LM-79 Distribution Test Report (Luminaire) Total Luminous Flux, Luminous Intensity Distribution and Electrical Characteristics of Customer Submitted Luminaire Report No.: PHM-RPT-171026-D04 Date: October 26, 2017 Page 1 of 9 LM-79 Distribution Testing: Total Luminous Flux, Luminous Intensity Distribution, Electrical Characteristics and Efficacy. Report Prepared for: Vantage Lighting 181 Narragansett Park Drive East Providence, RI 02916 USA Telephone: (860) 564-4512 Description : 3” Recessed Downlight Model Tested: V3CRRU-1335K-M WO/CS#: R3243 Report Issue Date: October 26, 2017 Date of Test(s): October 24-26, 2017 Total Number of Pages : 9 (including this page) Report Reviewed by: Mark Razvi Photometric Lab Supervisor Curtis-Straus

Transcript of LM-79 Distribution Testing: Total Luminous Flux, Luminous … · 2017-11-26 · 4.8 2738 2709 2700...

Page 1: LM-79 Distribution Testing: Total Luminous Flux, Luminous … · 2017-11-26 · 4.8 2738 2709 2700 2698 2708 2723 2735 2754 2782 2816 2827 2824 2821 2802 2762 2735 2713 4.9 2728 2708

LM-79 Distribution Test Report (Luminaire)

Total Luminous Flux, Luminous Intensity Distribution and Electrical Characteristics of Customer Submitted Luminaire

Report No.: PHM-RPT-171026-D04 Date: October 26, 2017

Page 1 of 9

LM-79 Distribution Testing: Total Luminous Flux, Luminous Intensity Distribution, Electrical

Characteristics and Efficacy.

Report Prepared for:

Vantage Lighting 181 Narragansett Park Drive East Providence, RI 02916

USA Telephone: (860) 564-4512

Description : 3” Recessed Downlight

Model Tested: V3CRRU-1335K-M WO/CS#: R3243

Report Issue Date: October 26, 2017 Date of Test(s): October 24-26, 2017 Total Number of Pages : 9 (including this page)

Report Reviewed by:

Mark Razvi

Photometric Lab Supervisor Curtis-Straus

Page 2: LM-79 Distribution Testing: Total Luminous Flux, Luminous … · 2017-11-26 · 4.8 2738 2709 2700 2698 2708 2723 2735 2754 2782 2816 2827 2824 2821 2802 2762 2735 2713 4.9 2728 2708

LM-79 Distribution Test Report (Luminaire)

Total Luminous Flux, Luminous Intensity Distribution and Electrical Characteristics of Customer Submitted Luminaire

Report No.: PHM-RPT-171026-D04 Date: October 26, 2017

Page 2 of 9

DETAILED PRODUCT DESCRIPTION Model # V3CRRU-1335K-M CS/WO# R3243 Description 3” Recessed Downlight Supply Voltage 120VAC Pictures:

Page 3: LM-79 Distribution Testing: Total Luminous Flux, Luminous … · 2017-11-26 · 4.8 2738 2709 2700 2698 2708 2723 2735 2754 2782 2816 2827 2824 2821 2802 2762 2735 2713 4.9 2728 2708

LM-79 Distribution Test Report (Luminaire)

Total Luminous Flux, Luminous Intensity Distribution and Electrical Characteristics of Customer Submitted Luminaire

Report No.: PHM-RPT-171026-D04 Date: October 26, 2017

Page 3 of 9

Curtis-Straus Photometric Testing Information

Photometric Testing is performed in accordance with the procedures outlined in LM-79-2008 . The test sample is evaluated for photometric and electrical characteristics using the equipment listed below.

Curtis-Straus Environmental Testing Info Curtis-Straus testing laboratory temperature during testing: 25.8 °C .

LIST OF EQUIPMENT USED: Description Model

No. Serial No. Asset

No Cal. Date

Cal. Due

Yokogawa Power Meter

WT-310E C2SK21021V

2356 12-19-16` 12-19-17

Chroma AC Power Supply

61603 616030001283 2355 04-21-17 04-21-18

LabSphere Type-C GoniometerSystem

PM-150-140 022172153 AA-30058-195

10-17-17 11-17-17

Digital Thermometer

4627 160739886 2392 09-07-16 09-07-18

This technical report may only be quoted in full. Any use for advertising purposes must be granted in writing. This report is the result of a single examination of the object in question and is not generally applicable evaluation of the quality of other products in regular production.

