Lecture - 9 - Microscopy
-
Upload
rachit-goyal -
Category
Documents
-
view
220 -
download
0
Transcript of Lecture - 9 - Microscopy
-
7/27/2019 Lecture - 9 - Microscopy
1/33
Lecture 9Image Analysis
-
7/27/2019 Lecture - 9 - Microscopy
2/33
Nature Material: Macro, micro, nano
10 centimtres
1X
1 centimtre
10X
100X
1 millimtre
100 microns
1000X
10 microns
10,000X
1 micron
100,000X
1,000,000X
100 nanomtres
10,000,000X
10 nanomtres
100,000,000X
1 nanomtre
-
7/27/2019 Lecture - 9 - Microscopy
3/33
-
7/27/2019 Lecture - 9 - Microscopy
4/33
Microstructural Features which concern us
Grain SizeGrain Shapes
Precipitate Size: mostly in the micron regime
Volume fraction and distribution of various phases
Defects such as cracks voids
Atom orientation within crystal
Microstructrue ranging from crystal structure to engine components (e.g Si3N4)
-
7/27/2019 Lecture - 9 - Microscopy
5/33
Characteristics Information : SEM
Crystallographic Information !!How the atoms are arranged in the object; direct relation
between these arrangements and material properties
-
7/27/2019 Lecture - 9 - Microscopy
6/33
Scanning Electron Microscope (SEM) The scanning electron microscope (SEM) is a type of electron microscope that takes images of the surface of
samples by focusing a high energy beam of electrons onto the sample.
The electrons interact with the atoms that make up the sample producing signals that give information aboutthe sample.
Electronic devices are used to capture or detect these signals and either allow them to expose film or, mostcommon today, create an image on a computer screen.
The type of signals made by an SEM can include secondary electrons, characteristic x-rays, and back scatteredelectrons.
In an SEM, these signals come from the beam of electrons striking the surface of the specimen and interactingwith the sample at or near its surface
The SEM has allowed researchers to examine a much bigger variety of specimens The scanning electron microscope has many advantages over traditional microscopes. The SEM has a large
depth of field, which allows more of a specimen to be in focus at one time.
The SEM also has much higher resolution, so closely spaced specimens can be magnified at much higher levels
Because the SEM uses electromagnets rather than lenses, the researcher has much more control in the degreeof magnification.
All of these advantages, as well as the actual strikingly clear images, make the scanning electron microscopeone of the most useful instruments in research today.
SEM micrographs have a very large depth of focus yielding a characteristic three-dimensional appearanceuseful for understanding the surface structure of a sample.
Characteristic x-rays are the second most common imaging mode for an SEM.
These characteristic x-rays are used to tell the chemical composition of the sample.
Back-scattered electrons (BSE) that come from the sample may also be used to form an image.
-
7/27/2019 Lecture - 9 - Microscopy
7/33
Sample Preparation for SEM SEM utilizes vacuum conditions and uses electrons to form an image, special preparations must be done
to the sample.
All water must be removed from the samples because the water would vaporize in the vacuum.
All metals are conductive and require no preparation before being used. All non-metals need to be made conductive by covering the sample with a thin layer of conductive
material. This is done by using a device called a "sputter coater.
The sputter coater uses an electric field and argon gas. The sample is placed in a small chamber that isat a vacuum. Argon gas and an electric field cause an electron to be removed from the argon, makingthe atoms positively charged.
Specimen at high vacuum requires sample fixation and dehydration or freezing.
Minimized charging by coating sample with metal or carbon or lowering the operating kV
-
7/27/2019 Lecture - 9 - Microscopy
8/33
-
7/27/2019 Lecture - 9 - Microscopy
9/33
How does SEM work?
What happens when the Electron Beam hits the sample?
SEM Setup
Electron/Specimen InteractionsWhen the electron beam strikes the sample, both photon and electron signals are emitted
-
7/27/2019 Lecture - 9 - Microscopy
10/33
Secondary Electron
Generation
-SEM-SE
-sample electrons ejected by the
primary beam [green line]
-low energy
-surface detail & topography
SEM Imaging Modes
SE
Backscattered Electron
Generation
-SEM-BSE
-primarybeam electrons
-high energy
-composition and topography
[specimen atomic number]
BSE
-
7/27/2019 Lecture - 9 - Microscopy
11/33
-
7/27/2019 Lecture - 9 - Microscopy
12/33
Identification of Fracture Mode
SEM micrographs of fracturedsurface of two BaTiO3 samples
D.Sarkar, B.Basu, M.J.Chu and S.J.Cho, R-Curve Behavior of Ti3SiC2, Ceramics International, 33 [5] 789 -793, 2007.
