Magnetic ï¬eld visualisation and inductance calculation of a

12
ENSE ˜ NANZA REVISTA MEXICANA DE F ´ ISICA E 52 (1) 1–12 JUNIO 2006 Magnetic field visualisation and inductance calculation of a simple configuration surface coil at low magnetic field A.O. Rodr´ ıguez * and R. Amador ** Centro de Investigaci´ on en Imagenolog´ ıa e Instrumentaci ´ on M´ edica, Universidad Aut´ onoma Metropolitana Iztapalapa, Av. San Rafael Atlixco 186, M´ exico, D.F., 09340. M´ exico. R. Rojas Louisiana State University, Health Science Center, School of Medicine, Department of Radiology, 1542 Tulane Ave., Room 308, New Orleans, LA 70112, USA. F.A. Barrios Instituto de Neurobiolog´ ıa, UNAM-Juriquilla, Juriquilla 76230, Quer´ etaro, M´ exico. Recibido el 28 de enero de 2004; aceptado el 30 de marzo de 2005 We propose a simple method based on the quasi-static approach to visualise the magnetic field created by a simple configuration coil using the calculation for coil inductance. Faraday’s law was used to calculate the inductance value and to simulate the magnetic field produced by circular-shaped and square-shaped coils. A software tool was developed to compute the coil inductance, and display the simulated magnetic field. We compared the predicted measured values of inductance, and performed a comparison plot of area versus inductance for both coils. Three dimensional plots of simulated fields are also shown. A circular-shaped coil was built using the method reported here, and brain images were acquired. This approach proved to be useful as an aid to visualise the magnetic field generated by simple coil geometries, and, simultaneously, calculate their coil inductance values. Keywords: Inductance; RF coil; simulation; magnetic field visualisation; magnetic resonance imaging; brain. Un m´ etodo simple basado en el enfoque cuasiest´ atico se propone para visualizar el campo magn´ etico creado por una antena, a trav´ es del alculo de la inductancia de la antena. La ley de Faraday se emplea para calcular el valor de la inductancia y simular el campo producido por una antena circular y una antena cuadrada. Una herramienta software se desarroll´ o para obtener la inductancia de la antena y visualizar su campo magn´ etico. Comparamos los valores experimentales y te´ oricos de la inductancia, graficamos el ´ area de la antena versus la inductanc´ ıa para ambos geometr´ ıas. Se obtuvieron gr´ aficas tridimensionales para mostrar la simulaci´ on del campo magn´ etico. Construimos una antena circular basada en este m´ etodo para generar im´ agenes cerebrales. Este m´ etodo muestra que es posible calcular la inductancia y visualizar el campo magn´ etico de antenas RF con geometr´ ıas simples. Descriptores: Inductancia; antena RF; simulaci ´ on; visualizaci ´ on campo magn´ etico; imagenolog´ ıa por resonancia magn´ etica; cerebro. PACS: 42.30.Va; 76.60.Lz; 76.60.Pc; 87.57.-s; 87.61.-c; 87.61.Cd; 87.63.-d 1. Introduction Radio frequency surface coils have become a widespread commodity in both Magnetic Resonance (MR) imaging and MR spectroscopy since their introduction in the early 1980’s [1]. A detailed description of the principles of mag- netic resonance imaging can be found in Ref. 2. Understand- ably so; these devices are simple, inexpensive and very versa- tile, and can be used in many different shapes such as rectan- gular, square, elliptical, circular and so forth. The popularity of surface receiving coils is due to their superior spatial se- lectivity and signal-to-noise ratio (SNR), compared to those coils able to surround the entire sample. Because of its sim- plicity of construction, the circular loop coil has received the widest attention and it is commonly used in a number of ap- plications ranging from imaging the heart, brain, and joints, for example. This particular type of coil, which can produce a higher SNR for a limited region, is usually made of copper substrate and a network that is necessary to maximise power transfer from the loop to the RF amplifier. This kind of array can be modelled as a coaxial transmission line. A great deal of effort has been expended developing Mag- netic Resonance Imaging and Spectroscopy (MRIS) coils, and much of the design technique is heuristic. Although the physical principles are established and software has been de- veloped to help researchers design RF coils, these tools are mainly used for bird-cage coil. Not much attention has been given to developing the aids for the study of surface coils for MRIS. The International Society for Magnetic Resonance in Medicine has contributed to this end with some educational material mainly focused on providing practical guidelines to build both surface and volume coils [3]. Scientists and stu- dents interested in starting to study the behaviour of surface coils can be help by the graphical tools to understand some of the characteristics governing the performance of MR surface coils. It is well known that the coil inductance for various con- figurations can be easily obtained, but these expressions can- not offer information on the form of the magnetic field gen- erated by the coil. In this paper, a method based on the quasi- static approach is introduced to estimate both the inductance and the magnetic field isolines associated with the most pop-

Transcript of Magnetic ï¬eld visualisation and inductance calculation of a

Page 1: Magnetic ï¬eld visualisation and inductance calculation of a

ENSENANZA REVISTA MEXICANA DE FISICA E 52 (1) 1–12 JUNIO 2006

Magnetic field visualisation and inductance calculation of a simple configurationsurface coil at low magnetic field

A.O. Rodrıguez∗ and R. Amador∗∗

Centro de Investigacion en Imagenologıa e Instrumentacion Medica, Universidad Autonoma Metropolitana Iztapalapa,Av. San Rafael Atlixco 186, Mexico, D.F., 09340. Mexico.

R. RojasLouisiana State University, Health Science Center, School of Medicine, Department of Radiology,

1542 Tulane Ave., Room 308, New Orleans, LA 70112, USA.

F.A. BarriosInstituto de Neurobiologıa, UNAM-Juriquilla,

Juriquilla 76230, Queretaro, Mexico.

Recibido el 28 de enero de 2004; aceptado el 30 de marzo de 2005

We propose a simple method based on the quasi-static approach to visualise the magnetic field created by a simple configuration coil usingthe calculation for coil inductance. Faraday’s law was used to calculate the inductance value and to simulate the magnetic field produced bycircular-shaped and square-shaped coils. A software tool was developed to compute the coil inductance, and display the simulated magneticfield. We compared the predicted measured values of inductance, and performed a comparison plot of area versus inductance for both coils.Three dimensional plots of simulated fields are also shown. A circular-shaped coil was built using the method reported here, and brainimages were acquired. This approach proved to be useful as an aid to visualise the magnetic field generated by simple coil geometries, and,simultaneously, calculate their coil inductance values.

