sensor cable must be disconnected, theinstrument can be put in the HOLD mode, inwhich all existing signal levels and outputcurrent levels and output current levels aremaintained. This may be useful to preventdisturbing an ON/OFF-control loop connectedto the relay contacts.

1. Press the UP ARROW and DOWNARROW at the same time to put theinstrument on HOLD.

2. If the security is activated, enter thesecurity identification (id) code and pressENTER to continue.

3. Hld, YES, flash briefly, then Hld flashesalternately with the last oxygen reading.The output signal will remain unchangeduntil the instrument returns to the RUNmode.

4. Press ENTER to return to the RUN mode,or continue with Section 4-4 DiagnosticsMode on page 4-8.

Page 52: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-KJune 2002

4-6 Calibration and Adjustments Rosemount Analytical Inc. A Division of Emerson Process Management

Model 7003M

CARETLIT VALUE DISPLAYED TYPICAL

VALUE1 ADJUSTABLE

% Raw Sensor Voltage 2 Yes°C Thermistor Voltage 2.43 No-3 Current Output 17.43 Yes

Use ARROW keys to make adjustments.

Table 4-7. Caret Readings

1 These typical values were obtained under the following conditions:

O2 Concentration: 20.95%Sensor Temperature: 26.07°CCurrent Output Range: 4-20mARange Endpoint, Low: 0%Range Endpoint, High: 25%Zero Offset Signal: NonePressure Compensation: None

2 Typical voltages for new sensors at 25°C:

TYPE OF SENSOR MINIMUM NOMINAL MAXIMUMDisposable 1.05V 1.39V 1.69VRechargeable 0.97V 1.29V 1.76V

Near the end of the life of the sensor, the sensor voltage is typically 0.53V (disposable) or 0.48V (rechargeable). At tempera-tures (T) other than 25°C, multiply the voltages above by (1.025)exp(T-25).

3 "-" is a minus sign displayed on the left of the display between AL 1 and AL 2.

Page 53: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-K

June 2002

Rosemount Analytical Inc. A Division of Emerson Process Management Calibration and Adjustments 4-7

Model 7003M

Figure 4-2. HOLD and DIAGNOSTICS Modes Flowchart

HOLD MODE

SECURITY ACCESS ROUTINE(see Figure 3-4)

+

DIAGNOSTICS MODE

Last displayedvalue held

ALARM1

ALARM2

%O2

°C

ENTER

ALARM1

ALARM2

%O2

°C

ALARM1

ALARM2

%O2

°C

+

ALARM1

ALARM2

%O2

°C

ALARM1

ALARM2

%O2

°C

DISPLAY

Adjust Parameters (see Section 4-4)

CARET LIT(see Table 4-7)

ALARM1

ALARM2

%O2

°C

Display Parameters

Sensor Voltage

ENTER

WARNINGTesting Relays will affect devices connected to them

ALARM

1

ALARM1

ALARM2

%O2

°C

ALARM

1

ALARM1

ALARM2

%O2

°C

ALARM

2

ALARM1

ALARM2

%O2

°C

ALARM

2

ALARM1

ALARM2

%O2

°C

ExitDIAGNOSTICMode

Return toHOLD

Page 54: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-KJune 2002

4-8 Calibration and Adjustments Rosemount Analytical Inc. A Division of Emerson Process Management

Model 7003M

4-4 DIAGNOSTICS MODE

Refer to Figure 4-2 on page 4-7, Table 4-7 onpage 4-6 and Table 4-8 below. TheDIAGNOSTICS mode is used to:

• Check sensor circuit voltages

• Check relays

• Adjust the output current

Enter DIAGNOSTICS ModeIf the instrument is already in HOLD mode,skip step 1 and proceed to step 2.

1. From the RUN mode, press the UPARROW and DOWN ARROW at thesame time to enter the HOLD mode.

2. If the security is activated, enter thesecurity identification (id) code and pressENTER to continue.

Hld, YES, flash briefly, then Hld flashesalternately with the last oxygen reading.The output signal will remain unchangeduntil the instrument returns to the RUNmode.

a. Raw Sensor Voltage

1. Press DISPLAY and ACK at the sametime to enter DIAGNOSTICS mode.

Tst, d-3, d-4 flash briefly, then the%O2 caret lights and the raw sensorvoltage is displayed.

