Low Mechanical Dissipation in Fused Silica S. D. Penn, G. M. Harry, A. M. Gretarsson, S. E....

20
Low Mechanical Dissipation in Fused Silica S. D. Penn, G. M. Harry, A. M. Gretarsson, S. E. Kittleberger, P. R. Saulson, J. J. Schiller, J. R. Smith, S. O. Swords Syracuse University LSC Meeting August 2000 LIGO-G000271-D

Transcript of Low Mechanical Dissipation in Fused Silica S. D. Penn, G. M. Harry, A. M. Gretarsson, S. E....

Page 1: Low Mechanical Dissipation in Fused Silica S. D. Penn, G. M. Harry, A. M. Gretarsson, S. E. Kittleberger, P. R. Saulson, J. J. Schiller, J. R. Smith, S.

Low Mechanical Dissipation in Fused

Silica

S. D. Penn, G. M. Harry, A. M. Gretarsson,

S. E. Kittleberger, P. R. Saulson, J. J. Schiller, J. R.

Smith, S. O. Swords

Syracuse University

LSC Meeting August 2000LIGO-G000271-D

Page 2: Low Mechanical Dissipation in Fused Silica S. D. Penn, G. M. Harry, A. M. Gretarsson, S. E. Kittleberger, P. R. Saulson, J. J. Schiller, J. R. Smith, S.

Experimental Apparatus

Page 3: Low Mechanical Dissipation in Fused Silica S. D. Penn, G. M. Harry, A. M. Gretarsson, S. E. Kittleberger, P. R. Saulson, J. J. Schiller, J. R. Smith, S.

Initial Q measurements

• 1998 measurement Q = 21 ‰ 106 at f = 732 Hz Measured in smaller bell jar

• post storage measurement Q = 37 ‰ 106 at f = 2868 Hz Measured in larger bell jar, after storage and measurement with calipers

• knocked against copper tube Q = 18 ‰ 106 at f = 1579 Hz Measured in larger bell jar, surface hit against copper tube multiple times

Page 4: Low Mechanical Dissipation in Fused Silica S. D. Penn, G. M. Harry, A. M. Gretarsson, S. E. Kittleberger, P. R. Saulson, J. J. Schiller, J. R. Smith, S.

Flame polishing

• Natural gas flame polished sample surface• Surface brought to transition temperature• Polished for 15 minutes• Cooled for 15 minutes• Quickly returned to lab, reattached to suspension, and pumped down to vacuum

Q = 57 ‰ 106 at f = 726 Hz

Page 5: Low Mechanical Dissipation in Fused Silica S. D. Penn, G. M. Harry, A. M. Gretarsson, S. E. Kittleberger, P. R. Saulson, J. J. Schiller, J. R. Smith, S.

Ringdow data

1 2 3 4 5 6 7 8

x 104

102

103

Time (s)

Am

plit

ude

Data envelopeFit to exponential

Page 6: Low Mechanical Dissipation in Fused Silica S. D. Penn, G. M. Harry, A. M. Gretarsson, S. E. Kittleberger, P. R. Saulson, J. J. Schiller, J. R. Smith, S.

Additional bob in suspension

• Added additional bob to to check for excess suspension loses• Replaced center bob with two smaller bobs• Sample carefully handled so surface was never touched

Q = 49 ‰ 106 at f = 726 Hz

Page 7: Low Mechanical Dissipation in Fused Silica S. D. Penn, G. M. Harry, A. M. Gretarsson, S. E. Kittleberger, P. R. Saulson, J. J. Schiller, J. R. Smith, S.

Graph of Q for all modes

Page 8: Low Mechanical Dissipation in Fused Silica S. D. Penn, G. M. Harry, A. M. Gretarsson, S. E. Kittleberger, P. R. Saulson, J. J. Schiller, J. R. Smith, S.

Dissipation vs diameter in fused

silica fibers

Page 9: Low Mechanical Dissipation in Fused Silica S. D. Penn, G. M. Harry, A. M. Gretarsson, S. E. Kittleberger, P. R. Saulson, J. J. Schiller, J. R. Smith, S.

Flame polish of rod stock sample

• Repeated experiment with 3 mm rod stock straight from Heraeus rather than drawn from larger bob

• Flame polished in same way, but only on surface

• Surface contained visible flaws even after flame polish

• Found Q = 20.4 ‰106 at 113 Hz

Page 10: Low Mechanical Dissipation in Fused Silica S. D. Penn, G. M. Harry, A. M. Gretarsson, S. E. Kittleberger, P. R. Saulson, J. J. Schiller, J. R. Smith, S.

