Macs 10 - test-ibscg.com
Transcript of Macs 10 - test-ibscg.com
www.sternmed.de
Macs 10Multi parameter patient monitor
www.sternmed.de
The Macs 10 is an affordable compact portable patient monitor with essen-tial basic parameters such as ECG, RESP, temperature, SPO2 and NIBP.
Macs 10Multi parameter patient monitor
© SternMed2
FEATURES
• 7” high resolution color TFT screen
• 5 or 7-channelwaveform display with trend coexistence, switchable to observing screen
• Automatic analysis of 20 arrhythmia waveforms
• S-T segment with automatic & manual analysis
• Monitoring standard parameters of ECG, Spo2, RESP, NIBP, TEMP & PR
• Optimized modes for adult, pediatric & neonate
• DigitalSpo2 technology with anti-motion & anti-low infusion
© SternMed3
• Trend data analysis for 6/24/120/480 hours
• Store & review 800 groups of NIBP list data & 24-hour ECG waveforms
• Audio and Visual alarms with 3-level classification
• Protection against interference from defibrillator & electrosurgical unit
• Removable rechargeable battery
• Network capability (optional)
© SternMed4
Monitoring mode: 0.5-40HzDiagnosis mode: 0.05-75Hz>89dB20-300bpm+1% or +2bpm, whichever is greater1/2 5mm/mV@+5%1 10mm/mV@+5%2 20mm/mV@+5%12.5mm/s, 25mm/s, 50mm/s@+10%
Intelligent oscillometric methodSystolic, diastolic & mean pressureManual, automatic & STAT (5 minutes)0-300 mmHgAdult (standard)Child/neonate (standard)<90sec<10sec (typical adult cuff)<2sec (typical adult cuff)Systolic 40-275mmHg | Mean 20-230mmHg | Diastolic 10-210mmHgSystolic 40-200mmHg | Mean 20-165mmHg | Diastolic 10-150mmHgSystolic 40-135mmHg | Mean 20-110mmHg | Diastolic 10-95mmHg300mmHg240mmHg150mmHgMaximum mean error +5 mmHgMaximum standard deviation 8 mmHg
Thoracic Impedance0-120bpm+5% or +2bom, whichever is greater0-120bpm
25.0-45.0·C+0.2°C< 150sec
Dual-wavelength LED35-100%+3% (50-100%) (RMS of difference)0.4-5%30bpm-240bpm+2% or +2bpm, whichever is greater
100-240VAC, 50/80Hz7 inches TFT color LCDAudible & visible alarmEthernet portRechargeable lithium battery
IEC 60601-1Type BF, CF applied partsClass I equipmentGroup I, Class B
ECGECG Amplifier Bandwidth
CMRRHeart Rate Range Accuracy ECG Gain
ECG Sweeping SpeedNIBPMeasurement MethodParametersWorking ModesPneumatic Pressure RangeCuff Type
Average Measurement TimeInflation TimeDeflation TimeMonitoring Range Adult Child Neonate Overpressure Protection Limit Adult Child NeonateNIBP Accuracy
RESPMeasurement MethodMeasurement RangeAccuracyAlarm Limit RangeTEMPRangeAccuracyResponding TimeSPO2
TransducerSpO2 Measuring RangeSpO2 Measuring RangeLow Perfusion CapabilityPulse rate Measuring RangeAccuracy of Pulse RateGENERAL SPECIFICATIONSPower SupplyDisplay ModeAlarming ModeCommunicationBatteryCLASSIFICATIONSafety StandardType of Protection Against Electric ShockDegree of Protection Against Electric ShockElectro-Magnetic Compatibility
© SternMed5
TECHNICAL SPECIFICATIONS Macs 10 | SternMed Multi parameter patient monitor