2. Press ENTER at any time to return toHOLD mode. Press ENTER again toreturn to RUN mode.

KEY DISPLAY WHAT HAPPENSDISPLAY "2.13" Scrolls among the DIAGNOSTIC displaysACK Nothing NothingALARM 1 "A-1" "On"/ "OFF" Toggles the Alarm 1 relay ON/OFFALARM 2 "A-2" "On" / "OFF" Toggles the Alarm 2 relay ON/OFFUP ARROW "2.13" →"2.14" Increases current output or span settingDOWN ARROW "2.14" →"2.13" Decreases current output or span settingCAL Nothing NothingENTER "hLd" Drops back to HOLD mode

Table 4-8. DIAGNOSTICS Mode Key Functions

b. Test Current Output

The current output can be tested(increased or decreased using theARROW keys) and the actual output willchange. When the instrument returns tothe RUN mode, the output will return tothe previous value.

1. Press DISPLAY until the minus sign (-) is displayed on left.

2. Use the ARROW keys to increase ordecrease the current output.

c. Linearizing Current Output

For some samples, current output doesnot track the oxygen reading displayed.The operator can linearize output andstore the compensating second-orderpolynomials in EEPROM. This data willremain and function there as long as theEEPROM is not erased. However, if a"cold boot" is performed, the operatormust repeat this linearization routine.

Page 55: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-K

June 2002

Rosemount Analytical Inc. A Division of Emerson Process Management Calibration and Adjustments 4-9

Model 7003M

To determine if current outputlinearization is necessary, do thefollowing:

1. Connect a separate current meter tothe output terminals (any device thatwill display an accurate currentmeasurement such as a computerinput will suffice).

NOTEDo not simply connect a currentmeasurement meter in parallel with theusual load because this will lead to in-correct readings.

2. Place the instrument in DIAGNOSTICmode by pressing the UP and DOWNARROWS, followed by the DISPLAYand ACK keys. The display will showconsecutively:

tSt, d-3, d-4, and oxygen inputvoltage

3. Press the DISPLAY key until theminus sign and a number aredisplayed. This reading is the currentoutput (in milliamps) as determined bythe Model 7003M.

4. Use the UP and DOWN ARROWkeys to increase the display reading,and occasionally, compare the currentdisplayed on both meters (theinstrument display and the separatemeter).

5. If the two readings are too disparate(different applications require lesseror greater tolerance), do the following3-point calibration to linearize thecurrent output:

6. Use the UP and DOWN ARROWkeys to set the current output to avalue halfway up its range. Theoperator may prefer to set it at acritical measurement point.

7. Press the CAL key. The instrumentwill flash CCA, illuminate the threeright-hand carets, and continuedisplaying the current output.

8. Use the UP and DOWN ARROWkeys to adjust the actual currentoutput (as noted on the separatemeter) to the correct amount. Notethat the instrument display does notchange.

9. Press ENTER, the instrumentdisplays diA, returns to theDIAGNOSTIC mode, and continues todisplay the current output.

10. Use the UP and DOWN ARROWkeys to adjust the current output to alower value, representing the lowrange (the operator may prefer to use4 mA to represent 0% oxygen).

11. Press CAL (the instrument displaysCCL for lower current output setting),and repeat steps 3 and 4 of thisprocedure.

12. Use the UP and DOWN ARROWkeys to adjust the current output to avalue higher than the initial calibrationvalue (step 1 of this procedure),representing the higher range (theoperator may prefer to set it at16.72mA to represent 20.9% oxygen.

13. Press CAL (the instrument displaysCCH for high current output setting),and repeat steps 3 and 4 of thisprocedure.

To return to the RUN mode, pressENTER twice.

At any time during this procedure, theoperator may press the ACK key to resetall calibration coefficients to defaultvalues. Pressing DISPLAY at any timeduring this procedure will cause theinstrument to display for a moment therelevant calibration efficient.

The operator may escape the currentCALIBRATION and DIAGNOSTIC modealtogether by pressing ENTER twice.Note, all coefficients will return to thedefault values, and the current calibrationprocedure must be started over tolinearize current output.

