Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical...

49
Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical Physics University of Wisconsin NCCAAPM Meeting Oct 11,2013

Transcript of Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical...

Page 1: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density

Larry A. DeWerd, Ph.D., FAAPMUW ADCL & Dept. of Medical Physics

University of Wisconsin

NCCAAPM Meeting Oct 11,2013

Page 2: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density

� Larry DeWerd has a partial interest in Standard Imaging

Page 3: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density

� Talks from people who have been in the field for awhile tend to be historical or philosophical

� My attempt at philosophical and hopefully informational material.

Page 4: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density
Page 5: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density

� I don’t have enough time� Why worry – it will all work out.� I don’t have to think about it – it is late.� Didn’t the manufacturer tell me it was

okay.

Page 6: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density

� Some Manufacturers harp on the time saved, etc but their instrument is not as precise and can result in errors.

� For example, some digital sampling electrometers (depending on their sampling algorithm) may miss some signal from pulses of the accelerator. The physicist should ask.

Page 7: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density

� Technicians use numbers, Medical physicists use understanding.

� Do it right – good enough is not good enough

Page 8: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density

� Transfer of Knowledge is Important.

� ADCLs transfer latest knowledge

ÂÂÂÂ\y \ {täx áxxÇ yâÜà{xÜ à{tÇ \y \ {täx áxxÇ yâÜà{xÜ à{tÇ \y \ {täx áxxÇ yâÜà{xÜ à{tÇ \y \ {täx áxxÇ yâÜà{xÜ à{tÇ Éà{xÜá? |à |á uç áàtÇw|Çz âÑÉÇ à{x Éà{xÜá? |à |á uç áàtÇw|Çz âÑÉÇ à{x Éà{xÜá? |à |á uç áàtÇw|Çz âÑÉÇ à{x Éà{xÜá? |à |á uç áàtÇw|Çz âÑÉÇ à{x á{ÉâÄwxÜá Éy z|tÇàáÊ á{ÉâÄwxÜá Éy z|tÇàáÊ á{ÉâÄwxÜá Éy z|tÇàáÊ á{ÉâÄwxÜá Éy z|tÇàáÊ f|Ü \áttv f|Ü \áttv f|Ü \áttv f|Ü \áttv axãàÉÇ? DIJIaxãàÉÇ? DIJIaxãàÉÇ? DIJIaxãàÉÇ? DIJI

Page 9: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density
Page 10: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density

� Just doing a measurement and applying ADCL calibration does not mean that it is right You need to understand the equipment and see through the measurement. Numbers no matter how precise cannot by themselves imply anything.

� Precision and accuracy of dose measurements and reporting of the measurement details should be sufficient to allow the work to be interpreted and repeated and to allow valid comparisons to be made, both in the same laboratory and by other laboratories.

Page 11: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density

� Ion chambers require calibration for the most precise measurements

� ADCLs willing to discuss measurements and methodology

� ADCL discuss the operation of instrumentation.

Page 12: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1

-6 -4 -2 0 2 4 6

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1

-6 -4 -2 0 2 4 6

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1

-6 -4 -2 0 2 4 6

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1

-6 -4 -2 0 2 4 6

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1

-6 -4 -2 0 2 4 6

a.u.

fre

qu

en

cy o

f re

sult

conventional true

value

primary standard

measurement

clinical

measurement

Page 13: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density

� NBS petitioned AAPM to create “Regional Calibration Laboratories” in 1975- In 1983 called ADCLs.

� NBS acknowledges traceability to primary standards (Proficiency tests)

� Agreement for Proficiency tests and round robins < 0.5%

� UWADCL founded 1981 by LAD

Page 14: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density

International &National Standards

Secondary StandardsADCL

Primary Calibration

Secondary Calibration

Hospital or ClinicalStandards

Tertiary Calibration

Operational Standards

Page 15: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density

� AAPM in conjunction with NIST set up secondary laboratories for medical ionizing radiation calibrations - For therapy applications the precision is within 0.5%.

� Traceability of radiation dose is dependent on standards that are the same throughout the world at primary labs.

Page 16: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density

� All calibrations (especially for medical applications) should have a traceability train to a standard

� This provides confidence in the dose given because it was measured and can be traced back to a national standard – at NIST

� However, the SSDL network is not the equivalent to the ADCL network

Page 17: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density

� The ADCLs have proven track records of providing precise calibrations of equipment for Therapy and Diagnostic applications

� Safety and treatment delivery or imaging is improved because of ADCL calibration and research.