Keywords: Inductance; RF coil; simulation; magnetic field visualisation; magnetic resonance imaging; brain.

Un metodo simple basado en el enfoque cuasiestatico se propone para visualizar el campo magnetico creado por una antena, a traves delcalculo de la inductancia de la antena. La ley de Faraday se emplea para calcular el valor de la inductancia y simular el campo producido poruna antena circular y una antena cuadrada. Una herramienta software se desarrollo para obtener la inductancia de la antena y visualizar sucampo magnetico. Comparamos los valores experimentales y teoricos de la inductancia, graficamos elarea de la antena versus la inductancıapara ambos geometrıas. Se obtuvieron graficas tridimensionales para mostrar la simulacion del campo magnetico. Construimos una antenacircular basada en este metodo para generar imagenes cerebrales. Este metodo muestra que es posible calcular la inductancia y visualizar elcampo magnetico de antenas RF con geometrıas simples.

Descriptores: Inductancia; antena RF; simulacion; visualizacion campo magnetico; imagenologıa por resonancia magnetica; cerebro.

PACS: 42.30.Va; 76.60.Lz; 76.60.Pc; 87.57.-s; 87.61.-c; 87.61.Cd; 87.63.-d

1. Introduction

Radio frequency surface coils have become a widespreadcommodity in both Magnetic Resonance (MR) imagingand MR spectroscopy since their introduction in the early1980’s [1]. A detailed description of the principles of mag-netic resonance imaging can be found in Ref. 2. Understand-ably so; these devices are simple, inexpensive and very versa-tile, and can be used in many different shapes such as rectan-gular, square, elliptical, circular and so forth. The popularityof surface receiving coils is due to their superior spatial se-lectivity and signal-to-noise ratio (SNR), compared to thosecoils able to surround the entire sample. Because of its sim-plicity of construction, the circular loop coil has received thewidest attention and it is commonly used in a number of ap-plications ranging from imaging the heart, brain, and joints,for example. This particular type of coil, which can producea higherSNRfor a limited region, is usually made of coppersubstrate and a network that is necessary to maximise powertransfer from the loop to the RF amplifier. This kind of arraycan be modelled as a coaxial transmission line.

A great deal of effort has been expended developing Mag-netic Resonance Imaging and Spectroscopy (MRIS) coils,and much of the design technique is heuristic. Although thephysical principles are established and software has been de-veloped to help researchers design RF coils, these tools aremainly used for bird-cage coil. Not much attention has beengiven to developing the aids for the study of surface coils forMRIS. The International Society for Magnetic Resonance inMedicine has contributed to this end with some educationalmaterial mainly focused on providing practical guidelines tobuild both surface and volume coils [3]. Scientists and stu-dents interested in starting to study the behaviour of surfacecoils can be help by the graphical tools to understand some ofthe characteristics governing the performance of MR surfacecoils.

It is well known that the coil inductance for various con-figurations can be easily obtained, but these expressions can-not offer information on the form of the magnetic field gen-erated by the coil. In this paper, a method based on the quasi-static approach is introduced to estimate both the inductanceand the magnetic field isolines associated with the most pop-

Page 2: Magnetic ï¬eld visualisation and inductance calculation of a

2 A.O. RODRIGUEZ, R. AMADOR, R. ROJAS, AND F.A. BARRIOS

ular surface coils shapes: circular and square-shaped coils.The geometry of the magnetic field produced by a surfacecoil is clearly an important factor in the behaviour of the in-ductance and the coil performance. A software tool basedon computation of the magnetic field generated by a surfacecoil and its inductance is proposed in this article. This soft-ware programme is intended to help scholars and researchersunderstand important characteristics of a surface coil. Thecomputational tool developed is able to generate 3D plots ofthe magnetic field as well as to compute the inductance oftwo types of receive coil: circular and square-shaped coils.Bringing these two aspects together in one tool allows usto study the two important aspects of single-loop coils be-haviour. This can be of particular interest to anyone startingto develop this kind of RF coil or trying to engage in morecomplicated arrays, such as array coils [4], SENSE coils [5],SMASH coils [6], and PERES coils [7].

The article is structured as follows: a brief review of thecharacteristics and functions of a conventional surface coilis presented. Special attention is given to coil inductancesince it also calculation of the magnetic field generated by thecoil possible. Mathematical expressions for calculating theinductance are derived from the classical theory of electro-magnetism for the circular-shaped and square-shaped coils.A software aid is then developed that simultaneously obtainsthe inductance, and it is able to generate a bi-dimensional plotof the magnetic field of an RF coil with an arbitrary size forthe geometries mentioned above. To test the validity of thisapproach, experimental inductance values were obtained andcompared to those predicted with the theory. Finally, a cir-cular coil was built with the aid of this software tool and wastested on a clinical MR imager. Brain images of a healthyvolunteer are shown. This research concludes with the tech-nical remarks concerning the validity of this approach in thestudy of the properties of receive surface coils and describessome prototypes that were built using this technique.

2. Radiofrequency coils for magnetic reso-nance imaging

Radio frequency (RF) coils have two important functions inan MR system, transmission and reception of the MR signal,however, there are also coils that can operate in both modes,the so-called transceiver coil. In the transmission case, theRF coil serves as a transducer which converts RF power into atransverse rotating RF magnetic field in the imaging volume.High efficiency for this transmit mode of operation corre-sponds to the maximum magnetic field in the sample volumefor minimum RF power. For the reception mode, the RF coiland its associated preamplifier serve as a transducer whichconverts a precessing nuclear magnetisation into an electricalsignal suitable for further signal processing. High efficiencyfor this reception mode corresponds to a minimal degradationof the inherent signal-to-noise ratio (SNR) of the sample vol-ume. In this case, coils are the component of the MR imagerdirectly in contact with the object to be imaged.

A well-designed coil can be highly efficient as both atransmitter and a receiver. For MRI it is also desirable forthe excitation and reception to be spatially uniform in theimaging volume. Unfortunately spatial uniformity and high-efficiency can not both be optimised simultaneously. Increas-ing the spatial uniformity will augment the required powerand decrease theSNR. In contrast an RF transmit coil needsto store magnetic energy temporarily with minimal dissipa-tion and preferably no radiation. Although the sample mate-rial may absorb significant RF energy, only a minuscule frac-tion of it is actually absorbed by the nuclear spins. Similarly,the RF receive coil detects the rotating nuclear magnetisa-tion without extracting any significant energy from the nu-clear spins. Such transfer of energy from the spins to the RFcoil will cause a shortening of the free induction decay. View-ing the RF coil as a magnetic storage device is a key to goodchoice design. Efficiency energy storage optimises both thetransmit and receive performances of a coil assembly. Reso-nant circuits are a natural choice for magnetic energy storage.

