Left Ventricular Assist Device (LVAD) HeartWare HVAD T . Schlöglhofer

28
Schulung HVAD Oktober 2012 Left Ventricular Assist Device (LVAD) HeartWare HVAD T. Schlöglhofer Center for Medical Physics and Biomedical Engineering, Department of Cardiac Surgery, Medical University of Vienna, Austria; Ludwig-Boltzmann-Cluster for Cardiovascular Research, Vienna, Austria

description

Left Ventricular Assist Device (LVAD) HeartWare HVAD T . Schlöglhofer Center for Medical Physics and Biomedical Engineering, Department of Cardiac Surgery, Medical University of Vienna, Austria; Ludwig-Boltzmann-Cluster for Cardiovascular Research, Vienna, Austria. HeartWare HVAD. - PowerPoint PPT Presentation

Transcript of Left Ventricular Assist Device (LVAD) HeartWare HVAD T . Schlöglhofer

Page 1: Left Ventricular Assist Device (LVAD) HeartWare  HVAD T .  Schlöglhofer

Schulung HVAD Oktober 2012

Left Ventricular Assist Device (LVAD)HeartWare HVAD

T. Schlöglhofer

 Center for Medical Physics and Biomedical Engineering,

Department of Cardiac Surgery, Medical University of Vienna, Austria; Ludwig-Boltzmann-Cluster for Cardiovascular Research, Vienna, Austria

Page 2: Left Ventricular Assist Device (LVAD) HeartWare  HVAD T .  Schlöglhofer

Schulung HVAD Oktober 2012

HeartWare HVAD

Page 3: Left Ventricular Assist Device (LVAD) HeartWare  HVAD T .  Schlöglhofer

Schulung HVAD Oktober 2012

• Implantierte Komponenten:– HVAD Pumpe– Ausstrom Graft– Transcutane Driveline

• Weitere Komponenten:– Controller– System Monitor– Energiequellen

• Akkus• AC Adapter• 12V Adapter

– Zubehör

Systemkomponenten-HeartWare HVAD

Page 4: Left Ventricular Assist Device (LVAD) HeartWare  HVAD T .  Schlöglhofer

Schulung HVAD Oktober 2012

HeartWare HVAD Pumpe

• Zentrifugal Pumpe mit kontinuierlichem Fluss

• Hydrodynamische Lagerung des Impellers

Page 5: Left Ventricular Assist Device (LVAD) HeartWare  HVAD T .  Schlöglhofer

Schulung HVAD Oktober 2012

• Einstrom: aus der linken Herzkammer

• Pumpe an der Herzspitze

• Rückfluss: in die aufsteigende Aorta

• Speed Bereich: 1800 - 4000 rpm

• Flow Bereich: bis 10 L/min

Pumpe/Systemlage

Page 6: Left Ventricular Assist Device (LVAD) HeartWare  HVAD T .  Schlöglhofer

Schulung HVAD Oktober 2012

Controller (Steuereinheit)

Page 7: Left Ventricular Assist Device (LVAD) HeartWare  HVAD T .  Schlöglhofer

Schulung HVAD Oktober 2012

Stromversorgung• Akku (ca. 4-6h Betriebszeit) • AC oder DC Adapter für Netzbetrieb

Page 8: Left Ventricular Assist Device (LVAD) HeartWare  HVAD T .  Schlöglhofer

Schulung HVAD Oktober 2012

• Die Leuchten bei jeder Bat. bedeuten:– Grün bei Ready: Bat. voll– Bei Status:

• Rot: Bat zu kalt oder heiß• Blinkend rot: Bat. Defekt• Gelb: Bat wird geladen

Ladegerät

Page 9: Left Ventricular Assist Device (LVAD) HeartWare  HVAD T .  Schlöglhofer

Schulung HVAD Oktober 2012

Zubehör• Umhängetasche/Bauchgurt

• Duschtasche

Page 10: Left Ventricular Assist Device (LVAD) HeartWare  HVAD T .  Schlöglhofer

Schulung HVAD Oktober 2012

Monitor

• Anzeige aller Pumpenparameter

• Drehzahländerung

• Anzeige von Alarmen bzw. Energiestatus des Controllers

Page 11: Left Ventricular Assist Device (LVAD) HeartWare  HVAD T .  Schlöglhofer

Schulung HVAD Oktober 2012

• Flow beeinflußt von Pre- und Afterload

• Pulsatilität: erzeugt vom nativen Herzen

• Pulsatilität: Änderung des Flow zwischen Systole/Diastole

Waveform/Pulsatilität

Page 12: Left Ventricular Assist Device (LVAD) HeartWare  HVAD T .  Schlöglhofer

Schulung HVAD Oktober 2012

Waveform/Pulsatilität

Page 13: Left Ventricular Assist Device (LVAD) HeartWare  HVAD T .  Schlöglhofer

Schulung HVAD Oktober 2012

Abhängigkeit Drehzahl vs. Druck

Page 14: Left Ventricular Assist Device (LVAD) HeartWare  HVAD T .  Schlöglhofer

Schulung HVAD Oktober 2012

Übersicht Hinweis-Alarme

Page 15: Left Ventricular Assist Device (LVAD) HeartWare  HVAD T .  Schlöglhofer

Schulung HVAD Oktober 2012

Übersicht Notfall-Alarme

Page 16: Left Ventricular Assist Device (LVAD) HeartWare  HVAD T .  Schlöglhofer

Schulung HVAD Oktober 2012

• Low Flow• High Watts• VAD gestoppt - Pumpe diskonnektiert• VAD gestoppt – beide Stromquellen

diskonnektiert

Wichtigste Alarme

Page 17: Left Ventricular Assist Device (LVAD) HeartWare  HVAD T .  Schlöglhofer

Schulung HVAD Oktober 2012

• Hypovolämie• Rechtsherzdekompensation• Hypertonie• Orthostase Probleme• Arrhythmie

