Lecture - 9 - Microscopy

download Lecture - 9 - Microscopy

of 33

Transcript of Lecture - 9 - Microscopy

  • 7/27/2019 Lecture - 9 - Microscopy

    1/33

    Lecture 9Image Analysis

  • 7/27/2019 Lecture - 9 - Microscopy

    2/33

    Nature Material: Macro, micro, nano

    10 centimtres

    1X

    1 centimtre

    10X

    100X

    1 millimtre

    100 microns

    1000X

    10 microns

    10,000X

    1 micron

    100,000X

    1,000,000X

    100 nanomtres

    10,000,000X

    10 nanomtres

    100,000,000X

    1 nanomtre

  • 7/27/2019 Lecture - 9 - Microscopy

    3/33

  • 7/27/2019 Lecture - 9 - Microscopy

    4/33

    Microstructural Features which concern us

    Grain SizeGrain Shapes

    Precipitate Size: mostly in the micron regime

    Volume fraction and distribution of various phases

    Defects such as cracks voids

    Atom orientation within crystal

    Microstructrue ranging from crystal structure to engine components (e.g Si3N4)

  • 7/27/2019 Lecture - 9 - Microscopy

    5/33

    Characteristics Information : SEM

    Crystallographic Information !!How the atoms are arranged in the object; direct relation

    between these arrangements and material properties

  • 7/27/2019 Lecture - 9 - Microscopy

    6/33

    Scanning Electron Microscope (SEM) The scanning electron microscope (SEM) is a type of electron microscope that takes images of the surface of

    samples by focusing a high energy beam of electrons onto the sample.

    The electrons interact with the atoms that make up the sample producing signals that give information aboutthe sample.

    Electronic devices are used to capture or detect these signals and either allow them to expose film or, mostcommon today, create an image on a computer screen.

    The type of signals made by an SEM can include secondary electrons, characteristic x-rays, and back scatteredelectrons.

    In an SEM, these signals come from the beam of electrons striking the surface of the specimen and interactingwith the sample at or near its surface

    The SEM has allowed researchers to examine a much bigger variety of specimens The scanning electron microscope has many advantages over traditional microscopes. The SEM has a large

    depth of field, which allows more of a specimen to be in focus at one time.

    The SEM also has much higher resolution, so closely spaced specimens can be magnified at much higher levels

    Because the SEM uses electromagnets rather than lenses, the researcher has much more control in the degreeof magnification.

    All of these advantages, as well as the actual strikingly clear images, make the scanning electron microscopeone of the most useful instruments in research today.

    SEM micrographs have a very large depth of focus yielding a characteristic three-dimensional appearanceuseful for understanding the surface structure of a sample.

    Characteristic x-rays are the second most common imaging mode for an SEM.

    These characteristic x-rays are used to tell the chemical composition of the sample.

    Back-scattered electrons (BSE) that come from the sample may also be used to form an image.

  • 7/27/2019 Lecture - 9 - Microscopy

    7/33

    Sample Preparation for SEM SEM utilizes vacuum conditions and uses electrons to form an image, special preparations must be done

    to the sample.

    All water must be removed from the samples because the water would vaporize in the vacuum.

    All metals are conductive and require no preparation before being used. All non-metals need to be made conductive by covering the sample with a thin layer of conductive

    material. This is done by using a device called a "sputter coater.

    The sputter coater uses an electric field and argon gas. The sample is placed in a small chamber that isat a vacuum. Argon gas and an electric field cause an electron to be removed from the argon, makingthe atoms positively charged.

    Specimen at high vacuum requires sample fixation and dehydration or freezing.

    Minimized charging by coating sample with metal or carbon or lowering the operating kV

  • 7/27/2019 Lecture - 9 - Microscopy

    8/33

  • 7/27/2019 Lecture - 9 - Microscopy

    9/33

    How does SEM work?

    What happens when the Electron Beam hits the sample?

    SEM Setup

    Electron/Specimen InteractionsWhen the electron beam strikes the sample, both photon and electron signals are emitted

  • 7/27/2019 Lecture - 9 - Microscopy

    10/33

    Secondary Electron

    Generation

    -SEM-SE

    -sample electrons ejected by the

    primary beam [green line]

    -low energy

    -surface detail & topography

    SEM Imaging Modes

    SE

    Backscattered Electron

    Generation

    -SEM-BSE

    -primarybeam electrons

    -high energy

    -composition and topography

    [specimen atomic number]

    BSE

  • 7/27/2019 Lecture - 9 - Microscopy

    11/33

  • 7/27/2019 Lecture - 9 - Microscopy

    12/33

    Identification of Fracture Mode

    SEM micrographs of fracturedsurface of two BaTiO3 samples

    D.Sarkar, B.Basu, M.J.Chu and S.J.Cho, R-Curve Behavior of Ti3SiC2, Ceramics International, 33 [5] 789 -793, 2007.