Many different coils have been developed and, accordingto their shape can be categorised into main groups. The firstgroup is called volume coils, which include Helmholtz coils,saddle coils, and high-pass and low-pass birdcage coils. Thesecond group is called the surface coil, which includes single-loop and multiple-loop coils of various shapes. These coilsare usually much smaller than the volume coils and, hence,have a higherSNRbecause they receive noises only fromnearby regions. However, they have a relatively poor mag-netic field uniformity and these are mainly used as receivecoils. Fig. 1 shows some schematics and photos of RF coilscommonly used for MRI.

RF coils are relatively complicated structures. It is dif-ficult and time consuming to analyse them exactly based onthe Maxwell’s equations. However, for RF coils whose sizeis a small fraction of a wavelength, we can use the so-calledequivalent circuit method to analyse them. The basic princi-ple of this method is, first to establish an equivalent lumped-circuit for the coil by modeling a conducting wire or stripas an inductor, then to analyse the equivalent circuit usingthe well-known Kirchhoff’s laws and, finally, to calculate themagnetic field using Biot-Savart’s law. This method is highlyefficient, reasonably accurate, and thus very practical for thedesign of RF coils. Some equivalent circuits of RF coils areshown in Fig. 2.

2.1. Radiofrequency subsystem of an mr imager

The function of this part of the apparatus is to apply RF mag-netic field pulses to the appropriate region of interest, and toreceive the weak MR signal emanating from the region ofinterest. The RF electronics of an MR system are designedto generate the pulsed RF output required to excite magneticresonance and to prepare the signals picked up by the coil forthe image processor. The signal paths are illustrated in Fig. 3.

Rev. Mex. Fıs. E52 (1) (2006) 1–12

Page 3: Magnetic ï¬eld visualisation and inductance calculation of a

MAGNETIC FIELD VISUALISATION AND INDUCTANCE CALCULATION OF A SIMPLE CONFIGURATION SURFACE COIL AT. . . 3

FIGURE 1. Schematics and photos of some RF coils used in vari-ous magnetic resonance imaging applications. a) both surface andvolume coils showing their use in practice. Surface coils: b) Pla-nar magnetron coil with 8 cavities [8], c) PERES coil with circularenvelope [9], d) Folded Crossed-Ellipse Coil[10], e) phased-arraycoils [4], Ampere coils for hand and arm imaging [11]. Volumecoils: popular birdcage coils.

FIGURE 2. Equivalent circuits of conventional surface coils: a)generic equivalent circuit, and b) equivalent circuit of a circular-shaped coil. The inductive reactancej2πfL cancels the capacitivereactance(j2πfC)−1 , and hence the resonant frequency becomesEq. (4).

RF coils are a very important part of the RF subsystem sincethey are responsible for receiving and transmitting the mag-netic resonance signals.

FIGURE 3. Diagram of the RF subsystem of a magnetic resonanceimager. The received and transmited paths are illustrated.

2.2. RF coils and the signal-to-noise ratio

While coils are, to a degree, frequency discriminating, theydo not discriminate between currents induced by the trans-verse magnetisation representing tissue MR behaviour (sig-nal), and the portion of black-body radiation generated bythe motion of molecules (thermal noise) that happens to bealigned in the transverse plane at the MR frequency. Thisnoise degrades image quality in all situations. The slice se-lective [2] approach to MRI causes signal to be generatedonly by part of the tissue in the sensitive volume of the coilbut noise comes from all of the coil volume. Therefore, itis best to select a coil with a sensitive volume that matchesthe target volume (useful field of view in 3 dimensions) asclosely as possible. As long as the target volume is includedin the coil sensitive volume, the smaller the coil the highertheSNR. Typical MRI imagers are available with body, head,and extremity coils and a range of anatomy specific surfacecoils, although the range and quality of coils varies betweenmanufacturers and MR systems (see Fig. 1). The MR radiog-raphers must choose the most appropriate coils for examina-tion to ensure optimum results. In day-to-day operation, theselection of an appropriately sized coil is the primary means

Rev. Mex. Fıs. E52 (1) (2006) 1–12

Page 4: Magnetic ï¬eld visualisation and inductance calculation of a

4 A.O. RODRIGUEZ, R. AMADOR, R. ROJAS, AND F.A. BARRIOS

of assuring the best image quality. A widely accepted formto measure the performance of an RF coil is theSNR.

An MRI experiment involves the combination of two im-portant parameters, signal and noise. This can be referred astheSNR. This parameter is an accepted standard for measure-ment of quality in MRI studies. The MR signal is determinedby the magnetic field generated by the coil. Noise alwaysaccompanies a signal that is not solely due to the coil itself,but the human body too. The biological sample (patient) pro-duces a resistance that is comparatively large, since it is apoor conductor, and contributes significantly in the noise fig-ure induced to the coil. To include the biological sample re-sistance, theSNRexpression is [12-13]:

SNR =Peak signalRMS noise

(1)

In particular, the quasi-staticSNRof a circular-shapedcoil as a function of a given depthzp is [14]:

SNRcircle =MV

√a√

(a2 + z2p)3

√3

16kT∆fσ(2)

whereM is the magnetisation density,V is the voxel,k is theBoltzmann constant,∆f is the receiver low-pass filter,T isthe temperature of the loss resistance,σ is the conductivityof the half-space,a is the coil radius, andzp is the depth ofthe observation point. Similarly, the squared-shaped coilSNRis [14],

SNRsquare=MV

√l√(

l2

2+ z2

p

)(l4

2+ z2

p

)√16πkT∆fσ

(1−√2

3 + ln(1 +√

2)) (3)

wherel is the length of one side of the square-shaped coil.Eqs. (2) and (3) are based on the model of a lossy half-spacein the quasi static case. The coils are assumed to have no con-ductive loss and the noise is dominated by the sample. It isworth mentioning that the model of a conductive half-spaceused here is adequate for cases in which the coils are smallcompared to the sample. Equivalent expressions can be foundfor volume coils like the traditional birdcage coil.