Page 56: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-KJune 2002

4-10 Calibration and Adjustments Rosemount Analytical Inc. A Division of Emerson Process Management

Model 7003M

NOTEThe low calibration point must be be-tween the initial calibration point (step1 of previous procedure) and zero, andthe high calibration point must be be-tween the initial calibration point and20mA.

If the third point calibrated (step 8,previous procedure) is actually an-other low range (between the initialcalibration point and CCL) it will over-ride the original CCL but won't affectthe initial calibration.

NOTEThe current output linearizing proce-dure uses second order polynomials toadjust the output linearity. The correc-tion effect is maximized at the calibra-tion point, and is smoothly diminishedtowards the endpoints.

d. Test Alarm Relays

1. Disconnect any alarm or ON/OFF-Control devices that are connected tothe analyzer, before testing.

NOTEWhen alarms are tested manually, anyalarm or ON/OFF-Control devices con-nected to the relays will respond asthey would in an alarm condition. Thealarm devices may have to be discon-nected from the relays for this test.

2. Toggle the relays ON and OFF usingthe ALARM 1 and ALARM 2 keys.

A-1 ON or A-2 OFF is displayed.The alarm contacts can be heardopening and closing.

3. Connect any alarm or ON/OFF-Control devices that weredisconnected in step 1.

4. Press ENTER to exit theDIAGNOSTICS mode.

"Hld" flashes alternately with the lastoxygen reading and the analyzerreturns to HOLD mode.

5. Press ENTER again to return to RUNmode.

4-5 ADJUSTMENTS AND SETTINGS

All operator settings and adjustments aremade using the front keypad except if thepower jumpers are to be changed. Refer tothe following sections:

2-5b Power Connections on page 2-4

3-5d Zero Offset Adjustment on page 3-15

4-4a Raw Sensor Voltage on page 4-8

4-4c Linearizing Current Output on page 4-8

Page 57: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-K

June 2002

Rosemount Analytical Inc. A Division of Emerson Process Management Theory 5-1

Model 7003M

SECTION 5 THEORY

5-1 PRINCIPLES OF OPERATION

The Model 7003M Percent Oxygen Analyzerconsists of an amperometric sensor andanalyzer interconnected by a multi-conductorshielded cable. The sensor responds to thepartial pressure of oxygen and produces acurrent signal. The analyzer amplifies andprocesses the signal, providing readout inpercent oxygen and current output.

a. Electrochemical Theory

With the sensor placed in the sample gas,a voltage is applied across the cathodeand anode. Sample gas diffuses throughthe membrane and is reduced at thecathode resulting in a current proportionalto the partial pressure of oxygen in thesample.

When no oxygen is present, a smallresidual sensor current, which can beoffset during instrument setup.

5-2 EFFECTS OF SAMPLE CONDITIONS

a. Barometric Pressure

If ambient air is used to calibrate thesensor, a change in barometric pressurecan be a significant factor. At 76 cm Hg,the partial pressure of oxygen is 15.9 cmHg (20.9% of 76 cm Hg). As atmosphericpressure deviates from 76 cm Hg, thepartial pressure of oxygen will varyproportionally. The highest partialpressure that can be read by the Model7003M is 19 cm Hg (25% oxygen at 76cm Hg).

b. Humidity

If calibrating with humid ambient air, theeffect of the partial pressure due to watervapor content should be taken intoconsideration. High humidity has theeffect of reducing the partial pressure of

oxygen without affecting the totalbarometric pressure.

This relationship can be expressed as :

P atm = Po2 + PH20 = P other,

where:

P atm = total barometric pressure,

Po2 = partial pressure of oxygen,

PH20 = partial pressure of watervapor,

P other = partial pressure of allgases other than oxygen and watervapor

c. Sample Temperature

The rate of oxygen diffusion through thesensor membrane varies withtemperature, with a coefficient of about+3% °C (5% °F), causing a correspondingchange in sensor current. The instrumentcontinuously monitors temperature with athermistor in the sensor body, andcompensates the output for both sampletemperature and membrane permeability.

Because of a slight time lag in theresponse of the temperature sensor, arapid temperature change will cause atemporary change in apparent oxygenlevel, such as might be encountered if thesensor is suddenly exposed to directsunlight from a shadow.