Page 4: LM-79 Distribution Testing: Total Luminous Flux, Luminous … · 2017-11-26 · 4.8 2738 2709 2700 2698 2708 2723 2735 2754 2782 2816 2827 2824 2821 2802 2762 2735 2713 4.9 2728 2708

LM-79 Distribution Test Report (Luminaire)

Total Luminous Flux, Luminous Intensity Distribution and Electrical Characteristics of Customer Submitted Luminaire

Report No.: PHM-RPT-171026-D04 Date: October 26, 2017

Page 4 of 9

LM-79-2008 Distribution Test Data

Model # V3CRRU-1335K-M CS/WO# R3243 Description 3” Recessed Downlight Supply Voltage 120VAC

Result of Tests

Photometric and Electrical Test Data

Input Voltage: 120 VAC

Base Orientation: Base Up

Lumens (lm) Watts (W) Volts (VAC) Amps (A) Efficacy PF Ithd (%)

1286.75 14.16 120.21 0.12 97.78 0.9889 0.24

Luminous Intensity Summary: Cd Illumination Cone of Light Mounting Ht: 10 ft.

Page 5: LM-79 Distribution Testing: Total Luminous Flux, Luminous … · 2017-11-26 · 4.8 2738 2709 2700 2698 2708 2723 2735 2754 2782 2816 2827 2824 2821 2802 2762 2735 2713 4.9 2728 2708

LM-79 Distribution Test Report (Luminaire)

Total Luminous Flux, Luminous Intensity Distribution and Electrical Characteristics of Customer Submitted Luminaire

Report No.: PHM-RPT-171026-D04 Date: October 26, 2017

Page 5 of 9

IsoCandela Plot

Zonal Lumens Summary Lumens per Zone

Page 6: LM-79 Distribution Testing: Total Luminous Flux, Luminous … · 2017-11-26 · 4.8 2738 2709 2700 2698 2708 2723 2735 2754 2782 2816 2827 2824 2821 2802 2762 2735 2713 4.9 2728 2708

LM-79 Distribution Test Report (Luminaire)

Total Luminous Flux, Luminous Intensity Distribution and Electrical Characteristics of Customer Submitted Luminaire

Report No.: PHM-RPT-171026-D04 Date: October 26, 2017

Page 6 of 9

Candela Table - Type C

0 22.5 45 67.5 90 112.5 135 157.5 180 202.5 225 247.5 270 292.5 315 337.5 360

0 2920 2920 2920 2920 2920 2920 2920 2920 2920 2920 2920 2920 2920 2920 2920 2920 2920

0.1 2914 2912 2918 2928 2928 2943 2937 2935 2953 2950 2936 2917 2902 2895 2880 2877 2888

0.2 2913 2916 2915 2930 2925 2937 2939 2935 2954 2948 2940 2914 2898 2891 2874 2880 2885

0.3 2912 2914 2913 2932 2923 2930 2940 2934 2950 2946 2939 2914 2895 2886 2872 2884 2881

0.4 2905 2911 2912 2934 2926 2924 2941 2936 2946 2942 2939 2913 2897 2884 2873 2889 2880

0.5 2898 2907 2914 2935 2929 2919 2934 2940 2941 2936 2935 2916 2900 2884 2874 2890 2883

0.6 2890 2899 2917 2936 2932 2917 2927 2945 2935 2930 2930 2922 2903 2883 2879 2891 2886

0.7 2888 2891 2918 2934 2930 2916 2920 2946 2929 2930 2924 2928 2910 2889 2885 2891 2889

0.8 2886 2885 2917 2928 2928 2917 2915 2944 2929 2930 2923 2932 2916 2895 2891 2885 2891

0.9 2886 2882 2916 2923 2924 2920 2909 2942 2931 2931 2922 2934 2920 2900 2891 2879 2893

1 2890 2879 2910 2914 2916 2922 2907 2936 2932 2935 2922 2936 2919 2903 2891 2874 2888

