Lecture 4 Beam restricting Devices, their Use and Maintenance Topic 1Importance of beam restricting...

23
Lecture 4 Beam restricting Devices, their Use and Maintenance Topic 1 Importance of beam restricting devices Topic 2 Types of beam restricting devices Topic 3 Quality assurance tests of beam restricting devices Topic 4 Compression of area of interest

Transcript of Lecture 4 Beam restricting Devices, their Use and Maintenance Topic 1Importance of beam restricting...

Page 1: Lecture 4 Beam restricting Devices, their Use and Maintenance Topic 1Importance of beam restricting devices Topic 2Types of beam restricting devices Topic.

Lecture 4 Beam restricting Devices, their Use and Maintenance

• Topic 1 Importance of beam restricting devices

• Topic 2 Types of beam restricting devices

• Topic 3 Quality assurance tests of beam restricting devices

• Topic 4 Compression of area of interest

Page 2: Lecture 4 Beam restricting Devices, their Use and Maintenance Topic 1Importance of beam restricting devices Topic 2Types of beam restricting devices Topic.

Topic 1 Importance of beam- restricting devices

Page 3: Lecture 4 Beam restricting Devices, their Use and Maintenance Topic 1Importance of beam restricting devices Topic 2Types of beam restricting devices Topic.

Topic 1 Importance of beam restricting devices

• These devices restrict the primary x-ray beam to the area of interest.

• The larger the area covered by primary x-ray beam, the greater the scattered radiation produced.

• Scattered radiation must be minimized. They increase radiation dose and add to film darkening thus reducing visualization of details.

Page 4: Lecture 4 Beam restricting Devices, their Use and Maintenance Topic 1Importance of beam restricting devices Topic 2Types of beam restricting devices Topic.

Topic 1 Importance of beam- restricting devices

Beam was not restricted to the chest – the area of interest. Entire abdomen and part of face of baby was exposed unnecessarily.

Page 5: Lecture 4 Beam restricting Devices, their Use and Maintenance Topic 1Importance of beam restricting devices Topic 2Types of beam restricting devices Topic.

Topic 1 Importance of beam- restricting devices

Unrestricted beam-considerable scatter

Restricted beam – less scattered radiation

Page 6: Lecture 4 Beam restricting Devices, their Use and Maintenance Topic 1Importance of beam restricting devices Topic 2Types of beam restricting devices Topic.

Topic 2 Types of beam-restrictingdevicesa. Attached to x-ray tube housing Aperture diaphragms - lead sheets with circular, square, or rectangular openings Cones – detachable metal tubes of different lengths, shapes, and sizes of opening

Variable aperture collimators – adjustable lead plates or shutters , contain "cross hairs", a light source, and a mirror to project the light

Page 7: Lecture 4 Beam restricting Devices, their Use and Maintenance Topic 1Importance of beam restricting devices Topic 2Types of beam restricting devices Topic.

Beam- restricting devices

Removablemetal cone

Variable aperture collimator

Page 8: Lecture 4 Beam restricting Devices, their Use and Maintenance Topic 1Importance of beam restricting devices Topic 2Types of beam restricting devices Topic.

Variable aperture, multiple shutter collimator

Diaphragm closer to window reduces off-focus x rays.

Smaller sized field reduces size of primary beam.

Multiple shutters

Multiple shutters

Page 9: Lecture 4 Beam restricting Devices, their Use and Maintenance Topic 1Importance of beam restricting devices Topic 2Types of beam restricting devices Topic.

Topic 2 Types of beam - restricting devicesb. Not attached to x ray tube housing

Lead blockers – used to divide a

cassette when more than one

exposure is made on a single film;

reduces unnecessary exposure of

film thus improving image quality

Page 10: Lecture 4 Beam restricting Devices, their Use and Maintenance Topic 1Importance of beam restricting devices Topic 2Types of beam restricting devices Topic.

Use of lead blockers• Place blocker on couch for lateral projections of thoracic,

lumbar, sacrum or

coccyx spine.• Position blocker next to patient's back to absorb radiation

that will not pass through spine.• Defective lead rubber aprons could be cut

into a range of sizes to serve as blockers.

Page 11: Lecture 4 Beam restricting Devices, their Use and Maintenance Topic 1Importance of beam restricting devices Topic 2Types of beam restricting devices Topic.

