Mark Powell 24 October 2016 October 24, 2016 Confidentiality Label 1 Making LC Connections

Transcript of Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections •...

Page 1: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

Mark Powell

24 October 2016

October 24, 2016

Confidentiality Label

1

Making LC Connections

Page 2: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

LC Connections

• Problems with improper connections • Source of leaks • Mistaken for chromatography issues

• Making connections can vary with technique

• Different manufacturers supply different types of fittings

• Capillary tubing choices

• Contribute to dispersion and extra-column volume

10/24/2016

Confidentiality Label

Page 3: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

Dispersion Reduces HPLC Performance

What is dispersion?

• Original sample concentration being diluted as it is carried through the system plumbing (extra-column volume)

What increases dispersion?

• Connecting tubing that is too long

• Connecting tubing that is too large in diameter

• Connections that have gaps and form small mixing chambers

Page 4: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

Extra-Column Volume

October 24, 2016

Confidentiality Label

4

• Flow cell

• Heat exchanger

• Switching valve

• Needle seat

• Connecting capillaries

Page 5: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

5

Extra-Column Volume

Page 5

Column: StableBond SB-C18, 4.6 x 30 mm, 3.5 mm Mobile Phase: 85% H2O with 0.1% TFA : 15% ACN Flow Rate: 1.0 mL/min Temperature: 35°C Sample: 1. Phenylalanine 2. 5-benzyl-3,6-dioxo-2-piperazine acetic acid 3. Asp-Phe 4. Aspartame

10 µL extra-column volume 50 µL extra-column volume

Time (min)0.0 0.5 1.0 1.5 2.0

4

32

1

Time (min)0.0 0.5 1.0 1.5 2.0

4

32

1

Page 6: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

Aris-Taylor Equation peak dispersion in cylindrical tubing

October 24, 2016

Confidentiality Label

6

Page 7: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

Tubing Volume

Tubing Length 10mm 50mm 100mm 150mm

Tubing ID Volume Volume Volume Volume

0.17mm (green) 0.227 uL 1.1uL 2.27 uL 3.3 uL 0.12mm (red) 0.113 uL 0.55uL 1.13 uL 1.65 uL

Page 8: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

As Efficiency Increases Peak Width Decreases System Dispersion Becomes More of a Factor

Columns: Eclipse Plus C18, as described below. Mobile Phase: A: water, B: MeOH, (15:85) Injection volume: 6uL Temperature: 25°C Flow: 1 mL/min. Detection: 310, 4 nm, 0.5 s response time, semi-micro flow cell, Sample: Sunscreens

min 0 2 4 6 8 10 12 14

mAU

0

20

40

60

min 0 2 4 6 8 10 12 14

mAU

0 20 40 60 80

100

min 0 2 4 6 8 10 12 14

mAU

0

50

100

150

4.6 x 100 mm, 3.5 µm

1

2 3 4

Rs3,2= 6.65

Rs3,2= 6.51

Rs3,2= 6.41

4.6 x 50 mm, 1.8 µm

4.6 x 150 mm, 5 µm

Page 9: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

Approximate Extra Column Volume Limits for This 4.6mm i.d. Column Example

5um 3.5um 1.8um

Peak (1) Width in uL 181 136 45 Injection (1/10 P.V.) 18 14 5

Extra Column (1/3 P.V.) 60 45 15 ECV minus Inj. Vol. 42 21 10 ECV Estimate for

2.1mm i.d. Column 8 4 2

Page 10: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

Optimizing Connecting Tubing Volume For UHPLC Columns

min 0.5 1 1.5 2 2.5

mAU

0

100

200

350

400

550

600 System Tubing Volume Optimized 0.12mm i.d. tubing

Peak width 0.018 min

Peak width 0.019 min

Resolution 1.902

min 0.5 1 1.5 2 2.5

mAU

0

100

200

300

400

System Tubing Volume Not Optimized 0.17mm i.d. tubing

Peak width 0.038 min

Peak width 0.037 min

Resolution 0.961

Page 11: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

11

Stainless Steel vs. Polymeric Fittings

Stainless Steel: • Agilent uses Swagelok type fittings with front and back

ferrules • Also available with a long nut PEEK (<400 bar system pressure):

• Connections are changed frequently • Connecting columns • Pressure is less critical • Fits on SS or PEEK tubing

Polyketone:

• Easy, hand tightened column connection • Used up to 600 bar (PN: 5042-8957) • Fits on SS Tubing

Page 11

Page 12: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

October 24, 2016

Confidentiality Label

12

Swagelok • Two piece ferrule • Used on Agilent LCs • Short nut • Also available with long nut 0.090 in.