Page 18: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density

0.9

94

0.9

95

0.9

96

0.9

97

0.9

98

0.9

99 1

1.0

01

1.0

02

1.0

03

1.0

04

1.0

05

0

2

4

6

8

10

12

14

16

N

u

m

b

e

r

0.9

94

0.9

95

0.9

96

0.9

97

0.9

98

0.9

99 1

1.0

01

1.0

02

1.0

03

1.0

04

1.0

05

Mean of ADCLs/ADCL

Page 19: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density

� These tests have been in place over 20 years

� NIST and ADCLs agree within 0.5% for Cobalt-60 beams

� NIST and ADCLs agree less than 2.0 % for x-ray beams between 20 kVp and 250 kVp

Page 20: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density

� Maintenance of accuracy and precision is very important for Medical Applications

� Knowledge of characteristics of chambers is very important

� Medical Applications rely heavily upon precise traceable calibrations.

Page 21: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density

� Calibration of chambers should be done across the energy range of use. Bracket your point of interest.

� Generally there is a reference energy in use that is the minimum energy point used.

Page 22: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density

� Because of variation in response with energy for chambers a range of calibration is necessary.

� The calibration should be done under correct criteria – e.g. TG 61

� There is no protocol using an absorbed dose to water calibration for x-ray beams. An SSDL offers absorbed dose in water but no protocol.

Page 23: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density

� Within + 1% for Mammography range: 24 kVp to 35 kVp, 0.15 to 1.0 mm Al

� Within + 2 % or 2.5% for entire diagnostic range, up to 150 kVp

� For superficial or orthovoltage (radiobiology) 1 % to 2 % is desirable

Page 24: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density

� There can be electron contamination and full buildup is needed for calibration of low – energy (< 100kV) clinical beams

� This requires an addition to the “window” to be put over a plane-parallel chamber

� The window of the chamber (for example 2.5 mg / cm2) should be subtracted from the values in the following table for the thickness of the foil for full buildup.

Page 25: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density

Tube potential (kV) Total wall thickness (mg – cm-2)

40 3.0

50 4.0

60 5.5

70 7.3

80 9.1

90 11.2

100 13.4

Page 26: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density

� Physicist should understand uncertainties as delivered by the ADCL

� For example, for cobalt calibrations

Page 27: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density

� The expanded uncertainty (at k=2) is expected to include the conventional true value.

� The chance of the population mean (the true value) being within the measured value has a 95% certainty.

� Or we can be sure that our measurement indicates the correct value within 95%

Page 28: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density

ADCL Standard Chamber

Parameter Type A Type BCharge 0.02 0.0Timing 0.04 0.05Air Density 0 0.1Ionic Recombination 0 0Distance from the source 0.1 0.07Beam Uniformity 0 0.1Air attenuation 0.1 0Quadratic Sum 0.110 0.166

Type A and B Quadratic Sum 0.198NIST Source Calibration 0.600

uNIST-ADCL Chamber Calibration 0.632

or at k=2 level, 1.26 %

Page 29: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density

Clinic Chamber

Parameter Type A Type BCharge 0.02 0.1Timing 0.04 0.05Air Density 0 0.1Ionic Recombination 0 0.1Distance from the source 0.1 0.07Beam Uniformity 0 0.1Quadratic Sum 0.110 0.218

Type A and B Quadratic Sum 0.244ADCL Electrometer Readout 0.057ADCL Transfer Calibration 0.632

uADCL-Clinic Chamber Calibration 0.680

or at k=2 level, 1.36 %

Page 30: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density

� Standards changing Absorbed dose to water done with precision and uniformly.

� Calibration of chambers and sources essential for radiation therapy: Majority traceable through Cobalt and Cesium.

� Unfortunately, some manufacturers improvise dosimetry to market their product

Page 31: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density

� Manufacturers need to pass some hoops to market product

� However, 510k from FDA does not mean that they have a calibration regimen for their machine.

� Also some manufacturers change without letting physicists know and assume everything still applies.

Page 32: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density
Page 33: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density

� What happens when manufacturers improvise for Standards? � Errors! NIST, ADCLs, AAPM should insist on a

standard

� There is a need to have traceability delivered by the Accredited Dosimetry Calibration Laboratories

� Hospital Physicists generally have requests

� Users must insist on traceable standards from NIST through the ADCLs

Page 34: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density

� Sr-90 applicators introduced in 1950. Plane or concave. 28 year half life

� Typical 8mm diameter with a beta shield� Extremely high dose rate� Calibrated by each mfr wrt Bragg-Gray

theory� Bizarre units like “reps, beta-Roentgens”� Hospitals asked UW ADCL accuracy of

dose.

