Long Term Performance of V-Cone Flow Meters
Mike Dyer Applications Engineer
New Zealand Gas Forum
Nov. 2017
![Page 2: Long Term Performance of V-Cone Flow Meters...• The wafer V-Cone offers 1% of rate accuracy and very little maintenance. • Straight run requirements are 0 to 4 Dia. up and 0 to](https://reader034.fdocuments.net/reader034/viewer/2022052017/6030a0fe88ec2b4340229bb4/html5/thumbnails/2.jpg)
Agenda • Powerco’s Pressure Control Module
• V-Cone Meter Principles of Operation
• Data Collection
• Test Facility
• Applications Test Meter
• Test Data
• Conclusions
• Wafer damage tests
• Straight run tests
![Page 3: Long Term Performance of V-Cone Flow Meters...• The wafer V-Cone offers 1% of rate accuracy and very little maintenance. • Straight run requirements are 0 to 4 Dia. up and 0 to](https://reader034.fdocuments.net/reader034/viewer/2022052017/6030a0fe88ec2b4340229bb4/html5/thumbnails/3.jpg)
Powerco Project
• Above ground pressure control
modules.
• They were not measuring flow but
wanted to verify their model.
![Page 4: Long Term Performance of V-Cone Flow Meters...• The wafer V-Cone offers 1% of rate accuracy and very little maintenance. • Straight run requirements are 0 to 4 Dia. up and 0 to](https://reader034.fdocuments.net/reader034/viewer/2022052017/6030a0fe88ec2b4340229bb4/html5/thumbnails/4.jpg)
Options
• Turbine meters offer good accuracy
but require maintenance.
• Other technologies did not fit.
• The wafer V-Cone offers 1% of rate
accuracy and very little maintenance.
• Straight run requirements are 0 to 4
Dia. up and 0 to 2 Dia. down.
![Page 5: Long Term Performance of V-Cone Flow Meters...• The wafer V-Cone offers 1% of rate accuracy and very little maintenance. • Straight run requirements are 0 to 4 Dia. up and 0 to](https://reader034.fdocuments.net/reader034/viewer/2022052017/6030a0fe88ec2b4340229bb4/html5/thumbnails/5.jpg)
Wafer V-Cone vertical up flow
![Page 6: Long Term Performance of V-Cone Flow Meters...• The wafer V-Cone offers 1% of rate accuracy and very little maintenance. • Straight run requirements are 0 to 4 Dia. up and 0 to](https://reader034.fdocuments.net/reader034/viewer/2022052017/6030a0fe88ec2b4340229bb4/html5/thumbnails/6.jpg)
Wafer V-Cone
![Page 7: Long Term Performance of V-Cone Flow Meters...• The wafer V-Cone offers 1% of rate accuracy and very little maintenance. • Straight run requirements are 0 to 4 Dia. up and 0 to](https://reader034.fdocuments.net/reader034/viewer/2022052017/6030a0fe88ec2b4340229bb4/html5/thumbnails/7.jpg)
Cone Meter Principles of
Operation High Pressure
Port
Low Pressure
Port
Flow
![Page 8: Long Term Performance of V-Cone Flow Meters...• The wafer V-Cone offers 1% of rate accuracy and very little maintenance. • Straight run requirements are 0 to 4 Dia. up and 0 to](https://reader034.fdocuments.net/reader034/viewer/2022052017/6030a0fe88ec2b4340229bb4/html5/thumbnails/8.jpg)
Equations
=Thermal Expansion Correction
=Meter Flow Coefficient
=Gas Expansion Factor (2001)
=Meter Geometric Constant
=Differential Pressure
=Fluid Density
DPKYCFQ Da 1
aF
DC
Y
1K
DP
![Page 9: Long Term Performance of V-Cone Flow Meters...• The wafer V-Cone offers 1% of rate accuracy and very little maintenance. • Straight run requirements are 0 to 4 Dia. up and 0 to](https://reader034.fdocuments.net/reader034/viewer/2022052017/6030a0fe88ec2b4340229bb4/html5/thumbnails/9.jpg)
Test Data Collection
• Customer’s were required by the EPA
to recalibrate meters.
• All of the calibrations took place at 3rd
party lab, CEESI Co. (air calibrated)
• This reduced the added uncertainty of
a second lab.
![Page 10: Long Term Performance of V-Cone Flow Meters...• The wafer V-Cone offers 1% of rate accuracy and very little maintenance. • Straight run requirements are 0 to 4 Dia. up and 0 to](https://reader034.fdocuments.net/reader034/viewer/2022052017/6030a0fe88ec2b4340229bb4/html5/thumbnails/10.jpg)
Test Facility
• Flow rates were dictated by
customers application but were
consistent for any given meter.