0.090 in.

0.130 in.

Parker • One piece ferrule • Short nut • Very similar to Swagelok

Waters • Longer nut • Used on Alliance systems

Type of Fittings

Page 13: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

Types of Fittings

October 24, 2016

Confidentiality Label

13

Page 14: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

Potential Fittings Issues

10/24/2016

• Leak

• Peak shape problem

• No dead volume

Confidentiality Label

Page 15: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

Fitting Mismatch

10/24/2016

Confidentiality Label

Page 16: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

Proper Fit

10/24/2016

Confidentiality Label

Page 17: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

Fitting Mismatch

10/24/2016

Confidentiality Label

Page 18: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

Proper Fit

10/24/2016

Confidentiality Label

Page 19: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

Polymeric Fittings

10/24/2016

Confidentiality Label

Page 20: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

Peak Shape

10/24/2016

Confidentiality Label

Page 21: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

Peak Shape

10/24/2016 Pag 21

min0 2 4 6 8 10

mAU

0

10

20

30

40

DAD1 A, Sig=254,4 Ref=360,100 (MERCURY 20... DATA; MERCURY 2010 ECLIPSE PLUS 3X100,1.8\USUYB0145300037.D)

min0 2 4 6 8 10

mAU

0

10

20

30

40

DAD1 A, Sig=254,4 Ref=360,100 (MERCURY 20...AND PRESS LOAD ECLIPSE PLUS C18 3X100, 1.8\USUYB0145500037.D)

Confidentiality Label

Page 22: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

Swaging Fittings

October 24, 2016

Confidentiality Label

22

Page 23: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

Swaging Fittings

October 24, 2016

Confidentiality Label

23

Page 24: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

Swaging Fittings

October 24, 2016

Confidentiality Label

24

Page 25: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

Swaging Fittings

October 24, 2016

Confidentiality Label

25

Page 26: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

Inspect the Position of the Ferrule

October 24, 2016

Confidentiality Label

26

Page 27: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

Tools

October 24, 2016

Confidentiality Label

27

Page 28: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

Tightening Fittings into a Column

October 24, 2016

Confidentiality Label

28

Page 29: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

Overtightened Fittings

October 24, 2016

Confidentiality Label

29

Page 30: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

Overtightened Fittings

October 24, 2016

Confidentiality Label

30

Page 31: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

A-Line Fitting: Unique Spring-loaded Design

October 24, 2016

Confidentiality Label

31

• The unique spring-loaded design applies a constant force to eliminate dead volume

• Fingertight to 1300 bar

Page 32: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

A-Line Quick Turn Fitting

A-Line Quick Connect Fitting

A-Line Quick Connect Fitting

10/24/2016

• Poor chromatography results • Broad or tailing peaks • Loss of resolution

• Expensive maintenance cost • Overtightening • Column damage • Leaks, added troubleshooting

Connection problems can lead to: • Downtime • High cost of operation

Confidential

Page 33: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

A-Line Fittings Last

October 24, 2016

Confidentiality Label

33

Chromatogram overlap before and after 200

reconnections

• No visible change of chromatogram after 200 reconnections.

Page 34: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

A-Line Fittings Last

October 24, 2016

Confidentiality Label

34

• Tailing factors and theoretical plates stayed constant within the experimental allowance through the reconnection procedures.

Page 35: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

UHPLC Fingertight Fitting from Another Vendor

October 24, 2016

Confidentiality Label

35

Page 36: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

Using the Quick Connect fitting

October 24, 2016

Confidentiality Label

36

Finger tighten the fitting until you feel resistance, then close the lever.