2.3. Equivalent circuit of a conventional surface coil

Surface coils are generally operated as resonant circuits.Therfore an equivalent circuit can be formulated to study theparameters affecting the design of an RF coil [15]. In a reso-nant circuit, equal amounts of energy are stored in the electricand magnetic fields, with the energy exchanging repeatedlybetween the electric and magnetic fields. Electric fields withlossy dielectrics (patients) lead to increased resistance in thecoil, the electric fields should be associated with the capac-itors used for tuning and matching the coil rather than thestray electric field of the coil. Surface coils can be regardedas a resonant circuit, being composed of a resistanceR, andinductanceL, and a capacitanceC. These circuits are com-monly calledRLCcircuits. The design of MRI surface coilsstrongly depends on these parameters. Consequently, usingKirchhoff’s law for circuits in Fig. 2b, it can be obtained thatthe resonant frequency,f is:

f =1

2π√

LC(4)

The simulation of an equivalent circuit to calculate thekey parameters of a MRI coil provides us with a way to save

a considerable amount of time and effort. Fig. 2 shows someequivalent circuits for RF coils.

2.4. Quality factor of an RF coil

To provide a quantitative way to measure the quality of thecircuit, a quality factor,Q can be defined as the energy storeddivided by the energy dissipated per period. TheQ of a res-onator can be expressed as [15]:

Q =fL

Rcoil(5)

From Eq. (5), it can be said that a highQ should havea small resistance. The higher theQ, the higher the ratio offlux density produced to power dissipated in the coil. Typicalvalues for loaded coils range from the order of magnitude of10 to that of 100. The relative values ofRcoil andRsample

can be determined by measuring theQ when the coil is emptyand when it is loaded by a patient or a phantom (water-filledbottle). The best indicator of the coil sensitivity is the ratio:

Qempty

Qloaded=

Rcoil + Rsample

Rcoil(6)

provided the dielectric losses do not contribute toRsample

(patient). This ratio can be 5 times or more for a magneticfield strength of 1.5 Tesla; thus coil losses contribute lessthan 11% of the observed noise value. Therefore, a goodcoil design should have aRsample À Rcoil. A much simplerapproach is to calculateQ according to the following expres-sion [16]:

Q =f

∆f(7)

Rev. Mex. Fıs. E52 (1) (2006) 1–12

Page 5: Magnetic ï¬eld visualisation and inductance calculation of a

MAGNETIC FIELD VISUALISATION AND INDUCTANCE CALCULATION OF A SIMPLE CONFIGURATION SURFACE COIL AT. . . 5

where∆f denotes the bandwidth, which can be measuredeasily with the aid of a network analyzer. The measuredQfactor includes the contribution of the capacitor’s resistance,and one must be aware that many capacitor manufacturersspecify theQ of their products. High quality chip capacitorsare thus recommended to be used in the design of an MRIcoil, to avoid unwanted resistance contributions coming fromthese electronic components which can drastically affect thequality factor.

3. Inductance calculation of a surface coil

The resistance and inductance depend on the shape of thecoil. Of particular interest is the inductance, because it allowsus to study the energy stored in a coil and its relation to themagnetic field. The physical inductance can be studied usingthe magnetic flux model as in classical electromagnetism. Itis well known that inductanceL is defined as

L =1i

s

B · dS, (8)

whereB is the magnetic field in the cross-section of areaSandi is current. Eq. (8) shows that the magnetic field in thisareaS is proportional to the current and inductance. It canalso be proved that, for a constant current, there is a direct re-lationship between the magnetic field of the surface coil andits inductance. This is an important aspect to be considered indeveloping surface coils for MR imaging and spectroscopy.This inductance formula (Eq. (8) provides a way to actuallymap the pattern of the dependence of spatial coordinates onthe magnetic field generated by the surface coil. The mea-

surement ofinductance and capacitance of an RF coil is animportant aspect when constructing coils dedicated to MRIS.In the case of simple geometries such as the circular andsquared-shaped coils, several equations to estimate the induc-tance of the coils have been proposed. Bueno and cowork-ers [17] have shown that all these mathematical expressionsof inductance are equivalent. In additon other methods havebeen introduced that are both analytical [18-19] and numer-ical [20] to compute the inductance of a surface coil. Theinductance of circular and square coils was obtained usingEq. (8), where the corresponding magnetic field was first cal-culated using the Biot-Savart law.

A position-dependent function is required to map, pointby point, the magnetic field generated by a surface coil. Thisapproach can give a spatial resolution to the magnetic fieldfor a particular shape of surface coil. Eq. (8) together withthe calculation of the magnetic field of a circular-shaped coil(see Appendix), an inductance formula can be derived:

Lcir =Rµ0

2

R∫

0

2π∫

0

(R− x cos(ω))(R2 + x2 − 2Rx cos(ω))3/2

dω dx (9)

whereR is the coil radius,x is the distance between the coilcentre and an arbitrary point,ω is the angle formed by the ra-diusR and the distancex, andµ0 is the permeability. Eq. (9)shows that coil inductance can be obtained via the calcula-tion of the coil magnetic field, which is a function of the coilradius and positionx.

The analytical expression to compute the inductance of asquare-shaped coil can similarly be found as in Eq. (9). Theinductance formula is:

Lsqr =µ0

x∫

0

y∫

0

λ

x

l − y√

(l − y)2 + x2

+y√

y2 + x2

+

λ

l − y

l − x√

(l − x)2 + (l − y)2+

x√(l − y)2 + x2

dxdy

+µ0

x∫

0

y∫

0

λ

l − x

y√

(l − x)2 + x2

+l − y√

(l − y)2 + (l − x)2

+

λ

y

[x√

x2 + y2+

l − x

y2 + (l − x)2

]dxdy (10)

where l is the square size. The magnetic field expression[Eq. (10)] on a square coil is dependent of the coil size(length of a side) and the position (x,y). Unlike the majorityof textbooks on electromagnetic theory (21-26), these induc-tance formulae above depend on the magnetic fields whichare functions depending on the spatial coordinates (x,y) andthe actual size of the coil. This approach provides a schemeto point-to-point map the magnetic field generated by the sur-face coil and calculate the inductance value for that particularconfiguration. It is necessary to find primitive functions forEqs. (9) and (10). This can be a very complicated task, sinceno primitive functions exist to compute these integrals.

Hence, a possible solution can be found by using a numericalapproach such as Simpson’s rule.