1.1 2893 2880 2902 2904 2907 2921 2908 2928 2936 2939 2928 2932 2918 2906 2889 2870 2882

1.2 2894 2882 2893 2895 2899 2918 2910 2920 2940 2943 2934 2926 2915 2904 2883 2866 2876

1.3 2893 2885 2886 2893 2893 2915 2911 2917 2941 2940 2938 2920 2909 2899 2876 2866 2870

1.4 2891 2883 2879 2892 2886 2906 2911 2914 2938 2938 2938 2917 2903 2894 2872 2868 2864

1.5 2885 2881 2873 2891 2885 2897 2912 2912 2935 2934 2937 2914 2901 2890 2869 2871 2859

1.6 2877 2878 2873 2893 2886 2889 2905 2915 2928 2927 2933 2913 2902 2886 2867 2873 2860

1.7 2870 2870 2872 2894 2886 2883 2898 2917 2919 2920 2926 2917 2902 2882 2869 2876 2861

1.8 2864 2861 2873 2892 2885 2876 2890 2917 2910 2915 2919 2921 2906 2885 2873 2878 2861

1.9 2859 2853 2874 2886 2885 2873 2881 2916 2906 2912 2914 2924 2910 2888 2878 2872 2863

2 2854 2846 2875 2881 2882 2875 2872 2914 2903 2909 2911 2926 2914 2891 2877 2866 2864

2.1 2856 2839 2870 2872 2874 2876 2867 2908 2900 2911 2908 2928 2913 2894 2876 2860 2860

2.2 2857 2837 2862 2861 2866 2874 2866 2899 2902 2914 2912 2925 2912 2896 2874 2853 2853

2.3 2858 2838 2854 2851 2857 2872 2864 2891 2904 2917 2916 2920 2908 2896 2867 2846 2846

2.4 2857 2839 2844 2846 2846 2869 2864 2884 2904 2915 2919 2914 2900 2889 2860 2842 2838

2.5 2857 2838 2834 2842 2836 2860 2864 2877 2901 2914 2919 2909 2893 2883 2854 2841 2829

2.6 2851 2837 2826 2838 2831 2850 2864 2871 2899 2911 2918 2904 2888 2876 2849 2841 2820

2.7 2841 2835 2823 2838 2828 2841 2859 2871 2892 2903 2914 2899 2885 2869 2843 2842 2819

2.8 2832 2827 2820 2838 2825 2832 2853 2872 2883 2895 2907 2901 2882 2862 2839 2843 2818

2.9 2824 2818 2818 2836 2825 2823 2847 2872 2873 2888 2900 2902 2884 2862 2835 2845 2817

3 2817 2810 2818 2832 2825 2816 2836 2871 2866 2882 2893 2904 2887 2863 2831 2839 2818

3.1 2809 2801 2817 2827 2824 2815 2825 2870 2860 2876 2886 2905 2889 2864 2828 2832 2819

3.2 2808 2791 2813 2819 2817 2813 2815 2865 2855 2875 2880 2907 2888 2865 2826 2826 2816

3.3 2808 2785 2808 2808 2810 2811 2811 2855 2854 2876 2880 2904 2888 2867 2824 2818 2810

3.4 2807 2782 2802 2797 2801 2809 2807 2845 2855 2878 2882 2897 2885 2866 2823 2811 2804

Page 7: LM-79 Distribution Testing: Total Luminous Flux, Luminous … · 2017-11-26 · 4.8 2738 2709 2700 2698 2708 2723 2735 2754 2782 2816 2827 2824 2821 2802 2762 2735 2713 4.9 2728 2708

LM-79 Distribution Test Report (Luminaire)

Total Luminous Flux, Luminous Intensity Distribution and Electrical Characteristics of Customer Submitted Luminaire

Report No.: PHM-RPT-171026-D04 Date: October 26, 2017

Page 7 of 9

3.5 2806 2779 2792 2788 2789 2806 2805 2836 2855 2877 2883 2890 2877 2860 2821 2804 2794

3.6 2805 2778 2780 2780 2778 2799 2805 2828 2852 2877 2881 2882 2869 2853 2821 2798 2784