Leak tight to 1300 bar

Page 37: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

Using the Quick Turn fitting

October 24, 2016

Confidentiality Label

37

Finger tighten for 600 bar

Wrench tighten for 1300 bar

Page 38: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

Quick Connect/Turn Fittings Use Specific Capillaries

October 24, 2016

Confidentiality Label

38

A wide variety of capillary lengths and inner diameters are available to meet HPLC and UHPLC needs

Page 39: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

Zero Dead Volume Connections

October 24, 2016

Confidentiality Label

39

Page 40: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

Not Just for Agilent Columns

October 24, 2016

Confidentiality Label

40

Page 41: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

A-Line FittingsTips

October 24, 2016

Confidentiality Label

41

A ZDV union can be used to adapt to heat exchangers but….

Page 42: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

October 24, 2016

Confidentiality Label

42

A-Line Fittings in Action

Low Dispersion Heat Exchanger G1316-80022

Works great with the latest low dispersion heat exchanger for G1315B/C TCC

Page 43: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

October 24, 2016

Confidentiality Label

43

A-Line Fittings in Action

Page 44: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

Replacing Capillaries and Ferrules

October 24, 2016

Confidentiality Label

44

The ferrules can also be replaced; just use pliers to remove the ferrule and insert a new one.

If the capillary is damaged, or another dimension is needed, just use a plier to remove the capillary and insert a new one.

Page 45: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

Quick Connect

October 24, 2016

Confidentiality Label

45

A-Line Quick Connect Assemblies 5067-5957 A-Line Quick Connect Assy ST 0.12x105mm 5067-5958 A-Line Quick Connect Assy ST 0.12x150mm 5067-5959 A-Line Quick Connect Assy ST 0.12x220mm 5067-5960 A-Line Quick Connect Assy ST 0.12x280mm 5067-5961 A-Line Quick Connect Assy ST 0.075x105mm 5067-6163 A-Line Quick Connect Assy ST 0.075x150mm 5067-6164 A-Line Quick Connect Assy ST 0.075x220mm 5067-6165 A-Line Quick Connect Assy ST 0.075x280mm 5067-6166 A-Line Quick Connect Assy ST 0.17x105mm 5067-6167 A-Line Quick Connect Assy ST 0.17x150mm 5067-6168 A-Line Quick Connect Assy ST 0.17x220mm 5067-6169 A-Line Quick Connect Assy ST 0.17x280mm 5067-6602 A-Line QC/QT Assembly 0.075x105mm 5067-6210 A-Line quick connect assy ST 0.25x105mm with a female connection 5067-6195 A-Line Quick Connect Assy ST 0.12x120 5067-6196 A-Line Quick Connect Assy ST 0.17x120

Note: Each assembly is equipped with a Quick Connect fitting, a capillary and a

Swagelok fitting or with a female connection where specified

Capillaries for Quick Connect fitting 5500-1170 A-Line Capillary ST 0.12mmx280mm 5500-1171 A-Line Capillary ST 0.12mmx220mm 5500-1172 A-Line Capillary ST 0.12mmx150mm 5500-1173 A-Line Capillary ST 0.12mmx105mm 5500-1174 A-Line Capillary ST 0.075mmx105mm 5500-1175 A-Line Capillary ST 0.075mm x 150mm 5500-1176 A-Line Capillary ST 0.075mm x 220mm 5500-1177 A-Line Capillary ST 0.075mm x 250mm 5500-1178 A-Line Capillary ST 0.075mm x 280mm 5500-1179 A-Line Capillary ST 0.12mm x 400mm 5500-1180 A-Line Capillary ST 0.12mm x 500mm 5500-1181 A-Line Capillary ST 0.17mm x 105mm 5500-1182 A-Line Capillary ST 0.17mm x 150mm 5500-1183 A-Line Capillary ST 0.17mm x 220mm 5500-1230 A-Line Capillary ST 0.17mm x 280mm 5500-1231 A-Line Capillary ST 0.17mm x 500mm 5500-1247 A-Line Capillary ST 0.12x120 5500-1248 A-Line Capillary ST 0.17x120 5500-1258 A-Line capillary ST 0.25mm x 105mm with a female connection 5500-1259 A-Line capillary ST 0.25mm x 150mm 5500-1260 A-Line capillary ST 0.25mm x 400mm 5500-1289 A-Line Capillary ST 0.12x150 M4 5500-1291 A-Line Capillary ST 0.17x150 M4