4. Magnetic field simulation and inductancecalculation

To compute the inductance of a surface coil with the formu-lae above [Eqs. (9) and (10)], it is necessary to first calculatethe point-by-point magnetic field inside the RF coil. Themagnetic field of the circular and square-shaped coils were

Rev. Mex. Fıs. E52 (1) (2006) 1–12

Page 6: Magnetic ï¬eld visualisation and inductance calculation of a

6 A.O. RODRIGUEZ, R. AMADOR, R. ROJAS, AND F.A. BARRIOS

numerically computed. To obtain the point-by-point mag-netic field within a circle as shown in Fig. 4a, the coil wascentered at the origin of a coordinate system, and then con-centric circles were then formed by increasingx until the en-tire area is covered (Fig. 4b,Swas fully covered, x=R). Themagnetic field generated by the square-shaped coil was car-ried out by selecting adxdyto form a differential of area in-side the square depicted in Fig. 5a. This procedure was againrepeated until the whole of the area was covered (Fig. 5b).In both cases, each different contribution represents the mag-netic field at each point within the area formed by the coilgeometry; therefore, the summation of all of them gives thetotal magnetic field inside the coil. Since the numerical so-lutions for the magnetic fields [Eqs. (A.7) and (A.20)] arehighly-computationally demanding, all the numerical calcu-lations of the magnetic field were done in Visual Basic (V.6,Microsoft Co., USA) with a 500 MHz Intel Pentium PC. Anumerical simulation of the field can be easily displayed us-ing these spatial data. To easily visualise the simulations ofthe magnetic fields, the data were imported into the spread

sheet MS Excel (Microsoft Office 2000) for plotting and fileswere created. A software tool was then developed to puttogether the output of the inductance calculation process andthe visualisation of the magnetic field.

5. Software tool

To facilitate the study of the behaviour of the magnetic fieldand the inductance of these two surface coils, a software toolwas developed. This programme can be used to calculatethe inductance value of both circular and square-shaped coilswhich can be used as a guideline to built a coil prototype, andto illustrate the pattern generated by the magnetic field of thesurface coil. Similarly, computer programme were speciallywritten in Visual Basic to compute the integrals in Eq. (9)and (10) using Simpson’s rule. The Excel application wasalso used to display the magnetic field in the form of three-dimensional plots. In Fig. 6, some plots produced by the pro-gramme are shown.

FIGURE 4. Calculation of the magnetic field of a circular coil. a) A vector diagram for an arbitrary point P which is selected inside thearea of circle, then the distance (r) from the point to a current element is measured. In order to obtain the contribution of this point, thedistance to the centre (x) and the circle radius (R) are obtained too. The sine and cosine rules can be used to yield an expression to calculate r.b) Magnetic fieldBcir(x, y) in a differential area ofdxdywithin the surfaceS (circle). All the contributions in the circle inside the area (S)should be considered until the entire area is scanned.

FIGURE 5. Calculation of the magnetic field of square-shaped coil. a) a side view of the square showing the vector diagram of the magneticfield applied on the pointP. b) Schematic showing the integration limits for the four sides of the square-shaped coil. c) Magnetic fieldBsqrt(x, y) in a differential area ofdxdywithin the surfaceS(square).

Rev. Mex. Fıs. E52 (1) (2006) 1–12

Page 7: Magnetic ï¬eld visualisation and inductance calculation of a

MAGNETIC FIELD VISUALISATION AND INDUCTANCE CALCULATION OF A SIMPLE CONFIGURATION SURFACE COIL AT. . . 7

FIGURE 6. Some windows of the software tool to calculate the inductance of the surface coil are shown. a) starting page, b) coil selectionpage, c) mapping of magnetic field generated by a circular-shaped coil, and d) magnetic-field isolines of square-shaped coil.

6. Experimental inductance

In order to compare these theoretical results of both circularand square-shaped coils against experimental inductance val-ues, coil prototypes of different sizes were built. Hence, res-onant circuits were constructed by attaching a non-magneticcapacitor (American Technical Ceramics: series ATC 100B,non-magnetic) in series to each coil: the capacitor valueswere the same for all cases. A 50Ω coaxial cable wasalso attached to the coil to conduct the signal. The induc-tance of a coil can be calculated by measuring its resonantfrequency with a network analyzer (Advantest R3753 AH,Tokyo, Japan) provided the capacitor value is known. Sinceall measurements were done with a network analyzer, lossreturn plots were generated and the resonant frequency wasmeasured for each coil. To finally obtain the coil inductance,the following formula was used:

Lexp =1

4π2f2C(11)

whereC is the capacitance andf the resonant frequency ofthe coil. Finally, to test the validity of this method, com-parison plots of calculated and experimentally measured datafor the two geometries were computed and shown in Fig. 7a

and 7b. The inductance plots show very good agreement be-tween the inductance values obtained with the simulation andthe experimentally-obtained inductance values. To comparethe inductance of both coils, a plot of area versus inductancein each case was computed and shown in Fig. 8.

7. Development of a surface coil prototype

To use an RF coil for MRI applications, it is necessary to tuneit to the specific resonant frequency and match it to 50Ω.The resonant frequency is determined by the species to beused, for example, 1H, 13C, 19F, 23N and 31P, which have aspecific resonant frequency depending on the magnetic fieldstrength [2]:

f =γ

2πB0 (12)

where the gyromagnetic ratio isγ, B0 is the uniform mag-netic field applied, andf is the Larmor frequency of the nu-cleus. In our case, we take the proton gyromagnetic ratio,42.58 MHz/T. The 1H species is the most significant nucleusfor most MRI studies because of this natural concentration inthe human body as part of the water molecule and its high

Rev. Mex. Fıs. E52 (1) (2006) 1–12

Page 8: Magnetic ï¬eld visualisation and inductance calculation of a

8 A.O. RODRIGUEZ, R. AMADOR, R. ROJAS, AND F.A. BARRIOS

FIGURE 7. Comparison of the theoretical and experimental valuesof inductance (a) circular coil for radii: 4.5, 5, 6.5, 7.5 and 8.5 cm.(b) square-shaped coil for sides: 4, 6, 10, and 16 cm.

FIGURE 8. Comparison plot of inductance values of both coils asa function of the area. It can be appreciated that inductance for thetow coils is approximately the same.

NMR sensitivity. The high sensitivity of the 1H species canbe intuitively understood by considering the fact that thereis typically only one electron shielding the magnetic momentthat the nucleus produces. This lack of shielding is even morepronounced for the hydrogen in water, where the nature ofthe O-H bond is covalent. This tends to remove the shieldingelectron from the 1H. The effect of this is to make the nu-cleus’ state more easily permutable from external influences.