3.7 2801 2777 2769 2773 2769 2790 2805 2819 2850 2875 2880 2874 2862 2845 2821 2791 2773

3.8 2791 2777 2763 2771 2764 2782 2801 2817 2845 2866 2876 2866 2855 2837 2821 2787 2768

3.9 2781 2771 2758 2769 2758 2771 2795 2815 2836 2857 2867 2865 2848 2829 2819 2786 2763

4 2772 2765 2753 2767 2752 2760 2789 2813 2826 2848 2857 2865 2848 2825 2816 2785 2759

4.1 2762 2758 2752 2763 2747 2751 2777 2812 2817 2839 2849 2864 2849 2822 2813 2784 2759

4.2 2753 2747 2751 2759 2742 2746 2765 2811 2807 2830 2840 2864 2850 2820 2804 2783 2759

4.3 2750 2735 2748 2752 2740 2741 2753 2807 2798 2825 2832 2865 2850 2821 2794 2781 2757

4.4 2747 2726 2742 2742 2738 2738 2747 2798 2793 2823 2829 2861 2849 2823 2785 2773 2751

4.5 2745 2720 2736 2732 2736 2736 2741 2789 2789 2820 2829 2852 2847 2823 2776 2763 2745

4.6 2744 2713 2726 2721 2727 2735 2737 2778 2786 2819 2828 2843 2838 2816 2768 2752 2736

4.7 2743 2710 2713 2709 2718 2729 2736 2766 2784 2818 2827 2834 2830 2810 2763 2743 2725

4.8 2738 2709 2700 2698 2708 2723 2735 2754 2782 2816 2827 2824 2821 2802 2762 2735 2713

4.9 2728 2708 2692 2695 2696 2716 2731 2749 2778 2806 2824 2815 2811 2791 2762 2727 2706

5 2719 2703 2684 2692 2684 2705 2725 2745 2770 2797 2815 2812 2802 2781 2761 2725 2699

5.1 2708 2697 2677 2689 2674 2692 2719 2741 2761 2786 2805 2809 2797 2773 2759 2722 2693

5.2 2697 2690 2675 2685 2668 2681 2708 2739 2751 2776 2796 2807 2793 2768 2757 2720 2691

5.3 2687 2679 2673 2682 2663 2672 2696 2736 2740 2766 2784 2807 2789 2762 2748 2718 2689

5.4 2682 2668 2670 2675 2659 2663 2683 2732 2729 2759 2773 2806 2789 2761 2738 2716 2686

5.5 2678 2658 2666 2665 2657 2657 2673 2723 2722 2754 2768 2803 2788 2760 2727 2709 2682

5.6 2674 2650 2661 2655 2654 2654 2663 2714 2717 2750 2766 2795 2787 2760 2717 2698 2677

5.7 2672 2641 2652 2644 2647 2651 2655 2704 2712 2747 2763 2787 2780 2754 2707 2688 2668

5.8 2670 2635 2640 2632 2639 2647 2654 2692 2710 2745 2761 2776 2773 2749 2699 2677 2656

5.9 2666 2633 2629 2620 2630 2642 2652 2680 2708 2743 2758 2765 2764 2741 2697 2666 2644

6 2657 2630 2618 2614 2617 2637 2648 2671 2706 2734 2754 2753 2752 2731 2695 2656 2634

6.1 2648 2625 2607 2609 2604 2627 2643 2664 2698 2725 2744 2746 2740 2720 2692 2651 2625

6.2 2637 2620 2597 2603 2592 2615 2638 2657 2691 2715 2734 2741 2732 2710 2689 2646 2616

6.3 2625 2615 2592 2600 2584 2603 2628 2655 2682 2703 2723 2735 2727 2700 2686 2641 2613

6.4 2613 2605 2587 2596 2576 2592 2615 2652 2668 2691 2712 2734 2721 2691 2678 2639 2610

6.5 2604 2594 2582 2591 2570 2580 2602 2649 2655 2680 2701 2732 2719 2686 2668 2636 2606

6.6 2597 2583 2578 2582 2567 2571 2591 2640 2644 2672 2693 2729 2717 2684 2658 2630 2602