  • 7/27/2019 Lecture - 9 - Microscopy

    13/33

    EDX EDX is an analytical technique used for the elemental analysis or chemical

    characterization of a specimen.

    it relies on the investigation of a sample through interactions betweenelectromagnetic radiation and matter, analyzing X-rays emitted by the matter inthis particular case.

    Each element has a unique atomic structure allowing x-rays to be uniquely

    distinguished from each other.

    A graph is plotted (x-ray energy vs. count rate). The peaks correspond to characteristic elemental emissions.

  • 7/27/2019 Lecture - 9 - Microscopy

    14/33

    Nanoscale MaterialsNanowires and Nanotubes

    Lateral dimension: 1 100 nm Nanowires and nanotubes exhibit novel

    physical, electronic andoptical properties due to Two dimensional quantum

    confinement

    Structural one dimensionality High surface to volume ratio

    Potential application in wide range ofnanodevices and systems Nanoscale sensors and actuators

    Photovoltaic devices solar cells Transistors, diodes and LASERs

    Nanowire Solar Cell: The nanowirescreate a surface that is able to absorb

    more sunlight than a flat surface

  • 7/27/2019 Lecture - 9 - Microscopy

    15/33

  • 7/27/2019 Lecture - 9 - Microscopy

    16/33

    Nanofluids

    Aluminum Oxide

    Particles in Water

    Nanofluids are engineered colloids = base fluid (water, organic liquid) + nanoparticles

    Nanoparticle size: 1-100 nm

    Nanoparticle materials: Al2O3, ZrO2, SiO2, CuO, Fe3O4, Au, Cu, C (diamond, PyC, fullerene) etc.

    Previous studies suggest significant enhancement of:

    - Thermal conductivity (+40%)

    - Single-phase convective

    heat transfer (+40%)

    - Critical Heat Flux (+100%)

  • 7/27/2019 Lecture - 9 - Microscopy

    17/33

    B i P i i l f TEM

  • 7/27/2019 Lecture - 9 - Microscopy

    18/33

    Basic Principle of TEM

    Electron source, electromagnetic lens system, sample holder, and imaging system

    Electron Source

    The electron source consists of a cathode and an anode. The cathode is

    a tungsten filament which emits electrons when being heated. A negative

    cap confines the electrons into a loosely focused beam. The beam is

    then accelerated towards the specimen by the positive anode. Electrons

    at the rim of the beam will fall onto the anode while the others at the

    center will pass through the small hole of the anode. The electron source

    works like a cathode ray tube.

    Electromagnetic lens system

    After leaving the electron source, the electron beam is tightly focused using electromagnetic lens and metal

    apertures. The system only allows electrons within a small energy range to pass through, so the electrons in the

    electron beam will have a well-defined energy.

    1.Magnetic Lens: Circularelectro-magnets capable of generating a precise circular magnetic field. The field acts

    like an optical lens to focus the electrons.

    2.Aperture: A thin disk with a small (2-100 micrometers) circular through-hole. It is used to restrict the electronbeam and filter out unwanted electrons before hitting the specimen.

    Sample Holder

  • 7/27/2019 Lecture - 9 - Microscopy

    19/33

    Sample Holder

    The sample holder is a platform equipped with a mechanical arm for holding the specimen and controlling its

    position.

    Imaging system

    The imaging system consists of another electromagnetic lens system and a screen. The electromagnetic lenssystem contains two lens systems, one for refocusing the electrons after they pass through the specimen, and the

    other for enlarging the image and projecting it onto the screen. The screen has a phosphorescent plate which

    glows when being hit by electrons. Image forms in a way similar to photography.

    Working principle

    TEM works like a slide projector. A projector shines a beam of light which transmits through the slide. The patterns

    painted on the slide only allow certain parts of the light beam to pass through. Thus the transmitted beam

    replicates the patterns on the slide, forming an enlarged image of the slide when falling on the screen.

    TEMs work the same way except that they shine a beam of electrons (like the light in a slide projector) through the

    specimen (like the slide). However, in TEM, the transmission of electron beam is highly dependent on the

    properties of material being examined. Such properties include density, composition, etc. For example, porous

    material will allow more electrons to pass through while dense material will allow less. As a result, a specimen with

    a non-uniform density can be examined by this technique. Whatever part is transmitted is projected onto aphosphor screen for the user to see.