7.1. Resonant frequency tuning

A circular-shaped coil was then built with the aid of thissoftware to resonate at 64 MHz. This is the resonant fre-quency of single protons when a 1.5 Tesla magnetic field isapplied. Hence, the theoretical inductance of a 1 cm-widestrip coil with coil radius of 15 cm is 408 nH. From this re-sult and Eq. (11), a chip capacitor (American Technical Ce-ramics: series ATC 100B non-magnetic) with a capacitancevalue 15.21 pF was attached in series to the coil to producethe desired resonant frequency. The circular-shaped coil wasmounted and fixed on a rigid surface to facilitate tuning andmatching of the coil prototype. Resonant frequency was mea-sured with a network analyzer as indicated in Sec. 6. Fig. 9ashows the resonant frequency of the coil prototype.

FIGURE 9. Coil characterisation: a) resonant frequency tuning andits bandwidth measured at 3 dB, and b) Smith chart showing 50Ωmatching. Quality factor of coil prototype was computed usingEq. (7), soQ = 31.4.

Rev. Mex. Fıs. E52 (1) (2006) 1–12

Page 9: Magnetic ï¬eld visualisation and inductance calculation of a

MAGNETIC FIELD VISUALISATION AND INDUCTANCE CALCULATION OF A SIMPLE CONFIGURATION SURFACE COIL AT. . . 9

FIGURE 10. Coronal brain images obtained with a circular coilprototype designed with the aid of the software tool. Images wereacquired with the following parameters: TE/TR = 171/3700 ms,FOV = 75 cm2, slice thickness = 4 mm, image size= 512× 512.

7.2. 50ΩΩΩ matching

The frequencies employed in MRI are in the same range asthose used for radio transmission, so to shield the signalscoming from a region of interest, a coaxial cable is normallyused. The type of cable imposes a constraint, the so-calledcharacteristic impedanceZ and only if, at the end of the ca-ble, there is a termination of resistanceR=Z0. If R differsfrom the valueZ0, the power is reflected. Similary, in gen-eral, if we measure the impedance between the conductorsof a cable terminated with an impedanceZ (i.e. Z is at theopposite end from our measurement), we will not record avalueZ. Only if Z=R=Z0 will that value be recorded, the ca-ble otherwise acts as an impedance transformer. Therefore,coils are usually matched to 50Ω to assure maximum energytransfer from the sample to the MR imager. The coil is con-nected to a network analyzer via a 50Ω coaxial cable to forma coaxial transmission line and to measure the resonant fre-quency and the 50Ω matching. Fig. 9b shows the resonantfrequency of the coil prototype. The 50Ω-coaxial transmis-sion line (quarter wavelength) was also used to carry the MRsignal to the preamplifier of the RF subsystem (see Fig. 3).The coil prototype was tested on a 1.5 Tesla Signa LX imager(General Electric Medical System, Milwaukee, WI, USA) to-gether with standard clinical sequences, and brain images ofa healthy volunteer were acquired. Fig. 10 shows brain coro-nal images obtained with a circular-shaped coil designed andbuilt with this method.

8. DiscussionA simple approach was used to calculate the inductance ofcircular-shaped and squared-shaped coils for MRI. Analyti-

cal inductance expressions for both coils were derived andtheir solutions were found via numerical solutions. These so-lutions offered not only a way to compute their correspondinginductance values as a function of the coil size, but also a wayto map the magnetic field generated by each coil configura-tion. Because of the highly computational nature of the solu-tion, a software tool to automatically compute the inductanceof both configurations was developed using widely availablesoftware.

Figure 6 shows the magnetic field associated to the cor-responding coil configuration and the inductance value for aparticular size (r, radius orl , square size). The encouragingresults in Fig. 6 indicate that this software tool might be ofsome aid in obtain a preliminary value of the inductance ofthe coil, as well as in providing an idea as to how the proto-type performs by plotting the magnetic field. The visualisa-tion of the magnetic field can also facilitate the understand-ing of the uniformity patterns of the coils. These magneticfield plots are in very good agreement with those reported bySobol in 1986 [26]. The magnetic field of a square-shapedcoil generates high peaks of the magnetic field at the cor-ners, see Fig. 6d. From Fig 6c & 6d, it can be seen that themagnetic field patterns of the circular coil and squared-coilshow important differences. These contribute negatively tothe image quality in the form of undesired artefacts. This canusually appear as image hyperintensities that mask out therelevant information. However, these defects can be easilydiminished by chopping the corners off to get a more roundedshape.

Figure 7 shows that numerically-acquired inductanceshows a very good agreement with the experimental induc-tance obtained with a network analyzer. It can be observedfrom Fig. 7 that calculation of coil inductance with the soft-ware tool developed here can serve as a guideline to calculatethe coil size too. Coil size is an important issue in MRI, sincecoils having big radii tend to capture higher noise levels com-ing from bigger regions of interest; this effect degrades theimage quality. It is then suggested small coils to be able togenerate highSNRimages. This approach is mainly used inphased-array coils, an example of this type of coil is shownin Fig. 1e. Knowing the inductance of a coil before buildinga coil prototype can save a considerable amount of time andeffort. Figure 8 shows that the coil inductance for a similararea is approximately the same. This implies that both coilconfigurations can be used indistinctly. However, an advan-tage of the square coil over the circular coil is that the fielduniformity of the former is slightly better than the latter one.Tuning, matching, and quality factor of coil prototype com-pare very well with those reported in the literature [27].

Since Microsoft products have been used to develop thissoftware tool, a number of restrictions are applied by thecompany. Potential users must purchase the correspondinglicenses to use Visual Basics and Excell programmes. Theaccess to a wide number of potential users is drastically re-duced due to license restrictions. Java language can be a goodalternative in order to avoid these restrictions, because it can

Rev. Mex. Fıs. E52 (1) (2006) 1–12

Page 10: Magnetic ï¬eld visualisation and inductance calculation of a

10 A.O. RODRIGUEZ, R. AMADOR, R. ROJAS, AND F.A. BARRIOS

be freely obtained and it is compatible with almost any com-puter platform available on the market. In addition, a numberof potential users can gain access to this software tool on theInternet.

We have proved that a method to simultaneously calcu-late the inductance of a surface coil and simulate its magneticfield can be elaborated for simple coil configurations. Thecombination of these physical parameters can serve as a toolin designing coil with complex geometry. It is also possible tovisualise the behaviour of the coil magnetic fields via the cal-culation of the coil inductance. These coils designed with thismethod did not require any further adjustments before oper-ating on a clinical imager. This computational programmeproved to be useful in building a coil prototype with a sim-ple geometry for a commercial MR imager. It is also com-patible with standard imaging sequences to generate qualitybrain images as shown in Fig. 10.