6.7 2589 2572 2573 2574 2563 2566 2579 2632 2634 2664 2688 2721 2716 2681 2646 2620 2597

6.8 2587 2560 2566 2562 2558 2560 2569 2622 2625 2659 2683 2713 2709 2677 2635 2610 2590

6.9 2584 2551 2556 2549 2551 2556 2565 2609 2619 2656 2680 2702 2702 2673 2624 2598 2577

7 2581 2545 2546 2535 2545 2551 2562 2597 2615 2652 2676 2690 2694 2667 2619 2586 2565

7.1 2572 2539 2534 2526 2533 2547 2558 2586 2610 2645 2673 2677 2682 2656 2613 2574 2553

7.2 2563 2535 2521 2518 2521 2539 2553 2576 2604 2638 2662 2667 2669 2645 2608 2566 2542

7.3 2553 2531 2509 2509 2509 2530 2549 2567 2598 2630 2651 2659 2658 2634 2605 2559 2530

7.4 2540 2527 2501 2505 2498 2520 2540 2562 2591 2616 2639 2650 2649 2622 2601 2553 2525

Page 8: LM-79 Distribution Testing: Total Luminous Flux, Luminous … · 2017-11-26 · 4.8 2738 2709 2700 2698 2708 2723 2735 2754 2782 2816 2827 2824 2821 2802 2762 2735 2713 4.9 2728 2708

LM-79 Distribution Test Report (Luminaire)

Total Luminous Flux, Luminous Intensity Distribution and Electrical Characteristics of Customer Submitted Luminaire

Report No.: PHM-RPT-171026-D04 Date: October 26, 2017

Page 8 of 9

7.5 2528 2519 2493 2501 2487 2508 2529 2558 2579 2602 2626 2647 2639 2610 2594 2550 2519

7.6 2517 2509 2485 2496 2478 2495 2517 2555 2567 2589 2613 2644 2634 2602 2583 2547 2515

7.7 2508 2498 2481 2488 2472 2483 2505 2547 2555 2579 2603 2641 2631 2598 2573 2542 2511

7.8 2499 2484 2477 2480 2467 2475 2492 2540 2543 2568 2596 2633 2627 2594 2561 2533 2507

7.9 2494 2470 2472 2470 2462 2467 2479 2531 2531 2561 2590 2624 2621 2589 2548 2523 2501

8 2491 2458 2464 2457 2457 2460 2473 2517 2522 2557 2586 2614 2615 2584 2535 2511 2490

8.1 2488 2451 2456 2444 2452 2457 2466 2503 2515 2553 2582 2600 2608 2579 2529 2497 2479

8.2 2479 2443 2445 2432 2442 2453 2460 2490 2508 2547 2579 2587 2596 2569 2522 2483 2467

8.3 2470 2438 2431 2421 2430 2446 2455 2479 2504 2540 2568 2575 2584 2558 2516 2474 2453

8.4 2460 2434 2418 2409 2418 2438 2451 2467 2500 2533 2556 2565 2572 2547 2511 2465 2439

8.5 2446 2430 2407 2404 2405 2429 2444 2461 2495 2519 2544 2555 2558 2533 2506 2457 2430

8.6 2432 2422 2396 2399 2392 2416 2434 2456 2484 2505 2530 2550 2545 2520 2499 2453 2422

8.7 2420 2413 2386 2394 2381 2402 2425 2451 2473 2492 2515 2546 2537 2510 2490 2450 2414

8.8 2409 2405 2381 2389 2373 2389 2412 2443 2461 2480 2503 2542 2531 2502 2481 2445 2410

8.9 2398 2392 2376 2383 2366 2379 2398 2435 2446 2469 2495 2535 2526 2494 2469 2436 2405

9 2392 2379 2370 2375 2360 2369 2384 2426 2432 2460 2488 2527 2522 2488 2456 2427 2400

9.1 2388 2368 2364 2362 2355 2361 2374 2413 2420 2455 2481 2518 2517 2484 2443 2416 2390

9.2 2383 2361 2357 2350 2350 2357 2364 2401 2411 2451 2476 2504 2513 2479 2433 2404 2380

9.3 2376 2355 2347 2337 2341 2352 2355 2388 2402 2444 2471 2490 2501 2470 2424 2391 2368