Low Flow Alarm – mögliche Ursachen

Page 18: Left Ventricular Assist Device (LVAD) HeartWare  HVAD T .  Schlöglhofer

Schulung HVAD Oktober 2012

• Pumpen Thrombus – bei Hämolyse (fHb) über 10mg/dl Kontakt mit dem Kunstherzteam aufnehmen!

• Pumpe defekt• Controller defekt• Kabelbruch

High Watts Alarm– mögliche Ursachen

Page 19: Left Ventricular Assist Device (LVAD) HeartWare  HVAD T .  Schlöglhofer

Schulung HVAD Oktober 2012

Waveforms

Page 20: Left Ventricular Assist Device (LVAD) HeartWare  HVAD T .  Schlöglhofer

Schulung HVAD Oktober 2012

Waveforms

Page 21: Left Ventricular Assist Device (LVAD) HeartWare  HVAD T .  Schlöglhofer

Schulung HVAD Oktober 2012

Waveforms

Page 22: Left Ventricular Assist Device (LVAD) HeartWare  HVAD T .  Schlöglhofer

Schulung HVAD Oktober 2012

Waveforms

Page 23: Left Ventricular Assist Device (LVAD) HeartWare  HVAD T .  Schlöglhofer

Schulung HVAD Oktober 2012

Waveforms

Page 24: Left Ventricular Assist Device (LVAD) HeartWare  HVAD T .  Schlöglhofer

Schulung HVAD Oktober 2012

MARCOUMAR: • Therapeutischer Bereich INR 2.0 - 3.0 (entspricht etwa einem

Thrombotest von 10 - 14%)• Achtung: bei Lupus-antikoagulans-positiven Patienten ist ein

höherer INR notwendig!• Wird bei Kontrolle der therapeutische Bereich unterschritten, ist

Lovenoxgabe meist notwendig - nach Rücksprache mit dem Kunstherzteam!

• Macoumar-Diät (Achtung bei Speisen mit Vitamin K!), die Patienten sind darauf hinzuweisen.

Thrombo-ASS : • Weiter nach Verordnung (typisch: 1x100mg) bzw. individuell

Gerinnungsmanagement

Page 25: Left Ventricular Assist Device (LVAD) HeartWare  HVAD T .  Schlöglhofer

Schulung HVAD Oktober 2012

VerbandswechselVorbereitungBeim Verbandwechsel der Kabelaustrittstelle bei Pumpenpatienten sind immer ein Mundschutz und Gummihandschuhe zu verwenden! Mundschutz für Schwester/Pfleger und Patient!

Schritt 1: Reinigen der Kabelaustrittstelle• Gegebenfalls Haare abrasieren, danach Kabelaustrittstelle mit Octenisept reinigen/desinfizieren (Keine alkoholischen Reinigungsmittel!)• Bei infizierten Kabelaustrittstellen Verwendung von Silbernitratlösung (ca. 10 Minuten mit befeuchtetem Tupfer einwirken lassen)

Page 26: Left Ventricular Assist Device (LVAD) HeartWare  HVAD T .  Schlöglhofer

Schulung HVAD Oktober 2012

Verbandswechsel

Schritt 2: Abdecken der Kabelaustrittstelle mit Verband• Die gereinigte Kabelaustrittstelle wird mit einem geschlitzten kleinen Tupfer bedeckt• Durchsichtige Katheterabdeckfolie (Smith&Nephew IV3000 1-Hand) wie abgebildet auf die Tupfer kleben

Page 27: Left Ventricular Assist Device (LVAD) HeartWare  HVAD T .  Schlöglhofer

Schulung HVAD Oktober 2012

Verbandswechsel

Schritt 3: Befestigen des Pumpenkabels• Das Pumpenkabel muss so nahe als moglich an der Austrittstelle mit der Schlauchbefestigung fixiert werden!

• Es muss sichergestellt sein, dass sich das Kabel an der Austrittstelle so wenig wie moglich bewegen kann, um eine mechanische Reizung zu verhindern!

Page 28: Left Ventricular Assist Device (LVAD) HeartWare  HVAD T .  Schlöglhofer

Schulung HVAD Oktober 2012

Kontaktdaten AKH Wien - Kunstherz

Verantwortliche Ärzte:Doz. Dr. Daniel Zimpfer 0664/243 72 22 oder 01/40400 6835Dr. Stephane Mahr 0676/377 96 19Dr. Thomas Haberl 0664/207 79 46Dr. Julia Riebandt 0650/523 83 77

Technische Betreuung:24h-Bereitschaftsnummer 01/40400 2200Prof. Heinrich Schima 0664/454 60 33 oder 01/40400 2001KT Zeno Hartner 0676/342 31 01 oder 01/40400 2001KT Johann Horvat 0664/504 56 26 oder 01/40400 2001BSc. Thomas Schlöglhofer 0650/711 91 29 oder 01/40400 2001