Chapter 11: Atomic Mass Spectrometry (Inorganic MS)
• Mass Spectrometers• ICP-MS• Spark Source MS• Glow-Discharge MS• Elemental Surface Analysis by MS• Laser Ablation ICP-MS
![Page 2: Laser Ablation ICP-MS • Glow-Discharge MS - UMass Lowellfaculty.uml.edu/david_ryan/84.314/Instrumental Lecture 11.pdf · • Spark Source MS • Glow-Discharge MS ... • Laser](https://reader030.fdocuments.net/reader030/viewer/2022032613/5aa56d347f8b9a7c1a8d528c/html5/thumbnails/2.jpg)
Atomic Mass Spec processes• Atomization (sample intro)• Conversion to ions• Separation based on m/z ratio• Detection
In other forms of MS (GC-MS or MS of organic compounds), sample introduction does not involve making atoms, just getting molecules into the high vacuum system
![Page 3: Laser Ablation ICP-MS • Glow-Discharge MS - UMass Lowellfaculty.uml.edu/david_ryan/84.314/Instrumental Lecture 11.pdf · • Spark Source MS • Glow-Discharge MS ... • Laser](https://reader030.fdocuments.net/reader030/viewer/2022032613/5aa56d347f8b9a7c1a8d528c/html5/thumbnails/3.jpg)
Basic MSdesigncomponents
Note highvacuum
![Page 4: Laser Ablation ICP-MS • Glow-Discharge MS - UMass Lowellfaculty.uml.edu/david_ryan/84.314/Instrumental Lecture 11.pdf · • Spark Source MS • Glow-Discharge MS ... • Laser](https://reader030.fdocuments.net/reader030/viewer/2022032613/5aa56d347f8b9a7c1a8d528c/html5/thumbnails/4.jpg)
Advantages of Atomic Mass Spec over Optical Atomic Spectrometry
1) Detection limits are better, sometimes several orders of magnitude better
2) Very simple spectra3) Ability to measure isotope ratiosDisadvantages1) Equipment cost2) Instrument drift3) Isotopic interferences
![Page 5: Laser Ablation ICP-MS • Glow-Discharge MS - UMass Lowellfaculty.uml.edu/david_ryan/84.314/Instrumental Lecture 11.pdf · • Spark Source MS • Glow-Discharge MS ... • Laser](https://reader030.fdocuments.net/reader030/viewer/2022032613/5aa56d347f8b9a7c1a8d528c/html5/thumbnails/5.jpg)
Atomic Mass Spec has been around for a long time but one of the most significant advances in this field occurred recently, the development of ICP-MS
![Page 6: Laser Ablation ICP-MS • Glow-Discharge MS - UMass Lowellfaculty.uml.edu/david_ryan/84.314/Instrumental Lecture 11.pdf · • Spark Source MS • Glow-Discharge MS ... • Laser](https://reader030.fdocuments.net/reader030/viewer/2022032613/5aa56d347f8b9a7c1a8d528c/html5/thumbnails/6.jpg)
Detectors = Transducers for Mass Spec1) Electron Multipliers – transducer of choice Discrete dynode electron multiplier works like
a photomultiplier tubeions � electrons � many electrons
typical current gain 107
Cu/Be surface
![Page 7: Laser Ablation ICP-MS • Glow-Discharge MS - UMass Lowellfaculty.uml.edu/david_ryan/84.314/Instrumental Lecture 11.pdf · • Spark Source MS • Glow-Discharge MS ... • Laser](https://reader030.fdocuments.net/reader030/viewer/2022032613/5aa56d347f8b9a7c1a8d528c/html5/thumbnails/7.jpg)
Continuous dynodeelectron multiplier – glass dopedWith Pb, gains from 105 to 108
![Page 8: Laser Ablation ICP-MS • Glow-Discharge MS - UMass Lowellfaculty.uml.edu/david_ryan/84.314/Instrumental Lecture 11.pdf · • Spark Source MS • Glow-Discharge MS ... • Laser](https://reader030.fdocuments.net/reader030/viewer/2022032613/5aa56d347f8b9a7c1a8d528c/html5/thumbnails/8.jpg)
2) Faraday Cup – historically important, not extremely sensitive, slow
![Page 9: Laser Ablation ICP-MS • Glow-Discharge MS - UMass Lowellfaculty.uml.edu/david_ryan/84.314/Instrumental Lecture 11.pdf · • Spark Source MS • Glow-Discharge MS ... • Laser](https://reader030.fdocuments.net/reader030/viewer/2022032613/5aa56d347f8b9a7c1a8d528c/html5/thumbnails/9.jpg)