Extended use of the sensor at lowtemperatures will shorten the life of thesensor. As the sensor reaches the end ofits useful life, distinct step changes in theoxygen reading will be evident as theoxygen level changes.

Page 58: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-KJune 2002

5-2 Routine Servicing and Troubleshooting Rosemount Analytical Inc. A Division of Emerson Process Management

Model 7003M

d. Interfering Gases

Gases that are reduced or oxidized in therange of 0.5 to 0.8VDC contribute tosensor current and can cause a readouterror. Only a few gases have thischaracteristic. Common gases thatshould be avoided include SO2, Cl2, andoxides of nitrogen (NOX). Low-levelconcentrations of hydrogen sulfide tend tocontaminate the sensor, but may notseriously affect oxygen measurement.

If you suspect that any streamcomponents are affecting the operation ofthe sensor, contact Rosemount AnalyticalTechnical Services.

If contaminated, the rechargeable sensormust be rejuvenated and the disposablesensor must be replaced. Refer to theinstructions supplied with the sensor.

5-3 CIRCUIT DESCRIPTIONS

a. Signal Board

The Signal Board is configured to matchthe type of sensor ordered with theanalyzer. If the type of sensor ischanged, or the Signal Board is repairedor replaced, verify the rocker switchposition (SW1):

Rechargeable Sensor: Close positions1,4, and 5

Disposable Sensor: Close positions 2,4,and 6

b. Power Supply Board

The Power Supply Board provides:

• ±12V (non-isolated) for analogsignaling function

• +5V (non-isolated) formicroprocessor and associatedlogic

• +24V (non-isolated) for theisolated current function

• +5V (isolated) for other functionsfor that circuit

There are isolated and non-isolatedgrounds on the Power Supply Board.TP1 is the non-isolated ground andshould be used with TP6 and TP7.

c. Microprocessor Board

The Microprocessor Board is thefunctional center of the instrument. Withthe exception of the program chip, nouser serviceable functions are on thisassembly. If the Microprocessor Board isto be repaired or replaced, note thepositions of the configuration jumpers forsetting jumpers on replacement board.

An instrument timer circuit monitorsinstrument power for a power failure,power surge or other outside conditioninterfering with the microprocessoroperation. When conditions return tonormal, the circuit will attempt a systemrestart. If the microprocessor is stillcapable of operating, both alarm relayswill be put into alarm condition, the screenwill clear, and the output current will dropto 0 mA. Then the instrument will gothrough a normal power-up routine andwill eventually return to routine operationwith alarms and output restored to theirnormal conditions. The appropriate errorcode (E-0, etc.) will be displayed until theACK button is pressed.

d. Alarm Circuits

Each alarm activation setpoint is designedto be at a different level from thedeactivation setpoint:

• To prevent relay contacts fromchattering if the oxygen level staysaround the setpoint.

• To indicate to the microprocessorwhether alarm is HIGH or LOW.

• To allow for a time lag between thetime the alarm is activated and thetime it is deactivated for ON/OFF-Control purposes.

Page 59: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-K

June 2002

Rosemount Analytical Inc. A Division of Emerson Process Management Routine Servicing and Troubleshooting 6-1

Model 7003M

SECTION 6ROUTINE SERVICING AND TROUBLESHOOTING

WARNING

ELECTRICAL SHOCK HAZARDServicing requires access to live partswhich can cause death or serious injury.Refer servicing to qualified personnel.

Relay contacts wired to separate powersources must be disconnected beforeservicing.

6-1 ROUTINE SERVICING

Most service and maintenance involvescleaning and recharging the sensor. Referto the instructions supplied with the sensorfor recharging and maintenance procedures.

Service is rarely required on the Model7003M Percent Oxygen Analyzer. Repair islimited to replacement of PCB boards,EPROM and fuses. See Section 7Replacement Parts.