B. Lunin’s data on fused silica

- 1 0 0 0 1 0 0 2 0 0 3 0 0

0

1 0

2 0

3 0

4 0

5 0

6 0

Th

e Q

-fa

cto

r, 1

06 , u

nits

T e m p e r a t u r e , 0

С

Page 11: Low Mechanical Dissipation in Fused Silica S. D. Penn, G. M. Harry, A. M. Gretarsson, S. E. Kittleberger, P. R. Saulson, J. J. Schiller, J. R. Smith, S.

Syracuse temperature data

0 50 100 150 200 2500.5

1

1.5

2

2.5

3

3.5

4

4.5

5x 10

6 Q vs Temperature

Temperature, °C

Q

mode 7 mode 8 mode 9 mode 10mode 11mode 12mode 13mode 15

Page 12: Low Mechanical Dissipation in Fused Silica S. D. Penn, G. M. Harry, A. M. Gretarsson, S. E. Kittleberger, P. R. Saulson, J. J. Schiller, J. R. Smith, S.

Possible indication of cause of dissipation

)11

()log(212

1

TTk

E

f

f

B

a

(Nowick and Berry, p. 59)

f1 = 8400 HzT1 = 105 C

f2 = 726 HzT1 = 25 C

Ea = 0.31 eV

A plausible value for activation energy

Page 13: Low Mechanical Dissipation in Fused Silica S. D. Penn, G. M. Harry, A. M. Gretarsson, S. E. Kittleberger, P. R. Saulson, J. J. Schiller, J. R. Smith, S.

Implications for LIGO II

50 100 200 500 1000

5.´ 10-25

1.´ 10-24

5.´ 10-24

1.´ 10-23

5.´ 10-23

NS Binary Inspiral range• Sapphire 202 Mpc• 30 million silica 166 Mpc• 57 million silica 193 Mpc

Page 14: Low Mechanical Dissipation in Fused Silica S. D. Penn, G. M. Harry, A. M. Gretarsson, S. E. Kittleberger, P. R. Saulson, J. J. Schiller, J. R. Smith, S.

Implications for LIGO II

• Low dissipation may be unobtainable with polished mirrors

• Optical coatings may greatly increase thermal noise

• Even obtaining 30 million Q may be challenging

• Further experiments are needed and some underway at Syracuse

Page 15: Low Mechanical Dissipation in Fused Silica S. D. Penn, G. M. Harry, A. M. Gretarsson, S. E. Kittleberger, P. R. Saulson, J. J. Schiller, J. R. Smith, S.

Suggestions for further research

• Thick flame polished drawn rods – 6 mm in vacuum, 10 mm

planned– can we flame polish mirror-

like geometries?

• Get further Q vs T data– 25-500 C oven built

• Explore alternatives/additions to physical polishing– chemical treatments – learn more about Russian

experience (B Lunin)– other possibilities?

Page 16: Low Mechanical Dissipation in Fused Silica S. D. Penn, G. M. Harry, A. M. Gretarsson, S. E. Kittleberger, P. R. Saulson, J. J. Schiller, J. R. Smith, S.

Coating experiment update

• Measured uncoated Q on 16.5 cm diameter, 2 cm thick disk– Q = 3.5 ‰ 106

• Disk being coated at REO• Using smaller, 7.5 cm

diameter, 2.5 mm thick, polished but not flat, disks– 1st welded directly to suspension

– 2nd connect by bonded ear

Page 17: Low Mechanical Dissipation in Fused Silica S. D. Penn, G. M. Harry, A. M. Gretarsson, S. E. Kittleberger, P. R. Saulson, J. J. Schiller, J. R. Smith, S.

Violin mode sensor

Page 18: Low Mechanical Dissipation in Fused Silica S. D. Penn, G. M. Harry, A. M. Gretarsson, S. E. Kittleberger, P. R. Saulson, J. J. Schiller, J. R. Smith, S.

Violin mode sensor

• Non vacuum mock up made with stainless steel fiber

• Feedback controls for all pendulum modes

• Laser noise eater working• Seismic noise seen

– may have to build isolation

• Preliminary results: thermal noise measured in one violin mode, in air, with steel wire

Page 19: Low Mechanical Dissipation in Fused Silica S. D. Penn, G. M. Harry, A. M. Gretarsson, S. E. Kittleberger, P. R. Saulson, J. J. Schiller, J. R. Smith, S.

Anelastic aftereffect experiment

Page 20: Low Mechanical Dissipation in Fused Silica S. D. Penn, G. M. Harry, A. M. Gretarsson, S. E. Kittleberger, P. R. Saulson, J. J. Schiller, J. R. Smith, S.

Future plans

• Ringdown experiment with silicate bonded fused silica rods (with J. Hough and S. Rowan)

• Ellipsometry experiment to measure coating on excess 40 m mirror

• Anelastic aftereffect measurements on sapphire including thermoelastic damping

• Further vs temperature work