Topic 3 Quality assurance tests of beam limiting devices• Collimation or limiting the x-ray beam to the area of

interest requires use of beam restricting devices. It is good practice to collimate such that the resulting image has collimated edges on all four sides of film

Page 12: Lecture 4 Beam restricting Devices, their Use and Maintenance Topic 1Importance of beam restricting devices Topic 2Types of beam restricting devices Topic.

Topic 3 Quality assurance tests of beam- restricting devices

• Bring tube to rest on table top to check that all four edges of square light beam diaphragm touch the table top when using vertical beam. Do several times during the day.• Do the same test with cone inserted in front of the x ray tube window.

Caution: Cones sometimes become bent when dropped or bumped.

Page 13: Lecture 4 Beam restricting Devices, their Use and Maintenance Topic 1Importance of beam restricting devices Topic 2Types of beam restricting devices Topic.

Topic 3 Quality Assurance tests of beam restricting devices• Do collimator-beam alignment check at least every six

months.

• Do check of alignment of centre of x ray

beam at least every six months.

Page 14: Lecture 4 Beam restricting Devices, their Use and Maintenance Topic 1Importance of beam restricting devices Topic 2Types of beam restricting devices Topic.

Collimator-beam alignment test

• Using 100 cm focus-film distance (FFD), set collimator at 20 cm x 20 cm field size on surface of loaded cassette.

• Place metal coins or paper clips on top of cassette as shown in next slide.

• Expose the film at 60 kVp and 4 – 8 mAs.• Process the film.• Check whether the distance between outer edges of image and of paper clips/coins is

within +/- 2% of FFD.

Page 15: Lecture 4 Beam restricting Devices, their Use and Maintenance Topic 1Importance of beam restricting devices Topic 2Types of beam restricting devices Topic.

Line diagram showing metal coins and field size

Coin to identify right upper corner.

Edge of light.

Cassette.

Page 16: Lecture 4 Beam restricting Devices, their Use and Maintenance Topic 1Importance of beam restricting devices Topic 2Types of beam restricting devices Topic.

Collimator-beam alignment test result

Page 17: Lecture 4 Beam restricting Devices, their Use and Maintenance Topic 1Importance of beam restricting devices Topic 2Types of beam restricting devices Topic.

Check of alignment of centre of x-ray beam• Place unexposed loaded cassette in centre of bucky tray

and centre tube to cassette.• Move tube to 100 cm FFD to bucky tray. Reduce

longitudinal collimators to a thin slit (e.g. 0.5 cm). Close lateral collimators. Expose using 60 kVp and 4 – 8 mAs.

• Do not remove cassette.

Page 18: Lecture 4 Beam restricting Devices, their Use and Maintenance Topic 1Importance of beam restricting devices Topic 2Types of beam restricting devices Topic.

Check of alignment of centre of x ray beam (cont)

• Close longitudinal collimators and open lateral collimators to a thin slit. Expose film again. Process film.

• Bend film in half and check that exposed "cross" is in film centre.

• Acceptable deviation: 1 cm on either side of centre.

• Can also be done for non-bucky radiography. Just move tube so 100 cm FFD is to cassette top.

Page 19: Lecture 4 Beam restricting Devices, their Use and Maintenance Topic 1Importance of beam restricting devices Topic 2Types of beam restricting devices Topic.

Centre of beam alignment test result

Page 20: Lecture 4 Beam restricting Devices, their Use and Maintenance Topic 1Importance of beam restricting devices Topic 2Types of beam restricting devices Topic.

Topic 3 Summary

Beam restricting devices• may be attached or not attached to tube housing • require periodic checks• are necessary for radiation protection and contribute to

good image quality • minimize primary radiation• minimize scatter radiation

Page 21: Lecture 4 Beam restricting Devices, their Use and Maintenance Topic 1Importance of beam restricting devices Topic 2Types of beam restricting devices Topic.

Topic 4 Compression of area of interest

• Reducing patient thickness by compression minimizes production of scattered radiation, thus reducing radiation dose.

• Device used : compression band.

Page 22: Lecture 4 Beam restricting Devices, their Use and Maintenance Topic 1Importance of beam restricting devices Topic 2Types of beam restricting devices Topic.

Topic 4 Compression of area of interest

Compression band over abdomen reduces its volume. This reduces scatter radiation.

Page 23: Lecture 4 Beam restricting Devices, their Use and Maintenance Topic 1Importance of beam restricting devices Topic 2Types of beam restricting devices Topic.

Lecture 4 Summary• Beam-restricting devices are important in

radiation protection.• These devices should be checked regularly.• Regular performance of quality assurance tests is in

accordance with ALARA.