Magnet Measurement Device for MICE F. Garnier , P-A. Giudici , F.Bergsma CERN

17
Magnet Measurement Device for MICE F. Garnier, P-A. Giudici, F.Bergsma CERN MICE Collaboration Meeting No. 30 6-9 July 2011 Oxford technical drawings by O.Jamet

description

Magnet Measurement Device for MICE F. Garnier , P-A. Giudici , F.Bergsma CERN MICE Collaboration Meeting No. 30 6-9 July 2011 Oxford technical drawings by O.Jamet. Layout CM29. Rails. Trolley with B-sensors. ¼ pipe of aluminum. Cradles. Service module. Tooth belt. - PowerPoint PPT Presentation

Transcript of Magnet Measurement Device for MICE F. Garnier , P-A. Giudici , F.Bergsma CERN

Page 1: Magnet Measurement Device for MICE  F.  Garnier , P-A.  Giudici ,  F.Bergsma CERN

Magnet Measurement Device for MICE

F. Garnier, P-A. Giudici, F.BergsmaCERN

MICE Collaboration Meeting No. 306-9 July 2011 Oxford

technical drawings by O.Jamet

Page 2: Magnet Measurement Device for MICE  F.  Garnier , P-A.  Giudici ,  F.Bergsma CERN

Layout CM29

Servo motor with encoder

Trolley with B-sensors

Rails

Cradles

¼ pipe ofaluminum

Tooth belt

Service module

Page 3: Magnet Measurement Device for MICE  F.  Garnier , P-A.  Giudici ,  F.Bergsma CERN

supports

supports

No support inside magnetindependent of magnet shape

Layout CM29

Revolving discstep size 5 degree

Page 4: Magnet Measurement Device for MICE  F.  Garnier , P-A.  Giudici ,  F.Bergsma CERN

5 m

2 m

400 mm

cradlesCradles need rectification

Page 5: Magnet Measurement Device for MICE  F.  Garnier , P-A.  Giudici ,  F.Bergsma CERN

sticks outstays inside 400 mmdiameter

Non-conducting materials usedagainst eddy currents

Page 6: Magnet Measurement Device for MICE  F.  Garnier , P-A.  Giudici ,  F.Bergsma CERN

sticks outnot inside400 mmdiam.

gearbox 1/12no clearance

Page 7: Magnet Measurement Device for MICE  F.  Garnier , P-A.  Giudici ,  F.Bergsma CERN

target support forsurvey

disk for spectrometersolenoid

disk for AFC

Page 8: Magnet Measurement Device for MICE  F.  Garnier , P-A.  Giudici ,  F.Bergsma CERN

removableextension plate

Page 9: Magnet Measurement Device for MICE  F.  Garnier , P-A.  Giudici ,  F.Bergsma CERN

detachableservice module

NMR test stand forB-sensors

Check on B-sensor calibration

For details on calibration see IMMW15 workshop

Page 10: Magnet Measurement Device for MICE  F.  Garnier , P-A.  Giudici ,  F.Bergsma CERN

Rotation in steps of 45 deg.over two axes

NMR

Page 11: Magnet Measurement Device for MICE  F.  Garnier , P-A.  Giudici ,  F.Bergsma CERN

5m version for spectrometer solenoids2m version for focus absorber solenoids.

Extension disk to by-pass internal flange AFCNeed to measure from both sides

2m version could be used to measure eventually coupling coils

Extension disk for AFC

dismountable

This flange cannot be removed andcannot be surpassed by rail support

Page 12: Magnet Measurement Device for MICE  F.  Garnier , P-A.  Giudici ,  F.Bergsma CERN

Mounting of B-sensors

main disk

Page 13: Magnet Measurement Device for MICE  F.  Garnier , P-A.  Giudici ,  F.Bergsma CERN

extension disk

Mounting of B-sensors

Page 14: Magnet Measurement Device for MICE  F.  Garnier , P-A.  Giudici ,  F.Bergsma CERN

Measurement grid

Disk can be rotated in steps of5 degree

180 mm sensor has to be removed for this region

/R140

Page 15: Magnet Measurement Device for MICE  F.  Garnier , P-A.  Giudici ,  F.Bergsma CERN
Page 16: Magnet Measurement Device for MICE  F.  Garnier , P-A.  Giudici ,  F.Bergsma CERN

Strategy mechanics:1 build 5 m version , test precision with laser tracker 2 if precision insufficient add encoder and rack and/or foresee laser tracker during measurements etc3 build 2 m version

DAQ:Build simple read-out for 7 B-sensorsAnticipate to have interface outside high field region

Accuracy:

< 0.5 mm longitudinal+/- 0.5mm radialTheta +/- 1 mradBx,By,Bz +/- 2 mT Check in situ with NMR on service module

very unlikely

Page 17: Magnet Measurement Device for MICE  F.  Garnier , P-A.  Giudici ,  F.Bergsma CERN

Status

Rail support send to company to rectifyWill come back this week => bench ready for final assembly

Motion: motor with gearbox tested in vertical bench with 2.5 kgf load and 7.5 kg mass. Works fine, clearance within limits

B-sensors ready and calibrated

Read-out: basics exist ( used in calibrator with 4 cards), parts exist duplicate electronics board

Ready at August 2011 for AFC (small version)

should be send to CERN !