Lecture 09 Ionizare

34
Types of X-ray Detectors C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 1 / 10

description

x-ray

Transcript of Lecture 09 Ionizare

Page 1: Lecture 09 Ionizare

Types of X-ray Detectors

Gas detectors

• Ionization chamber

• Proportional counter

• Geiger-Muller tube

Scintillation countersSolid state detectors

• Intrinsic semiconductor

• P-I-N junction

• Silicon drift

Charge coupled device detectors

• Indirect

• Direct coupled

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 1 / 10

Page 2: Lecture 09 Ionizare

Types of X-ray Detectors

Gas detectors

• Ionization chamber

• Proportional counter

• Geiger-Muller tube

Scintillation countersSolid state detectors

• Intrinsic semiconductor

• P-I-N junction

• Silicon drift

Charge coupled device detectors

• Indirect

• Direct coupled

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 1 / 10

Page 3: Lecture 09 Ionizare

Types of X-ray Detectors

Gas detectors

• Ionization chamber

• Proportional counter

• Geiger-Muller tube

Scintillation counters

Solid state detectors

• Intrinsic semiconductor

• P-I-N junction

• Silicon drift

Charge coupled device detectors

• Indirect

• Direct coupled

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 1 / 10

Page 4: Lecture 09 Ionizare

Types of X-ray Detectors

Gas detectors

• Ionization chamber

• Proportional counter

• Geiger-Muller tube

Scintillation countersSolid state detectors

• Intrinsic semiconductor

• P-I-N junction

• Silicon drift

Charge coupled device detectors

• Indirect

• Direct coupled

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 1 / 10

Page 5: Lecture 09 Ionizare

Types of X-ray Detectors

Gas detectors

• Ionization chamber

• Proportional counter

• Geiger-Muller tube

Scintillation countersSolid state detectors

• Intrinsic semiconductor

• P-I-N junction

• Silicon drift

Charge coupled device detectors

• Indirect

• Direct coupled

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 1 / 10

Page 6: Lecture 09 Ionizare

Types of X-ray Detectors

Gas detectors

• Ionization chamber

• Proportional counter

• Geiger-Muller tube

Scintillation countersSolid state detectors

• Intrinsic semiconductor

• P-I-N junction

• Silicon drift

Charge coupled device detectors

• Indirect

• Direct coupled

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 1 / 10

Page 7: Lecture 09 Ionizare

Types of X-ray Detectors

Gas detectors

• Ionization chamber

• Proportional counter

• Geiger-Muller tube

Scintillation countersSolid state detectors

• Intrinsic semiconductor

• P-I-N junction

• Silicon drift

Charge coupled device detectors

• Indirect

• Direct coupled

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 1 / 10

Page 8: Lecture 09 Ionizare

Types of X-ray Detectors

Gas detectors

• Ionization chamber

• Proportional counter

• Geiger-Muller tube

Scintillation countersSolid state detectors

• Intrinsic semiconductor

• P-I-N junction

• Silicon drift

Charge coupled device detectors

• Indirect

• Direct coupled

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 1 / 10

Page 9: Lecture 09 Ionizare

Types of X-ray Detectors

Gas detectors

• Ionization chamber

• Proportional counter

• Geiger-Muller tube

Scintillation countersSolid state detectors

• Intrinsic semiconductor

• P-I-N junction

• Silicon drift

Charge coupled device detectors

• Indirect

• Direct coupled

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 1 / 10

Page 10: Lecture 09 Ionizare

Types of X-ray Detectors

Gas detectors

• Ionization chamber

• Proportional counter

• Geiger-Muller tube

Scintillation countersSolid state detectors

• Intrinsic semiconductor

• P-I-N junction

• Silicon drift

Charge coupled device detectors

• Indirect

• Direct coupled

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 1 / 10

Page 11: Lecture 09 Ionizare

Types of X-ray Detectors

Gas detectors

• Ionization chamber

• Proportional counter

• Geiger-Muller tube

Scintillation countersSolid state detectors

• Intrinsic semiconductor

• P-I-N junction

• Silicon drift

Charge coupled device detectors

• Indirect

• Direct coupled

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 1 / 10

Page 12: Lecture 09 Ionizare

Types of X-ray Detectors

Gas detectors

• Ionization chamber

• Proportional counter

• Geiger-Muller tube

Scintillation countersSolid state detectors