Fittings 5067-5965 A-Line Quick Connect LC fitting 5043-0924 A-Line front ferrule

• Quick Connect fittings are sold as an assembly

• Comes with a Swagelok fitting for other end

• Replacement capillaries are also available

Page 46: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

Quick Turn

October 24, 2016

Confidentiality Label

46

Capillaries for Quick Turn fitting

5500-1188 Capillary ST 0.12mm x 105mm long socket 5500-1189 Capillary ST 0.12mm x 150mm long socket 5500-1190 Capillary ST 0.12mm x 200mm long socket 5500-1191 Capillary ST 0.12mm x 280mm long socket 5500-1192 Capillary ST 0.12mm x 500mm long socket 5500-1193 Capillary ST 0.17mm x 105mm long socket 5500-1194 Capillary ST 0.17mm x 150mm long socket 5500-1195 Capillary ST 0.17mm x 200mm long socket 5500-1196 Capillary ST 0.17mm x 280mm long socket 5500-1197 Capillary ST 0.17mm x 500mm long socket 5500-1198 Capillary ST 0.075mm x 105mm long socket 5500-1200 Capillary ST 0.12mm x 130mm long socket M4

5500-1205 Capillary ST 0.075mm x 500mm long socket 5500-1206 Capillary ST 0.075mm x 250mm long socket 5500-1232 Capillary ST 0.075mm x 150mm Long Socket 5500-1233 Capillary ST 0.12mm x 180mm Long Socket 5500-1234 Capillary ST 0.17mm x 180mm Long Socket 5500-1235 Capillary ST 0.17mm x 380mm Long Socket 5500-1236 Capillary ST 0.17mm x 400mm Long Socket 5500-1237 Capillary ST 0.17mm x 700mm Long Socket 5500-1243 Capillary SST 0.12x50mm long Socket 5500-1249 Capillary ST 0,12x120 SL/SL Long Socket 5500-1250 Capillary ST 0,17x120 SL/SL Long Socket 5500-1251 Capillary ST 0,12x400 SL/SL Long Socket 5500-1252 Capillary ST 0,17x400 SL/SL Long Socket 5500-1261 Capillary ST 0.25mm x 105mm long socket with a female connection

5500-1262 Capillary ST 0.25mm x 150mm long socket 5500-1263 Capillary ST 0.25mm x 400mm long socket 5500-1288 Capillary ST 0.12x150 long socket M4

5500-1290 Capillary ST 0.17x150 long socket M4

Fittings

5067-5966 A-Line Quick Turn LC Fitting

5043-0924 A-Line front ferrule

Quick Turn capillaries and fittings are sold separately

Page 47: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

A-Line Quick Turn Fitting A-Line Quick Connect Fitting

A-Line Fittings

10/24/2016

Agilent A-Line Quick Connect fittings Getting a perfect column connection, every time, by any operator.

Easy to use

No tools - Truly finger tight up to 1300 bar.

Use a Quick Connect fitting for the column inlet and a Quick Turn fitting for the column outlet

Compatible systems: Agilent portfolio - 1290, 1260, 1200, 1100 and the Waters Alliance systems

Works on Agilent, Waters, Phenomenex and Supelco Columns – Great results!

Fast column connection – make multiple column switches in seconds

Time and Cost Savings

Reliable and reusable multiple times – Capillary and Ferrule are user changeable

Eliminate leaks and avoid costly column damage

Reduce time spent troubleshooting. Avoid costly downtime resulting from bad connections

Better Chromatography

Zero dead volume connections

Application Flexible - capillary length and the internal diameter is application dependent

All typical lengths and internal diameters available for most connections

Confidential

Page 48: Making LC Connections - Agilent€¦ · LC Connections • Problems with improper connections • Source of leaks • Mistaken for chromatography issues • Making connections can

Who can use the Quick Connect Fitting?

• Any lab with a range of users, including newer chromatographers who may not have the experience to do traditional fittings well

• Any lab that is changing columns often

• Labs who are seeing variation in chromatographic results – broad or tailing peaks, and are spending time troubleshooting issues

• Any lab looking to make chromatography easier for everyone…. The lever locks the connection, so no tools are needed

October 24, 2016

Confidentiality Label

48