9.4 2369 2348 2335 2325 2331 2346 2350 2375 2397 2437 2462 2477 2488 2461 2415 2380 2355

9.5 2361 2340 2322 2312 2320 2339 2344 2362 2391 2430 2452 2465 2476 2451 2410 2370 2342

9.6 2348 2333 2310 2303 2307 2332 2337 2353 2386 2418 2442 2452 2462 2437 2405 2360 2331

9.7 2336 2322 2298 2296 2294 2320 2328 2347 2377 2406 2428 2444 2448 2423 2399 2355 2321

9.8 2323 2308 2285 2289 2281 2306 2319 2341 2368 2393 2414 2439 2437 2410 2391 2350 2312

9.9 2310 2294 2277 2284 2272 2293 2307 2334 2358 2380 2400 2433 2429 2401 2383 2345 2306

10 2297 2282 2269 2280 2263 2280 2292 2327 2344 2366 2391 2427 2422 2391 2372 2336 2301

15 1802 1802 1786 1787 1779 1782 1786 1798 1810 1818 1836 1850 1852 1835 1826 1809 1804

20 1334 1330 1336 1336 1335 1315 1313 1297 1306 1287 1279 1267 1271 1273 1289 1307 1332

25 842 846 862 887 887 880 870 835 811 774 750 747 761 770 797 831 845

30 379 386 399 427 422 422 414 398 362 336 327 335 349 351 361 369 379

35 101 98 97 102 104 106 107 106 102 96 95 98 101 101 100 99 101

40 45 43 43 43 44 45 45 46 45 45 45 46 47 47 48 47 45

45 26 25 25 26 25 26 25 26 26 27 27 28 28 28 28 27 26

50 17 17 16 17 16 17 16 17 17 18 19 19 19 18 18 18 17

55 11 10 10 11 10 11 11 11 12 12 14 13 12 12 12 11 11

60 7 7 8 8 7 7 7 7 8 8 34 22 8 8 8 8 7

65 6 6 7 7 6 6 6 6 6 8 12 11 7 7 7 6 6

70 6 6 6 7 6 6 6 6 7 8 8 7 7 6 6 6 6

75 6 6 6 6 6 6 6 6 7 7 7 7 7 6 6 6 6

80 6 6 6 6 6 6 6 6 7 7 7 7 7 7 7 6 6

Page 9: LM-79 Distribution Testing: Total Luminous Flux, Luminous … · 2017-11-26 · 4.8 2738 2709 2700 2698 2708 2723 2735 2754 2782 2816 2827 2824 2821 2802 2762 2735 2713 4.9 2728 2708

LM-79 Distribution Test Report (Luminaire)

Total Luminous Flux, Luminous Intensity Distribution and Electrical Characteristics of Customer Submitted Luminaire

Report No.: PHM-RPT-171026-D04 Date: October 26, 2017

Page 9 of 9

85 6 5 5 5 6 6 6 6 6 6 6 6 6 6 6 6 6

90 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 3 3

95 4 4 4 4 4 4 3 4 4 4 4 4 4 4 3 4 4

100 4 4 4 4 4 4 4 4 4 4 4 4 4 4 3 3 3

105 3 4 4 3 4 4 3 4 4 4 4 4 4 4 3 3 3

110 3 4 4 4 4 4 4 4 4 4 4 3 4 3 3 4 3

115 4 3 4 4 3 4 4 3 3 4 3 3 4 3 4 3 3

120 4 4 3 4 4 4 3 3 3 4 4 4 3 3 3 4 4

125 3 4 4 4 4 4 3 4 4 4 4 4 3 3 4 3 3

130 3 4 4 4 4 4 3 3 4 4 4 3 4 3 4 4 4

135 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4

140 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4

145 4 5 5 5 4 4 5 5 5 4 4 4 4 4 4 5 4

150 5 5 5 5 5 5 5 5 5 5 5 5 5 5 5 5 5

155 5 5 5 5 5 5 5 5 5 5 5 5 5 5 5 5 5

160 6 5 6 6 6 6 6 6 6 5 5 5 5 5 6 6 6

165 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6

170 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6

175 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6

180 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6