6-2 TROUBLESHOOTING

Troubleshooting should be referred toqualified service technicians. Refer toSection 5-3 Circuit Descriptions, on page 5-2, as an aide in isolating a problem.

a. Symptoms

Calibration error code displayedPossible causeExhausted Sensor

Check/TestInstall know good sensor and isolate

Corrective ActionRecharge or replace sensor

Cannot adjust signal to calibrationvaluePossible causeSensor voltage error

Check/TestInstall know good sensor and verifyvoltage

Corrective ActionRecharge or replace sensor

Readings not consistent with calibra-tion curvePossible causeIncorrect sensor used for application

Check/TestVerify sensor corresponds withapplication

Corrective ActionInstall new sensor and calibrate withspan gas

Possible causeSensor temperature incorrect

Check/TestSensor temperature should be 0 to 50°C

Corrective ActionInstall new sensor and calibrate withspan gas

Possible causeIncorrect span gas concentration

Check/TestVerify span gas concentration

Corrective ActionInstall new sensor and calibrate withspan gas

Page 60: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-KJune 2002

6-2 Routine Servicing and Troubleshooting Rosemount Analytical Inc. A Division of Emerson Process Management

Model 7003M

Error code E-0 displayedPossible causePower interruption, power surge or otheroutside interference has occurred

Corrective ActionPress ACK. If this symptom occursfrequently, surge-protection isrecommended.

Will not read low levels of oxygenPossible causeAmbient air leaking into system

Check/TestIncrease sample flow rate, block offflow of gas in and out of flowchamber, and not analyzer response

Corrective ActionIf signal increases with test, isolateand seal leak.If signal does not increase, replace orrecharge sensor.

Sudden change in output for no ap-parent reasonPossible causeSensor subjected to sudden change inconditions, as when a shadow movesand exposes the sensor to an increasein temperature

Corrective ActionLocate sensor in an area shaded fromsunlight and away from cold or warmdrafts.

Resolution degrading with timePossible causeSensor exposed to low temperatures.

Check/TestPer Section 4-4b on page 4-8, checkcurrent output of sensor.

If the sensor is degrading, the signal willchange in steps as the oxygen levelchanges instead of changingcontinuously.

Corrective ActionRelocate sensor or replace morefrequently.

Page 61: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-K

June 2002

Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts 7-1

Model 7003M

SECTION 7REPLACEMENT PARTS

7-1 CIRCUIT BOARD REPLACEMENT POLICY

In most situations involving a malfunction of acircuit board, it is more practical to replace theboard than to attempt isolation andreplacement of an individual component. Asstandard policy, rebuilt boards are availableon an exchange basis.

Because of the exchange policy coveringcircuit boards, the following sections onreplacement parts do not include individual,circuit board mounted components. Ifcircumstances necessitate replacement of anindividual component, which can be identifiedby inspection or from the schematic diagrams,obtain the replacement component from alocal source of supply.

7-2 MATRIX

7003M MODEL 7003M OXYGEN MONITOR*

Code Power01 115 VAC, 50/60 Hz05 230 VAC, 50/60 Hz

Code Mounting Options01 Panel Mount - Standard07 Wall Mounting Plate w/ Junction Box08 Two-Inch Pipe Mounting Bracket

Code Gain Setting50 Signal: High Level (Used with Rechargeable Sensor % Gas)51 Signal: Low Level (Used with Disposable Sensor % Gas)99 Signal: Special

7003M 01 01 50

Page 62: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-KJune 2002

7-2 Replacement Parts Rosemount Analytical Inc. A Division of Emerson Process Management

Model 7003M

7-3 REPLACEMENT PARTS - MODEL 7003M

WARNING

PARTS INTEGRITYTampering or unauthorized substitution of components may adversely affect safety of this product.Use only factory documented components for repair

PART NUMBER DESCRIPTION652630 Signal Board652640 Power Supply Board654605 Display/Keypad Board655195 EPROM655197 Microprocessor Board777156 Fuse, 1/4 A, 120V (pkg of 5)777360 Fuse, 1/8 A, 240 V (pkg of 5)858728 Arc Suppressor

When replacing Microprocessor Board PN 652650, replace with Microprocessor Board PN 655197 ANDEPROM 655195.