    The following movie will help you to understand more about the operation of a TEM:

    Imaging Mode of TEM: Bright Field Mode

    Dark Field Mode

    Diffraction Mode

  • 7/27/2019 Lecture - 9 - Microscopy

    20/33

    In a TEM, the specimen you want tolook at must be of such a low

    density that it allows electrons to

    travel through the tissue

    There are different ways to prepare

    your material for that purpose. You

    can cut very thin slices of your

    specimen from a piece of tissue

    either by fixing it in plastic or

    working with it as frozen material.

    Another way to prepare your

    specimen is to isolate it and study a

    solution of for example viruses or

    molecules in the TEM.

    Preparation of Specimen

  • 7/27/2019 Lecture - 9 - Microscopy

    21/33

    TEM Sample Prep for Materials

  • 7/27/2019 Lecture - 9 - Microscopy

    22/33

    How to Measure SAED (Selected Area Electron Diffraction)

    Au (111) ring [2.35 d-spacing]

    With 200KV, (111) ring should distance Rfrom the transmitted beam

    Rdhkl=lL

    R is measuredd inter atomic distancel is the electron wavelengthL is the camera length(lL is the camera constant)

    Think of TEM as a diffraction camera

    Transmitted Beam

    Diffracted Beam

    L

    R

  • 7/27/2019 Lecture - 9 - Microscopy

    23/33

    Electron DiffractionFour conditions in Back Focal Plane (BFP) of the

    objective lens: No sample No reflections (only transmitted beam)

    Amorphous Transmitted beam + random scattering

    Polycrystal Transmitted beam + rings

    Single crystal Transmitted beam + spots

    Metal particles Polymer mix

    Electron Diffraction

    Bright field TEM of FeNbO4 and (1 wt %) Pt

    impregnated FeNbO4. (Inset) SAED pattern

    of FeNbO4

    Particle Size : 30-60nm

  • 7/27/2019 Lecture - 9 - Microscopy

    24/33

    TiO2 Nanoscale Rods

    TEM images of Cu on ZnO (a model methanol catalyst)

  • 7/27/2019 Lecture - 9 - Microscopy

    25/33

    (b) bright field image

    TEM Characterization of TiO2 Nanoparticles

    The structure of all as-grownsamples is anatase.

    The particle sizes from TEM rangebetween 15 and 25 nm.

    (a) dark field image (c) diffraction patterns

    (d) lattice image

    20nm

    (d) Lattice Image

    Cr stallite Si e is Different than Particle Si e

  • 7/27/2019 Lecture - 9 - Microscopy

    26/33

    Crystallite Size is Different than Particle Size

    A particle may be made up of several different crystallites

    Crystallite size often matches grain size, but there are exceptions

    Grain boundaries are interfaces where

    crystals of different orientations meet.

    A grain boundary is a single-phase

    interface, with crystals on each side of

    the boundary being identical except in

    orientation. The term "crystallite

    boundary" is sometimes, though

    rarely, used. Grain boundary areas

    contain those atoms that have been

    perturbed from their original lattice

    sites, dislocations, and impurities that

    have migrated to the lower energygrain boundary.

    Crystallite is the average size of the particle whereas the

    particle size denotes the individual size of the particle

    Crystallite size is usually measured

    from X-ray diffraction patterns and

    grain size by other experimental

    techniques like transmission electron

    microscopy.

    Most materials are polycrystalline;

    they are made of a large number ofsingle crystals crystallites.

    S i P b Mi

  • 7/27/2019 Lecture - 9 - Microscopy

    27/33

    Scanning Probe Microscopy

    atomic force microscope

    sharp probe moves over surface of specimen atconstant distance

    up and down movement of probe as it maintainsconstant distance is detected and used to create image

    scanning tunneling microscope

    steady current (tunneling current) maintained betweenmicroscope probe and specimen

    up and down movement of probe as it maintains currentis detected and used to create image of surface ofspecimen

  • 7/27/2019 Lecture - 9 - Microscopy

    28/33

    1 2

    3 4

  • 7/27/2019 Lecture - 9 - Microscopy

    29/33

    5 6

    7 8

  • 7/27/2019 Lecture - 9 - Microscopy

    30/33

    Transmission Electron Microscopy Study of

    TiO2 Phase Transformation

    As-deposited 700 oC 800 oC

    TEM diffraction patterns

    for annealed and as-deposited 12-nm sample.

  • 7/27/2019 Lecture - 9 - Microscopy

    31/33

    Dark Field Imaging

    If the transmitted beamis excluded from theimage formation process

    off-axis imaging

    tilted beam imaging

  • 7/27/2019 Lecture - 9 - Microscopy

    32/33

    Bright Field Imaging

    If the main portion of the near-forwardscattered beam is used to form the image

    transmitted beam 000 beam

    zero-order beam

  • 7/27/2019 Lecture - 9 - Microscopy

    33/33