• Master meter Critical Flow Venturi
(CFV).
![Page 11: Long Term Performance of V-Cone Flow Meters...• The wafer V-Cone offers 1% of rate accuracy and very little maintenance. • Straight run requirements are 0 to 4 Dia. up and 0 to](https://reader034.fdocuments.net/reader034/viewer/2022052017/6030a0fe88ec2b4340229bb4/html5/thumbnails/11.jpg)
Test Facility
• Flow rate uncertainty 0.29% to 0.35%.
• Using the root sum square method
with sensitivity factors.
DPKYF
QC
a
D
1
![Page 12: Long Term Performance of V-Cone Flow Meters...• The wafer V-Cone offers 1% of rate accuracy and very little maintenance. • Straight run requirements are 0 to 4 Dia. up and 0 to](https://reader034.fdocuments.net/reader034/viewer/2022052017/6030a0fe88ec2b4340229bb4/html5/thumbnails/12.jpg)
Cd Test Error
2222
TEPEDPEQEerrorD STSPSDPSQC
EQ
EDP
EP
ET
S
= Flow Error (0.29% to 0.35%)
= DP Error (.07 to .75%)
= Pressure Error (.07%)
= Temperature Error (± 0.5 C)
= Sensitivity Factor for each
measurement (1 to .48)
uncertainty of approximately 0.4% to 0.5%. DC
![Page 13: Long Term Performance of V-Cone Flow Meters...• The wafer V-Cone offers 1% of rate accuracy and very little maintenance. • Straight run requirements are 0 to 4 Dia. up and 0 to](https://reader034.fdocuments.net/reader034/viewer/2022052017/6030a0fe88ec2b4340229bb4/html5/thumbnails/13.jpg)
Test Meter Applications
63 total calibrations, two of the 63 calibrations were identified as
outliers based on the Grubbs test and a 99.9% confidence.
![Page 14: Long Term Performance of V-Cone Flow Meters...• The wafer V-Cone offers 1% of rate accuracy and very little maintenance. • Straight run requirements are 0 to 4 Dia. up and 0 to](https://reader034.fdocuments.net/reader034/viewer/2022052017/6030a0fe88ec2b4340229bb4/html5/thumbnails/14.jpg)
Example of Test Data
7 calibrations yielded 21 data sets
Cal Year Ave Cd YBC Cd shift YBC Cd Shift YBC Cd Shift YBC Cd Shift YBC Cd Shift YBC Cd Shift
1995 0.844681 0
2001 0.849769 6 0.60% 0
2006 0.850767 11 0.72% 5 0.12%
2008 0.84994 13 0.62% 7 0.02% 2 -0.10%
2010 0.852034 15 0.87% 9 0.27% 4 0.15% 2 0.25%
2012 0.853277 17 1.02% 11 0.41% 6 0.29% 4 0.39% 2 0.15%
2013 0.859209 18 1.72% 12 1.11% 7 0.99% 5 1.09% 3 0.84% 1 0.70%
Meter A
![Page 15: Long Term Performance of V-Cone Flow Meters...• The wafer V-Cone offers 1% of rate accuracy and very little maintenance. • Straight run requirements are 0 to 4 Dia. up and 0 to](https://reader034.fdocuments.net/reader034/viewer/2022052017/6030a0fe88ec2b4340229bb4/html5/thumbnails/15.jpg)
• As DP meters wear they generate
less DP.
• We expected to see Cd values
increase over time.
DPKYF
QC
a
D
1
![Page 16: Long Term Performance of V-Cone Flow Meters...• The wafer V-Cone offers 1% of rate accuracy and very little maintenance. • Straight run requirements are 0 to 4 Dia. up and 0 to](https://reader034.fdocuments.net/reader034/viewer/2022052017/6030a0fe88ec2b4340229bb4/html5/thumbnails/16.jpg)
Cd Change vs Time
15 meters, 61 calibrations, 107 data sets
![Page 17: Long Term Performance of V-Cone Flow Meters...• The wafer V-Cone offers 1% of rate accuracy and very little maintenance. • Straight run requirements are 0 to 4 Dia. up and 0 to](https://reader034.fdocuments.net/reader034/viewer/2022052017/6030a0fe88ec2b4340229bb4/html5/thumbnails/17.jpg)
Cd vs Time by Meter
![Page 18: Long Term Performance of V-Cone Flow Meters...• The wafer V-Cone offers 1% of rate accuracy and very little maintenance. • Straight run requirements are 0 to 4 Dia. up and 0 to](https://reader034.fdocuments.net/reader034/viewer/2022052017/6030a0fe88ec2b4340229bb4/html5/thumbnails/18.jpg)
Conclusions
• Data shows cone meters have
extremely good long term
performance.