Separation of ions = Mass Analyzers1) Quadrupole Mass Analyzers
Importantdevice thatacts as a mass filterpassingonly ions ofcertain m/z
![Page 10: Laser Ablation ICP-MS • Glow-Discharge MS - UMass Lowellfaculty.uml.edu/david_ryan/84.314/Instrumental Lecture 11.pdf · • Spark Source MS • Glow-Discharge MS ... • Laser](https://reader030.fdocuments.net/reader030/viewer/2022032613/5aa56d347f8b9a7c1a8d528c/html5/thumbnails/10.jpg)
![Page 11: Laser Ablation ICP-MS • Glow-Discharge MS - UMass Lowellfaculty.uml.edu/david_ryan/84.314/Instrumental Lecture 11.pdf · • Spark Source MS • Glow-Discharge MS ... • Laser](https://reader030.fdocuments.net/reader030/viewer/2022032613/5aa56d347f8b9a7c1a8d528c/html5/thumbnails/11.jpg)
Separation of ions = Mass Analyzers1) Quadrupole Mass Analyzers
Importantdevice thatacts as a mass filterpassingonly ions ofcertain m/z
![Page 12: Laser Ablation ICP-MS • Glow-Discharge MS - UMass Lowellfaculty.uml.edu/david_ryan/84.314/Instrumental Lecture 11.pdf · • Spark Source MS • Glow-Discharge MS ... • Laser](https://reader030.fdocuments.net/reader030/viewer/2022032613/5aa56d347f8b9a7c1a8d528c/html5/thumbnails/12.jpg)
2) Time of Flight Mass Analyzers
Ions are accelerated into a field-free drift tube approximately 1 m long & separated based on kinetic energy ( 1 – 30 µs)
![Page 13: Laser Ablation ICP-MS • Glow-Discharge MS - UMass Lowellfaculty.uml.edu/david_ryan/84.314/Instrumental Lecture 11.pdf · • Spark Source MS • Glow-Discharge MS ... • Laser](https://reader030.fdocuments.net/reader030/viewer/2022032613/5aa56d347f8b9a7c1a8d528c/html5/thumbnails/13.jpg)
3) Double-Focusing Mass Analyzeremploys electrostatic analyzer &magnetic sector analyzer
![Page 14: Laser Ablation ICP-MS • Glow-Discharge MS - UMass Lowellfaculty.uml.edu/david_ryan/84.314/Instrumental Lecture 11.pdf · • Spark Source MS • Glow-Discharge MS ... • Laser](https://reader030.fdocuments.net/reader030/viewer/2022032613/5aa56d347f8b9a7c1a8d528c/html5/thumbnails/14.jpg)
ICP-MS showing varioussample introduction modes, ICPtorch, mass analyzer & detector
![Page 15: Laser Ablation ICP-MS • Glow-Discharge MS - UMass Lowellfaculty.uml.edu/david_ryan/84.314/Instrumental Lecture 11.pdf · • Spark Source MS • Glow-Discharge MS ... • Laser](https://reader030.fdocuments.net/reader030/viewer/2022032613/5aa56d347f8b9a7c1a8d528c/html5/thumbnails/15.jpg)
ICP-MS• Developed early 1980’s• Low detection limits (0.1 to 10 ppb)• High selectivity• Good precision (2 - 4%) & accuracy• m/z range 3 - 300 = 90% of periodic table• Resolution 1 m/z• Dynamic range 106
• Approximately 10 sec measurement time• Various sample introduction methods
![Page 16: Laser Ablation ICP-MS • Glow-Discharge MS - UMass Lowellfaculty.uml.edu/david_ryan/84.314/Instrumental Lecture 11.pdf · • Spark Source MS • Glow-Discharge MS ... • Laser](https://reader030.fdocuments.net/reader030/viewer/2022032613/5aa56d347f8b9a7c1a8d528c/html5/thumbnails/16.jpg)
ICP Mass Spectrum of rock sample producedby Laser Ablation
![Page 17: Laser Ablation ICP-MS • Glow-Discharge MS - UMass Lowellfaculty.uml.edu/david_ryan/84.314/Instrumental Lecture 11.pdf · • Spark Source MS • Glow-Discharge MS ... • Laser](https://reader030.fdocuments.net/reader030/viewer/2022032613/5aa56d347f8b9a7c1a8d528c/html5/thumbnails/17.jpg)
Comparison of ICP optical emission spectrum & ICP-MS spectrum for 100 ppmcerium (Ce)
6 Ce lines
