Laboratory Technical Issues

Presentation to:

KWWOA

April 9, 2014

Department for Environmental ProtectionEnvironmental & Public Protection Cabinet

To Protect and Enhance Kentucky’s Environment

Page 2: Laboratory Technical Issues Presentation to: KWWOA April 9, 2014 Department for Environmental Protection Environmental & Public Protection Cabinet To Protect.

2

Topics• IDC / ODC• MDL• DOC Worksheet• Significant Figures• Rounding• Reporting• Audit Process

Page 3: Laboratory Technical Issues Presentation to: KWWOA April 9, 2014 Department for Environmental Protection Environmental & Public Protection Cabinet To Protect.

3

IDC / ODC

Required by Drinking Water and Wastewater Laboratory Certification

Page 4: Laboratory Technical Issues Presentation to: KWWOA April 9, 2014 Department for Environmental Protection Environmental & Public Protection Cabinet To Protect.

4

Drinking Water Certification Manual• IDC – Chapter IV; Section 7.2.9• MDL – Chapter IV; Section 7.2.11• RLS – Chapter IV; Section 7.2.12

Page 5: Laboratory Technical Issues Presentation to: KWWOA April 9, 2014 Department for Environmental Protection Environmental & Public Protection Cabinet To Protect.

5

Wastewater Certification Manual• Chapter III; Section 7.1

–Demonstration of Capability – the laboratory shall demonstrate their capability to report compliance sample results at or below the required reporting limit (RRL); established by either the Commonwealth of Kentucky or the EPA• IDC/ODC• MDL• RLS

Page 6: Laboratory Technical Issues Presentation to: KWWOA April 9, 2014 Department for Environmental Protection Environmental & Public Protection Cabinet To Protect.

6

Initial Demonstration of Capability (IDC / ODC)

Used to demonstrate that the laboratory and analyst are capable of performing analysis with acceptable precision, accuracy, sensitivity, and specificity pertaining to that particular method

Page 7: Laboratory Technical Issues Presentation to: KWWOA April 9, 2014 Department for Environmental Protection Environmental & Public Protection Cabinet To Protect.

7

IDC / ODC Continued

Analyze 4-mid range concentration laboratory fortified blanks (LFB) on the same or different day.

Acceptance Criteria:• Each of the 4 replicates within 80 – 120% of

mean;• %RSD < 15%

Page 8: Laboratory Technical Issues Presentation to: KWWOA April 9, 2014 Department for Environmental Protection Environmental & Public Protection Cabinet To Protect.

8

Method Detection Limit (MDL)

Defined as the minimum concentration of a substance that can be measured and reported with 99% confidence that the target analyte concentration is greater than zero.

Page 9: Laboratory Technical Issues Presentation to: KWWOA April 9, 2014 Department for Environmental Protection Environmental & Public Protection Cabinet To Protect.

9

MDL Continued• Analyze 7-low level replicates over 2-3 non-

consecutive days;• Calculate the MDL using procedure specified

in 40 CFR 136 Appendix B;

Acceptance Criteria:• Spike level must be with 10 times MDL• MDL < MRL < RRL (Action Limit)

Page 10: Laboratory Technical Issues Presentation to: KWWOA April 9, 2014 Department for Environmental Protection Environmental & Public Protection Cabinet To Protect.

10

Reporting Limit Standard (RLS)

Demonstrate laboratory’s capability to report down to the minimum reporting level (MRL) with known accuracy (70-130% Recovery)

Page 11: Laboratory Technical Issues Presentation to: KWWOA April 9, 2014 Department for Environmental Protection Environmental & Public Protection Cabinet To Protect.

11

RLS Continued• RLS is typical a low calibration standard; or• Stand-alone verification standard

Acceptance Criteria:• RLS Standard < MRL• Calculated %Recovery = 70 - 130%

Page 12: Laboratory Technical Issues Presentation to: KWWOA April 9, 2014 Department for Environmental Protection Environmental & Public Protection Cabinet To Protect.

Detection Limit Terms

• MDL = Method Detection Limit• MRL = Minimum Reporting Limit• RRL = Required Reporting Limit

• RRL = Action Limit = Permit Limit

12

Page 13: Laboratory Technical Issues Presentation to: KWWOA April 9, 2014 Department for Environmental Protection Environmental & Public Protection Cabinet To Protect.