Page 63: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-K

June 2002

Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts 7-3

Model 7003M

7-4 REPLACEMENT PARTS - SENSORS

a. Rechargeable Sensors

Sensor:

MATERIAL PART NUMBERPolypropylene 623371Polypropylene 623370 (use with Fast Response Kit)Ryton 190408Ryton 190409 (use with Fast Response Kit)

Sensor Kits (Sensor not included):

DESCRIPTION PART NUMBER USE WITH SENSORSubmersion, Polypropylene 639904 623371In-Line Flow, Polypropylene 639905 623371Fast Response, Polypropylene 639906 623370Submersion, Ryton 646628 190408In-Line Flow, Ryton 646629 190408Fast Response, Ryton 646630 190409

Sensor Kits (Sensor included):

DESCRIPTION PART NUMBERSubmersion, Polypropylene 400011In-Line Flow, Polypropylene 400012Fast Response, Polypropylene 400013Submersion, Ryton 400021In-Line Flow, Ryton 400022Fast Response, Ryton 400023

Recharge KitRecharge Kit (10 recharges) 191755

Page 64: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-KJune 2002

7-4 Replacement Parts Rosemount Analytical Inc. A Division of Emerson Process Management

Model 7003M

b. Disposable Sensors

Sensor:

MATERIAL PART NUMBERPolypropylene 623742

Sensor Kits (Sensor not included):

DESCRIPTION PART NUMBER USE WITH SENSORSubmersion, Polypropylene 623715 623742In-Line Flow, Polypropylene 623716 623742

Sensor Kits (Sensor included):

DESCRIPTION PART NUMBERSubmersion, Polypropylene 500011In-Line Flow, Polypropylene 500012

Page 65: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-K

June 2002

Rosemount Analytical Inc. A Division of Emerson Process Management Return of Material 8-1

Model 7003M

SECTION 8RETURN OF MATERIAL

8-1 RETURN OF MATERIALIf factory repair of defective equipment isrequired, proceed as follows:1. Secure a return authorization from a

Rosemount Analytical Inc. Sales Office orRepresentative before returning theequipment. Equipment must be returnedwith complete identification in accordancewith Rosemount instructions or it will notbe accepted.Rosemount CSC will provide the shippingaddress for your instrument.In no event will Rosemount beresponsible for equipment returnedwithout proper authorization andidentification.

2. Carefully pack the defective unit in asturdy box with sufficient shock absorbingmaterial to ensure no additional damageoccurs during shipping.

3. In a cover letter, describe completely:• The symptoms that determined the

equipment is faulty.• The environment in which the

equipment was operating (housing,weather, vibration, dust, etc.).

• Site from where the equipment wasremoved.

• Whether warranty or non-warrantyservice is expected.

• Complete shipping instructions for thereturn of the equipment.

4. Enclose a cover letter and purchase orderand ship the defective equipment accord-ing to instructions provided in the Rose-mount Return Authorization, prepaid, tothe address provided by Rosemount CSC.

Rosemount Analytical Inc.Process Analytical DivisionCustomer Service Center

1-800-433-6076

If warranty service is expected, the defectiveunit will be carefully inspected and tested atthe factory. If the failure was due to theconditions listed in the standard Rosemountwarranty, the defective unit will be repaired orreplaced at Rosemount’s option, and anoperating unit will be returned to the customerin accordance with the shipping instructionsfurnished in the cover letter.For equipment no longer under warranty, theequipment will be repaired at the factory andreturned as directed by the purchase orderand shipping instructions.

8-2 CUSTOMER SERVICEFor order administration, replacement Parts,application assistance, on-site or factoryrepair, service or maintenance contractinformation, contact:

Rosemount Analytical Inc.Process Analytical DivisionCustomer Service Center

1-800-433-6076

8-3 TRAININGA comprehensive Factory Training Program ofoperator and service classes is available. Fora copy of the Current Operator and ServiceTraining Schedule contact the TechnicalServices Department at:

Rosemount Analytical Inc.Customer Service Center

1-800-433-6076

Page 66: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-KJune 2002

8-2 Return of Material Rosemount Analytical Inc. A Division of Emerson Process Management

Model 7003M

Page 67: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

WARRANTY

Goods and part(s) (excluding consumables) manufactured by Seller are warranted to be free fromdefects in workmanship and material under normal use and service for a period of twelve (12)months from the date of shipment by Seller. Consumables, glass electrodes, membranes, liquidjunctions, electrolyte, o-rings, etc., are warranted to be free from defects in workmanship andmaterial under normal use and service for a period of ninety (90) days from date of shipment bySeller. Goods, part(s) and consumables proven by Seller to be defective in workmanship and/ormaterial shall be replaced or repaired, free of charge, F.O.B. Seller's factory provided that thegoods, part(s) or consumables are returned to Seller's designated factory, transportation chargesprepaid, within the twelve (12) month period of warranty in the case of goods and part(s), and inthe case of consumables, within the ninety (90) day period of warranty. This warranty shall be ineffect for replacement or repaired goods, part(s) and the remaining portion of the ninety (90) daywarranty in the case of consumables. A defect in goods, part(s) and consumables of the com-mercial unit shall not operate to condemn such commercial unit when such goods, part(s) andconsumables are capable of being renewed, repaired or replaced.