9. Conclusions

A software tool to compute the inductance of an RF coil aswell as the visualisation of its magnetic field has been devel-oped for simple RF coil configurations. The computationalmethod proposed in this work offers an alternative approachto the trial-and-error method widely used in the developmentof RF coils for MRI. It has been proved that widely-availablecommercial software together with a first principle approachcan be used to build a computational tool to assist researchersand scholars in the development of RF coils for MRI.

Acknowledgements

We would like to thank Dr. Julian Sanchez-Cortazar, for-mer Medical Director of American British Cowdray MedicalCenter-Mexico for the support needed to carry out this work.R. A. is gratefull to National Council of Science and Tech-nology of Mexico for a graduate scholarship.

A Appendix: Calculation of surface-coil mag-netic field

A1. Circular coil

According to the Biot-Savart law in Fig. 1a, the magneticfield dB produced atP per current element time the changein magnetic field in lengthdl for a circular loop can be ex-pressed as:

dBcir =µ0

idl sin (θ)r2

(A.1)

wherer goes from current element,idl to the pointP, andθis the angle betweenr current element directions. If we takeR = R1 and apply the cosine rule, we obtain:

r2 = R + r2 − 2Rx cos(ω) (A.2)

x2 = R + r2 − 2Rx cos(ω) (A.3)

Also, if l = Rω thendl = R dω and taking

λ =µ0i

4π(A.4)

and sin(θ) = cos(α) (θ = 900 +α) and replacing Eq. (A.3) inEq. (A.1), the rate of magnetic field is:

dBcir = λdl cos (α)

r2(A.5)

Substituting Eqs. (A.2.a and b) in Eq. (A.6) and consid-ering thatdl = Rdω, the magnetic field,dBcan be rewritten,

dBcir = RλR− x cos(ω)

(R2 + x2 − 2Rr cos(ω))32dω (A.6)

Integrating Eq. (A.5), the magnetic field,B(x) of a circularcoil as shown in Fig. 1a is

dBcir =Rµ0i

2π∫

0

R− x cos (ω)

(R2 + x2 − 2Rr cos(ω))32dω (A.7)

An expression for the magnetic flux can be calculatedfrom Eq. (2)

φcir=∫

S

Rµ0i

2π∫

0

R−x cos(ω)

(R2+x2−2Rr cos(ω))32dω

dS (A.8)

wheredS= πre2- πr = π(e+1) (e-1) dr and

dx = r(e− 1), x ≈ re + 1

2, dS = 2π (A.9)

therefore inductance is:

Lcir=Rµ0

R∫

0

2π∫

0

(R−x cos(ω)) x

(R2+x2−2Rr cos(ω))32dωdx (A.10)

A2. Square-shaped coil

The inductance of a square-shaped coil can be calculated in asimilar way to the case of a circular loop. It is only necessaryto obtain an expression for the magnetic field,B generatedby a square loop, Fig 5a shows a square-shaped coil carryinga currentI and where

sin(θ) = cos(β) (A.11)

Hence, the magnetic fieldB of a square-shaped coil forpoints inside the coil is:

dBcir =λ cos3 (θ)

x2(A.12)

additionally,

dl = sec2(β)dβ (A.13)

replacing Eq. (A.13) in (A.12), we obtained:

dBsqr =x cos3 (β) sec2(β)

x2=

λ cos (β) dβ

x(A.14)

Rev. Mex. Fıs. E52 (1) (2006) 1–12

Page 11: Magnetic ï¬eld visualisation and inductance calculation of a

MAGNETIC FIELD VISUALISATION AND INDUCTANCE CALCULATION OF A SIMPLE CONFIGURATION SURFACE COIL AT. . . 11

(A.14) Integrating Eq. (A.13), an expression for the magnetic field of a square coil can be found:

B(x, y)sqr =

l−y∫

−y

λ cos (β) dβ

x=

∣∣∣∣λ

xsin(β)

∣∣∣∣l−y

−y

x

∣∣∣∣1√

x2 + l2

∣∣∣∣l−y

−y

(15)

Equation (A.15) represents the magnetic field at a pointP(x,y) due to a current I along a segment of lengthl. In thisparticular geometry, it is necessary to consider all four sides of the square (Fig. 5b). The superposition theorem is used tocalculate the total magnetic field at a particular point.

Equation (A.15) can be rewritten to include the four sides:

B(x, y)sqr =λ

x

∣∣∣∣1√

x2 + l2

∣∣∣∣l−y

−y

l − y

∣∣∣∣1√

x2 + l2

∣∣∣∣−x

l−x

l − x

∣∣∣∣1√

x2 + l2

∣∣∣∣y

y−l

y

∣∣∣∣1√

x2 + l2

∣∣∣∣x

x−l

(A.16)

(A.16) Developing Eq. (A.16), it can be

B(x, y)sqr =λ

x

[l − y√

(l − y)2 + l2+

y√x2 + y2

]+

λ

l − y

[l − x√

(l − x)2 + (l − y)2+

x√x2 + (l − y)2

]

l − x

[y√

y2 + (l − x)2+

l − y√(l − x)2 + (l − y)2

]+

λ

y

[x√

x2 + y2+

l − x√y2 + (l − x)2

](A.17)

Prior to obtaining an expression for the inductance of square-shaped coil, a formula for the magnetic flux should be found,then from Eq. (2) the magnetic flux is:

B(x, y)sqr =∫

S

x

[l − y√

(l − y)2 + l2+

y√x2 + y2

]+

λ

l − y

[l − x√

(l − x)2 + (l − y)2+

x√x2 + (l − y)2

])• dS

+∫

S

l − x

[y√

y2 + (l − x)2+

l − y√(l − x)2 + (l − y)2

]+

λ

y

[x√

x2 + y2+

l − x√y2 + (l − x)2

])• dS (A.18)

To solve Eq. (B.8) it is necessary to obtain the integral limits for each side of the square in Fig. 5b and 5c, so that itbecomes,

φsqr =µ0i

x∫

0

y∫

0

x

[l − y√

(l − y)2 + l2+

y√x2 + y2

]+

λ

l − y

[l − x√

(l − x)2 + (l − y)2+

x√x2 + (l − y)2

])dydx

+µ0i

x∫

0

y∫

0

l − x

[y√

y2 + (l − x)2+

l − y√(l − x)2 + (l − y)2

]+

λ

y

[x√

x2 + y2+

l − x√y2 + (l − x)2

])dydx (A.19)