2 Ce isotopes
![Page 18: Laser Ablation ICP-MS • Glow-Discharge MS - UMass Lowellfaculty.uml.edu/david_ryan/84.314/Instrumental Lecture 11.pdf · • Spark Source MS • Glow-Discharge MS ... • Laser](https://reader030.fdocuments.net/reader030/viewer/2022032613/5aa56d347f8b9a7c1a8d528c/html5/thumbnails/18.jpg)
Spectroscopic Interferences1) Isobaric interferences – element isotopes
with same m/z (e.g. 113In+ & 115In+ overlap with 113Cd+ & 115Sn+)
2) Polyatomic ion interferences – formed from interactions of species in plasma
3) Oxide & Hydroxide species interference –MO+ or MOH+ formed in plasma
4) Matrix effects – similar to optical atomic spectrometry
![Page 19: Laser Ablation ICP-MS • Glow-Discharge MS - UMass Lowellfaculty.uml.edu/david_ryan/84.314/Instrumental Lecture 11.pdf · • Spark Source MS • Glow-Discharge MS ... • Laser](https://reader030.fdocuments.net/reader030/viewer/2022032613/5aa56d347f8b9a7c1a8d528c/html5/thumbnails/19.jpg)
![Page 20: Laser Ablation ICP-MS • Glow-Discharge MS - UMass Lowellfaculty.uml.edu/david_ryan/84.314/Instrumental Lecture 11.pdf · • Spark Source MS • Glow-Discharge MS ... • Laser](https://reader030.fdocuments.net/reader030/viewer/2022032613/5aa56d347f8b9a7c1a8d528c/html5/thumbnails/20.jpg)
MultielementAnalysis
![Page 21: Laser Ablation ICP-MS • Glow-Discharge MS - UMass Lowellfaculty.uml.edu/david_ryan/84.314/Instrumental Lecture 11.pdf · • Spark Source MS • Glow-Discharge MS ... • Laser](https://reader030.fdocuments.net/reader030/viewer/2022032613/5aa56d347f8b9a7c1a8d528c/html5/thumbnails/21.jpg)
![Page 22: Laser Ablation ICP-MS • Glow-Discharge MS - UMass Lowellfaculty.uml.edu/david_ryan/84.314/Instrumental Lecture 11.pdf · • Spark Source MS • Glow-Discharge MS ... • Laser](https://reader030.fdocuments.net/reader030/viewer/2022032613/5aa56d347f8b9a7c1a8d528c/html5/thumbnails/22.jpg)
Good linearity
![Page 23: Laser Ablation ICP-MS • Glow-Discharge MS - UMass Lowellfaculty.uml.edu/david_ryan/84.314/Instrumental Lecture 11.pdf · • Spark Source MS • Glow-Discharge MS ... • Laser](https://reader030.fdocuments.net/reader030/viewer/2022032613/5aa56d347f8b9a7c1a8d528c/html5/thumbnails/23.jpg)
![Page 24: Laser Ablation ICP-MS • Glow-Discharge MS - UMass Lowellfaculty.uml.edu/david_ryan/84.314/Instrumental Lecture 11.pdf · • Spark Source MS • Glow-Discharge MS ... • Laser](https://reader030.fdocuments.net/reader030/viewer/2022032613/5aa56d347f8b9a7c1a8d528c/html5/thumbnails/24.jpg)
Isotope Ratio Measurements – ICP-MS makes measurement of isotope ratios for a sample become relatively easy in many cases. This is a powerful technique that can identify elements from different sources, allow use of tracers, etc.
![Page 25: Laser Ablation ICP-MS • Glow-Discharge MS - UMass Lowellfaculty.uml.edu/david_ryan/84.314/Instrumental Lecture 11.pdf · • Spark Source MS • Glow-Discharge MS ... • Laser](https://reader030.fdocuments.net/reader030/viewer/2022032613/5aa56d347f8b9a7c1a8d528c/html5/thumbnails/25.jpg)
Chapter 26: An Introductionto Chromatographic Separations
• Column Chromatography• Migration Rates
– Distribution Contstants– Retention Times– Selectivity Factor
• Zone Broadening & Column Efficiency• Optimizing Performance• Resolution
![Page 26: Laser Ablation ICP-MS • Glow-Discharge MS - UMass Lowellfaculty.uml.edu/david_ryan/84.314/Instrumental Lecture 11.pdf · • Spark Source MS • Glow-Discharge MS ... • Laser](https://reader030.fdocuments.net/reader030/viewer/2022032613/5aa56d347f8b9a7c1a8d528c/html5/thumbnails/26.jpg)