Page 35: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density

� “Wing-like” growth in conjunctiva

� Requires surgery to remove

� Surgery fails 90% of time

� One dose with Sr-90 prevents recurrence in roughly 90% of patients treated

Page 36: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density
Page 37: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density

� One Chicago hospital had an Amersham Model SIA-20 applicator calibrated at NBS: calibrations differed by 38%. Each lab stood by their calibration!

� NIST had to review their calibrations

Page 38: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density

� Sr-90 sources had only been inter-compared IN AIR at distances of 20 to 30cm

� Desired “dose rate” is IN CONTACT with Sr-90 on surface of silver matrix

� Chris Soares developed an extrapolation chamber for calibration.

Page 39: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density

� NIST traceable determination of the absorbed dose to water rate in the central 4 mm of the applicator including color enhanced contour plots and two dimensional dose profiles.

� Routine ADCL calibrations from UW ADCL using radiochromic film began in 1996.

� NIST quotes uncertainty of 7 %� Calibration disagreement (old versus

new) average -27 %.

Page 40: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density

9080

70

60

50

40

30

20

10

Example of a color enhanced contour plot of a

uniform Sr-90 ophthalmic

applicator. The dose weighted

isocenter is equidistant from the hash marks. The outer circle represents the

source physical diameter.

Page 41: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density

�2-Dimensional dose profiles for a uniform source.

0.00

0.10

0.20

0.30

0.40

0.50

-6.0 -4.0 -2.0 0.0 2.0 4.0 6.0

Horizontal Profile

Do

se

Ra

te (

Gy/s

ec)

Distance (mm)

0.00

0.10

0.20

0.30

0.40

0.50

-6.0 -4.0 -2.0 0.0 2.0 4.0 6.0

Vertical Profile

Do

se

Ra

te (

Gy/s

ec)

Distance (mm)

Page 42: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density

Example of a color enhanced contour plot of a skewed-Sr-90 ophthalmic applicator. Notice the offset or shift of the dose weighted isocenter from the physical source center, and the non-uniform dose distribution.

90

90

8070

60

50

40

30

20

10

Page 43: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density

0.00

0.10

0.20

0.30

0.40

-6.0 -4.0 -2.0 0.0 2.0 4.0 6.0

Horizontal Profile

Distance (mm)

0.00

0.10

0.20

0.30

0.40

-6.0 -4.0 -2.0 0.0 2.0 4.0 6.0

Vertical Profile

Distance (mm)

Page 44: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density

� The average difference between the old calibration and the present calibration is -27%.

� The distribution of source activity can be very unique for the individual source.

Page 45: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density

� NIST standard � Pd-103 has very short half-life (17.0 d)

so NIST traceable 109Cd source (half-life of 463.2d) was used as a reference source for 12 years, then replaced in 1997

� However, self-shielding of the source encapsulation was different between these two isotopes

� This resulted in a sudden 9% shift in calibration by letter to users in 1997

Page 46: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density
Page 47: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density

� There can be significant errors� If traceable to NIST, even if wrong,

everyone is consistent if traceable standard- at least traceable through ADCL

� NIST, ADCLs and CIRMS need to insist upon standards.

� Manufacturers try hard but they need to rely on NIST and ADCLs for standards

Page 48: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density

ÂÂÂÂg{Éáx ã{É wÉ ÇÉà g{Éáx ã{É wÉ ÇÉà g{Éáx ã{É wÉ ÇÉà g{Éáx ã{É wÉ ÇÉà ÜxÅxÅuxÜ à{x Ñtáà tÜx ÜxÅxÅuxÜ à{x Ñtáà tÜx ÜxÅxÅuxÜ à{x Ñtáà tÜx ÜxÅxÅuxÜ à{x Ñtáà tÜx vÉÇwxÅÇxw àÉ ÜxÑxtà vÉÇwxÅÇxw àÉ ÜxÑxtà vÉÇwxÅÇxw àÉ ÜxÑxtà vÉÇwxÅÇxw àÉ ÜxÑxtà |àAÊ|àAÊ|àAÊ|àAÊ

George Santayana, Harvard

Professor and poet

Page 49: Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical …chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/... · 2014-11-06 · Charge 0.02 0.0 Timing 0.04 0.05 Air Density

� Thanks are due to� Students and staff of the UW ADCL� All those who send us calibration

instruments that support the researchprogram of the UW Medical RadiationResearch Center.