Finally, the inductance formula is

Lsqr =µ0

x∫

0

y∫

0

x

[l − y√

(l − y)2 + l2+

y√x2 + y2

]+

λ

l − y

[l − x√

(l − x)2 + (l − y)2+

x√x2 + (l − y)2

])dydx

+µ0

x∫

0

y∫

0

l − x

[y√

y2 + (l − x)2+

l − y√(l − x)2 + (l − y)2

]+

λ

y

[x√

x2 + y2+

l − x√y2 + (l − x)2

])dydx (A.20)

Rev. Mex. Fıs. E52 (1) (2006) 1–12

Page 12: Magnetic ï¬eld visualisation and inductance calculation of a

12 A.O. RODRIGUEZ, R. AMADOR, R. ROJAS, AND F.A. BARRIOS

∗. Corresponding author: Alfredo O. Rodrıguez, Centro de In-vestigacion en Imagenologıa e Instrumentacion Medica, Uni-versidad Autonoma Metropolitana Iztapalapa, Av. San RafaelAtlixco 186, Mexico, D. F., 09340. Mexico. TelephoneNo.: (5255) 85024569, Fax No.: (5255) 804-4631, e-mail:[email protected].

∗∗. Present Address: Institute of Imaging Science, Vanderbilt Uni-versity, Nashville TN 37232-2675, USA.

1. J.J.H. Ackerman, T.H. Grove, G.G. Wong, D.G. Gadian, andG.K. Radda,Nature283(1980) 167.

2. A.O. Rodriguez,Rev. Mex. Fıs.50 (2004) 272.

3. The International Society for Magnetic Resonance in Medicine,RF Bootcamp: Essential of RF coil design, construction andinterface.(Knowledge Access Publishing, Connecticut, 1999),CD ROM.

4. P.B. Roemer, W.A. Edelstein, C.E. Hayes, S.P. Souza, and O.M.Muller, Magn. Reson. Med.16 (1990) 192.

5. K.P. Pruessmann, M. Weiger, M.B. Scheidegger, and P. Boe-siger,Mag. Reson. Med.42 (1999) 952.

6. D.K. Sodickson and W.J. Manning,Mag. Reson. Med.38(1998) 591.

7. S. Hidalgoet al., Proc. 9th Ann. Meet. of ISMRM, 112 (2001).

8. M.A. Lopez Terrones,Antena superficial tipo magnetron planode 8 cavidades para Espectroscopıa por Resonancia Magneticaa 1.5 T, 3 T y 7 T, Master’s Thesis, Biomedical EngineeringGraduate Programme, Universidad Autonoma MetropolitanaIztapalapa, 2004, Mexico.

9. S.E. Solıs Najera, Antena superficial PERES con envolventecircular para IRM, Master’s Thesis, Biomedical EngineeringGraduate Programme, Universidad Autonoma MetropolitanaIztapalapa, 2004, Mexico.

10. R. Hernandez, A.O. Rodriguez, J. Sanchez Cortazar, P. Sal-gado, and F. A. Barrios, Folded Crossed-Ellipse Coil for HandMRI. Proc. 25th Ann. Int. Con. IEEE Eng. Med. Biol. Cancun,Mexico, (2003) 497.

11. O. Marrufo,Jornadas del Posgrado de Ciencias Basicas, UAMIztapalapa, (2004).

12. D.I. Hoult, Prog. NMR Spect.,12 (1978) 41.

13. D.I. Hoult and R.E. Richardson,J. Magn. Reson.24 (1976) 71.

14. J. Wang, A. Reykowski, and J. Dickas,IEEE Trans. Biom. Eng.42 (1995) 908.

15. J. Jin,Electromagnetic Analysis and Design(CRC Press, BocaRaton, 1999) 137.

16. J.D. Jackson,Classical Electrodynamics(John Wiley, NewYork, 1999), 3rd Edition, 215-223.

17. M. Bueno and A.K.T. Assis,Can. J. Phys.75 (1997) 357.

18. F.E. Terman,Radio engineer’s handbook(MacGraw-Hill, NewYork, 1943).

19. A. Ouldkhal, B. Bandelier, S. Kan, and F. Rious-Damidau,IEEE on Mag.31 (1995) 1590.

20. M. Gyimesi and D. Ostergaard,IEEE on Mag.35 (1999) 1119.

21. C.A. Balanis,Antenna theory: analysis and design(John Wi-ley, New York, 1997), 2nd Edition, 468.

22. R.P. Feynman, R.B. Leighton, and M. Sands,The Feynman lec-tures on physics: The electromagnetic field(Addison-Wesley,Reading, 1964), Vol. II, 1.

23. E.M. Purcell,Electricity and magnetism, (MacGraw-Hill, NewYork, 1965) Berkely Physics Course, Vol. II, 250.

24. E.C. Jordan,Electromagnetic waves and radiating systems(Prentice Hall, Englewood Cliffs, 1958) 211.

25. J.R. Reitz and F.J. Milford,Foundations of electromagnetic the-ory (Addison-Wesley, Reading, 1960) 170.

26. W.T. Sobol,Rev. Magn. Reson. Med.1 (1986) 181.

27. C.N. Chen and D.I. Hoult,Biomedical Magnetic ResonanceTechnology(Adam Hilger, Institute of Physics Publishing,Britain, 1989). pp. 117-226.

28. C.E. Hayes, W.A. Edelstein, and J.F. Schenck,Radio FrequencyCoils, in S.R. Thomas and R.L. Dixon, NMR in Medicine: TheInstrumentation and Clinical Applications (American Instituteof Physics Publishing, New York, 1986) 142.

29. E. Krestel, Imaging systems for medical diagnostics; fun-damentals and technical solutions; X-ray diagnostics, com-puted tomography, nuclear medical medicine diagnostics, mag-netic resonance imaging, sonography, biomagnetic diagnostics(Siemens-Aktienges, Berlin, 1090) 479.

30. R. Amador Baheza,Desarrollo de antenas superficiales de ra-diofrequencia para imagenologıa por resonancia magnetica,Master’s Thesis, Biomedical Engineering Graduate Pro-gramme, Universidad Autonoma Metropolitana Iztapalapa,(2001) Mexico.

Rev. Mex. Fıs. E52 (1) (2006) 1–12