• Intrinsic semiconductor

• P-I-N junction

• Silicon drift

Charge coupled device detectors

• Indirect

• Direct coupled

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 1 / 10

Page 13: Lecture 09 Ionizare

Types of X-ray Detectors

Gas detectors

• Ionization chamber

• Proportional counter

• Geiger-Muller tube

Scintillation countersSolid state detectors

• Intrinsic semiconductor

• P-I-N junction

• Silicon drift

Charge coupled device detectors

• Indirect

• Direct coupled

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 1 / 10

Page 14: Lecture 09 Ionizare

Gas Detector Curve

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 2 / 10

Page 15: Lecture 09 Ionizare

Ionization Chamber

Useful for beam monitoring, flux measurement, fluorescence measurement,spectroscopy.

• Closed (or sealed) chamber oflength L with gas mixtureµ =

∑ρiµi

• High voltage applied to plates

• Calculate fraction of beamabsorbed I/Io = e−µL

• When x-ray interacts with gasatom, photoionized electronsswept rapidly to positiveelectrode and current (nanoAmperes) is measured.

• Count rates up to 1011

photons/s/cm3

• 22-41 eV per electron-holepair (depending on the gas)makes this useful forquantitative measurements.

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 3 / 10

Page 16: Lecture 09 Ionizare

Ionization Chamber

Useful for beam monitoring, flux measurement, fluorescence measurement,spectroscopy.

• Closed (or sealed) chamber oflength L with gas mixtureµ =

∑ρiµi

• High voltage applied to plates

• Calculate fraction of beamabsorbed I/Io = e−µL

• When x-ray interacts with gasatom, photoionized electronsswept rapidly to positiveelectrode and current (nanoAmperes) is measured.

• Count rates up to 1011

photons/s/cm3

• 22-41 eV per electron-holepair (depending on the gas)makes this useful forquantitative measurements.

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 3 / 10

Page 17: Lecture 09 Ionizare

Ionization Chamber

Useful for beam monitoring, flux measurement, fluorescence measurement,spectroscopy.

• Closed (or sealed) chamber oflength L with gas mixtureµ =

∑ρiµi

• High voltage applied to plates

• Calculate fraction of beamabsorbed I/Io = e−µL

• When x-ray interacts with gasatom, photoionized electronsswept rapidly to positiveelectrode and current (nanoAmperes) is measured.

• Count rates up to 1011

photons/s/cm3

• 22-41 eV per electron-holepair (depending on the gas)makes this useful forquantitative measurements.

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 3 / 10

Page 18: Lecture 09 Ionizare

Ionization Chamber

Useful for beam monitoring, flux measurement, fluorescence measurement,spectroscopy.

• Closed (or sealed) chamber oflength L with gas mixtureµ =

∑ρiµi

• High voltage applied to plates

• Calculate fraction of beamabsorbed I/Io = e−µL

• When x-ray interacts with gasatom, photoionized electronsswept rapidly to positiveelectrode and current (nanoAmperes) is measured.

• Count rates up to 1011

photons/s/cm3

• 22-41 eV per electron-holepair (depending on the gas)makes this useful forquantitative measurements.

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 3 / 10

Page 19: Lecture 09 Ionizare

Ionization Chamber

Useful for beam monitoring, flux measurement, fluorescence measurement,spectroscopy.

• Closed (or sealed) chamber oflength L with gas mixtureµ =

∑ρiµi

• High voltage applied to plates

• Calculate fraction of beamabsorbed I/Io = e−µL

• When x-ray interacts with gasatom, photoionized electronsswept rapidly to positiveelectrode and current (nanoAmperes) is measured.

• Count rates up to 1011

photons/s/cm3

• 22-41 eV per electron-holepair (depending on the gas)makes this useful forquantitative measurements.

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 3 / 10

Page 20: Lecture 09 Ionizare

Ionization Chamber

Useful for beam monitoring, flux measurement, fluorescence measurement,spectroscopy.

• Closed (or sealed) chamber oflength L with gas mixtureµ =

∑ρiµi

• High voltage applied to plates

• Calculate fraction of beamabsorbed I/Io = e−µL

• When x-ray interacts with gasatom, photoionized electronsswept rapidly to positiveelectrode and current (nanoAmperes) is measured.

• Count rates up to 1011

photons/s/cm3

• 22-41 eV per electron-holepair (depending on the gas)makes this useful forquantitative measurements.