• Recalibrating not necessary, however
we did not test abrasive fluids.
• We do recommend periodic visual
inspections.
![Page 19: Long Term Performance of V-Cone Flow Meters...• The wafer V-Cone offers 1% of rate accuracy and very little maintenance. • Straight run requirements are 0 to 4 Dia. up and 0 to](https://reader034.fdocuments.net/reader034/viewer/2022052017/6030a0fe88ec2b4340229bb4/html5/thumbnails/19.jpg)
![Page 20: Long Term Performance of V-Cone Flow Meters...• The wafer V-Cone offers 1% of rate accuracy and very little maintenance. • Straight run requirements are 0 to 4 Dia. up and 0 to](https://reader034.fdocuments.net/reader034/viewer/2022052017/6030a0fe88ec2b4340229bb4/html5/thumbnails/20.jpg)
Damaged Cone
![Page 21: Long Term Performance of V-Cone Flow Meters...• The wafer V-Cone offers 1% of rate accuracy and very little maintenance. • Straight run requirements are 0 to 4 Dia. up and 0 to](https://reader034.fdocuments.net/reader034/viewer/2022052017/6030a0fe88ec2b4340229bb4/html5/thumbnails/21.jpg)
0.845
0.850
0.855
0.860
0.865
0.870
0.875
0.880
0.885
0.890
0.895
0.900
0 20000 40000 60000 80000 100000 120000 140000 160000
Flo
wm
ete
r C
f
Re
baseline no damage
heavily damaged cone
![Page 22: Long Term Performance of V-Cone Flow Meters...• The wafer V-Cone offers 1% of rate accuracy and very little maintenance. • Straight run requirements are 0 to 4 Dia. up and 0 to](https://reader034.fdocuments.net/reader034/viewer/2022052017/6030a0fe88ec2b4340229bb4/html5/thumbnails/22.jpg)
Obstruction Test Data
• 0.45 betas are very large cone and
condition flow better than a .75 beta.
• 0.45 beta test data with elbows out of
plane and half open gate valves.
• 0.75 beta with elbows out of plane.
• 0.65 beta Wafer V-Cones with
reducers.
![Page 23: Long Term Performance of V-Cone Flow Meters...• The wafer V-Cone offers 1% of rate accuracy and very little maintenance. • Straight run requirements are 0 to 4 Dia. up and 0 to](https://reader034.fdocuments.net/reader034/viewer/2022052017/6030a0fe88ec2b4340229bb4/html5/thumbnails/23.jpg)
Straight Run Data
![Page 24: Long Term Performance of V-Cone Flow Meters...• The wafer V-Cone offers 1% of rate accuracy and very little maintenance. • Straight run requirements are 0 to 4 Dia. up and 0 to](https://reader034.fdocuments.net/reader034/viewer/2022052017/6030a0fe88ec2b4340229bb4/html5/thumbnails/24.jpg)
Gate Valve Up and Down
Stream
3 dia. up and 0 dia. down
![Page 25: Long Term Performance of V-Cone Flow Meters...• The wafer V-Cone offers 1% of rate accuracy and very little maintenance. • Straight run requirements are 0 to 4 Dia. up and 0 to](https://reader034.fdocuments.net/reader034/viewer/2022052017/6030a0fe88ec2b4340229bb4/html5/thumbnails/25.jpg)
Elbows 3 dia. Upstream
![Page 26: Long Term Performance of V-Cone Flow Meters...• The wafer V-Cone offers 1% of rate accuracy and very little maintenance. • Straight run requirements are 0 to 4 Dia. up and 0 to](https://reader034.fdocuments.net/reader034/viewer/2022052017/6030a0fe88ec2b4340229bb4/html5/thumbnails/26.jpg)
2” x 4” Reducing Adapters 2” x 4” Concentric Reducers
![Page 27: Long Term Performance of V-Cone Flow Meters...• The wafer V-Cone offers 1% of rate accuracy and very little maintenance. • Straight run requirements are 0 to 4 Dia. up and 0 to](https://reader034.fdocuments.net/reader034/viewer/2022052017/6030a0fe88ec2b4340229bb4/html5/thumbnails/27.jpg)
Reducer Data
![Page 28: Long Term Performance of V-Cone Flow Meters...• The wafer V-Cone offers 1% of rate accuracy and very little maintenance. • Straight run requirements are 0 to 4 Dia. up and 0 to](https://reader034.fdocuments.net/reader034/viewer/2022052017/6030a0fe88ec2b4340229bb4/html5/thumbnails/28.jpg)
Questions?