-
7/27/2019 Lecture - 9 - Microscopy
13/33
EDX EDX is an analytical technique used for the elemental analysis or chemical
characterization of a specimen.
it relies on the investigation of a sample through interactions betweenelectromagnetic radiation and matter, analyzing X-rays emitted by the matter inthis particular case.
Each element has a unique atomic structure allowing x-rays to be uniquely
distinguished from each other.
A graph is plotted (x-ray energy vs. count rate). The peaks correspond to characteristic elemental emissions.
-
7/27/2019 Lecture - 9 - Microscopy
14/33
Nanoscale MaterialsNanowires and Nanotubes
Lateral dimension: 1 100 nm Nanowires and nanotubes exhibit novel
physical, electronic andoptical properties due to Two dimensional quantum
confinement
Structural one dimensionality High surface to volume ratio
Potential application in wide range ofnanodevices and systems Nanoscale sensors and actuators
Photovoltaic devices solar cells Transistors, diodes and LASERs
Nanowire Solar Cell: The nanowirescreate a surface that is able to absorb
more sunlight than a flat surface
-
7/27/2019 Lecture - 9 - Microscopy
15/33
-
7/27/2019 Lecture - 9 - Microscopy
16/33
Nanofluids
Aluminum Oxide
Particles in Water
Nanofluids are engineered colloids = base fluid (water, organic liquid) + nanoparticles
Nanoparticle size: 1-100 nm
Nanoparticle materials: Al2O3, ZrO2, SiO2, CuO, Fe3O4, Au, Cu, C (diamond, PyC, fullerene) etc.
Previous studies suggest significant enhancement of:
- Thermal conductivity (+40%)
- Single-phase convective
heat transfer (+40%)
- Critical Heat Flux (+100%)
-
7/27/2019 Lecture - 9 - Microscopy
17/33
B i P i i l f TEM
-
7/27/2019 Lecture - 9 - Microscopy
18/33
Basic Principle of TEM
Electron source, electromagnetic lens system, sample holder, and imaging system
Electron Source
The electron source consists of a cathode and an anode. The cathode is
a tungsten filament which emits electrons when being heated. A negative
cap confines the electrons into a loosely focused beam. The beam is
then accelerated towards the specimen by the positive anode. Electrons
at the rim of the beam will fall onto the anode while the others at the
center will pass through the small hole of the anode. The electron source
works like a cathode ray tube.
Electromagnetic lens system
After leaving the electron source, the electron beam is tightly focused using electromagnetic lens and metal
apertures. The system only allows electrons within a small energy range to pass through, so the electrons in the
electron beam will have a well-defined energy.
1.Magnetic Lens: Circularelectro-magnets capable of generating a precise circular magnetic field. The field acts
like an optical lens to focus the electrons.
2.Aperture: A thin disk with a small (2-100 micrometers) circular through-hole. It is used to restrict the electronbeam and filter out unwanted electrons before hitting the specimen.
Sample Holder
-
7/27/2019 Lecture - 9 - Microscopy
19/33
Sample Holder
The sample holder is a platform equipped with a mechanical arm for holding the specimen and controlling its
position.
Imaging system
The imaging system consists of another electromagnetic lens system and a screen. The electromagnetic lenssystem contains two lens systems, one for refocusing the electrons after they pass through the specimen, and the
other for enlarging the image and projecting it onto the screen. The screen has a phosphorescent plate which
glows when being hit by electrons. Image forms in a way similar to photography.
Working principle
TEM works like a slide projector. A projector shines a beam of light which transmits through the slide. The patterns
painted on the slide only allow certain parts of the light beam to pass through. Thus the transmitted beam
replicates the patterns on the slide, forming an enlarged image of the slide when falling on the screen.
TEMs work the same way except that they shine a beam of electrons (like the light in a slide projector) through the
specimen (like the slide). However, in TEM, the transmission of electron beam is highly dependent on the
properties of material being examined. Such properties include density, composition, etc. For example, porous
material will allow more electrons to pass through while dense material will allow less. As a result, a specimen with
a non-uniform density can be examined by this technique. Whatever part is transmitted is projected onto aphosphor screen for the user to see.