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 3 / 10

Page 21: Lecture 09 Ionizare

Ionization Chamber

Useful for beam monitoring, flux measurement, fluorescence measurement,spectroscopy.

• Closed (or sealed) chamber oflength L with gas mixtureµ =

∑ρiµi

• High voltage applied to plates

• Calculate fraction of beamabsorbed I/Io = e−µL

• When x-ray interacts with gasatom, photoionized electronsswept rapidly to positiveelectrode and current (nanoAmperes) is measured.

• Count rates up to 1011

photons/s/cm3

• 22-41 eV per electron-holepair (depending on the gas)makes this useful forquantitative measurements.

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 3 / 10

Page 22: Lecture 09 Ionizare

Scintillation Counter

Useful for photon counting experiments

• NaI(Tl), Yttrium Aluminum Perovskite (YAP) or plastic which,absorb x-rays and fluoresce in the visible spectrum.

• Light strikes a thin photocathode which emits electrons into thevacuum portion of a photomultiplier tube.

• Photoelectrons are accelerated in steps, striking dynodes andbecoming amplified.

• Output voltage pulse is proportional to initial x-ray energy.

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 4 / 10

Page 23: Lecture 09 Ionizare

Scintillation Counter

Useful for photon counting experiments

• NaI(Tl), Yttrium Aluminum Perovskite (YAP) or plastic which,absorb x-rays and fluoresce in the visible spectrum.

• Light strikes a thin photocathode which emits electrons into thevacuum portion of a photomultiplier tube.

• Photoelectrons are accelerated in steps, striking dynodes andbecoming amplified.

• Output voltage pulse is proportional to initial x-ray energy.

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 4 / 10

Page 24: Lecture 09 Ionizare

Scintillation Counter

Useful for photon counting experiments

• NaI(Tl), Yttrium Aluminum Perovskite (YAP) or plastic which,absorb x-rays and fluoresce in the visible spectrum.

• Light strikes a thin photocathode which emits electrons into thevacuum portion of a photomultiplier tube.

• Photoelectrons are accelerated in steps, striking dynodes andbecoming amplified.

• Output voltage pulse is proportional to initial x-ray energy.

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 4 / 10

Page 25: Lecture 09 Ionizare

Scintillation Counter

Useful for photon counting experiments

• NaI(Tl), Yttrium Aluminum Perovskite (YAP) or plastic which,absorb x-rays and fluoresce in the visible spectrum.

• Light strikes a thin photocathode which emits electrons into thevacuum portion of a photomultiplier tube.

• Photoelectrons are accelerated in steps, striking dynodes andbecoming amplified.

• Output voltage pulse is proportional to initial x-ray energy.

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 4 / 10

Page 26: Lecture 09 Ionizare

Scintillation Counter

Useful for photon counting experiments

• NaI(Tl), Yttrium Aluminum Perovskite (YAP) or plastic which,absorb x-rays and fluoresce in the visible spectrum.

• Light strikes a thin photocathode which emits electrons into thevacuum portion of a photomultiplier tube.

• Photoelectrons are accelerated in steps, striking dynodes andbecoming amplified.

• Output voltage pulse is proportional to initial x-ray energy.

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 4 / 10

Page 27: Lecture 09 Ionizare

Solid State Detectors

Open circuit p-n junction has a natural depletion region

When reverse biased, the depletion region growscreating a higher electric field near the junction

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 5 / 10

Page 28: Lecture 09 Ionizare

Solid State Detectors

Open circuit p-n junction has a natural depletion region

When reverse biased, the depletion region grows

creating a higher electric field near the junction

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 5 / 10

Page 29: Lecture 09 Ionizare

Solid State Detectors

Open circuit p-n junction has a natural depletion region

When reverse biased, the depletion region growscreating a higher electric field near the junction

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 5 / 10

Page 30: Lecture 09 Ionizare

Ge Detector Operation

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 6 / 10

Page 31: Lecture 09 Ionizare

Silicon Drift Detector

Same principle as intrinsic or p-i-n detector but much more compact andoperates at higher temperatures

Relatively low stopping power is a drawback

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 7 / 10

Page 32: Lecture 09 Ionizare

CCD Detectors - Direct

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 8 / 10

Page 33: Lecture 09 Ionizare

CCD Detectors - Indirect

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 9 / 10

Page 34: Lecture 09 Ionizare

CCD Detectors - Lightpipe Taper

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 10 / 10