Calibration Curve and RRL - Ideal

13

MDL

MRL

High Std

Concentration

Inst

rum

ent R

espo

nse

Range of Quantita

tion

Est. (J)

RRLCalibration Standards

Page 14: Laboratory Technical Issues Presentation to: KWWOA April 9, 2014 Department for Environmental Protection Environmental & Public Protection Cabinet To Protect.

Calibration Curve and RRL - Issue

14

MDL

MRL

High Std

Concentration

Inst

rum

ent R

espo

nse

Range of Quantita

tion

Est. (J)

RRL

Calibration Standards

Page 15: Laboratory Technical Issues Presentation to: KWWOA April 9, 2014 Department for Environmental Protection Environmental & Public Protection Cabinet To Protect.

15

KY DOC Calculator

DOW provides a DOC Calculator for use by drinking water and wastewater laboratories:

http://water.ky.gov/permitting/Pages/labcert.aspx

Calculator

Page 16: Laboratory Technical Issues Presentation to: KWWOA April 9, 2014 Department for Environmental Protection Environmental & Public Protection Cabinet To Protect.

Significant Figures• Significant figures are critical when reporting scientific data.

1) ALL non-zero numbers (1,2,3,4,5,6,7,8,9) are ALWAYS significant. • The number 123.678 has 6 significant digits.

2) ALL zeroes between non-zero numbers are ALWAYS significant. • The number 103 has 3 significant digits.

3) ALL zeroes which are SIMULTANEOUSLY to the right of the decimal point AND at the end of the number are ALWAYS significant. • The number 111.0 has 4 significant digits.

4) ALL zeroes which are to the left of a written decimal point and are in a number >= 10 are ALWAYS significant.• The number 100.4 has 4 significant digits.

5) Trailing zeroes that do nothing but set a decimal point are not significant• The number 100 has 1 significant digits.

16

Page 17: Laboratory Technical Issues Presentation to: KWWOA April 9, 2014 Department for Environmental Protection Environmental & Public Protection Cabinet To Protect.

Rounding

• Rounding is the last step in calculating results. Only round the final result, not at each step of the calculation. Then follow the following rules:

17

Page 18: Laboratory Technical Issues Presentation to: KWWOA April 9, 2014 Department for Environmental Protection Environmental & Public Protection Cabinet To Protect.

Rounding• Rule 1. Select the digit you want to round and look to the right side of it. If the digit

is 0,1,2,3 or 4, do not change the rounding digit. When rounding whole numbers, all digits that are on the right-hand side of the requested rounding digit will become 0. When rounding numbers involving decimals, all digits to the right of the rounded number are dropped.

18

Page 19: Laboratory Technical Issues Presentation to: KWWOA April 9, 2014 Department for Environmental Protection Environmental & Public Protection Cabinet To Protect.

RoundingRule 2. Select the digit you want to round and look to the right of it. If the digit is 6,7,

8 or 9, your rounding digit rounds up by one number. When rounding whole numbers, all digits that are on the right-hand side of the requested rounding digit will become 0. When rounding numbers involving decimals, all digits to the right of the rounded number are dropped.

19

Page 20: Laboratory Technical Issues Presentation to: KWWOA April 9, 2014 Department for Environmental Protection Environmental & Public Protection Cabinet To Protect.

RoundingRule 3. Select the digit you want to round and look to the right of it. If the digit is 5,

your rounding digit rounds up to the even number. When rounding whole numbers, all digits that are on the right-hand side of the requested rounding digit will become 0. When rounding numbers involving decimals, all digits to the right of the rounded number are dropped.

20

Page 21: Laboratory Technical Issues Presentation to: KWWOA April 9, 2014 Department for Environmental Protection Environmental & Public Protection Cabinet To Protect.

Reporting• Drinking Water

–Hard copy• Must use DOW form

– If Laboratory creates its own forms they must be identical to the DOW form

–Electronic• Must use DOW format (CSV file)

• Wastewater–New DMR Manual is currently being prepared

21

Page 22: Laboratory Technical Issues Presentation to: KWWOA April 9, 2014 Department for Environmental Protection Environmental & Public Protection Cabinet To Protect.

Audit Process• Audit announcement• Day of the audit• Exit• Audit findings• Laboratory response to the findings• DOW accepts/rejects responses• Audit completion certificate issued