The Seller shall not be liable to the Buyer, or to any other person, for the loss or damage directlyor indirectly, arising from the use of the equipment or goods, from breach of any warranty, or fromany other cause. All other warranties, expressed or implied are hereby excluded.

IN CONSIDERATION OF THE HEREIN STATED PURCHASE PRICE OF THE GOODS,SELLER GRANTS ONLY THE ABOVE STATED EXPRESS WARRANTY. NO OTHER WAR-RANTIES ARE GRANTED INCLUDING, BUT NOT LIMITED TO, EXPRESS AND IMPLIEDWARRANTIES OR MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.

Limitations of Remedy. SELLER SHALL NOT BE LIABLE FOR DAMAGES CAUSED BY DE-LAY IN PERFORMANCE. THE SOLE AND EXCLUSIVE REMEDY FOR BREACH OF WAR-RANTY SHALL BE LIMITED TO REPAIR OR REPLACEMENT UNDER THE STANDARDWARRANTY CLAUSE. IN NO CASE, REGARDLESS OF THE FORM OF THE CAUSE OF AC-TION, SHALL SELLER'S LIABILITY EXCEED THE PRICE TO BUYER OF THE SPECIFICGOODS MANUFACTURED BY SELLER GIVING RISE TO THE CAUSE OF ACTION. BUYERAGREES THAT IN NO EVENT SHALL SELLER'S LIABILITY EXTEND TO INCLUDE INCIDEN-TAL OR CONSEQUENTIAL DAMAGES. CONSEQUENTIAL DAMAGES SHALL INCLUDE, BUTARE NOT LIMITED TO, LOSS OF ANTICIPATED PROFITS, LOSS OF USE, LOSS OF REVE-NUE, COST OF CAPITAL AND DAMAGE OR LOSS OF OTHER PROPERTY OR EQUIPMENT.IN NO EVENT SHALL SELLER BE OBLIGATED TO INDEMNIFY BUYER IN ANY MANNERNOR SHALL SELLER BE LIABLE FOR PROPERTY DAMAGE AND/OR THIRD PARTY CLAIMSCOVERED BY UMBRELLA INSURANCE AND/OR INDEMNITY COVERAGE PROVIDED TOBUYER, ITS ASSIGNS, AND EACH SUCCESSOR INTEREST TO THE GOODS PROVIDEDHEREUNDER.

Force Majeure. Seller shall not be liable for failure to perform due to labor strikes or acts beyondSeller's direct control.

Page 68: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M

Operator Manual748223-KJune 2002

© Rosemount Analytical Inc. 2001

Emerson Process ManagementRosemount Analytical Inc.Process Analytic Division1201 N. Main St.Orrville, OH 44667-0901T (330) 682-9010F (330) 684-4434E [email protected]

ASIA - PACIFICFisher-RosemountSingapore Private Ltd.1 Pandan CrescentSingapore 128461Republic of SingaporeT 65-777-8211F 65-777-0947http://www.processanalytic.com

Fisher-Rosemount GmbH & Co.Industriestrasse 163594 HasselrothGermanyT 49-6055-884 0F 49-6055-884209

EUROPE, MIDDLE EAST, AFRICAFisher-Rosemount Ltd.Heath PlaceBognor RegisWest Sussex PO22 9SHEnglandT 44-1243-863121F 44-1243-845354

LATIN AMERICAFisher - RosemountAv. das Americas3333 sala 1004Rio de Janeiro, RJBrazil 22631-003T 55-21-2431-1882

Model 7003M

Page 69: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M
Page 70: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M
Page 71: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M
Page 72: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M
Page 73: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M
Page 74: Manual: 7003M Percent O2 Analyzer - Rev K - · PDF fileOperator Manual 748223-K June 2002 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model 7003M