The following movie will help you to understand more about the operation of a TEM:
Imaging Mode of TEM: Bright Field Mode
Dark Field Mode
Diffraction Mode
-
7/27/2019 Lecture - 9 - Microscopy
20/33
In a TEM, the specimen you want tolook at must be of such a low
density that it allows electrons to
travel through the tissue
There are different ways to prepare
your material for that purpose. You
can cut very thin slices of your
specimen from a piece of tissue
either by fixing it in plastic or
working with it as frozen material.
Another way to prepare your
specimen is to isolate it and study a
solution of for example viruses or
molecules in the TEM.
Preparation of Specimen
-
7/27/2019 Lecture - 9 - Microscopy
21/33
TEM Sample Prep for Materials
-
7/27/2019 Lecture - 9 - Microscopy
22/33
How to Measure SAED (Selected Area Electron Diffraction)
Au (111) ring [2.35 d-spacing]
With 200KV, (111) ring should distance Rfrom the transmitted beam
Rdhkl=lL
R is measuredd inter atomic distancel is the electron wavelengthL is the camera length(lL is the camera constant)
Think of TEM as a diffraction camera
Transmitted Beam
Diffracted Beam
L
R
-
7/27/2019 Lecture - 9 - Microscopy
23/33
Electron DiffractionFour conditions in Back Focal Plane (BFP) of the
objective lens: No sample No reflections (only transmitted beam)
Amorphous Transmitted beam + random scattering
Polycrystal Transmitted beam + rings
Single crystal Transmitted beam + spots
Metal particles Polymer mix
Electron Diffraction
Bright field TEM of FeNbO4 and (1 wt %) Pt
impregnated FeNbO4. (Inset) SAED pattern
of FeNbO4
Particle Size : 30-60nm
-
7/27/2019 Lecture - 9 - Microscopy
24/33
TiO2 Nanoscale Rods
TEM images of Cu on ZnO (a model methanol catalyst)
-
7/27/2019 Lecture - 9 - Microscopy
25/33
(b) bright field image
TEM Characterization of TiO2 Nanoparticles
The structure of all as-grownsamples is anatase.
The particle sizes from TEM rangebetween 15 and 25 nm.
(a) dark field image (c) diffraction patterns
(d) lattice image
20nm
(d) Lattice Image
Cr stallite Si e is Different than Particle Si e
-
7/27/2019 Lecture - 9 - Microscopy
26/33
Crystallite Size is Different than Particle Size
A particle may be made up of several different crystallites
Crystallite size often matches grain size, but there are exceptions
Grain boundaries are interfaces where
crystals of different orientations meet.
A grain boundary is a single-phase
interface, with crystals on each side of
the boundary being identical except in
orientation. The term "crystallite
boundary" is sometimes, though
rarely, used. Grain boundary areas
contain those atoms that have been
perturbed from their original lattice
sites, dislocations, and impurities that
have migrated to the lower energygrain boundary.
Crystallite is the average size of the particle whereas the
particle size denotes the individual size of the particle
Crystallite size is usually measured
from X-ray diffraction patterns and
grain size by other experimental
techniques like transmission electron
microscopy.
Most materials are polycrystalline;
they are made of a large number ofsingle crystals crystallites.
S i P b Mi
-
7/27/2019 Lecture - 9 - Microscopy
27/33
Scanning Probe Microscopy
atomic force microscope
sharp probe moves over surface of specimen atconstant distance
up and down movement of probe as it maintainsconstant distance is detected and used to create image
scanning tunneling microscope
steady current (tunneling current) maintained betweenmicroscope probe and specimen
up and down movement of probe as it maintains currentis detected and used to create image of surface ofspecimen
-
7/27/2019 Lecture - 9 - Microscopy
28/33
1 2
3 4
-
7/27/2019 Lecture - 9 - Microscopy
29/33
5 6
7 8
-
7/27/2019 Lecture - 9 - Microscopy
30/33
Transmission Electron Microscopy Study of
TiO2 Phase Transformation
As-deposited 700 oC 800 oC
TEM diffraction patterns
for annealed and as-deposited 12-nm sample.
-
7/27/2019 Lecture - 9 - Microscopy
31/33
Dark Field Imaging
If the transmitted beamis excluded from theimage formation process
off-axis imaging
tilted beam imaging
-
7/27/2019 Lecture - 9 - Microscopy
32/33
Bright Field Imaging
If the main portion of the near-forwardscattered beam is used to form the image
transmitted beam 000 beam
zero-order beam
-
7/27/2019 Lecture - 9 - Microscopy
33/33